1 Introduction

Thyroid cancer has been shown responsible for 0.2—0.5% of cancer-related deaths. In particular, well-differentiated thyroid cancer (DTC) is considered the most common type making up around 85–95% of all thyroid cancers [1, 2]. In DTC management, total thyroidectomy is deemed the curative-intent therapy of DTC followed by adjuvant radioiodine therapy (RIT) to eradicate occult microscopic thyroid foci, extra-thyroidal extensions, involvement of lymph nodes, and distant metastases [3]. 131I is a β- emitting radionuclide that decays to stable xenon-131 with a physical half-life of 8.02 days, a principal γ-ray of 364 keV, maximum β- energy of 0.61 MeV, and linear energy transfer of 0.25 keV/μm [4]. The therapeutic activity is often a standardized activity chosen empirically with less focus on personalized dosimetry. After oral administration, radioiodine systemically circulates through the entire body with mounting uptake in thyroid residues and iodine avid lesion. In the meantime, well-perfused organs such as red marrow and kidneys are exposed to the risk of radiation-induced malfunctions [5, 6].

Furthermore, the main salivary glands (parotid, submandibular, and sublingual) secret about 90% of the saliva in the oral cavity, while only 5% is released from minor lingual glands [7]. In RIT, salivary glands concentrate radioiodine 7–700 times the serum level due to overexpression of sodium/iodide symporter (NIS). This status leads to the most prevalent side effect (sialadenitis) occurring in 10–60% of the treated patients with RI in addition to sensation of a dry mouth (xerostomia), and an increased risk of oral infections and dental caries [8,9,10,11,12]. It was also found that 20% of salivary glands became dysfunctional after 5 years of a single 131I dose and exposure to about 45 Gy may cause significant damage to salivary glands [13,14,15]. So far, lemon sucking and chewing gum are routinely used to reduce the iodine residence in salivary glands. More recently, phenylephrine has been shown effective to alleviate 131I radiation damage by maintaining mitochondrial homeostasis [16].

On the other hand, the finite resolution of the nuclear imaging systems constituted an obstacle to providing exemplary dose calculation to lesions. Thus, the trend was directed to administer maximum permissible activity limited by the dose boundaries of the organs at risk (e.g. red marrow in RIT) [17]. Presently, lesions detection and physical modeling of the image have been improved with the advent of hybrid technology (SPECT/CT and PET/CT) [18, 19]. Nevertheless, obtaining reliable dose estimates after administration of large 131I activity remains challenging due to high radiation flux and count loss. Alternatively, quantitative imaging of iodine tracer was vastly utilized for dose estimation prior to therapy. However, there is a lack of understanding of the consistency between the dose measures from the tracer and those actually received after therapy in addition to the vulnerability of 131I bio-kinetics to change owing to the so-called ‘stunning effect’ [20, 21].

In light of that, the current study aimed to verify the absorbed doses to salivary glands and pathologic lesions after radioiodine therapy using patients’ SPECT/CT scans and earlier phantom investigations in an attempt to resolve dosimetry degrading factors practically.

2 Materials and methods

2.1 Calibration experiment

Acrylic glass body phantom (PET Phantom NEMA 2012 / IEC 2008) was used for the system calibration. A total of 1983.2 MBq 131I was injected into the whole phantom. Thereafter, 20 SPECT/CT scans (≈ 2 scans/ week) were subsequently acquired with Symbia™ T scanner (SIEMENS) until the activity decayed to a neglected counts loss [< 30 MBq]. The acquisition settings were selected as step and shoot mood, 64 projections/π, 25 s/projection, 15% photo-peak window at 364 keV with ± 7% scatter windows, and 128 × 128 matrix size. OSEM algorithm was used for image reconstruction with 8 subsets and different numbers of iterations (10, 20, 30, and 40), followed by low pass Gaussian filtering-9 mm. Regions of interest (ROIs) were drawn over the phantom, middle cylinder, and the whole frame. The counts/activity conversion factor was generated based on the CT scan and 30% is-contour. For count loss correction, a correction factor was derived from the ratio between the measured and true counts of the phantom for every single SPECT/CT scan. Then, a set of mathematical equations were derived for count loss correction to be applied on the patients scans [22].

For partial volume effect correction (PVC), the spheres of 26 cm3, 11.5 cm3, 5.6 cm3, 2.6 cm3, and 0.5 cm3 volume were filled with 4:1 activity to background ratio at neglected dead time. ROIs were delineated manually on the CT images and also automatic 30%-isocontour for different quantification methods. Afterward, the measured activity was divided by the true activity mount for obtaining recovery coefficients (RCs) to be later used in lesions dose estimation.

2.2 Counts loss correction

As expected, the measured counts increased non-linearly with larger 131I activity due to machine dead time. Accordingly, a set of bi-compartmental equations combining the whole frame counts and the correction factors were generated for counts loss correction, as follows:

  • - whole frame counts ≥ 47 kcps,

    $$\begin{array}{cc}\mathrm{Y}=0.0009{\mathrm{X}}^{3}-0.1039{\mathrm{X}}^{2}+3.9189\mathrm{X}-13.309& \mathrm{R}-\mathrm{squared}=0.98\end{array}$$
    (1)
  • -whole frame counts ≤ 47 kcps,

    $$\begin{array}{cc}\mathrm{Y}=5670.9{\mathrm{X}}^{-1.33}& \mathrm{R}-\mathrm{squared}=0.93\end{array}$$
    (2)

    y: correction factor, X: whole frame counts (kcps).

At neglected dead time, two conversion factors (counts/MBq) were generated; the first was CT- based, and the latter was based on automated iso-thresholding over SPECT images. The CT-based conversion factor was slightly higher than 30% iso-contour in order of 4% at 10it. × 8subsets, and 1% at 40it. × 8subsets.

2.3 Partial volume effect (PVE)

Recovery coefficients (RCs) were generated from the phantom study to account for the activity underestimation in small objects. The obtained RCs were varied according to the measured volumes from CT and 30%-isocontour methods as demonstarted in Fig. 1. The impact of increasing iterations numbers on the quantification accuracy was shown in Fig. 2. It was revealed that the recovery coefficients improved relatively with larger iterations. Thus, the SPECT/CT scans reconstruction was made by 40it. × 8sub and 9-mm Gaussian filtering for all patients. And the numerical functions shown below were used;

Fig. 1
figure 1

Volumes versus recovery coefficients from CT scans (A) and 30% iso-threshold (B) with 40 iterations and 8 subsets

Fig. 2
figure 2

Recovery coefficients improvement with larger iterations number over A) CT scan, and B) SPECT with 30%-iso contour

for PVE correction in salivary glands:

$$\begin{array}{cc}\mathrm{RC}=0.1726\mathrm{ln}\left(\mathrm{v}\right)+0.3208& \mathrm{R}-\mathrm{squared}=0.97\end{array}$$
(3)

for PVE correction in thyroid residues and lesions:

$$\begin{array}{cc}\mathrm{RC}=0.5583\times {\mathrm{v}}^{0.0989}& \mathrm{R}-\mathrm{squared}=0.96\end{array}$$
(4)

RC: recovery coefficient, v: volume (cm3).

The real volumes of the spheres were well-correlated with those estimated by both CT and 30%-iso contour, as illustrated in Fig. 3. Accordingly, 30% iso-thresholding was employed to measure the volume and counts of thyroid remnants and distal lesions in the patients’ study.

Fig. 3
figure 3

The consistency between the actual volumes (cm3) of the spheres and those measured by CT and SPECT with 30%-isocontour

2.4 Clinical study

25 DTC patients (n = 25; 6 males and 19 females) undergoing radioiodine therapy were accepted to join this study. Informed consent was received from all the participant patients according to the ethical review board at Cerrahpasa Medical School. The activity to administer was empirically prescribed (by the referring physicians) and broadly variable based on the clinical case. A range of 3497 to 9250 MBq (median: 5550 MBq) was administered to the participant patients. The median weight was 72 kg (range: 45–114 kg) and the median age was 44 years (range: 19–71). Following oral 131I administration, SPECT/CT scans were subsequently acquired at 4 ± 2, 24 ± 2, and 168 ± 3 h with the settings summarized in Table1.

Table 1 SPECT/CT settings used in the patients’ study

The participant patients were prepared by thyroid hormone withdrawal (≈ 4 weeks) and the TSH level was above 30 mIU/L. The routinely followed protocol implies lemon stimulation and chewing gum to reduce the salivary glands’ uptake.

CT images were used to draw ROIs over the salivary glands, while automated 30% iso-contour was found superior for thyroid residues and distant lesions quantification. The partial volume effect (PVE) and counts loss were corrected using the equations derived from the calibration experiment (Eqs. 14). The Uptake values (%) of the targets were calculated at the first and second measurements in proportion to the administered activity. The activity in the target organ Ai was estimated at a given time by:

$${\mathrm{A}}_{\mathrm{i}}=\left[\mathrm{C}\times \left(1+\mathrm{CLCF}\right)\right]/\mathrm{RC}\times \mathrm{CF}$$
(5)

C: scatter corrected counts, CLCF: Counts loss correction factor, RC: Recovery coefficient, CF: calibration factor.

The total number of disintegrations (Ã) in the salivary glands was estimated from the mono-exponential fitting of the time-activity plots, while the trapezoidal rule was used to generate à in the lesions. The time-integrated activity coefficient (ā) was then calculated by:

$$\overline{\mathrm{a} }=\widetilde{\mathrm{A}}/{\mathrm{A}}_{0}$$
(6)

Ao: administered activity, ã: time-integrated activity coefficient (TIAC).

The absorbed dose to salivary glands was calculated by IDAC-Dose 2.1 program. This code is equipped with specific absorbed dose fractions [reported by the International Commission on Radiological Protection (ICRP)] and a computational framework of internal dose assessment given for reference adults in ICRP Publication 133 [23]. However, the absorbed dose to thyroid residues/ lesions was calculated by the unit sphere density model built in the IDAC-Dose 2.1 program. The dose fractions from radioiodine activity for spherical volumes, derived from Monte Carlo simulation methods, have been reported earlier [24].

2.5 Statistical analysis

Mann–Whitney test was used to compare the absorbed doses to submandibular and parotid glands via IBM SPSS Statistics 20.

3 Results

3.1 Patients study

The absorbed dose to submandibular glands, parotid glands, thyroid residues, and distal metastases was estimated for 25 patients undergoing radioiodine therapy as reported in table 2. The median volume of the right parotid glands was 26.7 cm3 (range: 13.0–70.0 cm3), and 26.2 cm3 (range: 15.2–62.0 cm3) for the left parotid glands. The median volumes of the right and left submandibular glands were 12.5 cm3 (range: 8.00–22.5 cm3) and 12.0 cm3 (range: 7.00–21.6 cm3), respectively. Furthermore, 22 thyroid residues were visualized on SPECT/CT images with a median volume of 5.42 cm3 (range: 2.82–12.70 cm3) and 10 pathologic lesions with a median volume of 5.95 cm3 (range: 2.16–11.50 cm3). It was difficult to identify such a lesion or thyroid foci in patient number 12 on SPECT/CT.

Table 2 The measured volumes (cm3) of parotid glands and submandibular glands, thyroid residues, and metastases

As summarized in Table 3, the median 131I uptake (%) in the thyroid residues at 4 ± 2 h was 0.82 (range: 0.01–10.9) and gradually increased to 1.20 (range: 0.04–18.3) after 24 ± 2 h. The median uptake in the right parotid glands was 0.18 (range: 0.03–0.57) at 4 ± 2 h and 0.13 (range: 0.002–0.47) at 24 ± 2 h. The median uptake in the left parotid glands was 0.18 (range: 0.02–0.54) at 4 ± 2 h and 0.12 (range: 0.003–0.44) at 24 ± 2 h. Whereas, lower median uptake was manifested in the submandibular glands as 0.10 (range: 0.04–0.21) on the left-sided and 0.07 (range: 0.002–0.15) on the right-sided at 4 ± 2 and 24 ± 2 h, respectively.

Table 3 Uptake values (%) of the submandibular glands, parotid glands, thyroid residues, and metastases

Additionally, the interpatient variability in the 131I kinetics was comprehended from the variation in the time-integrated activity coefficients (TAICs). The median TIAC in the left parotid glands was 5.21 min (range: 0.19–25.4) and 4.45 min (range: 0.21–20.4) in the right parotid glands. However, right and left submandibular glands showed lower TIAC values with a median of 2.62 (0.70–8.95) and 2.59 min (range: 1.10–12.2), respectively.

The present work demonstrated that the absorbed dose per unit administered activity to parotid glands varied from 0.04 to 0.97 Gy/GBq with a median of 0.26 Gy/GBq, and the absorbed dose to submandibular glands was lower with a median of 0.14 Gy/GBq ranging from 0.05 to 0.56 Gy/GBq, as illustrated in Fig. 4. A significant difference (Pvalue < 0.05) was demonstrated between the absorbed doses to parotid and salivary glands.

Fig. 4
figure 4

whisker plot of the absorbed dose values for parotid and submandibular glands

On the other hand, the absorbed dose to thyroid remnants varied among the patients as revealed in Table 4. The absorbed dose to thyroid remnants ranged from 0.55 to 399.5 Gy/GBq (median: 21.8 Gy/GBq), and the distal lesions likewise exhibited a broad dose range of 0.78—28.0 Gy/GBq (median: 3.12 Gy/GBq). It was also deduced that only 41% of the thyroid residues received absorbed dose > 80 Gy and 18% between 70–80 Gy following administration of 3497–9250 MBq (median:150 MBq) 131I. Whereas 18% of the patients received doses amidst 40–70 Gy, and the remaining 23% showed absorbed doses less than 40 Gy. In contrast, 18% of the metastases had absorbed doses above 80 Gy, 9% between 40–60 Gy, and the vast majority (64%) exhibited absorbed doses less than 40 Gy.

Table 4 Absorbed dose (Gy/GBq) to thyroid residues and metastases

4 Discussion

Radioiodine therapy is deemed an essential standard of care for DTC with a favorable impact on overall survival [25, 26]. In metastatic DTC patients, the survival rate was significantly decreased and larger activity administration is more demanded [27, 28]. Commonly, a single dose of 7.4 GBq is applied for metastatic DTC, while higher activity (up to 9 GBq) was administered for advanced DTC therapy [27, 28]. In 2015, less intense treatment strategies for many DTC patients including observation or thyroid lobectomy without RIT were recommended by the American Thyroid Association (ATA) making controversial and significant differences in clinical practice [29, 30]. More recently, a joint statement has been released (in 2019) by leading nuclear medicine organizations involving the ATA, European Association of Nuclear Medicine (EANM), Society of Nuclear Medicine and Molecular Imaging (SNMMI), and European Thyroid Association (ETA) that defined the indications and the optimal prescribed activity in RIT by reporting a set of collective principles (Martinique Principles) [31, 32]. Overall, the administered RAI activity can be determined empirically using a "one-dose-fits-all" regime or based on a dosimetric approach. The suggested 131I activities for postoperative ablation/adjuvant RIT were outlined as 1.1–2.2 GBq in low-risk DTC (< 2 cm), 2.2–3.7 GBq in low-risk (> 2 cm), and intermediate-risk DTC, and ≥ 3.7 GBq in high-risk DTC [31]. The empirical approach is simple and effective, however, significant adverse reactions may arise like reduced bone marrow reserve, lung function impairment, and progressive loss of efficacy after repeated treatment [31]. In contrast, the latter dosimetric approach is more complicated and provides patient-tailored activities with several advantages of high success rates in a single treatment, avoiding suboptimal or excessive 131I administration, the possibility to plan the activity to the target volume, and reducing radiation toxicity [31, 33,34,35]. The integrated CT in hybrid SPECT/CT was a key element for resolving dosimetry problems and improving quantitative accuracy, however, a few inherited uncertainties in dose calculation are still matters of concern [36, 37]. Though it is challenging to confirm a preference between the empiric or dosimetric approach for RIT, the latter strategy may be more effective in subjects with critical conditions (e.g., renal failure, diffuse lung metastases, and reduced bone marrow reserve) [31].

On the other hand, long-term sequelae and elevation of morbidity rates are associated with 131I treatment as well as salivary glands deterioration. Unlike tracer-based dosimetry, this study sought to provide clinically validated method for post-therapeutic dose estimates. From the technical side, the calibration tests with IEC PET phantom and SPECT/CT were made to improve the dose calculation accuracy. The triple energy windows method was utilized for scatter correction with a feasible compromise between accuracy and noise as recommended in past 131I studies [38, 39]. For PVE correction, recovery coefficients were derived to compensate for the activity underestimation in anatomic structures (e.g. salivary glands and lesions) approximated by spherical shape. However, this method might not be a typical model due to interference of other factors such as shape, signal-to-noise ratio, and position in the image. Alternatively, the geometric transfer matrix has been shown effective to improve SPECT quantitative accuracy by using the point spread function to calculate regional contribution fractions [40, 41]. In addition, the accuracy of the area under time-activity curve is associated with several factors like the number of samples, sampling time, and the fitting function [41]. Unfortunately, it was impossible to increase the number of SPECT/CT acquisitions in this study because of the global covid-19 pandemic.

From the clinical standpoint, earlier salivary glands assessment and follow-up showed a wide absorbed dose range from 0.03 to14 Gy/ GBq [42]. In another report, the absorbed dose range to a single parotid and submandibular gland was found 0.10–0.30 Gy/GBq and 0.20–1.20 Gy/GBq, respectively [43]. However, these studies like others were based on serial planar images with several estimation errors related to volume determination, background ROIs, and scatter contribution. Exceptionally, Jentzen and co-authors estimated the absorbed dose to submandibular glands (0.35 ± 0.14 Gy/GBq) and parotid glands (0.33 ± 0.09 Gy/GBq) using 124I PET/CT under lemon-juice stimulation [44]. In contrast, the current study demonstrated that the absorbed dose per unit administered activity to parotid glands varied from 0.04 to 0.97 Gy/GBq with a median of 0.26 Gy/GBq. Whereas, the median dose to submandibular glands was 0.14 Gy/GBq ranging from 0.05 to 0.56 Gy/GBq. The agreement between the current findings obtained after 131I therapy and the aforementioned 124I PET/CT study ascertain the accuracy of the implemented method for post-therapeutic dose estimation.

Clinically, the computed absorbed doses per administered activity are indeed small to cause significant damage to salivary glands. However, the oversensitivity to low radiation levels is attributed to the unique combination of mucous and serous acinar cells in the salivary glands [42]. In radiotherapy, a complete loss of salivary glands function has been shown at 60 Gy. However, the dysfunction of salivary glands occurs in a dose-dependent manner after radioiodine therapy [4244]. For example, a single dose of 6 GBq resulted in a partial function loss of 30%, while cumulative administration of 35 GBq caused a complete glandular failure [42]. Moreover, mild dysfunction was observed in 60% of patients following RIT with 5143–5550 MBq including decreased uptake and delayed excretion in parotid glands. A similar degree of dysfunction was also manifested in the submandibular glands for 45% of patients. In addition, moderate to severe dysfunction and decreased uptake were noted with more iodine therapies [45,46,47]. It has been recently supported that the mean parotid dose should be kept below 26 Gy for the preservation of salivary gland function in radiotherapy and a mean dose < 20 Gy might increase the local recurrence risk for head and neck cancer [48]. On the basis of our findings, a cumulative activity of 27 GBq is probably adequate to deliver the preservation dose (26 Gy) of the parotid gland, and 46 GBq is expected to induce parotid glands failure according to the parotid glands dose of 0.97 Gy/GBq and maximum tolerated dose of 45 Gy.

In lesion dosimetry, the present work exhibited notably disparate dose values for thyroid remnants (range: 0.5–399.5 Gy/GBq) and metastases (range: 0.78–15 Gy/GBq). In contrast, previous 131I tracer-based data demonstrated variable absorbed doses to thyroid remnants and metastases ranging from 5 to 1000 Gy [49,50,51]. This consistency represents further proof of the accuracy of the dosimetry method used here.

Moreover, it was revealed that only 41% of the thyroid residues received an absorbed dose > 80 Gy, 18% had an absorbed dose between 70–80 Gy, 18% were with a dose range of 40–70 Gy, and 23% received absorbed dose below 40 Gy following administration of 5402 ± 1665 MBq131I. Whereas, 18% of the distal lesions exhibited absorbed doses above 80 Gy and 9% between 40–60 Gy, while the vast majority of lesions (73%) received doses less than 40 Gy. In dosimetry-oriented therapy, thyroid ablation was successfully achieved with absorbed doses > 49 Gy, 90 Gy, and 300 Gy, while partial to complete response was reported for distal metastases after receiving doses higher than 40 Gy to 80 Gy [17, 51].

Herein, the safety and efficacy of 131I therapy are exquisitely favorable outcomes for DTC patients that must be achieved with personalized data and well-validated dosimetry methods. Prospectively, this method might be feasibly used for establishing a dose–effect relationship relying on reliable dose estimates after RI therapy.

5 Conclusion

Dose estimation to salivary glands and pathologic lesions, identified by SPECT/CT, has been shown to be feasible following RIT, together with good agreement with 131I-tracer and 124I-PET/CT-based studies. A significant salivary glands uptake has been shown in the early hours that may affect the accuracy of the measured thyroid uptake by the standard probe. A broad and sub-effective dose range was estimated for thyroid residues and distal lesions. The feasibility of the current dosimetry method may be useful to establish a dose–effect relationship in RIT based on post-therapeutic dose estimates.