Visual patterns of breast attenuation artifacts in women and men with an upright and supine cadmiun-zinc-telluride camera

  • Original Article
  • Published:
Journal of Nuclear Cardiology Aims and scope

Abstract

Background

Breast attenuation artifacts occurring with upright cadmium-zinc-telluride (CZT) cardiac imaging systems have not been well characterized.

Methods

216 consecutive patients with Single Photon Emission Computerized Tomography myocardial perfusion imaging and no angiographically significant obstructive coronary artery disease were identified. All upright and supine SPECT images as well as coronary angiograms were reviewed and analyzed in blinded fashion.

Results

In women imaged upright, more visual false positive defects were noted in the inferior wall compared to the anterior wall (26 vs. 10 at rest, p = 0.006, and 33 vs. 13 at stress, p < 0.001). Visual inferior wall defects were more common in the upright than supine position at stress (33 vs. 23, p = 0.018) and rest (26 vs. 14, p = 0.011), and most apparent in non-obese women (13 vs. 8, at stress, p = 0.059 and 11 vs. 5, at rest, p = 0.014).

Conclusions

With upright CZT myocardial perfusion imaging, women often have visible inferior wall attenuation artifact defects, likely from pendant breast tissue. These inferior wall attenuation artifacts may be seen in non-obese female patients.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Figure 1
Figure 2
Figure 3

Similar content being viewed by others

Abbreviations

BMI:

Body mass index

CZT:

Cadmium-zinc-telluride

LVEF:

Left ventricular ejection fraction

MPI:

Myocardial perfusion imaging

SPECT:

Single photon emission computerized tomography

SRS:

Summed rest score

SSS:

Summed stress score

TID:

Transient ischemic dilation

References

  1. DePuey EG, Rozanski A. Using gated technetium-99m-sestamibi SPECT to characterize fixed myocardial defects as infarct or artifact. J Nucl Med 1995;36:1995.

    Google Scholar 

  2. DePuey EG, Morley J, Leykekhman A. Shifting breast attenuation artifact mimicking ischemia: Resolved with attenuation correction. J Nucl Cardiol 2019;26:2142-7.

    Article  Google Scholar 

  3. Ben-Haim S, Almukhailed O, Neill J, Slomka P, Allie R, Shiti D, et al. Clinical value of supine and upright myocardial perfusion imaging in obese patients using the D-SPECT camera. J Nucl Cardiol 2014;21:478-85.

    Article  Google Scholar 

  4. Gambhir SS, Berman DS, Ziffer J, Nagler M, Sandler M, Patton J, et al. A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med 2009;50:635-43.

    Article  Google Scholar 

  5. Gimelli A, Bottai M, Genovesi D, Giorgetti A, Di Martino F, Marzullo P. High diagnostic accuracy of low-dose gated SPECT with solid-state ultrafast detectors: Preliminary clinical results. Eur J Nucl Med Mol Imaging 2012;39:83-90.

    Article  CAS  Google Scholar 

  6. Nakazato R, Tamarappoo BK, Kang X, Wolak A, Kite F, Hayes SW, et al. Quantitative upright-supine high-speed SPECT myocardial perfusion imaging for detection of coronary artery disease: Correlation with invasive coronary angiography. J Nucl Med 2010;51:1724-31.

    Article  Google Scholar 

  7. Jameria ZA, Abdallah M, Fernandez-Ulloa M, O’Donnell R, Dwivedi AK, Washburn E, et al. Analysis of stress-only imaging, comparing upright and supine CZT camera acquisition to conventional gamma camera images with and without attenuation correction, with coronary angiography as a reference. J Nucl Cardiol 2018;25:540-9.

    Article  Google Scholar 

  8. Grossman GB, Garcia EV, Bateman TM, Heller HV, Johnson LL, Folks RD. Quantitative Tc-99m sestamibi attenuation-corrected SPECT: Development and multicenter trial validation of myocardial perfusion stress gender-independent normal databases in an obese population. J Nucl Cardiol 2004;11:263-72.

    Article  Google Scholar 

  9. Takazakura R, Takahashi M, Nitta N, Murata K. Diaphragmatic motion in the sitting and supine positions: Healthy subject study using a vertically open magnetic resonance system. J Magn Reason Imaging 2004;19:605-9.

    Article  Google Scholar 

  10. Chowla D, Rahaby M, Amin AP, Vashista R, Alyyousef T, Martinez HX, et al. Soft tissue attenuation patterns in stress myocardial perfusion SPECT images: A comparison between supine and upright acquisition systems. J Nucl Cardiol. 2011;18:281-9011.

    Article  Google Scholar 

  11. Doukky R, Rahaby M, Chawla D, Vashistha R, Alyousef T, Amin AP. Soft tissue attenuation patterns associated with upright acquisition SPECT myocardial perfusion imaging: A descriptive study. Open Cardiovasc Med J 2012;6:22-7.

    Article  Google Scholar 

  12. Athar MW, Waqar F, Dwivedi AK, Ahmad S, Sanghvi S, Scott E, et al. Effects of gender and defect reversibility on detection of coronary disease with an upright and supine cadmium-zinc-telluride camera. J Nucl Cardiol 2019. https://doi.org/10.1007/s12350-019-01878-7.singh.

    Article  PubMed  Google Scholar 

  13. Singh B, Bateman TM, Case JA, Heller G. Attenuation artifact, attenuation correction, and the future of myocardial perfusion SPECT. J Nucl Cardiol 2007;14:153-64.

    Article  Google Scholar 

Download references

Acknowledgement

The authors thank Marge Duke for her invaluable assistance with data management.

Funding

Funded by a grant from the John R. Strauss Fund for Research and Education in Cardiovascular Imaging.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Myron C. Gerson.

Ethics declarations

Disclosures

None of the authors have any conflicts of interest. The study is funded by a grant from the John R. Strauss Fund for Research and Education in Cardiovascular Imaging.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

The authors of this article have provided a PowerPoint file, available for download at SpringerLink, which summarises the contents of the paper and is free for re-use at meetings and presentations. Search for the article DOI on SpringerLink.com.

The authors have also provided an audio summary of the article, which is available to download as ESM, or to listen to via the JNC/ASNC Podcast.

Supplementary Information

Below is the link to the electronic supplementary material.

Supplementary material 1 (PPTX 1712 kb)

Supplementary material 2 (DOCX 5157 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Waqar, F., Athar, M.W., Dwivedi, A.K. et al. Visual patterns of breast attenuation artifacts in women and men with an upright and supine cadmiun-zinc-telluride camera. J. Nucl. Cardiol. 29, 1976–1984 (2022). https://doi.org/10.1007/s12350-021-02632-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12350-021-02632-8

Keywords

Navigation