Background

Coronary artery disease (CAD) is a major burden for health care systems and economies worldwide [1]. Percutaneous coronary intervention (PCI) is a common approach in CAD to reduce the incidence of myocardial infarction (MI) and to reduce mortality rates in acute coronary syndrome [2]. PCI originated in the 1980s and has become increasingly popular [3]. Currently, more than 500,000 PCI procedures are performed annually worldwide [4]. Although many stent implantation operations are carried out successfully every day, the occurrence of a rare but potentially fatal situation during PCI should always be kept in mind [5].

Case presentation

A 44-year-old male was admitted to the hospital due to chest pain after exercise and positive of the exercise treadmill test. His medical history was limited to moderate hypertension without medication. The patient had no past surgical/interventional history. After the patient was admitted, the electrocardiogram (ECG) and echocardiogram showed no obvious abnormalities. High-sensitivity troponin I (hsTnI) was normal. Blood lipid tests indicated moderate hyperlipidemia. A coronary angiography showed a balanced coronary dominant circulation, with diffuse and severe stenosis in the middle right coronary artery (RCA) and 90% stenosis by diameter in the proximal RCA (Fig. 1a, Supplementary Video 1), which was consistent with a diagnosis of CAD. Due to 90% stenosis in the RCA and typical ischemic symptoms with chest pain after exercise, as well as positive of the exercise treadmill test, PCI was indicated [6, 7]. With the consent of the patient and his family, PCI in the RCA was performed. Heparin (100 U/kg) was used during PCI. A 6Fr SAL 0.75 guiding catheter and SION-BLUE wire (Asahi, Tokyo, Japan) were used. First, the stenotic area was dilated with a 2.0 mm*20 mm compliance balloon (Fig. 1b, Supplementary Video 2). Then, a 3.0 mm*33 mm drug-eluting stent (Firebird, MicroPort, China) was implanted in the middle of the RCA (Fig. 1c, Supplementary Video 3). Finally, a 4.0 mm*33 mm drug-eluting stent (Firebird, MicroPort, China) was implanted in the proximal RCA (Fig. 1d, Supplementary Video 4).

Fig. 1
figure 1

Coronary angiography and stent implantation in the RCA. a Coronary angiography showed diffuse arteriosclerosis with severe stenosis in the middle RCA, and the proximal RCA had 90% stenosis by diameter. b The stenotic region was dilated with a 2.0 mm*20 mm compliance balloon. c A 3.0 mm*33 mm drug-eluting stent (DES) was implanted in the middle of the RCA. d A 4.0 mm*33 mm drug-eluting stent was implanted in the proximal RCA

Under normal conditions, the balloon is inflated to expand the stent, and the stent compresses the atherosclerotic plaque against the vascular wall and restores blood flow to the myocardium. Then, the balloon is withdrawn after deflation. However, after we had dilated the stent balloon with a pressure of 14 atmospheres, the stent balloon could not be deflated. Although the indeflator had achieved negative pressure, the stent balloon still kept expanding and did not deflate. We suspected that the indeflator was broken, so we immediately replaced it with a new one. After the stent balloon was connected, negative pressure was exerted, but the stent balloon still kept expanding. Because the 4.0 mm*33 mm stent balloon continued to block the ostial RCA, resulting in myocardial ischemia, the patient began to complain of general sweating with chest pain, progressing to hypotension and a low heart rate. ECG monitoring in Hemodynamic Recording System showed that the arterial pressure was 63/34 mmHg, and the heart rate was 42 beats/minute, with ST segment elevation in lead II. Immediate action was required, otherwise, the patient's life would be in danger. We contacted the cardiac surgery team by telephone as soon as possible, to activate bystander cardiac surgery. Meanwhile, we forced the stent balloon out of the RCA (Fig. 2a, Supplementary Video 5). When the stent balloon was extracted from the RCA and was in the ascending aorta, the patient's discomfort immediately improved, and blood pressure and heart rate gradually returned to normal.

Fig. 2
figure 2

Stent balloon deflation failure and the treatment. a The stent balloon was pulled out of the RCA. b The 4.0 mm*33 mm stent balloon was pulled to the right radial artery

Another problem was removing the dilated stent balloon from the body. To prevent the balloon from traveling into the cerebral artery or descending aorta, we first pulled the stent balloon to the right radial artery (Fig. 2b, Supplementary Video 6). Then, we tried again to deflate the balloon. Two approaches were considered: one was to dilate the balloon at a pressure greater than the rated burst pressure (RBP). In this way, the volume of the balloon would be reduced after the contrast agent flowed out. Because of the fear of radial artery damage with balloon explosion, we rejected this approach; Another method was to puncture the balloon with a hard guide wire to allow the contrast agent to flow out. However, repeated attempts to puncture the balloon with a Conquest Pro guide wire (Asahi, Tokyo, Japan) failed. Thus, we forced the indeflator with a continuous negative pressure, so that the defect balloon could be retracted into the guiding catheter due to a little partially deflation. Finally, the balloon in the expanded state and the guiding catheter were successfully pulled out of the body.

We reviewed the radiography of the RCA and confirmed that the stent was still inside the proximal RCA and had not slipped away (Fig. 3a). Intravascular ultrasound (IVUS) was performed (Fig. 3b, Supplementary Video 7), because we were concerned that the RCA or the stent had been damaged when the stent balloon was forced out of the RCA. Angiography of the RCA was finally performed, and re-examination showed that the stent had expanded completely and was well apposed, with the stenosis of the RCA opened (Fig. 3c, Supplementary Video 8). There was no obvious intimal lesion on either side of the stent. We tested the stent balloon in vitro, and the stent balloon again failed to deflate, confirming that this was not a case of transitory failure of balloon deflation (Supplementary Video 9). After the operation, the patient returned to the ward and was discharged three days later. At the 6-month follow-up, the patient did not have any discomfort.

Fig. 3
figure 3

Post-PCI angiography of the RCA with IVUS detection. a Radiography of the RCA was performed to confirm that the stent was still inside the proximal RCA. The red arrow points to the stent. b Intravascular ultrasound was performed to check the stents in the RCA. Angiography was finally performed to check the RCA

Discussion

Deflation failure of a stent balloon is an extremely rare complication of PCI [8]. The deflation failure of a balloon following inflation occurs mostly due to a mechanical obstruction in the shaft of the balloon [9, 10]. These obstructions block the channel, the patency of which is required for the development of negative pressure adequate for balloon deflation. In this case, the stent company inspected the stent balloon after the operation. It was found that the balloon weld, which marks the site where the balloon is laser welded to the shaft, was compromised by a manufacturing defect. Thus, a one-way valve had been formed in the lumen of the balloon. When positive pressure was applied for inflation, the contrast media could pass through, but the negative pressure of deflation would block the lumen. Therefore, balloon deflation failure occurred. According to the evaluation of the stent company, this defect of the balloon was an isolated case, and did not occur again in the following days. Thus, the company did not perform a recall or warning.

There are various stepwise strategies that may be adopted by the operator to approach the dilemma of balloon deflation failure. The first option is the application of negative suction pressure. The second option is to inflate balloon beyond maximal rated pressure to facilitate its bursting. The third option is transection of the balloon shaft. The fourth option involves extracting all systems. The fifth option is to puncture the balloon using the stiff end of a guide wire, either a 0.014″ or 0.018″ guide wire [11, 12]. Of course, the last option is to send the patient for surgical retrieval of the balloon if the other methods proved ineffective [13]. Generally, these options are included in our normal plan for dealing with balloon deflation failure. However, in this case, the patient's hemodynamics were very unstable, and the situation was urgent. We did not have enough time to try the methods above, which were time consuming and would cause some inevitable delay. Therefore, given the emergency, we directly pulled the balloon out of the proximal RCA with gentle force. Indeed, this approach was not always recommended. Especially when an undeflatable balloon were entrapped more distally in the RCA or other coronary arteries, the cardiac surgery should be activated, because more serious risks such as coronary rupture, tamponade, malignant arrhythmias and stent emigration might occur. Fortunately, once the balloon was pulled out of the RCA and into the radial artery, the coronary artery blood flow was immediately restored. Thus, the hemodynamics improved, which kept the patient safe. We then had enough time to try various methods for getting the balloon out of the body. Finally, we were successful, and surgery was avoided.

Conclusions

Cardiologists engaged in cardiovascular intervention should be prepared for complications such as balloon deflation failure. In patients with hemodynamic stability, the technique of puncturing the balloon with a stiff guide wire may be considered. However, if the patient is unstable, it is urgent to immediately restore blood flow in the blocked coronary artery. After the patient's hemodynamics are stable, the remaining related complications can be managed.