Abstract
Fast neutron beams (E\(_n > \) 1 MeV) are of relevance for many scientific and industrial applications. This paper explores fast neutron production using a TANDEM accelerator at the Legnaro National Laboratories, via an energetic ion beam (90 MeV \(^{14}N\)) onto a lithium target. The high energy models for nuclear collision of FLUKA foresee large neutron yields for reactions of this kind. The experiment aimed at validating the expected neutron yields from FLUKA simulations, using two separate and independent set-ups: one based on the multi-foil activation technique, and the other on the time of flight technique, by using liquid scintillator detectors. The results of the experiment show clear agreement of the measured spectra with the FLUKA simulations, both in the shape and the magnitude of the neutron flux at the measured positions. The neutron spectrum is centered around the 8 MeV range with mild tails, and a maximum neutron energy spanning up to 50 MeV. These advantageous results provide a starting point in the development of fast neutron beams based on high energy ion beams from medium-sized accelerator facilities.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Neutron beams have important applications in many different fields. Most neutron beams are available at reactors and have cold or thermal spectra; such neutrons are mainly suited to the study of the structure and dynamics of materials at the atomic level. On the other hand, few facilities can provide beams with neutron energies greater than 1 MeV; the access is limited and costs are high.
Fast neutron sources are of importance not only for basic nuclear physics. The list of other fields of applications includes: dosimetry, neutron detector development, fast neutron oncology, radiation protection shielding materials for accelerator based oncology and Space missions, and the study of single-neutron induced effects in digital and power electronics.
Tandem accelerators are still widely used in many small-medium size laboratories and could provide cost effective neutrons on existing beam lines. In addition to direct applications to fundamental and applied physics mentioned above, important is also the possibility of training students in neutron physics.
2 Fast neutrons at the LNL Tandem
The Legnaro National Laboratories (LNL) of the Italian National Institute for Nuclear Physics is developing new sources of fast neutrons that would integrate the existing one at the CN 6 MV Van de Graaff accelerator and the one under design at the SPES 70 MeV proton cyclotron [1, 2]. To produce fast neutrons, we studied reactions involving heavy ion beams on light targets, as inverse kinematics can produce forwardly focused high energy neutrons in the laboratory reference frame. Due to the high velocity of the center of mass, neutrons are emitted at higher energy with respect to the direct kinematics. Moreover, this booster effect of the center of mass collimates the emitted products in the forward direction.
Energetic ions beams are available at the Tandem-PIAVE-ALPI accelerator complex [3]. The experiment described below used the maximum energy available at the Tandem XTU accelerator; a followup experiment will use the PIAVE-ALPI heavy ion injector and superconducting linac to deliver higher energy beams.
As a solid target material, the lightest possible elements to be considered are lithium or beryllium. With regards to the beam and the performance of the Tandem, a balance between ion mass and available current gives nitrogen as the best compromise.
The initial target choice is lithium, routinely used at the BELINA neutron facility at the CN accelerator; beryllium will soon be evaluated too.
2.1 The experimental reaction
The cross section and yield for the \(^7\)Li(\(^{14}\)N,xn)X reaction are not available in literature. We used the Monte Carlo simulations to study the neutron production spectra; the calculated neutron spectra are based on theoretical models since no data are available for the reaction.
Two independent measurement methods were used to measure the energy spectrum and validate the results of the simulations: the multi-activation foil technique and time-of-flight (ToF).
The ToF part of this experiment is actually part of a broad research program for characterizing Liquid Scintillators developed for Neutrino Physics and Rare Event Astrophysics. In general, a detailed understanding of neutron signals in detectors is crucial to construct realistic simulations and predictions of natural phenomena; this is the immediate application of this new neutron source at LNL.
2.2 FLUKA simulations
Monte Carlo simulations of the nuclear reaction were performed using the models available in FLUKA [4]. Evaporation including heavy ion fragments and coalescence hadronic models were activated in the simulations.
Figure 1 shows the expected neutron yield at zero and 10 degrees, corresponding to the two positions where the fluence was measured. The position of the activation foil was tilted 10\(^\circ \) from the beam axis and was located 2.5 cm from the lithium target, covering a solid angle of 0.193 sr. The position of the ToF detector was located 4.56 m away from the lithium target in the forward (0\(^\circ \)) beam direction, covering a solid angle of \(0.235\times 10^{-3}\) sr. The estimated yield at the activation position is lower than the one at the ToF detector position, confirming the strong forward angle dependence of the neutron production.
To study the presence of charged particles, an extensive simulation was made of the setup with the target thickness and Cu backing, including the cadmium cover for the activation foils at their position; we also included a scoring volume (a phantom detector) in the forward direction at a distance of 10 cm. The presence of protons, alphas, and ions was checked for at both the activation foil position and the scoring volume. Protons do arrive at these positions and they have high energy (so are hard to shield against), but they are a small contamination, amounting to about 1\(\%\) of the number of neutrons. On the other hand, there are almost no alphas while the ions are stopped in the copper.
For the activation foil measurement, this proton contamination does not significantly affect the determination of the neutron spectrum as the reaction products from (n,X) reactions are different from the reaction products of (p,X). For what concerns the ToF measurement, only protons with energies greater than \(\sim \) 30 MeV can make it from the target into the distant active scintillation volume; lower energy protons are stopped by intervening material (see Sect. 4.1). For such high energies, the detection efficiency for neutrons is 20\(\%\) (Fig. 14), therefore the proton contamination in the higher energy data sample is estimated, conservatively, to be \(\sim \)5\(\%\) (assuming full proton detection efficiency). For this reason, proton signal rejection was deemed unnecessary.
2.3 The beam
Negative Boron Nitride (BN) molecules were injected into the Tandem, accelerated by a 13.7 MV terminal voltage, and dissociated by one carbon stripping foil to produce positive N ions that were then further accelerated. Magnetic Analysis selected \(^{14}\)N\(^{+6}\) 90 MeV ions that were then transported along the 30\(^\circ \) beam line in Hall 1 (Fig. 2) towards the lithium neutron production target.
The beam was pulsed (800 ns repetition rate and a pulse time width sigma of 1 ns) in order to perform ToF measurements, thereby providing a cross-check of the simultaneous multi-foil activation measurement.
2.4 The neutron production target
The target uses metal lithium encapsulated into a copper backing beam stopper designed to have low mass and minimal material in order to have low neutron spectra perturbability in the forward direction.
The target system is produced in an inert atmosphere inside an argon glovebox. The lithium is attached to the 300 \(\upmu \)m thick hemispherical Cu backing by pressing them between male and female molds, made with a Computerized Numerical Control machine, thereby producing a lithium layer of the desired thickness, in this case 200 \(\upmu \)m, with an uncertainty smaller than ± 10\(\%\) (Fig. 3). This technique is used regularly to make thin lithium targets used to produce neutrons at the CN accelerator. By dedicated FLUKA simulations, this uncertainty in thickness responds to a variation of 6.5 % in the neutron yield.
Although the range of 90 MeV \(^{14}\)N ions in lithium is 400 \(\upmu \)m, we chose 200 \(\upmu \)m in order to proceed confidently (to avoid the risk of melting the Li in thicker targets) as we were particularly interested in the production of high energy neutrons. According to FLUKA simulations, the forward yield, normalized to the maximum possible value, depends quadratically on the lithium thickness: at 90 MeV the yield from a 200 \(\upmu \)m target of lithium is about 70\(\%\) of that obtainable by completely stopping the beam in lithium (Figs. 4 and 5). At this energy, the contribution from the copper backing is an additional 2–3\(\%\); the neutrons produced by the copper are present along the whole energy range, but are only appreciable on the low energy side of the most probable energy (8 MeV), and even there they are less than a few percent of that part of the spectrum.
The copper backing is electrically insulated from the beam line. Using a picoammeter,Footnote 1 the current at the target is acquired directly from the copper backing. A two-stage tantalum collimator of 5 mm diameter is placed 10 cm upstream of the copper backing and the current delivered on it was also acquired. Figure 6 shows the time-dependent ion current on the copper backing during the irradiation; the total charge accumulated during the experiment amounted to 0.153 mC; the error on the integrated charge is negligible. During the experiment, the current on the copper backing was between 0.5\(-\)1.5 nA; for 100 consecutive readings (once a second), the current values were normally distributed with a 0.02 nA sigma, corresponding to a 1–3\(\%\) uncertainty.
3 The multi-foil activation technique
The activation technique is a well-established and very powerful method that can be used to measure neutron spectra. The main goal with this setup was to obtain broad spectral information including the low energy range below few MeV and especially to provide an absolute value for the yield at the measured position.
3.1 Experimental setup
For the multi-foil activation measurement, a series of samples especially sensitive to neutrons in the energy range of interest were selected (i.e., with large cross sections or with thresholds ranging from hundreds of keV to tens of MeV). A list of these activation foils and the reactions used are shown in Table 1. The activation foils were purchased from ShieldWerx [5]. These samples were arranged in a stack configuration and were surrounded with two cadmium covers, to minimize the activation due to low energy neutrons in the Au and In foils. The particular ordering in the arrangement of samples was chosen based on the relative attenuation or scattering produced by each of the samples within the stack. These alterations in the neutron flux at each position were later corrected via dedicated Monte Carlo simulations using MCNP6.2 [6]. Figure 7 shows the correction factor to the flux for each of the samples used for the unfolding of the spectrum. Additional samples of Au were placed outside the Cd covers (Au-a sample) and in the last position (Au-c sample) and used for cross-checking the thermal contamination and possible scattering back from the experimental hall. However, these samples were not included in the unfolding of the neutron spectrum.
Figure 8 shows (top) the aluminum support and holder at its tilted position during the experiment; and (bottom) a detail of the sample holder. The stack of samples was mounted on an aluminum support and placed 2.5 cm from the center of the lithium target. The aluminum support was perforated near the tip of the holder so that the stack could fit in there. The samples were secured in that position using a fine layer of kapton foil. The samples were tilted by 10\(^\circ \) with respect to the beam axis in order to reduce the interference with the measurements by the ToF detectors located in the forward direction at 0\(^\circ \) in a farther position. Given the close distance between the neutron source and the samples, the difference in solid angle covered by the first and last samples in the stack was not negligible. This effect was also incorporated into the unfolding by correcting the activation depending on their distance to the source.
3.2 Activation measurements
The activity of each isotope produced by irradiation of the samples was measured using Gamma spectroscopy with a High Purity Germanium (HPGe) detector. It was placed inside a lead well and cooled every 12 h with liquid nitrogen. A shaping amplifier ORTEC 672 was used and pulses were digitized with an ORTEC 928 MCB and the MAESTRO 7.01 DAQ. The efficiency \(\varepsilon (E_\gamma )\) of the HPGe detector was calibrated with several certificated radioactive gamma sources (22Na, 57Co, 60Co, 88Y, 133Ba and 137Cs) for that purpose. Several positions at different distances (3–15 cm) between the detector and samples were used to keep a balance between detection efficiency and dead-time losses. A dedicated low perturbating sample holder has been designed and 3D printed to guarantee the accuracy of the efficiency calibration and measurements. The uncertainty of the efficiency calculation was estimated to be 3%, dominated by the calibration of the gamma sources. Figure 9 shows the setup of the gamma spectroscopy station. MCNP simulations of the detector geometry (HPGe including a detection dead layer plus Al encapsulation) were carried out to account for the different sizes of the calibration radioactive sources and the measured irradiated samples, obtaining correction factors to the efficiency, \(\kappa _\varepsilon \). The low activity of the irradiated samples impelled in most cases to close the distance down to 3 cm to increase the count rates. Even if generally small, dead time corrections were taken into account. The self-shielding correction factor, \(\kappa _s\), was also computed in dedicated simulations for every gamma line of each sample. The combined effect of both \(\kappa _\varepsilon \) and \(\kappa _s\) are shown in Fig. 10 for the gamma lines used in the analysis.
After the end of irradiation, a schedule of measurements was followed in order to obtain the activity data with the best precision, given the time constraints. Special regards had to be considered given the relatively short half-lives of some of the produced isotopes, which were assessed by measuring some of the samples jointly (e.g., Al (containing \(^{24}\)Na), In (containing \(^{110}\)In, \(^{115m}\)In and \(^{116m}\)In) and Co (containing \(^{56}\)Mn). Figure 11 shows some of the measured spectra for these samples.
The samples were measured after some cooling time, \(t_c\), from the end of irradiation (EOI). The activity of isotope k at the end of irradiation, \(A_{k}^{EOI}\), was extrapolated from the measured counts in a certain gamma line, \(C_\gamma \), during the measuring time, \(t_m\) using the following expression:
where energy \(E_\gamma \), relative intensities of the gamma rays \(n_\gamma \), and the decay constants \(\lambda _k\) of the produced isotopes were taken from ENSDF [7]. At least one gamma line was used for each reaction. More than one was employed when available. An average value between the activity values from each gamma line was used in these cases.
3.3 Unfolding routine
In order to recover the spectral information from the activity of the samples, an unfolding procedure has to be performed. In the case of a steady-state neutron field, the activation rate equals the decay rate after several half-lives of the produced isotope. Therefore the saturated activity, \(A_S\), can be used directly as input to the unfolding (\(A_S = n\sigma \dot{\varPhi }\)). However, if the neutron flux is not stable in time (or not enough compared to the half-lives of the isotopes of interest), a correction factor has to be applied. In this sense, instead of keep using the saturated activity outside of its true definition, the total number of activations per atom in the samples, R/n, can be used. This in turn can be done by integrating in time (using the information from the irradiation history). This leads to an activations-to-fluence relation in order to perform the deconvolution:
In a real experiment, the fluence measured by a foil, \(\tilde{\varPhi }\), is not exactly the true fluence in that position, \(\varPhi \), due to multiple scattering and self-shielding effects (and other effects from the surrounding support and other samples in the case of a stack), therefore, a correction factor \(\kappa \) has to be introduced to the response function, which is dominated by the cross-section:
The energy spectrum can be subsequently divided into several energy groups, \(\varPhi _j\), that provide information on the shape of the spectrum and can be chosen depending on the relative contribution to the activations on sample i, \(\left( \frac{R}{n}\right) _i\), mediated by the response function, M:
Under this approach, the response matrix includes an effective cross-section within each energy group, \(\sigma _{ij}\), that has to be calculated and included in the response function as:
For the computation of the response function, cross-section data spanning from the low energy range up to above tens of MeV is needed. For this reason, the TENDL-19 extended cross-section evaluation database was used (including up to 60 MeV or 200 MeV depending on the isotope and reaction) [8]. In the cases where the cross-section data was not available above 60 MeV, linear extrapolation was applied. These matrix elements are thus dependent on the intra-group shape of the neutron spectrum. The selection of any spectrum for the calculation of this matrix impacts the final unfolding, which should be chosen carefully. This impact can be assessed by performing variations in the guess spectrum and including that in the unfolded flux uncertainty.
In order to perform the unfolding of the spectrum, several algorithms are available, including iterative methods such as the SAND-II code [9]. Lately, there have been efforts to use other methods to manage the uncertainties more comprehensively. One such strategy uses a Bayesian unfolding routine coupled to Markov Chain Monte Carlo simulations (MCMC) using the code JAGS [10]. This methodology is based on the one presented by Chiesa et al. [11, 12]. Different sources of uncertainty can be handled, including the uncertainty in the experimental data from the activation of the samples, the aforementioned uncertainty in the spectrum shape, and the uncertainties in the cross-section data, which prove to be very relevant in the MeV energy range. This approach has recently been used to unfold the spectrum of the newly built NEAR station of n_ToF at CERN [13].
4 The ToF technique
To provide a complementary and separate measurement of the spectral shape with higher energy resolution a fast liquid scintillation detector was deployed in the neutron beam to allow ToF measurements, especially for the high energy part of the spectrum. In the following paragraph, both the detector used and the associated electronics are briefly introduced.
4.1 Experimental setup
A custom-made liquid scintillator detector was used for the measurements presented here. Figure 12 shows a conceptual drawing of this device. A 3x3 inch cylindrical cell of borosilicate glass with 1 mm thin optical windows was filled with a Linear alkylbenzene (LAB) based scintillation cocktail with admixtures of 2.5 g/l PPO (2,5-Diphenyloxazole) and 3 mg/l BisMSB (1,4-bis-(o-methylstyryl)-benzene). LAB was chosen as the detector medium for safety reasons, due to its low toxicity and high flashpoint (\(\sim \) 140 \(^{\circ }\)C). To protect the scintillator from oxygen, it was previously flushed extensively with high-purity nitrogen. After filling, the glass cell was sealed gas-tight under moderate overpressure (a few mbar) of nitrogen. To enhance the light collection efficiency the mantle surface of the cylindrical cell was previously covered by a directly evaporated aluminum mirror protected by a layer of plasma polymer. This vessel is coupled to two fast 3 inch 9821B photomultiplier tubes (PMTs) housed in mu-metal shields provided by ET Enterprises [14]. The detector was housed in a thin-walled, fully enclosed aluminum dark box that serves as both a light shield and a Faraday cage against electromagnetic noise. This assembly was placed in the beam axis in a distance of 4.76 m behind the Li-target (distance with respect to liquid scintillators (LS) target cell center).
The readout electronics of the ToF setup are based on a staged coincidence circuit. The first coincidence stage is realized between the two PMT branches and the second between the output of the first stage and the discriminated RF signal of the accelerator’s beam buncher. The block diagram of the used electronics can be found in Fig. 13. For ToF measurements described here, the PMT signals are fed to a CAEN N978 10x Fast Amplifier [15]. While one of the two outputs of each stage is directly connected to the waveform digitizer Agilent U1065A Acqiris DC282 (ADC), the other output is connected to a CAEN N844 LTD (Low Threshold Discriminator) [16].
The sinusoidal RF Signal used for the XTU Tandem’s Beam Buncher is discriminated on the falling edge of the negative half-wave by a CAEN N842 CFD (Constant Fraction Discriminator) [17]. The resulting logic signal is multiplied by a CAEN N625 Fan-In Fan-Out [18] and sent to the ADC on the one hand and to the second stage of the double coincidence module on the other hand. In case of a coincident event in both PMTs in coincidence with the buncher signal, the waveform digitizer is triggered. While the LS detector reaches an internal time resolution of \(\sigma _{LS}\) = (\(320\pm 5\)) ps for energy depositions of \(\sim \)300 keV\(_{ee}\), the buncher of the XTU tandem was capable of producing beam bunches with a nearly gaussian shape and a width of \(\sigma _{Bunch}\) = (\(1.08\pm 0.02\)) ns. Therefore, the beam properties are fully dominating the time and with that the energy resolution of the ToF spectrum.
4.2 Unfolding routine
The efficiency and the ToF-to-energy resolution function of the ToF detection setup were estimated from simulations with the MCNP6.2 code [6]. Several simulations with monoenergetic neutron beams were carried out to compute the total detection efficiency, shown in Fig. 14 for the energy range covered in the experiment. Also, time discretization in the experimental time window of 200 ns was included in the simulations in order to fill the resolution function matrix, as shown in Fig. 15. The resolution function matrix provides a means to convert ToF distributions to neutron energy spectra, including the total flight path from the neutron source to the actual detection inside the liquid scintillator. Also, the time width of the pulse was reproduced in the simulations by using the distribution observed in the detection of the \(\gamma \)-flash. For the calculations of the efficiency and resolution function, elastic scattering of neutrons by hydrogen and charged particle reactions from both hydrogen and carbon were included. Elastic scattering on carbon was not included due to the heavy quenching that prevented the detection of the carbon recoils. The unfolding routine to convert ToF to energy spectra was a bayesian-based iterative algorithm using the maximum likelihood expectation-maximization (MLEM) method. This method has been previously used for comparable applications, for instance the unfolding of \(\beta \)-decay total absorption spectra [19] and for \(\gamma \)-ray spectroscopy decomposition [20].
5 Results
5.1 Results from activation and ToF measurements
The activation produced during the irradiation was measured via Gamma spectroscopy, from which the activity at the end of irradiation, \(A^{EOI}\), was inferred. Table 2 lists the \(A^{EOI}\) for each sample and reaction. Subsequent spectral results from the analysis using the activation unfolding method and the total neutron yield are shown in the following section, compared to the FLUKA simulations.
The experimental ToF spectrum measured with the LS detectors is shown in Fig. 16. This ToF spectrum includes the \(\gamma \)-flash from the target and the neutron distribution. The long ToF tail is altered due to event missing in triggering due to the fast bunch rate used in the experiment. In short, we had to make the gates shorter given that the beam frequency was high. Also there was an arbitrary shift between the gate for the RF (Buncher) and the gate from the detector, which resulted in a limited and reduced time window for the triggering. In addition to that, the end of the gate actually has an influence on the trigger efficiency there, thus altering the tail of the spectrum. This made us reject the low energy tail in the reconstruction. Also, this part of the experiment lacked a concise and accurate determination of the ion current used to produce the spectrum, and therefore only the shape of the spectrum is retained. The unfolded energy spectrum is shown in the next section.
5.2 Comparison between FLUKA simulations and the experimental data
Table 3 reports the measured neutron flux from unfolding at the activation foils position, compared with the FLUKA simulation; the unfolded neutron spectrum is shown in Fig. 17. The integral neutron flux of \((4.04\pm 0.42) \times 10^{5}\) n/cm\(^2\)/s/pnA at that position, confirms the estimates by FLUKA ( \((3.79\pm 0.10) \times 10^{5}\) n/cm\(^2\)/s/pnA), in good agreement within uncertainties. The unfolded spectrum obtained from activation data is shown in blue in Fig. 17. Five energy groups above 0.1 MeV show good agreement both in shape and in the absolute value of the production yield.
Moreover, the unfolded data from the ToF measurement, covering the high energy part of the spectrum (above 8 MeV), shows a spectral shape with increased energy resolution and also in excellent agreement with the simulation. The low energy limit for the reconstruction from the ToF spectrum was set at 8 MeV due to the presence of artifacts from the unfolding, which surpassed the attainable uncertainties in that range. The results of the unfolding are shown also in Fig. 17 in comparison with the activation and simulated data. The spectral data from ToF has been scaled to the activation data in that range as this part of the experiment did not include an independent normalization.
6 Prospects
At LNL, the ALPI booster can be used to reach higher beam energies: according to FLUKA simulations, the energy spectrum can be stiffened and the neutron yield increased by a factor 5 by raising both the beam energy up to 170 MeV and the lithium thickness to 1 mm (Fig. 18). We are scheduled to perform an ALPI version of this experiment in 2024 in order to verify the energy dependence, especially for the higher energy neutrons; the proton contamination will also be measured with a silicon diode.
Using thicker lithium will require modifying target fabrication and re-evaluating the cooling system. For this reason, we will also investigate the possibility of using a beryllium easier to install self-standing foil target instead of the lithium targets fabricated in a glovebox (described in Sect. 2.4).
Indeed beryllium has better physical properties (higher melting point, thermal conductivity, and lower linear thermal expansion coefficient) and it is stable in air making it much more manageable compared to lithium. However, the issue of radiation damage to the Be foil must be evaluated with great care: Be is toxic and the risk of polluting the beam line might be too great.
7 Summary
The \(^7\)Li(\(^{14}\)N,xn)X reaction is cost effective and easy to implement at Tandem machines. LNL can use this reaction to immediately produce fast neutrons in an interesting energy range while waiting for the construction and commissioning of a dedicated neutron irradiation facility at the SPES proton cyclotron. The experimental neutron energy distribution measured is broad and peaked at 8 MeV, ranging from hundreds of keV up to tens of MeV. Activation foils and ToF technique have been used to measure the produced energy spectra, showing a good agreement between them and the heavy ion fragmentation and coalescence models available in FLUKA.
Notes
Keithley 6485.
References
P. Mastinu, D. Bisello, R.A. Barrera, I. Porras, G. Prete, L. Silvestrin, J. Wyss, J. Neutron Res. 22(2–3), 233 (2020)
L. Silvestrin, D. Bisello, J. Esposito, P. Mastinu, G. Prete, J. Wyss, Eur. Phys. J. Plus 131(72) (2016). https://doi.org/10.1140/epjp/i2016-16072-0
C. Signorini, F. Cervellera, G. Bezzon, Nuclear instruments and methods in physics research section A: Accelerators. Spectrometer. Detect. Assoc. Equip. 244(1), 27 (1986). https://doi.org/10.1016/0168-9002(86)90731-X
C. Ahdida, D. Bozzato, D. Calzolari, F. Cerutti, N. Charitonidis, A. Cimmino, A. Coronetti, G. D’Alessandro, A. Donadon Servelle, L. Esposito, R. Froeschl, R. García Alía, A. Gerbershagen, S. Gilardoni, D. Horváth, G. Hugo, A. Infantino, V. Kouskoura, A. Lechner, B. Lefebvre, G. Lerner, M. Magistris, A. Manousos, G. Moryc, F. Ogallar Ruiz, F. Pozzi, D. Prelipcean, S. Roesler, R. Rossi, M. Sabaté Gilarte, F. Salvat Pujol, P. Schoofs, V. Stránský, C. Theis, A. Tsinganis, R. Versaci, V. Vlachoudis, A. Waets, M. Widorski, Front. Phys. 9(788253) (2022). https://doi.org/10.3389/fphy.2021.788253
Shieldwerx, a division of Bladewerx LLC. Activation foils. https://www.shieldwerx.com/activation-foils.html
C.J. Werner, J.S. Bull, C.J. Solomon, F.B. Brown, G.W. McKinney, M.E. Rising, D.A. Dixon, R.L. Martz, H.G. Hughes, L.J. Cox, A.J. Zukaitis, J.C. Armstrong, R.A. Forster, L. Casswell, MCNP6.2 Release Notes - report LA-UR-18-20808 (Los Alamos National Laboratory (LANL), 2018)
Evaluated Nuclear Structure Data Files (2023). https://www.nndc.bnl.gov/ensdf/. Last accessed 21 July 2023
A. Koning, D. Rochman, J.C. Sublet, N. Dzysiuk, M. Fleming, S. van der Marck, Nuclear Data Sheets 155, 1 (2019). https://doi.org/10.1016/j.nds.2019.01.002. Special Issue on Nuclear Reaction Data
P. Griffin, J. Kelly, J. VanDenburg, User’s Manual for SNL-SAND-II Code (1994). https://rsicc.ornl.gov/codes/psr/psr3/psr-345.html
M. Plummer, JAGS Version 4.3.0 user manual (2017). https://martynplummer.wordpress.com/2017/07/18/jags-4-3-0-is-released/
D. Chiesa, E. Previtali, M. Sisti, Ann. Nucl. Energy 70, 157 (2014). https://doi.org/10.1016/j.anucene.2014.02.012
D. Chiesa, M. Nastasi, C. Cazzaniga, M. Rebai, L. Arcidiacono, E. Previtali, G. Gorini, C.D. Frost, Nuclear instruments and methods in physics research section A: accelerators. Spectrom. Detect. Assoc. Equip. 902, 14 (2018). https://doi.org/10.1016/j.nima.2018.06.016
M.E. Stamati, P. Torres-Sánchez, P. Pérez-Maroto, M. Cecchetto, S. Goula, M. Mastromarco, S. Chasapoglou, C. Beltrami, D. Chiesa, A. Manna, R. Mucciola, N. Patronis, J. Praena, C. Guerrero, N. Colonna, A. Mengoni, C. Massimi, O. Aberle, V. Alcayne, S. Altieri, S. Amaducci, H. Amar Es-Sghir, J. Andrzejewski, V. Babiano-Suarez, M. Bacak, J. Balibrea, S. Bennett, A.P. Bernardes, E. Berthoumieux, D. Bosnar, M. Busso, M. Caamaño, F. Calviño, M. Calviani, D. Cano-Ott, A. Casanovas, D.M. Castelluccio, F. Cerutti, G. Cescutti, E. Chiaveri, P. Colombetti, P.C. Console Camprini, G. Cortés, M.A. Cortés-Giraldo, L. Cosentino, S. Cristallo, M. Di Castro, D. Diacono, M. Diakaki, M. Dietz, C. Domingo-Pardo, R. Dressler, E. Dupont, I. Durán, Z. Eleme, S. Fargier, B. Fernández-Domínguez, P. Finocchiaro, S. Fiore, V. Furman, F. García-Infantes, A. Gawlik-Ramiega, G. Gervino, S. Gilardoni, E. González-Romero, F. Gunsing, C. Gustavino, J. Heyse, D.G. Jenkins, E. Jericha, A. Junghans, Y. Kadi, T. Katabuchi, T. Katabuchi, I. Knapová, M. Kokkoris, Y. Kopatch, M. Krticka, D. Kurtulgil, I. Ladarescu, C. Lederer-Woods, J. Lerendegui-Marco, G. Lerner, T. Martínez, A. Masi, P. Mastinu, F. Matteucci, E.A. Maugeri, A. Mazzone, E. Mendoza, V. Michalopoulou, P.M. Milazzo, F. Murtas, E. Musacchio-Gonzalez, A. Musumarra, A. Negret, A. Oprea, J.A. Pavón-Rodríguez, M.G. Pellegriti, J. Perkowski, C. Petrone, L. Piersanti, E. Pirovano, S. Pomp, I. Porras, N. Protti, J.M. Quesada, T. Rauscher, R. Reifarth, D. Rochman, Y. Romanets, F. Romano, C. Rubbia, A. Sánchez, M. Sabaté-Gilarte, P. Schillebeeckx, D. Schumann, A. Sekhar, A.G. Smith, N.V. Sosnin, M. Spelta, G. Tagliente, A. Tarifeño-Saldivia, D. Taríro, N. Terranova, S. Urlass, S. Valenta, V. Variale, P. Vaz, D. Vescovi, V. Vlachoudis, R. Vlastou, A. Wallner, P.J. Woods, T. Wright, P. Zugec, n_TOF Collaboration. EPJ Web of Conf. 284, 06009 (2023). https://doi.org/10.1051/epjconf/202328406009
ET Enterprises Electron Tubes, 78 mm (3’’) photomultiplier 9821B series data sheet (ET Enterprises Limited, Uxbridge, United Kingdom, 2012)
CAEN, N978 - 4 Channel Variable Gain Fast Amplifier. CAEN, URL: https://www.caen.it/products/n978/ (2022)
CAEN, N844 - 8 Channel Low Threshold Discriminator. CAEN, URL: https://www.caen.it/products/n844/ (2022)
CAEN, N842 - 8 Channel Constant Fraction Discriminator. CAEN, URL: https://www.caen.it/products/n842/ (2022)
CAEN, N625 - Quad Linear FAN-IN FAN-OUT. CAEN, URL: https://www.caen.it/products/n625/ (2022)
J. Tain, D. Cano-Ott, Nuclear instruments and methods in physics research section A: accelerators. Spectrom. Detect. Assoc. Equip. 571(3), 728 (2007). https://doi.org/10.1016/j.nima.2006.10.098
J.T. Matta, A.J. Rowe, M.P. Dion, M.J. Willis, A.D. Nicholson, D.E. Archer, H.H. Wightman, IEEE Trans. Nucl. Sci. 69(6), 1212 (2022). https://doi.org/10.1109/TNS.2022.3162986
Acknowledgements
This work has been supported by the Cluster of Excellence PRISMA\(^{+}\), the Cluster of Excellence ORIGINS as well as the Collaborative Research Center “Neutrinos and Dark Matter in Astro- and Particle Physics” (SFB 1258) and the University of Granada Chair “Neutrons for Medicine”. Moreover, we are grateful for the support from the DFG Research Unit JUNO 2319 and 5519 (Project Number: 498394246), the Spanish Junta de Andalucía (Projects FEDER no. A-FQM-371-UGR18 and B-FQM-156-UGR20) and Fundación Científica AECC. I.P. acknowledges financial support from the Spanish Ministerio de Ciencia, Innovación y Universidades for his research stay at LNL under contract PRX18/00551. Furthermore, we would like to thank Francisco Ogallar Ruiz for his helpful advice and suggestions on the FLUKA simulations.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Code availability
This manuscript has no associated code/software. [Authors’ comment: Code/Software sharing not applicable to this article as no code/software was generated or analysed during the current study.]
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Funded by SCOAP3.
About this article
Cite this article
Torres-Sánchez, P., Steiger, H.T.J., Mastinu, P. et al. Fast neutron production at the LNL Tandem from the \(^7\)Li(\(^{14}\)N,xn)X reaction. Eur. Phys. J. C 84, 372 (2024). https://doi.org/10.1140/epjc/s10052-024-12658-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1140/epjc/s10052-024-12658-0