Abstract
State-of-the-art physics experiments require high-resolution, low-noise, and low-threshold detectors to achieve competitive scientific results. However, experimental environments invariably introduce sources of noise, such as electrical interference or microphonics. The sources of this environmental noise can often be monitored by adding specially designed “auxiliary devices” (e.g. microphones, accelerometers, seismometers, magnetometers, and antennae). A model can then be constructed to predict the detector noise based on the auxiliary device information, which can then be subtracted from the true detector signal. Here, we present a multivariate noise cancellation algorithm which can be used in a variety of settings to improve the performance of detectors using multiple auxiliary devices. To validate this approach, we apply it to simulated data to remove noise due to electromagnetic interference and microphonic vibrations. We then employ the algorithm to a cryogenic light detector in the laboratory and show an improvement in the detector performance. Finally, we motivate the use of nonlinear terms to better model vibrational contributions to the noise in thermal detectors. We show a further improvement in the performance of a particular channel of the CUORE detector when using the nonlinear algorithm in combination with optimal filtering techniques.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
To meet increasingly demanding physical sensitivities, experiments require low threshold detectors with low noise. Despite the excellent energy resolutions and low thresholds obtained for modern detectors, the intrinsic noise limit is often far out of reach for many of these experiments. Regardless of the information carrier used (e.g. heat, light, charge), detectors are often subject to unavoidable sources of noise when deployed in the field. These may include vibrations from nearby equipment, seismic activity, and/or electronic interference from ground loops or cross-talk from other electronics used in the experiment. For example, gravitational-wave data from the Laser Interferometer Gravitational-Wave Observatory (LIGO) [1] and Virgo [2] experiments are subject to anthropogenic microphonic noise and are sensitive to transient noise during earthquakes [3]. Analyses of these data are typically conducted over a range of frequencies above \(\sim \)20 Hz [4], though this cutoff could be reduced with the reduction of anthropogenic noise below 20 Hz [5]. In the case of charge-collecting detectors used in modern gamma ray spectroscopy experiments, vibrational disturbances can change the capacitance of the system, leading to microphonic noise which degrades the detector performance, including its energy resolution [6]. In these cases, the signals of interest can have rise times on the order of hundreds of nanoseconds [7], resulting in signal bandwidths that are tens of MHz wide [8]. In the case of low-temperature calorimetric detectors, non-negligible amounts of noise can arise from the vibrations induced by the cryogenic facility which dissipate power, thereby adding microphonic and thermal noise to the system [9]. In experiments employing such detectors (e.g. [10,11,12,13,14]), the relevant signal bandwidths range from tens to hundreds of Hz [15]. Analyses of the CUORE data have already suggested that a large component of the noise in several channels is due to vibrations [16], warranting further study of microphonic noise.
The aforementioned sources of noise can be monitored with auxiliary devices which are specially designed to measure particular sources of interest. Deploying such devices to monitor transient noise in an experiment is a common practice. For example, the LIGO and Virgo collaborations use seismometers, accelerometers, magnetometers, and other devices to monitor non-Gaussian transient noise near their detectors [3, 4]. Compared to the cost of typical detectors used in experiments today, these devices are typically affordable while providing sufficiently high fidelities to monitor the relevant noise sources. When the auxiliary data is correlated with the noise in the detectors of interest, denoising algorithms can be applied to the data. The result is a relatively low-cost way to reduce noise without major hardware interventions that can seriously impact an ongoing experiment. Noise cancellation algorithms using data from accelerometers have been demonstrated [8] for possible use in the GRETINA [17] and GRETA [18] experiments. Noise cancellation techniques were also developed and applied to data from the CUORICINO [19] and CUORE-0 [20] experiments by decorrelating the noise from different detector channels [21, 22].
Here, we implement a denoising algorithm based on a multi-input, single-output model. In Sect. 3, we demonstrate the algorithm on simulated data. In Sect. 4, we apply the algorithm to a cryogenic calorimeter operated as a light detector instrumented with neutron transmutation doped (NTD) germanium [23, 24] as our detector of interest. We use an antenna, three accelerometers, and a separate NTD vibration sensor as auxiliary devices and demonstrate that our algorithm improves both the signal-to-noise ratio (SNR) of the detector and its energy resolution. Finally, in Sect. 5 we expand this algorithm to include nonlinear terms and apply this nonlinear version of the algorithm to a channel of the CUORE experiment. Using accelerometers and seismometers as auxiliary devices, we again show an improvement in SNR and energy resolution of the detector.
2 Denoising algorithm
The starting point of this algorithm draws heavily from the multiple-input, single-output model outlined in [25]. We begin with n timestreams \(x_i[t]\) digitized at a sampling frequency of \(f_s\). Here \(i = 0,.., n-1\) indexes the different inputs, i.e. the auxiliary devices described in Sect. 1. We also require that the timestream of the detector we seek to denoise, y[t], be sampled at \(f_s\).Footnote 1 The goal is to find the array of transfer functions \(H_{x_iy}[t]\) which describes the multiple-input single-output system. In practice, y[t] will contain noise and signals of interest for the experiment. (Henceforth, we will refer to this signal of interest as a pulse, while “signal” will be reserved to refer to digital signals such as \(x_i\) and y.) One must therefore subdivide the timestreams \(x_i[t]\) and y[t] into events of length \(N = T f_s\), where T, the length of the filter, is a tunable parameter of the algorithm. The subdivision is done by triggering the raw data stream, after which we can divide the data into two subsets. In the ideal case, one subset contains only events which are free of pulses, \(\{y^{(n)}[t]\}\), which we will call noise events. The other subset, \(\{y^{(p)}[t]\}\), contains all of the pulses present in the data. We take the set of noise events in the detector of interest as well as the corresponding set of events \(\{x^{(n)}_i[t]\}\), which occur at the same times as the elements of \(\{y^{(n)}[t]\}\). We then take the discrete Fourier transform of all noise events giving us the frequency domain sets \(\{Y^{(n)}[f]\}\) and \(\{X^{(n)}_i[f]\}\). With these two sets, we construct the following objects:
The expectation values are taken over the sets of noise events from each device that occur at the same time. For all these objects, f is defined within the range \([0, f_{s})\) and discretized into N bins of width \(\Delta f = \frac{1}{T}\). For convenience, we will drop [f] hereafter and it will be implied that we are operating in the frequency domain unless otherwise noted. It is worthwhile to describe these objects in terms of familiar quantities. \(G_{yy}\) is the average noise power spectrum (ANPS) of the output signal, which is a real non-negative \(1 \times N\) vector. \(G_{x_iy}\) is a complex \(n\times N\) matrix containing the cross-spectral densities of each input with the output. \(G_{x_ix_j}\) is a complex \(n \times n \times N\) 3-dimensional array. For each frequency it contains a Hermitian matrix of the cross-spectral densities between the input signals. The on-diagonal terms \(G_{x_ix_i}\) comprise the ANPS of the input signals. After averaging over all noise events, the transfer functions from the inputs to the detector \(H_{x_iy}\) can be calculated:
\(H_{x_iy}\) is the same size as \(G_{x_iy}\) and also complex. The denoised signal, \(Y'\), is the difference between the original signal and the sum of the auxiliary signals filtered with the transfer function:
This process can then be applied to pulse events to predict the noise which is correlated with the input devices during the event. Assuming that the noise in the detector of interest has components which are correlated with the auxiliaries, the result will be a denoised version of the original set of pulse events. This assumption is usually true given a careful choice of auxiliary devices used in the experiment, however in cases where the matrix \(G_{x_ix_j}\) is dominated by uncorrelated noise in the auxiliary devices, then this denoising technique will not work.
One can verify the success of the algorithm by comparing the noise power spectra of the output signal and the denoised signal, \(G_{yy}\) and \(G_{y'y'}\). One can then proceed to analyze the denoised data, leading to more precise measurements. It is important to note that this algorithm is not a replacement for other filtering techniques such as matched filtering. Rather, this algorithm can be used first to improve the SNR, resulting in new coefficients for the matched filter which can then be applied to the data. This procedure and the implementation of many common filters including the matched filter are linear operations so the two can be combined into a single step [22].
3 Simulation of calorimeter data
To measure the efficacy of our multivariate algorithm under known conditions, we simulate a multi-input, single output system to generate mock data of a cryogenic calorimeter. This mock signal of interest contains both noise and pulses. The simulated auxiliary timestreams represent three accelerometers measuring vibrations and one antenna measuring electromagnetic interference. We first create two signals which represent the original sources of electrical and vibrational noise. The electrical source signal is a sum of many sinusoids at harmonics of 60 Hz, each with a randomized phase.Footnote 2 The vibrational noise is modeled as the sum of three independent sources. The first source signal is a series of sinusoids at harmonics of 1.4 Hz representing the noise from pulse tube (PT) coolers such as the Cryomech PT 415 cryocooler [26], four of which are used in the CUORE experiment. The second source signal is another sinusoid at 100 Hz, simulating a pump or fan with a particular operating frequency. The third source signal is a randomly distributed set of infrequent impulses passed through band-pass filters at three low-frequency values. The impulses represent sudden disturbances in the system, e.g. earthquakes, and the narrow-band responses represent mechanical resonances in the system that are excited by these disturbances.
To create our mock antenna signal, \(x_0[t]\), we add 1/f noise and Gaussian white noise to the electrical source signal. The addition of white noise simulates thermal noise in the antenna, and the 1/f noise represents the flicker noise often present in electronic amplifiers. We generate three accelerometer timestreams by individually passing the vibration source signal through unique band-pass filters and adding 1/f noise and white noise to each output. The band-pass filters represent the response function of each accelerometer, which we slightly vary for each signal. The resulting signals \(x_{1}[t],x_{2}[t],x_{3}[t]\), correspond to accelerometer signals in the \({\hat{x}}\), \({\hat{y}}\), and \({\hat{z}}\) directions. An example of each input signal and their power spectra are shown in Figs. 1 and 2.
To create the mock detector signal, we again generate random white noise and 1/f noise. These terms are similar to some of the noise terms used in [15]. We then take the vibrational source signal and pass it through a unique band-pass filter. This models the detector’s response to vibrations, which should be different from that of the auxiliaries. We add the result and the electrical source signal to the detector signal. Finally, we take a template pulse shape and generate a train of pulses which are randomly distributed in time with the caveat that the minimum time between pulses is longer than the length of the template to prevent pileup events. The pulse amplitude is randomly chosen to be 1 or 2 units with equal probability. The pulse amplitudes represent physical events with energies \(E_0\) and \(E_1 = 2E_0\) assuming linearity of the detector response. After generating the set of valid noise events and pulse events, we apply the noise decorrelation algorithm to the detector signal. Using the simulated antenna and accelerometer signals as auxiliary inputs, we construct the transfer functions \(H_{x_iy}\) following Eqs. 2, 3, and 4, where the expectation values are taken over 5-s long noise windows. The results of the noise decorrelation on the output signal are shown in Fig. 3.
We now seek to show that the amplitude reconstruction and resolution of the simulated data is improved with the addition of the denoising algorithm. To perform our amplitude estimate, we use the detector noise power spectrum and a time-domain average of the pulses from the detector timestream to construct the matched filter, also called the optimal filter (OF) [27,28,29,30]. Applying the OF to the detector signal produces a new signal for which the signal to noise ratio is maximized [31]. Provided that the OF has unity gain, the maximum values of the filtered pulses are the optimal estimators of the pulse amplitudes [32]. We can empirically measure the OF noise resolution by choosing an amplitude value from the same random time point within each noise window. The standard deviation of these values is a good approximation for the OF noise resolution of the detector.
To analyze the results of the noise decorrelation algorithm, we run two analyses in parallel. First we construct the OF using the original simulated signal and use it to reconstruct the pulse amplitudes and OF noise resolution. Separately, we construct a new OF using the denoised data and apply it to the denoised data. The results before and after denoising are shown in Table 1. One important ramification of the algorithm is that it can change the mean value of the OF amplitudes. This is expected because the denoising removes non-stationary noise from the system and because the average pulse created after the denoising is less correlated with the remaining noise. In principle, the denoised average pulse is a better representation of the physical pulse, and in practice the algorithm serves to unbias the data. The amplitude resolution is improved at the noise peak and both pulse amplitudes. In this particular iteration of the simulation, the amplitude resolution consistently improves by 22–23%. Both before and after denoising, the amplitude resolutions across all three values are compatible with each other. This is also expected because no energy-dependent resolution scaling is built into the simulation.
4 Denoising experimental data
4.1 Experimental setup
We performed the measurements to demonstrate the denoising algorithm in an Oxford Instruments Triton 400 dilution refrigerator cooled with a cryogen-free PT cooler at an operating temperature of \(\sim \)10 mK. Inside the refrigerator, we mounted a low-temperature light detector which serves as our detector of interest in this demonstration. The light detector is composed of a 45\(\times \)45\(\times \)0.5 mm\(^3\) silicon wafer that absorbs the light increasing its temperature. This temperature change is measured by a 3\(\times \)3\(\times \)1 mm\(^3\) NTD germanium thermistor which is epoxied to the wafer. The silicon wafer was rigidly connected to a square copper frame that was connected to the experimental stage of the dilution refrigerator. The front-end electronics used in this setup are identical to those used in CUORE [33]. We ran a fiber-optic cable to near this detector with the axis of the cable approximately normal to the substrate surface. We connected the fiber to a commercially available LED with a wavelength of (\(600 \pm 12\)) nanometers to generate a train of LED light pulses with durations and separation times that can be easily controlled by a signal generator with pulse width modulation (PWM). This allowed us to create periodic light signals which we could use to characterize our detectors.
Alongside the light detector, we also mounted another NTD-Ge thermistor which was adhered to a small shard of a silicon wafer. This “NTD vibration sensor” was not rigidly connected to the copper frame, but instead left free to vibrate during the data-taking. Because the NTD vibration sensor was adhered to a relatively small wafer, we expected that the rate of pulses it observed would be negligible compared to that of the light detector. We could therefore use it as another auxiliary device while making cuts to the data in the rare case of a pulse event on the auxiliary device. This was a useful addition to the setup because we expected that the vibrations it measured would be closely correlated to those measured in the light detector due to their close proximity. We also expected that the electrical noise on each of the NTD signals will be similar since they are being read out in identical ways.
To measure the electrical noise present in our system, we attached an antenna to our DAQ and placed it inside the Faraday cage which houses the experiment. The antenna we used was a hand-made Helmholtz coil with \(\sim 50\) turns and a radius of \(\sim 30\) cm, and we amplified the antenna signal by a factor of 5000. To measure the vibrational noise, we used three PCB 393B31 accelerometers [34] mounted triaxially to the 300 K plate of the dilution refrigerator. The accelerometer signals were amplified with a PCB 482C15 signal conditioning pre-amplifier [35] and read out to the same DAQ system as the antenna, NTD vibration sensor, and light detector. The ANPS of the auxiliary devices are shown in Fig. 4.
As a starting ground for testing the algorithm, we collected data while a ground loop persisted in our readout chain. This gave rise to noise peaks at 60 Hz and subsequent harmonics from the AC power line with sufficient power to dominate the detector signal. In addition to the electrical noise, we observe noise peaks at 1.4 Hz and subsequent harmonics due to vibrations from the PT which vibrate the internal cryogenic structure, including the detector and the cold electronics. To construct an average noise template, we first took a long run of data without injecting any pulses to ensure a sufficient number of noise events free of pulses. The data from these runs were used to construct the transfer functions, \(H_{x_iy}\), which are shown in Fig. 5. After, we acquired two runs of data, each with a set of LED pulses at a fixed frequency (1.7 Hz). The first set of LED pulses had a duration of 100 \(\mu \)s, and the second set had a duration of 200 \(\mu \)s and thus twice the energy of the first set. We split the PWM signal and acquired it with our DAQ to efficiently trigger on the light pulses with few backgrounds. We denoised these LED runs by applying the previously calculated transfer functions to the auxiliary data from the LED runs and subtracting the result from the light detector signal.
4.2 Analysis techniques
Since we were able to trigger the LED pulses, we could still analyze our signal despite the low SNR in the system. We used the same analysis procedure described in Sect. 3, but this time we conducted three separate analyses. We again analyze the original data and denoised data, but we also analyze the data after applying a notch filter centered at 60 Hz to the signal. Since the signal is dominated by 60 Hz noise, one might expect that notch filtering the data will lead to an improvement comparable to that of the noise decorrelation algorithm. We selected 0.5-s windows between LED pulses and making RMS-based cuts to veto windows with spurious events (e.g. caused by cosmic rays) either on the detector signal or NTD vibration sensor. We then construct the average noise power spectrum for each of the three signals. We use the LED triggers to build an average pulse which is also 0.5 s long. Together, the noise events and average pulse template allow us to denoise the data and build the OF. We separately construct the OF using the original data, the band-pass filtered data, and the denoised data.
To align the gains of the three filters, we perform a pseudo-calibration assuming that the second set of LED pulses has twice the energy of the first one. We align the LED pulse amplitudes to arbitrary energy values of \(E_0\) and \(2E_0\). We construct histograms of the amplitudes obtained by each analysis of each LED event, then fit the amplitudes to a normal distribution using an unbinned fit. We then use a second-order polynomial fixed at the origin to align the measured amplitude values to a common scale. Following the work of [36], one can model the energy resolution of the detector as \(\sigma (E) = \sqrt{\sigma _0^2 + f(E)\sigma _E^2}\), where \(\sigma _0\) is the baseline resolution of the detector, \(\sigma _E\) is a scale factor that arises due to the stochastic nature of the LED signals, and f(E) is a monotonically increasing function. We expect this behavior of the detector response because the magnitude of the Poisson fluctuations in the number of photons captured by the light detector will increase as the amplitude of the LED pulse increases. With this model, we expect that the denoising algorithm does not affect f(E) or \(\sigma _E\), but does lower the value of \(\sigma _0\). We thus expect to see the greatest relative improvement in energy resolution at the noise peak where the energy-dependent effects are small. We define the ratio of resolutions \(R(E) = {\hat{\sigma }}_{denoised}(E) / {\hat{\sigma }}_{original}(E)\) and calculate it at each energy for the three signal processing techniques.
4.3 Results of denoising
After denoising, there was a visible improvement in the signal-to-noise ratio of the detector as seen in Fig. 6. We also observed a reduction in both vibrational and electrical noise in the light detector power spectrum (see Fig. 7). The vibrational peak at 1.4 Hz was eliminated, and many of the electrical peaks ranging from 60 Hz to 5 kHz were significantly reduced. The 60 Hz noise peak was reduced by 33 dB, and the full width at half-max (FWHM) was reduced by approximately 10%. This peak was extremely wide due to small modulations of the ground loop frequency, but these modulations were measured by the antenna and the NTD vibration sensor so we could successfully remove them from the data. This is reflected in the wide-band region of high gain near 60 Hz in the transfer functions for both of these devices as was shown in Fig. 5. Outside of this region, the main effect of the denoising is a reduction in noise peaks. The algorithm does not have an apparent effect on the broadband detector noise. Before denoising, high frequency noise with a large 60 Hz component contaminates the average pulse. Applying the notch filter induces a ringing in all pulses which coherently add together when constructing the average pulse. When the denoising algorithm is applied, the noise in the average pulse is reduced without inducing any ringing (see Fig. 8).
We next examine the effect of the denoising algorithm on the resolution of the noise and LED pulse events. As expected, the resolution increases monotonically with energy. At all amplitudes, the denoising improves the measured detector resolution (see Table 2). We see the greatest relative improvement at lower pulse amplitudes, which was also expected. It should be noted that the improvement in amplitude resolution at the first LED peak \((E = E_0)\) is compatible with the improvement at the noise peak, though it has a greater uncertainty.
We also observe that the denoising algorithm performs as well as or better than the notch filtering approach (see Fig. 9). This effect is especially visible at the noise peak, where the improvement in the detector resolution from the denoising (4.1%) is 1.7 times that of from the notch filtering (2.4%). At the highest-amplitude peak, we do not expect the denoising to have as much of an effect on the resolution, but we still observe an improvement. At this amplitude, we see that the denoising and notch filtering have similar effects on the data These results are summarized in Table 3.
It is important to note that the improvement in resolution is \(\le 5\%\) in all cases. This is also expected since the OF is by definition the filter resulting in the highest SNR given a particular choice of noise and pulse template. The improvements from the denoising are thus limited by the amount of noise that can be removed from the noise and signal templates, which can in turn be limited by the “SNR” of the auxiliary devices, i.e. the ratio of the amount of noise correlated with the detector of interest to the amount of uncorrelated noise. Having verified the efficacy of the denoising technique on a controlled experiment, we applied the method to data from the CUORE experiment.
5 Denoising CUORE data
5.1 The CUORE experiment
CUORE (Cryogenic Underground Observatory for Rare Events) [10] is an ongoing experiment at the Gran Sasso National Laboratory (LNGS) primarily searching for the lepton number violating process of neutrinoless double beta decay \((0\nu \beta \beta )\) [37, 38]. The discovery of such a process would be a clear experimental signature of physics beyond the Standard Model and may provide insight into the nature of the absolute neutrino mass scale [39,40,41,42]. The sensitivity to such a discovery requires detectors that have very good energy resolution. For example, in the case of a finite, non-zero number of expected background counts, the sensitivity to the half-life of \(0\nu \beta \beta \) scales as \((\Delta E)^{-\frac{1}{2}}\) [43], where \(\Delta E\) is the energy resolution of the system at the expected peak energy \(Q_{\beta \beta }\).
CUORE employs cryogenic calorimeters consisting of a 5 \(\times \) 5 \(\times \) 5 cm\(^3\) TeO\(_2\) crystal, a silicon resistive heater, and an NTD-Ge thermistor that measures the small temperature rise in the crystal when energy from a nuclear decay or particle interaction is deposited in it. Each of these calorimeter signals is then read out as an individual channel and digitized with a sampling frequency of 1 kHz. As mentioned before, CUORE uses four Cryomech PT 415 cryocoolers with fundamental frequencies of 1.4 Hz to cool the cryostat. The vibrations induced by the PTs are a large source of noise due to CUORE’s relatively small bandwidth (\(\sim \)80 Hz), and active steps are already taken to reduce this PT noise during data-taking [44]. Nonetheless, this noise persists in many channels in the CUORE data.
To measure sources of microphonic noise, we installed three triaxially-oriented accelerometers identical to the ones described in Sect. 4.1 on the structural support frame of the cryostat and read the three signals out to the CUORE DAQ system. We also installed a triaxial Sara Electronics SS-10 seismometer [45] to the CUORE suspension system and read it out to the same DAQ. We use these devices as the six input signals when we apply the noise decorrelation algorithm. We also motivate the use of a nonlinear version of the algorithm for thermal detectors. Using 24 h of data from one CUORE detector channel, we demonstrate both the linear and nonlinear versions of the denoising algorithm using the accelerometer and seismometer signals as inputs. We then analyze the noise events and so-called “pulser” events to evaluate the performance of the algorithm.
5.2 CUORE pulser events
To account for changes in the detector response due to thermal fluctuations, short pulses of known voltages are applied to the CUORE detector at regular intervals using the silicon heaters mentioned in Sect. 5.1. This results in pulser events in the CUORE channels. Since the injection method is practically identical for each pulser event, we expect their resolutions to be comparable to the noise resolution of the channel, so they make good candidates for verifying the performance of the denoising algorithm. The particular channel we choose has pulser amplitudes during this run. The two amplitudes are comparable to physical pulses with energy depositions of 960 keV and 3500 keV. These energy values were determined during the online CUORE data analysis. We will use the two pulser event energies and the noise peak to calibrate the detector.
5.3 The nonlinear algorithm
It is expected that quadratic mixing of acceleration signals should occur in the CUORE thermal detectors. This is motivated both by the non-zero bicoherence observed in CUORE (see Appendix A) and also from first principles. A simplified explanation for this is that any power incident on the detector due to vibrations heats a thermal detector regardless of directionality of the vibration. This power is non-trivially related to the square of the waveform describing this vibration (see Appendix B for a detailed explanation). Here we use the square of each auxiliary signal \(x_i^2[t]\) as a proxy for the thermal response to vibrational noise. Approaches similar to this which analyze nonlinear systems by including higher-order powers of the input signals are well described in [46, 47]. We still keep the original auxiliary signals \(x_i[t]\) since they may be separately correlated with the detector noise, for example due to capacitive pickup in the detector wiring. We proceed with the denoising algorithm using the new set of input signals, which is twice as large as the original set. In practice the Fourier amplitudes \(X_i[f]\) and \(X_i^2[f]\) can be strongly correlated, so one must ensure that the matrices are well-conditioned before proceeding. While we do not expect this model to capture the full nonlinear behavior of the thermal response to the input signals, we do expect to see an improvement in the noise decorrelation over the linear model.
5.4 Results of denoising
The CUORE data have a much higher SNR than the data used to demonstrate the algorithm in Sect. 4. We are therefore able to trigger on pulse events directly and create a set of noise events. The only possible contamination of this set is from very low-amplitude pulses not detected by the CUORE triggering algorithm. This effect is expected to be small because the pulse rate is small (\(\sim \)3 mHz) and the small amount of signal power due to these pulses is negligible. We first run the algorithm with 50-second long noise windows using the accelerometer and seismometer signals as inputs, then we repeat the algorithm on the original data using the same signals and their squares as inputs. We conduct an independent analysis after each of the different denoising methods is applied to the raw data. We show that the denoising produces a cleaner pulse template as well as a reduced noise power spectrum. Comparing the noise power spectra, we see that the denoising performs excellently on the PT noise, but does not substantially impact the noise below 1 Hz and does little to change the broadband continuum noise below the various peaks (see Fig. 10). The nonlinear method is more effective than the linear one, though this is limited to specific bands in frequency space. The most noticeable differences between the linear and nonlinear algorithms are seen at the harmonics of 1.4 Hz and in a region around 280 Hz with many noise peaks (see Fig. 11). This suggests the presence of a nonlinearity in the system from the auxiliary devices to the CUORE channel. This result is discussed further in Appendix A.
To evaluate the performance of the different denoising methods, we first examine the distributions of noise events which are shown in Fig. 12. Before the OF is applied, the denoising algorithm reduces the RMS of a typical noise event by more than a factor of 2. We build the OF using a 10-s long signal template. To make the calculation of the OF coefficients as straightforward as possible, we also use 10-s long noise windows to construct the OF as opposed to the 50-s long noise windows we use to perform the denoising. The OF naturally further reduces the RMS noise, but combining the OF with the denoising produces a smaller mean noise event RMS. This value is further reduced when the OF is combined with the nonlinear algorithm. We confirm this effect by building the ANPS of the detector after each signal processing technique is applied (see Fig. 13). Taking the square root of the integral of each of the ANPS gives the expected RMS detector noise, the values of which are tabulated in Table 4. The denoising improves the noise RMS of the OF signal by 1.0 mV (\(\sim \)8%). The quadratic version of the algorithm again out-performs the linear one, improving the noise RMS by 1.2 mV (\(\sim \)10%).
We now analyze the noise events and pulser events together using the same analysis techniques as described in Sect. 4.2. Fitting the OF amplitude spectrum and calibrating using the known pulser energies gives us our final energy spectrum. The obtained resolutions are reported in Table 5, while the ratios of resolutions are reported in Table 6. Looking at the noise peak, where the statistical uncertainty on the resolution is smallest, the relative improvement in resolution is \(9\%\) when the linear denoising is applied and \(10\%\) when the nonlinear version is applied. This again shows the better performance of the nonlinear method. This \(10\%\) improvement in energy resolution is similar to the improvement in the noise RMS, verifying that the calibration method does not qualitatively change the effects of the denoising. The pulser resolutions are comparable to the noise resolution as expected, though the resolution of the 3500 keV pulser is slightly worse than expected. Even so, the relative improvement in energy resolution is compatible at all energies (see Fig. 14). Together, these results support the hypothesis that the resolution of the pulsers is approximately the same as that of the noise. An energy-dependent contribution to the pulser resolution should not be removed by the denoising algorithm because the auxiliary device signals are completely uncorrelated to the pulse energy. If a non-negligible energy-dependent contribution were present, then the relative change in the resolution of the pulsers after denoising would be different as a function of energy. We observe that the improvements in pulser resolution are compatible at all measured energies, which suggests that the energy-dependent contribution to the pulser energy resolution is negligible. In summary, the denoising algorithm consistently improves the energy resolution of noise events and pulser events by \(\sim \)10%. The nonlinear version of the algorithm adds another \(\sim \)1% improvement, though this improvement is within the uncertainty of the measurement.
We can also estimate the timing resolution of the detector before and after the denoising techniques are applied. For a signal of amplitude A, the timing resolution of the digital signal is given by [32]:
where T is the length of the signal in seconds, \(s(f_n)\) is the Fourier transform of the average pulse, and \(N(f_n)\) is the ANPS. The timing resolution is inversely proportional to the amplitude of the signal, and it is a single number for a given average pulse and ANPS. We therefore use this proportionality constant \(A\sigma _{t_0}\) as our figure of merit for the different signal processing techniques. We calculate it using the average pulse with amplitude 1 mV and multiply by the value of the calibration function at \(A = 1\) mV to report the value in keV\(\cdot \)s (see Table 7). We find that the linear denoising improves the timing resolution by 1.4% while the quadratic version improves the timing resolution by 1.9%. We note that the improvement in timing resolution is significantly less than that of the energy resolution. This suggests that in this channel, vibrational noise does not contribute significantly to the timing resolution.
The main limitations of the algorithm are likely related to the auxiliary devices. First, it is important that the auxiliary devices be sensitive to noise in relevant signal bands. The magnitude of the OF applied to the raw timestream peaks near 1.4 Hz, where the PT noise is very well measured. At frequencies below \(\sim \)1 Hz, the accelerometer responses roll off significantly, thus the accelerometers are not sensitive to the sub-Hz noise observed in the CUORE data. The seismometers used in CUORE are designed to measure sub-Hz noise with better sensitivity, so the algorithm can reduce the sub-Hz noise peaks in the CUORE data. These peaks are reduced by 4–5 dB and are far from eliminated. The performance of the denoising is also directly tied to the coherence of the auxiliary devices with the CUORE channel, i.e. the fraction of the power in the auxiliary signals that is correlated with the detector at a particular frequency. It is possible that the coherence of sub-Hz noise between the seismometer signal and detector noise can be increased by moving the seismometers elsewhere on the cryostat, as their locations have not yet been optimized. It is also likely that the algorithm can be improved by introducing auxiliary devices in close proximity to the CUORE detector. This suggests that introducing cryogenic vibration sensors into future cryogenic bolometric experiments is to be investigated further. A basic version of this technique using an NTD vibration sensor was demonstrated in Sect. 4.1.
Finally, it is instructive to compare the effects of this denoising technique with the noise decorrelation technique used in CUORICINO [21]. In CUORICINO, neighboring bolometer channels were used as auxiliary devices to remove noise that was correlated between them. The CUORICINO technique effectively removed sharp peaks from the noise power spectrum much like the technique presented here. However, the CUORICINO technique was effective at removing continuum noise below 1 Hz, which our technique fails to do. This suggests that the noise measured in the auxiliary devices used in CUORE is not correlated with the continuum noise in the CUORE detector. In [21], it was found that most CUORICINO channels do not exhibit more than a 5% improvement in OF noise resolution, though the largest improvements were more than 50%. It is unclear whether the CUORE channel showed here is one of the most-improved or least-improved CUORE channels using our noise correlation algorithm, but the 10% improvement in OF resolution we report here out-performs the vast majority of channels in the CUORICINO analysis.
It should also be noted that the CUORICINO noise decorrelation was done using 11 bolometer channels as auxiliary channels. In practice, this creates a problem due to “side-pulses,” i.e. pulses present on neighboring bolometer channels. The presence of side-pulses creates difficulties as one must find a way to remove them, either by fitting the pulses on the auxiliary channels and subtracting them or by removing the channel(s) containing side-pulses and re-computing the transfer functions between the remaining channels. The probability of encountering side-pulses grows as the number of channels used in the decorrelation increases. Furthermore, neighboring bolometer channels are the most likely to have correlated noise, but they are also the most likely to have coincident pulses due to multi-site physics events. The denoising technique presented here does not incur such problems since there are no side-pulses on the auxiliary devices used in CUORE. Furthermore, this technique denoises bolometer channels against auxiliary devices, which has not been attempted before in bolometric experiments, and it introduces nonlinear terms motivated by detector response.
6 Conclusions
We presented an algorithm based on a multi-input, single-output model which can be used in a variety of settings to remove noise from detector signals using multiple auxiliary devices such as microphones, accelerometers, and seismometers. The algorithm can be applied to detectors with a wide range of bandwidths and offers an inexpensive way to reduce noise without major experimental interventions. We validated the approach by implementing the multivariate noise decorrelation algorithm on simulated calorimeter data with simulated noise contributions from electromagnetic interference and microphonic vibrations. We then implemented the algorithm on a low-temperature light detector using accelerometers, an antenna, and an NTD vibration sensor as auxiliary devices. The result was a reduction in the noise RMS of the detector and a 4% improvement in energy resolution at the noise peak. We then presented the case for using a nonlinear algorithm which accounts for quadratic mixing terms from the auxiliary device signals to denoise thermal detectors. We demonstrated the nonlinear version of the algorithm with one channel from the CUORE experiment, showing an improvement in the OF noise RMS and the energy resolution of noise events and pulser events. The denoising algorithm improved these by approximately \(10\%\). The nonlinear version of the algorithm outperformed the linear version, adding a further improvement of 1–2%. The main effect of the denoising was a reduction in noise peaks but there was no clear reduction in the broadband noise. The timing resolution of the channel was also estimated to improve by 1–2%, with the greatest estimated improvement again coming from the nonlinear version of the algorithm.
Work is ongoing to integrate this denoising algorithm into the analysis framework of CUORE in future searches for rare processes such as neutrinoless double beta decay. Looking ahead, we plan to conduct a rigorous study of the effects of the denoising algorithm, particularly its impact on the timing resolution, energy resolution, and energy threshold of the entire CUORE detector over multiple years of data-taking. This will allow for a better comparison of the denoising technique outlined here with the technique used in CUORICINO. We are also in the process of expanding our array of auxiliary devices in the CUORE experiment by adding antennas and more accelerometers to our DAQ system. Beyond this, we will further investigate the use of bicoherence and higher-order frequency mixing terms to more accurately model the effects of vibrations in thermal detector signals, extending the model proposed in [16]. This will lay the ground work as we build toward a multivariate nonlinear adaptive filter based on the single-input model demonstrated in [6]. Active filters such as these have the potential to be integrated into the online data-taking of future experiments including CUPID [48].
Data availability statement
The CUORE data used in this paper are a small subset of the much larger CUORE dataset which will be analyzed in future works. The code used to complete this analysis and the raw data analyzed here (including waveform data from the NTD-Ge detector, the CUORE channel, and all auxiliary devices) are not currently publicly available but they can be made available upon request.
Notes
Technically, the common sampling frequency requirement is a simplification. Because the transfer functions are calculated in the frequency domain, if the sampling rates differ, the transfer functions can still be calculated by truncating the signal with higher the sampling rate in the frequency domain.
Here we are simulating electrical noise in data taken in North America, where the standard AC frequency is 60 Hz. Later, in Sect. 5.1, we will use data taken in Europe, where the standard AC frequency is 50 Hz.
References
J. Aasi et al., “Advanced LIGO”, Class. Quantum Gravity 32(7), 074001 (2015)
M. Mantovani, Virgo and the gravitational interferometry. EPJ Web Conf. 280, 08005 (2023)
F. Acernese et al., The Virgo O3 run and the impact of the environment. Class. Quantum Gravity 39(23), 235009 (2022)
B.P. Abbott et al., A guide to LIGO-Virgo detector noise and extraction of transient gravitational-wave signals. Class. Quantum Gravity 37(5), 055002 (2020)
B. Berger, LIGO Scientific Collaboration Team, “Saga of mysterious 15 Hz ground motion at LIGO Livingston Observatory,” in APS April Meeting Abstracts, Volume 2022 of APS Meeting Abstracts, (2022)
S. Zimmermann, Active microphonic noise cancellation in radiation detectors. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 729, 404–409 (2013)
T. Stezelberger et al., Data Acquisition and Signal Processing for the Gamma Ray Energy Tracking Array (GRETA). arXiv preprint arXiv:2011.00129, (2020)
S. Zimmermann et al., Development of next-generation nuclear physics integrated readout electronics for GRETINA, in 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC) (IEEE, 2013), p. 1–7
S. Pirro et al., Vibrational and thermal noise reduction for cryogenic detectors. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 444(1), 331–335 (2000)
D.Q. Adams et al., Search for Majorana neutrinos exploiting millikelvin cryogenics with CUORE. Nature 604(7904), 53–58 (2022)
C. Alduino et al., CUORE-0 detector: design, construction and operation. J. Instrum. 11(07), P07009 (2016)
E. Armengaud et al., The CUPID-Mo experiment for neutrinoless double-beta decay: performance and prospects. Eur. Phys. J. C 80(1), 1–15 (2020)
O. Azzolini et al., Final result on the neutrinoless double beta decay of \(^{82}{{\rm Se}}\) with CUPID-0. Phys. Rev. Lett. 129, 111801 (2022)
H.B. Kim et al., Status and performance of the AMoRE-I experiment on neutrinoless double beta decay. J. Low Temp. Phys. 209(5–6), 962–970 (2022)
M. Carrettoni, M. Vignati, Signal and noise simulation of CUORE bolometric detectors. J. Instrum. 6(08), P08007 (2011)
D. Adams et al., An energy-dependent electro-thermal response model of CUORE cryogenic calorimeter. J. Instrum. 17(11), P11023 (2022)
I. Lee et al., GRETINA: a gamma ray energy tracking array. Nucl. Phys. A 746, 255–259 (2004)
M. Deleplanque et al., GRETA: utilizing new concepts in \(\gamma \)-ray detection. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 430(2), 292–310 (1999)
C. Arnaboldi et al., New limit on the neutrinoless \(\beta \beta \) decay of \(^{130}{{\rm Te}}\). Phys. Rev. Lett. 95, 142501 (2005)
L. Canonica et al., Results from the CUORE-0 experiment. J. Phys. Conf. Ser. 718(6), 062007 (2016)
C. Mancini-Terracciano, M. Vignati, Noise correlation and decorrelation in arrays of bolometric detectors. J. Instrum. 7, P06013 (2012)
J. Ouellet, The Search for Neutrinoless Double Beta Decay in\(^{130}\)Te With CUORE-0. Ph.D. thesis, University of California, Berkeley (2015)
E. Haller et al., NTD germanium: a novel material for low temperature bolometers, in Neutron Transmutation Doping of Semiconductor Materials (Springer, 1984), p. 21–36
N. Wang et al., Electrical and thermal properties of neutron-transmutation-doped Ge at 20 mK. Phys. Rev. B 41(6), 3761 (1990)
J.S. Bendat, A.G. Piersol, Random Data: Analysis and Measurement Procedures (Wiley, New York, 2011)
Cryomech. https://www.cryomech.com/products/pt415/
G. Turin, An introduction to matched filters. IRE Trans. Inf. Theory 6(3), 311–329 (1960)
D. North, An Analysis of the factors which determine signal/noise discrimination in pulsed-carrier systems. Proc. IEEE 51(7), 1016–1027 (1963)
E. Gatti, P.F. Manfredi, Processing the Signals From Solid State Detectors in Elementary Particle Physics. Riv. Nuovo Cim. 9N1, 1–146 (1986)
A. Branca et al., Performance of the low threshold optimum trigger on CUORE data. J. Phys. Conf. Ser. 1468, 012118 (2020)
S.D. Domizio et al., Lowering the energy threshold of large-mass bolometric detectors. J. Instrum. 6(02), P02007 (2011)
S.R. Golwala, Exclusion limits on the WIMP-nucleon elastic-scattering cross-section from the Cryogenic Dark Matter Search. Ph.D. thesis, University of California, Berkeley (2000)
C. Arnaboldi et al., A front-end electronic system for large arrays of bolometers. J. Instrum. 13, P02026 (2018)
PCB Piezotronics. https://www.pcb.com/products?m=393b31
PCB Piezotronics. https://www.pcb.com/products?m=482c15
V. Singh et al., Large-area photon calorimeter with Ir-Pt bilayer transition-edge sensor for the CUPID experiment. Phys. Rev. Appl. 20, 064017 (2023)
S.R. Elliott, P. Vogel, Double beta decay. Ann. Rev. Nucl. Par. Sci. 52(1), 115–151 (2002)
M.J. Dolinski et al., Neutrinoless double-beta decay: status and prospects. arXiv preprint arXiv:1902.04097 (2019)
J. Schechter, J.W. Valle, Neutrinoless double-\(\beta \) decay in SU (2)\(\times \) U (1) theories. Phys. Rev. D 25(11), 2951 (1982)
J.D. Vergados et al., Neutrinoless double beta decay and neutrino mass. Int. J. Mod. Phys. E 25(11), 1630007 (2016)
W. Rodejohann, Neutrinoless double beta decay and neutrino physics. J. Phys. G 39, 124008 (2012)
F. Deppisch et al., Neutrinoless double beta decay and physics beyond the Standard Model. J. Phys. G 39, 124007 (2012)
I. Nutini, Performance of the CUORE bolometers-first CUORE results. Il Nuovo Cimento C 41(3), 1–10 (2018)
A. D’Addabbo et al., An active noise cancellation technique for the CUORE Pulse Tube cryocoolers. Cryogenics 93, 56–65 (2018)
Sara Electronic Instruments. https://www.sara.pg.it/prodotti.php?id=60
H. Rice, J. Fitzpatrick, A generalised technique for spectral analysis of non-linear systems. Mech. Syst. Signal Process. 2(2), 195–207 (1988)
J.S. Bendat, A.G. Piersol, Spectral analysis of non-linear systems involving square-law operations. J. Sound Vib. 81(2), 199–214 (1982)
K. Alfonso et al., CUPID: The Next-Generation Neutrinoless Double Beta Decay Experiment. J. Low Temp. Phys. 1–9 (2022)
Acknowledgements
The authors would like to thank Dr. Steven M. Kay for his guidance and suggestions regarding the spectral analysis of nonlinear systems as well as Dr. Arnaud Allézy for his work on modeling the vibrational modes of the CUORE cryostat. We also thank the CUORE Collaboration for helping to make this work possible. This work was supported by the US Department of Energy (DOE), Office of Science under Contract No. DE-AC02-05CH11231, and by the DOE Office of Science, Office of Nuclear Physics under Contract No. DE-FG02-00ER41138. This work was also supported by the Istituto Nazionale di Fisica Nucleare (INFN). This work makes use of both the DIANA data analysis and APOLLO data acquisition software packages, which were developed by the CUORICINO, CUORE, LUCIFER, and CUPID-0 Collaborations.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Code availability statement
Code/software will be made available on reasonable request. [Author’s comment: The code/software generated during and/or analysed during the current study is in fact available from me on reasonable request].
Appendices
Appendix A: Bicoherence
Using the same notation as in Sect. 2 where possible, we define the bispectrum of a signal x[t]:
and we define the bicoherence, b using the following normalization convention:
where the expectation values are taken over the same noise samples used to compute the denoising transfer function. The bicoherence is restricted to be between 0 and 1, and it measures the proportion of the signal power at any frequency pair that is phase coupled to its sum frequency. This is sometimes interpreted to be a measurement of the nonlinearity present in a system, however there are subtleties related to this. Consider a signal of the form:
where \(A_i\) are all real coefficients. There is no nonlinear mixing in this system. Now consider three frequencies \(f_i, f_j, f_k\) where \(f_i + f_j = f_k\). If there is no phase coupling between \(f_i\), \(f_j\), and \(f_k\), then \(b^2(f_i, f_j)\) approaches 0 after averaging over a sufficiently long period of time. If the phases are coupled such that over a long time \(\big \langle \phi _k - (\phi _i + \phi _j)\big \rangle \ne 0\), then \(b^2(f_i, f_j) > 0\). This phase coupling is expected in systems that are driven by an external periodic source of noise. The PTs in CUORE are an example of such an external source, so we expect non-zero bicoherence at harmonics of the fundamental PT frequency (1.4 Hz) even without the presence of a nonlinear mixing term.
Now we analyze the bicoherence of a system with a quadratic mixing term. Consider a simple signal of containing two frequencies with a nonlinear mixing term. Using \(\omega _i = 2\pi f_i\), we have:
The bicoherence of this signal at \((\omega _1,\omega _2)\) is:
In this system, all of the power at \(f = f_1 + f_2\) is due to the mixing of the signals at \(f_1\) and \(f_2\), so the bicoherence attains the maximum possible value.
Analysis of these two signals shows that while a non-zero bicoherence is a natural consequence of a quadratic nonlinearity in a system, it does not guarantee the existence of such a nonlinearity. The bicoherence is therefore useful for examining the phase-coupled components of a signal, but one should be cautious before attributing such phase coupling to an underlying nonlinearity. We discuss why this quadratic mixing model may still be appropriate for the CUORE system in Sect. 5.3 and Appendix B.
The bicoherence of a single CUORE channel is plotted in Fig. 15 for regions of frequency space where the bicoherence is non-negligible. Note that the bicoherence is an auto-quantity; it only depends on the bolometer signal, not a pair of signals. We see a large amount of bicoherence between the different PT frequencies, but we expect the harmonics of the PT phases to be coupled already. An interesting region exists around \(f_1 \approx 280\) Hz, \(f_2 < 10\) Hz. There, we see several points with a large bicoherence term. The distances between these points are integer multiples of 0.6 and 1.4 Hz, both of which appear on their own in the bolometer ANPS. Here, there is no obvious harmonic effect that could contribute to the bicoherence, so a nonlinear mixing term may be at play. We also note that it is at these frequencies near 280 Hz where we see the greatest difference in noise power between the linearly denoised and quadratically denoised signal, as discussed in Sect. 5.4 and shown in Fig. 11. This is further evidence that the noise power at these frequencies is due to a nonlinear effect.
Appendix B: Frequency mixing in power signals
Consider an object of mass m vibrating in 1-dimension whose acceleration is described by a discrete signal a[t]. The power supplied by this oscillation is equal to \(p[t] = ma[t]v[t] = m a[t] \sum _{t' < t} a[t'] dt' \). (We will drop the m so that p has units of \(t(a[t])^2\). Taking the Fourier transform of the power, we have:
where \(\Delta \omega ' = \omega - \omega '\) and \(\circledast \) represents the convolution. One can replace \((\Delta \omega ')\) with \((2\pi f_{s} + \Delta \omega ')\) if \(\omega ' > \omega \) when working on the non-negative frequency domain. Now consider the square of the acceleration signal \((a[t])^2\). Its Fourier transform is:
The transfer function from this signal to the power signal is:
Plugging in, we see the transfer function is averaged over all mixing frequencies, the \(P[\omega ]\) term of which is averaged in a frequency-weighted way:
The bispectrum \({\mathcal {B}}^2[\omega ', \Delta \omega ']\) of the original acceleration signal is now a useful quantity, since it tells us which frequency pairs will contribute to the expectation value of the above sum. If the matrix is sparse, then the number of frequency pairs over which one must sum becomes significantly smaller, and evaluating the sum for each \(\omega \) becomes significantly less computationally expensive. Nonetheless, the transfer function \(H_{x^2P}\) still provides a “frequency-smeared” approximation of the expected value of the power signal given the square of the accelerometer signal. In the limit that a single frequency pair \((\omega _0, \omega - \omega _0)\) dominates the sums in Eqs. B6 and B7, we can approximate the signals as:
Plugging this into B9, the factors of \(\big \langle (A[\omega _0] A[\omega - \omega _0])^*(A[\omega _0] A[\omega - \omega _0])\big \rangle \) cancel assuming there is no uncorrelated noise in the acceleration signal at \(\omega \). The transfer function is:
In this case, the transfer function accurately describes the frequency mixing behavior while accounting for the difference in units between P and \((a[t])^2\). Of course, this approximation of a single perfectly bicoherent frequency pair contributing to the sum in B9 is not valid in the CUORE system since there are multiple frequency pairs exhibiting bicoherence that sum to the same frequency (see Fig. 15). Still, using the frequency-smeared coefficients from the squares of the auxiliary signals in the denoising algorithm results in an improvement over the linear version.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Funded by SCOAP3.
About this article
Cite this article
Vetter, K.J., Beretta, M., Capelli, C. et al. Improving the performance of cryogenic calorimeters with nonlinear multivariate noise cancellation algorithms. Eur. Phys. J. C 84, 243 (2024). https://doi.org/10.1140/epjc/s10052-024-12595-y
Received:
Accepted:
Published:
DOI: https://doi.org/10.1140/epjc/s10052-024-12595-y