Abstract
Because the human eye cannot visually detect the results of direct bilirubin test papers accurately and quantitatively, this study proposes four different highly collimated mini light-emitting diodes (HC mini-LEDs) as light sources for detection. First, different concentrations of bilirubin were oxidized to biliverdin by FeCl3 on the test paper, and pictures were obtained with a smartphone. Next, the red, green, and blue (RGB) channels of the pictures were separated to average grayscale values, and their linear relationship with the direct bilirubin concentration was analyzed to detect bilirubin on the test paper noninvasively and quantitatively. The experimental results showed that when green HC mini-LEDs were used as the light sources and image analysis was performed using the G channel, for a direct bilirubin concentration range of 0.1–2 mg/dL, the G channel determination coefficient (R2) reached 0.9523 and limit of detection was 0.459 mg/dL. The detection method proposed herein has advantages such as rapid analysis, noninvasive detection, and digitization according to RGB grayscale changes in the images of the detection test paper.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Bilirubin is a yellow pigment produced by the redox reaction of hemoglobin in the red blood cells. In the human body, it normally circulates through the blood to the liver, where it combines with glucuronic acid by the UDP-glucuronosyltransferase (UGT) system. Then, it is excreted with bile into the intestine, where it is finally converted to urobilinogen and expelled via urine or feces [1, 2]. Bilirubin before and after binding with the liver cells is called as indirect or unconjugated bilirubin and direct or conjugated bilirubin, respectively, while the total bilirubin is given by the sum of indirect and direct bilirubin [3, 4]. A high indirect bilirubin index can affect brain development, which is a condition known as jaundice or hyperbilirubinemia. This is partly attributed to the inability of the liver cells to convert indirect bilirubin to direct bilirubin quickly, resulting in excessively high indirect bilirubin in the blood. If the direct bilirubin index is too high, it is clinically referred to as conjunctive or bile-retentive jaundice, and the possibility of biliary atresia should be considered in neonates [5]. Neonatal jaundice affects 60% and 80% of full-term and preterm neonates, respectively. Therefore, detection of bilirubin is a routine test item for each newborn [6, 7]. Newborns often have physiological jaundice because their liver is not fully developed, and the liver cells lack the ability to bind bilirubin. This problem disappears when the body’s ability to bind bilirubin matures [8,9,10]. In addition, neonates may develop pathological jaundice due to hepatitis or biliary atresia. In biliary atresia, early diagnosis and surgery are necessary to avoid the eventual need for a liver transplant [11,12,13]. The normal concentration range of total bilirubin in human serum is approximately 0.2–1.2 mg/dL. If the bilirubin index remains too high, it may lead to conditions such as hepatitis, athetosis, and even brain damage [14, 15]. Currently, the most common approach for detecting bilirubin in clinical medicine is the diazo method [16], which uses serum and diazo reagents in a strong acidic environment. Through the bilirubin reaction, diazo salts are combined to form dark-red azo compounds, which is then measured using a photometer to obtain the fluorescence intensity and proportion of bilirubin concentration for colorimetric determination [17,18,19]. However, the diazo reagents are easily affected by the pH of the environment, which can cause detection errors [20,21,22]. Bilirubin oxidase (BOD) is a multi-copper oxidase belonging to the class of oxidoreductases. It uses metal ions to catalyze the oxidation of bilirubin to biliverdin and has been widely employed as a reagent for the detection of bilirubin [23]. The laboratory mainly uses enzymatic methods to measure the direct and serum bilirubin concentrations. When direct bilirubin is oxidized by BOD under acidic conditions to form biliverdin, the original concentration of bilirubin is determined according to the generated bilirubin [24,25,26,27]. Although BOD has been widely used as a reagent for the detection of bilirubin, the enzymatic method still suffers from low sensitivity, high cost, and insufficient stability [28]. Fluorescence detection mainly uses metal nanodots or quantum dots to calibrate the detection object, and it uses the change in fluorescence intensity to achieve analysis [29,30,31,32,33,34]. Although the fluorescence method achieves the goal of quantitative detection of bilirubin, it has shortcomings, such as limited applications of fluorescence imaging in vivo and in vitro, and easy interference with samples containing relatively complex components so difficulty with using in clinical practice [35]. The blood test by venipuncture is currently the most accurate method for the clinical detection of human bilirubin in hospitals, but blood drawing is invasive and increases the risk of infection for special patients, time-consuming, expensive, and painful for the patients [36, 37]. Therefore, some noninvasive methods have also been proposed [38, 39]. However, with the current test-paper detection method, the change in color can only be distinguished through qualitative visual judgment by the human eye, resulting in low detection accuracy when the concentration of bilirubin in the urine is low or when the color change is not apparent. Compared with light sources used in traditional analytical instruments, the full-width at half maximum (FWHM) of light-emitting diodes (LEDs) is narrow and usually less than 30 nm, which can improve the utilization efficiency of light. Therefore, LEDs have been applied in color analyses, capillary absorption detection, photoacoustic spectroscopy, microscopic fluid control devices, and fluorescence spectroscopy [40]. Recently, smartphones have developed rapidly in terms of size, processing speed, and camera resolution, so the application of the smartphone as a high-performance image detection instrument has become possible. Moreover, the popularity of the smartphone also avoids the traditional need for complex and expensive equipment [41]. Pairing a red, green, and blue (RGB) analysis app or software with the smartphone offers advantages for developing cost-effective, portable, and easy-to-use methods that enable real-time and on-site detection of the analytes [42].
To solve the problem of the inability of the human eye to visually discriminate color changes to the test paper, this study proposes image spectral analysis with highly collimated mini-LEDs (HC mini-LEDs) as the light sources for quantitative detection of direct bilirubin. The measurement results showed that when using green HC mini-LED as the light source and using the G channel for image analysis, the G channel determination coefficient (R2) reached 0.9523 in the direct bilirubin concentration range of 0.1–2 mg/dL, with a detection limit (LOD) of 0.459 mg/dL. The detection method proposed in this paper has the advantages of rapid analysis, non-invasive detection, rapid detection, and quantitative analysis.
2 Materials and methods
2.1 Experimental materials
Direct bilirubin powder (Bilirubin Conjugate, Ditaurate, Disodium Salt – Calbiochem, Merck Millipore 201102, Inc., USA) is the current standard used by medical institutions to simulate direct bilirubin. The certificate of analysis (CoA) of the direct bilirubin powder was obtained according to Regulation (EC) No. 1907/2006. After using a precision scale to weigh the direct bilirubin powder, 25 mL of deionized water was added to 0.5 mg of the powder and stirred magnetically at 500 rpm for 5 min until evenly mixed to form 2 mg/dL of the solution. Then, 2 mg/dL of the direct bilirubin solution was added to deionized water and diluted into 12 standard solutions of different concentrations ranging from 0.1 to 2 mg/dL. This study used Fouchet’s reagent (S-Y Fouchet’s reagent-SY8076-1–11011019-Shih-Yung Instruments Co., Ltd., Taiwan) as the oxidant of bilirubin, and the prepared direct bilirubin solutions were oxidized with Fouchet’s reagent. In the reaction, bilirubin was oxidized and reduced to blue/green biliverdin and Fe3+. The bilirubin redox reaction is shown in Eq. (1).
Direct bilirubin test paper (S-Y U-B Test Kit Lot No. 1110412-SY8076-11011019-Shih-Yung Instruments Co., Ltd.) of diameter 30 mm was used as the analytical direct bilirubin carrier. The four HC mini-LEDs used in this study were provided by EVERLIGHT Corporation, New Taipei, Taiwan. The model, viewing angle, and peak wavelength of the four LEDs are as follows: cool white (334-15/F1C1-1XZA), 10°, 446.8 nm; red (333-2SURC/S400-A8), 8°, 636.1 nm; green (333-2SUGC/S400-A5), 12°, 518.5 nm; blue (333-2SUBC/C470/S400-A6), 8°, 452.9 nm. The luminous intensity distribution curves of the four HC mini-LEDs are shown in Fig. 1.
2.2 Recording images and RGB analysis
Images were captured with an iPhone 8 Plus camera, with the test strip at 90° to the camera lens. The camera’s shutter speed and focal length were 1/15 s and 10 mm, respectively, and the vertical distance between the camera lens and HC mini-LED light source as well as the test strip sample were 10 cm and 15 cm, respectively. P(λ) and R(λ) are the light source distribution and reflection spectra, respectively. After multiplying P(λ) and R(λ), the spectrum of biliverdin on the test paper can be obtained. Finally, the captured image is analyzed using MATLAB software to obtain the grayscale values of the RGB channels. Then, the grayscale values of the RGB channels are divided by the same sampling pixel number, and the average grayscale is used to represent its color composition for analysis.
Converting RGB images to grayscale images is achieved through the transformation of Y, U, and V, where Y represents luminance, U represents chrominance, and V represents chroma. The formula for converting R, G, B to Y, U, V is given as Eq. (2):
Y = R * 0.2989 + G * 0.5870 + B * 0.1140, which is the grayscale conversion formula [43]. The schematic diagram of the image analysis is shown in Fig. 2.
2.3 Experimental process
In this study, HC mini-LEDs and image grayscale analysis were used to detect direct bilirubin. The HC mini-LEDs were used as light sources, and images were captured with a smartphone. MATLAB software was used to analyze the captured images. Before we start the experimental process, we need to confirm the voltage and current used for the LED, taking into consideration the camera’s ISO sensitivity setting to ensure that the captured images are not overly exposed. Then, process the captured images. We begin by analyzing the grayscale values for the lowest concentration since higher concentration grayscale values will be lower. Therefore, as long as the grayscale values for the lowest concentration do not experience excessive exposure (with values closer to 255 being preferable), it ensures that higher concentrations will also avoid overexposure. Then, the experimental process as below. Step 1 involves pretreatment and zero calibration, where we prepared a total of 12 different concentrations of the direct bilirubin standard solutions from 0.1 to 2.0 mg/dL and added 0.1 mL of Fouchet’s reagent as the oxidant to each solution. Then, the parameters, such as current, voltage, distance of the light source, and focal length of the smartphone, were corrected and fixed. Step 2 involves image processing and image analysis on images captured by the four HC mini-LED light sources. After calibration of the measurement parameters, image acquisition of the 12 direct bilirubin test papers for different concentrations was performed. Step 3 entails finding the best linear R2 relationship between the image RGB values. Cool white LEDs were used as the detection light source, and three kinds of monochromatic light HC mini-LEDs were used to analyze the values of the direct bilirubin test papers for different concentrations. The color compositions were represented by the average grayscale values, and the linear relationship between the direct bilirubin concentration and RGB average grayscale was analyzed through R2. Step 4 involves comparing the image analysis results captured by the four HC mini-LEDs. The influence of direct bilirubin detection on linearity was compared using different mini-LED light sources. We analyzed the influence of RGB monochromatic and cool white light LEDs on the detection of direct bilirubin.
3 Results and discussion
3.1 Light source and experimental architecture
In this study, four HC mini-LEDs were used as detection light sources, and their normalized spectra are shown in Fig. 3.
The experimental architecture for capturing images with an iPhone 8 Plus is shown in Fig. 4. The smartphone and light sources are fixed with brackets, and the test strip and lens are fixed at a 90° angle for image capture. H1 and H2 are the distances from the camera lens to test paper sample and light source to the test paper, respectively. The values of H1 and H2 are 10 cm and 15 cm, respectively.
3.2 Image detection of bilirubin test paper using cool white LEDs as light source
In this study, cool white LEDs were used as the light source to detect the direct bilirubin concentration from the test paper. The light source current and voltage were 10 mA and 2.7 V, respectively. Direct bilirubin test papers containing samples with 12 different concentrations ranging from 0.1 to 2.0 mg/dL were studied. In the sample picture shown in Fig. 5a, the camera ISO is set to 25. It is found from the actual sample picture that when the concentration of direct bilirubin is higher, the color of the pattern is darker. This is because higher concentration of bilirubin produces greater proportion of biliverdin, and as the biliverdin deposited on the test paper increases, the color changes. Once the captured images are divided into their R, G, and B channels, MATLAB is used for image processing to analyze the average R, G, and B grayscale values. The average grayscale and direct bilirubin concentration were considered for the linearity analysis. The bilirubin test paper values in the R, G, and B channels corresponded to direct bilirubin R2 values of 0.8912 (y = − 21.373x + 142.28), 0.9289(y = − 11.972x + 166.73), and 0.9001 (y = − 14.932x + 207.43), respectively. The experimental results showed that when using cool white LEDs as the light sources, the best linear relationship was observed for the G channel, where the direct bilirubin concentration range was 0.1–2.0 mg/dL, and the linear equation, R2, and LOD for the G channel were y = − 11.972x + 166.73, 0.9289, and 0.567 mg/dL, respectively, as shown in Fig. 5b.
3.3 Image detection of bilirubin test paper using red, green, and blue HC mini-LEDs as light source
The current, voltage, and FWHM of the red HC mini-LED were 30 mA, 1.9 V, and 21 nm, respectively. The direct bilirubin test paper results are shown in Fig. 6a, with the camera ISO set to 200. When using the red HC mini-LED, the best linear relationship was in the R channel when the direct bilirubin concentration range was 0.1–2.0 mg/dL, and the linear equation, R2, and LOD were y = − 39.708x + 250.92, 0.9365, and 0.534 mg/dL, respectively, as shown in Fig. 6b.
The current, voltage, and FWHM of the green HC mini-LED were 30 mA, 2.8 V, and 38 nm, respectively. The direct bilirubin test paper results are shown in Fig. 7a, with the camera ISO set to 100. When using the green HC mini-LED, the best linear relationship was in the G channel when the direct bilirubin concentration range was 0.1–2.0 mg/dL, and the linear equation, R2, and LOD in the G channel were y = − 15.669x + 240.34, 0.9523, and 0.459 mg/dL, respectively, as shown in Fig. 7b.
Finally, using the blue HC mini-LED as the light source, the current was 5 mA, voltage was 2.6 V, and FWHM was 30 nm. The direct bilirubin test paper results are shown in Fig. 8a, with the camera ISO set to 25. When using the blue HC mini-LED as the light source, the best linear relationship was in the B channel for a concentration range of direct bilirubin of 0.1–2.0 mg/dL. In the B channel, the linear equation, R2, and LOD were y = − 16.253x + 174.29, 0.9224, and 0.595 mg/dL, respectively, as shown in Fig. 8b.
Table 1 shows the specifications of the four types of HC mini-LEDs used and the analytical test paper results. The test paper images were captured with 178,101 pixels, and the RGB average grayscale was taken to analyze the linear relationship. We compared the cool white and monochromatic LEDs for detection of the direct bilirubin test papers in the concentration range of 0.1–2.0 mg/dL. The R2 detected by the monochromatic LEDs increased from 0.8912, 0.9289, and 0.9001 to 0.9364, 0.9523, and 0.9224 in the R, G, and B channels, respectively. These results show that the linearity of the bilirubin test paper could be improved when monochromatic light LEDs are used as light sources for image analysis. Moreover, they show that the detection result of the green LED was the best, and its LOD was 0.459 mg/dL. The R2 values for Figs. 5, 6, 7 and 8 are 0.9289, 0.9365, 0.9523, and 0.9224, respectively. We analyzed the three RGB channels after converting them to grayscale for analysis. The main source of error in this process is the uniformity of coloration in the test paper. In order to reduce this error, we have also standardized the reaction time and size of the test paper, minimizing the magnitude of this discrepancy as much as possible.
4 Conclusion
In this study, we propose image spectral analysis with HC mini-LEDs as the light sources for quantitative detection of direct bilirubin. Redox-labeled bilirubin was used to capture images with a smartphone, and image processing was used to perform RGB analysis to achieve quantitative detection of images of the direct bilirubin test paper. The experimental results showed that when green HC mini-LED was used as the light source and the average RGB grayscale of the image was analyzed in the concentration range of 0.1–2.0 mg/dL, the bilirubin concentration and G channel had the best linear relationship. The linear equation, R2, and LOD were y = − 15.669x + 240.34, 0.9523, and 0.459 mg/dL, respectively. Moreover, it was possible to quantitatively analyze direct bilirubin concentrations higher than 1.530 mg/dL. The proposed analysis method has the advantages of noninvasiveness, fast detection, and quantitative numerical analysis, and the detected data can be used to more accurately to analyze changes in the patient’s direct bilirubin index so as to provide a reference for hospitals to achieve their telemedicine goals. This study proposes Spectral Analysis with Highly Collimated Mini-LEDs as Light Sources for Quantitative Detection of Direct Bilirubin powder. Real human direct bilirubin detection has not been conducted yet, and future steps involve conducting human trials in collaboration with medical institutions.
Data availability
The data presented in this study are available on request from the all authors.
Abbreviations
- Mini-LEDs:
-
Mini-light-emitting diodes
- BOD:
-
Bilirubin oxidase
- LOD:
-
Limit of detection
- CoA:
-
Certificate of analysis
- FWHM:
-
Full width at half maximum
- R2 :
-
Correlation of determination
References
Kwo PY, Cohen SM, Lim JK. ACG clinical guideline: evaluation of abnormal liver chemistries. Am J Gastroenterol. 2017. https://doi.org/10.1038/ajg.2016.517.
Hooda V, Gahlaut A, Gothwal A, Hooda V. Bilirubin enzyme biosensor: potentiality and recent advances towards clinical bioanalysis. Biotechnol Lett. 2017. https://doi.org/10.1007/s10529-017-2396-0.
Hamoud AR, Weaver L, Stec DE, Hinds TD Jr. Bilirubin in the liver-gut signaling axis. Trends Endocrinol Metab. 2018. https://doi.org/10.1016/j.tem.2018.01.002.
Yao Q, Chen R, Ganapathy V, Kou L. Therapeutic application and construction of bilirubin incorporated nanoparticles. J Control Release. 2020. https://doi.org/10.1016/j.jconrel.2020.08.054.
Sokol RJ, Shepherd RW, Superina R, Bezerra JA, Robuck P, Hoofnagle JH. Screening and outcomes in biliary atresia: summary of a National Institutes of Health workshop. Hepatology. 2017. https://doi.org/10.1002/hep.21790.
Cayabyab R, Ramanathan R. High unbound bilirubin for age: a neurotoxin with major effects on the developing brain. Pediatr Res. 2019. https://doi.org/10.1038/s41390-018-0224-4.
Bell JG, Mousavi MPS, Abd El-Rahman MK, Tan EKW, Homer-Vanniasinkam S, Whitesides GM. Paper-based potentiometric sensing of free bilirubin in blood serum. Biosens Bioelectron. 2019. https://doi.org/10.1016/j.bios.2018.10.055.
Hansen TWR, Wong RJ, Stevenson DK. Molecular physiology and pathophysiology of bilirubin handling by the blood, liver, intestine, and brain in the newborn. Physiol Rev. 2020. https://doi.org/10.1152/physrev.00004.2019.
Olusanya BO, Kaplan M, Hansen TWR. Neonatal hyperbilirubinaemia: a global perspective. Lancet Child Adolesc Health. 2018. https://doi.org/10.1016/s2352-4642(18)30139-1.
Brits H, Adendorff J, Huisamen D, Beukes D, Botha K, Herbst H, Joubert G. The prevalence of neonatal jaundice and risk factors in healthy term neonates at National District Hospital in Bloemfontein. Afr J Prim Health Care Fam Med. 2018. https://doi.org/10.4102/phcfm.v10i1.1582.
Scottoni F, Davenport M. Biliary atresia: potential for a new decade. Semin Pediatr Surg. 2020. https://doi.org/10.1016/j.sempedsurg.2020.150940.
Ortiz-Perez A, Donnelly B, Temple H, Tiao G, Bansal R, Mohanty SK. Innate Immunity and Pathogenesis of Biliary Atresia. Front Immunol. 2020. https://doi.org/10.3389/fimmu.2020.00329.
Sokol RJ, Mack C, Narkewicz MR, Karrer FM. Pathogenesis and outcome of biliary atresia: current concepts. J Pediatr Gastroenterol Nutr. 2003. https://doi.org/10.1097/00005176-200307000-00003.
Ebbesen F, Vandborg PK, Donneborg ML. The effectiveness of phototherapy using blue-green light for neonatal hyperbilirubinemia—Danish clinical trials. Semin Perinatol. 2021. https://doi.org/10.1016/j.semperi.2020.151358.
Watchko JF, Maisels MJ. The enigma of low bilirubin kernicterus in premature infants: Why does it still occur, and is it preventable? Semin Perinatol. 2014. https://doi.org/10.1053/j.semperi.2014.08.002.
Xiao W, Zhi D, Pan Q, Liang Y, Zhou F, Chen Z. A ratiometric bilirubin sensor based on a fluorescent gold nanocluster film with dual emissions. Anal Methods. 2020. https://doi.org/10.1039/d0ay01781g.
Dhungana N, Morris C, Krasowski MD. Operational impact of using a vanadate oxidase method for direct bilirubin measurements at an academic medical center clinical laboratory. Pract Lab Med. 2017. https://doi.org/10.1016/j.plabm.2017.05.004.
Ponhong K, Teshima N, Grudpan K, Motomizu S, Sakai T. Simultaneous injection effective mixing analysis system for the determination of direct bilirubin in urinary samples. Talanta. 2011. https://doi.org/10.1016/j.talanta.2011.09.048.
Tan W, Zhang L, Doery JCG, Shen W. Study of paper-based assaying system for diagnosis of total serum bilirubin by colorimetric diazotization method. Sens Actuators B Chem. 2020. https://doi.org/10.1016/j.snb.2019.127448.
Thangavel B, Lingagouder J, Berchmans S, Ganesh V. Electrochemical grafting of 4-aminobenzoic acid onto toray carbon—interfacial investigation and fabrication of non-enzymatic bilirubin sensor. Sens Actuators B Chem. 2021. https://doi.org/10.1016/j.snb.2021.130292.
Yan D, Domes C, Domes R, Frosch T, Popp J, Pletz MW, Frosch T. Fiber enhanced Raman spectroscopic analysis as a novel method for diagnosis and monitoring of diseases related to hyperbilirubinemia and hyperbiliverdinemia. Analyst. 2016. https://doi.org/10.1039/c6an01670g.
Mano N, Edembe L. Bilirubin oxidases in bioelectrochemistry: features and recent findings. Biosens Bioelectron. 2013. https://doi.org/10.1016/j.bios.2013.07.014.
Rawal R, Kharangarh PR, Dawra S, Bhardwaj P. Synthesis, characterization and immobilization of bilirubin oxidase nanoparticles (BOxNPs) with enhanced activity: application for serum bilirubin determination in jaundice patients. Enzyme Microb Technol. 2021. https://doi.org/10.1016/j.enzmictec.2020.109716.
Okada H, Kawada K, Itoh S, Ozaki M, Kakutani I, Arai T, Koyano K, Yasuda S, Iwase T, Murao K, Kusaka T. Effects of bilirubin photoisomers on the measurement of direct bilirubin by the bilirubin oxidase method. Ann Clin Biochem. 2018. https://doi.org/10.1177/0004563217716474.
Batra B, Lata S, Rana JS, Pundir CS. Construction of an amperometric bilirubin biosensor based on covalent immobilization of bilirubin oxidase onto zirconia coated silica nanoparticles/chitosan hybrid film. Biosens Bioelectron. 2013. https://doi.org/10.1016/j.bios.2012.12.034.
Rahman MA, Lee KS, Park DS, Won MS, Shim YB. An amperometric bilirubin biosensor based on a conductive poly-terthiophene-Mn(II) complex. Biosens Bioelectron. 2008. https://doi.org/10.1016/j.bios.2007.09.005.
Chauhan N, Rawal R, Hooda V, Jain U. Electrochemical biosensor with graphene oxide nanoparticles and polypyrrole interface for the detection of bilirubin. RSC Adv. 2016. https://doi.org/10.1039/c6ra15671a.
Pan X, Li L, Lin H, Tan J, Wang H, Liao M, Chen C, Shan B, Chen Y, Li M. A graphene oxide-gold nanostar hybrid based-paper biosensor for label-free SERS detection of serum bilirubin for diagnosis of jaundice. Biosens Bioelectron. 2019. https://doi.org/10.1016/j.bios.2019.111713.
Nandi S, Biswas S. A recyclable post-synthetically modified Al(iii) based metal-organic framework for fast and selective fluorogenic recognition of bilirubin in human biofluids. Dalton Trans. 2019. https://doi.org/10.1039/c9dt01180c.
Xia C, Cao M, Xia J, Zhou G, Jiang D, Zhang D, Wang J, Li H. An ultrafast responsive and sensitive ratiometric fluorescent pH nanoprobe based on label-free dual-emission carbon dots. J Mater Chem C. 2019. https://doi.org/10.1039/c8tc05693e.
Anjana RR, Anjali Devi JS, Jayasree M, Aparna RS, Aswathy B, Praveen GL, Lekha GM, Sony G. S, N-doped carbon dots as a fluorescent probe for bilirubin. Mikrochim Acta. 2017. https://doi.org/10.1007/s00604-017-2574-8.
Aparna RS, JS AD, John N, Abha K, Syamchand SS, George S. Blue emitting copper nanoclusters as colorimetric and fluorescent probe for the selective detection of bilirubin. Spectrochim Acta A Mol Biomol Spectrosc. 2018. https://doi.org/10.1016/j.saa.2018.03.045.
Zhang M, Xu L, Ma Q, Yu H, Fang H, Lin Z, Zhang Q, Chen Z. A pH-controlled kit for total and direct bilirubin built on mimetic peroxidase CoFe2O4-DOPA-catalyzed fluorescence enhancement. ACS Appl Mater Interfaces. 2018. https://doi.org/10.1021/acsami.8b17554.
Jayasree M, Aparna RS, Anjana RR, Anjali Devi JS, John N, Abha K, Manikandan A, George S. Fluorescence turn on detection of bilirubin using Fe (III) modulated BSA stabilized copper nanocluster; a mechanistic perception. Anal Chim Acta. 2018. https://doi.org/10.1016/j.aca.2018.05.026.
Ahmmed E, Mondal A, Sarkar A, Chakraborty S, Lohar S, Saha NC, Dhara K, Chattopadhyay P. Bilirubin quantification in human blood serum by deoxygenation reaction switch-triggered fluorescent probe. ACS Appl Bio Mater. 2020. https://doi.org/10.1021/acsabm.0c00170.
Rizvi MR, Alaskar FM, Albaradie RS, Rizvi NF, Al-Abdulwahab K. A novel non-invasive technique of measuring bilirubin levels using bilicapture. Oman Med J. 2019. https://doi.org/10.5001/omj.2019.05.
Miah MMM et al. Non-invasive bilirubin level quantification and jaundice detection by sclera image processing. In: 2019 IEEE global humanitarian technology conference; 2019. https://doi.org/10.1109/GHTC46095.2019.9033059
Parnianchi F, Kashanian S, Nazari M, Peacock M, Omidfar K, Varmira K. Ultrasensitive electrochemical sensor based on molecular imprinted polymer and ferromagnetic nanocomposite for bilirubin analysis in the saliva and serum of newborns. Microchem J. 2022. https://doi.org/10.1016/j.microc.2022.107474.
AuYoung B, Gutha Ravichandran A, Patel D, Dave N, Shah A, Wronko-Stevens B, Bettencourt F, Rajan R, Menon N. A low-cost paper-based device for the colorimetric quantification of bilirubin in serum using smartphone technology. Front Chem. 2022. https://doi.org/10.3389/fchem.2022.869086.
Bui DA, Hauser PC. Analytical devices based on light-emitting diodes–a review of the state-of-the-art. Anal Chim Acta. 2015. https://doi.org/10.1016/j.aca.2014.09.044.
Tan W, Zhang L, Doery JCG, Shen W. Three-dimensional microfluidic tape-paper-based sensing device for blood total bilirubin measurement in jaundiced neonates. Lab Chip. 2020. https://doi.org/10.1039/c9lc00939f.
Chu S, Wang H, Ling X, Yu S, Yang L, Jiang C. A portable smartphone platform using a ratiometric fluorescent paper strip for visual quantitative sensing. ACS Appl Mater Interfaces. 2020. https://doi.org/10.1021/acsami.9b20458.
Xin L, Jin H, Tu Y, Yuan Z, Lv Z, Wang C. Defect detection and characterization of RTV silicone rubber coating on insulator based on visible spectrum image. IEEE Trans Power Deliv. 2020. https://doi.org/10.1109/tpwrd.2020.2995071.
Acknowledgements
This work was supported by NSTC 112-2622-E-194-004 and Ditmanson Medical Foundation Chia-Yi Christian Hospital.
Funding
This work was supported by This work was supported by the National Science and Technology Council of Taiwan (NSTC 112-2622-E-194-004).
Author information
Authors and Affiliations
Contributions
ZTY and CWT are responsible for the structure and conception of the entire article. ZTY and SFT is responsible for the analyze measurement data of the article. ZTY, SXT, and CWT are responsible for the initial writing of the manuscript.
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
All authors consent to the publication of this manuscript.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ye, Z.T., Tseng, S.F., Tsou, S.X. et al. Spectral analysis with highly collimated mini-LEDs as light sources for quantitative detection of direct bilirubin. Discover Nano 19, 13 (2024). https://doi.org/10.1186/s11671-024-03957-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s11671-024-03957-2