FormalPara Key Summary Points

Why carry out this study?

Understanding the nature of postoperative intraocular lens (IOL) shift, refractive change, and margin of error in predicted postoperative refraction (PPR) is crucial for the accuracy of PPR.

We aimed to assess the relationships between different fixation methods of IOLs with different haptic designs and describe the correlation between IOL shift, postoperative refraction, margin of error in PPR, and preoperative biometric variables.

What was learned from the study?

The relationship between the postoperative IOL shift and refraction change varied among IOLs with different haptic designs; preoperative anterior chamber depth, lens thickness, and axial length may have predictive roles.

The study results indicate that capsulorhexis may be more stable than in-the-bag IOL fixation, leading to improved accuracy of PPR and better refractive outcomes.

Introduction

Cataract surgery has become a routine part of refractive surgery. One of the main causes of postoperative patient dissatisfaction is residual refractive error [1]. An ultrasonography-based A-scan was traditionally used to measure axial length (AL), which is the most common cause of error in intraocular lens (IOL) calculations [2]. The use of partial coherence interferometry improved the accuracy of AL measurements to five times that of the A-scan [3]. Additionally, new IOL calculation formulae have increased the accuracy of predicted postoperative refraction (PPR) [4]. However, postoperative refraction changes are sometimes observed, even in the early postoperative period. This makes choosing the IOL power of the fellow eye challenging.

Previous studies have reported a relationship between postoperative IOL shift and postoperative refraction. Nejima et al. indicated that postoperative IOL shift could induce postoperative refractive changes [5]. Erickson demonstrated the influence of the postoperative IOL position on the postoperative refractive error [6]. The results of these studies suggest that the stability of IOLs may be an important factor in improving the accuracy of PPR. Remón et al. investigated the influence of material and haptic design on the mechanical stability of IOLs and concluded that haptic design had more influence on stability than material properties [7]. Regarding IOL placement methods, in-the-bag fixation is the most common procedure, while capsulorhexis has been recently introduced [8, 9, 10].

Given that postoperative refraction in the operated eye is often considered when choosing the IOL power of the fellow eye, knowledge of the postoperative IOL shift and its influence on postoperative refraction changes in the early postoperative period is crucial for improving the accuracy of PPR.

The objective of this study was to investigate the postoperative IOL shift in three different IOLs with different haptic designs and different IOL fixation methods, as to the best of our knowledge no previous study has done so. In addition, we aimed to measure their effect on postoperative refraction changes in the early postoperative period. The correlations between IOL shift, postoperative refraction changes, and preoperative biometric variables were also analyzed.

Methods

This single-center, open-label study was conducted at the Yokosuka Chuo Eye Clinic, and operations were performed between May 2018 and November 2019. Detailed information about this research was provided and informed consent was obtained from all participants. All research procedures were performed in accordance with the Declaration of Helsinki. This study was approved by the Ethics Committee of the Yokosuka Chuo Eye Clinic.

Patient and IOL Selection

The study included 69 eyes of 69 patients. The AcrySof IQ (model SN60WF; Alcon, Fort Worth, TX, USA), FineVision Pod F (PhysIOL, Liège, Belgium), and FEMTIS (Oculentis, Berlin, Germany) lenses were used in 27, 24, and 18 eyes, respectively. All operations were performed by the same surgeon at the Yokosuka Chuo Eye Clinic. All participants had age-related cataracts with a nucleus grading of > 2 based on the Emery–Little classification and were randomly chosen by an ophthalmologist. Patients with pseudoexfoliation glaucoma, uveitis, proliferative diabetic retinopathy, or a history of intraocular surgery, laser treatment, and retinal problems, or subjects with corrected postoperative visual acuity 20/40, were excluded.

The eye specialist informed all patients about the characteristics of each IOL. The IOLs were selected according to patient demands. FEMTIS is a haptic-designed, hydrophilic, acrylic plate with four flanges for capsulorhexis fixation (Fig. 1a). It has an anteroposterior, aspheric, biconvex optic with a diameter of 5.7 mm, an overall length of 10.5 mm, and no haptic angulation (0°). FineVision is a hydrophobic, acrylic, single-piece IOL with double C-loop haptics (Fig. 1b). It has an anteroposterior, aspheric, biconvex optic with a diameter of 6.0 mm, an overall length of 11.4 mm, and a haptic angulation of 5°. AcrySof IQ is a hydrophobic, acrylic, single-piece IOL with a single C-loop haptic (Fig. 1c). It has an anteroposterior, aspheric. biconvex optic with a diameter of 6.0 mm, an overall length of 13.0 mm, and a haptic angulation of 0°.

Fig. 1
figure 1

Types of IOLs: AcrySof IQ (a), FineVision (b), FEMTIS (c)

Pre- and Postoperative Examinations

All patients underwent complete preoperative ophthalmic examinations. Nucleus grading was classified according to slit-lamp examination findings by the same eye specialist. The vision test for subjective refraction was performed by the same technician for all patients. Swept-source anterior segment optical coherence tomography imaging was performed using the CASIA2 system (Tomey Corp., Nagoya, Japan) to detect anatomical abnormalities. Anterior chamber depth (ACD), lens thickness (LT), and AL were measured using an IOLMaster 700 device (Carl Zeiss, Oberkochen, Germany). The PPR and recommended IOL power were calculated using the IOLMaster 700.

In addition to the ophthalmic examinations, the IOL position was measured using CASIA2 on postoperative day 1, week 1, and month 1. The lens position was measured from the apex of the corneal surface to the center of the IOL (Fig. 2). Subjective postoperative refraction was evaluated by the same technician. Refraction was reported as the spherical equivalent (SE). Due to a possible margin of error caused by different examiners, vision and postoperative IOL positions were measured by the same experienced examiner.

Fig. 2
figure 2

Measurement of the postoperative IOL position (corneal F to lens F)

Surgical Technique

All operations involved femtosecond laser-assisted cataract surgery using the LenSx platform (Alcon Laboratories, Inc., Fort Worth, TX, USA). The same surgeon performed the laser and manual operations. Before the laser process, topical anesthesia with 0.4% oxybuprocaine eye drops (Benoxyl; Santen Pharmaceutical Co., Osaka, Japan) was administered after maximal dilation of the pupil with 0.5% tropicamide and 0.5% phenylephrine hydrochloride (Mydrin-P, Santen Pharmaceutical, Osaka, Japan). Preoperative planning of capsulotomy and lens fragmentation was performed using the LenSx platform. A 5.0 mm capsulotomy and chop-and-cylinder fragmentation pattern was selected, with three chops, each measuring 6 mm in length. The posterior offset was set at 800 μm. No dock-related issues were observed while using SoftFit™ (Alcon Laboratories, Inc., Fort Worth, TX, USA) in all cases. A 5.0 mm capsulotomy centered on white-to-white was created with 8.0 mJ of energy (spot and layer separations: 9 μm each). Nuclear fragmentation was performed using the chop-and-cylinder technique with 8.0 mJ of energy (spot and layer separations: 9 μm each). The anterior and posterior delta values were 350 μm.

A temporal clear corneal incision of 2.40 mm was made using an ophthalmic knife (MANI Ophthalmics, Tochigi, Japan). The anterior chamber was filled with a dispersive viscoelastic VISCOAT (Alcon Laboratories, Fort Worth, TX, USA) and a cohesive viscoelastic OPELEAD (Senju Pharmaceutical Co., Ltd., Osaka, Japan) using the soft-shell technique. Capsulotomy was carefully performed using capsule forceps, and cortico-cleaving hydrodissection was performed. Emulsification and aspiration of the fragmented lens and aspiration of the cortex were conducted using the Centurion® Vision System (Alcon Laboratories, Inc., Fort Worth, TX, USA). After filling the capsule bag with OPELEAD, the IOLs were inserted into the bag using an IOL injector. The attachment of the enclaved flanges to the rim of the capsulorhexis was performed using the reverse optic capture technique. The viscoelastic material was removed before watertight corneal wound closure was confirmed.

Data Analysis

The postoperative IOL longitudinal shift was calculated as follows: the IOL shift between 1 day and 1 week was the IOL position on postoperative day 1 minus the position at postoperative week 1; the IOL shift between 1 week and 1 month was defined as the IOL position on postoperative week 1 minus the position on postoperative month 1. A positive IOL shift indicates that the IOL showed a forward shift, whereas a negative number indicates a backward shift during a particular period.

The postoperative refraction change was calculated as follows: the refraction change between 1 day and 1 week was the PR in postoperative week 1 minus the PR on postoperative day 1; the refraction change between 1 week and 1 month was defined as the PR in postoperative month 1 IOL minus the PR in postoperative week 1. These were analyzed in terms of both the real and absolute values. In real values, a positive number refraction change indicates a hyperopic shift, whereas a negative number indicates a myopic shift during a particular period. In absolute values, the numbers indicate the degree of refraction change during a particular period.

The margin of error of postoperative refraction, which was calculated as the postoperative refraction minus the predicted postoperative refraction (PR-PPR) in three different types of IOLs, was also analyzed at 1 day, 1 week, and 1 month after surgery. These results are shown as real and absolute values. When analyzing real values, a positive PR-PPR value indicates that the postoperative refraction (PR) showed a hyperopic shift compared to the PPR, whereas a negative PR-PPR value indicated that the postoperative refraction showed a myopic shift compared to the PPR.

We performed an analysis of variance (ANOVA) test for the numerical values that followed a normal distribution (assessed with the Shapiro–Wilk test), the Kruskal–Wallis test for numerical values not following a normal distribution, and Fisher’s exact test to analyze data from categorical variables. We tested the following numerical variables: IOL shift between 1 day and 1 week and 1 week and 1 month postoperatively, and the PR-PPR at 1 day, 1 week, and 1 month postoperatively. We investigated whether there were statistically significant differences in terms of eye laterality (right/left), sex, age, preoperative ACD, LT, and AL; IOL shift and change in PR between 1 day and 1 week and 1 week and 1 month among the types of IOLs; and PR-PPR difference at 1 day, 1 week, and 1 month postoperatively among the categories (types) of IOLs.

Furthermore, we used Dunn’s test [11] to perform multiple comparisons with a Bonferroni adjustment of the p-value. We determined which categories (types) of IOLs had significant differences in IOL shifts and changes in PR between 1 day and 1 week and 1 week and 1 month. Moreover, we determined the PR-PPR differences at 1 day, 1 week, and 1 month postoperatively.

Spearman's rank correlation coefficients were used to examine whether preoperative ACD, preoperative LT, and preoperative AL correlated with the following variables during each study period: postoperative IOL shifts, the degree of change in PR, and the degree of margin of error. The above tests were only performed for two types of IOLs (FineVision and AcrySof IQ) to ensure consistency in the types of lenses being compared (i.e., in-the-bag-fixation IOLs) and also because of the small number of patients in the capsulorhexis (FEMTIS) group.

The significance level was set at p < 0.05. All statistical analyses were performed using R (R Foundation for Statistical Computing, Vienna, Austria; https://www.R-project.org/).

Results

The descriptive statistics and the results of the ANOVA, Kruskal–Wallis test, and Fisher's exact test, according to the three IOL types, are presented in Table S1. Biometric measurements are presented in Table 1. The proportion of male patients in the AcrySof IQ, FineVision, and FEMTIS lens groups was 35%, 42%, and 43%, respectively.

Table 1 Clinical characteristic of the eyes in each intraocular lens group

Eye laterality (right/left) and sex were not significantly different (p > 0.05) among the different IOL groups. Age was significantly different (x2 = 6.2706, p = 0.043) between the IOL groups, which was not confirmed by Dunn’s test with the Bonferroni adjustment (all p values > 0.05).

The results of the ANOVA for preoperative ACD and LT and the Kruskal–Wallis test for AL also showed that they were not significantly different (p > 0.05) among the different types of IOLs.

The IOL shift from 1 day to 1 week postoperatively, tested in both real and absolute values, showed significant differences (for both cases: p < 0.01) among the IOL types (Fig. 3a, b, respectively). The IOL shift was significantly lower with FEMTIS than with both FineVision and AcrySof IQ  (Z = −5.199565, p < 0.01; and Z = −5.9472881, p < 0.01, respectively). The IOL shift in real values was negative for the FEMTIS and positive for both FineVision and SN60WF (Fig. 3a).

Fig. 3
figure 3

a, b Graph showing mean lens shift (in mm) in real (a) and absolute (b) values between 1 day and 1 week postoperatively by lens type. Square dots represent the median values of the shifts, and the upper and lower whiskers represent the upper and lower percentiles

The IOL shift from 1 week to 1 month postoperatively, tested in both real and absolute values, showed significant differences (p < 0.01) among the IOL types (Fig. 4a, b). Dunn’s test showed that the IOL shift from 1 week to 1 month postoperatively was significantly smaller with FEMTIS than with FineVision and AcrySof IQ  (Z = 3.474281, p < 0.01; and Z = 6.099305, p < 0.01, respectively). In addition, the IOL shift showed significant differences (Z = 2.754178, p < 0.05) between the FineVision and AcrySof IQ  IOL types. The IOL shift in real values was negative for all three FEMTIS, FineVision, and AcrySof IQ  IOLs (Fig. 4a).

Fig. 4
figure 4

Graph showing mean lens shift (in mm) in real (a) and absolute (b) values between 1 week and 1 month postoperatively by lens type. Square dots represent the median values of the shifts, and the upper and lower whiskers represent the upper and lower percentiles

The overall mean postoperative refraction (PR) was −0.090 ± 0.153. The change between 1 day and 1 week, tested in both real and absolute values, showed significant differences (p < 0.01) among the IOL types (Fig. 5a, b), with the FEMTIS IOL showing smaller changes than both the FineVision and AcrySof IQ  IOLs. The PR change between 1 week and 1 month, tested in both real and absolute values, showed significant differences (p < 0.01) between the IOL types (Fig. 6a, b), with the FEMTIS IOL showing smaller changes than both the FineVision and AcrySof IQ  IOLs.

Fig. 5
figure 5

Graph showing the mean change in PR in real (a) and absolute (b) values between day 1 and week 1 postoperatively by lens type. Square dots represent the median values of the shifts, and the upper and lower whiskers represent the upper and lower percentiles

Fig. 6
figure 6

Graph showing the mean change in PR in real (a) and absolute (b) values between week 1 and month 1 postoperatively by lens type. Square dots represent the median values of the shifts, and the upper and lower whiskers represent the upper and lower percentiles

The PR-PPR difference was −0.059 ± 0.063. On day 1, tested in absolute values, significant differences (p < 0.01) were found among the IOL types (Fig. 7a), with the AcrySof IQ  IOL having the highest absolute values at this time point. Dunn’s test showed that the absolute values for the PR-PPR differences at 1 day after the operation were not significantly different (Z = −1.24, p > 0.05) between the FEMTIS and FineVision lenses, while they were significantly different between the FEMTIS and AcrySof IQ  lenses (Z = −4.53, p < 0.01) and the FineVision and AcrySof IQ  lenses (Z = −3.54, p < 0.01; Table S2). The PR-PPR difference at 1 week, tested in absolute values, showed significant differences (p < 0.05) among the IOL types (Fig. 7b). The absolute value of the PR-PPR difference at 1 week after the operation was significantly different (Z = −3.38, p < 0.01) between the FEMTIS and FineVision lenses, while it was not significantly different between the FEMTIS and AcrySof IQ  lenses (p > 0.05) or the FineVision and AcrySof IQ  lenses (p > 0.05). The PR-PPR difference at 1 month, tested in absolute values, showed significant differences (p < 0.05) among the IOL types (Fig. 7c). Dunn’s test showed that the absolute value of the PR-PPR difference at 1 month after the operation was significantly different (Z = −3.33, p < 0.01) between the FEMTIS and FineVision lenses, while it was not significantly different between the FEMTIS and AcrySof IQ  lenses (p > 0.05) or the FineVision and AcrySof IQ  lenses (p > 0.05).

Fig. 7
figure 7

Graph showing PR-PPR difference at 1 day (a), 1 week (b), and 1 month (c) postoperatively by lens type. Square dots represent the median values of the shifts, and the upper and lower whiskers represent the upper and lower percentiles

Spearman’s rank correlation coefficient tests for the two in-the-bag IOLs (FineVision and AcrySof IQ) are shown in Tables S3–S5. The IOL shift from 1 day to 1 week showed a strong positive correlation with both ACD and AL (rho = 0.83, rho = 0.90, respectively; p < 0.01), and a strong negative correlation with LT (rho = −0.90; p < 0.01). The changes in PR (absolute values) from 1 day to 1 week showed a strong positive correlation with both ACD and AL (rho = 0.71, rho = 0.77, respectively; p < 0.01), and a strong negative correlation with LT (rho = −0.77; p < 0.01).

The PR-PPR difference on day 1 in absolute values showed a weak positive correlation with both ACD and AL (rho = 0.44, rho = 0.36, respectively; p < 0.05), and a weak negative correlation with LT (rho = −0.36; p < 0.05). The PR-PPR difference at week 1 in absolute values did not show any significant correlation with ACD, AL, or LT (p > 0.05). The PR-PPR difference at month 1 in absolute values did not show any significant correlation with ACD (p > 0.05), but it did show a weak positive correlation with AL (rho = 0.28, p < 0.05) and a weak negative correlation with LT (rho = −0.31; p < 0.05).

Discussion

Outcomes following IOL implantation have become an important research area [12, 13]. Based on our clinical experience, we wondered why some patients showed refraction changes after surgery and why these changes varied from patient to patient and by IOL type. Erickson insisted that postoperative longitudinal shifts might be the principal contributors to postoperative refractive errors [6]. Considering that one of the main causes of patient dissatisfaction after IOL implantation is postoperative refractive errors [1], ophthalmologists need to understand postoperative IOL longitudinal shifts and their influence on postoperative refractive changes [14]. This knowledge can also help clinicians better predict postoperative refraction in the early postoperative period and choose a suitable IOL power for the fellow eye [14]. Furthermore, understanding the relationship between preoperative biometric variables, postoperative IOL shifts, and postoperative refraction changes is also useful for improving the accuracy of PPR.

In this study, the FEMTIS IOL, which was fixed to the anterior capsule, showed the least postoperative IOL longitudinal shift, the least change in postoperative refraction, and the smallest difference between PPR and actual postoperative refraction in the early postoperative period. These results are in line with previous studies that also used a capsulorhexis-fixated IOL with slightly different designs and indicated greater stability with FEMTIS. Ganesh et al. and Dick et al. reported similar results in terms of postoperative IOL longitudinal shift and postoperative refraction changes [8, 9]. Kim et al. investigated lens stability and refractive outcomes after cataract surgery using primarily posterior capsulorhexis fixation and concluded that this fixation method was useful in stabilizing the postoperative IOL position and minimizing postoperative refractive changes [15].

The success of the capsule fixation technique requires a consistent size and shape of the capsulorhexis [16]. However, ensuring a consistent size and shape of the capsulorhexis is challenging even for an experienced surgeon. Recently released cutting-edge devices, such as femtosecond laser systems, selective capsulotomy laser systems, and precision pulse capsulotomy, can be used to tackle these challenges [17,18,19,20,21]. Among these, the femtosecond laser system is the only device that is used to create the capsulorhexis on the visual axis. One previous study reported that aspheric IOLs can be very sensitive in quality of vision due to decentration depending on the aberration correction concept of the IOL design. In aberration-correcting IOLs, coma aberration and astigmatism significantly increase with increasing decentration [22]. Kránitz et al. showed that IOL decentration was six times as likely to occur when the capsulorhexis was created manually as when it was created using the femtosecond laser technique [20]. Some studies have reported that IOL decentration can also induce postoperative astigmatism [6]. The results of these studies imply that a combination of anterior capsule fixation of the IOL and capsulorhexis creation with a femtosecond laser may be ideal for stabilizing the postoperative IOL position and postoperative quality of vision.

Several studies have investigated postoperative refractive changes and postoperative IOL longitudinal shifts in in-the-bag fixated IOLs [5, 23,24,25]. In our study, the capsulorhexis-fixation IOL FEMTIS was the most stable throughout the study period. In contrast, in-the-bag-fixation IOLs AcrySof IQ  and FineVision showed significant forward shifts between 1 day and 1 week, which means that these IOLs showed myopic shifts during this period. The degree of the shift between 1 day and 1 week was the most significant compared to the other periods. One of the main differences between previous research and our work are the periods of the assessments. While most previous studies assessed shifts of in-the-bag-fixation IOLs during a relatively long postoperative period, we compared them in the early postoperative period, up to 1 month after surgery. Similar to our study, Koeppl et al. [26] investigated postoperative IOL longitudinal shifts and postoperative refraction changes among in-the-bag-fixation IOLs in the early postoperative period; they reported that the changes in both were the most remarkable between 1 day and 1 week after the operation, after which they became more stable, which corresponds to our findings. Both their and our results indicate that IOL longitudinal shifts and refraction changes may be observed even in the early postoperative period. Therefore, it may be wise to consider these tendencies when choosing the IOL power of the fellow eye.

In our study, AcrySof IQ and FineVision showed similar overall effects on postoperative IOL longitudinal shift, but the magnitudes of the effects were different. These two in-the-bag-fixation IOLs have different specifications, including the material they are made of, their length, edge designs, and haptic designs, all of which can cause differences in the magnitude of the postoperative shift. In fact, previously published studies reported that capsular bag shrinkage, IOL design (haptic or optic), and IOL material may influence changes in postoperative longitudinal IOL shifts. Koeppl et al. insisted that significant shrinkage of the capsular bag in the first postoperative week was correlated with a forward IOL shift [26]. Petternel et al. assessed the influence of optic-edge design and optic-haptic angulation of IOLs on postoperative longitudinal shifts [27] and found that the influence of optic-haptic angulation was significantly greater than that of edge design. Wirtitsch et al. investigated the effect of haptic design on postoperative IOL longitudinal shifts and reported that single-piece IOLs were more stable than three-piece IOLs [14]. It has also been reported that the IOL material influences postoperative IOL shifts [23]. Thus, the various specifications and characteristics of different IOLs may cause postoperative IOL shifts. Therefore, it is difficult to conclude which IOL specification and characteristics primarily resulted in the differences in magnitudes of the postoperative IOL shifts in both acrylic single-piece IOLs with different haptics. These issues will be addressed in future studies.

Our results also indicated significant differences in the margin of error of predicted postoperative refraction (PR-PPR) between in-the-bag-fixation IOLs and capsulorhexis-fixation IOLs. Although these differences were not consistent over time, the capsulorhexis-fixation IOL FEMTIS showed a smaller margin of error than the in-the-bag-fixation IOLs throughout the study period. In particular, the PR-PPR with FEMTIS was significantly smaller than that of the in-the-bag-fixation IOLs at 1 day postoperatively.

As a result, FEMTIS was most stable in all aspects in this study including postoperative IOL shift, change in postoperative refraction (PR), and PR-PPR. Taken together, these results suggest that optic capture in capsulorhexis fixation may be more stable in terms of the postoperative IOL position and possibly refractive outcomes than in-the-bag fixation. Therefore, capsulorhexis fixation may be associated with improved PPR accuracy. This notion is supported by previous research on the mechanism of postoperative movement of in-the-bag fixated IOLs. They reported that postoperative IOL decentration and dislocation can be attributed to a combination of the high flexibility of in-the-bag-fixation IOLs, their haptic configuration, and the postoperative fibrosis and shrinkage of the capsular bag [28,29,30]. Therefore, we can theoretically assume that since the optics of the capsule-fixated IOLs are fixed in the capsulorhexis, they may be less influenced by postoperative fibrosis and shrinkage of the bag.

Interestingly, when comparing the two in-the-bag IOLs with each other, the AcrySof  IOL had a significantly higher absolute PR-PPR difference (and thus margin of error) than the FineVision lens at day 1, week 1, and month 1. Of note, the margin of error with the AcrySof  IOL showed a hyperopic shift that was the most pronounced on day 1. Given the fact that AcrySof  showed significant postoperative IOL shifts and changes in refraction in the very early postoperative period, it may be wiser for eye surgeons to wait until week 1 to choose the IOL power.

Limited research on the correlation between preoperative biometric variables and postoperative IOL longitudinal shift motivated us to analyze the relationship between such variables, such as ACD, LT, and AL [31], and postoperative IOL longitudinal shift, refraction change, and PR-PPR. In analyzing the in-the-bag IOLs, we found significant positive correlations between IOL shifts and changes in PR from day 1 to week 1 postoperatively in terms of preoperative ACD and AL. We also observed a significant negative correlation between IOL shifts from postoperative day 1 to week 1.

Our results indicate that the larger the preoperative ACD and the longer the preoperative AL, the more the IOLs shifted and the more the PR changed in the early postoperative period. In addition, the smaller the preoperative LT, the more the IOLs shifted, and the more the PR changed. Therefore, preoperative ACD, AL, and LT appear to be predictive factors for estimating the degree of IOL shift and changes in PR. These results indicate that these preoperative variables may be promising predictive factors. A larger preoperative ACD, longer preoperative AL, and smaller preoperative LT were all associated with a larger margin of error of postoperative refraction, but to a lesser degree than observed with IOL shift and PR changes.

There were some limitations to our study. First, the detailed mechanism of postoperative IOL longitudinal shift in the in-the-bag fixated IOLs SN60WF and FineVision was not investigated. In addition, the effect of IOL material, which may potentially play an important role, was not assessed. Future research will be needed to assess whether the differences can be explained, even in part, due to IOL material. Second, this study investigated only one type of capsulorhexis-fixation IOL. Different capsulorhexis-fixation IOLs with different materials and designs should be investigated in the future. Third, one of the main research objectives of this study was to compare postoperative IOL behaviors between in-the-bag and capsulorhexis fixations. However, since the three types of IOLs are associated with various factors that may affect their postoperative behavior, this study did not precisely investigate the difference between in-the-bag and capsulorhexis fixation. To precisely compare different IOL fixation methods, the same IOL should be used in different placements. Finally, only a single center was used for the recruitment of patients in this study, which may have affected the generalizability of our results. Further studies across multiple institutions are needed to validate our results.

Conclusions

When eye surgeons choose in-the-bag-fixation IOLs, it is vital that they understand the nature of potential postoperative refraction changes and the margin of postoperative refraction error. Preoperative ACD, LT, and AL may be useful factors for predicting IOL shift, change in PR, and to a lesser extent the degree of margin of error in clinical practice. Familiarity with these points could also contribute to improved accuracy of PPR. Additionally, our results indicate that in the early postoperative stage, capsulorhexis-fixation is more stable than in-the-bag IOLs in terms of postoperative IOL shifts and changes in postoperative refraction; it also showed the lowest margin of error of predicted postoperative refraction. Capsulorhexis fixation may be the most stable and promising IOL fixation method for improving the accuracy of PPR.