Background

Cataracts are the most common cause of impaired vision worldwide, and cataract surgery is the most common surgical procedure in the field of ophthalmology [1, 2]. Cataract surgery with conventional monofocal intraocular lens (IOL) implantation has shown very successful results for distant vision, but patients often require spectacle correction for near vision [3]. Multifocal IOLs have been developed to meet patients’ need for near vision, but there is a limit to the increased incidence of subjective visual disturbance, including halos and glare [4,5,6].

Currently, extended computer use and younger age at cataract surgery also give rise to growing needs for intermediate vision [7,8,9]. The newly developed TECNIS Eyhance ICB00 (Johnson & Johnson Vision Care, Inc.), a monofocal with a higher-order aspheric anterior surface IOL to enhance intermediate function, sought to meet those needs while sparing distant vision and visual disturbance. It shares the same geometry with the conventional monofocal TECNIS 1-piece ZCB00 IOL (Johnson & Johnson Vision Care, Inc.) about 85% of the surface except for the modified aspherical anterior surface of the optics [10]. This unique anterior surface is intended to create a continuous power change from the periphery to the center inducing the continuous power profile created with a higher-order asphere and improves intermediate vision. It is based on the refractive technology, without diffractive rings or zones, and it is visually indistinguishable from the TENIS 1-piece ZCB00 IOL. Thus, we wanted to know how the visual performance improved as the anterior surface profile changed from the ZCB00 to ICB00.

The purpose of this study was to compare clinical outcomes in terms of distance, near, and intermediate visual acuities, visual disturbances, and spectacle independence between patients who underwent bilateral implantation of ZCB00 IOLs and patients who underwent bilateral implantation of ICB00 IOLs. Besides, we also evaluated the optical performance of two IOLs through optical bench testing.

Methods

Study population

This retrospective case-control study included patients who underwent cataract surgery with either the ZCB00 or ICB00 IOLs implanted bilaterally at the Korea University College of Medicine between March and October 2020. Patient who had a postoperative corrected distance visual acuity (CDVA) of 20/40 or better in the operated eye were included. Eyes with traumatic cataracts, a previous history of ocular surgery, eventful surgery (eg, anterior capsule tear), or postoperative complications were excluded. This study adhered to the tenets of the Declaration of Helsinki, and was approved by both the Institutional Review Board of Korea University Guro Hospital (IRB no. 2020GR0525) and that of Korea University Ansan Hospital (IRB no. 2020AS0344). The Institutional Review Board of Korea University Medicine waived the need for written informed consent from the participants, because of the study’s retrospective design.

Patient examination

Preoperative uncorrected distance visual acuity (UDVA) was measured at 4 m. The preoperative corneal power, anterior chamber depth (ACD), and axial length (AL) were measured using an IOLMaster 500 (Carl Zeiss Meditec AG, Jena, Germany). The IOL power was calculated based on the predicted refraction by Haigis formula, and targeted between 0 and − 0.50 D. IOL constants of a0, a1, and a2 for the Haigis formula were − 1.302, 0.210, and 0.251, respectively.

Surgical technique

All phacoemulsification and IOL implantations were performed by one of two experienced surgeons (S.J.S. and E.Y.) in one of our two institutions under topical anesthesia with 0.5% proparacaine hydrochloride (Alcaine; Alcon Laboratories Inc., Fort Worth, Tx or Paracaine; Hanmi Pharm, Seoul, Korea). After making a 2.20- or 2.75-mm clear corneal incision, a 26-gauge needle and a capsulorrhexis forceps were used to create a continuous curvilinear capsulorrhexis slightly smaller than the IOL optic size. The phacoemulsification was performed with either the stop-and-chop or phaco-chop technique. The IOL was folded for implantation using an insertion system, and inserted into the capsular bag through a clear corneal incision. All patient was treated with topical 1.5% levofloxacin (Cravit; Santen Pharmaceutical, Osaka, Japan) and topical steroid eyedrop (1% prednisolone acetate (Pred Forte®; Allergan, Inc., Irvine, CA) or 0.1% fluorometholone (Santen Pharmaceutical)) 4 times daily, and 0.1% bromfenac sodium (Bronuck; Taejoon Pharm, Seoul, Korea) twice daily from 3 days before cataract surgery to 4 weeks after cataract surgery.

Patient evaluation

Postoperative monocular and binocular uncorrected and corrected distance visual acuity (CDVA) at 4 m, binocular uncorrected and distance corrected near visual acuity (UNVA, DCNVA) at 40 cm, and binocular uncorrected and distance corrected intermediate visual acuity (UIVA, DCIVA) at 66 cm were measured at postoperative visits 4 weeks after surgery. The distance corrected defocus curves were obtained binocularly at 4 m to measure the visual acuity with each defocus between − 3.00 D and + 1.00 D in 0.50 D intervals [6].

The refractive prediction error was defined as the difference between postoperative achieved refraction and preoperative targeted refraction (i.e., refractive prediction error = achieved spherical equivalent – targeted spherical equivalent). Mean absolute error (MAE) was defined as the mean absolute value of refractive prediction error and median absolute error (MedAE) was defined as the median absolute value of refractive prediction error [11].

Postoperative incidence of photic phenomena (glare, starburst, and halos), patient satisfaction for overall, near, intermediate, and distance vision, and dependence on glasses for near, intermediate, and distance vision were assessed with a questionnaire answered at 4 weeks after cataract surgery [6, 12]. A questionnaire with illustrations was used to assess whether patients experienced photic phenomena after cataract surgery. Patient satisfaction was rated on a 1-5 scale: 1, Very dissatisfied; 2, Dissatisfied; 3, Neutral; 4, Satisfied; 5, Very satisfied. Spectacle dependence was rated on a 1-5 scale: 1, Never; 2, Seldom; 3, About half the time; 4, Usually; 5, Always (Table 1) [12].

Table 1 Questionnaire used for comparison between monofocal and new monofocal with enhanced intermediate function intraocular lenses

Optical bench system

The optical bench system used in this study consisted of a LED light, the 1951 United States Air Force (1951 USAF) resolution test chart, an artificial pupil, a pupil camera, trial lens, model eye, and complementary metal-oxide-semiconductor (CMOS) camera (BFS-U3-120S4M-CS; FLIR Systems Inc., Wilsonville, OR) (Fig. 1) [13]. The model eye composed of an aberration-free artificial cornea and a wet cell which was made of N-BK7 (DG100X100-600) and filled with a balanced salt solution, and they were mounted on the XYZ translation stage [14]. The IOL was fixed using an aspheric lens adapter which was mounted on the XYZ translation stage and then positioned in a wet cell. After that the lenses center and the camera were precisely aligned so that the image focus was on the camera’s sensor. The 1951 USAF resolution chart was illuminated by 555 nm LED light, and the image formed by the model eye was obtained by the CMOS camera [14, 15]. The trial lens was placed in front of the model eye to obtain defocus image between − 2.50 D and + 1.00 D in 0.50 D intervals [16]. These measurements were repeated at different pupil size from 2.0 mm to 5.0 mm in 1.0 mm increments using the artificial pupil with scale.

Fig. 1
figure 1

Schematic diagram of the optical bench system used in this study (LED = light emitting diode; USAF = United States Air Force; IOL = intraocular lens; CMOS = complementary metal–oxide–semiconductor)

The images of the 1951 USAF target formed by two IOLs of 21.0 D ZCB00 and ICB00 IOLs were compared following International Organization for Standardization (ISO) 11,979-2 requirements and test methods (International Organization for Standardization, 2014. Ophthalmic Implants – Intraocular Lenses – Part 2: Optical Properties and Test Methods). The element 3 of the group 2 in the 1951 USAF resolution test chart was set to be 15 cycle per degree (CPD), which is approximately equivalent to 20/40 vision. The images were converted to through-focus modulation transfer function (MTF) in the horizontal and vertical directions using the MatLab (Mathworks, Inc., Natick, MA) [17]. The root mean square of the horizontal and vertical MTF (MTFRMS) was calculated using the following formula [15]:

$$MTF=\frac{I_{max}-{I}_{min}}{I_{max}+{I}_{min}}\kern0.5em {MTF}_{RMS}=\sqrt{\frac{MTF_{Horizontal}^2+{MTF}_{Vertical}^2}{2}}$$

Statistical analysis

Data analysis was performed using the Statistical Package for the Social Sciences (version 20.0, SPSS, Inc.). Student’s t tests, Mann-Whitney U tests, and Fisher’s exact tests were performed to compare the clinical characteristics, IOLMaster 500 measurements, implanted IOL power, monocular UDVA, monocular and binocular CDVA, binocular CDVA, UNVA, DCNVA, UIVA, and DCIVA, incidence of photic phenomena, patient satisfaction score, and spectacle dependence score between the ZCB00 and ICB00 groups. A P value less than 0.05 was considered statistically significant.

Results

Clinical study

This study evaluated 60 eyes of 30 patients who underwent uncomplicated phacoemulsification with bilateral implantation of the TECNIS 1-piece ZCB00 (30 eyes of 15 patients) or TECNIS Eyhance ICB00 (30 eyes of 15 patients) IOL. The mean age of the 30 enrolled patients was 70.1 ± 7.1 years (range: 52–83 years). Among the total study population, there were 19 women (63.3%). There was no significant difference in mean age, sex ratio, preoperative UDVA, or mean corneal power between the two groups. On the other hand, the ZCB00 group showed a shallower ACD, shorter AL, and higher IOL power compared to the ICB00 group (Table 2).

Table 2 Baseline characteristics of patients with cataract and their eyes in a comparison between monofocal and new monofocal with enhanced intermediate function intraocular lenses

Mean targeted and achieved refraction of the ICB00 group were − 0.23 ± 0.31 and − 0.16 ± 0.37 D, respectively, and those of the ZCB00 group were − 0.27 ± 0.19 and − 0.26 ± 0.34 D, respectively. There was no significant difference in MedAE and MAE between the ICB00 and ZCB00 groups (Table 3).

Table 3 Comparison of median absolute error and mean refractive prediction error calculated using the Haigis formula between the ICB00 and ZCB00 groups

Table 4 shows the comparison of postoperative visual acuity between the ICB00 and ZCB00 groups. The mean monocular UDVA and CDVA, and binocular CDVA at 4 m was 0.06, − 0.01, and − 0.04 logMAR for the ICB00 group, respectively, 0.07, − 0.02, and − 0.05 logMAR for the ZCB00 group, respectively, with no significant differences between the two groups. On the other hand, the mean binocular UNVA was significantly better in the ICB00 group (0.09 logMAR) compared to the ZCB00 group (0.35 logMAR) (p <  0.001). Besides, the binocular UIVA and DCIVA were also better in the ICB00 group (0.03 and 0.01 logMAR) compared to the ZCB00 group (0.25 and 0.13 logMAR) (p <  0.001, p = 0.031, respectively).

Table 4 Comparison of postoperative visual acuity between the ICB00 and ZCB00 groups

Figure 2 shows binocular distance-corrected defocus curves at 4 weeks after cataract surgery of the two groups. Both curves peaked at 0.00 D of defocus and decreased with increasing negative defocus. However, the ICB00 group achieved a smooth landing area with a less abrupt decrease in visual acuity, especially within the intermediate defocus range up to − 1.50 D. The mean visual acuity was 0.1 logMAR or more between + 1.00 D to − 1.50 D of defocus in the ICB00 group. The mean visual acuity of the ICB00 group was significantly better than that of the ZCB00 group at between − 1.00 D to − 3.00 D of defocus.

Fig. 2
figure 2

Mean binocular distance-corrected defocus curves at 4 weeks after cataract surgery of the ICB00 (solid line) and ZCB00 (dashed line) groups. An asterisk indicates p < 0.05 by the Student’s t-test

There was no significant difference in the incidence of glare, starburst, and halos between the ICB00 and ZCB00 groups (Table 5). The mean patient satisfaction score for near and intermediate vision was significantly higher in the ICB00 group (4.0 ± 0.9 and 4.4 ± 0.6, respectively) than in the ZCB00 group (3.2 ± 0.9 and 3.7 ± 0.8, respectively) (P = 0.026 and P = 0.017, respectively; Fig. 3). However, there was no significant difference in the mean patient satisfaction score for overall and distance vision between the two groups. The spectacle dependence score for near and intermediate vision was significantly better in the ICB00 group (1.7 ± 1.3 and 1.1 ± 0.3, respectively) compared to the ZCB00 group (3.2 ± 1.4 and 2.3 ± 1.4, respectively), but not that for distance vision (Fig. 3).

Table 5 Comparison of postoperative incidence of photic phenomena between the ICB00 and ZCB00 groups
Fig. 3
figure 3

a Patient satisfaction score rated on a five-point scale as follows for overall, near, intermediate, and distance vision: 1, very dissatisfied; 2, dissatisfied; 3, neutral; 4, satisfied; 5, very satisfied. b Spectacle dependence score rated on a five-point scale as follows for near, intermediate, and distance vision: 1, never; 2, seldom; 3, about half the time; 4, usually; 5, always. An asterisk indicates p < 0.05 by the Student’s t-test

Optical bench performance

Figure 4 shows captured images of the 1951 USAF resolution test chart from two IOLs. For the ZCB00 IOL, as the minus diopters were added, the image was gradually blurred, and the image became indistinguishable from − 1.00 D. However, for the ICB00 IOL, the image was identifiable until − 2.00 D was added.

Fig. 4
figure 4

Captured images of the 1951 United States Air Force resolution test chart using the ICB00 and ZCB00 intraocular lenses. Minus diopter defocus represents near distance

MTF analysis showed that the ICB00 IOL had higher MTFRMS values at − 0.50 D, − 1.00 D, − 1.50 D, and − 2.00 D of defocus (0.459, 0.320, 0.315, and 0.168, respectively) compared to the ZCB00 IOL (0.280, 0.104, 0.098, and 0.075, respectively) (Fig. 5A). This result is consistent with the defocus curves obtained in the clinical study.

Fig. 5
figure 5

A comparison of the root mean square of modulation transfer function (MTF) values in the horizontal and vertical directions (MTFRMS). a A comparison of MTFRMS values measured at a 3.0-mm pupil size between the ICB00 (solid line) and ZCB00 (dashed line) intraocular lenses. b A comparison of MTFRMS values obtained with the ICB00 intraocular lens according to pupil size

The highest MTFRMS values of the ICB00 IOL at − 0.50 D of defocus (0.690) was obtained in the pupil size of 2.0 mm (Fig. 5B and Table 6). On the other hand, the MTFRMS values of the ICB00 IOL at between − 1.00 D to − 2.50 D of defocus was similar between the pupil size of 2.0 mm and 3.0 mm. This result indicates that best optical performance of the ICB00 IOL in the intermediate distance could obtained at 2.0 mm pupil size.

Table 6 A comparison of modulation transfer function (MTF) values in the horizontal and vertical directions and the root mean square of MTF values in those two directions obtained with the ICB00 intraocular lens according to pupil size

Discussion

This study showed that after the Eyhance ICB00 IOLs were implanted bilaterally, patients could obtain better intermediate and near vision than the ZCB00 IOL while maintained the distant vision. Moreover, there was no difference in the incidence of photic phenomena between the ICB00 and ZCB00 IOL. In the optical bench test, the Eyhance ICB00 IOL showed an excellent MTF curve at intermediate distances, and it was affected by the pupil size.

Although cataract surgery with monofocal IOL implantation is still the most common option, there is a gap between postoperative results and patient’s expectations as it limits to improve only distant vision. At this point, many investigators have developed other IOL designs to improve near and intermediate vision, but those IOLs also have limitations as they accompany unwilling phenomena such as reduced contrast sensitivity, halos, and glares [18,19,20]. Besides, as the optical technologies differ, the subjects to apply those advanced IOLs are limited. Therefore, the Eyhance ICB00 IOL could be a good option as it shares the same geometry with the monofocal 1-piece IOL and provides an improved intermediate and near vision, and better spectacle independence as well, without accompanying other unwilling phenomena.

Previous studies reported the Eyhance ICB00 IOL’s effectiveness as it provided improved intermediate vision compared to the monofocal 1-piece IOL [10, 21, 22]. In those previous studies, the Eyhance ICB00 IOL yielded better UIVA, higher spectacle independence at intermediate distance, and similar optical quality compared to the classic monofocal IOL, but provided similar distance and near visual acuities [10, 21, 22]. The results of this study also showed that the Eyhance ICB00 IOL produced better intermediate VA and similar optical quality compared to the ZCB00 IOL. However, binocular UNVA and spectacle independence at near distance for the ICB00 group were better than those of the ZCB00 group in this study. The reason for the discrepant results between the previous studies and this study in terms of near vision improvement by the Eyhance ICB00 IOL compared to the ZCB00 IOL may be due to race and study design. The previous two studies indicating that the Eyhance ICB00 IOL provides better intermediate vision but not near vision were conducted in a prospective manner in Europe. On the other hand, this study was conducted in a retrospective manner in Asia. In line with this study, a previous study conducted in a retrospective manner in Asia showed that the Eyhance ICB00 IOL provided superior intermediate and near vision compared to the ZCB00 IOL [23]. Therefore, the effect of the Eyhance IOL for improving near vision needs to be confirmed through a large-scale study.

Optical bench tests in this study supported clinical outcomes. The MTF curves of two IOLs at distances showed similar good results. In contrast, the Eyhance ICB00 IOL showed better outcomes at intermediate distances, especially within the range up to − 1.50 D. Besides, when evaluating the effect of pupil size on the MTF curves of the Eyhance ICB00 IOL, it showed the best results with the 2.0 mm pupil size at intermediate distances. At the same time, it had a minimal impact at distances. The unique anterior surface lies on the center deviation of the Eyhance ICB00 IOL may explain those results [24]. Previous study showed that the through-focus MTF curve of the ICB00 shifted to a myopic defocus of − 0.50 D at a 2.0 mm pupil size and the maximum MTF value was obtained at − 0.5 D defocus [25]. However, in this study, the maximum MTF value was obtained at 0.0 D defocus without myopic shift of the MTF curve at a 2.0 mm pupil size. Considering the myopic shift of the through-focus MTF curve in the previous study, it is thought that a relatively lower MTF value (0.459) was measured at − 0.5 D defocus without myopic shift of the MTF curve in this study.

One limitation of the current study is that the sample size was relatively small, the follow-up period was relatively short, and the study design retrospectively reviewed the medical records. The other is that we did not evaluate whether patients’ pupil size had impacts on visual acuity.

Conclusions

In conclusion, the TECNIS Eyhance ICB00 IOL may be a good option for both clinicians and patients. It provided better intermediate and near vision than the TECNIS 1-piece IOL while maintaining excellent distance vision without worsening the visual symptoms. Moreover, the results of the optical bench test, which showed that the Eyhance ICB00 IOL had higher MTF values between − 0.50 D and − 2.00 D of defocus compared to the ZCB00 IOL, supported these clinical outcomes.