Utilization of a 3D printed dental splint for registration during electromagnetically navigated mandibular surgery

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

A dental splint was developed for non-invasive rigid point-based registration in electromagnetically (EM) navigated mandibular surgery. Navigational accuracies of the dental splint were compared with the common approach, that is, using screws as landmarks.

Methods

A dental splint that includes reference registration notches was 3D printed. Different sets of three points were used for rigid point-based registration on a mandibular phantom: notches on the dental splint only, screws on the mandible, contralateral screws (the side of the mandible where the sensor is not fixated) and a combination of screws on the mandible and notches on the dental splint. The accuracy of each registration method was calculated using 45 notches at one side of the mandible and expressed as the target registration error (TRE).

Results

Average TREs of 0.83 mm (range 0.7–1.39 mm), 1.28 mm (1.03–1.7 mm), 2.62 mm (1.91–4.0 mm), and 1.34 mm (1.30–1.39 mm) were found, respectively, for point-based registration based on the splint only, screws on the mandible, screws on the contralateral side only, and screws combined with the splint.

Conclusion

For dentate patients, rigid point-based registration performs best utilizing a dental splint with notches. The dental splint is easy to implement in the surgical, and navigational, workflow, and the notches can be pinpointed and designated on the CT scan with high accuracy. For edentate patients, screws can be used for rigid point-based registration. However, a new design of the screws is recommended to improve the accuracy of designation on the CT scan.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. ter Braak TP, Brouwer de Koning SG, van Alphen MJA et al (2020) A surgical navigated cutting guide for mandibular osteotomies: accuracy and reproducibility of an image-guided mandibular osteotomy. Int J CARS 15:1719–1725. https://doi.org/10.1007/s11548-020-02234-8

    Article  Google Scholar 

  2. Brouwer de Koning SG, Geldof F, van Veen RLP, van Alphen MJA, Karssemakers LHE, Nijkamp J, Schreuder WH, Ruers TJM, Karakullukcu MB. Electromagnetic surgical navigation in patients undergoing mandibular surgery. Int J CARS (Submitted)

  3. Naujokat H, Rohnen M, Lichtenstein J, Birkenfeld F, Gerle M, Florke C, Wiltfang J (2017) Computer-assisted orthognathic surgery: evaluation of mandible registration accuracy and report of the first clinical cases of navigated sagittal split ramus osteotomy. Int J Oral Maxillofac Surg 46(10):1291–1297

    Article  CAS  Google Scholar 

  4. Eggers G, Muhling J, Marmulla R (2006) Image-to-patient registration techniques in head surgery. Int J Oral Maxillofac Surg 35(12):1081–1095

    Article  CAS  Google Scholar 

  5. Soteriou E, Grauvogel J, Laszig R, Grauvogel TD (2016) Prospects and limitations of different registration modalities in electromagnetic ENT navigation. Eur Arch Otorhinolaryngol 273(11):3979–3986

    Article  Google Scholar 

  6. Kristin J, Mucha D, Schipper J, Klenzner T (2012) Registration strategies for the application of the navigation system Fiagon at the lateral scull base. Laryngorhinootologie 91(5):306–310

    Article  CAS  Google Scholar 

  7. West JB, Fitzpatrick JM, Toms SA, Maurer CR Jr, Maciunas RJ (2001) Fiducial point placement and the accuracy of point-based, rigid body registration. Neurosurgery 48(4):810–817

    CAS  PubMed  Google Scholar 

  8. Shamir RR, Joskowicz L (2011) Geometrical analysis of registration errors in point-based rigid-body registration using invariants. Med Image Anal 15(1):85–95

    Article  Google Scholar 

  9. Lee SJ, Yang HJ, Choi MH, Woo SY, Huh KH, Lee SS, Heo MS, Choi SC, Hwang SJ, Yi WJ (2019) Real-time augmented model guidance for mandibular proximal segment repositioning in orthognathic surgery, using electromagnetic tracking. J Craniomaxillofac Surg 47(1):127–137

    Article  Google Scholar 

  10. Wang J, Shen Y, Yang S (2019) A practical marker-less image registration method for augmented reality oral and maxillofacial surgery. Int J Comput Assist Radiol Surg 14(5):763–773

    Article  Google Scholar 

Download references

Acknowledgements

The Netherlands Cancer Institute received a research grant from ATOS Medical Sweden, which contributes to the existing infrastructure for health-related, quality-of-life research in the Department of Head and Neck Oncology and Surgery of the Antoni van Leeuwenhoek Hospital. In addition, the research was partly funded by the Verwelius foundation (Huizen, the Netherlands).

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

All authors have contributed to the research described in this manuscript.

Corresponding author

Correspondence to R. L. P. van Veen.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Availability of data and materials

The authors are willing to share the data and material for further research purposes.

Code availability

The authors are willing to share the code and material for further research purposes.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Brouwer de Koning, S.G., Riksen, J.J.M., ter Braak, T.P. et al. Utilization of a 3D printed dental splint for registration during electromagnetically navigated mandibular surgery. Int J CARS 15, 1997–2003 (2020). https://doi.org/10.1007/s11548-020-02271-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-020-02271-3

Keywords

Navigation