Background

Airway hyperreactivity (AHR) is a characteristic feature of bronchial asthma, and methacholine challenge testing (MCT) is well established to quantitate AHR in patients with unexplained symptoms such as cough, chest tightness and/or dyspnea, when the diagnosis of asthma is uncertain [110]. Compared to control subjects the provoked bronchial obstruction appears earlier and at lower provocation doses, and is more intensive in patients with asthma, a functional feature serving as rationale for the underlying mechanisms of AHR [4, 11, 12]. Both the European Respiratory Society (ERS) [1] and the American Thoracic Society (ATS) [4] recommend bronchial provocation tests by inhalation of aerosolized methacholine (MCH), considering this approach to be a reproducible and relatively easy to perform test in adults and children.

Following MCT, the cumulative provocation dose (PD) at which the forced expiratory volume in 1 s (FEV1), measured with spirometry, decreases at least 20% compared to baseline (PD−20FEV1), is currently the most commonly used outcome of assessing AHR by detection of flow limitation in consequence of airway narrowing [9]. Alternatively, whole-body plethysmography can also be used to determine AHR, measuring changes in airway mechanics either by a 35% [2, 7], a 40% [4, 5], or a 50% [5] reduction in specific airway conductance (sGaw) or an increase of specific airway resistance (sRaw) of 100% [8, 10] from baseline. To our knowledge, studies comparing reliability of spirometric parameters with those obtained by whole-body plethysmography, or even a combination of both are rare, and the so-called effective specific airway conductance (sGeff) has never been evaluated as a target parameter.

Therefore, the purpose of the present study was to compare diagnostic value of the response characteristics to MCH assessed by FEV1 and forced expiratory flow at 50% of vital capacity (FEF50), in relation to sGeff differentiating different diagnostic and functional groups and to encounter possible differences of AHR in relation to pulmonary hyperinflation and/or the phenomenon of dysanapsis [1317].

Methods

Study population and ethics

The study was conducted in the Division of Pulmonary Medicine, Clinic Barmelweid, CH-5017 Barmelweid, Switzerland, recruiting patients referred to the pulmonary function laboratory for baseline pulmonary function and MCT in patients with symptoms suggestive for asthma, such as cough, shortness of breath, wheezing or chest tightness. According to the classical definition of the GINA dissemination committee report [18], and based on the characteristics of the case histories and clinical findings, two experienced pulmonary physicians (Co-authors MF and TS) have classified the cases previously into a group of patients with proven bronchial asthma (n = 337; 69.8%), and a group of non-asthmatic subjects (n = 147; 30.4%). Bronchial asthma was diagnosed when subjects presented with a medical history of atopy (allergic rhinitis, hay fever, exercise-induced, partly also an eosinophilic inflammation, infection-induced, occupational or intrinsic), and documented with a positive MCT (PD−20FEV1). The groups of non-asthmatic subjects was composed of 4 subgroups according to the potential origin of pulmonary symptoms differentiating either between (i) an “upper airway syndrome” (UACS), previously also termed as post-nasal-drip-syndrome [19] (n = 55; 37.4%), (ii) a gastroesophageal reflux disease (GERD, n = 40; 27.2%), (iii) “persistent chronic cough” lasting more than 8 weeks [20] (n = 30; 20.4%), and patients with (iv) a “symptom complex” such as unexplained dyspnoea, chest tightness, hyperventilation, or somatisation (n = 22; 15.0%), which could not be attributed to a clear diagnosis. There were no patients with chronic obstructive pulmonary disease (COPD), bronchiectasis, cystic fibrosis, obstructive sleep apnoea syndrome (OSAS), or interstitial lung disease. Short-acting ß2-agonists were withdrawn for 8 h, long-acting beta-agonists for 48 h, and leukotriene receptor antagonists for 24 h prior to the lung function testing. Inhaled corticosteroids were withdrawn 7 days before MCT.

We conducted this study retrospectively as a case controlled study in order to compare diagnostic accuracy of the MCT assessed by different target parameters between the diagnosed asthmatic patients and non-asthmatic subjects. Inclusion criteria were reproducible base-line measurements, at least 5 plethysmographic tidal breath efforts as well as at least 3 reproducible forced expiratory manoeuvres at each provocation level. The study was planned according to the Federal Law of Human Research, conceptualized according to the Swiss Ethics Committee on Research involving humans, and approved by the Governmental Ethics Committee of the State of Berne. Master-files haven been stored and secured in the REDCap-system of the Clinical Trial Unit, Medical Faculty, University of Berne, Switzerland.

Pulmonary function procedures

Spirometry and plethysmographic measurements were performed using standard techniques according to ATS-ERS recommendations [21, 22] previously established and subsequently extended (http://www.atsjournals.org/doi/suppl/10.1164/rccm.200407-948OC/suppl_file/online_methods.pdf) [23]; http://www.atsjournals.org/doi/suppl/10.1164/rccm.200603-423OC/suppl_file/onlinesup200603-423ocr2.pdf) [24]; http://www.biomedcentral.com/content/supplementary/1465- 9921-10-106-S1.doc) [25].

Using a Jaeger MasterLab whole-body plethysmograph (CareFusion, Würzburg, Germany), measurements of airway mechanics, slow spirometry and hence assessment of static lung volumes were carried out first and only afterwards forced flow-volume loops were performed according to the ATS-ERS recommendations (best value of the 3 trials). The sReff, and its reciprocal value the sGeff resp., were assessed while the subject breathed tidally, in an upright position, through filter and measuring head in the body plethysmograph [26], without requiring any special breathing manoeuvres or efforts against a closed shutter. Both measurement techniques (whole-body plethysmography and spirometry) were performed with the subjects in sitting position within the whole-body plethysmograph cabin.

Assessment of airway mechanics

In order to obtain a parallel synoptical presentation of the MCT, we routinely present the reaction of FEV1 and FEF50 together with sGeff. sGeff is computed as the ratio between the integral of the area of the tidal flow-volume loop as numerator (∮VdV T ) and the integral of the area enclosed by the specific resistive work of breathing (sWOB = ∮ΔV box dV T ) [27] according the equation:

$$ s{G}_{eff} = \frac{1}{P_{amb}-{P}_{H_2O}}\ast \frac{{\displaystyle \oint }{V}^{\prime }d{V}_T}{{\displaystyle \oint}\Delta {V}_{box}d{V}_T} = \frac{1}{s{R}_{eff}} $$

where Pamb is the barometric pressure, PH2O the saturated vapour water pressure at body temperature. The mathematical background to obtain sGeff as the reciprocal value of the effective specific resistance (sReff), has been given previously [2325, 28], and is presented synoptically in Fig. 1. An Additional file 1 furthermore provides details regarding the paradigm shift in the assessment of airway mechanics. During quiet breathing at end-expiratory lung volume (EELV) measurements of sGeff were measured in a first phase of testing. In a second phase of testing functional residual capacity (FRCpleth) was assessed by normal resting breathing against the closed shutter (no panting). In a third phase a slow maximal vital capacity manoeuvre, carefully linked with the shutter-measurement was performed to get the static lung volumes, functional residual capacity (FRCpleth), residual volume (RV) and total lung capacity (TLC). However, only in a forth phase of testing 3 consecutive forced flow-volume-loops were recorded. The median of at least 5 consecutive single measurements of sGeff and 3 measurements of FRCpleth were computed, and from the triplet of forced expiratory manoeuvres the best effort was analysed. The pulmonary function test data were expressed as a percentage of predicted normal values, and z-transformed accordingly [21, 29, 30].

Fig. 1
figure 1

Print-screen, originally depicted from the Jaeger infant whole-body plethysmograph showing breath-by-breath tracings from which effective specific resistance (sReff), and its reciprocal value, the effective, specific airway conductance (sGeff) are computed using the integral of the tidal flow-volume area (upper tracing) and the integral of the plethysmographic shift-volume versus tidal volume area, the latter representing the specific resistive work of breathing (middle tracing). The crossbar clearly demonstrates that the zero-flow points are perfectly in phase

The assessment of the dose–response characteristics to MCH, observed not only by forced spirometry, but additionally by whole-body plethysmography, was chosen because a certain paradigm shift at least in the technical approach of the assessment of airway mechanics is established since several years, albeit the combined assessment of AHR by MCH by spirometry and whole-body plethysmography is not yet well validated and established in its clinical use. However, improved reliability of test results can be expected by assessing the bronchial reaction to sGeff, which is computed real-time by an integral, multi-dimensional approach. Details are given in the Additional file 1. Our anticipated advantages, therefore, were to elaborate, whether or not in comparison to the spirometric approach, the new plethysmographic technology could provide (i) better diagnostic accuracy, (ii) leading by independency of deep inspirations and hence changes in the volume-history [10, 3136] to less “a priori” modulation of the airway calibre, moreover, (iii) by independency from the subject’s cooperation (need of forced breathing manoeuvres) circumvention of inadvertent change of the airway responsiveness during test procedure [34, 37, 38].

The aerosol provocation system (APS)

MCT was performed according to a protocol routinely used in several pulmonary function laboratories in Europe, applying a modified dose oriented, single concentration sequence [39] with the Aerosol Provocation System (APS) using the SideStream nebulizer of Philips/Respironics previously described [38, 40]. Technical details and advantages of the APS are given in the Additional file 2, presenting the similar physical characteristics likewise they were recently outlined by Kannan et al. [41]. In their novel efforts they defined high fidelity computational simulations, performed over several breathing cycles, to get information regarding regional deposition for different particle sizes and an algorithm accounting for the re-entry of particles during the exhalation phase.

Methacholine challenge test (MCT)

After baseline measurements, MCH (5%) was administered in 3 steps of increasing dose. However, in contrast to a previously proposed 4-step procedure [39], a one-concentration-3-step protocol with increasing single doses of MCH (0.2 mg, 0.8 mg, 1.2 mg) was performed, consisting of 3 consecutive levels of cumulative doses defined as CD1: 0.2 mg; CD2: 1.0 mg; CD3: 2.2 mg. The measurements of whole-body plethysmography and spirometry were performed two minutes after each MCH-inhalation. On the basis of the inhalation time and the nebulizer output provided by the APS and the known concentration of the MCH, the applied doses of MCH was individually computed at each provocation level. Response to MCH was assessed by linear regression against the logarithmic doses of MCH giving the individual PD causing either a 20% decrease of FEV1 (PD−20FEV1), a 40%, 45% or 50% decrease of sGeff (PD−40sGeff, PD−45sGeff, PD−50sGeff, resp.), or a 20% decrease of FEF50 (PD−20FEF50). A synopsis of the assessment of AHR by this technique is presented in the Additional file 3. Break off criterion were when a decrease in FEV1 > 20% was reached or symptoms occurred. After having defined PD−40sGeffas the most discriminating factor and in order to distinguish the degree of AHR to MCH, patients were classified into 4 functional severity groups: (1) non-reactive if PD−40sGeff ≥ 2.0 mg, (2) low-reactive if PD−40sGeff ≥ 1.0 mg but < 2.0 mg, (3) medium-reactive if PD−40sGeff ≥ 0.2 mg but < 1.0 mg, or (4) severe AHR, if PD−40sGeff < 0.2 mg. Each provocation test was terminated with 2 puffs of salbutamol inhaled from spacer device.

Data analysis and statistical methods

The discriminative power of each lung function parameter was evaluated by measures of diagnostic accuracy such as sensitivity, specificity, positive predictive values (PPV), negative predicting values (NPV), likelihood ratios (LR+; LR-), the area under receiver operating curves (ROC), the Youden’s index (J) and diagnostic odds ratios (DOR). Statistical comparisons were performed applying McNemar testing. Using these different statistical procedures several aspects of diagnostic accuracy, such as predictive ability and/or discriminative property of the MCT, could be evaluated [42]. The Additional file 4 (Statistical Approach) provides details of all the statistic methods applied. Diagnostic accuracy was tested in a first step by cross-tabulation comparing the proportions of positive and negative reaction to the MCH challenge for each parameter (sGeff, FEV1 and FEF50) using Chi-squared tests. Statistical analysis was performed with the IBM SPSS version 24.0 (SPSS Inc., Chicago, IL). The limit of significance was a p-value of 0.05.

Results

Patient characteristics

A total of 484 patients (199 males, 51.1%; 285 females, 58.9%; age-range 9.11–87.1 years) were eligible for inclusion in the study stratifying patients into a group of asthmatic patients (n = 337; 69.6%) and non-asthmatic subjects (n = 147; 30.4%) previously diagnosed. Table 1 shows that in the gender distribution, more females than males were found, especially in the group of asthmatics. Non-asthmatic subjects were slightly older than asthmatics. The distribution within age-classes (not shown) revealed that the younger the collective, the more asthmatics, and the older the collective, the more non-asthmatic subjects were found. The stratification into different functional groups assessed at baseline prior to MCT shows for plethysmographic measurements normal lung function in 83.1% of asthmatic patients and 80.3% of non-asthmatic subjects. Interestingly, a considerable percentage of patients presented with a pulmonary hyperinflation (FRCpleth > +1.645 SD), without or in combination with bronchial obstruction (sGeff < −1.645 SD). Regarding spirometry, normal flow-volume curve indices were found in 90.5% of asthmatic patients and 93.9% of non-asthmatic subjects. Only a very minor percentage of asthmatic patients (7.7%) presented with flow limitation (FEV1 < −1.645 SD) at baseline.

Table 1 Subject’s characteristics (N = 484; 199 males (41.1%, 285 females (58.9%), age-range (9.1–87.1 years

Dose–response curves obtained by using different lung function parameters

In Fig. 2 the dose-responses to MCH at each provocation level, expressed as percent changes from baseline (BL set to zero) for each lung function parameter are synoptically presented, comparing response in asthmatic patients with non-asthmatic subjects. It can be shown that the response assessed by sGeff was much quicker and much more pronounced than the response assessed by FEV1 or FEF50. The cumulative percent-responses at each cumulative dose- (CD)-level demonstrate completely different response characteristics regarding the 3 target parameters. In asthmatic patients the cumulative percent-response of sGeff was at the 1st PD-level (0.2 mg MCH) 3.7 times more pronounced than with FEV1, 1.6 times than with FEF50 resp. At the 2nd CD-level (1 mg MCH) the percent-response of sGeff was 2.4 times more pronounced than with FEV1, 1.4 times more than with FEF50 resp., and at the 3rd CD-level (2.2 mg MCH) the percent-response of sGeff was 2.0 times more pronounced than with FEV1, 1.3 times more than with FEF50 resp. By nature the percent-response in non-asthmatic subjects was 1.5 to 2.9 times lower than in asthmatic patients, but the differences in percent-responses between the lung function parameters were even more pronounced.

Fig. 2
figure 2

Dose–response curves of methacholine at each consecutive provocation levels (PD1: 0.2 mg; PD2: 1.0 mg; PD3: 2.2 mg methacholine and reversibility to a ß2-agonist) referred for effective specific conductance (sGeff), forced expiratory volume in one second (FEV1) and forced expiratory flow at 50% vital capacity (FEF50) in relation to the PD-level −20% (PD-20) for FEV1 and FEF50, PD-level −40% (PD-40) for sGeff resp., regarding asthmatic patients (A = in comparison with non-asthmatic subjects (non-A)

Diagnostic accuracy of assessing AHR by using different lung function parameters

All indices contributing to the assessment of diagnostic accuracy (Additional file 4: Statistical Approach) are summarized in Table 2, differentiating between asthmatic patients and non-asthmatic subjects. Since the cut-off level for sGeff is not yet clearly defined, we first compared the 3 potentially valuable thresholds PD−40sGeff, PD−45sGeff, and PD−50sGeff. Optimal determination as represented by the highest sensitivity, highest negative predictive values, highest diagnostic odds ratios (DOR) [43], lowest negative likelihood ratios, and highest diagnostic effectiveness was found for PD−40sGeff, indicating that PD−40sGeff is the most appropriate threshold for evaluating AHR by sGeff. Whereas the PD−40sGeff level was reached in all subjects by sGeff, the PD−20FEV1 was not reached by FEV1 in 9.1% of subjects and in 6.4% PD−20FEF50 by FEF50 respectively. Therefore, PD−40sGeff measures AHR in comparison with PD−20FEV1 and PD−20FEF50 with better properties of diagnostic accuracy.

Table 2 Measures of MCH-challenge procedures computed for response characteristics of sGeff, FEV1 and FEF50

Test-duration and MCH-doses to achieve PD-levels

We found it of clinical relevance to compare the test duration and MCH-doses to achieve PD-levels by the different target lung function parameters. Table 3 shows that the test duration was significantly shorter for PD−40sGeff compared to PD−20FEV1, or PD−20FEF50 (14:45 ± 5:54 min. versus 17:46 ± 5:16 min; 16:17 ± 5:36 min respectively). Consequently, the provocation doses to which subjects are exposed to MCH were significantly lower for PD−40sGeff compared to PD−20FEV1, or PD−20FEF50 (0.495 ± 0.491 mg versus 0.739 ± 0.615 mg; 0.625 ± 0.588, respectively).

Table 3 Test-duration and provocation-doses of methacholine (mg) needed to achieve the different provocation-dose levels

Influence of pulmonary hyperinflation prior to or during MCT

Changes of end-expiratory lung volume (EELV) during MCT cannot be assessed by spirometry, especially not a shift of the flow-volume curve towards total lung capacity in consequence of dynamic hyperinflation, which results in an elevated residual volume and a decrease of inspiratory capacity and vital capacity as well. However, a shift of EELV toward total lung capacity may have an influence on the magnitude of FEV1. In 63 of 337 asthmatic patients (18.7%) pulmonary hyperinflation was observed prior to testing, or was developed during MCT. In Fig. 3 response to MCH is compared in all subjects differentiating those with pulmonary hyperinflation which those without pulmonary hyperinflation. There was a much more pronounced AHR in patients with pulmonary hyperinflation (mean diff in PD−40sGeff: 7.5%, n.s.; PD−20FEV1: 5.8% p < 0.001; PD−20FEF50: 7.6%, n.s.). Noteworthy to realise that pulmonary hyperinflation influenced AHR measured by PD−20FEV1 significantly. Moreover interestingly, differences were not significant for PD−40sGeff. This may be a formal proof, that the assessment of AHR using the integral method of sGeff, evaluates changes of airway mechanics concomitantly with changes of EELV during MCT.

Fig. 3
figure 3

Response to methacholine in asthmatic patients comparing those with initial or developed pulmonary hyperinflation (n = 78;16.1%) with those without pulmonary hyperinflation (n = 404;83.9%), assessed by forced expiratory volume in one second (FEV1) and forced expiratory flow at 50% vital capacity (FEF50) and the effective specific conductance sGeff

Receiver operating characteristics

Receiver operating curves (ROC) describing the relationship between sensitivity and specificity of percent changes during MCT of the 3 lung function parameters are given in Fig. 4. The area under the curve (AUC) for absolute change of sGeff yielded the greatest value for PD−40sGeff (.859; 95% CI: .857-.897) compared to PD−45sGeff and PD−50sGeff on one hand, and compared to PD−20FEV1 (.749; 95% CI: .705-.793) and PD−20FEF50 (.729;) 5% CI: .680-.778) on the other hand. Considerable differences are also found regarding the Youden J-index and the likelihood ratios, both estimates of diagnostic test performance.

Fig. 4
figure 4

Receiver operating curves (ROC) describing the relationship between sensitivity and specificity of percent changes obtained at the corresponding provocation dose levels, PD-40sGeff for a 40% fall in effective specific conductance, PD-20FEV1 for a 20% fall in forced expiratory volume in one second, and PD-20FEF50 for a 20% fall in forced expiratory flow at 50% vital capacity during methacholine challenge represented by methacholine

Discussion

Key findings

The principal goal of our study was to demonstrate that the performance of MCTs using the plethysmographic technique in addition to the spirometric approach offers some fundamental benefits, such as independence from deep inspiration, and hence, modulation of the airway calibre [10, 3136], avoidance of forced expiratory manoeuvres, and hence the subject’s cooperation and coordination, which is known to change the airway responsiveness during the test procedure [34, 37, 38]. There are two important denouements, which are achieved, if the spirometric assessment of AHR is combined with whole-body plethysmography. First, MCTs based on a plethysmographic approach offers improved diagnostic accuracy. Second, the broncho-provocating process by MCH in relation to the development of pulmonary hyperinflation and/or the phenomenon of dysanapsis, the ratio FEF25–75/FVC thought to be a surrogate measure of airway size relative to lung size significantly associated with AHR [44]. The present study positively confirms the findings of Nensa et al. [10] reinforcing sReff, and its reciprocal value sGeff respectively, as the most useful target parameters in the detection of AHR. Due to our observations of much higher percent-response to MCH by PD−40sGeff than by PD−20FEV1, we suggest, that some false negative PD−20FEV1 tests resulted in an underestimation of the severity of AHR, and hence potentially missed the diagnosis of asthma. In the following discussion we would like to focus on some important aspects of MCT performed by whole-body plethysmography, if combined with the spirometric assessment. Furthermore, we evaluate the physiological aspects of the so-called dysanapsis [13, 14], which was found to be an important determinant of AHR to MCH, and hence airway narrowing during MCT by different author-groups [1517, 45].

Diagnostic accuracy

There is an increasing interest to define proximity of measurement results comparing different test procedures. Recently Porpodis et al. [46] compared the diagnostic validity of MCH with mannitol. Based on their spirometric PD−20FEV1-results they found that both challenge tests were equivalent in diagnosing asthma. In the present study, however, we first look at the accuracy of different target lung function parameters, which could be discriminative between “asthma” and “non-asthma”.

The most remarkable finding of the present study is that the sGeff with a provocation level of PD−40presented the highest diagnostic accuracy (sensitivity, specificity likelihood ratio, diagnostic odds ratios, ROC curves) for the diagnosis of asthma. In comparison with PD−20FEV1 and PD−20FEF50 the response-pattern to MCH evaluated by PD−40sGeff was consistently different. Identical agreement regarding severity of AHR was only found in 159 (32.9%) cases (44 severe AHR, 50 moderate AHR, 19 low AHR, 46 no AHR). In comparison to PD−40sGeff, AHR-responses of PD−20FEV1 were observed in 83 (24.9%) cases only one PD-level later, in another 73 (21.9%) cases only two PD-levels later, and in 20 (6.0%) cases even only three PD-levels later.

Furthermore, in contrast to PD−40sGeff, the level of PD−20FEV1was not reached in 65 cases (13.5%). It follows that in 67.1% of asthmatic patients, a disagreement regarding MCH-response between PD−40sGeff and PD−20FEV1 was found. Since the diagnosis of asthma was primarily based on the results of AHR evaluated by PD−20FEV1 the question remains open whether or not some asthmatics are hidden in the collective of patients diagnosed as chronic cough disease. This question, however, can only be answered by a properly planned prospective study.

Deep inspiration during MCT

It has well been recognised that deep inspirations play a major role in modulating airway calibre and airway responsiveness. In healthy subjects, deep inspirations reduce the level of pharmacologically induced airway obstruction by bronchodilation [45], whereas prohibition of taking deep breaths enhances the reaction to a broncho-constrictor agent [33]. Moreover, it has been recognised that inhalations to total lung capacity are broncho-protective, particularly in subjects with borderline AHR and non-asthmatics [34]. However, it has also been shown, that in asthmatics with mild AHR, deep inhalations do not produce a significantly lower response to MCH [37, 47] than tidal breathing [40].

Site of action and interrelationship with dysanapsis

For many years it has been recognised that one factor determining the presence of AHR to broncho-provocative agents such as MCH is airway size. It varies from one individual to another, and some of this variability cannot be explained by differences in lung size between individuals [15, 16]. The term “dysanapsis” was initially proposed by Green et al. [13] to describe this disproportionate, but physiologically normal and gender-specific growth between airways and lung parenchyma. The wide variation in maximal expiratory flow rates between individuals with similar lung size was interpreted to mean that there is no consistent association between lung and airway size. Mead [14] determined the association between airway size and lung size in adult women and men. He found that healthy adult men have airways that are larger in diameter than the airways of women. Moreover, it was concluded that women and boys have airways that are smaller relative to lung size compared to men, and therefore, the apparent gender-based differences occur late in the growth period. These findings were confirmed by Shell et al. [48], demonstrating evidence for gender-specific dysanapsis as shown by computer tomography (CT) imaging. There are significant male–female differences in the luminal areas of the larger and central airways, which are not accounted for by differences in lung size.

Forced expiratory flow between 25 and 75% of vital capacity (FEF25–75) is considered as the most effort-independent part of the flow volume curve. This fraction is very sensitive regarding flow-limitation in peripheral airways, when chronic airflow obstruction is present [13]. Since direct measures of airway mechanics and lung size are often not available concomitantly, and FEF25–75 and forced vital capacity (FVC) are measured routinely during forced breathing manoeuvres, the ratio between FEF25–75 and FVC has been taken as a surrogate measure of dysanapsis [7, 13, 14, 17, 44]. Parker et al. [17] showed that when subjects were classified into four groups according to their FEF25–75/FVC ratio, subjects with lowest ratio also had the lowest PD−20FEV1, supporting the notion that subjects who are more sensitive to MCH have smaller airway calibres in relation to their lung size. They concluded that baseline FEF25–75/FVC ratio could be a determinant of AHR to MCH. Moreover, Mirsadraee et al. [7] showed that the diagnostic accuracy of PC−20FEF25–75/FVC and PD−20FEF50 were similar to PD−35sGaw and superior to PD−20FEV1.

The analysis in a subgroup of our asthmatic patients (n = 363) regarding dysanapsis expressing the FEF25–75/FVC ratio as age- and gender-specific quartiles showed significant association to the severity of AHR assessed by PD−40sGeff, PD−20FEV1, and PD−20FEF50. However, within these quartiles of dysanapsis the response-pattern of the 3 lung function parameters were completely different. The differences between severe and medium AHR can be much better presented by PD−40sGeff (36.7%, 46.7%, resp.) than by PD−20FEV1 (8.2% 23.3%, resp.), or by PD−20FEF50 (23.2%, 34.9%, resp.), if related to the phenomenon of “airway narrowing” by FEF25–75/FVC ratios < 25%ile, 25 to 50%ile. These associations existed for both male and female asthmatics, indicating that as opposed to Gaw and sGaw, AHR assessed by sGeff represents a much wider range of central and peripheral airways within the bronchial tree. In line with Parker et al. [17] we therefore conclude that AHR assessed by PD−40sGeff is in strong association with dysanapsis, and may well provide a new conceptual approach, and hence better understanding of mechanisms that predispose an individual to asthma. Within the group of non-asthmatic subjects (n = 148) 14 (9.5%) (8 males, 6 females) presented with a PD−40sGeff to be labelled as “severe”, within the subgroup of dysanapsis with a ratio of FEF25–75/FVC < 25%ile, and the question remains open whether in these patients the diagnosis “asthma” was missed, because the diagnosis “asthma” was only expressed when PD−20FEV1 demonstrated AHR.

Limitations of method

The present study has some limitations. Firstly, because this was a retrospective study, there is a potential risk that some asthmatic patients might have been under-diagnosed. As standard for the diagnosis “asthma” PD−20FEV1was taken and it could well be that a certain number of patients with cough-related diseases should be attributed to the groups of “asthmatics”, if in fact PD−40sGeffindicated a moderate or high AHR, and the subject presented with a low FEF25–75/FVC ratio. Secondly, the database from which the evaluation was performed offered only a small number of really healthy subjects. It is not totally excluded that some “potential false positive” responses in the non-asthmatic group might have been more prevalent compared to healthy controls, and the differences between the asthmatics and false positive responders might be smaller than with healthy subjects. Thirdly, usually a retrospective study looks backwards and examines exposures to suspected risks or protection factors in relation to an outcome that is established at the start of the study. In our case–control study, however, we evaluated individual lung functions within 2 diagnostic groups, and for each parameter the chance to obtain a comprehensive result regarding AHR was the same. Great care was taken to avoid biases for one or the other parameter and confounding factors such as the development of pulmonary hyperinflation or dysanapsis were carefully addressed.

Conclusion

Important findings in regard to the performance of MCT as a hallmark in the diagnosis of bronchial asthma could be discovered, if the spirometric approach is combined with the assessment by whole-body plethysmography. Where diagnostic accuracy is required, there are striking differences regarding the choice of target parameters. Although the spirometric assessment based on changes in FEV1 is regarded as the gold standard for evaluation of MCT, this study shows some important physiological inadequacies, to be considered as prerequisites, which could well be important for diagnosing “asthma”. Deep inspiration in the set-up of test-sequences should be carefully monitored, as it is inevitably linked to forced breathing manoeuvres. Using the APS technology within the Jaeger MasterLab plethysmograph, MCTs are by default based on plethysmographic tidal breathing analysis, applying breathing loops with an implementation of real-time multi-level integral computation of all data points throughout the whole breathing cycle. In this light the study demonstrates that the plethysmographic sGeff features a highly sensitive and reproducible target parameter for the assessment of AHR by MCTs. Furthermore, the combined assessment of AHR by spirometry and whole-body plethysmography offers the possibility to include changes of EELV at FRCpleth (and hence pulmonary hyperinflation), as well as the phenomenon of dysanapsis to be included into the differentiation between “asthmatic” and “non-asthmatics”.