Background

Work-related musculoskeletal disorders (WMSD) are injuries or a group of painful disorders of muscles, tendons, skeleton, nerves, and related tissues that occur in work environments that significantly promote the condition or are made worse due to continuous extended activity. Surgery is one of many professions that have this issue. Several studies show that WMSD associated with performing surgery is incredibly common [1, 2]. Gutierrez-Diez et al. [3] showed that 90% of surgeons reported musculoskeletal disorders (MSDs) when performing minimally invasive surgery.

Most common injuries experienced by surgeons occur to the neck (82.9%), lower back (68.1%), shoulder (57.8%), and hands (45%) [4, 5]. To cope with this, surgeons use ergonomic interventions in the operating room, such as adjustable tables and instrument handles near the elbow level, to reduce discomfort and shoulder strain. Other interventions include improved instrument interface, proper monitor placement, adaptable imaging equipment, robotics, camera systems, and integrated hand controls [6,7,8].

However, one of the main ergonomic obstacles that surgeons face is the hand-held instrument design. Studies show that hand-held surgical instruments may cause musculoskeletal disorders. Specifically, prevalent instrument design issues are associated with laparoscopic surgery [9,10,11]. Berguer et al. [12] found that in a sample of 149 surgeons, the muscular work of the forearm and thumb muscles is increased when a laparoscopic device is used. In other words, the ergonomic coupling of the surgeon’s hand to the instrument is inadequate due to the handle configuration. Likewise, Trejo et al.’s [13] study revealed that a high proportion of surgeons had concerns such as stiffness, discomfort, and difficulty to perform precise movements when using laparoscopic tools. Twenty-nine percent of surgeons experienced numbness of the fingers or thumb after surgery while 66% faced neck pain when using traditional laparoscopic instruments.

This problem is not limited to laparoscopic or minimally invasive instruments. Surgeons within various surgical specialties (general surgery, plastics, orthopedic, and otolaryngology—head and neck surgery) also report pain and discomfort when using surgical instruments [5, 14,15,16] and attribute this to the design of the tool. Fram et al. [17] found that 48% of surgeons believe that instruments are not designed for them.

The solution is not simply to build a smaller tool. The majority of surgical instruments are built to perform a specified purpose often with little regard for ergonomics or for the ease of use with which the tools may be handled by the operator, which makes them hard to hold. Rather than the instrument adapting to the operator, the surgeon needs to adapt their operating style, possibly contorting their bodies, to the instrument. Many medical scissors, for example, are constructed with little ergonomic care for comfortable holding; therefore, many surgeons face problems with thumb flexibility and movements when using these types of scissors [5, 14,15,16].

There are many factors that must be considered when designing a surgical instrument. The design of functional medical devices requires a thorough grasp of human physical capabilities and limitations. A fundamental understanding of the numerous scientific disciplines involved in proper design such as engineering, psychology, anatomy, and physiology is required to build a functional medical device [18].

In this systematic review, the objective was to evaluate the association between sex, hand size, surgical instruments, and MSKIs by answering the following PICO question: “Do female surgeons or surgeons with small hand size, who use surgical hand-held instruments have an increased risk of musculoskeletal disorders and discomfort compared to male or large handed surgeons?”. Specifically, the goal was to evaluate every field of expertise and tool and analyze the instruments that female and small-handed surgeons encounter challenges and musculoskeletal injuries (MSKIs) with.

Methods

Search methodology

This study adopts a systematic review methodology to describe and analyze the effects of the use of unfit medical instruments on surgeons. The study protocol was registered on PROSPERO (ID: CRD42022283378) (https://www.crd.york.ac.uk/prospero/). The systematic review was conducted using the Cochrane methodology [19], to ensure a rigorous and comprehensive analysis of the available evidence. Hence, the 2020 Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) reporting guideline was followed [20].

Information sources

Author (R.H.) conducted the database search. The following databases were searched: (1) PubMed, (2) Embase, (3) CINAHL Plus with Full Text (EBSCOhost), (4) Scopus, for an ergonomics and health perspective, and (5) Web of Science Core Collection (including Science Citation Index Expanded, Emerging Sources Citation Index, and Social Sciences Citation Index), for a cross-disciplinary perspective. All databases were searched from inception to 03/17/2023 (Additional file 1).

Study eligibility criteria, screening, and selection

Inclusion criteria consisted of original, cross-sectional surveys on the ergonomics and human factors of surgical instruments, in relation to hand size, gender, and musculoskeletal symptoms. Publications included from the inception of the selection were required to be written in English and published in scholarly, peer-reviewed journals. Published conference abstracts were excluded, as these abstracts frequently did not get the same level of investigation as original papers. Any articles that had tools that required multiple surgeons to operate were excluded as the focus is on handheld tools by one surgeon. Articles that did not include gender or sex, hand size, or MSKI outcomes were excluded. Finally, articles that had non-handheld tools such as purely robotic instruments were excluded.

One reviewer (A.B.) screened titles and abstracts to identify articles for full-text review. Two reviewers (A.B., H.W.) independently examined the full text of the published journal article. Full-text analysis was carried out to find if the articles pertaining to the information were consistent with the PICO question under consideration. The disputes for article inclusion were settled through consensus-building discussions and, if necessary, a third reviewer (Q.W.) would give the final say.

The following items were abstracted for included articles: title, author names, gender or sex, glove size, medical equipment, medical outcomes, results, population size, medical field, type of article, date of study, age, anthropometric measurements, handiness (use of right or left hand), and time in the operating room.

A test for heterogeneity was conducted and it was found that it had an I2 of 80% when comparing studies in relation to gender. It also had an I2 of 83% when comparing results in relation to hand size. Due to the high heterogeneity, we decided that we will not conduct a meta-analysis and that a systemic review would be sufficient.

Risk of bias

The Checklist for Appraising Surveys tool was used to assess risk of bias in cross-sectional studies [21]. This method included thirteen detailed questions and was introduced in Foster and Jewell [21] Assembling the pieces of a systematic review: a guide for librarians book (Additional file 2). This method was used to rate each study as having a low risk of bias, medium risk of bias, or high risk of bias. Two reviewers (A.B., H.W.) did the analysis of the included articles.

Results

A total of 2165 articles were identified (PubMed, 361; Embase, 990; CINAHL, 111; Scopus, 442; and Web of Science combined, 261). There were 895 duplicates, and after removal, 1270 scientific papers were included for a title and abstract screening. A total of 1122 articles were excluded, leaving 148 articles for the full-text analysis. An additional 129 articles were excluded that did not meet the criteria, and an additional study was not considered as their full-text paper was unavailable, yielding a total of 19 papers to be included in this review (Fig. 1).

Fig. 1
figure 1

PRISMA flow diagram for systematic review

Risk of bias

The results of the risk of bias are shown in Table 1. Four articles were identified as low risk, 11 articles as medium risk, and 4 articles as high risk.

Table 1 Risk of bias results

Study characteristics

The key characteristics of the study are presented in Table 2. All included studies [17, 22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39] were from the past 19 years. The oldest study [23] was from 2004 and the most recent [17] was published in 2023. The sample size for the surgeon population varied across the studies; six studies [22, 24, 29, 34, 36, 38] had a small sample size (n ≤ 100), ten studies [17, 25,26,27,28, 30,31,32, 35, 39] had a moderate sample size (between 100 and 400), while three studies [23, 33, 37] had a large sample size (n ≥ 400). All studies provided detailed populations based on gender or sex except one [29]. Thirty-one percent of all the surgeons surveyed were female. All studies discuss the glove size of surgeons but only four [17, 23, 25, 30] had detailed descriptions of the glove size for each participant. These detailed glove sizes are shown in Fig. 2.

Table 2 Included articles’ key characteristics
Fig. 2
figure 2

Glove size by sex

Discussion

The results from the studies were mostly conclusive. As seen by most authors, medical instrument design is one major reason why female surgeons or surgeons with small hands face musculoskeletal injuries when practicing. The majority of survey surgeon report MSKIs and up to 87% of female surgeons report this has to do with instruments [23]. The authors described musculoskeletal issues that surgeons face differently based on their interests. Most prominently musculoskeletal injuries, pain, and fatigue were addressed the most, followed by the difficulty of use and stress. Medical instruments tested in the studies varied due to different fields of surgery, but mostly laparoscopic and endoscopic instruments had the most focus in the articles.

Impact of gender or sex

Gender was a big determinant in having musculoskeletal issues. Studies [17, 22, 24, 26,27,28, 31, 33,34,35,36, 38, 39] showed that around 77% of females report musculoskeletal issues ranging from difficulty of use to pain compared to 64% of males. Female surgeons (59 to 100%) compared to male surgeons (13 to 56%) had more problems when using surgical instruments. In all included studies, the average glove size for women (6.0 to 6.5) was significantly smaller than that for males (7.0 to 8.0), which was a big factor in having difficulty in using the instruments, so there is a link between gender and hand size [30]. Women were more likely to describe surgical hand-held instruments as “usually difficult” [22, 36, 37] and requiring the use of two hands [22]. Female surgeons were much more likely to report having negative views regarding orthopedic surgical tools and finding some devices to be challenging or painful to use [17]. Morais, Pawa, and Yong [31, 33, 38] mentioned that women also have a higher risk of MSKI due to differences in hand size and grip power. Endoscopic movements require greater strength and effort on the part of women [33]. This raises the possibility of developing a repetitive strain injury. According to Sutton [35], female surgeons are more likely to treat the hands more than their male counterparts related to doing the same surgical procedures, which include the wrist, thumb, and fingers. Kono et al. [40] dove deeper in another study and stated that since the required operating force exceeds the maximal grip force, it is physically difficult for the majority of Japanese women surgeons to fire the stapler by clutching the proximal side of the lever, which likely explains the tension these women felt and expressed.

Impact of hand size

Small glove size on its own was also cited as a reason for developing musculoskeletal issues when using surgical hand-held instruments. Studies [22, 23, 28, 31, 34, 37] showed that around 73% of surgeons who wear ≤ 6.5 glove size report musculoskeletal issues ranging from difficulty of use to pain compared to 31% of surgeons who wear a glove size > 6.5. Berguer [23] found that the small glove-size group had a larger percentage of participants who reported difficulty using any laparoscopic device. Shepard [34] also indicated that people with gloves smaller than 6.5 were more likely to report moderate to severe symptoms connected to handle dimensions and more likely to have worse symptoms connected to laparoscopic surgery activities.

Prevalence and type of instruments not fitting

Twelve of the articles included detailed reported problems when using surgical hand-held instruments. For example, Kono [28], Berguer [23], Weinreich [37], and Adams [22] all mentioned that the current design of the stapler causes pain and discomfort for surgeons especially female and small glove-handed surgeons. As a result, it was found that instrument use was a problem for females more than for males. These three studies [22, 23, 28] suggest that between 78 and 92% of females deem the stapler as hard to use compared to 41–56% of males. Filisetti [25] showed that the needle holder had a greater difficulty score in all hand-size groups. Morais [31], Pawa [33], and Yong [38] reported that around 72.5% of endoscopists faced at least one musculoskeletal injury due to the use of endoscopes. Moreover, Shepherd [34], Sutton [35], and Green [36] discussed the unfit use of some of the laparoscopic instruments that caused moderate to serious symptoms ranging from discomfort to back and neck problems.

However, in three of the included studies, there seems to be a high level of satisfaction with the fit of some of the surgical instruments that have been studied. Franasiak [26] reported instrument fit as “just right” for bipolar devices, graspers, and monopolar devices. Park [32] also mentions high percentage levels of instrument handle size being adequate for graspers, laparoscopic needle drivers, and staplers. Lucas-Hernandez [30] had results showing mixed reviews of the laparoscopic dissector but a high level of satisfaction for the laparoscopic needle holder. To clarify, these results are specific only to the fit and handle size of some of the mentioned instruments. Strength nor power or posture needed to use was discussed. Franasiak [26] reported a very high 88.1% of strain among the participating surgeons of the frequently used laparoscopic devices. Lucas-Hernandez [30] reported that 68% of surgeons take uncomfortable or forced postures when using the laparoscopic dissector and 61% reported they take an uncomfortable or forced posture when using the laparoscopic needle holder.

Nevertheless, there is minimal discussion of the biological factors between men and women with the same hand size in the included studies. Only Sutton [35] discussed the differences between both sexes with the same hand size; the author explained that women with big or small hand size reported more problems than men with the same glove size. This may be due to different factors but it certainly has to be investigated more. It is an area of study that is slightly neglected but exposed that the problems that face women when using medical instruments are more than just a difference in size problems.

Effects on training

Trainees also reported having problems when using surgical hand-held instruments. Kroon [29] stated that female trainees reported injuries to their dominant hand. Morais [31] also mentioned that 78.2% of surgeons believed that hand size affected endoscopy learning. Moreover, Pawa [39] found that half of the participants in a national survey of gastroenterology trainees reported having at least one endoscopic-related injury (ERI) and findings imply that certain ERI vulnerabilities manifest during training and should be checked. Improving these medical tools to accommodate more female surgeons and those with small hands will help in attracting more of these trainees into surgical professions as some might have decided against certain medical specialties due to the discomfort they had when practicing with these types of equipment.

Ergonomic suggestions

Three of the studies included in this systemic review suggest that improvements should be made to surgical hand-held instruments to help in lowering musculoskeletal issues and some propose ways on how to ergonomically improve some of the instruments. For instance, practitioners in the Shepherd [34] study suggested that rotating mechanisms, smaller handle dimensions, and softer handles as improvements to laparoscopic instruments. Lucas-Hernandez proposed enhancing the laparoscopic dissector and needle holder’s design mechanism to make it easier to use by making it lighter while keeping or boosting the sensitivity of the distal surgical instrument. Furthermore, 79% of those polled for Park’s [32] article claimed that they would use different-length instruments if they were available.

Strengths and limitations

This systematic review has several strengths. First, after a lengthy literature search, it is determined that this is the only review that has been done that tackles specifically the issue of difficulty of hand-held surgical instruments in relation to the combination of gender, sex, and hand size. Second, this study’s methodology was thorough. Moreover, the inclusion of detailed glove size analysis and instrument analysis in this study, along with a critical appraisal of the existing literature, helps to highlight the unique contributions of this study compared to previous systematic reviews [41, 42]. By carefully examining and evaluating the limitations or gaps in the existing literature, we were able to identify areas where this study can make a significant contribution to the field.

On the other side, there are certain limitations to this systematic review. There was significant heterogeneity among the studies in terms of definitions of MSKIs and thus precluded a meta-analysis. Additionally, terms used such as discomfort, pain, and difficulty of use are subjective. Furthermore, there is significant bias in how populations were selected for surveys and many of the studies did not publish methodology that was clear enough for reproducibility. Further research examining the impact of MSKI needs to move beyond subjective assessment and provide objective measurements that can then be compared across studies. Furthermore, it must be mentioned that 18 of the included studies that discussed hand-size dealt with glove size and not anthropometric hand measurement sizes. These glove sizes are determined by surgeons themselves, which may lead to some surgeons wearing bigger or smaller glove sizes.

While recognizing the limitations of this systematic review, one can observe that the papers included in this review span a period of 19 years. It is important to understand that, despite the advancements in technology and medicine over this period, many of the medical instruments included are still the ones that are used today. Even though some of the included studies were carried out on older dates, they can nevertheless provide insightful information. Additionally, the fact that fairly recent publications still report problems using the same tools as old published journal articles suggests a persistent issue that has not dissipated over the 19-year span. Notably, comparing the oldest paper [23] in the review with one of the newest [36] reveals a noteworthy observation: both studies highlight issues with the same laparoscopic tools, indicating that despite potential advancements, certain challenges persist in the use of these instruments.

Conclusion

As this systematic review points out, many female and small hand surgeons are having major issues when dealing with medical instruments. Thus, there is a need to develop new ergonomic designs for some of the surgical hand-held instruments that surgeons use most often. The number of female surgeons has risen substantially in the last decade, so there is an imperative need to address the major challenges that they face when operating. Having a productive environment through a complete health and safety evaluation of the design and potential redesign of medical instruments will not only reflect positively on the surgeon’s health and ability but also could create scenarios for better patient health outcomes as well.