Introduction

Endotracheal intubation (ETI) is performed in many patients admitted to the intensive care unit (ICU), the proportion being 22% in a multicenter study [1]. The circumstances that surround ETI in the ICU are associated with a high complication rate of up to 50% [1,2,3,4]. The serious complications are severe hypoxemia (26%), severe hypotension (25%), cardiac arrest (1–3%), and death (0.5–3%) [1, 4, 5]. Severe hypoxemia, which can be fatal [7], is more common when the patient is hypoxemic before intubation and/or intubation is difficult [8, 9].

Predicting the occurrence of hypoxemia during ETI is challenging [10]. Preoxygenation reduces the risk by prolonging the safe apnoea time and is therefore universally recommended [11, 12]. Preoxygenation consists in giving pure oxygen to wash out the nitrogen contained in the lungs at functional residual capacity. The effectiveness of preoxygenation is evaluated experimentally and in the operating room by the expired fraction of oxygen (FeO2), but this marker has limitations in the emergency setting (sensitivity to leaks) and is not available in the ICU [13]. Recent studies have evaluated various preoxygenation devices with contradictory results [6, 14,15,16], and currently no device ensures that desaturation will not occur during intubation [17]. Pulsed oxygen saturation (SpO2) measured by pulse oximetry—which is the parameter monitored in intensive care—only detects hypoxemia at a late stage. Thus, at present, the effectiveness of preoxygenation cannot be evaluated in intensive care. The absence and/or low quality of preoxygenation was associated with the occurrence of cardiac arrest during the peri-intubation period in a retrospective analysis of a multicentre prospective database that had 1847 ETI procedures [7]. Thus, optimising preoxygenation is a crucial goal. To optimise preoxygenation, the availability of a reliable predictor of desaturation would be valuable.

The oxygen reserve index (ORI) is a new oxygenation monitoring parameter measured continuously and noninvasively by a specific pulse oximeter device manufactured by Masimo (Irvine, CA). It provides a dimensionless index from 0.0 (PaO2  ≥  100 mm Hg) to 1.0 (PaO2  ≥  200 mm O2). Thus, the ORI supplies information beyond the range explored by SpO2.

The objective of this prospective observational pilot study was to evaluate the potential role for ORI in providing early warning that hypoxemia will occur during ETI of patients in the ICU.

Patients and methods

Study design

We conducted a prospective observational proof-of-concept single-centre study. We collected the data in the medical ICU of the Nantes University Hospital (Nantes, France), which has 25 ICU beds and five intermediate-care beds.

The study was approved by our ethics committee on September, 2018 (CPP Ile de France 1, ID-RCB: 2018-A01288-47) and was open to inclusion on ClinicalTrials.gov on January 30, 2019 (#NCT03600181). Written informed consent to participation was obtained from each patient or proxy. According to French legislation, patients who were not competent and had no proxy available were included if they met the selection criteria then asked for their consent as soon as they recovered competency.

Oxygen reserve index

The oxygen reserve index (ORI) is a nondimensional index that ranges from 1 (high reserve) to 0 (no reserve) and is measured by optically detecting changes in mixed venous oxygen saturation (SvO2) after oxygen saturation (SaO2) reaches 100%. It is measured by a multi-wavelength pulse co-oximeter placed on the tip of a finger (Rainbow SET, Masimo) [18]. The device analyses variations in the pulsatile blood absorption of incident light at both the arterial and venous levels. It is a relative indicator of PaO2 changes in the moderate hyperoxemia range. When pure oxygen is administered, SaO2 reaches 100% when PaO2 reaches 100 mmHg. Beyond that, PaO2 continues to increase, both SaO2 and SpO2 remain at 100%, and the ORI increases non-linearly from 0.00 (PaO2–100 mm Hg) to 1.00 (PaO2–200 mmHg).

Patient selection

Inclusion criteria were ICU admission with a need for ETI and a SpO2/FiO2 ratio above 214. The SpO2/FiO2 ratio was measured during non-invasive ventilation (NIV) or high-flow oxygen therapy. For conventional oxygen therapy, the fraction of inspired oxygen (FiO2) was calculated as follows: FiO2 =  0.21  +  O2 flow·0.03 [19].

Exclusion criteria were age younger than 18 years, indication to use an alternate tool to perform ETI (unstable spinal cord injury for example), insufficient time to include the patient (e.g., cardiac arrest), pregnancy or breastfeeding, being a correctional services inmate, being under guardianship, or not being covered by the French statutory health insurance system.

Patients were secondarily excluded in the event of an ORI device malfunction, failure of ORI recording by the computer, ORI remaining constant at 1 throughout preoxygenation, and SpO2  <  97% throughout preoxygenation [20].

Data collection

Standardised forms were used to record the following data: patient’s baseline characteristics [including difficult intubation criteria: Mallampati score, thyromental distance  <  65 mm, mouth opening  <  35 mm, limited cervical mobility, sleep apnoea, body mass index (BMI)  >  35], main reason for ETI (neurological, respiratory, cardiovascular, or other), use of a bougie and/or other devices (e.g., laryngeal mask airway or videolaryngoscope), duration of preoxygenation, total ETI duration (from anaesthesia induction to capnography over more than three cycles indicating proper endotracheal tube position), and baseline SpO2. We also collected SpO2 and ORI at four time points: beginning of preoxygenation, end of preoxygenation, during ETI and just after successful ETI. SpO2 drops below 97%, 90%, and 80% between anaesthesia induction and successful ETI were recorded. Complications were recorded as death, cardiac arrest, and systolic blood pressure drop to less than 90 mmHg [21]. Case-report forms were completed in real time by a dedicated clinical research nurse. Each case report form was then introduced into an electronic file (Excel, Microsoft Corporation, Redmond, WA).

Outcomes

The primary endpoint was the time between the ORI decrease below 0.4 and the SpO2 decrease below 97% during ETI (between the end of preoxygenation and confirmation that the tube was in the trachea). The value of 0.4 for ORI was chosen in agreement with the only available study [22].

The secondary objective of our study was to determine whether a decline in ORI during preoxygenation predicted the occurrence of SpO2  <  97% during ETI.

Intubation procedure

Once the decision to perform ETI was made and consent obtained, an ORI device (Rainbow® Lite SET-1 Adt sensors, Revision M, Masimo) was applied to the patient’s 3rd or 4th fingertip on the contralateral side of the non-invasive blood pressure monitoring device. The sensor was covered to protect it from light. SpO2 and ORI values displayed on the Rad7® monitor were recorded every 2 s and transferred to a laptop computer throughout the ETI procedure. The patient’s physicians were not aware of the ORI values.

ETI was performed according to the standardised protocol used in the ICU of the Nantes University Hospital [21]. The choice of the preoxygenation device, anaesthesia induction agents, and ETI device was at the discretion of the physician in charge. The recommended preoxygenation duration was 3 min.

The procedure began with the initiation of preoxygenation and ended with confirmation that the tube was in the trachea. The end of preoxygenation was defined as the induction of anaesthesia. The duration of ETI was the time from the end of preoxygenation and confirmation by capnography of correct tube position. Neither apneic oxygenation nor apneic ventilation [23] was part of our local protocol during management of the study patients.

Sample size

In the absence of critical care data, we estimated the time to an SpO2 decrease to 97% at 60  ±  30 s [22, 24] and we assumed that the ORI would decrease below 0.4 (warning cut-off determined by the manufacturer and consistent with data from a paediatric study [22]), 30 s before SpO2 decreased below 97%. We estimated the median time between these two decreases, with the 95% confidence interval (95% CI) and an accuracy of 10 s. Assuming a standard deviation of 30 s, 35 patients were needed [25]. According to Szmuk et al. [22], it was expected that 28% of patients would have a secondary exclusion criterion, and we therefore decided to include 50 patients.

Statistical analysis

Qualitative data were described as frequency and percentage and quantitative data as mean  ±  SD and median [interquartile range].

The median time between the ORI decrease below 0.4 and the SpO2 decrease below 97% during ETI was estimated with its 95% confidence interval (95% CI).

The performance of the ORI for predicting SpO2  <  97% during ETI was estimated by the area under the ROC curve with its 95% CI. To evaluate the informative value of the ORI signal, a multivariate logistic regression model with the occurrence of SpO2  <  97% as a predictor variable was constructed. The variables included by step-down selection were the ORI and the variables selected a priori as clinically relevant [10] (age, gender, BMI, and ETI duration).

All tests were two-tailed with a significance level of 0.05. No imputation strategy was used. The statistical analyses were done using Stata® statistical software (version 13; StataCorp LP, College Station, TX).

Results

Patients

Between February 2019 and July 2020, 56 patients were included. Main reasons for non-inclusion were SpO2/FiO2  <  214 (n  =  114, 76%) followed by omission (n  =  21, 14%) (Additional file 1: Figure S1). Among the 56 included patients, 5 were secondarily excluded, leaving 51 for in the analysis. Table 1 reports their main features. No patient had a history of difficult intubation, the Mallampati score was 1/4 in 36 (70.6%) patients, and 22 (43%) patients met none of the criteria for difficult ETI.

Table 1 Patients’ baseline characteristics

Procedure

Preoxygenation was performed with a bag valve mask (n  =  40, 78.4%) or NIV (n  =  11, 21.6%) and ETI with a standard laryngoscope (n  =  48, 94.1%) or a videolaryngoscope (n  =  3, 5.9%). A bougie was used in 16 (31.4%) patients. The anaesthetic agents used for induction were etomidate in 43 (84.3%) patients and propofol in 8 patients (15.7%), for hypnosis; for neuromuscular blockade, succinylcholine was used in 41 (80.4%) patients and rocuronium in 9 (17.6%) patients, with the remaining patient receiving no neuromuscular blocker. The operator was a resident for 48 (94.1%) patients; residents were closely supervised by senior intensivists.

ETI was achieved in all patients, with a median number of attempts of 1 [1, 2] and a proportion of successful first attempts of 74.5% (n  =  38). In the remaining 13 patients, ETI was achieved after at least one additional attempt; the reasons for first-attempt failure were poor glottis visibility in 5 (38.5%) patients and tracheal catheterisation failure in 8 (61.5%) patients. ETI was difficult in 3 (5.9%) patients (i.e., required at least three attempts). The only patient who needed more than three attempts was intubated on the seventh attempt. At the time of exposure, head elevation was required in 7 (13.7%) patients, backwards-upwards-rightwards pressure (BURP) on the larynx in 14 (27.5%) patients, and the Sellick manoeuvre in 3 (5.9%) patients. The Cormack–Lehane grade was 1 or 2 in 45 (88.2%) patients and 3 or 4 in 6 (11.8%) patients. The median percentage of glottic opening seen (POGO) was 80 [70–100]. In 6 (11.8%) patients, face-mask ventilation was required after the first attempt due either to desaturation (n  =  3) or to first-attempt failure (n  =  3). The median lowest SpO2 during intubation was 98% [91–99] in patients who did not need face-mask ventilation and 72% [55–82] in those who did require face-mask ventilation before ETI. SpO2 fell below 97% in 20 patients.

ETI-related complications occurred in 24 (24/51, 47.1%) patients. By decreasing order of frequency, they consisted of arterial hypotension (n  =  18, 75%), moderate hypoxemia (SpO2  <  90%, n  =  9, 37.5%), severe hypoxemia (SpO2  <  80%, n  =  4, 16.7%), oesophageal intubation (n  =  2, 8.3%), and aspiration (n  =  1, 4.1%). No patient had cardiorespiratory arrest or death due to ETI.

Overall, 9 (17.6%) patients died in the ICU. On day 28, 42 (82.4%) patients were alive. The median ICU length of stay was 5.36 days [2.9–15.8].

Primary outcome

Of the 51 patients, 20 experienced an SpO2 fall below 97% between the end of preoxygenation and successful ETI. An ORI fall below 0.4 preceded this SpO2 fall under 97% in 10 of these patients. In the other 10 patients, when the SpO2 fall below 97% occurred, the ORI had not increased above 0.4. In the 10 patients with a warning ORI fall below 0.4 before the SpO2 fall below 97%, the median time between these two events was 81 s [34–146] and the median ETI duration was 303 s [180–648].

Secondary outcomes

As noted above, of the 51 patients, 20 (39.2%) experienced SpO2  <  97% during intubation. Table 2 and Figs. 1 and 2 show the changes in SpO2 and ORI from the start of preoxygenation to the end of intubation. Additional file 2: Figure S2 shows changes in SpO2 and ORI in the subset of patients who had moderate hypoxemia and Additional file 3: Figure S3 shows changes in all patients.

Table 2 Comparison of peripheral oxygen saturation (SpO2) and oxygen reserve index (ORI) in patients with and without SpO2  <  97% during endotracheal intubation (ETI)
Fig. 1
figure 1

SpO2 changes during preoxygenation and endotracheal intubation (ETI)

Fig. 2
figure 2

Changes in oxygen reserve index (ORI) in patients with and without hypoxemia during endotracheal intubation (ETI)

The areas under the ROC curve for ORI during preoxygenation and at the end of preoxygenation predicted the occurrence of SpO2  <  97% during intubation (0.73; 95% CI 0.58–0.88 and 0.70; 95% CI 0.54–0.85, respectively). In contrast, the area under the ROC curve of SpO2 at the end of preoxygenation was poorly predictive of SpO2  <  97% during intubation (0.54; 95% CI 0.40–0.67).

By univariate analysis, longer ETI duration was associated with the occurrence of SpO2  <  97% during intubation (odds ratio, 1.00; 95% CI 1.00–1.01; P  =  0.0474), whereas a higher ORI value during preoxygenation was associated with a less frequent occurrence of SpO2  <  97% (odds ratio, 0.09; 95% CI 0.01–0.69; P  =  0.0199). The highest ORI value during preoxygenation remained significantly associated with a lower risk of SpO2  <  97% during ETI after adjustment for ETI duration and BMI (odds ratio, 0.76; 95% CI 0.61–0.95; P  =  0.0141; Table 3).

Table 3 Multivariable analysis: factors associated with SpO2  <  97% during endotracheal intubation (ETI)

Discussion

The main findings from our study are that in critically ill patients who require ETI, are not hypoxemic, and whose ORI is above 0.4 during preoxygenation, the median time between the ORI decrease below 0.4 and the SpO2 decrease below 97% was 81 s [34–146] and that, by multivariate analysis, a higher ORI value during preoxygenation was associated with a lower risk of desaturation below 97% during ETI, after adjustment on ETI duration. The median highest ORI value during preoxygenation was higher in the group without desaturation below 97%. In contrast, no differences between the two groups were found for the SpO2 values at the beginning or end of preoxygenation or for the change in SpO2 between the beginning and end of preoxygenation.

Our results are consistent with those found in patients undergoing elective surgery, which were prospective observational studies with small sample sizes. Reported time intervals between the ORI decline and desaturation were 31.5 [19.0–34.3] s [22], 32.5 [18.8–51.3] s [20], and 48.4 [40.4–62.0] s [22]. Thus, the warning time in patients without critical illness was shorter than in our cohort. Possible explanations are the variations in cut-offs chosen to define hypoxemia, inclusion of children in some studies, use of the latest sensor model (revision M of revision L) in our study, possible continuation of oxygenation during the apneic period in previous studies, and/or successful ETI in some patients after a failed first attempt but before the occurrence of decreases in ORI and SpO2.

SpO2 changes during preoxygenation serve as a proxy for PaO2 changes, but neither evaluate the oxygen reserve nor predict hypoxemia during ETI [26]. SpO2 values may even be falsely reassuring: in our study, the median SpO2 at the end of preoxygenation was 100% [99–100] in both groups. The fraction of oxygen in expired air (FeO2) can also provide information on the oxygen reserve [27]: an FeO2 of 90% is taken to indicate denitrogenation of the functional residual capacity (FRC). In the operating room, preoxygenation for 3 min is usually sufficient to bring FeO2 up to 90%. However, this technique has limitations in the emergency setting, as it is sensitive to leaks, and is not available in the ICU [13]. Moreover, in critically ill patients, particularly those with acute hypoxemic respiratory failure (who were excluded for this proof-of concept study), FeO2 may not reliably reflect the effectiveness of preoxygenation: the reduction in functional lung volume leads, on the one hand, to a reduction in FRC and, on the other hand, to shunting that impairs the efficiency of the alveolar–capillary interface [28]. The result is that both SpO2 and FeO2 can be high despite PaO2 being low. PaO2 can be considered the reference standard for evaluating the effectiveness of preoxygenation but, unfortunately, cannot currently be obtained at the bedside in real-time in clinical practice [13].

We chose the SpO2/FiO2 ratio as the inclusion criterion because it was simpler to use than the PaO2/FiO2 ratio, as obtaining an arterial blood sample may be difficult in emergency situations. The 214 cut-off was chosen as corresponding to a PaO2/FiO2 ratio greater than 180 [29]. Including patients with more severe hypoxemia may not have allowed for ORI increases to occur in both groups during preoxygenation. We defined mild hypoxemia as SpO2 below 97%, as this value corresponds to the inflection point towards a rapid decrease in SpO2 during the apneic period [24] and to an ORI value of 0 indicating an absence of oxygen reserves [18].

The clinical implications of our findings may be important. The 81-s forewarning may allow immediate intubation, early face-mask ventilation, insertion of a supraglottic device, or a call for help in the event of intubation difficulties. In addition, ORI monitoring can help identify patients who do not increase their oxygen reserve despite preoxygenation and are therefore at risk of desaturation during ETI. These patients may benefit from a longer preoxygenation period and/or a change in device. An ORI decline might lead to the detection of a fault in the preoxygenation technique such as an insufficient oxygen flow rate or major leaks. The clinical impact of an ORI-guided airway management and preoxygenation strategy remains to be evaluated prospectively.

Our work has several limitations. First, the sample size is small. We had estimated that at least 35 patients had to be included in the analysis to compute the median time from the ORI decrease to the SpO2 decrease. However, only 10 of our patients exhibited both ORI  <  0.4 and SpO2  <  97% during intubation, and 10 patients did not have an ORI increase  >  0.4 during preoxygenation. Second, 5 patients, representing 9% of the initial cohort, were excluded secondarily due to either a malfunction of the ORI system or a failure of the Rad7® monitor to record the ORI. Third, at the beginning of preoxygenation, the ORI value had already started to increase and was therefore not equal to 0 in either group. However, there was no significant difference in ORI values at this time point between the two groups. Fourth, our results cannot be generalised to all ICU patients. We selected patients with SpO2  <  97% and no haemodynamic failure who required ETI mainly to protect the upper airway due to neurological failure. The incidence of hypoxemia during ETI is lower in this population than in patients with respiratory failure. Studies in patients with hypoxemia are therefore necessary. Fifth, as with pulse oximetry, ORI measurement can be unreliable in situations where the peripheral perfusion is impaired, such as in the event of shock or high-dose vasopressor therapy. Sixth, significant inter-individual variability is likely, as some factors may affect the calculation of ORI and therefore alter the relationship between PaO2 and ORI. These factors include oxygen consumption, cardiac output, temperature, pH, PaCO2, presence of abnormal haemoglobin, and venous pulsatility.

Conclusion

The median time between the ORI decrease below 0.4 and the SpO2 decrease below 97% during the apneic period was 81 s [34–146]. A higher ORI during preoxygenation was independently associated with a lower risk of mild hypoxemia (SpO2  <  97%). Whether these findings apply to hypoxemic patients, and the clinical impact of a preoxygenation strategy based on ORI monitoring, remain to be evaluated prospectively.