Introduction

Quantitative phase microscopy (QPM) is a powerful technique which provides the measurement of different parameters of cells including their dynamics such as thickness and refractive index fluctuations. QPM is also used to determine cell’s dry mass density, i.e., non-aqueous content1. High space-bandwidth product (SBP) with high spatial phase sensitivity is indispensable for QPM system. The SBP in QPM is defined as the product of the area of coherence limited interference field of view (IFOV) and the area of the spatial frequency band of an interferometric image2. To achieve high SBP in single-shot QPM systems, a highly temporally and spatially coherent (such as laser) light source is generally utilized to obtain the interference pattern easily throughout the FOV of the camera3,4. However, high temporal and spatial coherence (SC) properties of laser light source (Fig. 1a) degrade the image quality due to coherent noise and parasitic fringe formation due to multiple reflections from different surfaces of the optical components5. As a consequence, it reduces the spatial phase sensitivity and height measurement accuracy of the system.

Figure 1
figure 1

Effect of coherence properties of different light sources in QPM: (ad) Exhibit light field illumination wavefront profiles in case of laser, halogen lamp, LED and PTLS, respectively. (e) Schematic Linnik interferometer based QPM system. BS beam splitter, L lens, M mirror, MO1,2 microscope objectives. (f) Interference field of view (IFOV) and noise comparison of different light sources employed in QPM system.

The spatial phase sensitivity of QPM can be improved by employing broadband light sources like white light (halogen lamp) and light emitting diodes (LEDs) (Fig. 1b,c)6,7,8. These light sources provide high phase sensitivity due to their low temporal coherence (TC) length compared to narrow-band lasers. It is well known that the interference pattern occurs only when the optical path length difference between the object and the reference beam is within the coherence length of the light source9,10. Therefore, it becomes difficult to obtain interference pattern quickly while employing low TC length light sources (coherence length lc ~ 1–6 µm). Moreover, high fringe density of the interference signal over the whole camera FOV cannot be obtained in case of low TC equipped off-axis QPM systems based on Linnik and Mach Zehnder interferometers (Fig. 1b,f). This overall reduces the SBP of QPM system due to coherence limited interference field of view (IFOV). However, the SBP can be improved with on-axis digital holography which can utilize whole FOV of camera at the cost of low acquisition speed. On axis holography requires multiple interferograms for phase recovery of specimens11,12,13. This limits the ability to study the live cell dynamics of the biological cells like sperm and adds complexity to the system. In addition, the interferometric system becomes more sensitive to the external vibrations and introduce fringe like modulation error in the reconstructed phase maps12. Moreover, the use of spectrally broad band light sources in interferometry systems requires dispersion compensation and chromatic aberration corrected optical components9,10.

The high fringe density over the whole camera FOV can be achieved by coherence plane tilting even for broadband light sources7,14. The optical configurations utilize diffraction grating into the beam path for coherence plane tilting7,14. The white light diffraction phase microscopy (WL-DPM) is one of the techniques which works on the same principle to generate high fringe density with white light at the detector14. In WL-DPM, grating splits the input beam into several orders: zero, + 1 and higher orders. One of the beams (say 0 order) is then spatially filtered to generate reference beam which interferes with the object beam (+ 1 order) to form modulated intensity patterns. To generate the reference beam, a pinhole of large diameter (~ 200 µm) is inserted into the beam path which does not completely block the object information and leads to the formation of halo in the recovered phase maps14. The imaging FOV in off-axis interference microscopy system has also been increased using τ interferometer in the past15. However, the recovered phase images suffered from the coherent noise generated due to the implementation of laser diode in the interferometer3. Thus, providing high sensitivity (< 5 mrad) together with large FOV in a single-shot QPM is still unaccomplished.

Recently, pseudo-thermal light source (PTLS) has attracted attention of the researchers in the field of QPM16,17,18,19. It is also called dynamic speckle pattern interferometry20,21. Such light sources have been implemented in the quantitative phase imaging of optical waveguides and biological specimens previously19,20,21,22,23. The PTLS is generated by passing a highly temporal coherent laser beam through a rotating diffuser or a rough surface (Fig. 1d). The output of the diffuser acts as a highly temporally coherent (depends on the spectral bandwidth of the laser) and partially spatially coherent (depends on the source size) light source. The experimental scheme to obtain interference pattern corresponding to all types of light sources is depicted in Fig. 1e. PTLS carries the advantages of both broadband such as halogen lamp, LEDs in terms of high spatial phase sensitivity and narrowband such as laser sources in terms of the generation of high-density interference fringes over whole camera FOV and high photon degeneracy (Fig. 1f), which is not convincingly covered in the literature. PTLS does not suffer from chromatic aberration of the optical components and dispersion issues from the biological specimens as it has narrow temporal spectral bandwidth like laser. Here, we exhibit pros and cons of the coherence properties of different types of light sources in the field of QPM for enhancing the SBP, phase sensitivity and temporal resolution. We demonstrate that PTLS can deliver single-shot phase imaging over large FOV without sacrificing the spatial phase sensitivity and the temporal resolution with imaging speed limited only by the camera acquisition speed. In addition, the interference fringes can be formed over the extended range of OPD between the object and the reference arm of QPM system unlike low TC light sources (Fig. 1f).

In Refs.19,21, it has been implemented with a spectrally broad bandwidth (Δλ = 17 nm and 170 nm) light source (Ti:sapphire laser and supercontinuum light) which restricted the OPD adjustment within the temporal coherence length of the light source. The spectral bandwidths (17 nm and 170 nm) of the light sources are quite large compared to the bandwidth of narrowband laser and cannot be considered spectrally pure. This would arise the problem of optical dispersion due to the biological samples and also require chromatic aberration free optical components to obtain aberration free images. In addition, identical objective lenses must be used in both the object and reference arm of the phase microscopy system to minimize/compensate chromatic aberration.

With this type of large spectral bandwidth light sources, non-identical objective lenses in the object and the reference arm will disturb the interference condition, i.e., optical path difference less than the temporal coherence length, which is required to obtain interference fringes in QPM system. In addition, it might generate different amount of chromatic aberration and fails to compensate its effect in the interferometric images. This would lead to the degradation of interference fringe contrast and subsequently affects the spatial phase sensitivity of the system due to the reduced dynamic range of the recorded interferograms. Moreover, the use of identical objective lenses in the object and the reference arm fixes the field of view and resolution and decided by the pre-chosen imaging objective lens. Further, the use of highest possible numerical aperture (NA) oil and water immersion objective lenses might be tricky due to the horizontal reference arm in the reported optical configurations. This would add difficulty in holding the immersion media in reference arm for a very long time and add instability to the system.

In transmission mode QPM, a narrowband light source (He–Ne) has been implemented previously20. In transmission geometry, the situation is different from the reflection geometry and requires precise positioning of the identical objective lenses from the sensor plane. Therefore, identical objective lenses are placed at equal distances from the sensor in both the object and the reference arm of the system. This will limit the application of the system and removes the system’s user friendliness as highlighted in the previous paragraph. In addition, in both reflection and transmission geometries, it is mentioned that the high fringe density interferograms can only be observed with the implementation of grating in the light beam path. This will generate fixed spatial frequency interferogram for a chosen combination of grating and lenses in QPM system. In addition, the grating pitch must be sufficiently high to avoid the loss of resolution due to insufficient sampling of the object arm speckle field. This overall restricts the widespread penetration of PTLS in the field of phase microscopy.

In the present work, first, a comparison of different light sources white light (WL), filtered white light (FWL), laser and PTLS in terms of an optical image quality is systematically done. It is observed that PTLS generates speckle free images of USAF chart unlike in case of laser. The quality of the optical images is found to be comparable to the WL/FWL sources. Further, the spatial phase sensitivity of the QPM employed with FWL, laser, speckle fields generated by stationary diffuser and PTLS is compared. The phase sensitivity in case of PTLS is found to be comparable to the FWL source, which provides the maximum phase sensitivity in any QPM system12,13. However, the short TC length of WL/FWL confines the high-density interference fringes (required for single shot QPM) in a small FOV of the detector. Contrary to this, PTLS can form high density interference fringe over the whole camera FOV with superior quality. The PTLS increases the SBP of the QPM system ~ 18 times compared to that for WL source. This makes the PTLS ideal for high-speed quantitative phase imaging with advantages of high spatial phase sensitivity, single shot, scalable field of view (FOV) and resolution and high space bandwidth product.

The present article enables the use of non-identical objective lenses in the object and the reference arm of the system provided the above conditions are satisfied. Thus, it provides flexibility to the user to achieve scalable FOV and resolution in the system. Moreover, user defined fringe density (low or high) can be easily obtained in PTLS-QPM system without using grating in the beam path and provide an ease to implement either single-shot or multi-shot QPM. The use and applicability of non-identical objective lenses in the object and the reference arm of the system is demonstrated experimentally by performing experiments on human sperm cells using water immersion objective lens (60 × /1.2NA) for single shot, high speed and highly spatial sensitive phase imaging of fast-moving sperm cells. In addition, large area FOV phase imaging is also demonstrated on placenta tissue sample using air objective lens (20 × /0.45NA) keeping 10 × /0.25NA objective lens in the reference arm.

Theory

Space bandwidth product

The SBP of an optical microscope is the product of FOV and the area of spatial frequency information collected by the microscope and defined as follows2:

$${\text{SBP}}_{{Microscope}} = {\text{FOV}} \times \pi \left(\text{NA}/\lambda\right)^{2}.$$
(1)

In QPM, the FOV of the reconstructed phase maps is limited by the interference FOV. Therefore, the SBP of an interference microscope can be calculated by employing the following modified expression:

$${\text{SBP}}_{QPM}=\text{Area of interference FOV} \times \uppi {\left(\text{NA}/\uplambda\right)}^{2},$$
(2)

where, \({{\uppi}}\left(\text{NA}/\lambda\right)^{2}\) is the area of the spatial frequency band, NA is the numerical aperture of the objective lens and λ is the illumination wavelength.

Temporal coherence

In the theory of the optical fields, temporal coherence defines the correlation between the light vibrations at two different moments of time. According to Wiener–Khinchin theorem, temporal coherence function also called autocorrelation ‘\(\Gamma \left(\Delta t\right)=\langle E\left(t\right){E}^{*}\left(t-\Delta t\right)\rangle\)’ between the light fields is the Fourier transform of source power spectral density and can be expressed as follows:

$$\Gamma \left(\Delta t\right)=\underset{-\infty }{\overset{\infty }{\int }}S\left(\nu \right)\,\text{exp}\left(i2\pi \nu \Delta t\right)\,d\nu ,$$
(3)

where, \(S\left(\nu \right)\) is the source spectral distribution function, \(\Gamma \left(\Delta t\right)\) is the temporal coherence function, and \(\Delta t\) is the time delay between the optical fields \(E\left(t\right)\) and \({E}^{*}\left(t-\Delta t\right)\).

The full width half maximum (FWHM) of temporal coherence function provides information about the temporal coherence length. Thus, larger spectral bandwidth light sources such as WL, FWL and LEDs etc. lead to smaller temporal coherence length or vice versa.

Spatial coherence

The spatial coherence governs the phase relationship between the light field vibrations reaching at two different points in space at the same instance of time. In transverse spatial coherence, the correlation function \(\Gamma \left(\overrightarrow{{r}_{1}},\overrightarrow{{r}_{2}},\Delta t=0\right)=\langle E\left(\overrightarrow{{r}_{1}},t\right){E}^{*}\left( \overrightarrow{{r}_{2}}, t\right)\rangle\) between the light fields originated from source \(S\left(u,v\right)\) at two different spatial points \({P}_{1}\left( \overrightarrow{{r}_{1}}, z\right)\) and \({P}_{2}\left( \overrightarrow{{r}_{2}}, z\right)\) located in \(\left(x,y\right)\) plane is considered. Analogous to Wiener- Khinchin theorem, the transverse spatial coherence function ‘\(\Gamma \left(\overrightarrow{{r}_{1}},\overrightarrow{{r}_{2}},\Delta t=0\right)\)’ and spatial angular frequency spectrum of the light source forms Fourier transform pairs. According to Van-Cittert-Zernike theorem, the correlation function can be defined as follows24,25,26:

$$\Gamma \left(\overrightarrow{{r}_{1}},\overrightarrow{{r}_{2}},\Delta t=0\right)=\underset{-\infty }{\overset{\infty }{\int }}S\left({k}_{x},{k}_{y}\right)\,\text{exp}\left[i\left({k}_{x}\delta x+{k}_{y}\delta y\right)\right]\,d{k}_{x}\,d{k}_{y},$$
(4)

where, \(S\left({k}_{x},{k}_{y}\right)\) is the source’s angular spatial frequency spectrum, \(\overrightarrow{{r}_{1}}\) and \(\overrightarrow{{r}_{2}}\) are the position vectors of points \({P}_{1}\) and \({P}_{2}\) located in the plane of observation at points \(\left(x,y\right)\) with respect to light source, \({k}_{x}\) and \({k}_{y}\) represents the spatial frequencies of the light field along \(x\) and \(y\) directions respectively.

For a plane wave propagating along the direction \(\overrightarrow{N}\left(\text{cos}\,{\theta }_{x}, \text{cos}\,{\theta }_{y}\right)\), the spatial frequency components \({k}_{x}\) and \({k}_{y}\) can be given by the following expressions24:

$${k}_{x}=\frac{2\pi }{{\Lambda }_{x}}=\frac{2\pi }{\lambda }\text{cos}\,{\theta }_{x}=\frac{\omega }{c}\text{cos}\,{\theta }_{x},$$
(5)
$${k}_{y}=\frac{2\pi }{{\Lambda }_{y}}=\frac{2\pi }{\lambda }\text{cos}\,{\theta }_{y}=\frac{\omega }{c}\text{cos}\,{\theta }_{y},$$
(6)

where, \({\Lambda }_{x}\) and \({\Lambda }_{y}\) are the spatial periods of the light field along \(x\) and \(y\) directions. \({\theta }_{x}\) and \({\theta }_{y}\) are the angles of the propagating field direction from \(x\) and \(y\) axes, respectively, and \(\lambda\) is the wavelength of light. The transverse spatial coherence lengths along \(x\) and \(y\) direction can be expressed as follows:

$${\rho }_{x}=\frac{2\pi }{\Delta {k}_{x}}, \quad {\rho }_{y}=\frac{2\pi }{\Delta {k}_{y}},$$
(7)

where, \(\Delta {k}_{x}\) and \(\Delta {k}_{y}\) are the transverse spatial frequency range along \(x\) and \(y\) axis.

The expressions for transverse coherence lengths of the optical field along \(x\) and \(y\) directions in terms of the angles can be defined as follows24:

$$\frac{1}{{\rho }_{x}}\approx \left(\frac{2}{\lambda }+\frac{1}{{l}_{c}}\right)\text{sin}\,{\theta }_{x}, \quad \frac{1}{{\rho }_{y}}\approx \left(\frac{2}{\lambda }+\frac{1}{{l}_{c}}\right)\text{sin}\,{\theta }_{y},$$
(8)

where, \({l}_{c}\) is the temporal coherence length of the light source and given by \(l_{c} \approx {\raise0.7ex\hbox{${\lambda ^{2} }$} \!\mathord{\left/ {\vphantom {{\lambda ^{2} } {\Delta \lambda }}}\right.\kern-\nulldelimiterspace} \!\lower0.7ex\hbox{${\Delta \lambda }$}}\). \(\Delta \lambda\) is the spectral bandwidth of the light source.

It can be seen from Eq. (8) that the transverse coherence of the source is governed by both the angular spatial frequency spectrum and temporal frequency spectrum of the optical field. The first and second terms in Eq. (8) represent the spatial and the temporal coherence part of the light field, respectively. For WL and FWL, second term dominates over the first term and confines interference fringes in a limited FOV due to low TC length. Whereas, PTLS has a narrow temporal spectral bandwidth depending on the laser’s temporal spectral bandwidth and wide angular spatial frequency spectrum depending on the source size. In case of PTLS, confinement of the interference fringes depends on the spatial term rather than on the temporal term. The extremely wide angular spatial frequency spectrum (i.e., extended) light source leads to the generation of short spatial coherence length or vice versa and also limits the confinement of the fringes in a limited FOV. Thus, adequate source size of PTLS has to be kept for the formation of fringes over wider FOV.

Results and discussion

Optical imaging of USAF chart

Most of the optical imaging systems frequently use different types of conventional light sources like halogen lamp, LEDs and lasers for the illumination of the sample. In bright field microscopy, halogen lamp or LED light sources being temporally low coherent are used to generate speckle free images of the specimens. The spatial coherence of such light sources is improved in the past by introducing a pinhole (50–100 µm) into the beam path at the expense of huge optical power loss from several W to mW. Thus, these light sources have very low photon degeneracy, i.e., average number of photons per unit coherence volume, compared to narrowband light sources like lasers27. High photon degeneracy of the lasers leads to formation of coherent and speckle noise in the images and significantly degrades their quality. Figure 2a,b illustrate the optical images of USAF resolution test target (Thorlabs: part #R3L3S1N) acquired using 10 × /0.25NA objective lens corresponding to WL and FWL of 10 nm spectral bandwidth. The experimental setup details are given in the materials and methods section (Fig. 8). A narrow band pass filter of peak wavelength 632.8 nm and spectral bandwidth of 10 nm (Thorlabs: part # FL632.8-10) is inserted into the WL beam path to generate FWL.

Figure 2
figure 2

Different light sources for microscopic imaging. Different light sources are compared in terms of the coherent noise in non-interferometric images of USAF resolution chart. Low temporal coherent light sources like white light (WL) and filtered white light (FWL) produce noise free optical images, whereas direct laser generates coherent noise in the images. On the contrary, pseudo-thermal light source generates images comparable to the WL/FWL keeping the advantages of monochromatic light source.

It can be clearly visualized that the low TC light sources generate speckle free images of the specimens. Next, a narrowband laser Cobolt Flamenco™ (λ = 660 nm; Δλ = 0.001 pm) is used to illuminate the same sample. The optical image of the resolution chart for laser illumination is depicted in Fig. 2c. The image clearly showed the unwanted parasitic fringe pattern noise which is not related to the test object. Thus, laser degraded the quality the images significantly. The optical imaging quality can be improved by employing relatively low TC and SC light source like diode laser (λ = 638 nm; Δλ = 0.7 nm) as exhibited in Fig. 2d. On the contrary, PTLS which has low SC and high TC properties, can generate the quality of the images comparable to WL or FWL light source. The microscopic image obtained by PTLS of the resolution chart is illustrated in Fig. 2e. Thus, the implementation of PTLS in bright field optical microscopy outperforms the conventional light sources in terms of photon degeneracy (compared to WL and FWL) and coherent noise reduction (compared to lasers).

Spatial phase sensitivity comparison

In the previous section, it is demonstrated that PTLS can reduce the coherent noise from the optical images significantly compared to the direct laser. In this section, the effect of the coherence properties of different light sources on the spatial phase sensitivity of QPM system is systematically studied. An optically flat silicon wafer is placed under the QPM system to record the interferometric images corresponding to FWL (bandwidth = 10 nm), direct laser, speckle field (stationary diffuser) and PTLS (rotating diffuser) as illustrated in Fig. 3a,d,g,j, respectively. It can be clearly seen that FWL and PTLS generate coherent noise free interferograms. The high TC length of the laser formed unwanted interference fringes due to the superposition of multiple reflections coming from different surfaces of the optical components degrading the image quality. Conversely, PTLS equipped QPM generate coherent noise free interferograms like in case of FWL or LEDs. It can be visualized from Fig. 3g, the interferogram is filled with the speckle noise when the diffuser is kept stationary. The size of each interferogram is 512 × 512 pixels.

Figure 3
figure 3

PTLS prevents speckle noise formation. Four differently generated light sources were used: filtered white light, narrow band laser, speckle field and pseudo-thermal light source. A bandpass filter of 10 nm bandwidth at central wavelength 632 nm is inserted into the white light beam path to generate FWL. Speckle field is generated by passing a narrow band laser beam through a stationary diffuser. PTLS is synthesized by passing a highly temporal coherent beam through a rotating diffuser followed by a multimode fiber. (a,d,g,j) Interferometric images of a standard Si wafer. (b,e,h,k) Corresponding reconstructed phase images while employing FWL, laser, speckle field and PTLS, respectively. (c,f,i,l) Histograms of the recovered phase images of Si wafer to illustrate the spatial phase noise while using aforementioned light sources. The PTLS effectively suppressed spatial phase noise, behaving similarly to the FWL but very differently from conventional lasers. The color bars are in mrad.

The interferograms depicted in Fig. 3a,d,g,j are further post processed to recover corresponding phase maps using Fourier transform fringe analysis method28. The corresponding recovered phase images are illustrated in Fig. 3b,e,h,k, respectively. Further, the effect of the coherent noise on the recovered phase images is calculated in terms of a quantitative matric called spatial phase sensitivity12. In order to measure the spatial phase sensitivity of the QPM system for FWL, laser, speckle field and PTLS, the standard deviation of the recovered phase maps is quantified. The standard deviation of the recovered phase maps is given in Table 1. The histogram plots corresponding to FWL, laser, speckle field and PTLS are presented in Fig. 3c,f,i,l, respectively. It can be visualized that the spatial phase sensitivity of PTLS is comparable to the FWL/LEDs and much higher than that for laser. Thus, PTLS is capable to provide highly spatial phase sensitive images of the specimens while keeping a high TC length comparable to laser and subsequently enables the generation of high-density fringes over large FOV. PTLS provides ~ 17 times more spatial phase sensitivity than that for laser. We also performed systematic study to understand the effect of rotating diffuser speed on the spatial phase sensitivity of the QPM system and given in the Supplementary Information.

Table 1 Spatial phase sensitivity comparison of different light sources: Standard deviation calculated from the recovered phase maps corresponding to FWL, laser, speckle field and PTLS.

Demonstration of high spatial phase sensitivity: phase imaging of waveguide of 8 nm rib height

Next, the experiment is conducted on a rib optical waveguide having core material of silicon nitride (Si3N4), refractive index n ~ 2.04 and with a rib height of only ~ 8 nm. The fabrication procedure of the waveguide can be found elsewhere29. Here, a diode laser (λ = 638 nm; Δλ = 0.7 nm) is used instead of extremely high TC length Cobolt Flamenco laser. The diode laser is preferred over Cobalt laser as it generates less coherent noise due to short TC length. This makes a fair comparison between PTLS and diode laser in terms of the phase measurement sensitivity enhancement.

The rib waveguide is placed under the QPM to acquire high fringe density interferograms corresponding to direct laser and PTLS both as depicted in Fig. 4a,e, respectively. It is depicted from Fig. 4a, direct laser’s interferogram suffers from the coherent noise generated due to high TC length of the laser. In contrast, PTLS produces coherent and speckle noise free interferogram as illustrated in Fig. 4e. The insets of Fig. 4a,e clearly depict the fringe quality difference in the interferometric images generated from laser and PTLS. The recorded interferograms are further post-processed for the phase recovery of rib waveguide. Figure 4b,f illustrate the full FOV reconstructed phase maps of the rib waveguide corresponding to laser and PTLS, respectively. Figure 4c,g illustrate the zoomed views of the region marked with yellow dotted box in Fig. 4b,f. Subsequently, the recovered phase maps can also be used to measure the height maps of rib waveguide as the refractive index of the material is known from ellipsometry experiments30.

Figure 4
figure 4

PTLS produces high quality phase images of optical waveguide of 8 nm rib height. (a,e) Interferometric images of Si3N4 optical waveguide while using diode laser (at 638 nm) and PTLS, respectively. The insets depict the zoomed view of the region marked with yellow dotted box. (b,f) Reconstructed full FOV phase maps of an optical waveguide (H ~ 8 nm) corresponding to diode laser and PTLS, respectively. (c,g) Zoomed views of the regions marked with yellow dotted boxes in the reconstructed phase images depicted in (b) and (f), respectively. (d,h) The histogram plots of the phase images shown in (c) and (g), respectively. It can be visualized that the PTLS produces superior quality phase images of the waveguide compared to diode laser. The thin rib height of ~ 8 nm is not visible and embedded in the coherent noise of the QPM system in case of diode laser. The color bars are in rad.

The histogram plots of the recovered phase maps are depicted in Fig. 4d,h, respectively. It can be observed from the histogram plots that laser generates large coherent noise in the recovered phase maps which is significantly very less in case of PTLS. In PTLS histogram plot, the peaks are well resolved, whereas the peaks are not distinct in case of laser. The phase information of rib waveguide is completely embedded in the phase noise for laser. The difference between the peaks shown in Fig. 4h is equal to the difference between the waveguide’s foreground and background phase values. The bandwidth of the peaks, i.e., standard deviation exhibits the phase variation over the waveguide surface. The FWHM of the peak is measured to be equal to 0.027 rad. The little large value of standard deviation could be due to following possible reasons: optical thickness variation of the waveguide, phase noise and optical aberration present in the QPM system. It is quite evident from Fig. 4d,h that the coherence property of the light source plays an important role in the phase measurement sensitivity of the QPM system. Thus, PTLS provides approximately 10-fold increase in the spatial phase sensitivity compared to diode laser.

Demonstration of high space bandwidth product: interferometric imaging of USAF chart

Here, we demonstrate the advantages of PTLS over the WL/FWL source in terms of enhancing the SBP. For this, experiments are conducted on USAF resolution test target by sequentially employing WL, FWL, laser and PTLS in QPM system. The interferograms of USAF chart corresponding to WL, FWL, laser and PTLS are sequentially recorded using developed QPM system as depicted in Fig. 5a,d,g,j, respectively. Figure 5b,e,h,k represents the zoomed views of the regions marked with green color boxes depicted in Fig. 5a,d,g,j, respectively. The insets of Fig. 5b,e,h,k represent the line profiles of the interferograms along red dotted lines corresponding to WL and PTLS. The line profiles of the whole camera FOV corresponding to WL, FWL, laser and PTLS along blue dotted lines are illustrated in Fig. 5c,f,i,l, respectively.

Figure 5
figure 5

PTLS generates high space bandwidth product interferometric images without speckle noise. Four light sources with different degrees of temporal and spatial coherence were used: a white light (WL), filtered white light (FWL), narrow band laser and pseudo-thermal light source (PLTS). (a,d,g,j) Interferometric images of USAF resolution chart while using WL, FWL, Laser and PTLS, respectively. (b,e,h,k) Zoomed view of the region marked with green color box in (a), (d), (g) and (j), respectively. The insets depict the line profiles along red dotted horizontal lines. (c,f,i,l) The line profiles over the whole camera FOV along blue dotted lines shown in (a), (d), (g) and (j), respectively. PTLS produces coherent noise free interference fringes over the whole camera FOV unlike WL/FWL and narrowband laser.

It is evident from Fig. 5c that the interference fringes are confined only in a limited region of the camera FOV for WL source. This is due to the low TC length of WL source. The area of the interference FOV can be increased by inserting a narrow spectral band-pass filter (Δλ = 10 nm) into the beam path at the expense of large intensity loss. It can be visualized that narrow band-pass filter of 10 nm spectral bandwidth is not sufficient to produce interference fringe over whole camera sensor. On the contrary, the interference fringes of the same quality as that for WL/FWL are observed over the whole camera FOV while using PTLS as illustrated in Fig. 5k. The area of the interference FOV for WL is found to be 18 times less than that for PTLS. The normalized interference fringe contrast value below 0.1 is used to select the boundary of the interference FOV. Thus, the implementation of the PTLS increases the SBP of QPM system by 18-fold compared to WL, without sacrificing the temporal resolution. It can be seen that direct laser can obviously forms the high-density fringes over whole camera FOV at the cost of large coherent noise (Fig. 5i). In addition, the fringe visibility is not constant over whole camera FOV. This could be due to the gaussian intensity profile of the light beam coming out from a single mode fiber. On the contrary, the line profile exhibited in Fig. 5l clearly demonstrates that the fringe visibility does not vary over whole camera FOV in case of PTLS.

Demonstration of high-speed QPM with high spatial phase sensitive: QPM of fast-moving living human sperm cells

Next, the experiment is conducted on the live human sperm cells using water immersion (60 × /1.2NA) objective lens. The reference objective is not changed and kept 10 × /0.25NA. The explanation of obtaining interference fringes in such unbalanced optical configuration is provided in the Supplementary Information. Live human sperm cells are always challenging to image in their natural environment because of the following reasons: the mean swimming speed of sperm cells is 20–160 µm/s31,32,33 and the tail of the sperm is very thin and generates minute optical thickness34. According to the World Health Organization (WHO) criteria, tail of the healthy sperm cell should be thinner than the midpiece and almost of uniform thickness over its length of about 45 µm35. Thus, to image healthy sperm cells, the QPM system should be capable to acquire highly spatial sensitive interferometric/phase images at very high acquisition speed more than 120 fps. As mentioned in the previous sections, the implementation of PTLS in QPM system is capable to provide highly spatial sensitive phase images at very high speed.

To prepare the sperm sample for phase imaging, first, a 170 µm thick polydimethylsiloxane (PDMS) chamber is placed on top a reflecting silicon wafer. The diluted sperm sample’s volume of 10–15 µl is placed in a ~ 4 mm × 4 mm opening area of the PDMS. The sample is then covered from the top using a cover glass (thickness: 170 µm) to avoid any air current in the sample. The sample is placed under the QPM system to perform high speed and high-resolution phase imaging of live human sperm cells. The interferometric images of the sperm cells are acquired at 30 fps, which is the maximum speed of our camera at full resolution. This can be obviously improved by using a high speed camera such as 200–1000 fps, supported by camera pco.dimax HS4, Phantom T134036.

The interferometric image of the live sperm is depicted in Fig. 6a. The inset represented the zoomed view of the region marked with red dotted box. The live interferometric movie of the live sperm cells can be seen through Visualization 1. The reconstructed phase map of one interferometric frame of the movie is illustrated in Fig. 6b. The phase movie of the live sperm cells can be found in Visualization 2. The zoomed view of the region marked with red dotted box in Fig. 6b is shown in Fig. 6c. Due to high sensitivity, we can also visualize the phase map of thin tail region of the sperm cells which is around 100 nm34. Thus, PTLS enables both high-speed imaging with high spatial sensitive quantitative phase imaging. Thanks to the high-density fringes formed over large FOV, single-shot PTLS based QPM systems can be deliver quantitative phase maps of thin biological samples moving at fast speeds. We believe PTLS equipped QPM can be implemented to visualize and quantify abnormalities in live sperm cells like defects of head neck attachment, dysplasia of fbrosis sheath (DFS) and primary ciliary dyskinesia (PCD) in future.

Figure 6
figure 6

PTLS provides high resolution video rate phase images of fast moving transparent biological cells. A water immersion objective lens with 60 × /1.2 NA is used to image fast moving human sperm cells. (a) Interferometric image of live human sperm cells (see Visualization 1). The inset depicts the zoomed view of the region marked with red dotted box. (b) Reconstructed high-resolution phase image of the sperm cells (Visualization 2). (c) The zoomed view of the phase image of sperm cells enclosed by red dotted box shown in (b). The scale bars are in micron. The color bar is in rad.

Demonstration of large FOV together with high spatial phase sensitive: QPM of placenta tissue sample

To demonstrate the large FOV phase imaging capability of PTLS equipped QPM system, the experiments are conducted on human placenta tissue samples of 1 μm thickness. The images are acquired with 20 × /0.45 NA objective lens which provides FOV approximately equal to 660 × 660 μm2. The sequentially acquired interferometric images of the whole tissue sample are illustrated in Fig. 7a,b. The sample stage has motorized translation along X and Y axis to acquire the interferometric images of side-by-side FOV. The overlap between the images is kept approximately equal to 20%. It is demonstrated that with just two interferometric images acquired in 500 ms are sufficient to cover the whole tissue samples as shown in Fig. 7. The interferograms are further utilized to recover the phase maps of the tissue sample. Figure 7c illustrates the stitched phase image of the whole tissue sample. The total FOV of the stitched image is 660 × 1200 μm2. The FOV is just limited by the camera chip size and the magnification of the objective lens employed, while the temporal resolution (imaging speed) is limited by the camera sensor speed.

Figure 7
figure 7

Large FOV phase imaging. (a,b) Interferometric images of the placenta tissue sample. The inset depicts the zoomed view of a small region of the interferogram marked with yellow dotted box. (c) Stitched phase image of the whole tissue sample of 1 μm thickness. The color bar is in rad.

Conclusions

PTLS offers many folds increase in SBP compared to WL source without sacrificing the spatial phase sensitivity and temporal resolution. In addition, PTLS enables the use of single-shot phase recovery algorithms without introducing coherent noise in the recovered phase images which is otherwise a common problem while working with lasers. The comparison between different light sources, in terms of coherent noise, spatial phase sensitivity and SBP, is demonstrated on USAF chart and on an optical waveguide with rib height of 8 nm. It is observed that PTLS outperforms the use of conventional light sources like WL, FWL, LEDs and Lasers in QPM system. Being, a single shot technique video rate quantitative phase imaging is successfully demonstrated on live human sperm cells, with speed only limited to the camera speed. This became possible due to the presence of high-density fringes over the whole camera FOV, which enabled single shot phase recovery of the fast-moving sperm cells. Thus, the FOV obtained by PTLS based QPM is just limited by the camera chip size and the magnification of the objective lens employed, while the temporal resolution (imaging speed) is limited by the camera sensor speed.

Evaluating sperm cell quality with PTLS based QPM system has a potential to be applied in IVF clinic. The concept of large FOV phase imaging capability of PTLS based QPM system is also demonstrated on placenta tissue samples. In pathology, generally a large area of sample sections has to be scanned to provide a holistic understanding about diseases and their progression. We believe that the capabilities of PTLS paves the way to achieve its wider penetration in the field of quantitative phase imaging in life sciences application and would possibly also find applications in metrology. The present technique may find applications in the high speed QPM and can be integrated with other functionalities like optical waveguide trapping3,37, microfluidics38, optical tweezers39 and multi-modal optical imaging techniques40.

Materials and methods

Experimental setup

The experimental scheme of QPM system is illustrated in Fig. 8, which is based on the principle of non-common-path Linnik based interference microscopy. The light beam coming from a laser was split into two beams using beam splitter BS1. One of the beams directed towards microscope objective MO1 which coupled the light beam into a single mode fiber (SMF) which spatially filtered and generated a temporally and spatially high coherent light beam. The other one directed towards microscope objective MO2 which illuminated the rotating diffuser (RD) with a diverging beam of spot size at diffuser plane ~ 1–2 mm. The grit size and diameter of the circular diffuser (Edmund optics, stock # 62-620) are 220 and 100 mm. The laser beam was hit at a distance of ~ 40 mm from the center of the diffuser. The speed of the diffuser is kept 240 RPM to provide sufficient averaging of the speckle noise within the exposure time (30 ms) of the camera. The RD generated temporally varying speckle field and leads to the reduction in speckle contrast significantly41. The scattered photons at the output of RD are coupled into a multi-mode fiber (MMF) with the help of beam splitter BS2 and lens L1. The core diameter of MMF is 1 mm. The loss of power of the input light beam after every component is given in the Supplementary Information. The RD followed by MMF generated uniform illumination, i.e., speckle free field, at the output port of MMF, which acts as an extended purely monochromatic light source namely PTLS. Thus, generated a temporally high and spatially low coherent light source corresponding to narrow-band laser. The MMF also carried the light coming from a halogen lamp as illustrated in Fig. 8. The output of MMF was both temporally and spatially incoherent for WL. To generate a filtered white light (FWL), a narrow bandpass filter of spectral bandwidth ~ 10 nm is inserted into the WL beam path after MMF output.

Figure 8
figure 8

Schematic diagram of the QPM system. MO1–4 Microscope objectives, BS1–4 beam splitters, L1–4 Lenses, RD rotating diffuser, MMF multi-mode fiber, SMF single mode fiber, A bandpass filter, M mirror, CMOS complementary metal oxide semiconductor camera.

The MMF carried light coming from all types of light sources to the interference microscopy system as illustrated in Fig. 8. The output facet of MMF acted as a light source for phase microscopy system and its image is relayed to the back focal plane of the microscope objective MO3 (10 × /0.25NA) using lenses L2 and L3. Thus, it generated a uniform illumination at the test specimen placed under the microscope. The input light beam is split into the object and the reference beam using beam splitter BS4. The object beam is travelled towards the sample and passed through it twice before reaching the camera due to the reflection geometry of the present microscope. The reference beam is reflected from the reference mirror (M) and recombined with the object beam at the same beam splitter BS4. The recombined beams were finally overlapped at the camera plane using lens L4 to generate interference pattern which carried information about the test specimens. The reference mirror is attached to a kinematic mount to control the angle between the object and the reference beam and subsequently the fringe density of the observed interference fringes at Complementary Metal Oxide Semiconductor (CMOS) camera. The maximum angle between both the beams is kept in a way such that the Nyquist sampling criteria is satisfied. WL, FWL, Laser, and PTLS are sequentially implemented to acquire corresponding interferometric data for different specimens.

Sample preparation

Sperm cells

The semen samples were collected according to the guidelines of WHO from men who attended the IVF clinic for the investigation/volunteers. The high motility sperm fraction was isolated using gradient centrifugation method. Complete details of the whole sample preparation can be found elsewhere23. Later, the sperm sample was diluted to a concentration of 5.0 × 106 cells/ml in culture medium. For QPM measurements the sperm cells were placed in a PDMS chamber on a reflective Si substrate and covered with standard #1.5 thickness coverslip. Informed consent was obtained from all participants and the ethical clearance was approved by The Regional Committee for Medical and Health Research Ethics of Norway (REK_nord) and all methods were performed in accordance with the relevant guidelines and regulations.

Human placenta tissue sections

The placenta tissue sample was collected immediately after delivery, sectioned and rinses in physiological saline to remove any contamination. Placental samples with size 1 mm3 were immersed in 5 ml 8% formaldehyde in PHEM buffer and incubated at 4 °C overnight. Tissue samples were immersed in 0.12% glycine at 37 °C (1 h), infiltrated with 2.3 M sucrose at 4 °C overnight to prevent freeze damage/structure deformation and transferred to liquid Nitrogen tank for storage. The samples were cut into 1 μm thick cryosections (EMUC6 ultramicrotome, Leica Microsystems, Vienna, Austria), and placed on Si substrate. Thereafter, samples were washed and mounted with PBS before imaging.