Skip to main content
Log in

Development and implementation of a quality control protocol for B-mode ultrasound equipment

  • Published:
Journal of Ultrasound Aims and scope Submit manuscript

Abstract

Purpose

Quality assurance (QA) of ultrasound (US) equipment is currently required in only a few countries around the world. In Greece, no national or other norms exist for regulating the use of US equipment. However, to obtain accreditation for the radiology department of a Greek hospital, the establishment and implementation of a quality control (QC) protocol and a QA programme for US equipment was required.

Materials and methods

A literature review regarding US QC/QA procedures was performed. The information collected was used as a guide to create a QC/QA protocol and to obtain an appropriate US QC phantom. Drafting and testing of the initial protocol lasted 6 months. Its final version was implemented for 18 months in two US systems and five US transducers.

Results

The QC tests included in the protocol evaluate mechanical and electrical safety, image display, uniformity, penetration depth, distance accuracy, greyscale display, anechoic object imaging, geometric distortion, and axial/lateral resolution. The only QC test that failed was the test for uniformity since intense non-uniformities were observed that led to the replacement of two linear transducers.

Conclusion

US imaging is considered safe and, where appropriate, is preferred over imaging modalities that use ionizing radiation. However, the lack of QC/QA implies that US image quality is not routinely monitored. Therefore, the possibility of malfunctions that may go undetected and lead to wrong diagnosis cannot be excluded. A QC/QΑ programme can contribute to the elimination of such errors and ensure that performance is maintained over time.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11

Similar content being viewed by others

References

  1. World Health Organization (2011) Manual of diagnostic ultrasound. Geneva, Switzerland

  2. Bushberg JT, Seibert JA, Leidholdt EM (2012) The essential physics of medical imaging, 3rd edn. Lippincott Williams & Wilkins, Philadelphia

    Google Scholar 

  3. Boote EJ (2003) AAPM/RSNA physics tutorial for residents: topics in US, B-mode US: basic concepts and new technology. Radiographics 23:1315–1327

    Article  Google Scholar 

  4. Hangiandreou NJ (2003) AAPM/RSNA physics tutorial for residents: topics in US Doppler US techniques: concepts of blood flow detection and flow dynamics. Radiographics 23:1019–1033

    Article  Google Scholar 

  5. United States Food and Drug Administration. https://www.fda.gov/radiation-emitting-products/medical-imaging/ultrasound-imaging. Accessed 15 Sept 2020

  6. Kofler JM (2001) Quality assurance of ultrasound imagers: procedures, expectations and philosophies. AAPM, 43rd Annual Meeting

  7. Dudley N, Russell S, Ward B, Hoskins P, BMUS QA working party (2014) The BMUS guidelines for regular quality assurance testing of ultrasound scanners. Ultrasound 22:6–7

    Article  Google Scholar 

  8. Hangiandreou N, Stekel S, Tradup D, Gorny K, King D (2011) Four-year experience with a clinical ultrasound quality control program. Ultrasound Med Biol 37(8):1350–1357

    Article  Google Scholar 

  9. Balbis S, Meloni T, Tofani S, Zenone F, Nucera D, Guiot C (2012) Criteria and scheduling of quality control of B-mode and Doppler ultrasonography equipment. J Clin Ultrasound 40(3):167–173

    Article  Google Scholar 

  10. Goodsitt M, Carson P, Witt S, Hykes D, Kofler J (1998) Real time B-mode ultrasound quality control test procedures. Report of AAPM ultrasound task group No1. Med Phys 25(8):1385–1406

    Article  CAS  Google Scholar 

  11. IPEM (2010) Report No102. Quality assurance of ultrasound imaging systems

  12. User’s Guide, 410 HE Series, Multi-Purpose Accreditation Phantom, Gammex Inc, Middleton, USA, 2017

  13. AIUM (2008) American Institute of Ultrasound in Medicine (AIUM), Routine quality assurance for diagnostic ultrasound equipment, Laurel, USA

  14. Dudley N, Russel S, Ward B, Hoskins P, BMUS QA working party (2014) BMUS guidelines for the regular quality assurance testing of ultrasound scanners by sonographers. Ultrasound 22(1):8–14

    Article  Google Scholar 

  15. Kollmann C, deKorte C, Dudley N, Gritzmann N, Martin K, Evans D (2012) Guidelines for technical quality assurance of ultrasound devices (B-mode)—EFSUMB technical quality assurance group, US-TQA/B, Version 1 July 2012. Ultraschall Med 33:544–549

    Article  CAS  Google Scholar 

  16. Sipilä O, Mannila V, Vartieinen E (2011) Quality assurance in diagnostic ultrasound. Eur J Radiol 80:519–525

    Article  Google Scholar 

  17. AS Filho (2020) Ultrasound quality control. https://leanpub.com/cqus_guidelines/read#leanpub-auto-image-uniformity. Accessed 15 Sept 2020

  18. Choi JI, Jung SE, Kim PN, Cha SH, Jun JK, Lee HY et al (2014) Quality assurance test of screening ultrasound examination of hepatocellular carcinoma using a standardized phantom and standard images: a 3-year nationwide survey in Korea. J Ultrasound Med 33:985–995

    Article  Google Scholar 

  19. Dudley NJ, Woolley DJ (2016) A simple uniformity test for ultrasound phased arrays. Phys Med 32(9):1162–1166

    Article  Google Scholar 

Download references

Funding

The authors did not receive support from any organization for the submitted work. No funding was received to assist with the preparation of this manuscript.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Virginia Tsapaki.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tsapaki, V., Tsalafoutas, I.A., Triantopoulou, S.S. et al. Development and implementation of a quality control protocol for B-mode ultrasound equipment. J Ultrasound 25, 155–165 (2022). https://doi.org/10.1007/s40477-021-00579-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s40477-021-00579-7

Keywords

Navigation