137Cs transmission imaging and segmented attenuation corrections in a small animal PET scanner

  • Published:
Radiological Physics and Technology Aims and scope Submit manuscript

Abstract

Attenuation correction (AC) is required for accurate quantitative evaluation of small animal PET data. Our objective was to compare three AC methods in the small animal Clairvivo-PET scanner. The three AC methods involve applying attenuation coefficient maps generated by simulating a cylindrical map (SAC), segmenting the emission data (ESAC), and segmenting the transmission data (TSAC), imaged using a 137Cs single-photon source. Investigation was carried out using a 65 mm uniform cylinder and an NEMA NU4 2008 mouse phantom, filled with water or tungsten liquid, to mimic bone. Evaluation was carried out using the difference of the segmented map volume from the known cylindrical phantom volume, the recovery of the radioactivity concentration, and the line profiles. The optimal transmission scan time for achieving accurate AC using TSAC was determined using 5, 10, 15, 20, and 25 min transmission scan time. The effects of scatter correction and reconstruction algorithms on ESAC were investigated. SAC showed the best performance but was unable to correct for different tissues and the scanner bed, and faced difficulty with correct positioning of the attenuation coefficient map. ESAC was affected by scatter correction and reconstruction algorithm, and may result in poor boundary delineation, and hence was unreliable. TSAC showed reasonable performance but required further optimization of the default segmentation setting. A minimum transmission scan time of 20 min is recommended for Clairvivo-PET using 137Cs source to ensure that sufficient transmission counts are obtained to generate accurate attenuation coefficient map.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Zaidi H, Hasegawa B. Determination of the attenuation map in emission tomography. J Nucl Med. 2003;44:291–315.

    PubMed  Google Scholar 

  2. El Ali HH, Bodholdt RP, Jørgensen JT, Myschetzky R, Kjaer A. Importance of attenuation correction (ac) for small animal PET imaging. Diagnostics. 2012;2:42–51.

    Article  PubMed  PubMed Central  Google Scholar 

  3. deKemp RA. Attenuation correction in PET using single photon transmission measurement. Med Phys. 1994;21(6):771.

    Article  CAS  PubMed  Google Scholar 

  4. D’Ambrosio D, Zagni F, Spinelli AE, Marengo M. Attenuation correction for small animal PET images: a comparison of two methods. Comput Math Methods Med. 2013;2013:1–12. doi:10.1155/2013/103476.

    Article  Google Scholar 

  5. Lehnert W, Meikle SR, Siegel S, Newport D, Banati RB, Rosenfeld AB. Evaluation of transmission methodology and attenuation correction for the microPET Focus 220 animal scanner. Phys Med Biol. 2006;51:4003–16.

    Article  CAS  PubMed  Google Scholar 

  6. Chow PL, Rannou FR, Chatziioannou AF. Attenuation correction for small animal PET tomographs. Phys Med Biol. 2005;50:1837–50.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Sato K, Shidahara M, Watabe H, Watanuki S, Ishikawa Y, Arakawa Y, Nai Y, Furumoto S, Tashiro M, Shoji T, et al. Performance evaluation of the small-animal PET scanner ClairvivoPET using NEMA NU 4-2008 standards. Phys Med Biol. 2016;61:696–711.

    Article  CAS  PubMed  Google Scholar 

  8. Mizuta T, Kitamura K, Iwata H, Yamagishi Y, Ohtani A, Tanaka K, Inoue Y. Performance evaluation of a high-sensitivity large-aperture small-animal PET scanner: ClairvivoPET. Ann Nucl Med. 2008;22:447–55.

    Article  PubMed  Google Scholar 

  9. Defrise M, Kinahan PE, Townsend DW, Michel C, Sibomana M, Newport DF. Exact and approximate rebinning algorithms for 3-D PET data. IEEE Trans Med Imaging. 1997;16:145–58.

    Article  CAS  PubMed  Google Scholar 

  10. Tanaka E, Kudo H. Subset-dependent relaxation in block-iterative algorithms for image reconstruction in emission tomography. Phys Med Biol. 2003;48:1405.

    Article  PubMed  Google Scholar 

  11. Kinouchi S, Yamaya T, Yoshida E, Tashima H, Kudo H, Suga M. GPU implementation of list-mode DRAMA for real-time OpenPET image reconstruction. In: IEEE nuclear science symposium & medical imaging conference. 2010. pp. 2273–2276.

  12. Lubberink M, Kosugi T, Schneider H, Ohba H, Bergström M. Non-stationary convolution subtraction scatter correction with a dual-exponential scatter kernel for the Hamamatsu SHR-7700 animal PET scanner. Phys Med Biol. 2004;49:833.

    Article  PubMed  Google Scholar 

  13. Otsu N. A threshold selection method from gray-level histograms. Automatica. 1975;11:23–7.

    Google Scholar 

  14. Loening AM, Gambhir SS. AMIDE: a free software tool for multimodality medical image analysis. Mol Imaging. 2003;2(3):131–7.

    Article  PubMed  Google Scholar 

  15. Chow PL, Bai B, Siegel S, Leahy RM, Chatziioannou AF. Transmission imaging and attenuation correction for the microPET® P4 tomograph. In: IEEE nuclear science symposium conference record. 2002. pp. 1298–1302.

  16. Karp JS, Muehllehner G, Qu H, Yan X-H. Singles transmission in volume-imaging PET with a 137Cs source. Phys Med Biol. 1995;40:929.

    Article  CAS  PubMed  Google Scholar 

  17. Bilger K, Adam LE, Karp JS. Segmented attenuation correction using 137 Cs single photon transmission. In: IEEE nuclear science symposium conference record. 2001. pp. 2095–2099.

  18. Watson CC, Schaefer A, Luk WK, Kirsch CM. Clinical evaluation of single-photon attenuation correction for 3D whole-body PET. IEEE Trans Nucl Sci. 1999;46:1024–31.

    Article  Google Scholar 

  19. Bailey DL, Livieratos L, Jones WF, Jones T. Strategies for accurate attenuation correction with single photon transmission measurements in 3D PET. In: IEEE nuclear science symposium. 1997. pp. 1009–1013.

Download references

Acknowledgements

The authors would like to thank Dr. Tetsuro Mizuta and Dr. Takayuki Kikuchi from Shimadzu Corp., Japan for their help and advice in this study. This study was funded by the Grants-in-Aid for Scientific Research (B) (No. 26293133) and (C) (No. 15K08707) from the Ministry of Education, Culture, Sports, Science and Technology (MEXT), Japanese Government.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hiroshi Watabe.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Statement of human and/or animal rights

This article does not contain any studies with human participants or animals performed by any of the authors.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nai, YH., Ose, T., Shidahara, M. et al. 137Cs transmission imaging and segmented attenuation corrections in a small animal PET scanner. Radiol Phys Technol 10, 321–330 (2017). https://doi.org/10.1007/s12194-017-0407-4

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12194-017-0407-4

Keywords

Navigation