Abstract
Background
The aim of this study was examining the accuracy of accelerometer-based portable navigation systems (HipAlign) when measuring leg length changes using two-dimensional (2D) and three-dimensional (3D) methods.
Methods
Inclusion criteria were patients ≥ 20 years old with symptomatic hip disease who underwent primary total hip arthroplasty (THA) in the supine position using HipAlign between June 2019 and April 2020. The exclusion criteria were patients who underwent THA via a posterior approach. We examined correlations between the leg length change measurement with HipAlign and either 2D or 3D measurement. We performed a multivariate analysis to determine which factors may have influenced the absolute error results.
Results
This study included 34 patients. The absolute error in leg length change between the HipAlign and 3D measurement (4.0 mm) was greater than the HipAlign and 2D measurement (1.7 mm). There were positive correlations between leg length change with HipAlign and 2D and 3D measurements. Male patients had larger errors with 2D measurement. No significant factors were identified for 3D measurement.
Conclusion
HipAlign provided acceptable measurement accuracy for leg length changes.
Similar content being viewed by others
Background
Leg length discrepancy is a common concern following total hip arthroplasty (THA). Furthermore, it remains one of the most common reasons for litigation against the orthopedic community in the US [1]. To equalize leg length, surgeons perform preoperative planning using radiological templates or intraoperative measurement methods, including calipers, radiographs, and navigation systems [2,3,4,5,6].
Navigation systems have been used for accurate cup placement to reduce complications from cup malpositioning, including impingement and dislocation [7,8,9]. Using navigation systems, surgeons can also detect a change in leg length. Computed tomography (CT)-based navigation can be used; however, the associated high cost and radiation exposure are problematic [10]. Large console image-free navigation does not require preoperative CT but is expensive [11]. In recent years, portable navigation systems have been developed, which are lower in cost. The accuracy of acetabular cup placement using a portable navigation system (HipAlign, OrthAlign, Aliso Viejo, CA, USA) has been studied. Previous studies demonstrated improved accuracy of cup inclination and anteversion using navigation systems [12,13,14,15,16]. Only two studies have demonstrated the accuracy of leg length change measurements using HipAlign [17, 18]. These studies examined measurement accuracy using 2D radiographs. To the best of our knowledge, there are no studies that examine measurement accuracy using 3D evaluation. The purpose of this study was to examine the accuracy of leg length change measurements by the portable navigation system (HipAlign) using 3D evaluation. Comparisons were also made between the 2D and 3D evaluations. To ensure accurate 2D and 3D measurements, we previously performed a reliability study using the intra-class and inter-class correlation coefficients [19, 20]. The intra-class and inter-class correlation coefficients for the 2D measurements were 0.98 and 0.92, respectively. The intra-class and inter-class correlation coefficients for the 3D measurements were 0.97 and 0.94, respectively [19].
We hypothesized that 2D evaluation would show better accuracy for leg length change using HipAlign.
Materials and methods
Patients
Inclusion criteria were patients ≥ 20 years old with symptomatic hip disease who underwent primary THA in the supine position under general anesthesia using HipAlign between June 2019 and April 2020. The exclusion criteria were patients who underwent THA via a posterior approach, such as hips with a high degree of dislocation (Crowe groups 3 and 4) [21] requiring subtrochanteric osteotomy and revision hip arthroplasty.
All hips were implanted with cementless components by the same surgeon (MH). The hips were exposed via a direct anterior approach (DAA) on a traction table or with a modified Watson-Jones approach (anterolateral supine approach: ALS), with the patient in a supine position. Twenty-six patients were treated using DAA, and the remaining eight patients were treated using ALS. ALS was selected in cases with severe deformity or excessive anteversion of the femoral neck. Two pins, 4.5 mm in diameter, were placed at the iliac crest, and the navigation unit was fixed to these pins. The location of the bilateral anterior superior iliac spine (ASIS) and pubic symphysis were recorded. HipAlign can measure the leg length change using a femoral registration screw, probe, laser module, and thigh plate for a vertical laser target (Fig. 1). Prior to neck osteotomy, a small reference screw was placed in the proximal femur for femoral registration. However, it can be difficult to fix this screw to the proximal femur in patients with DAA or ALS. A hole was made using a 2.0 mm diameter Kirschner wire at the proximal femur, and a registration probe was placed in this hole (Fig. 2). For intraoperative leg length measurement, the laser module was attached to the bracket of the navigation unit and the thigh plate was placed on the distal anterior thigh. A vertical laser target was attached to the thigh plate. The limb was positioned in the neutral reference position, and the laser projection on the vertical target was traced using a marking pen [18]. The surgeon realigned the vertical laser target to the laser projection and registered the marker (Fig. 1).
Leg length change was defined as the change in length measured before the neck cut and after the reduction with the same six hip directions (flexion/extension, abduction/adduction, and external rotation/internal rotation). Surgeons can detect 2D leg length changes on the monitor (Fig. 3). A G7 PPS Finned BoneMaster Limited Hole Shell (Zimmer Biomet, Warsaw, IN, USA) was used. Regarding stems, Taperloc Microplasty stems (Zimmer Biomet, Warsaw, IN) or AMIStem (Medacta, Castel San Pietro, Switzerland) were used. CT was performed preoperatively from the pelvis to knee joint. Neck length was determined according to preoperative planning using 3D software (ZedHip, LEXI Co. Ltd, Tokyo, Japan). If instability after trial reduction was found, the surgeons selected a longer neck than was decided in preoperative planning.
Evaluation
The anteroposterior radiographs of the hips were performed in the supine position preoperatively and postoperatively. The leg length discrepancy on 2D measurement was measured as the distance between the horizontal line connecting the bilateral tear drops and the medial apex of the lesser trochanter. Leg length change on 2D measurement was defined as the difference between the preoperative and postoperative leg length. The leg length change on 2D measurement was subtracted from the leg length discrepancy pre-and -post-surgery. CT was performed from the pelvis to the knee preoperatively and 2 weeks postoperatively. The ZedHip was used to examine pre- and postoperative leg length discrepancies and compared with 3D measurement, which was assessed as the distance from the ASIS to the intercondylar fossa of the femur. Pre- and 2 weeks postoperative CT digital images were superimposed using the ZedHip. The leg length change measured by ZedHip was subtracted from the leg length discrepancy pre-and -post-surgery. To evaluate the accuracy of HipAlign based on 2D and 3D measurements, intraoperative leg length change was compared with 2D and 3D measurements [19]. Hip measurements with absolute error values within 3, 5, and 10 mm were determined using 2D and 3D measurements. All patients were followed up post-THA and examined for potential complications.
Statistical analysis
Power analysis was performed, and a minimum of 27 patients were required to provide appropriate power (β = 0.80) at a significance level of 0.05. The absolute errors in leg length change were compared between the 2D and 3D measurements using the Wilcoxon signed-rank test. Correlation analyses were performed using Spearman’s rank correlation test. The correlations between leg length change measured with HipAlign and 2D or 3D measurements were evaluated. The correlation between leg length change measured using 2D and 3D measurements were determined. Multivariate analysis was performed to determine the factors affecting absolute errors, including age (≥ 65 years and < 65 years), sex, BMI (≥ 25 kg/m2 and < 25 kg/m2), approach, and Crowe group. P-values of < 0.05 were considered statistically significant. Data were statistically analyzed using the EZR software program, version 1.61 (Saitama Medical Center, Jichi Medical University, Saitama, Japan) [22].
Results
This study included 34 patients. There were 3 men and 31 women, with a mean age of 66.9 years (range, 48–82 years) and mean body mass index (BMI) of 23.4 kg/m2 (range, 18.9–28.2 kg/m2). The preoperative diagnosis for all patients was osteoarthritis of the hip. Thirty hips were classified as Crowe group 1 [21]. Four hips were classified as Crowe group 2 [21].
The absolute error for leg length change between the HipAlign measurement and 2D measurement was 1.71 ± 1.92 mm (range, 0–5 mm). These errors were within 3 mm for 79% of patients (27 of 34 patients), and the remaining patients had errors within 5 mm.
The absolute error in leg length change between HipAlign measurement and 3D measurement was 4.03 ± 3.51 mm (range, 0.4–17.7 mm). These errors were within 3 mm for 53% of patients (18 patients), within 5 mm for 82% of patients (28 patients), and within 10 mm for 94% of patients (32 patients).
The errors in leg length change for the 3D measurements were greater than those for the 2D measurements (p < 0.01). Positive correlations of leg length changes were found between HipAlign and 2D (r = 0.84, p < 0.01, Fig. 4A) and 3D measurements (r = 0.56, p < 0.01, Fig. 4B). Comparing the 2D and 3D measurements, there was a positive correlation with leg length change (r = 0.52, p < 0.01, Fig. 4C). Multivariate analysis demonstrated that sex was a significant contributing factor to 2D measurement errors (p < 0.01). Male patients demonstrated larger 2D errors. There were no significant factors for the 3D measurements (Table 1).
There were no early postoperative complications, including fracture, infection, and nerve injury.
Discussion
The most significant finding of this study was that the absolute error for leg length change was greater for HipAlign evaluated with 3D measurement than 2D measurement. This confirmed our hypothesis that 2D measurement is a more accurate evaluation of leg length change when using HipAlign. An explanation for the smaller error in 2D measurement than that in 3D measurement is that HipAlign shows 2D leg length change. There was a positive correlation for leg length change between 2 and 3D measurements.
An acceptable measure of leg length discrepancy has yet to be defined. However, several studies have shown that a leg length discrepancy of up to 10 mm is well tolerated by most patients [4, 23]. Some patients, especially young patients, have complained of minor discrepancies.
Table 2 shows the results of previous reports on the accuracy of leg length measurements using navigation systems. Ogawa et al. [5] reported an absolute error of 2.4 mm with CT-based navigation. Absolute errors with portable navigation (HipAlign) were reported to be 2.3 and 3.1 mm in studies by Tanino et al. [18] and Anjiki et al. [17], respectively. The 2D errors within 5 mm were 81% and 100% in the present study and the study by Anjiki et al. [17], respectively. The accuracy of the leg length measurement in the present study was equivalent to that of previous studies, including large console navigation systems.
A small reference screw is often difficult to fix to the proximal femur in patients with DAA or ALS. In our experience, the screw was often loose. The screw was not required if we could register the same point before the neck cut and after implantation. We utilized an easy technique whereby we created a small hole in the proximal femur and performed registration at this hole. Assessing intraoperative leg length change using navigation is based on the measurement of the distance between fixed points on the pelvis and the femur. However, these techniques are unreliable as they require accurate femur repositioning. Five degrees of abduction/adduction malpositioning may lead to leg length errors of up to 8 mm [24]. Using a laser enables surgeons to reposition the femur in HipAlign accurately.
Male patients in our study had larger errors when 2D measurements were used. As men generally have larger muscle volumes than women, this could explain the increase in error value; placing the registration probe on the hole was sometimes difficult without retracting the gluteus medius, gluteus minimus, and tensor fascia latae muscle in ALS and tensor fascia latae muscle and rectus femoris in DAA. Errors in leg length measurement can occur with 2D measurements when the hips have flexion contracture or are in the pelvic position. The 2D measurement was defined as the distance between the horizontal line connecting bilateral tear drops and the medial apex of the lesser trochanter. The 3D measurement was defined as the distance from the ASIS to the intercondylar fossa of the femur [19]. The importance of 3D measurement was reported in femoral offset [25]; however, the usefulness of 3D measurement in leg length change was not confirmed by comparison with 2D measurement [19]. Regarding leg length change, 2D evaluation alone is useful if accurate radiographs can be obtained [19]. Although 3D measurement of femoral offset is clinically important, 3D measurement of leg length change did not have a clinically important role.
The limitations of this study include the small sample size, the use of a new registration method, no assessment of the offset, and the use of CT. The offset is an important factor requiring intraoperative assessment, however, HipAlign cannot determine the offset intraoperatively. CT has several drawbacks, including radiation exposure and cost [26]. Recently, low-dose CT has been used for preoperative planning and postoperative assessment of THA [27].
Conclusions
The absolute error in leg length change for the HipAlign and 3D measurement (4.03 mm) was greater than the HipAlign and 2D measurement (1.71 mm). Positive correlations between leg length changes were found between HipAlign and 2D and 3D measurements. HipAlign used in the supine position provided acceptable accuracy for leg length change measurement. 3D measurement is not recommended for the evaluation of leg length change after THA using HipAlign.
Availability of data and materials
The datasets during and/or analyzed during the current study are available from the corresponding author on reasonable request.
Abbreviations
- 2D:
-
Two-dimensional
- 3D:
-
Three-dimensional
- THA:
-
Total hip arthroplasty
- CT:
-
Computed tomography
- BMI:
-
Body mass index
- DAA:
-
Direct anterior approach
- ALS:
-
Anterolateral supine approach
- ASIS:
-
Anterior superior iliac spine
References
Hofmann AA, Skrzynski MC. Leg-length inequality and nerve palsy in total hip arthroplasty: a lawyer awaits! Orthopedics. 2000;23(9):943–4.
Murphy SB, Ecker TM. Evaluation of a new leg length measurement algorithm in hip arthroplasty. Clin Orthop Relat Res. 2007;463:85–9.
Kitada M, Nakamura N, Iwana D, Kakimoto A, Nishii T, Sugano N. Evaluation of the accuracy of computed tomography-based navigation for femoral stem orientation and leg length discrepancy. J Arthroplasty. 2011;26(5):674–9.
Desai AS, Dramis A, Board TN. Leg length discrepancy after total hip arthroplasty: a review of literature. Curr Rev Musculoskelet Med. 2013;6(4):336–41.
Ogawa K, Kabata T, Maeda T, Kajino Y, Tsuchiya H. Accurate leg length measurement in total hip arthroplasty: a comparison of computer navigation and a simple manual measurement device. Clin Orthop Surg. 2014;6(2):153–8.
Renkawitz T, Sendtner E, Schuster T, Weber M, Grifka J, Woerner M. Femoral pinless length and offset measurements during computer-assisted, minimally invasive total hip arthroplasty. J Arthroplasty. 2014;29(5):1021–5.
McCarthy TF, Alipit V, Nevelos J, Elmallah RK, Mont MA. Acetabular Cup Anteversion and Inclination in Hip Range of Motion to Impingement. J Arthroplasty. 2016;31(9 Suppl):264–8.
Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60-A:217–20.
Seagrave KG, Troelsen A, Malchau H, Husted H, Gromov K. Acetabular cup position and risk of dislocation in primary total hip arthroplasty. Acta Orthop. 2017;88(1):10–7.
Beckmann J, Stengel D, Tingart M, Götz J, Grifka J, Lüring C. Navigated cup implantation in hip arthroplasty. Acta Orthop. 2009;80(5):538–44.
Snijders T, van Gaalen SM, de Gast A. Precision and accuracy of imageless navigation versus freehand implantation of total hip arthroplasty: A systematic review and meta-analysis. Int J Med Robot. 2017;13(4). https://doi.org/10.1002/rcs.1843.
Kamenaga T, Hayashi S, Hashimoto S, Matsumoto T, Takayama K, Fujishiro T, Hiranaka T, Niikura T, Kuroda R. Accuracy of cup orientation and learning curve of the accelerometer-based portable navigation system for total hip arthroplasty in the supine position. J Orthop Surg (Hong Kong). 2019;27(2):2309499019848871.
Takada R, Jinno T, Miyatake K, Hirao M, Yoshii T, Okawa A. Portable imageless navigation system and surgeon’s estimate for accurate evaluation of acetabular cup orientation during total hip arthroplasty in supine position. Eur J Orthop Surg Traumatol. 2020;30(4):707–12.
Tanino H, Nishida Y, Mitsutake R, Ito H. Portable Accelerometer-Based Navigation System for Cup Placement of Total Hip Arthroplasty: A Prospective, Randomized. Controlled Study J Arthroplasty. 2020;35(1):172–7.
Okamoto M, Kawasaki M, Okura T, Ochiai S, Yokoi H. Comparison of accuracy of cup position using portable navigation versus alignment guide in total hip arthroplasty in supine position. Hip Int. 2021;31(4):492–9.
Hasegawa M, Naito Y, Tone S, Wakabayashi H, Sudo A. Accuracy of acetabular cup insertion in an anterolateral supine approach using an accelerometer-based portable navigation system. J Artif Organs. 2021;24(1):82–9.
Anjiki K, Kamenaga T, Hayashi S, Hashimoto S, Kuroda Y, Nakano N, Fujishiro T, Hiranaka T, Niikura T, Kuroda R, Matsumoto T. Effectiveness of an accelerometer-based portable navigation for intraoperative adjustment of leg length discrepancy in total hip arthroplasty in the supine position. J Orthop Sci. 2022;27(1):169–75.
Tanino H, Nishida Y, Mitsutake R, Ito H. Accuracy of a portable accelerometer-based navigation system for cup placement and intraoperative leg length measurement in total hip arthroplasty: a cross-sectional study. BMC Musculoskelet Disord. 2021;22(1):299.
Tone S, Hasegawa M, Naito Y, Wakabayashi H, Sudo A. Accuracy of image-free navigation in intraoperative leg length change from total hip arthroplasty using evaluations from 2D and 3D measurements. BMC Musculoskelet Disord. 2021;22(1):1021.
Montgomery AA, Graham A, Evans PH, Fahey T. Inter-rater agreement in the scoring of abstracts submitted to a primary care research conference. BMC Health Serv Res. 2002;2(1):8.
Crowe JF, Mani VJ, Ranawat CS. Total hip replacement in congenital dislocation and dysplasia of the hip. J Bone Joint Surg Am. 1979;61(1):15–23.
Kanda Y. Investigation of the freely available easy-to-use software “EZR” for medical statistics. Bone Marrow Transplant. 2013;48(3):452–8.
White TO, Dougall TW. Arthroplasty of the hip. Leg length is not important. J Bone Joint Surg Br. 2002;84(3):335–8.
Sarin VK, Pratt WR, Bradley GW. Accurate femur repositioning is critical during intraoperative total hip arthroplasty length and offset assessment. J Arthroplasty. 2005;20(7):887–91.
Tone S, Hasegawa M, Naito Y, Wakabayashi H, Sudo A. Comparison between two- and three-dimensional methods for offset measurements after total hip arthroplasty. Sci Rep. 2022;12(1):12644.
Brenner DJ, Hall EJ. Computed tomography–an increasing source of radiation exposure. N Engl J Med. 2007;357(22):2277–84.
Su AW, Hillen TJ, Eutsler EP, Bedi A, Ross JR, Larson CM, Clohisy JC, Nepple JJ. Low-Dose Computed Tomography Reduces Radiation Exposure by 90% Compared With Traditional Computed Tomography Among Patients Undergoing Hip-Preservation Surgery. Arthroscopy. 2019;35(5):1385–92.
Acknowledgements
Not applicable.
Funding
Not applicable.
Author information
Authors and Affiliations
Contributions
M.H. was the main investigator and wrote the manuscript. Y.N. and S.T. helped with data analysis. A.S. helped with the interpretation of the data and results. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study was approved by the institutional review board of Mie University (H2018-083), and all patients completed written informed consent prior to participation. All procedures were performed in accordance with the principles of the Declaration of Helsinki.
Consent for publication
Not applicable.
Competing interests
M.H. is a member of the editorial board (Associate Editor) of this journal. Other authors declare that they have no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Hasegawa, M., Naito, Y., Tone, S. et al. Two- and three-dimensional measurements of leg length change using an accelerometer-based portable navigation system in total hip arthroplasty. BMC Musculoskelet Disord 24, 872 (2023). https://doi.org/10.1186/s12891-023-07006-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12891-023-07006-4