Abstract
Background
Intra-operative breakage of laparoscopy instruments is a rare occurrence. It entails a risk of the foreign body being retained in the abdomen and may lead to medico-legal implications. The foreign body migrates due to bowel movements. Therefore, the retrieval of such foreign bodies can be challenging even for a highly skilled and experienced surgeon. Fluoroscopy can guide in locating the missing foreign body. The only preventive measure is the vigilance of the operating team and thorough inspection of all the instruments prior to completion of surgery.
Case Report
We present a case report of a 50-year-old lady who underwent total laparoscopic hysterectomy with bilateral salpingoopherectomy at a private nursing home. During the closure of the vault, the tip of the needle holder broke inside the abdomen and the primary surgeon was unable to retrieve it. The missing tip of the needle holder was removed from the right hypochondriac region, by a second laparoscopic surgery by an advanced laparoscopic surgeon under fluoroscopic guidance.
Conclusion
Intra-operative breakage of laparoscopy instruments can occur, demanding a great deal of expertise for the retrieval of such foreign bodies. Fluoroscopy can guide in locating the missing foreign body which can be retrieved by an expert laparoscopic surgeon.
Similar content being viewed by others
References
Katara AN, Bhandarkar DS, Shah RS, Udwadia TE. Breakage of fascial closure device during laparoscopic surgery. J Mini Access Surg. 2005;1(2):79.
Sang AX, Lal R, August A, Danzer E, Buckingham B, Mueller CM. Fluoroscopic-assisted laparoscopic retrieval of retained glucose sensor wire from the omentum. Clin Case Rep. 2019;7(9):1717–20.
Rani BD, Jarugulla S. Retained surgical instrument in minimally invasive surgery. J Obstet Gynecol India. 2019; 1–2.
Lynch CM, Powers AK. Management of a broken needle at the time of laparoscopic burch. JSLS J Soc Laparoendosc Surg. 2000;4(3):275.
Duhan N, Malik R, Hooda MR. A broken cautery tip lost in the abdomen at laparoscopic ovarian cystectomy. Open J Clin Med Case Rep. 2016;2(2):1067.
Kandioler-Eckersberger D, Niederle B, Herbst F, Wenzl E. A magnetic probe to retrieve broken metallic parts of instruments during laparoscopic procedures. Surg Endos Other Interv Tech. 2002;16(1):208–9.
Yasuhara H, Fukatsu K, Komatsu T, Murakoshi S, Saito Y, Uetera Y. Occult risk of broken instruments for endoscopy-assisted surgery. World J Surg. 2014;38(11):3015–22.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interests.
Informed Consent
Patient’s identity is not revealed as only X-ray and fluroscopic images are being sent for publication.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Sanjay Brahmbhatt M.D., Amrita Makhija M.S., Jayna Brahmbhatt M.D., Yagnesh V Patel D.N.B.: Department of Gynecological Endoscopy, Himalaya Hospital, Anand, India.
Rights and permissions
About this article
Cite this article
Brahmbhatt, S., Makhija, A., Brahmbhatt, J. et al. Intra-abdominal Breakage of Laparoscopic Needle Holder Tip and Its Retrieval Under Fluoroscopic Guidance. J Obstet Gynecol India 71, 452–455 (2021). https://doi.org/10.1007/s13224-020-01394-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13224-020-01394-6