Introduction

Fractures of the upper extremity are associated with individual physical impairment and often high socioeconomic relevance. While distal radius fractures are among the most frequent trauma consequences in any emergency department with particularly high incidence in elderly patients1,2,3, dislocations and fractures in the elbow region are more common in younger populations4. Plain radiographs usually represent the primary imaging method if traumatic injuries of the upper limb are suspected5. Radiograms offer a compromise between fast scan time, ubiquitous availability, and low radiation dose on the one side, and limited 2D image information on the other side6. Considering that trauma mechanisms can be vastly heterogeneous, ranging from simple falls on the outstretched hand from standing height to high-velocity injuries in sports activities and motor vehicle accidents2,7, 2D imaging may be insufficient for trauma evaluation in some patients. Especially more severe injuries, e.g., multi-fragmentary, intraarticular and displaced fractures of the wrist and elbow require additional 3D assessment for presurgical planning8 and image fusion with intraoperative fluoroscopy9,10. In addition, computed tomography scans allow for reliable exclusion or detection of radiographically occult fractures in case of divergent clinical presentation11. Even for certain chronic conditions or postoperative follow-up imaging, CT can provide valuable diagnostic information with the typical drawback of increased radiation dose12,13.

Despite the predominance of multidetector CT in most radiology departments, the value of another technical approach to computed tomography is increasingly recognized for skeletal imaging tasks: Combining a flat-panel detector build with pyramid-shaped beam geometry, cone-beam computed tomography (CBCT) has become a powerful tool for 3D imaging in trauma patients. Detailed visualization of bone microarchitecture14, superior positioning options and potential for reduced radiation dose are among the major advantages compared to conventional multidetector CT15. Having been considered an integral part of maxillofacial imaging for a long time16,17,18,19, access to dedicated scanners for the appendicular skeleton has led to increasing relevance of CBCT in musculoskeletal imaging during the last decade20.

The twin robotic x-ray system in this study represents a novel approach to the CBCT formula that uses two telescopic arms for 2D and 3D image acquisition instead of a conventional gantry-based approach. Whilst the benefits of the CBCT option for fracture detection have previously been analyzed, data regarding the actual dose reduction potential of this multi-use system is lacking. Therefore, in this work, various combinations of acquisition and reconstruction parameters were evaluated for the CBCT mode of the twin robotic x-ray system in cadaveric wrist and elbow scans, aiming to define the best possible trade-off between image quality and radiation dose.

Material and methods

Cadaveric phantoms

For this experimental study, we received permission from the institutional review board of the University of Würzburg, Germany (IRB number 20200506 01). Two formalin-fixed cadaveric specimens were obtained from the anatomical institute of the local university. Body donors had voluntarily donated their bodies to the anatomical institute for study and scientific purposes. Thus, the ethics committee of the University of Würzburg waived additional written informed consent. Bilateral wrist and elbow examinations were performed for both cadaveric specimens in supine position using the twin robotic x-ray system’s tableside scan trajectory. Scanning process for upper extremity imaging is displayed in Fig. 1 by a staff member (informed consent for publication was obtained). All procedures were conducted in accordance with the relevant guidelines and regulations.

Figure 1
figure 1

Rendering of staff member demonstrating the scan position for 3D cone-beam CT imaging of upper extremity joints using the tableside trajectory of the twin robotic x-ray system.

Cone-beam CT scans

For image acquisition, the multifunctional radiography system (Multitom Rax, Siemens Healthcare GmbH, Erlangen, Germany) uses two telescopic arms connected to ceiling rails within the x-ray suite. One of the arms holds a quadratic flat-panel detector (edge length 426 mm), while the other carries the x-ray tube. Different motion trajectories for both arms allow for 2D and 3D imaging in specified positions. CBCT scans are performed with an asymmetric source-to-image distance of 1150 mm, an input field of 213 × 213 mm2 and a 2D matrix of 1440 × 1440 un-binned pixels in high-resolution mode, which results in an effective pixel size of 149 µm. The current software version VF 11 (Siemens Healthcare GmbH) supports acquisition of 26 projection images per second with a sweep angle of 200 degrees, hence resulting in a total scan time of 14 s.

Before each CBCT scan, system settings can be manually adjusted to achieve target dose levels. Using a clinically established full-dose protocol (FD) as reference, four other scan protocols with lower radiation doses were evaluated in this study. Images were either acquired with different fixed combinations of tube voltage and tube current–time product (low-dose [LD], ultra-low-dose [ULD]) or with the scanner’s automatic dose modulation feature (modulated low-dose [mLD], modulated ultra-low-dose [mULD]): In this scan mode, a sensor implemented in the flat-panel detector measures the incoming radiation for each projection image. Dependent on the previously determined dose level, the tube current–time product is then adjusted continuously throughout the acquisition process. Scan parameters and corresponding radiation dose values associated with each CBCT examination protocol are summarized in Table 1. Dose-area products for all examinations were obtained from the automatically created scan report. To calculate volume CT dose indices (CTDIvol), multiplication by a linear scaling factor that we computed in advance for every combination of tube current–time product and tube voltage was needed. Therefore, dose-length product measurements were performed in the five chambers of a conventional 16 cm PMMA dosimetry phantom (IEC 60601-2-44:2009 compliant). After standard weighting schemes were applied to obtain volume dose-length product values (DLPvol(16 cm)), these values were divided by the field of view in z-direction (equal to the beam width) to compute the CTDIvol(16 cm). Finally, dividing the CTDIvol(16 cm) by the DAP resulted in the required scaling factor.

Table 1 Scan protocols and radiation dose.

Reconstruction kernels

Each set of projection data was reconstructed with dedicated 3D processing software (syngo via, Siemens Healthcare GmbH) using different modulation transfer functions characterized by standard convolution kernels: very sharp (Ur77): ρ10 = 25.4 lp/cm, ρ50 = 16.7 lp/cm; sharp (Br69): ρ10 = 15.1 lp/cm, ρ50 = 12.3 lp/cm; intermediate (Br62): ρ10 = 11.3 lp/cm, ρ50 = 9.2 lp/cm (all Siemens Healthcare GmbH). Reformatting was performed in orthogonal planes (axial, coronal and sagittal) with slice thickness of 1.0 mm, increment of 0.5 mm, image matrix of 1024 × 1024 pixels and field of view of 80 mm. Window settings were 3000 and 1000 Hounsfield units (width and center) by default with observers being allowed to modify the contrast manually on screen for their reads.

Image analysis

After image reconstruction, all datasets were analyzed in randomized and blinded fashion with standard picture archiving and communication software (Merlin, Phoenix-PACS, Freiburg im Breisgau, Germany) by five independent radiologists with 8, 7, 6, 5 and 4 years of experience in musculoskeletal imaging. First, each observer was tasked to evaluate whether the reviewed scan was sufficient for diagnostic use. Second, readers assessed overall image quality and gave dedicated image quality ratings for cortical bone, cancellous bone and soft tissue using a seven-point Likert scale (7 = excellent, 6 = very good, 5 = good, 4 = satisfactory, 3 = fair, 2 = poor, 1 = very poor image quality).

Statistics

All tests and analyses were conducted using specialized software (SPSS Statistics Version 27 for Mac, IBM, Armonk, New York, USA) with p values less than 0.05 indicating statistical significance. Categorical items are reported in form of percentages, frequencies and median values. Wilcoxon signed rank and Friedman tests were carried out to compare two or more paired non-parametric variables. Interrater reliability was assessed by calculation of Fleiss’ kappa with interpretation of agreement following Landis and Koch (1.00–0.81 = almost perfect; 0.80–0.61 = substantial; 0.60–0.41 = moderate; 0.40–0.21 = fair; 0.20–0.00 = slight; < 0.00 = poor agreement)21.

Results

Dose comparison between scan protocols

Operating with fixed tube current–time products, CTDIvol(16 cm) of low-dose and ultra-low-dose scans was 4.5 and 1.2 mGy, corresponding to 74.2% and 93.1% of dose reduction compared to the standard full-dose scan protocol used in clinical routine (17.4 mGy). With automatic dose regulation activated, modulated low-dose (mean CTDIvol(16 cm) 0.6 mGy) and modulated ultra-low-dose acquisition (mean CTDIvol(16 cm) 0.3 mGy) allowed for dose reduction of 96.6% and 98.3% (p < 0.001). Scan parameters and dose values were equal for wrist and elbow scans.

Image quality assessment

Independent of the convolution kernel used for image reconstruction, overall image quality of every FD, LD and ULD scan in this study was deemed suitable for diagnostic evaluation in clinical routine by all five radiologists. In contrast, readers considered 40/45/30% of mLD and 100/97/95% of mULD (Ur77/Br69/Br62 kernel) insufficient for clinical imaging. Comparing reconstructions with the same kernel, studies with higher dose levels produced superior overall image quality in upper extremity scans (all p < 0.001). Detailed observer ratings are outlined in Table 2. The best overall image quality was reported for FDUr77 examinations (median scale value 7). Evaluation of LDUr77 studies delivered favorable results compared to FDBr62 (p < 0.001) and comparable image quality to FDBr69 scans (p = 0.157). No distinction was stated between LDBr69 and FDBr62 (p = 0.475). For examinations with the ULD scan protocol, at least satisfactory image quality (scale value ≥ 4) was declared in 50/58/40% of studies with no significant difference between image processing algorithms (median 3.5/4/3; p = 0.072). Fleiss’ kappa was 0.618 (95% confidence interval 0.594–0.641; p < 0.001), indicating substantial interrater reliability for image quality assessment. Figures 2 and 3 contain representative CBCT images in axial and coronal orientation for visualization of image quality.

Table 2 Overall image quality assessment.
Figure 2
figure 2

Axial reformatting of 3D wrist scans with varying acquisition parameters and reconstruction kernels: (A) Full-dose mode (A1: very sharp; A2: sharp; A3: intermediate). (B) Low-dose mode (B1: very sharp; B2: sharp; B3: intermediate). (C) Ultra-low-dose mode (C1: very sharp; C2: sharp; C3: intermediate). (D) Modulated low-dose mode (D1: very sharp; D2: sharp; D3: intermediate). (E) Modulated ultra-low-dose mode (E1: very sharp; E2: sharp; E3: intermediate).

Figure 3
figure 3

Coronal reformatting of 3D elbow scans with varying acquisition parameters and reconstruction kernels: (A) Full-dose mode (A1: very sharp; A2: sharp; A3: intermediate). (B) Low-dose mode (B1: very sharp; B2: sharp; B3: intermediate). (C) Ultra-low-dose mode (C1: very sharp; C2: sharp; C3: intermediate). (D) Modulated low-dose mode (D1: very sharp; D2: sharp; D3: intermediate). (E) Modulated ultra-low-dose mode (E1: very sharp; E2: sharp; E3: intermediate).

Comparing the effect of each convolution kernel on the display of bone microarchitecture, visualization of cortical and cancellous bone was best in very sharp reconstructions of full-dose and low-dose examinations (p < 0.001), with LDUr77 scans also providing favorable bone depiction compared to FDBr69 and FDBr62 (p < 0.001). Accordingly, sharper ULDUr77 (p = 0.005) and ULDBr69 (p = 0.009) reconstructions were deemed more suitable for differentiation of trabecula and fatty marrow than softer ULDBr62 studies. The effect of the different convolution kernels on image sharpness is exemplified in Fig. 4. Quality of cortex display in ultra-low-dose scans was independent of image processing (p = 0.764). For soft tissue analysis, no difference between kernels was identified in full-dose (p = 0.302) and low-dose image acquisition (p = 0.129). In ultra-low-dose studies, both ULDBr69 (p = 0.008) and ULDBr62 (p < 0.001) provided superior soft tissue assessability compared to ULDUr77 with no difference between the two softer kernels (p = 0.739). Detailed image quality evaluation for bone and soft tissue is summarized in Table 3.

Figure 4
figure 4

Cinematic volume rendering technique (syngo.via, Siemens Healthcare GmbH) demonstrates the difference in image sharpness provided by the three convolution kernels for full-dose elbow studies: (A) Very sharp kernel (Ur77). (B) Sharp kernel (Br69). (C) Intermediate kernel (Br62).

Table 3 Detailed image quality assessment for bone and soft tissue.

Discussion

In this multi-observer study, we evaluated the image quality provided by 15 combinations of acquisition protocols and reconstruction parameters for the CBCT scan mode of a twin robotic x-ray system in cadaveric wrist and elbow imaging. We were able to show that dose reduction of more than 90% is feasible for certain imaging tasks, while maintaining diagnostic image quality with a dedicated ultra-low-dose scan protocol. Reducing the radiation dose even further by implementing detector-based dose modulation resulted in a considerable amount of non-diagnostic studies. Therefore, we postulate that the ultra-low-dose protocol with reduced tube voltage represents the best trade-off between required dose and image assessability for upper extremity scans.

A recent meta-analysis by Nardi et al. identified an average dose of 7.1 µSv for CBCT scans of the appendicular skeleton, stating considerable dose reduction potential in comparison to multidetector CT imaging22. Compared to conventional radiography, however, radiation dose of CBCT remains significantly higher in most studies23. While scan protocols without a focus on dose reduction may have the potential to overcome the image quality of state-of-the-art multidetector CT scanners, we believe that CBCT can also challenge radiography as the primary method of fracture diagnosis24. Due to distance to the radiation-sensitive body trunk, effective dose of extremity imaging is generally lower than for other body regions25. Providing diagnostic 3D image information with approximately four times the effective dose of standard 2D radiograms, the ultra-low-dose protocol in this study appears promising for primary use in particular imaging tasks, e.g., carpal bone14,23 or radial head evaluation26. Furthermore, pediatric radiology could benefit from the reduced radiation dose in cases that require 3D assessment to ascertain the injury pattern, e.g., in distal humerus fractures26. To investigate which scan protocol is appropriate for different pathologies should be part of future research. From a technical point of view, the maintained image quality in ultra-low-dose scans may foremost be attributed to the small effective pixel size of 149 µm realized by the combination of asymmetric acquisition geometry (the detector rotates in a smaller radius around the isocenter than the x-ray tube to balance the effect of the focal spot size) and un-binned readout of the flat-panel detector in high-resolution mode.

With very sharp reconstructions of low-dose examinations providing favorable image quality over full-dose scans with softer reformatting, additional dose reduction potential can be realized through image postprocessing. In particular, the superior display of cancellous bone in studies reconstructed with the two sharper convolution kernels may aid fracture assessment due to better discrimination of trabecula and fatty bone marrow. In contrast, the inferior visualization of bone microarchitecture provided by the intermediate convolution kernel might be attributed to the fact that the realized modulation transfer function of this kernel is far below the system’s resolution limit without a defined kernel.

Limitations

The two formalin-fixed cadaveric specimens were selected without knowledge on body donor age, time of fixation and bone density of the examined wrist and elbow region. Thus, preexisting osteopenia and the demineralizing effect of formalin on bone tissue could have had a negative effect on subjective image quality ratings27,28. With examinations in this study limited to cadaveric limbs, the potential effect of motion artifacts in future patient scans could not be estimated. Given the limited number of projection images acquired per scan, the influence of off-center positioning on image quality ought to be evaluated in further studies. However, with motion compensation algorithms continuously improving and comfortable positioning being facilitated by the open scanner architecture, the authors do not expect a considerable loss of image quality in a realistic clinical scenario. Finally, despite being blinded to scan and reconstruction parameters, radiologists could have gained a certain amount of involuntary training during their reading sessions.

Conclusion

With the cone-beam CT scan mode of a twin robotic x-ray system, dose reduction of more than 90% can be realized for certain upper extremity joint examinations, while maintaining diagnostic image quality. Application of sharper convolution kernels for image reconstruction facilitates superior display of bone microarchitecture, potentially aiding in detection of microfractures in future patient studies.