Introduction

Infrared (IR) laser ablation is an efficient means for ablating particulates from the surfaces of water-containing samples, such as plant and animal tissues [1]. IR laser ablation is utilized in several ambient mass spectrometry techniques, such as laser ablation electrospray ionization (LAESI) [2], matrix-assisted laser desorption electrospray ionization (MALDESI) [3], infrared laser ablation metastable-induced chemical ionization (IR-LAMICI) [4], and laser ablation atmospheric pressure photoionization (LAAPPI) [5]. A major advantage of IR laser ablation is that in contrast to UV laser ablation, which is used in, e.g., matrix-assisted laser desorption/ionization (MALDI), IR laser ablation does not require the addition of a light-absorbing matrix. Instead, the radiation of the IR laser at 2.94 μm is efficiently absorbed by water molecules, which are naturally abundant in tissue samples. Absorption of the IR laser radiation by water leads to explosive evaporation of water and ablation of the target. Ionization at this step is inefficient and, therefore, the neutral analytes within the plume are typically ionized by varying post-ionization methods, such as electrospray ionization (e.g., LAESI and MALDESI), metastable-induced chemical ionization (e.g., IR-LAMICI), or photoionization (e.g., LAAPPI).

Owing to the efficient ablation of tissues, IR laser ablation has been extensively utilized in MS imaging of plant and animal tissues [614]. In MS imaging, the size of the structures that can be distinguished is dependent on the spatial resolution of the desorption/ablation technique in question (i.e., in case of laser ablation, the size of the ablated spot). Typically, spot sizes that have been reported with IR laser ablation are in the range of 220–400 μm [2, 4, 9, 15, 16]. Smaller spot size has been achieved by delivering the mid-IR laser pulses through an etched tip of a glass fiber, which is placed in close proximity of the sample [17]. The tip confines the laser beam to the diameter of the tip, and roughly 30–40 μm ablation spot diameters have been achieved by this method in LAESI. This, however, requires very careful positioning of the fiber tip and exposes it to ablated sample material. On the other hand, this setup may decrease sensitivity due to partial blockage of the ejected plume by the fiber. In atmospheric pressure-infrared laser desorption/ionization, 30 μm diameter spots have been reported with a complicated optical setup including a coaxially lined IR objective lens [18]. Spot sizes of 10–20 μm have been achieved in LAESI using transmission geometry and an objective lens with a high numerical aperture [19]. This approach has been used for analysis of adherent mammalian cells in situ. Another approach to reach better resolution without the need for matrix is to use a shorter pulse length than in typically applied nanosecond lasers. Lateral resolution of 100 μm has been achieved using picosecond IR laser pulses [20]. Such lasers, however, are very expensive and available to few researchers. Furthermore, using a 25 μm pinhole to filter the beam of a UV laser has been reported to lead to 5 μm laser spot size in MALDI imaging [21], and would possibly function also with an IR laser.

In this contribution, we present an easy and inexpensive method for improving the ablation spot size in IR laser ablation to at least 44 μm by simply increasing the length of the laser beam path. This method will be useful for researchers working with different IR-laser based ambient mass spectrometry imaging methods.

Experimental

All animal experiments were approved by the National Animal Experiment Board of Finland (License number ESAVI/11198/04.10.07/2014). Mice from a triple mixed genetic background (129Ola/ICR/C57bl6) were anesthetized with an overdose of sodium pentobarbital (100 mg/kg, i.p.). Anesthetized animals were transcardially perfused with warm phosphate buffered saline (PBS) for 4 min followed by 4% paraformaldehyde (PFA) in 0.1 M phosphate buffer, pH 7.4, for at least 5 min. The brains were removed and postfixed in the same PFA solution overnight in room temperature and then stored in phosphate buffer containing 20% sucrose at 4 °C. Coronal striatal sections were cut into 30 μm thickness with a cryomicrotome (Leica CM 3050; Leica Microsystems Nussloch GmbH, Nussloch, Germany). The sections were saved in a serial order at –20 °C in cryoprotectant buffer containing glycerol and ethylene glycol. Before the MS analysis, the sections were flushed with Milli-Q water (Merck Millipore, Molsheim, France).

For laser energy and spot size measurements, a mid-IR laser beam was delivered on the sample surface using a gold-coated mirror (PF10-03-M01; Thorlabs, Newton, MA, USA) and focused by an anti-reflection coated 100 mm focal length plano-convex CaF2 lens (LA5817-E, Thorlabs). The mirror and the lens both had a diameter of 1 inch. The laser was an IR Opolette HE 2940 fixed wavelength and flashlamp pumped Nd:YAG laser operating at 2.94 μm wavelength and 20 Hz repetition rate (Opotek, Carlsbad, CA, USA). The laser energy was set to 6 mJ/pulse. The ablation holes were created at 1–18 m laser aperture to lens distances, the beam was focused with the lens, and the spots were measured using a Leica DMC2900 camera module attached to a Leica DM6000 B research microscope with 2.5×–5× objectives (Wezlar, Germany). The reported ablation spot diameters are the averages of maximum and minimum Feret diameters of the spots created with a specific distance. The laser energy was measured using a thermal power sensor (Thorlabs S310C, d = 20 mm), which was connected to an energy meter (Thorlabs PM100D). The sensor was first attached to the laser output aperture for measurement of the maximum output, after which the laser energy was measured at distances of 1–18 m. The measurements were carried out in normal laboratory lighting, and therefore the background energy was always measured first.

The mass spectra for the MS images were collected using a Bruker MicrOTOF mass spectrometer (Bremen, Germany). The commercial ion source of the mass spectrometer was replaced by a self-built LAAPPI source, similar to the one described in literature [5, 6]. The mass spectrometer was equipped with a modified extension capillary. An aluminum shield was attached to the outer part of the extension capillary to direct the drying gas to the sides, away from the ionization area. The drying gas temperature was 250 °C, and the gas was supplied at 5 L/min. The capillary exit and skimmers 1 and 2 voltages were set to 240, 80, and 23 V, respectively. The data acquisition time was set to 1 Hz, and the dwell-time was 35 s. The ablation plume was intercepted by a hot acetone (Sigma-Aldrich, HPLC grade [99.8%], Steinheim, Germany) vapor jet, which was directed toward the inlet of the mass spectrometer. The jet was produced using an all-glass heated nebulizer microchip, described previously in detail [22]. The liquid solvent (acetone) was introduced into the heated nebulizer microchip at 10 μL/min using an Agilent 1100 capillary LC pump (Waldbronn, Germany), and vaporized with the aid of nitrogen gas flow (100 mL/min) and high temperature (4.0 W heating power producing ~350 °C jet temperature at 5 mm of the chip nozzle [23]). The tissue sections were placed on a microscope glass slide mounted on a Peltier cooling stage and positioned in front of the mass spectrometer ~3 mm below the inlet orifice. The temperature of the Peltier stage was kept at 5 °C in order to minimize dehydration and to optimize the amount of water for the ablation process. The sample plume and solvent jet were irradiated in the ambient air with 10.0 and 10.6 eV photons produced by a krypton discharge vacuum ultraviolet (VUV) photoionization lamp (PKR 100; Heraeus Noblelight, Cambridge, UK). The tissue sample was rastered by moving the cooled stage in the xy-plane using two computer-controlled motorized 100 mm linear translation stages (NRT100/M; Thorlabs). The mass spectra were collected using Compass 1.3 for micrOTOF (Bruker). The XY stage and the IR laser were controlled by a self-written MATLAB-based program. Time and stage position data were recorded during the imaging experiment. This information was later used to split the measurement file into pixel files. Each pixel was created by selecting the measurement time interval of the corresponding pixel, and forming an average mass spectrum from the collected spectra. The measurement file was converted to mz5 file format [24] using ProteoWizard’s msconvert tool [25] before processing the data with MATLAB. Heatmaps were created with a MATLAB based MSiReader software [26].

Results and Discussion

The spatial resolution of laser ablation was observed to improve significantly by increasing the distance of the laser aperture from the focusing lens. Figure 1 shows the ablation holes caused by 20 IR laser shots (20 Hz) on 30 μm thick mouse brain tissue at distances of 1–18 m between the laser aperture and the focusing lens. While at 1 m distance the average Feret diameter of the spots was 440 μm, at 14 m, an average spot diameter of 55 μm was achieved. We suggest that this phenomenon is caused by the divergence of photons within the noncollimated, multimodal laser beam. According to the laser manufacturer (OPOTEK), the laser beam has a divergence of 10 mrad in the x-axis and 5 mrad in the y-axis; the beam profile is shown in Supplementary Figure S1 in the Supporting Material. At short distance, the focusing lens collects most of the photons within the laser beam, and thus the aberration is larger since the angles at which the photons arrive to the lens have a wide distribution (Fig. 2). As the distance to the focusing lens increases, only the most collimated center of the laser beam is collected, while the diverted photons are lost. Since the lens now focuses more uniform light, the aberration and the spot size are reduced.

Figure 1
figure 1

IR laser ablation holes created on a 30 μm thick mouse tissue at 1–18 m distance between the laser aperture and the focusing lens. Note: a Peltier cooling stage was not used in this measurement series

Figure 2
figure 2

At short distance, photons with different flight paths are efficiently captured by the focusing lens and, as a consequence, the resulting ablation area is large (left). At long distance, diverted photons fail to hit the focusing lens, and the photons hit the lens in a similar angle. Thus, the resulting ablation area is very small (right)

Figure 3 shows the experimentally measured spot diameters, laser energies, and fluencies at 1–18 m laser aperture-focusing lens distances (Supplementary Tables S1S3 show the measured and calculated values). The spot size decreases first exponentially, but then levels off, probably since most of the diverted photons are already lost at the first meters. Supplementary Figure S2 shows the trendlines calculated based on the experimentally defined values for the spot and hole diameters and the laser beam travel distance. A good correlation (R2 = 0.98) between the fitted curve and measured values is observed in both cases.

Figure 3
figure 3

Effect of the distance between the laser aperture and the focusing lens on the average Feret diameters of the ablation spots and holes, the energy reaching the focusing lens, and the fluence on target. (Note: the presented values for laser energy and fluence are based on measurements with 20 mm o.d. thermal power sensor and are thus underestimations of the amount of photons arriving at the lens with 25.4 mm o.d.)

Decreasing the spot size below ~40 μm becomes impractical due to space limitations, although very long optical paths can be achieved with suitably tilted mirrors, between which the laser beam is reflected several times. An example of such a setup (also used in the MS imaging experiment described below) is shown in Supplementary Figure S3. Of course, several reflections reduce the laser energy due to imperfect transmission. However, for a MS imaging experiment, 50 μm spot diameter is usually a practical compromise between the amount of structural information and measurement time, and with the current setup it can be reached with 14 m distance (or less, if a cooling stage is used, see discussion below).

As the distance that the laser beam needs to travel increases, a smaller proportion of photons will finally reach the focusing lens and the sample, and less energy will be available for the ablation process. The plot of laser energy reaching the focusing lens (mJ) follows largely a similar curve as the spot size (Fig. 3), as shown by thermal power sensor measurements. However, since the laser beam can now be focused to a smaller area, the laser fluence (J/cm2) is in the same order of magnitude with all the distances, and efficient ablation is achieved even at the longest laser distances. Some fluctuations are observed in the laser fluence (Fig. 3). This is partly explained by the uncertainty of the measurement of the irregularly formed spot diameters. Further, as shown in Supplementary Figure S1, the multimodal laser beam contains intensity variations, so-called “hot spots”, and the most intense part of the laser is not in the middle of the beam. As a result, increasing the beam travel distance also changes the intensity profile of the focused beam, which in turn leads to small intensity and fluence variations across different aperture-focusing lens distances. Moreover, at fixed optical set-up, the produced ablation spots were quite uniform as presented in Fig. 4a, which shows a microscope image of an array of ablated spots on the analyzed tissue area.

Figure 4
figure 4

(a) Partial microscope picture of the imaged area showing the spots created with laser ablation using a 60 μm step size. The measured average ablation spot and hole diameters were 55 μm and 30 μm, respectively. (b) Partial total ion chromatogram showing the typical signal intensities achieved from the ablation spot in the imaging experiment with LAAPPI. (c) A single mass spectrum obtained in the MS imaging experiment with LAAPPI. (d) Distributions of ions at m/z = 264.28 and m/z = 401.35 measured from a 1.56 × 3.54 mm size area of a 30 μm thick mouse brain tissue with LAAPPI-MS using a 9 m laser beam-focusing lens distance and a Peltier cooling stage. The ion intensities were scaled against the total ion current. (e) Microscopic image showing the rastered tissue area

To demonstrate the utility of the small ablation spot sizes in MS imaging of a clinical sample, a rectangular area on a coronal mouse brain tissue slice close to the bregma line [27] was imaged using IR laser ablation with 9 m laser beam travel distance resulting in 55 μm spot size, and LAAPPI. Microscopic image of the ablated spots, total ion chromatogram and mass spectrum obtained from the tissue, as well as MS images of two ions are shown in Fig. 4a–e. A slightly smaller spot size with this distance was achieved here than in the experiment shown in Figs. 1 and 3 because here the sample was cooled using a Peltier stage, which is thought to reduce heat propagation in the sample. Beam instability or vibrations were not observed during the measurement, regardless of the long laser beam travel distance (Fig. 4a). As an example, Fig. 4d shows the heatmaps of two ions with highly distinct distributions. The ion at m/z 264.28, which is highly abundant in the corpus callosum (CC), dorsal striatum (dSTR), and septum areas (Fig. 4e), is suggested to be a C18H34N+ fragment of sphingosines. The ion at 401.35, on the other hand, which shows an intense signal from the cortex (CTX) area and around the region of lateral ventricle, is suggested to be oxidized cholesterol ([M – H + O]+), or the protonated molecule of ketocholesterol [5]. Furthermore, very intense signals compared with the background were obtained from the ablated spots, as shown in the total ion chromatogram in Fig. 4b, and therefore it is proposed that the signals of even smaller spots are measurable with the current mass spectrometric setup.

Conclusions

A significant improvement to the spatial resolution in IR laser ablation was achieved by simply increasing the distance between the IR laser and the focusing lens. The setup is easy to reproduce, and it does not require expensive optics or fragile fibers. Even 44 μm spot size was achieved, which is sufficiently small for most biological imaging applications. Even smaller spot sizes are achievable using a cooling stage under the sample, as was shown in the MS imaging experiment of a tissue sample. The introduced technique for improving the spatial resolution can be utilized in, e.g., LAESI, LAAPPI, IR-LAMICI, and MALDESI. Current work in our laboratory is focused on MS imaging by LAAPPI and LAESI with high spatial resolution, and a more detailed description of the imaging experiments with whole tissue heatmaps will be published in upcoming work.