Abstract
Purpose
Measurement of cortisol concentrations is method dependent. The study aimed to establish assay-specific cut-off limits for cortisol after adrenocorticotropic hormone (ACTH) stimulation, comparing Roche Elecsys Cortisol II immunoassay to liquid chromatography-mass spectrometry (LC-MS/MS), and to assess the impact of patient characteristics, estrogen containing oral contraceptives as well as relation to other adrenocortical steroid hormone dynamics.
Methods
One hundred healthy participants underwent a 250 μg ACTH-test, with plasma samples analyzed using ElecsysCortI, ElecsysCortII, and LC-MS/MS. Cortisone, corticosterone, 17-OH-progesterone, dehydroepiandrosterone sulfate (DHEAS), androstenedione, and testosterone were additionally analyzed with LC-MS/MS. Cut-off limit for a normal cortisol response to the ACTH-test was defined as: 2.5th percentile–1.96 × SE.
Results
ElecsysCort II measured cortisol concentrations 21% (95% CI: 19–22%) lower than ElecsysCort I. Cut-off limits for cortisol 30 and 60 min after ACTH were 426 and 485 nmol/L (ElecsysCort II) and 411 and 470 nmol/L (LC-MS/MS). Cut-offs were unaffected by gender, or body-composition. The ACTH-test resulted in significantly increased adrenocortical steroid hormones, except for decreased cortisone concentrations (both sexes), and decreased testosterone in men (1.9 nmol/L, 95% CI: 1.3–2.5). Testosterone was increased in women (0.07 nmol/L, 95% CI: 0.02–0.13).
Conclusion
ElecsysCort II has high analytical performance and yields significantly lower cortisol concentrations than prior polyclonal immunoassays. This clinically relevant difference underscores the necessity for revised cut-off limits for improved diagnostic precision. Suggested 30-minute cortisol cutoff limits are 411 nmol/L (LC-MS/MS) and 426 nmol/L (ElecsysCort II). Adrenocortical steroids increased upon ACTH stimulation, except for cortisone in both sexes and testosterone in men, both of which decreased.
Similar content being viewed by others
Introduction
The adrenocorticotropic hormone (ACTH) stimulation test and the insulin tolerance test (ITT) are the diagnostic tests usually performed to confirm or reject the clinical suspicion of adrenal insufficiency [1, 2]. Both tests include assessment of the cortisol concentration in serum or plasma, although assessment in saliva has been suggested in recent years [3]. Cut-off limits defining insufficiency depend on test circumstances, the type of stimulation test, and not least the assay used for measurement of cortisol [4,5,6,7]. The ACTH test is a validated and safe test [2, 8, 9], and reliable in most circumstances, although a disadvantage is its inability to detect newly onset central adrenal insufficiency [8].
Most often, cortisol concentrations are measured using immunoassays performed in automated analyzers because they are reasonably priced and have a short turnaround time, which allows for the same day or day-to-day diagnosis. However, the results are largely dependent on the antibody sensitivity and specificity in the chosen assay, and a disadvantage can thus be that the antibodies used cross-react and bind to more than the intended antigen [10, 11]. Within recent years, studies have tested different 2nd generation cortisol immunoassays using more specific antibodies and suggested cortisol cut-off limits for the ACTH test ranging from 350–460 nmol/L [12,13,14,15,16,17,18], which is lower than the formerly recommended cut-off limit at 500–550 nmol/L [19].
As a consequence of differences in performance and standardization of immunoassays, cut-off limits for cortisol deficiency should ideally be identified for each stimulation test and assay. The 2nd generation immunoassay Roche Elecsys Cortisol II (ElecsysCort II) is improved by using monoclonal antibodies to identify cortisol, which has increased its specificity compared with Roche Elecsys Cortisol I (ElecsysCort I), which uses polyclonal antibodies [20]. As given by the manufacturer, this resulted in cortisol concentrations approximately 20% lower when measured by the new more specific immunoassay. Thus, a switch to this new and more specific assay has significant clinical implications, as it entails the need for an adjustment of the cut-off limits defining adrenal insufficiency which guide the clinicians’ decision whether to treat or not.
The aim of the study was to perform a direct method comparison between the gold standard LC-MS/MS and the 1st and 2nd generation Roche cortisol immunoassays, in order to establish test and assay specific cut-off limits for the standard 250 μg ACTH test, defined in a large well characterized group of healthy participants. In addition, we aimed to address the potential impact of gender and body composition, and to assess the impact of patient characteristics, estrogen containing oral contraceptives as well as relation to other adrenocortical steroid hormone dynamics.
Materials and methods
The participants consisted of 50 healthy men and 63 healthy women. Thirteen of the women were using oral contraceptives (OC) containing ethinyl estradiol in combination with progestins. Inclusion criteria were healthy individuals aged >16 years. Healthy participants were defined as participants with no medical conditions known to potentially affect the outcome at the time of study participation. Exclusion criteria were ongoing treatment with glucocorticoids or spironolactone, pregnancy, breastfeeding, adrenal and pituitary disease [4].
Sample collection and handling
Participants rested 15 min before testing after inserting an indwelling catheter in a large forearm vein. An ACTH test was performed between 0800 and 1000 h, after an overnight fast, administering 250 µg iv ACTH1–24 (Synacthen; Novartis Healthcare, Copenhagen, Denmark). All participants were tested in a supine position with sampling at baseline and at 30 and 60 min. The blood samples were centrifuged, and plasma samples were stored at −80 °C until analysis. All samples from each participant were analyzed in triplicate by ElecsysCort I and II as well as LC-MS/MS for the assessment of P-cortisol concentrations. Samples were also analyzed for cortisone, androstenedione, testosterone, 17-OH-progesterone, and dehydroepiandrosterone sulfate (DHEAS) by LC-MS/MS. Forty-one participants underwent a dual-energy X-ray scan (DXA) (model XP-26/XR-46; Norland Medical Systems, Fort Atkinson, WI) for the examination of body composition of both total and regional fat mass. The DXA scan had an in-house intra-operator variation of 5%.
Laboratory analysis
Cortisol was quantified by the two commercial immunoassays Roche Elecsys Cortisol I and II (on Cobas 8000 e-module) (Roche GmbH, Germany) and by LC-MS/MS.
Elecsys Cortisol uses electrochemiluminescent detection. ElecsysCort I has interassay CVs (Coefficients of Variability) of 4.3%, 3.8%, and 2.5% at concentrations of 101, 436 and 1095 nmol/L, respectively, while ElecsysCort II has interassay CVs of 2.1%, 1.6%, and 1.8% at concentrations of 161, 532 and 837 nmol/L [21].
The LC-MS/MS was adapted from Phenomenex Inc. (USA) application (20655) with another column and gradient and performed using a Waters TQ-S instrument, with an i-class HPLC and a Waters Cortecs T3, 2.1 × 50 mm, 1.7 µm column. The mobile phase A was water, 0.1% formic acid, and 2 mM ammonium acetate, while mobile phase B was pure methanol. The gradient was 55% phase A decreasing linearly to 20% at 1.6 min further decreasing to 1% at 1.7 min. The rinse between samples was 99% phase B for 1.3 min and back at 3 min ending at 3.8 min. The column temperature was 40 °C. The LC-MS/MS method was ESI+ with the parameters shown in Online Resource 1. The sample was prepared by mixing with 3 parts acetonitrile containing the internal standard (9,11,12,12-D4-Hydrocortison - Sigma 705594), centrifuged and 2 µL of the supernatant was injected. The method was part of an external quality control program from UK-NEQAS, and the trueness of the calibration was verified by measuring NIST SRM 921.
Statistics
Statistical analyses were performed using SAS Enterprise Guide Version 8.3 (SAS Institute Inc., Cary, NC, USA), R Version 4.3.1 (R Foundation for Statistical Computing, Austria), and Microsoft Excel Version 16.75.2 (Microsoft Corporation, USA). P-cortisol was log Gaussian distributed and thus log-transformed before analyses. Baseline reference intervals were calculated as the 2.5th and 97.5th percentile; mean plasma cortisol ± 1.96 × SD (standard deviation). The adjusted 2.5th percentile was calculated as the 2.5th percentile-1.96 × SE (standard error of the mean) to exclude false positive cases. The cut-off limit for a normal adrenal function was defined as the adjusted 2.5th percentile of P-cortisol concentrations 30 min after Synacthen® injection. Comparison of the 2.5th percentile and adjusted 2.5th percentile of P-cortisol concentrations between sexes and between women using OC and women not using OC were evaluated with unpaired t-tests. Assay differences were demonstrated by creating histograms, Deming regression, and Bland-Altman plots with limits of agreement (LOA) defined as mean difference ±2 SD. The response to ACTH stimulation by the other six adrenal hormones was examined with linear mixed models. The association of body composition and cortisol concentrations during the ACTH test was examined with Spearman correlation and univariate regression analyses. Observations with missing values were excluded from analyses. A difference was considered significant when P < 0.05.
Results
The characteristics of the participants are summarized in Table 1.
Comparison of ElecsysCort I and II
ElecsysCort II measured P-cortisol concentrations 21% lower than ElecsysCort I as indicated by the slope of the scatter (0.79 (95% CI: 0.78–0.81)) (Fig. 1a). The agreement between the two immunoassays from Roche is illustrated in Fig. 2a–c. The Bland-Altman plots show the numeric difference in P-cortisol between the two methods which increased as the mean value of the measurements increased as indicated by the bias line (Fig. 2a–c). ElecsysCort II measured mean P-cortisol to be 124 nmol/L, 185 nmol/L, and 200 nmol/L lower than ElecsysCort I at baseline, 30 min, and 60 min, respectively.
Comparison of ElecsycCort II and LC-MS/MS
The P-cortisol concentrations measured by ElecsysCort II and LC-MS/MS were highly concordant with a mean difference of 8% (1.08 (95% CI: 1.05–1.10)) (Fig. 1b). The numeric differences between the methods at baseline and post ACTH stimulation are illustrated in Fig. 2d–f. The numeric differences of P-cortisol concentrations exhibited less variation, although increasing with higher cortisol concentrations as indicated by the bias line. ElecsysCort II measured mean P-cortisol 9 nmol/L higher compared with LC-MS/MS 30 min post ACTH stimulation, but 4 nmol/L and 35 nmol/L lower than LC-MS/MS at baseline and 60 min post ACTH stimulation, respectively.
Reference intervals and cut-off limits
The distributions of 30 min P-cortisol concentration from the 100 participants (excluding women on OC) as measured by ElecsysCort II versus ElecsysCort I (Fig. 3a) and ElecsysCort II versus LC-MS/MS (Fig. 3b).
The reference intervals and cut-off limits for normal adrenal function with each assay are shown in Table 2. The cut-off limit was 574 nmol/L with ElecsysCort I, 426 nmol/L with ElecsysCort II, and 411 nmol/L with LC-MS/MS 30 min after ACTH stimulation. There was no significant difference between the 2.5th percentiles of men and women who were not on OC (Table 3).
Cortisol measurement and oral contraceptives
Women on OC (N = 13) had significantly higher 2.5th and adjusted 2.5th P-cortisol cut-off limits at all timestamps during ACTH stimulation regardless of the method used (Table 3).
The median increase in P-cortisol during the ACTH test from 0 to 30 min was 150 nmol/L (IQR: 119–252 nmol/L) with ElecsysCort I, 149 nmol/L (IQR: 117–237 nmol/L) with ElecsysCort II, and 192 nmol/L (IQR: 160–235 nmol/L) with LC-MS/MS. From 0 to 60 min, the median increase in P-cortisol was 316 nmol/L (IQR: 192–350 nmol/L) with ElecsysCort I, 310 nmol/L (IQR: 228–360 nmol/L) with ElecsysCort II, and 309 nmol/L (IQR: 134–386 nmol/L) with LC-MS/MS. The cortisol increase from 0–30 min was not significantly different in women taking OC compared to women who did not take OC: ElecsysCort II (mean difference 32 nmol/L, 95% CI: 11–75 nmol/L, P = 0.4), and LC-MS/MS (mean difference 25 nmol/L, 95% CI: 7–57 nmol/L, P = 0.9). nor from 0–60 min: ElecsysCort II (22 nmol/L, 95% CI: 1.5–45 nmol/L, P = 0.4), and LC-MS/MS (42 nmol/L, 95% CI: 0.8–83 nmol/L, P = 0.1).
Influence of body composition on cortisol
Data concerning the influence of body mass index (BMI), waist-hip ratio (WHR), abdominal fat mass (ABD), and total fat mass (TFM) on baseline cortisol, stimulated cortisol, and the increase in cortisol during ACTH test as measured by LC-MS/MS are available as supplemental files (Online Resource 2, 3).
Baseline and 30- and 60-minute stimulated cortisol concentrations did not correlate with BMI, WHR, ABD, and TFM in the total group of subjects (P > 0.05) (see data in Online Resource 2).
In men, the increase in cortisol from 0 to 30 min correlated positively to WHR (P = 0.03) but was otherwise unrelated to the other body composition variables. In women, the increase in cortisol from 0 to 30 min correlated positively with WHR (P = 0.02), ABD (P = 0.001), and TFM (P < 0.01). These body composition variables remained significantly correlated to the increase in cortisol from 0 to 60 min (see data in Online Resource 2, 3).
ACTH stimulation of other adrenocortical steroids
The baseline reference intervals, and increase/decrease 30 and 60 min after the ACTH test are shown in Table 4. Androstenedione, 17-OH-progesterone, and DHEAS increased significantly in both men and women after ACTH stimulation, whereas cortisone decreased in both sexes reaching significantly lower concentrations after 30 min in women (mean difference: 55 nmol/L, SD: 9.7; P < 0.0001) and after 60 min in men (mean difference: 57 nmol/L, SD: 7.3; P < 0.02). Testosterone decreased in men (mean: 1.9 nmol/L, 95% CI: 1.3–2.5; P < 0.0001) but increased in women (mean: 0.07 nmol/L, 95% CI: 0.02–0.13; P = 0.01). Delta cortisol was positively correlated with delta corticosterone (r = 0.60, P < 0.0001), androstenedione (r = 0.59, P < 0.0001), and DHEAS (r = 0.34; P < 0.01) in women, whereas positively correlated with corticosterone (r = 0.65, P < 0.0001) and androstenedione (r = 0.70; P < 0.0001) in men. Delta cortisol was not correlated with delta cortisone or testosterone in neither women nor men. Delta testosterone in men was positively correlated with delta DHEAS (r = 0.53; P < 0.0001), and delta cortisone (r = 0.31; P = 0.03), but not to any of the other steroid hormones. No such correlations were seen in women.
Discussion
This study compared 1st and 2nd generation immunoassays ElecsysCort I, and ElecsysCort II, and the gold standard LC-MS/MS in measuring P-cortisol concentrations during ACTH stimulation in a large cohort of healthy individuals, in whom DXA was performed allowing for assessment of the potential relation to body composition. The P-cortisol concentrations measured with ElecsysCort II and LC-MS/MS were highly concordant with a mean difference of 8%, while ElecsysCort II measured P-cortisol concentrations 21% lower than ElecsysCort I. The cut-off limits for normal adrenal function (defined as the adjusted 2.5th percentile) were 574 nmol/L with ElecsysCort I, 426 nmol/L with ElecsysCort II, and 411 nmol/L with LC-MS/MS 30 min post ACTH stimulation. This data illustrates that the application of cut-off limits for adrenal insufficiency defined by first generation immunoassays can mislead to a large proportion of false positive cases when using more specific immunoassays such as ElecsysCort II or LC-MS/MS. The high concordance with the gold standard method LC-MS/MS measuring baseline and stimulated cortisol concentrations was also reported by others [14, 18, 20] and underlines that with ElecsysCort II we now have an immunoassay that produces more reliable results for the establishment of a cortisol cut-off limit. The explanation for this is the change from using polyclonal antibodies in ElecsysCort I to more specific monoclonal antibodies in ElecsysCort II, which has reduced the variability and enhanced the specificity of the immunoassay as cortisol now binds to a single site on the target antigen reducing the possibility of cross-reactivity with other substances. This is an important improvement as the use of immunoassays is more widespread than the use of LC-MS/MS. The strength of LC-MS/MS is its capability to quantify compounds with a high degree of sensitivity and selectivity based on unique mass/charge transition of each compound of interest. This method is however far more labor intensive, expensive, and not generally available.
In recent studies, the cut-off limit for a normal cortisol response to 250 μg ACTH varied from 400–440 nmol/L with ElecsysCort II [14, 17, 18] and from 400–412 nmol/L with LC-MS/MS [14, 18], which is in agreement with the results from the present study. Studies using the low-dose ACTH test, ITT, and glucagon stimulation test appeared to find lower cut-off limits ranging from 350–375 nmol/L with ElecsysCort II [12, 13, 15]. Thus, differences in cut-off limits should also be considered depending on stimulation tests and not only by assay variations [1, 5]. However, opinions were divided as older studies suggested the use of ITT, low-dose, and standard-dose ACTH to be equal and some even recommended low-dose ACTH test as standard method in screening for adrenal insufficiency [2, 6, 22, 23]. However, the low-dose ACTH test could be influenced by technical details e.g., loss of ACTH through tubing hereby decreasing the accuracy and specificity [24].
As anticipated, women on OC had significantly higher P-cortisol concentrations compared to women not on OC both before and after ACTH stimulation [4], due to estrogen induced elevation of cortisol binding globulin in plasma which challenges the diagnosis in OC users [10]. The increase in cortisol during ACTH stimulation was parallel to that observed in non-OC users and it could be speculated if the increase in P-cortisol during stimulation could be used to suggest or exclude adrenal insufficiency in this challenging situation. Such an approach should eventually take the baseline concentration into account as a lower increase was seen with higher baseline concentration in healthy subjects not on OC [4, 7].
Body composition did not affect baseline, 30- and 60-min stimulated cortisol concentrations during the ACTH test in men and women, which is in accordance with previous data [25, 26]. Another study found that obese men had the same increase in cortisol to ACTH stimulation compared with a normal weight control group which supports our findings [27]. On the other hand, we found that women with higher WHR and ABD had a higher increase in cortisol during the ACTH stimulation, which was also described by others [28].
The adrenocortical steroids increased upon ACTH stimulation, except for testosterone in men and cortisone in both sexes. The rise in 17-OH progesterone in women is inconclusive due to the lack of information on blood sample timing with the menstrual cycle [29]. Men experienced a significant decrease in testosterone, which is probably caused by the excessive presence of ACTH that can interfere with the usual signaling by luteinizing hormone, leading to a partial yet definite decrease in androgen production in the testes [30]. The significant decrease in cortisone, the inactive form of cortisol, may be linked to the enzyme 11-beta hydroxysteroid dehydrogenase restoring cortisone to biologically active cortisol [31]. Similarly, increased ratio of cortisol to cortisone was seen in acute-phase response to increasing CRP [32] and in intensive care patients [33].
A limitation of this study was the number of participants, albeit the number of healthy subjects assessed by ElecsysCort II was the largest so far. Achieved cut-off limits would always be more reliable if more participants were included. However, studies that included fewer or slightly more participants (ranging from 39 to 137) to define the cut-off limit for cortisol with ElecsysCort II or LC-MS/MS found very similar cut-off limits as this study [13, 14, 17, 18] supporting robustness of the suggested cut-offs. The evidence for such robustness ensures that data can be compared across centers to a larger extent than previously. The definition of a cut-off for adrenal insufficiency is historically defined based on healthy individuals’ ability to increase cortisol during i.e. surgical stress, and this response was later translated into the response of the ACTH test. Studies performed in patients with pituitary disease [17] supported the cut-off threshold including patients judged as adrenal sufficient by the treating physicians.
The development of high-performance assays allows for a more valid comparison of results across centers using different assays. As it stands now, the newer assays on the market are becoming more uniform in the measurement of cortisol in blood samples. Some of the frequently used cortisol immunoassays have been documented to measure similar P-cortisol cut-off limits (30 min post ACTH stimulation) as ElecsysCort II e.g., 423 nmol/L with Immulite 2000 by Siemens [14], 427 nmol/L with Access by Beckman Coulter [18], and 364 nmol/L with Architect Cortisol by Abbott [16]. This makes the interpretation of cortisol more consistent regardless of the assay used and the test results easily commutable when the patients are being transferred from one hospital to another. A suggestion for future guidelines in this area would be to define cortisol cut-off limits based on a specific assay combined with a specific stimulation test and a specific laboratory and including cortisol insufficient patients for comparison. However, several factors contribute to variations in results across laboratories. These include differences in assay types, often inadequate cohort sizes, variations in cohort compositions, unexplained technical disparities among laboratories, and various interfering factors such as discrepancies that may arise in methods of measurement and correction for binding proteins. Notably, such cross-laboratory comparisons, especially for different cortisol measurement methods and stimulation techniques, can potentially be detected through participation in quality control programs.
In conclusion, ElecsysCort II has a high analytical performance and measures significantly lower cortisol concentrations compared to the previous polyclonal immunoassay ElecsysCort I, but similar concentrations compared to LC-MS/MS. This difference is clinically relevant, and cut-off limits must be changed to improve diagnostic precision. We recommend a 30 min cortisol cutoff of 411 nmol/L (LC-MS/MS) and 426 nmol/L (ElecsysCort II). Cut-of limits were unaffected by gender and body composition. Whether delta cortisol can be used to suggest adrenal sufficiency warrant more data. As anticipated several adrenocortical steroids increased upon ACTH stimulation, except for testosterone in men and cortisone in both sexes which decreased.
References
Y. Simsek, Z. Karaca, F. Tanriverdi, K. Unluhizarci, A. Selcuklu, F. Kelestimur, A comparison of low-dose ACTH, glucagon stimulation and insulin tolerance test in patients with pituitary disorders. Clin. Endocrinol. 82, 45–52 (2015). https://doi.org/10.1111/cen.12528
T.A.M. Abdu, T.A. Elhadd, R. Neary, R.N. Clayton, Comparison of the Low Dose Short Synacthen Test (1 µg), the Conventional Dose Short Synacthen Test (250 µg), and the Insulin Tolerance Test for Assessment of the Hypothalamo-Pituitary-Adrenal Axis in Patients with Pituitary Disease. J. Clin. Endocrinol. Metab. 84, 839–843 (1999). https://doi.org/10.1210/jcem.84.3.5535
M.H. Laudat, S. Cerdas, C. Fournier, D. Guiban, B. Guilhaume, J.P. Luton, Salivary Cortisol Measurement: A Practical Approach to Assess Pituitary-Adrenal Function. J. Clin. Endocrinol. Metab. 66, 343–348 (1988). https://doi.org/10.1210/jcem-66-2-343
M. Klose, M. Lange, A.K. Rasmussen, N.E. Skakkebæk, L. Hilsted, E. Haug, M. Andersen, U. Feldt-Rasmussen, Factors influencing the adrenocorticotropin test: Role of contemporary cortisol assays, body composition, and oral contraceptive agents. J. Clin. Endocrinol. Metab. 92, 1326–1333 (2007). https://doi.org/10.1210/jc.2006-1791
H.Y. Cho, J.H. Kim, S.W. Kim, C.S. Shin, K.S. Park, S.W. Kim, H.C. Jang, S.Y. Kim, Different cut-off values of the insulin tolerance test, the high-dose short Synacthen test (250 μg) and the low-dose short Synacthen test (1 μg) in assessing central adrenal insufficiency. Clin. Endocrinol. 81, 77–84 (2014). https://doi.org/10.1111/cen.12397
J. Gonzálbez, C. Villabona, J. Ramón, M.A. Navarro, O. Giménez, W. Ricart, J. Soler, Establishment of reference values for standard dose short synacthen test (250 μg), low dose short synacthen test (1 μg) and insulin tolerance test for assessment of the hypothalamo-pituitary-adrenal axis in normal subjects. Clin. Endocrinol. 53, 199–204 (2000). https://doi.org/10.1046/j.1365-2265.2000.01028.x
N. El-Farhan, A. Pickett, D. Ducroq, C. Bailey, K. Mitchem, N. Morgan, A. Armston, L. Jones, C. Evans, D.A. Rees, Method-specific serum cortisol responses to the adrenocorticotrophin test: Comparison of gas chromatography-mass spectrometry and five automated immunoassays. Clin. Endocrinol. 78, 673–680 (2013). https://doi.org/10.1111/cen.12039
J. Lindholm, H. Kehlet, Re-evaluation of the clinical value of the 30 min ACTH test in assessing the hypothalamic-pituitary-adrenocortical function. Clin. Endocrinol. 26, 53–59 (1987). https://doi.org/10.1111/j.1365-2265.1987.tb03638.x
J. Wood, A. Frankland, V.H.T. James, J. Landon, A rapid test of adrenocortical function. Lancet 234, 243–245 (1965). https://doi.org/10.1016/s0140-6736(65)91526-6
M.H. Choi, Clinical and Technical Aspects in Free Cortisol Measurement. Endocrinol. Metab. 37, 599–607 (2022). https://doi.org/10.3803/EnM.2022.1549
G. Casals, F.A. Hanzu, Cortisol measurements in Cushing’s syndrome: Immunoassay or mass spectrometry? Ann. Lab. Med. 40, 285–296 (2020). https://doi.org/10.3343/alm.2020.40.4.285
G.A. Kline, J. Buse, R.D. Krause, Clinical implications for biochemical diagnostic thresholds of adrenal sufficiency using a highly specific cortisol immunoassay. Clin. Biochem. 50, 475–480 (2017). https://doi.org/10.1016/j.clinbiochem.2017.02.008
V. Raverot, C. Richet, Y. Morel, G. Raverot, F. Borson-Chazot, Establishment of revised diagnostic cut-offs for adrenal laboratory investigation using the new Roche Diagnostics Elecsys® Cortisol II assay. Ann. Endocrinol. 77, 620–622 (2016). https://doi.org/10.1016/j.ando.2016.05.002
G.A. Ueland, P. Methlie, M. Øksnes, H.B. Thordarson, J. Sagen, R. Kellmann, G. Mellgren, M. Ræder, P. Dahlqvist, S.R. Dahl, P.M. Thorsby, K. Løvås, E.S. Husebye, The Short Cosyntropin Test Revisited: New Normal Reference Range Using LC-MS/MS. J. Clin. Endocrinol. Metab. 103, 1696–1703 (2018). https://doi.org/10.1210/jc.2017-02602
G. Grassi, V. Morelli, F. Ceriotti, E. Polledri, S. Fustinoni, S. D’agostino, G. Mantovani, I. Chiodini, M. Arosio, Minding the gap between cortisol levels measured with second-generation assays and current diagnostic thresholds for the diagnosis of adrenal insufficiency: a single-center experience. Hormones 19, 425–431 (2000). https://doi.org/10.1007/s42000-020-00185-y
L. Zha, J. Li, S.M. Krishnan, M.R. Brennan, Y.V. Zhang, P. Povse, R. Kerlin, K. Shively, F. Oleksik, J.A. Williams, E. Sykes, Q. Sun, New Diagnostic Cutoffs for Adrenal Insufficiency After Cosyntropin Stimulation Using Abbott Architect Cortisol Immunoassay. Endocr. Pr. 28, 684–689 (2022). https://doi.org/10.1016/J.EPRAC.2022.04.003
H. Husni, M.S. Abusamaan, R. Dinparastisaleh, L. Sokoll, R. Salvatori, A.H. Hamrahian, Cortisol values during the standard-dose cosyntropin stimulation test: Personal experience with Elecsys cortisol II assay. Front Endocrinol. 13, 1–7 (2022). https://doi.org/10.3389/fendo.2022.978238
B.R. Javorsky, H. Raff, T.B. Carroll, A. Algeciras-Schimnich, R.J. Singh, J.M. Colón-Franco, J.W. Findling, New Cutoffs for the Biochemical Diagnosis of Adrenal Insufficiency after ACTH Stimulation using Specific Cortisol Assays. J. Endocr. Soc. 5, 1–11 (2021). https://doi.org/10.1210/jendso/bvab022
P. Agnieszka, S. Pearce, Adrenal insufficiency – Recognition and management. Clin. Med. 17, 258–262 (2017). https://doi.org/10.7861/clinmedicine.17-5-480a
M. Vogeser, J. Kratzsch, Y. Ju Bae, M. Bruegel, U. Ceglarek, T. Fiers, A. Gaudl, H. Kurka, C. Milczynski, C. Prat Knoll, A.C. Suhr, D. Teupser, I. Zahn, R.E. Ostlund, Multicenter performance evaluation of a second generation cortisol assay. Clin. Chem. Lab Med. 55, 826–835 (2017). https://doi.org/10.1515/cclm-2016-0400
J.M. Hawley, L.J. Owen, S.J. Lockhart, P.J. Monaghan, A. Armston, C.A. Chadwick, H. Wilshaw, M. Freire, L. Perry, B.G. Keevil, Serum cortisol: An up-to-date assessment of routine assay performance. Clin. Chem. 62, 1220–1229 (2016). https://doi.org/10.1373/clinchem.2016.255034
K. Tordjman, A. Jaffe, Y. Trostanetsky, Y. Greenman, R. Limor, N. Stern, Low-dose (1 μg) adrenocorticotrophin (ACTH) stimulation as a screening test for impaired hypothalamo-pituitary-adrenal axis function: Sensitivity, specificity and accuracy in comparison with the high-dose (250 μg) test. Clin. Endocrinol. 52, 633–640 (2000). https://doi.org/10.1046/j.1365-2265.2000.00984.x
N.S. Ospina, A. Nofal, Al, I. Bancos, A. Javed, K. Benkhadra, E. Kapoor, A.N. Lteif, N. Natt, M.H. Murad, ACTH stimulation tests for the diagnosis of adrenal insufficiency: Systematic review and meta-analysis. J. Clin. Endocrinol. Metab. 101, 427–434 (2016). https://doi.org/10.1210/jc.2015-1700
M. Wade, S. Baid, K. Calis, H. Raff, N. Sinaii, L. Nieman, Technical details influence the diagnostic accuracy of the 1 μg ACTH stimulation test. Eur. J. Endocrinol. 162, 109–113 (2010). https://doi.org/10.1530/EJE-09-0746
L.G. Mancillas-Adame, A. Sánchez-García, R. Rodriguez-Gutierrez, C. Gonzalez-Velazquez, F.J. Lavalle-Gonzalez, J.A. Zuñiga-Hernandez, A.G. Rios-Ortega, J.G. González-González, Low-Dose ACTH Stimulation Test in Obesity: A Randomized Dose Assessment. Int. J. Endocrinol. 2022, (2022). https://doi.org/10.1155/2022/7860272
I.A. Odeniyi, O.A. Fasanmade, A.O. Ogbera, A.E. Ohwovoriole, Body mass index and its effect on serum cortisol level. Niger. J. Clin. Pr. 18, 194–197 (2015). https://doi.org/10.4103/1119-3077.151040
R. Pasquali, L. Gagliardi, V. Vicennati, A. Gambineri, D. Colitta, L. Ceroni, F. Casimirri, ACTH and cortisol response to combined corticotropin releasing hormone-arginine vasopressin stimulation in obese males and its relationship to body weight, fat distribution and parameters of the metabolic syndrome. Int J. Obes. 23, 419–424 (1999). https://doi.org/10.1038/sj.ijo.0800838
R. Pasquali, S. Cantobelli, F. Casimirri, M. Capelli, L. Bortoluzzi, R. Flamia, A.M.M. Labate, L. Barbara, The hypothalamic-pituitary-adrenal axis in obese women with different patterns of body fat distribution. J. Clin. Endocrinol. Metab. 77, 341–346 (1993). https://doi.org/10.1210/jcem.77.2.8393881
I.H. Thorneycroft, D.R. Mishell, S.C. Stone, K.M. Kharma, R.M. Nakamura, The relation of serum 17-hydroxyprogesterone and estradiol-17-beta levels during the human menstrual cycle. Am. J. Obstet. Gynecol. 111, 947–951 (1971). https://doi.org/10.1016/0002-9378(71)90951-3
I.Z. Beitins, F. Bayard, A. Kowarski, C.J. Migeon, The effect of ACTH administration on plasma testosterone, dihydrotestosterone and serum LH concentrations in normal men. Steroids 21, 553–563 (1973). https://doi.org/10.1016/0039-128x(73)90044-5
M. Vogeser, R. Zachoval, K. Jacob, Serum cortisol/cortisone ratio after Synacthen stimulation. Clin. Biochem. 34, 421–425 (2001). https://doi.org/10.1016/s0009-9120(01)00251-x
M. Vogeser, R. Zachoval, T.W. Felbinger, K. Jacob, Increased Ratio of Serum Cortisol to Cortisone in Acute-Phase Response. Horm. Res. 58, 172–175 (2002). https://doi.org/10.1159/000065486
E. Boonen, H. Vervenne, P. Meersseman, R. Andrew, L. Mortier, P.E. Declercq, Y.-M. Vanwijngaerden, I. Spriet, P.J. Wouters, S. Vander Perre, L. Langouche, I. Vanhorebeek, B.R. Walker, G. Van den Berghe, Reduced Cortisol Metabolism during Critical Illness. N. Engl. J. Med. 368, 1477–1488 (2013). https://doi.org/10.1056/nejmoa1214969
Funding
This study was supported by unrestricted grants from Eva Maduras Foundation. S.O. was supported by a fellowship from Copenhagen University Hospital, Rigshospitalet, and the Novo Nordisk Foundation (Project: Double Edge, Grant: NNF20OC0063280). U.F.-R.’s research salary was sponsored by Kirsten and Freddy Johansen’s Fund. M.K.’ research salary was sponsored by Skibsreder Per Henriksen, R. & Hustrus Fund. Open access funding provided by National Hospital.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study’s conception and design. Material preparation, data collection, and analysis were performed by M.K., L.H., L.F.H., L.E.P. and S.O. The first draft of the manuscript was written by S.O. and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Ethics approval
This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the local ethics committee.
Informed consent
Informed and written consent was obtained from all individual participants included in the study.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Okutan, S., Jørgensen, N.T., Pedersen, L.E. et al. Determination of cortisol cut-off limits and steroid dynamics in the ACTH stimulation test: a comparative analysis using Roche Elecsys Cortisol II immunoassay and LC-MS/MS. Endocrine (2024). https://doi.org/10.1007/s12020-024-03752-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s12020-024-03752-0