Introduction

Hemoglobin is a hemeprotein with two linked pairs of globin chains. Each chain is bound to a heme residue in its center. Hemoglobin molecules are in red blood cells (RBCs) and regulate O2 and CO2 concentrations in tissues. Structural differences in the globin chains account for the different hemoglobin types in circulating blood. In normal adults, hemoglobins include HbA (α2β2), HbA22δ2), and HbF (α2γ2), accounting for 95–97.5%, 1.5–3.5%, and up to 2% of total hemoglobin, respectively (Iyer et al. 2015; Greene et al. 2015; Kohne 2011).

Hemoglobinopathies or hemoglobin disorders are common genetic diseases that have an autosomal recessive inheritance pattern. Hemoglobinopathies are divided into quantitative and qualitative defects in globin synthesis. Quantitative defects lead to thalassemia syndromes, often with normal globin structure. Qualitative defects lead to hemoglobin variants with point mutations in globins (Kohne 2011; Payandeh et al. 2014; Brancaleoni et al. 2016).

Laboratory diagnoses of hemoglobinopathies are based on complete blood count (CBC) results and hemoglobin measurement by syan-methemoglobin, electrophoresis, and chromatography. Red blood cell morphology varies from normochromic normocytic to hypochromic microcytic, and hemoglobin concentration can reduce or normal (Kohne 2011).

In clinical laboratories, various errors may occur pre-analytically, analytically, and post-analytically. Lack of awareness of these errors can lead to confusion in diagnosis and treatment; therefore, identification of the potential error source is required to prevent it. Most laboratory errors are pre-analytical (32–75%). Pre-analytical errors include patient identification errors, inappropriate sample collection, poor sample quality, wrong transfer of sample, and others (Marin et al. 2014; Plebani and Carraro 1997; Tapper et al. 2017). Hemolysis is a pre-analytical problem that reduces sample quality for measurement of many analytes. In hemolysis, RBC membranes are destroyed and their contents released. Hemolysis can occur due to disease, centrifugation, prolonged storage on ice, and other causes (Ji and Meng 2011).

The aim of this study was to determine the effect of hemolysis on hemoglobin measurement by capillary electrophoresis.

Materials and methods

Specimens

Blood samples were obtained from Pardis clinical laboratory between December 2017 and April 2018. Samples from 311 female and 189 male subjects with normal and abnormal Hb electrophoresis patterns were studied. Three milliliters of whole blood was obtained from all subjects in EDTA tubes. To hemolyze the samples, 1.5 ml from each tube was removed and frozen for 24 h. The remaining 1.5 ml of non-hemolyzed blood was stored at 4 °C for 24 h. After 24 h, hemoglobin was measured in both the hemolyzed and non-hemolyzed samples by capillary electrophoresis.

Procedure

Sebia 2-flex piercing capillary electrophoresis was used to evaluate the hemoglobin. The instrument was fully automated, and Hb was measured using a Hemoglobin(e) kit purchased from the instrument manufacturer. In this method, an electrical field is established and the charged hemoglobins are separated by their electrophoretic mobility. Separation occurs according to the electrolyte pH and electroosmotic flow. The electrical charge of each hemoglobin depends on its structure (Jenkins and Ratnaike 2003).

Hemoglobin in hemolyzed and non-hemolyzed samples was measured at 415 nm. Before analyzing the samples, normal HbA2 control blood from Sebia was used for internal quality control.

Result and discussion

Data were analyzed by linear regression with SPSS16 software. The detection limits and number of samples are given in Table 1. The method was linear for HbA (y = 0.9986x + 0.3331, R2 = 0.9984), HbA2 (y = 0.9843x − 0.0851, R2 = 0.9796), and HbF (y = 1.0081x − 0.2448, R2 = 0.9984). Regression analysis is shown in Fig. 1.

Table 1 Number of samples and limit of detection for hemoglobin A, A2, and F
Fig. 1
figure 1

Regression curve of hemolyzed and non-hemolyzed samples for hemoglobin measurement. a Compression of hemolyzed and non-hemolyzed sample for HbA, HbA2, and HbF are shown in graphs a, b, and c, respectively

The coefficient of variation (%CV) in internal control for 50 different runs (between run) was 2.9%, and acceptable values were up to 3.5%.

According to these results, method comparison demonstrated good agreement between non-hemolyzed and hemolyzed samples for HbA, HbA2, and HbF. Because the Sebia system uses hemolysate solution and RBC membranes are destroyed, it seems hemolysis has no effect on hemoglobin measurement.

Capillary electrophoresis is an accurate instrument for screening and diagnosing hemoglobinopathies. Recently, the FDA recommended using capillary automated system even in routine laboratory diagnoses (Keren et al. 2008; Lippi and Plebani 2016).

Hemoglobinopathies are generally analyzed by capillary electrophoresis in reference laboratories. Other labs send their samples to these centers. Usually, sample hemolysis is due to inappropriate sampling or transferring. The risk of sample hemolysis is increased in laboratories that do not perform the test and send it at low temperature. One concern is that sample quality will be reduced during the transmission. Therefore, the aims of this study were to determine the effect of hemolysis on sample quality and finally hemoglobin measurement.

Conclusion

This assay demonstrates excellent linearity and good agreement between non-hemolyzed and hemolyzed samples. Thus, hemolysis has no effect on hemoglobin measurement and use of hemolyzed samples to measure hemoglobin patterns does not lead to erroneous results.