Abstract
Imaging large areas of a sample non-destructively and with high resolution is of great interest for both science and industry. For scanning coherent X-ray diffraction microscopy, i. e., ptychography, the achievable scan area at a given spatial resolution is limited by the coherent photon flux of modern X-ray sources. Multibeam X-ray ptychography can improve the scanning speed by scanning the sample with several parallel mutually incoherent beams, e. g., generated by illuminating multiple focusing optics in parallel by a partially coherent beam. The main difficulty with this scheme is the robust separation of the superimposed signals from the different beams, especially when the beams and the illuminated sample areas are quite similar. We overcome this difficulty by encoding each of the probing beams with its own X-ray phase plate. This helps the algorithm to robustly reconstruct the multibeam data. We compare the coded multibeam scans to uncoded multibeam and single beam scans, demonstrating the enhanced performance on a microchip sample with regular and repeating structures.
Similar content being viewed by others
Introduction
X-ray ptychography has developed into a robust imaging technique with applications in many fields of science, allowing to visualize complex processes in nano-structured materials with high sensitivity and spatial resolutions down to below 10 nm1,2,3,4. One of the remaining challenges in ptychography is the fast scanning of larger sample areas in a shorter time while preserving the highest resolution in the reconstructed image5,6,7. One approach to tackle this challenge is to increase the rate at which data is taken, for example by minimizing overheads and continuously (fly-)scanning. With this approach of speeding up the data rates of conventional ptychography, the available coherent flux will eventually pose the upper limit on how quickly data can be recorded while preserving the resolution in the reconstruction. Since the coherent fraction of the monochromatic beam at third-generation synchrotron radiation sources is in the one percent range, most of the beam remains unused in this technique. This limits the field of view and resolution in ptychography, typically trading the one off against the other. Experiments that require both, are typically time consuming and thus sometimes not feasible.
For instance, to understand reactions in heterogeneous catalysis the chemistry of relevant structures on the single-digit nano-scale need to be followed under working conditions over time and extended regions inside a reactor8,9. Here, besides a high spatial resolution10 also chemical sensitivity11,12 is needed. Both require high fluence, currently limiting the field of view. However, such experiments are essential to derive structure-activity relationships, leading to the design of more sustainable and efficient materials for purposes such as emission control, synthesis of industrial-scale chemical products, and conversion or storage of energy-related chemicals. Decreasing the time needed to scan larger representative samples will enable the tracking of dynamic changes in heterogeneous catalysts as a function of the environment and potentially allowing proper kinetic studies such as reactor rates or activity9. Similar requirements are found for other projects, such as the RAVEN program of IARPA13, where millimeter-scale microchips need to be imaged with resolutions down to the transistor level.
One way to overcome the current limitations is to increase the brightness of the source, which is pursued by most synchrotron radiation sources around the world and will lead to significant advances in the next decade14,15,16. Another complementary way is to exploit the multibeam approach in ptychography17. In this scheme, the object is scanned simultaneously with several spatially separated mutually incoherent probes. This parallel acquisition of several parts of the sample increases the field of view while preserving the fluence and thus the sensitivity and resolution.
Multibeam ptychography was demonstrated with visible light17,18 and with X-rays using spatially separated slits19, Fresnel zone plates20 and refractive lenses21. Best use of the incident X-rays can be made if the optics that generate the probes cover a large area of the incident beam. This was demonstrated using an array of 3D printed refractive lenses22 that captured 92 % of the beam irradiating the optics, creating up to six spatially separated probes21. In this experiment, the probes were all very similar and arranged in an array with a distance of 46 μm, which is small compared to the detector pixel size of 75 μm. This makes it difficult in the reconstruction to disentangle the diffraction patterns created by the different probes, in particular when the sample is weakly scattering. As a result, artifacts can appear, such as superimposed ghost images of regions scanned by neighboring probes.
This ambiguity in the inverse problem can be avoided by scanning the sample with a set of very different probes. In this work, we present multibeam ptychography with coded probes, which allows us to disentangle the contributions from multiple probes robustly, accelerating the ptychographic reconstruction and utilizing large fraction of the beam irradiating the optics. The wavefront of each probe was individually adjusted by tailor-made phase plates23,24 that were 3D printed together with the focusing optics. This results in highly diverse coded probes and significantly improves the reconstruction quality and speed compared to the same measurement but with similar probes. This approach solves the problem of probe ambiguity in multibeam ptychography and makes the technique reliably applicable in practice.
Results
The X-ray-optical experiments were carried out at the microprobe end-station of beamline P0625 of the synchrotron radiation source PETRA III at DESY in Hamburg. The experimental scheme is illustrated in Fig. 1a. We used an array of 3D-printed compound refractive lenses22 to create multiple probes (photon energy range: 7–9 keV). As our primary goal was to show a practical solution to the problem of disentangling the scattered signal from neighboring beams, for better reader understanding, we designed the lens array to use vertical and horizontal arrangements with only two and three beams. Some of these lens stacks were equipped with phase plates to create vortex beams with topological charges of + 1, − 1, or + 2.
Slits (not shown) in front of the lens array ((A) in Fig. 1a) and a pinhole array ((B) in Fig. 1a) between the lenses and the sample ((C) in Fig. 1a) were used to select various combinations of lenses, defining the arrangement, number and shape of different probes. A Siemens star resolution chart made by NTT-AT (model XRESO-50HC) with smallest features of 50 nm was imaged to prove the performance enhancement. A microchip manufactured by Infineon Technologies AG was imaged to validate the technique with a real-life sample.
Method verification
To investigate the influence of coded probes as compared to similar ones with the most basic case of multibeam ptychography – a two-beam case, we recorded a two-beam ptychogram of a sample (Siemens star) with similar probes and one with coded probes with topological charges of + 1 and − 1, respectively. The same region of the Siemens star was scanned during both scans, using identical scan parameters with probes arranged horizontally (Fig. 1b and 1c.) The size of the scan area was chosen to have an overlap area of 10 μm × 4 μm (v × h), which was scanned by both probes. The total area covered by either probe during the scan was 10 μm × 50 μm (v × h). The scans were recorded in a step-settle-expose mode and took around 15 minutes each, including time for the movement and settling of the scanning stage. Scanning the object with the same fluence using a single beam would have taken 1.9 times longer. Afterwards, both scans were reconstructed using the same algorithm.
Figure 2 shows the reconstructed phase images of the Siemens star after 1000 iterations together with the corresponding complex wave fields of the probes used to illuminate the sample. In the first case using identical probes, the presence of large scale image artifacts on the object is evident (see Fig. 2a): the copy image of the center of the Siemens star located on the right side (marked with the red arrow) and the missing vertical bar beneath the center of the Siemens star (marked with the purple arrow). The center of the reconstructed Siemens star is shown in Fig. 2a1 – the 50 nm spokes of the innermost ring are not properly resolved. In the central area of the scan, which was illuminated by both probes, the artifacts are strongly suppressed. The reconstruction of this scan, with uncoded, similar probes, has revealed two types of artifacts that originate from two different sources: missing or blurry features due to position errors of the scanning stage and ghost copies from other scan regions due to the ambiguity in disentangling the information from similar probing beams.
The first type of artifacts, namely, positioning errors, causing the missing bar (purple arrow in Fig. 2) and blurry 50 nm spokes of the innermost ring (Fig. 2a1), is a well-studied problem in single-beam ptychography, and position refinement26 can compensate for it. One of its drawbacks is that it creates more degrees of freedom in the solution of the inverse problem and significantly increases the total reconstruction time. In cases of severe position errors, the increase in time required for the reconstruction is especially high, as the search for the correct positions has to be performed over a large area. In parallel, the position errors complicate disentangling the diffraction signal from the different probes. As a result, artifacts resulting from the positioning errors show up in corresponding places of sample sub-regions for every beam, affecting the reconstructed image’s quality.
The use of coded probes greatly reduced the ambiguity in the diffraction signal from different regions of the sample affecting both types of artifacts which notably improved the reconstruction quality: Fig. 2b shows the reconstructed image of the Siemens star in the coded probe case. The comparison between Fig. 2a and b clearly shows that the artifacts are strongly reduced. Figure 2b1 illustrates the reconstruction quality: the 50 nm spokes of the innermost ring are well resolved.
Additionally to the improved image quality, we have observed an increase in the convergence speed of the ptychographic reconstruction. Figure 3 shows cropped intermediate reconstructed images from the scans with uncoded and coded probes after 200 iterations and the evolution of the normalized error of both reconstructions. The conspicuous difference in reconstruction quality can be seen when comparing between uncoded - Fig. 3a and coded - Fig. 3b probes: the scan with coded probes has no observable image artifacts while the one with uncoded probes has a ghost image of the Siemens Star center and blurry vertical bar.
The significantly improved reconstruction quality in the case of coded probes allows a direct comparison between multi-beam and single-beam ptychography. For this purpose we have performed a scan with a single beam with the same step size as for the two beam case. Figure 4a and b compare the quality of the reconstructed object using a single probe and using two coded probes, respectively. To quantitatively evaluate the resolution, two line profiles in Fig. 4 were fitted with an error function. The measured widths agree within 10 %, indicating that the resolution and the quality of the reconstruction for the multi-beam scan reach the level of a single-beam scan. Both scans were recorded with the same fluence, the single beam scan covered the area of 10 μm × 10 μm (v × h) with a speed of 18 μm2 min−1 and the two-beam scan covered the area of 10 μm × 52 μm (v × h) with a speed of 34 μm2 min−1.
Microchip imaging
To apply the method to a real-world use case, we imaged a microchip structure produced by Infineon Technologies AG. We used three vertically-arranged coded probes (see Fig. 1d) to image an area of 82 μm × 80 μm with a similar overlap area between neighboring probes as in other measurements. The scan took around 1 hour and 14 minutes, covering the same area with a single probe would have taken roughly 2.7 times longer. Figure 5 shows the entire reconstructed image of the microchip and two enlarged areas to evaluate the reconstruction quality.
To estimate the achieved resolution, the scan was repeated with identical scan parameters, field of view and coded probing beams. The two scans were reconstructed separately using identical reconstruction parameters. The two reconstructions were used to calculate the Fourier ring correlation (FRC)27,28,29 between them, estimating resolution of 95.23 nm (see Fig.6a). To verify that the two scans did not just reconstruct the same artifacts in the same place (something not caught by the FRC, as it measures the similarity of two images, but can not evaluate the correctness of either image), a line profile across a prominent edge (see Fig. 6b) was extracted and fitted with an error function, resulting in a resolution estimate of 114.1 nm (see Fig.6c).
Discussion
The need to image large samples with the highest spatial resolution exists in many fields in science and industry and will increase with future needs in materials design. Due to the limited coherent flux at modern synchrotron radiation sources, multi-beam ptychography is a desirable solution. It enables the usage of the so far discarded incoherent fraction of the beam, increasing photon utilization in the experiment. While the coherent flux is greatly enhanced at the next generation synchrotron radiation sources with ultra-low emittance, the coherent fraction of hard X-rays at these sources will still be far from 100 %. Thus a scanning with multiple beams in ptychography will still be a highly desirable option.
Recent studies showed the potential to increase the acquisition speed or enlarge the scan area by covering the incident beam with closely packed lens arrays with a spacing smaller than the detector’s pixel size21. However, the improved photon utilization and scanning speed, as a result, came along with a decreased image quality. Due to the high ambiguity of the diffraction signal from several similar probes with a small separation and experimental uncertainties, the object reconstruction often contains artifacts and has a reduced resolution. These image artifacts may outweigh the benefit of parallel scanning and prevent the broader use of multi-beam scanning. In addition, the multi-beam reconstruction is more sensitive to unavoidable experimental uncertainties, such as positioning errors or mechanical vibrations.
In the current work, we have demonstrated the practical solution to the drawbacks in multi-beam ptychography by introducing diversity in the probes. We further reduced the probe separation to 26 μm, which positively affected the scanning speed as it allowed to perform a scan of a desired object’s area with fewer steps compared to measurements with larger probe separation. Using individually coded probes for multi-beam ptychography, we could image a larger area of the sample faster and with similar quality as compared to conventional ptychography. Compared to uncoded multi-beams, the unique topological charge of each beam accelerates the reconstruction and leads to faster convergence, here by a factor of three. To assess the reconstructed image quality, we have directly compared the results of the coded multi-beam and conventional single-beam ptychography, achieving a similar resolution. Furthermore, we used coded multi-beams to scan a microchip sample that contains regular and repeating structures30 that are challenging for multi-beam ptychography with similar probes. We imaged an area of 82 μm × 80 μm with no visible artifacts in just 1 hour and 14 minutes in a step-settle-expose scanning mode, while a similar scan using only a single beam would have taken nearly four hours. The proposed method enables rapid imaging of large sample areas with high resolution and high image quality, which is in high demand in many fields of science and industry5,6.
Combining multi-beam ptychography with continuous (fly)-scanning31,32,33 will allow to utilize most of the available X-ray beam and minimize any dead time during the scan. The continuous movement of the sample during each exposure can lead to smearing in the recorded diffraction patterns. The influence of this smearing on the reconstruction quality of multi-beam data is an open question.
Besides the smearing from fly-scanning, other factors could limit reconstruction quality. Mutual partial coherence between adjacent probing beams and thus interference effects between the signal from adjacent beams must be considered, or even better, avoided. By choosing a suitable sample–detector distance, coherence effects can be suppressed through alias cloaking or auto correlation filtering17,34. To avoid partial mutual coherence in the first place, it is possible to either design the X-ray optics aperture/spacing to match the transverse coherence length or, if the beamline allows for it, adjust the coherence length to fit the aperture. This could be done for example with slits to create a secondary source35,36 or by using prefocusing optics37.
It has been shown that the 3D printed X-ray optics can deteriorate and shrink in intense X-ray beams38,39. To counter this, the lenses could be designed with the shrinking in mind, or even pre-treated40 to avoid shrinking during the experiment. Another option is to exchange the optics frequently. As the manufacturing process is inexpensive, it is straightforward to print the same design multiple times. In previous experiments, we have experienced that 3D printed optics can be used for one week of beamtime without showing any deterioration22,41. The actual durability will vary from experiment to experiment and depend on the resist used and the energy and intensity of the photon beam.
The scan area was larger than the distance between adjacent probing beams in all multi-beam scans. Therefore, the scan regions of the individual beams overlap. Nevertheless, this overlap is not necessary for a successful reconstruction. To demonstrate this point, we increased the probe separation in a new reconstruction of the 2-beam scan of the Siemens star. This places the two scan regions so far apart that each scan region only interacts with one beam. The successful reconstruction is shown in the supplementary information (S1). Both sample regions were faithfully reconstructed without sharing any information in the sample plane. The reconstructed object is now split into two parts, and it requires stitching to obtain a single image of the whole sample. Overlapping the scan regions from adjacent probing beams avoids this post-processing as it forces the scan regions to match in the overlapping areas. This increases the data redundancy in the whole dataset, which also helps the reconstruction.
Methods
Scanning setup
The experimental studies were performed at the microprobe end-station of beamline P0625 of the synchrotron radiation source PETRA III at DESY in Hamburg, Germany. The experimental scheme is illustrated in Fig. 1a. We used an array of 3D-printed compound refractive lenses22 (A in Fig. 1a) to create multiple probes by focusing X-rays. A single lens has a square footprint of 25 µm × 20 μm and is 290 μm high. The single-lens aperture was illuminated by the X-ray beam with a flux of 2.2 × 107 photons/sec. The lenses are placed in a 26 μm raster to avoid printing defects between neighboring lenses, leaving a 1 μm gap between them. Such an approach to forming probe arrays has a significant advantage as it leaves very little space not occupied by the optics and, as a result, allows higher photon utilization: (lens width/probe spacing)2 × 100%. Here, the experiments were performed with the probes arranged inline, leaving no gaps in the second direction. This yields a photon utilization of 50/51 × 100% = 98% for the two-beam case, and 75/77 × 100% = 97.4% for the three-beam case. These percentages are relative to the aperture of the lens stacks used and the spacing between them and thus represent the effectiveness of the arrangement of the X-ray optics. Assuming a homogeneous illumination of the lens stack, the number of photons utilized to probe the sample scales with the number of beams used, namely 196% for the two-beam case and 292.2% for the three-beam case.
Each lens comprises six single parabolic surfaces with an apex curvature of 2.5 μm. At the photon energy of 7 keV these lens stacks had a focal length of about 250 mm. Some of the lenses were equipped with phase plates creating beams with orbital angular momentum (OAM) with topological charges of + 1, − 1, or + 2, respectively24. This was realized by vortex phase plates with a smooth helical phase ramp. The phase plates and the lenses were printed together as one solid piece and are thus made from the same material. An individual phase plate covers the whole aperture of the lens it is placed on. The height of the phase plates spiral was designed to match a phase shift of \(2\pi\) at 7 keV.
The upstream slits in front of the lens array were used to select different lens combinations. Background scattering from the lenses was reduced by placing a pinhole array between the lens array and the sample (B in Fig. 1a). The pinhole array was made from 100 μm thick platinum foil by laser ablation, with a pinhole size of 20 μm. As samples (C in Fig. 1a), a Siemens star resolution chart made by NTT-AT (XRESO-50HC) with smallest features of 50 nm and a microchip manufactured by Infineon Technologies AG was used. For the measurements with the Siemens star sample and microchip a photon energy of 7keV and 9keV were used, respectively.
The sample was placed into the approximate focal plane, and the diffracted beam was recorded with a single-photon counting, custom made in-vacuum Eiger 4M detector (pixel size 75 μm)42. The detector was mounted at the end of an evacuated flight tube to eliminate absorption and scattering in air4,25. The distance from the sample to the detector was 8200 mm for the Siemens star sample and 7100 mm for the microchip.
Figure 1b and c illustrate the uncoded and the coded lens combinations used in the measurements with the test sample, while (d) shows the coded three beam combination that was used to image the microchip. Prior to the multi-beam measurement, we verified the lenses by characterizing each beam individually using single-beam ptychography. Fig. 1e illustrates the complex wavefields of the probes in the object plane.
In the single beam measurement, the Siemens star was scanned in a grid of 21 × 21 = 441 (v × h) positions in 0.5 μm steps, with an exposure time of 0.5 s per point, covering a total area of 10 μm × 10 μm. The scan took around 5 minutes, including time for the movement of the scanning stage.
For the two-beam measurements with the Siemens star, the sample was scanned in a grid of 21 × 61 = 1281 (v × h) positions in 0.5 μm steps, with an exposure time of 0.5 s per point. Since the lenses were placed in a 26 μm raster, the scan areas of the two probes overlap on a 10 μm × 4 μm (v × h) wide area. The scan took around 15 minutes, including time for the movement of the scanning stage. With both probes the total area covered by the scan was 10 μm × 52 μm. The gain in scanning speed over the conventional single-beam scanning, in this case, was 1.9.
For the three beam measurement of the microchip, the sample was scanned in a grid of 61 × 161 = 9861 (v × h) positions in 0.5 μm steps with an exposure time of 0.2 s per point. The scan areas of the neighboring probes overlap on a 10 μm × 80 μm (v × h) wide area. With all three probes the total area covered by the scan was 82 μm × 80 μm. The gain in scanning speed over the conventional single-beam scanning, in this case, was 2.7. The total thickness of the imaged chip was 150 μm, while all visible features were in stacked embedded layers with a total thickness of 15 μm. The lateral size of the shown repeating features in these different layers varied between 2.5 μm and 250 nm.
Reconstruction approach
For the reconstructions, the recorded far-field diffraction patterns were cropped to 512 × 512 pixels centered around the beam axis. For the single beam scan, the image was reconstructed with the standard ePIE algorithm43 for 500 iterations, followed by the modified coupled update44 for another 500 iterations. The pixel size in the reconstruction is 37.8 nm. To compensate for positioning errors of the scanning stage, the scan positions were numerically refined every 50 iterations to improve the reconstruction26.
In the two-beam scan, the experimental parameters were identical to the single beam scan, resulting in the same pixel size in the reconstruction. Following the same procedure as in Wittwer, et al.21, the image was reconstructed using 500 iterations of a modified multiprobe sPIE45 followed by 500 iterations of the modified coupled update44. We again used a local position refinement every 50 iterations.
For the three-beam scan of the microchip, the photon energy and detector distance were different, resulting in a pixel size of 25.4 nm in the reconstruction. Because we performed the experiment at a higher energy, the size of the probes illuminating the sample increased. We changed the reconstruction scheme for this scan: for the first 500 iterations we used the modified multiprobe sPIE41,45 with a 2 × 2 upsampling followed by 1500 iterations of the modified coupled update44. Also here, a local position refinement was used every 50 iterations.
References
Pfeiffer, F. X-ray ptychography. Nat. Photon. 12, 9–17. https://doi.org/10.1038/s41566-017-0072-5 (2018).
Rodenburg, J. & Maiden, A. Ptychography, 2–2 (Springer, 2019).
Schropp, A. et al. Hard x-ray scanning microscopy with coherent radiation: Beyond the resolution of conventional x-ray microscopes. Appl. Phys. Lett. 100, 253112 (2012).
Reinhardt, J. et al. Beamstop-based low-background ptychography to image weakly scattering objects. Ultramicroscopy 173, 52–57. https://doi.org/10.1016/j.ultramic.2016.11.005 (2017).
Holler, M. et al. High-resolution non-destructive three-dimensional imaging of integrated circuits. Nature 543, 402–406. https://doi.org/10.1038/nature21698 (2017).
Guizar-Sicairos, M. et al. High-throughput ptychography using eiger: Scanning x-ray nano-imaging of extended regions. Opt. Express 22, 14859–14870. https://doi.org/10.1364/OE.22.014859 (2014).
Jiang, Y. et al. Achieving high spatial resolution in a large field-of-view using lensless x-ray imaging. Appl. Phys. Lett. 119, 124101. https://doi.org/10.1063/5.0067197 (2021).
Grunwaldt, J.-D. & Schroer, C. G. Hard and soft x-ray microscopy and tomography in catalysis: Bridging the different time and length scales. Chem. Soc. Rev. 39, 4741. https://doi.org/10.1039/c0cs00036a (2010).
Meirer, F. & Weckhuysen, B. M. Spatial and temporal exploration of heterogeneous catalysts with synchrotron radiation. Nat. Rev. Mater. 3, 324–340. https://doi.org/10.1038/s41578-018-0044-5 (2018).
Becher, J. et al. Mapping the pore architecture of structured catalyst monoliths from nanometer to centimeter scale with electron and x-ray tomographies. J. Phys. Chem. C 123, 25197–25208. https://doi.org/10.1021/acs.jpcc.9b06541 (2019).
Beckers, M. et al. Chemical contrast in soft x-ray ptychography. Phys. Rev. Lett. 107, 208101 (2011).
Hoppe, R. et al. High-resolution chemical imaging of gold nanoparticles using hard x-ray ptychography. Appl. Phys. Lett. 102, 203104 (2013).
RAVEN BAA.https://www.iarpa.gov/index.php/research-programs/raven/raven-baa (2015).
Tavares, P. F., Leemann, S. C., Sjöström, M. & Andersson, Å. The MAX VI storage ring project. J. Synchrotron Rad. 21, 862–877 (2014).
Raimondi, P. ESRF-EBS: The extremely brilliant source project. Synchr. Rad. News 29, 8–15 (2016).
Schroer, C. G. et al. PETRA IV: The ultra-low emittance source project at DESY. J. Synchrotron Rad. 25, 1277–1290 (2018).
Bevis, C. et al. Multiple beam ptychography for large field-of-view, high throughput, quantitative phase contrast imaging. Ultramicroscopy 184, Part A, 164–171 https://doi.org/10.1016/j.ultramic.2017.08.018.(2017).
He, X. et al. High-speed ptychographic imaging based on multiple-beam illumination. Opt. Express 26, 25869–25879. https://doi.org/10.1364/OE.26.025869 (2018).
Hirose, M., Higashino, T., Ishiguro, N. & Takahashi, Y. Multibeam ptychography with synchrotron hard X-rays. Opt. Express 28, 1216. https://doi.org/10.1364/OE.378083 (2020).
Yao, Y. et al. Multi-beam X-ray ptychography for high-throughput coherent diffraction imaging. Sci. Rep. 10, https://doi.org/10.1038/s41598-020-76412-8(2020).
Wittwer, F. et al. Upscaling of multi-beam x-ray ptychography for efficient x-ray microscopy with high resolution and large field of view. Appl. Phys. Lett. 118, 171102. https://doi.org/10.1063/5.0045571 (2021).
Lyubomirskiy, M. et al. Ptychographic characterisation of polymer compound refractive lenses manufactured by additive technology. Opt. Express 27, 8639. https://doi.org/10.1364/OE.27.008639 (2019).
Seiboth, F. et al. Perfect X-ray focusing via fitting corrective glasses to aberrated optics. Nat. Commun. 8, 14623. https://doi.org/10.1038/ncomms14623 (2017).
Seiboth, F. et al. Refractive hard x-ray vortex phase plates. Opt. Lett. 44, 4622–4625. https://doi.org/10.1364/OL.44.004622 (2019).
Schroer, C. G. et al. Ptychographic Nano-Analytical Microscope (PtyNAMi) at PETRA III: Signal-to-background optimization for imaging with high sensitivity. In X-Ray Nanoimaging: Instruments and Methods IV, vol. 11112, 111120D, (International Society for Optics and Photonics, 2019) https://doi.org/10.1117/12.2529096.
Schropp, A. et al. Full spatial characterization of a nanofocused x-ray free-electron laser beam by ptychographic imaging. Scientific Reports 3, https://doi.org/10.1038/srep01633(2013).
van Heel, M. & Schatz, M. Fourier shell correlation threshold criteria. J. Struct. Biol. 151, 250–262. https://doi.org/10.1016/j.jsb.2005.05.009 (2005).
Banterle, N., Bui, K. H., Lemke, E. A. & Beck, M. Fourier ring correlation as a resolution criterion for super-resolution microscopy. J. Struct. Biol. 183, 363–367. https://doi.org/10.1016/j.jsb.2013.05.004 (2013).
Horstmeyer, R., Heintzmann, R., Popescu, G., Waller, L. & Yang, C. Standardizing the resolution claims for coherent microscopy. Nat. Photon. 10, 68–71. https://doi.org/10.1038/nphoton.2015.279 (2016).
Tuitje, F. et al. Reliability of ptychography on periodic structures. OSA Continuum 3, 1691–1702. https://doi.org/10.1364/OSAC.394384 (2020).
Clark, J. N., Huang, X., Harder, R. J. & Robinson, I. K. Continuous scanning mode for ptychography. Opt. Lett. 39, 6066–6069. https://doi.org/10.1364/OL.39.006066 (2014).
Deng, J. et al. Continuous motion scan ptychography: characterization for increased speed in coherent x-ray imaging. Opt. Express 23, 5438–5451. https://doi.org/10.1364/OE.23.005438 (2015).
Huang, X. et al. Artifact mitigation of ptychography integrated with on-the-fly scanning probe microscopy. Appl. Phys. Lett. 111, 023103. https://doi.org/10.1063/1.4993744 (2017).
Wittwer, F. Development and Study of Refractive Phase Retrieval and X-ray Multibeam Ptychography. Ph.D. thesis, Staats-und Universitätsbibliothek Hamburg Carl von Ossietzky (2020).
Björling, A. et al. Ptychographic characterization of a coherent nanofocused x-ray beam. Opt. Express 28, 5069–5076. https://doi.org/10.1364/OE.386068 (2020).
Kahnt, M. et al. Measurement of the coherent beam properties at the CoSAXS beamline. J. Synchr. Radiat. 28, 1948–1953. https://doi.org/10.1107/S1600577521009140 (2021).
Schroer, C. G. & Falkenberg, G. Hard X-ray nanofocusing at low-emittance synchrotron radiation sources. J. Synchr. Radiat. 21, 996–1005. https://doi.org/10.1107/S1600577514016269 (2014).
Petrov, A. K. et al. Polymer x-ray refractive nano-lenses fabricated by additive technology. Opt. Express 25, 14173–14181. https://doi.org/10.1364/OE.25.014173 (2017).
Barannikov, A. et al. Optical performance and radiation stability of polymer X-ray refractive nano-lenses. J. Synchr. Radiat. 26, 714–719. https://doi.org/10.1107/S1600577519001656 (2019).
Sharipova, M. et al. Effect of pyrolysis on microstructures made of various photoresists by two-photon polymerization: Comparative study. Opt. Mater. Express 11, 371–384 (2021).
Wittwer, F. et al. Ptychography with a virtually enlarged illumination. Microsc. Microanal. 24, 48–49. https://doi.org/10.1017/S1431927618012667 (2018).
Johnson, I. et al. Eiger: A single-photon counting x-ray detector. J. Instrum. 9, C05032–C05032. https://doi.org/10.1088/1748-0221/9/05/C05032 (2014).
Maiden, A. M. & Rodenburg, J. M. An improved ptychographical phase retrieval algorithm for diffractive imaging. Ultramicroscopy 109, 1256–1262. https://doi.org/10.1016/j.ultramic.2009.05.012 (2009).
Thibault, P. et al. High-resolution scanning X-ray diffraction microscopy. Science 321, 379–382. https://doi.org/10.1126/science.1158573 (2008).
Batey, D. J. et al. Reciprocal-space up-sampling from real-space oversampling in x-ray ptychography. Phys. Rev. A 89, 043812. https://doi.org/10.1103/PhysRevA.89.043812 (2014).
Acknowledgements
We acknowledge DESY (Hamburg, Germany), a member of the Helmholtz Association HGF, for the provision of experimental facilities. Parts of this research were carried out at PETRA III within experiment 11006769 and we would like to thank Gerald Falkenberg for assistance in using the beamline P06. This research was supported in part through the Maxwell computational resources operated at DESY and by Ptychography 4.0, a project of the Helmholtz Incubator on Information and Data Science. M.S. acknowledges funding from the Deutsche Forschungsgemeinschaft (DFG) - SFB 1441 - Project-ID426888090.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
M.L., F.W. and M.K. conceived the experiments, F.K., A.K. and C.D manufactured the optics. M.L., F.W., M.K., M.S., K.V.F. and J.G. conducted the experiments, M.L., F.W., M.K. and M.S. analysed the results. M.L, F.W., M.K. and C.G.S wrote the article. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lyubomirskiy, M., Wittwer, F., Kahnt, M. et al. Multi-beam X-ray ptychography using coded probes for rapid non-destructive high resolution imaging of extended samples. Sci Rep 12, 6203 (2022). https://doi.org/10.1038/s41598-022-09466-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-09466-5
- Springer Nature Limited