1 Introduction

Elevated measures of arterial stiffness (central and peripheral) have been shown to negatively impact LV systolic and diastolic function, impair coronary blood flow, and have been identified as a risk factor for cardiovascular disease and mortality [1,2,3,4,5,6,7,8,9]. One possible mechanism for these findings is that increased arterial stiffness can create additional LV work while simultaneously making coronary perfusion less efficient. The hydraulic load (IE: work) encountered during LV ejection is directly increased when arterial stiffening (less compliance) results in higher aortic input impedance. The size and timing of reflected waves will also change in the setting of a stiff arterial tree, often with a larger positive reflected wave returning to the proximal aorta in systole instead of diastole. This alteration of reflected wave properties can indirectly add to the increased LV hydraulic load, while also causing diminished coronary perfusion.

Endovascular aneurysm repair (EVAR) is well-established and frequently used treatment modality for aortic pathology, including abdominal aortic aneurysms (AAA). Accordingly, much research has been generated on the local effects of the graft with respect to the mechanisms of endoleak, graft migration and aneurysm sac remodeling. There has also been interest in the arterial physiologic effects of placing artificial graft material into the aorta, as well as the potential implications on cardiac function. A dog model demonstrated that proximal thoracic aortic stenting results in increased overall aortic input impedance, ascending aorta characteristic impedance (Z0) and pulse-wave velocity (PWV). These alterations occurred in conjunction with decreased total arterial compliance, alterations in reflected wave properties and reduced LV output (measured as lower aortic flow) [10]. An increased PWV has subsequently been demonstrated in human patients following implantation of an EVAR endograft [11,12,13,14,15,16]. Although this suggests an endograft induced increase in aortic stiffness, a comprehensive analysis of the biomechanical impact of endograft implantation has not been possible due to the invasive measures that would be required.

Recent improvements in medical imaging, combined with the rapid growth in computing power, has resulted in an emergence of non- or minimally-invasive methods for performing more nuanced determinations of aortic mechanical properties. Ultrasound images of the aorta can be acquired with intraoperative transesophageal echocardiography (TEE). These images can then be analyzed with speckle tracking (STE), a validated method which is able to follow tissue motion and quantify position, strain, and velocity [17, 18]. Though mostly used for quantifying ventricular function, this technology has been used to quantify aortic properties- including correlation with histologic and physical biomechanical testing of subsequently excised aortic tissue [19,20,21].

We report a technique using STE to assist in non-invasively determining the immediate effects of abdominal endograft deployment on aortic input impedance, ascending aortic compliance, systemic vascular resistance, reflected wave properties and LV function.

2 Materials and Methods

This study received approval from our institutional ethic committee. Participants were selected from adult (age > 18 years) patients scheduled to have elective EVAR for AAA. Exclusion criteria included patient refusal, contraindication to trans-esophageal echocardiography, prior aortic surgery, aortic valve abnormality (stenosis, regurgitation or prosthesis), atrial fibrillation and severe aortic calcification (> 2/3 circumference). After being identified and enrolled in the study demographic, medical information and current medications were collected. All procedures were performed under general anesthesia. Pre-implantation data was acquired once the patient was in a plane of anesthesia with stable hemodynamics. Post-implantation data was acquired in the operating theatre, shortly after endograft deployment when the patient had stable hemodynamics.

2.1 Data-Collection and Analysis

2.1.1 Echocardiography

Trans-esophageal echocardiogram (TEE) was performed in each patient using a GE Vivid 7 machine (General Electric, Fairfield, Connecticut, USA). The following views were obtained (Fig. 1):

Fig. 1
figure 1

Images demonstrating the standard views and Doppler quantification acquired from each patient before and after EVAR procedure. a Mid-esophageal 4-chamber view. b Mid-esophageal aortic valve long axis view. c Trans-mitral pulse wave Doppler flow. d Tissue Doppler imaging of the basal septal and basal lateral LV walls. e Deep transgastric aortic valve/LVOT view. f LVOT pulse wave Doppler flow

  1. 1.

    Mid-esophageal 4-chamber view.

  2. 2.

    Tissue-Doppler imaging (TDI) and pulsed-wave Doppler (PWD) of the lateral and septal mitral valve annuli.

  3. 3.

    PWD of the trans-mitral flow.

  4. 4.

    Mid-esophageal long axis view of the aortic valve, including the aortic root and proximal ascending aorta

  5. 5.

    Deep-transgastric view of the aortic valve with PWD of the left ventricular outflow tract (LVOT), just proximal to the aortic valve.

2.1.2 Speckle-Tracking Echocardiography

Analysis of the TEE images were performed post-acquisition offline using a PC workstation and EchoPac, Version 7.2.0 software (General Electric, Fairfield, Connecticut, USA). The value for peak mitral annulus TDI velocity in early diastole was obtained for both the septal and lateral annuli (E’sept and E’lat) at the same times peak trans-mitral flow velocity in early diastole (E) using PWD.

Measurements of proximal aortic dimensions were taken from the standard TEE mid-esophageal aortic long-axis view using the caliper function (Fig. 2). The aorta was divided into 2 segments. The first segment was the aortic root while the second segment was the proximal ascending aorta. The anterior and posterior lengths of the aortic root were measured as 2 straight lines spanning, but not including, the walls of the Sinus of Valsava from the aortic valve to the sinotubular junction. The second segment was measured as 2 straight lines along the anterior and posterior aortic walls, from the sinotubular junction extending 2 cm into the ascending aorta. The lengths of the second segment were shortened in the event of poor image quality or if the boundary of the image prevented a full 2 cm segment. The anterior and posterior segment lengths were averaged to obtain a single length for our calculation of aortic volume. The diameters of each segment were taken inner edge to inner edge, at the midpoint of the segment. The diameter of the LVOT was also measured from this view. All measurements were taken at end-systole.

Fig. 2
figure 2

Caliper measurements of the ascending aorta used to calculate compliance. Taken from the mid-esophageal long axis view of the aorta at end-systole. A1 posterior length of segment 1 (aortic root), A2 anterior length of segment 1, B1 posterior length of segment 2 (ascending aorta), B2 anterior length of segment 2, C segment 1 diameter, D segment 2 diameter. See Methods section for details

Speckle tracking imaging (STI) analysis was performed using the same EchoPac software. In the mid- esophageal four-chamber view the endocardial border of the left ventricle (LV) was traced. Appropriate tissue tracking was confirmed with the program’s own quality control feature, as well as visually by the investigators by observing a full cine loop to ensure the ventricular walls were tracked throughout the cardiac cycle. The STI-derived peak values for longitudinal tissue velocity, longitudinal strain rate and radial strain rate of the basal-septal and basal-lateral walls were measured in early diastole (Fig. 3).

Fig. 3
figure 3

Speckle tracking imaging of the LV. Peak longitudinal velocity, peak longitudinal strain rate (Fig. 3a) and peak radial strain rate (Fig. 3b) were taken in early diastole (E’). The basal septal wall is depicted in yellow, the basal lateral wall is depicted in red and the peak values are identified with a white arrow

STI analysis was further applied to the ascending aorta using the EchoPac software’s pre-set for two- chamber LV images using a method we have previously described [4]. This method includes adjusting the software’s pre-set spatial and timing filters to the lowest level. The walls of the aortic root and ascending aorta were traced from the aortic valve annulus to the most distal portion of the ascending aorta that was visible on the acquired images. Proper aortic tissue tracking was confirmed in the same fashion as mentioned above. The region of interest (ROI) was made as narrow as possible to best conform to the width of the aorta. The ROI borders were further adjusted so that the software’s pre-set labels for identifying LV segments matched the aortic segments described above. The ROI for the LV basal segments spanned the aortic root while the LV mid segments spanned the proximal 2 cm of ascending aorta (Fig. 4). The LV apical segments did not track any aortic tissue and were not included in our analysis. Radial velocity and longitudinal strain rates were taken for the anterior and posterior walls of the aortic root and ascending aorta (Fig. 5). The graphs produced by the EchoPac software were converted into digital form using Digitize-It (version 1.5.8b, Bormisoft, Braunschweig, Germany) and saved as a text file.

Fig. 4
figure 4

The aortic root and proximal ascending aorta are traced using speckle tracking imaging software. The region of interest was adjusted so that the width incorporated the aortic wall. The segments were adjusted to include the posterior root (yellow), anterior root (red), posterior ascending aorta (cyan) and proximal ascending aorta (blue)

Fig. 5
figure 5

Speckle tracking imaging analysis of the aortic root and proximal ascending aorta. Region of interest was adjusted as described in the text and Fig. 4. Radial velocity and longitudinal strain rates were taken throughout the cardiac cycle in order to measure aortic volume change

2.1.3 Arterial Pressure Acquisition and Aortic Pressure Derivation

A right radial artery catheter placed prior to the induction of anesthesia. The catheter was connected to the transducer via a short length of pressure tubing (12 cm) to optimize the resonant frequency of the system. The filter of the arterial pressure channel on the Datex/Ohmeda S5 monitor was set to a frequency of 44 Hz and the analogue waveforms (radial artery pressure wave and the ECG) were acquired simultaneously through the data-out port at 1000 Hz using an analogue–digital converter (National Instruments, Austin, Texas, USA) and LabVIEW (National Instruments, Austin, Texas, USA). A pop-test was performed prior to every recording in order to calculate the system’s resonant frequency and dampening coefficient. Ten to twelve pressure waves were averaged from the peak of the QRS complex of the ECG. The arterial pressure wave was then calibrated using the systolic and diastolic pressure recorded at the time of the acquisition. The pressure wave was then corrected for the resonance characteristics of the system and converted to an aortic pressure waveform using a previously validated transfer function [22].

2.1.4 Calculation of Aortic Parameters

Off-line analysis was performed using Maple (version 18.02, MapleSoft, Waterloo, Ontario, Canada). The aortic pressure wave was used to calculate the rate of pressure change (dP/dt) during the cardiac cycle. The STI based graphs of aortic radial velocity and longitudinal strain rate were used to calculate aortic volume and its rate of change during the cardiac cycle (dV/dt). All signals were filtered using a Butterworth, 4-element filter with a cut-off frequency of 16 Hz. The minor differences in the duration of the dP/dt and dV/dt waves were corrected through interpolation of the dV/dt waveforms. The ECG- gating from the TEE machine and anesthesia monitor were not identical resulting in a timing shift of the graphs of dP/dt and dV/dt. Therefore, we aligned the two graphs at the point where both dP/dt and dV/dt were initially positive. The interactions of pressure and volume change were used to generate pressure–volume loops for each patient, both before and after stent implantation (Fig. 6). We then produced graphs displaying total proximal aortic compliance (combined aortic root + ascending aorta) over the entire cardiac cycle. To calculate proximal aortic systolic compliance, the start of systole was initially identified by considering the time point in the cardiac cycle where dP/dt first becomes positive (Fig. 7a). The value of this time point was then used on the generated compliance versus time curve to identify the approximate beginning of systole. An asymptote near the start of systole and a flat nadir at the end of systole were consistent findings in all the study subjects (Fig. 7b). There was high variability in both the data values near the asymptote and the length of the nadir. Therefore, to obtain consistent measurements between patients, the first post-asymptote peak on the compliance curve in systole was chosen as the starting point and the beginning of the nadir as the end-point for measuring early systolic aortic compliance. Thus, proximal aortic systolic compliance (CAortaSyst) was defined as the average compliance over this period (Fig. 7b).

Fig. 6
figure 6

By combining STI-based aortic volume change with invasively measured arterial pressure pressure–volume loops were generated for each patient before and after deployment of the EVAR endograft. Arrows depict the direction of the loops during the cardiac cycle

Fig. 7
figure 7

The determination of proximal aortic systolic compliance was completed in a 2-step process. The start of systole was determined by identifying the time point where dP/dt first became positive (Fig. 7a). This time value was then used on the compliance curve to locate the start of systole (Fig. 7b). A consistent finding was an asymptote in this area. The mean compliance in the period from the first post-asymptote peak until the beginning of the nadir was used to calculate proximal aortic systolic compliance (CAortaSyst) in our analysis. This period is represented by the shaded box

The LVOT flow acquired from the intra-operative TEE using PWD was digitized (DigitizeIt version 1.5.8b, Bormisoft, Braunschweig, Germany) at 2000 Hz, resampled at 1000 Hz and filtered. Stroke volume (SV) and cardiac index (CI) were calculated:

$${\text{SV}} = \pi \times {\text{LVOT}}\;{\text{radius}}^{2} \times {\text{LVOT}}\;{\text{velocity}}\;{\text{time}}\;{\text{integral}}$$
(1)
$${\text{CI}} = {\text{SV}} \times {\text{HR/BSA}}$$
(2)

By combining LVOT flow and central aortic pressure, aortic input impedance was plotted across the frequency spectrum up to 15 Hz. The zero modulus was taken as systemic vascular resistance (SVR). The aortic characteristic impedance (Z0) was calculated from the average impedance in the 5–15 Hz frequencies, excluding any values that were more than 2 standard deviations above the average. The left ventricular pulsatile power, including mean and oscillatory power, was calculated from the pressure and flow waves using methodology previously described [23]. Additional use of the aortic pressure included wave form analysis allowing identification of the forward (incident) and backward (reflected) components (Fig. 8). This was done according to previously accepted methods [24,25,26,27]. We calculated the total forward, total backward, systolic forward and systolic backward waves. The reflection coefficient (PB/PF) was determined by the total reflected wave divided by the total forward wave.

Fig. 8
figure 8

Example from one of our study subjects of our wave form analysis using the aortic pressure wave (solid; red) allowing identification of the forward (hashed; blue) and backward (hashed-dot; pink) components. A difference can be seen in the size and timing of the reflected wave in a pre-EVAR compared to b post-EVAR

2.2 Statistical Analysis

A sample size was calculated that would allow for comparison of post-EVAR changes in both LV diastolic function and aortic characteristic impedance. We used data from a previous study which measured characteristic impedance in both normotensive and hypertensive middle-aged patients, similar to our patient population. Their results showed a mean Z0 of 166 dynes s cm−5 with a standard deviation of 69 dynes s cm−5 [28]. Eighteen patients would allow for a detection of a post-implantation increase in impedance of 40% with 80% power and a p value < 0.05 [24]. A 40% relative change in impedance was felt to be clinically significant based on a similar change described by Dobson et al. [10]. This assertion is supported by studies showing that a 20–40% difference in valvuloarterial impedance impacts the clinical outcomes in patients with aortic valve stenosis [29].

With regards to diastolic function, STI values in middle-aged patients with mild or moderate diastolic dysfunction, the average longitudinal early diastolic strain rate was 1.15%/s with a standard deviation of 0.26%/s [30]. Twenty-one patients would allow for detecting a 20% decrease in diastolic function following EVAR with a power of 80% and a p value < 0.05 [31]. This change is likely clinically relevant based on the findings of Tsioufis et al., where a 20% difference in diastolic function was observed between normotensive and hypertensive patients [32].

All statistical analysis was performed using Analyze-It software version 3.90.7 (Analyze-It Software Ltd, Leeds, UK). Descriptive statistics were used for demographic and intra-operative data. The values for LV diastolic function and all measures of aortic function were analyzed for changes following endograft implantation using a Student’s T test.

3 Results

Twenty-one patients were included in the study. Three of the patients had incomplete data sets due to equipment failure or inadequate TEE images, therefore a total of 18 patients’ data was analyzed. The demographic information and medical information is summarized in Table 1. Intraoperative hemodynamic details and endograft information is listed in Table 2. There was a near-significant trend towards higher systolic blood pressure after the EVAR procedure increasing from a mean (SD) of 109.5 (17.1) mmHg to 119.0 (20.3) mmHg (p = 0.06). There was no significant difference after EVAR in diastolic blood pressure (p = 0.59), mean arterial pressure (p = 0.35), heart rate (p = 0.19), and a trend towards reduced cardiac index (p = 0.11).

Table 1 Patient demographics, comorbidities and pre-operative medications
Table 2 Intra-operative clinical and procedural data

Pre-operatively the peak systolic mean (SD) STE measures of aortic longitudinal strain rates were 1.36 (0.80) %/s for the anterior segments and 1.57 (0.73) %/s for the posterior segments while the STE aortic radial velocities were 3.30 (1.12) cm/s for the anterior segments and 3.02 (0.95) cm/s for the posterior segments. Post-operatively peak systolic aortic longitudinal strain rates were 1.69 (1.10) %/s for the anterior segments and 1.78 (1.00) %/s for the posterior segments while the aortic radial velocities were 3.52 (0.99) cm/s for the anterior segments and 3.22 (1.18) cm/s for the posterior segments.

Full results of the comparative analysis of aortic and left ventricular diastolic function post-EVAR are included in Tables 3 and 4. Following deployment of the endovascular stent graft proximal aortic systolic compliance (p = 0.60) and calculated SVR (p = 0.21) were not significantly altered by EVAR endograft deployment while the aortic characteristic impedance (Z0) had a borderline significant post-EVAR decrease from a mean (SD) of 133.8 (63.6) dynes s/cm5 to 102.3 (77.5) dynes s/cm5 (p = 0.06). The magnitude of forward and reflected waves did not change significantly, however the reflection coefficient increased from a mean (SD) of 0.85 (0.08) to 0.88 (0.07) (p = 0.03). This was due to a change in the timing of positive reflected waves, with the systolic component increasing significantly (p = 0.002). The left ventricular pulsatile power (hydraulic load) increased from a mean (SD) of 160.1 (108.3) mW to 206.3 (93.8) mW (p = 0.05). There was a non-significant trend towards increased LV diastolic impairment post-EVAR, in both TDI and STI measured parameters (Table 4).

Table 3 Aortic properties, LV pulse power and wave analysis results
Table 4 Pre-operative and post-operative measures of LV diastolic function

4 Discussion

Our results show that immediately after deployment of an EVAR endograft there was no increase in aortic characteristic impedance, ascending aortic compliance or systemic vascular resistance. The positive reflected wave remained the same size, however the timing of its reflection was altered, with a greater portion now returning during systole. We found that the LV experienced an increased pulsatile load, though the LV stroke volume and cardiac index were not affected. Finally, there may have been a small impairment of LV diastolic function, but not to a great enough extent to meet statistical significance.

As the aorta stiffens pulse waves originating in the heart conduct to the periphery with increasing velocity. As a result, pulse wave velocity (PWV) can be used as an indirect measure of the relative stiffness of the arterial system. It has been frequently used in clinical studies quantifying changes in arterial stiffness due to aging or disease states. More importantly, increasing arterial stiffness as detected by elevated PWV has been linked to subclinical target organ disease and impaired cardiovascular health, including increased mortality [5,6,7,8,9]. The potential value of PWV as a clinical tool for improving long-term cardiovascular patient care has led to its inclusion in both the American Heart Association and European Society of Cardiologists’ most recent guidelines on the management of arterial hypertension and aortic disease [33, 34].

The endografts used in EVAR procedures are made with material that is stiffer than the native aorta, even in the setting of atherosclerotic or aneurysmal disease. Therefore, it is not surprising that an increased PWV measured across the EVAR site has been a consistently reported finding. Lantelme et al. assessed 50 patients following either open or endovascular AAA repair. After a median follow-up interval of 47 days they demonstrated an increase in mean PWV from a baseline of 11.3 to 12.3 m/s, or approximately 10% (p = 0.001) [13]. Of note, the hemodynamic alterations were different between the surgical graft prosthesis and the endograft, suggesting device material cannot be ignored. Post-EVAR increase in PWV was confirmed by Kadoglou, et al. with a 25% increase in mean PWV (13.11 m/s to 16.41 m/s) after 6 months [11]. The post-op intervals, population demographics and surgical specifics were not identical in each study, which may explain the difference in relative PWV change. Importantly, the finding of increased PWV post-EVAR was consistent and both authors suggest that this elevation may have clinical implications given the clear association of elevated PWV and cardiovascular outcomes in non-surgical populations. Their concern is further supported by the results of subsequent studies where an increased post-EVAR PWV was also associated with higher levels of inflammatory markers, increased LV mass, and LV diastolic dysfunction [12, 15, 16]. In one study, increased LV mass was detected as early as one week following an EVAR procedure [16].

Studies in animal models shed light onto the implications of increased PWV with regards to concurrent changes in aortic mechanical properties and possible mechanisms for the negative LV effects observed in humans. Using a canine model Dobson et al. implanted an endograft in the thoracic aorta [10]. There was a significant elevation in PWV post-deployment (418 cm s−1 vs. 755 cm s−1, p < 0.05), a finding consistent with the EVAR patient studies mentioned above. This change occurred alongside other changes in the ascending aortic mechanical properties. Significant post-stent increases occurred in proximal aortic impedance (Z0 = 0.250 mmHg mL−1 s−1 vs. 0.414 mmHg mL−1 s−1, p < 0.05). Additionally, total arterial compliance decreased (0.446 mL mmHg−1 vs. 0.276 mL mmHg−1, p < 0.05) alongside a near- significant increase in SVR (5.8 mmHg mL−1 s−1 vs. 7.7 mmHg mL−1 s−1, p = 0.06). There was also an alteration of reflected wave forms, with the reflected wave coefficient becoming dominated by a large negative reflected wave (0.09 vs − 0.49, p < 0.05). Due to endograft-aorta impedance mismatch one would expect to see larger positive and negative reflected waves, from the proximal and distal ends of the graft respectively which would typically cancel each other. The authors suggest the change in wave co-efficient in their study was due to increased proximal aortic stiffness suppressing the positive reflected wave, unmasking the negative wave. Another interesting finding was a measurable decrease in aortic flow (IE cardiac output) which neared statistical significance (0.89 L min−1 vs. 0.66 L min−1, p = 0.09), possibly as a result of the described changes in the aortic properties causing an increased LV hydraulic load. An alteration of aortic input impedance by thoracic grafting was similarly found in Mekkaoui et al. using a porcine model [35]. Of particular note, these changes resulted in an impairment of the natural diastolic augmentation of blood flow to the coronary and supra-aortic arteries. The findings of these papers suggest that (1) increased PWV may reflect other concurrent changes in aortic mechanics, including proximal characteristic impedance, (2) the net effect of these changes may have a deleterious effect on coronary blood flow, LV hydraulic load and cardiac output and (3) these results provide possible mechanistic explanations for the findings in post-EVAR human studies.

Previous studies in human subjects following EVAR had failed to demonstrate any “upstream” alteration in aortic stiffness [13, 36]. But these studies did not assess all of the aspects of aortic mechanical behaviour, nor did they measure aortic stiffness in the same fashion as the animal studies. This methodological discordance could easily explain the discrepant findings. Our study did not have such limitations. Even though our study methods closely approximated those in the animal studies, we did not find the changes in ascending aortic compliance or characteristic impedance identified by Dobson and Mekkaoui. An obvious explanation is that these changes simply do not occur in humans. Failed translation of animal models to human clinical studies is not uncommon. Other possible explanations do exist. Our patients, unlike the animal studies, had endografts deployed more distally in the aorta (abdomen vs thoracic) which may be too remote to induce changes in the ascending aorta. Alternatively, inhaled volatile anesthetic’s suppression of the sympathetic nervous system may have counteracted the proposed mechanism of sympathetic activation for the changes in local aortic mechanics following endograft implantation [37,38,39]. Our measured SVR was lower post-EVAR, likely reflecting the cumulative effect of general anesthesia on decreased sympathetic/vascular tone at the level of the small arteries and arterioles (considered to be the largest contributors to SVR) [40]. It should be noted that neither set of animals in the comparison studies were anesthetized in a similar fashion to our patients (IE general anesthesia with inhaled volatile anesthetics).

Reflected waves occur whenever a point of impedance mismatch occurs along the arterial tree. A transition point to higher or lower impedance results in positive and negative reflected waves respectively. There are naturally occurring reflection points in the normal human arterial system, a negative wave reflects near the origins of the abdominal vessels while a positive one originates near the femoral arteries. As the arterial tree stiffens due to aging and vascular disease the size and timing of the positive reflected wave may change. Instead of providing diastolic blood flow augmentation, the wave arrives during systole. This results in added hydraulic load on the left ventricle, impaired coronary blood flow, LV diastolic dysfunction and contribution to long term adverse cardiovascular outcomes [41,42,43]. Stiffer endograft material also creates an impedance mismatch which may affect wave reflection properties. Our data suggests that although the absolute size of the positive reflected wave did not change, the timing was altered with an increased systolic component. This systolic reflected wave augmentation likely explains the increased fractional pulse pressure and trend towards higher systolic blood pressure, with no corresponding change in mean or diastolic pressures. It may also explain the increased LV hydraulic load and trend towards lower cardiac output despite the marginally decreased aortic impedance. The mechanism for previously reported PWV associated changes in LV mass and diastolic function following EVAR may in fact be a result of endograft induced changes in reflected wave properties, not increased aortic impedance.

Diastolic function has been shown to deteriorate in the setting of increases in measures of arterial stiffness (including PWV), as well as fractional pulse pressure and reflected waves [2, 32, 44,45,46,47,48,49]. If implantation of an abdominal endograft in an EVAR procedure alters any or all of these aortic properties, there may be clinical implications on left ventricular diastolic function. Several of our measures of LV diastolic function suggest relaxation is impaired following EVAR, though they did not meet statistical significance. More time may be required for diastolic function to be affected following EVAR. Takeda et al. found no change in diastolic function in the setting of increased PWV post-EVAR. However, at a 1 year follow-up over half the patients had deterioration in exercise tolerance and reduced diastolic function [15]. Therefore, although we did see statistically significant changes in diastolic function, there is both substrate (IE immediate increase in LV hydraulic load) and evidence to suggest that given enough time or a larger sample size we may see diastolic dysfunction in these patients.

4.1 Limitations

Our parameters, though comprehensive, were only measured immediately following endograft deployment. Therefore, we do not know how any of these might progress over time. Do those that become altered, such as reflected wave properties and LV hydraulic load, remain the same, become more prominent, or eventually normalize? Do factors that did not change, such as proximal aortic compliance and diastolic function change over time? During intra-operative data collection our patients were under general anesthesia which impacts LV preload, afterload and sympathetic tone, potentially affecting our results. We were unable to include measurements of PWV concurrently with our other parameters. This limits the ability to put our findings into context with those from studies where increased PWV was measured post-EVAR. Given the ubiquity of elevated PWV following EVAR it is reasonable to assume we would have had a similar increase as well. We are not able to comment on the effects of different graft manufacturers or materials on the parameters measured because a majority of our patients had the same endograft implanted. Previous studies have suggested that differences between graft materials affect measures of aortic stiffness [12, 13, 37]. Finally, we have no control group with which to compare. One study that did include matched controls found that although EVAR grafts acutely increased PWV and reflected waves, these changes remained stable over time. The control group had gradual increases in PWV and reflected waves over time, and eventually the two groups were similar [14]. This suggests that the immediate changes in aortic stiffness, wave reflections and cardiac function following EVAR may just accelerate changes to the cardiovascular system which would have occurred over time in a non-surgical cohort. Whether it is detrimental to EVAR patients to have these changes occur acutely and sooner than would naturally occur is a question that requires further study.

5 Conclusion/Future Directions

Primarily, our study demonstrates that speckle tracking echocardiography represents a promising tool to non-invasively quantify aortic biomechanics in a real-time clinical setting.

In the setting of EVAR for abdominal aneurysms, measures of aortic stiffness that directly affect LV afterload, such as aortic impedance, ascending aortic compliance and SVR do not immediately increase following endograft implantation. The graft does appear to create an impedance mismatch which results in changes to the timing of positive reflected waves resulting in a greater systolic component. This timing shift is associated with an increased LV hydraulic load, which showed a trend towards decreased cardiac output and impaired LV diastolic relaxation [23].

To better describe the natural progression of aortic and cardiac changes following EVAR further studies should focus on the following: (1) Measure the corresponding changes in proximal aortic impedance, ascending aortic compliance, pulse wave analysis, PWV, LV hydraulic load, and LV systolic/diastolic function immediately following the procedure (IE after effects of general anesthesia are gone), then again at mid- and long-term intervals. (2) Examine these same parameters after placing stents in more proximal locations such as the thoracic aorta, aortic arch and ascending aorta. (3) Perform a similar comprehensive analysis with a variety of graft materials and designs (both surgical and endovascular) to determine inter-graft differences. These additional studies would allow for a more thorough description of the changes that occur after endovascular graft procedures, the clinical implications of these changes, and further explore ways to minimize any potential negative effects with different graft designs or materials.