Background

Chronic obstructive pulmonary disease (COPD) is characterized by persistent and not fully reversible airflow limitation, with high economic and social burden, high morbidity and mortality worldwide [1,2,3,4]. Recently, Pre-COPD highlighted by the Global Initiative for Chronic Obstructive Lung Disease (GOLD) strategy indicated that there are patients with normal lung function, but with structural abnormalities such as emphysema [2]. Emphysema is characterized by air trapping and dynamic hyperinflation (DH) during exercise [5]. DH is implicated as a major cause of dyspnea during physical activity in COPD, which might cause limited exercise capacity [6, 7]. Exercise is a cornerstone of pulmonary rehabilitation for patients with COPD [8], however, the presence of DH during exercise makes it difficult for patients to tolerate higher training intensities [9].

Therefore, reversing DH is important for patients with COPD or Pre-COPD to reduce dyspnea and improve exercise capacity [10]. Pursed lip breathing (PLB) could theoretically provide positive expiratory pressure (PEP), reduce dynamic compression during expiration, attenuate expiratory airflow restriction, and reduce inspiratory threshold load in hyperinflated lungs in patients with COPD [11]. Thus, PLB has been recommended to reduce dyspnea [12]. However, the increase in airway pressure is due to the increased mouth expiratory resistance, which varies with the rate of airflow, is unstable and cannot be quantified [13]. Therefore, the evidence that PLB is beneficial for dyspnea, exercise endurance, and dynamic hyperinflation remains uncertain. Non-invasive ventilation (NIV) has been used as an aid to improve exercise tolerance in patients with COPD [14]. Among NIV modalities, PEP performs the major role in reversing DH, which can be offered by less costly and elementary devices. The PEP, which is believed to have similarities with PLB, could increase intraluminal airway pressure, counteract intrinsic positive end-expiratory pressure (PEEPi) and reduce airway collapse during expiration by moving the airway equal pressure point from the peripheral airway, which is prone to collapse, to a more solid airway in the center. Thus, PEP could reduce airway collapse and DH, and thus possibly reduce dyspnea and improve exercise capacity [15].

PEP devices have been wildly used to produce a reduction of breathing frequency and expiratory flow limitation, change breathing pattern, improve gas exchange, as well as result in less airway collapse and air-trapping in patients with COPD [16,17,18,19,20,21]. Russo D et al. had subjects breathe using different levels of PEP (1 cmH2O and 10 cmH2O) during 6MWT and found that PEP could improve exercise tolerance [18]. Nicolini A et al. had subjects breathe using a simple device with 5 cmH2O PEP in 6MWT and found that PEP significantly improved exercise capacity (walking distance), oxygen saturation and heart rate [16]. Cardoso et al. had subjects breathe with PEP device for 20 min and showed that patients had a better physiological breathing pattern with PEP, which was effective in reducing assisted inspiratory muscle activity and improving ventilation and dyspnea [17]. Therefore, previous studies basically used PEP for breathing only during exercise, a brief process.

However, to our knowledge, few studies to date have explored the effects of long-term domiciliary use of PEP devices on exercise capacity in patients with COPD or Pre-COPD. Therefore, this study aimed to evaluate the effects and safety of long-term domiciliary use of PEP delivered by a spring-loaded resistor face mask in patients with COPD or Pre-COPD.

Methods

Study subjects

Subjects were consecutive screened from the Pulmonary Clinic of a tertiary care teaching hospital from July 2021 to May 2022. Inclusion criteria: (I) Age range: 18 to 80 years; (II) emphysema on CT scan (HU ≥ − 900); (III) no other imaging changes of lung disease such as occupancy, exudation and interstitial changes on CT scan; (IV) no history of lung cancer, lung resection, cystic fibrosis; (V) signed informed consent. Exclusion criteria: (I) with underlying lung disease such as asthma, bronchiectasis and interstitial lung disease; (II) with heart failure; (III) unwilling to participate. All participants were clinically stable at the time of recruitment. All subjects have not recently, or were currently, participating in pulmonary rehabilitation during the time of the study. Informed consent was obtained from all subjects.

Study procedures

All enrolled subjects were randomized into intervention group (with PEP device) or control group (with sham-PEP device) by a sequence of computer-generated randomized numbers. An independent third-party researcher was responsible for generating the allocation sequence (i.e., computer-generated random numbers), and the outcome assessors were unaware of patients’ allocation.

The PEP device was a silicone face mask (PEP device, Xiamen Kangbo Medical facility Company) loaded with a spring linear pressure resistor in a unidirectional expiratory valve and was placed on the subject’s face and held in place with a 4-point strap around the head. The spring linear pressure resistor can be adjustable to provide flow-independent PEP between 5 and 20 cmH2O (Fig. 1). During the inspiratory period, the tiny valve opened to lower the inspiratory pressure. In this trial, the resistor was adjusted to perform a load around 5 cmH2O and was kept the same for the duration of the study, based on previous studies that used these values and encountered some clinical benefits in patients with COPD during exercise[15, 16]. All subjects needed to do was to put on the mask, inhale through the nose and exhale through the nose. But the subjects needed to make sure that the mask was airtight and did not leak. The sham PEP device was not loaded with a spring linear pressure resistor; thus it did not provide PEP. Subjects were advised to use the PEP device at any time of the day, either continuously or intermittently. Subjects were required to ensure a total of 4 h of use per day for a total of two months and were advised to use the device while sitting or walking slowly, and not during strenuous activities such as running.

Fig. 1
figure 1

A showed a subject in the intervention group using PEP device (PEP = 5cmH2O). B showed a spring linear pressure resistor. The sham PEP device used in the control group was not loaded with a spring linear pressure resistor

During the first visit (T0), the self-administered questionnaire was used to record the demographic and clinical characteristics, including age, weight, gender, medical history, tobacco history, pulmonary function tests, COPD grade, drug treatment, 6-min walk test (6MWT), The partial pressure of end-tidal carbon dioxide (PETCO2), and Modified British Medical Research Council score (mMRC score). The breathlessness in daily life was measured by a mMRC score [2]. The PETCO2 was measured by portable PETCO2 capnography monitor (Kingst®, KMI800). All subjects sat for 10 min to reach a steady state before measuring PETCO2, and then were measured PETCO2 in a seated position. The pulmonary function was performed (Master Screen Body, Jager, Germany) according to American Thoracic Society/European Respiratory Society (ATS/ERS) guidelines and measured the forced vital capacity (FVC), the forced expiratory volume in one second (FEV1), FEV1/FVC, inspiratory capacity (IC) and maximum voluntary ventilation (MVV) [22]. The 6MWT was performed in a 30-m long, flat, straight, indoor corridor supervised by a well-trained researcher according to the ATS/ERS guidelines [23]. Instructions were provided prior to the 6MWT and all 6MWTs were conducted by the same investigator. The 6-min walk distance (6MWD) was recorded. Heart rate (HR), oxygen saturation (SPO2) before and after the 6MWT (Initial SPO2 and End SPO2) and the minimal SPO2 during the 6MWT were measured continually through a wrist-worn pulse oximeter (WristOx2, Nonin Medical, Plymouth, MN). Subjects were shown the modified Borg scale before and after the 6MWT to rate their dyspnea and fatigue from 0 to 10, with higher scores indicating worse dyspnea and fatigue. All subjects performed two 6MWTs following the standard procedure mentioned in the guidelines with at least one hour between them and recording the highest 6MWD. After 2 months (T1), during the second visit, the same outcomes were collected and the comfort sensation of using PEP device by a Likert scale from 0 to 10 (0: no discomfort, 10 maximum discomfort) were also collected. For analysis, the ratings were divided into 3 categories: 0 to 3 = ‘not or slightly uncomfortable’, 4 to 6 = ‘moderately uncomfortable’ and 7 to 10 = 'uncomfortable or very uncomfortable’[24]. Primary efficacy outcomes were 6MWD, Borg-Dyspnea and Borg-Fatigue. Primary safety outcomes were PETCO2, PEP device-related discomfort and all adverse events. Adverse events associated with the use of the EPAP device were defined as asphyxia, pressure ulcers, acute deterioration, and death. Secondary outcomes were pulmonary function and mMRC score.

Subjects were contacted by phone every 3 days to confirm their adherence to the device. The subjects were asked about their feelings on using the EPAP device and whether they were still using it every day. All subjects were asked this: “Does anything about using this equipment make you uncomfortable?”, “Do you regularly use this device for 4 h every day?”, and “Do you encounter any issues while using this device?” All subjects received optimal medical therapy as usual according to GOLD guidelines at the time of enrollment. Changes to medication were not permitted throughout the study, unless required for safety [2]. If in doubt, subjects were instructed to consult the investigator any time in addition to scheduled phone interviews. All enrolled subjects were explained that two different forms of devices would be compared, but no further details were disclosed, thus, all enrolled subjects were blind on group allocation. All data analyses were performed with the evaluator blinded to the subjects’ condition.

Statistical analysis

SPSS version 22.0 (SPSS Inc., Chicago, IL, USA) was used for data processing and analysis. Data are presented as mean and standard deviation (mean ± SD), median and interquartile [M(P25, P75)] or frequency counts and percentages. The baseline characteristics were compared using independent samples t-test, Mann–Whitney U test or chi-square test. Paired t-test or Wilcoxon test was used for before-and-after comparisons within the intervention or control groups. Independent sample t test was performed for comparisons between groups. P < 0.05 was statistically significant. Subjects were a sample of convenience and the power for this study was 0.82.

Results

Twenty-five consecutive eligible subjects were randomly assigned to either the intervention group (n = 13) or the control group (n = 12), without dropouts and relevant adverse events occurring during the study (Fig. 2). There were no significant differences between two groups at baseline (P > 0.05) (Table 1) and no medication changes for all subjects during the whole procedure. Of all subjects who met the inclusion criteria, all current smokers were advised to quit smoking. Until the end of the trial, all subjects were compliant with the training regime.

Fig. 2
figure 2

Flow of recruitment of study subjects. A total of 25 subjects were eligible for inclusion and underwent randomization in the study, with 12 subjects in the control group (sham- PEP device) and 13 subjects in the intervention group (PEP device)

Table 1 Baseline characteristics

Compared with the control group, 6MWD (− 71.67 ± 8.70 m (95% CI [− 89.67, − 53.67]), P < 0.001), end-dyspnea (P = 0.002) and end-fatigue (P = 0.022) were significantly improved in the intervention group and PETCO2 was significantly lower in the intervention group (2.33 ± 0.43 mmHg, (95% CI [1.53, 3.12]), P < 0.001). All subjects adopted the PEP device stated that using PEP device in daily life was well acceptable. (Likert scale 1(0.5,1.25)) and there were no complications or adverse events. More details were shown in Table 2 and Fig. 3.

Table 2 Primary efficacy and safety outcomes
Fig. 3
figure 3

Primary efficacy and safety outcomes. A Showed the 6MWD at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). B Showed the Borg-dyspnea score at the end of the 6MWT at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). C Showed the Borg-fatigue score at the end of the 6MWT at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). D Showed the PETCO2 at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). T0 refers to the time when subjects were enrolled. T1 refers to the time when subjects were followed up at 2 months after enrollment. (*: P < 0.05; **: P < 0.001)

Compared with the control group, the initial SpO2 (1.13 ± 0.39% (95% CI [0.34, 1.93]), P = 0.007), minimal SpO2 (1.45 ± 0.36% (95% CI [0.70, 2.20]), P = 0.001), FEV1/ FVC (7.00 ± 2.85% (95% CI [1.10, 12.90]), P = 0.022) and mMRC (P = 0.003) was higher in the intervention group. More details were shown in Table 3 and Fig. 4.

Table 3 Secondary outcomes
Fig. 4
figure 4

Secondary outcomes. A Showed the SpO2 at the beginning of the 6MWT at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). B Showed the minimal SpO2 during the 6MWT at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). C Showed the SpO2 at the end of the 6MWT at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). D Showed the FEV1/FVC at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). E Showed the MVV at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). F Showed the mMRC score at baseline (T0) and two months later (T1) in the intervention group (solid circles) or control group (solid squares). T0 refers to the time when subjects were enrolled. T1 refers to the time when subjects were followed up at 2 months after enrollment. (*: P < 0.05; **: P < 0.001)

Discussion

To the best of our knowledge, this is the first randomized controlled trial to evaluate the effects of using PEP device during activities of daily living on exercise capacity in patients with COPD or Pre-COPD. The major findings of the present study are as follows: (1) Patients in the intervention group showed significant improvement in exercise tolerance after one month of using the PEP device (5 cmH2O), as evidenced by improved walking distance, significant improvement in dyspnea and fatigue after 6MWT and significant improvement in minimal SPO2 during 6MWT (all P < 0.05). Meanwhile, the use of the PEP device resulted in a mean 6MWD change of 71.67 ± 8.70 m and a mean Borg scale change in dyspnea at the end of the exercise of 1 (0.5–1.75) units, which were above the established minimal clinically important difference (MCID) [25, 26]; (2) Patients in the intervention group showed significant improvement in FEV1/FVC (P < 0.05); (3) There was a significant and clinically relevant decrease in activity-related dyspnea in the intervention group (P < 0.05), measured with the mMRC scale.

Dyspnea is a common symptom reported by patients with COPD, despite optimal medical management, the use of supplemental oxygen and PLB [27]. PEP could increase bronchial pressure, which can counteract intrinsic positive end-expiratory pressure and move the equal pressure point back to more central and solid airways, alleviate early dynamic airway compression on expiration and attenuate expiratory airflow limitation, which may result in more homogenous ventilation and improved gas exchange [15, 21, 27]. PEP could reduce the inspiratory threshold load of hyperinflated lungs in COPD during exercise, promoting an enhancement in neuromuscular coupling [28]. PEP could change breathing patterns and prolong expiratory time, and thereby reduce lung volumes and DH [19]. PEP devices we used were lower-cost, which facilitated their use in daily life for patients.

Monteiro et al. indicated that a small improvement in DH observed with PEP might lead to greater exercise endurance and reduced dyspnea [21]. However, they suggested that PEP devices increased expiratory effort, which may increase dyspnea and counterbalance the benefits of DH improvement [21]. We observed an improvement in mMRC score and a significant reduction in end-dyspnea after two months of PEP use, and subjects did not experience an increase in dyspnea. It is possible that PEP devices in our study were used during daily life rather than during strenuous exercise on a treadmill as in Monteiro et al. [21]. Another study conducted by Wibmer et al. noted that nasal PEP device reduced DH during 6MWT, but without an improvement in walking distance, possibly because the reduction of DH resulting from short application of the PEP device may not result in an acute improvement in exercise endurance, as subjects might require time to derive an advantage from their improved ventilatory capacity resulting from the reduction in DH [29]. The subjects in our study used the PEP device for 4 h per day for two months to improve DH, and 6MWD was significantly improved. Moreover, the PEP device used by Wibmer et al. provided the PEP of 10 cmH2O, and it has been shown that lower level of PEP may be more beneficial [18, 29]. The PEP used in our study was 5 cmH2O, which was similar with the pressure produced by PLB [15]. PEP of 5 cmH2O improved the exercise tolerance and did not cause significant discomfort such as dyspnea. Vanderschans GP et al. conducted a randomized cross-over controlled trial of eight male patients, all of whom performed two incremental cycling exercises while breathing with or without a PEP device, respectively [15]. They found that PEP may increase dyspnea, which may be related to the high intensity of exercise in their study. The high intensity exercise would make subjects more likely to exertional breathing and make PEP effects underestimated. In another study, subjects in the intervention group was breathing through a flow-dependent PEP device, which may produce a positive pressure of 4–20 cmH2O during leg extension exercise. The results showed that the subjects had improved DH, but did not reach statistical significance [20]. They noted that such leg exercising may involve an element of anaerobic metabolism and consequently they may have underestimated the benefit of PEP during purely aerobic exercise such as walking.

Nicolini et al. indicated that PEP (5 cmH2O) improved 6MWD and post-exercise SpO2, which was similar to our findings, and we also found a significant improvement in minimal SpO2 during 6MWT [16]. The PEP device could prevent the airway collapse, which results in more homogeneous ventilation. Thus, improved ventilation-perfusion distribution probably results in reduced dead space waste, improved alveolar ventilation, and improved SpO2 [20]. In our study, we observed an improvement in initial SpO2, which showed that that gas exchange improved after two months. Exercise intolerance in COPD is also associated with reduced IC and occurs with a neuromuscular “power reserve”, which is associated with resting FEV1/FVC [30]. IC, which represents the true operating limits for tidal volume expansion in patients with expiratory flow limitation during exercise, is an important surrogate measurement for COPD [31]. The FEV1/FVC, one of the spirometry measurements, was associated with expiratory flow limitation and used to classify the airflow limitation severity of patients with COPD [2]. We found there was a significant improvement in FEV1/FVC for subjects with EPAP device after 2 months and an improvement in IC without reaching statistical significance in our study, which may be due to IC at baseline having a higher value. Resting IC as a percentage of the predicted value was above 80% for all subjects in our study, which may be due to the fact that the majority of subjects had pre-COPD or mild COPD [7]. Monteiro et al. demonstrated an increase in IC in patients with moderate to severe COPD treated with EPAP at 5 to 10 cmH2O after submaximal treadmill exercise [21].

Regarding the safety of PEP devices, our study showed that daily use of PEP devices did not cause carbon dioxide (CO2) retention, but rather reduced the PETCO2, suggesting that the EPAP device alleviated airway obstruction and improved ventilation. VanderSchans et al.[15] showed that CO2 retention were increased in patients with COPD who breathed through a PEP device at 5 cmH2O during exercise. They hypothesized that insufficient positive pressure was generated to reduce airway closure and that using higher positive expiratory pressure would be more effective during exercise. Padkao et al.[20] applied conical-PEP with 13 cmH2O for patients, and the PETCO2 and SPO2 was not significantly altered, indicating that conical-PEP device with 13 cmH2O allowed appropriate gas exchange. Meanwhile, we observed an improvement in SPO2 in patients with PEP device. These indicated that PEP device used in our study did not have any adverse effects on CO2 retention or SPO2. In addition, the PEP device used in our study proved to be acceptable to the patients when used in daily life. More than 80% of those eligible patients were willing to try it, and of those who were willing, all found it acceptable. When used daily for 4 h a day, there were no adverse effects.

This study has some important limitations. First, the subject sample in our study was small, including subjects with a wide range of disease severity. It can be argued that the heterogeneity of the subjects could have affected the response to the intervention. Therefore, future studies should focus on a larger sample with a more homogeneous population, such as those with consistent disease severity, to avoid underestimating or overestimating the true effect of PEP devices. Second, the recommendation of 4 h of the use of PEP device has no real scientific background. Subjects may accept and tolerate an average of 4 h of PEP device use per day. However, further studies are needed to decide the duration of optimal use. Third, we do not know whether a longer study duration could have resulted in further improvement in lung function, and we expect future studies to clarify these clinically important questions. Fourth, the PEP device used in this study cannot currently be remotely monitored.

Conclusions

In conclusion, regular daily use of PEP device is safe and may improve exercise capacity in patients with COPD or pre-COPD. Meanwhile, the PEP device may be used in the future to assist COPD patients with breathing retraining and symptom management due to its efficiency, safety, simplicity and low cost when compared to other devices and approaches (such as NIV).