Introduction

The dental sector is in continuous evolution towards the use of the latest technological advances [1,2,3,4,5]. Nowadays computer-aided design and manufacturing (CAD/CAM), which includes additive manufacturing (AM) techniques, are often used for the digital workflow in dentistry. Such technologies are used for the design of dental arches and appliances through data recording with intraoral scanners (IOS) [6, 7]. Furthermore, IOS was used in [8, 9] to acquire the dental images, which were designed to realize prosthetic dentistry. CAD/CAM technologies have also been proposed for the evaluation and planning of surgical interventions and for the design of dental devices [10,11,12]. Some of the most common examples are the design of teeth [13, 14] and orthopaedic devices [15] and the planning of dental movements with a high degree of accuracy [16, 17]. Schweiger et al. showed to what extent 3D printing can be used in dental laboratories and dental practices at present [18]. Similarly, 3D printing and AM can be applied to a wide range of medical applications [19]. This way, Abid Haleem et al. performed a review of the applications of 3D printing in bone tissue engineering [20], Genoa et al. studied the main factors that are critical for bioprinting in bone tissue engineering [21] and Daniela et al. proposed a novel synchronized dual bioprinting approach for mechanically and biologically improved substitutes for cartilage tissue engineering [22]. Finally, different additive manufacturing techniques applied in the design, diagnosis and planning of a locked intramedullary nail used in the diaphyseal femoral fractures were studied by Fernandes et al. [23].

The mechanical properties of biocompatible materials have to be taken into account to properly design dental devices with 3D printing technology. An inappropriate choice of material can lead to breakage or high deformation, causing poor performance of the manufactured device. In addition, the printing direction of the device has to be carefully selected due to its influence on the mechanical properties of the manufactured part.

CAD/CAM technologies have been boosted in recent years by the emergence of biocompatible materials for 3D printing in the dental sector [24]. In recent times, 3D printing has acquired fundamental importance in the design of devices and dental pieces. New biocompatible materials are appearing to be used with this technology and the accuracy and speed of the manufacturing process are continuously increasing. Seven AM categories have been identified according to the American Society for Testing and Materials (ASTM) Committee F42 on AM technologies. These are the following: stereolithography (SLA), material jetting (MJ), material extrusion or fused deposition modelling (FDM), binder jetting, powder bed fusion, sheet lamination, and direct energy deposition. The following printing technologies have been used in this paper: Polyjet, Selective Laser Sintering (SLS), Digital Light Processing (DLP), and Stereolithography (SLA) [25, 26]. The most commonly used AM technologies in dentistry nowadays are SLA and MJ due to the appearance of novel photopolymer and biocompatible materials and their high resolution [27,28,29].

The use of 3D printing is widely extended in the manufacturing of models for the production of invisible teeth aligners. In recent years, great advances have been made in biocompatible materials that allow the manufacturing of complete functional devices such as mandibular advancement devices, mouth guards, surgical splints, hearing aids, and cutting guides [30]. In this sense, the ASTM has already set technical standards for a wide range of applications, designs, materials and processes, terminology, and test methods related to additive manufacturing. In this paper, tests have been conducted according to general testing ASTM standards for traction and bending tests. According to these standards, the anisotropic behaviour has to be considered to have a better knowledge of the mechanical characteristics of a material. This anisotropy is especially important in 3D printing techniques, where parts are created through the superposition of layers of material. Therefore, different properties are expected in each direction.

Materials used in thermoformed invisible splints have been thoroughly studied. They are well-known for their mechanical behaviour over time [31,32,33,34]. However, this is not the case with new biocompatible materials for 3D printing. This paper aims to analyse the behaviour of a wide range of biocompatible materials used in different additive manufacturing technologies. To this end, their mechanical properties for different layer orientations have been determined using standardised tensile and bending tests. The analysis of the results can help producers select the most suitable material, printing system, and printing orientation according to the desired characteristics of the device to be manufactured.

The study of the change of the properties of the manufactured parts over time depending on the biocompatible materials utilized and the environmental conditions where they are used is out of the scope of this paper. It is worth mentioning that there are many studies referring to the different existing technologies, their precision in terms of layer thickness, and the influence of post-printing treatments [35,36,37]. Several works have also studied the effect of print orientation in the printing process on the mechanical properties of manufactured specimens. [38,39,40]. However, to the best of our knowledge, there is no information about the mechanical behaviour of materials used for the development of dental devices in which different additive manufacturing technologies, biocompatible materials, and layer orientation have been taken into account. These data are of great interest to researchers to select the most appropriate material and printing method to manufacture dental devices.

Materials and methods

In this paper, the main biocompatible materials existing on the market for 3D printing that can be used for dental applications have been selected. Thus, the materials tested in this work were one for milling technology (PMMA) and eight for different 3D printing technologies: one for Polyjet, two for Selective Laser Sintering (SLS), three for Digital Light Processing (DLP), and two for laser-based Stereolithography (SLA). These materials are certified biocompatible in compliance with ISO 10993-1:2010 [41]. This way, selected materials meet the requirements for dental applications in terms of cytotoxicity, sensitivity, and irritation according to this international standard. In the case of the milling technology, the initial workpiece was a cylinder with a diameter of 12 cm and a thickness of 40 mm. These dimensions made it impossible to manufacture a specimen in the thickness direction (Z). Therefore, only specimens manufactured along the X and Y directions were studied.

This way, the tested materials were classified according to their manufacturing technology. Table 1 summarizes the manufacturing technology, printer model, material commercial denomination and type, and the denomination throughout the paper of the materials used to carry out the tests in this work.

Table 1 Specimen manufacturing technologies, printer model, materials, and material denomination

The specimens were created with the material recommended by each manufacturer for their printer. Manufacturers indicate the printing parameters and optimize the performance of their printers according to the materials they supply. The combination of each printer and material provided specimens that showed different behaviour. For example, DLP and SLA are similar 3D printing techniques, although the former uses a digital projector screen to flash an ultraviolet light which cures the photopolymer resins, and the latter a directed laser. However, different mechanical behaviours of the specimens were observed from the results of the study.

Tensile tests and bending tests were conducted on a population of specimens of the aforementioned materials and manufacturing technologies. Tensile tests were performed following the ASTM D638 standard [42]. Bending tests were conducted according to the ISO 178:2019 standard [43], which is equivalent to ASTM D790 [44]. Differences between the two directions on the plane of the printer tray, X and Y, and the perpendicular direction to it, Z, were evaluated. Therefore, it was necessary to have specimens for each manufacturing direction according to the X, Y, Z axes. Figure 1 shows three different positions of a part on the printer tray. The red faces represent the printing layers in the three previously mentioned directions during the manufacturing process.

Figure 1
figure 1

Orientation of the specimens on the print tray

Five specimens of each material manufactured in the three printing directions were tested. This way, 30 specimens were tested for each material and 270 specimens should have been tested. However, as mentioned before, the PMMA specimens could only be manufactured in the X and Y directions. Furthermore, due to supply problems, specimens manufactured with SLA2 material were not evaluated in traction tests in the Y printing direction. Consequently, the results obtained with 255 specimens are included in this paper. As indicated before, specimens were supplied by the printer manufacturers. This way, specimens were created by the printer manufacturer with the material supplied and recommended for their printer. Manufacturers optimized their printer’s performance according to the indicated materials. The only common parameter in all cases was layer thickness, which was equal to 100 μm in all printed specimens.

A test bench designed by this research team was used to carry out this work. The bench is composed of a frame with a fixed and a mobile plate. Specimens were tested as shown in Figs. 2 and 3. The movable plate is the component used to apply the deformation to the specimens. The excitation to the parts to be tested is generated by a hydraulic cylinder controlled by a servo-valve. This way, a controlled displacement or force is transmitted to the movable plate. A high accuracy displacement sensor and a force sensor were used to measure de elongation and applied forces, respectively.

Figure 2
figure 2

Tensile test. a Specimen in the test bench. b Rupture of the specimen. c Specimen measurements

Figure 3
figure 3

Bending test. a Specimen in the test bench. b Maximum allowable specimen deflection. c Specimen, supports, and loading pin measurements

Tensile tests

In the tensile tests, a progressively increasing force was applied to the ends of the specimens by means of two holder clamps. The lower clamp remained fixed while the upper one moved at a constant speed of 0.1 mm/min (Fig. 2a). The applied force and the elongation were continuously recorded with a load cell and a linear variable differential transformer (LVDT), respectively. The tensile tests continued until the specimens broke (Fig. 2b). This way, the stress that the material could withstand before complete failure took place was measured.

Three key parameters to analyse material behaviour were obtained with these tests:

  1. 1.

    Maximum allowable stress.

  2. 2.

    Stress recorded at the moment of fracture, called rupture stress.

  3. 3.

    Deformation suffered by the specimen before fracture.

All the specimens were manufactured according to the ASTM D638 standard (see Fig. 2c). The dimensions of the specimens are shown inTable 2.

Table 2 Specimen dimensions for the tensile test according to ASTM D638-02a

Bending tests

In the bending tests, each specimen was supported at both ends as shown in Fig. 3a. Next, a progressively increasing force was applied to the midpoint of the top face of the specimen, producing continuous transversal displacement and causing the curvature of the sample (Fig. 3b). The tests continued until specimen fracture took place. The force acting on the material and the displacements of the midpoint were recorded simultaneously.

The specimens and the supports were manufactured according to the ISO 178 standard (Fig. 3c). The thickness (h), width (b) and length (l) of the specimens were 4, 10, and 80 mm, respectively. The dimensions of the supports and the loading pin were the following: the length between the supports (L) was 64 mm, the punch and supporting piece radius (R1 and R2) were 5 mm and the relief angle (A) was ≥ 5°.

In this case, four key parameters to analyse material properties were obtained:

  • 1. Maximum flexural stress.

  • 2. Stress recorded at failure in bending, called rupture flexural stress.

  • 3. Flexural modulus.

  • 4. Deformation suffered by the specimens before fracture in the transversal plane.

Results and discussion

A total of nine materials suitable to be used in the dental industry were tested. Eight of them are used in 3D printing technologies and one is used in milling technology. As mentioned before, the specimens were subjected to tensile and bending tests. Results are shown next.

Tensile test results

In these tests, total force applied to the specimens as well as elongation were recorded. The initial dimensions of the narrow section of each of the specimens were measured with a Vernier caliper. In addition, it was verified that all dimensional tolerances were fulfilled. Figure 4 shows a sample of the results obtained with a specimen of each material.

Figure 4
figure 4

Tensile test results with a specimen of each material

The total deformation of the specimens were recorded in these traction tests, i.e. the change in distance between grips were measured. These tests were conducted as an initial evaluation of the specimens. As we did not measure the deformations in the narrow section, we cannot include the strain in these figures. On one hand, it can be seen that SLA2 and PMMA materials had the highest values for allowable stress and rupture stress. On the other hand, the SLS1 material provided lower values for rupture and maximum allowable stress. Besides, SLS2 showed great tenacity in all directions, but its maximum allowable stress was medium–low. Interestingly, the SLS1 tenacity was high in the X and Y direction but very low in the Z direction. The average maximum allowable stress and rupture stress of the specimens of each material and printing direction as well as the maximum and minimum values obtained in the tests are shown in Figs. 5 and 6.

Figure 5
figure 5

Average maximum allowable stress for each material in the X, Y and Z printing directions

Figure 6
figure 6

Average rupture stress for each material in the X, Y and Z printing directions

It can be observed that SLA2 and PMMA reached the maximum allowable stress and rupture stress. Both materials showed brittle behaviour. There were no big differences between the three printing directions. DLP materials showed a quite homogeneous performance, being slightly better in the Z direction.

Tables 3 and 4 show the average results of maximum allowable stress and rupture stress obtained in tensile tests of specimens of each one of the materials in the three printing directions X, Y, Z as well as their standard deviations (Values are presented as means ± SD). As mentioned before, no results are available for the Y-axis of the SLA2 material due to lack of supply of the specimen and for the Z-axis of the PMMA material due to the impossibility of producing milled specimens in this direction.

Table 3 Maximum allowable stress (UTS) and standard deviation (SD) of the tensile test
Table 4 Average fracture stress (FS) and standard deviation (SD) of the tensile test

Concerning deformation, Table 5 shows the average elongation at maximum stress (EMS) and elongation at the rupture point (ERP) in the three printing directions. In brittle specimens, both values are quite similar, such as in DLP and SLA specimens. As shown in Fig. 4, the highest elongations were achieved with the SLS manufacturing technology.

Table 5 Elongation at maximum stress (EMS) and elongation at rupture point (ERP)

Bending test results

In these three-point flexural tests, total force applied to specimens as well as deflections of test specimens at midspan were recorded. With these data, flexural stresses and flexural strains were obtained. Figure 7 shows a sample of the results obtained in these flexural tests with a specimen of each material.

Figure 7
figure 7

Flexural test results with a specimen of each material

Figures 8 and 9 show the average maximum stress and rupture stress for each of the materials, as well as the maximum and minimum values obtained in these tests. The average values and standard deviations are shown in Tables 6 and 7.

Figure 8
figure 8

Average maximum flexural stress for each material in the X, Y and Z printing directions

Figure 9
figure 9

Average stress at maximum strain for each material in the X, Y and Z printing directions

Table 6 Average results of maximum stress and standard deviation for each manufacturing direction of the specimens
Table 7 Average results of stress at maximum strain and standard deviation for each manufacturing direction of the specimens

It should be noted that the material with the highest maximum stress and rupture stress in the bending tests was SLA2 and PMMA, having similar values in all printing directions. SLS materials reached low maximum stress values, but rupture did not take place before maximum measured strain. This way, the SLS material again proved to be the most ductile of all the tested materials.

Figure 10 and Table 8 show the flexural modulus (FM) of the materials. It can be seen that the SLA2 material provided a high flexural modulus value regardless of the printing directions. On the other hand, the SLS1 printing material had the lowest flexural modulus value of all the tested materials and printing technologies.

Figure 10
figure 10

Flexural modulus for each material in the X, Y and Z printing directions

Table 8 Average results of the flexural modulus and standard deviation values for each manufacturing direction of the specimens

Table 9 shows the bending deformation measured in these tests. The maximum strain was limited to 15%. In this table, XM, YM and ZM columns include the deformations measured at maximum stress with the specimens manufactured in the X, Y and Z printing directions, respectively. Similarly, XR, YR and ZR are the deformations measured at maximum strain. It can be seen that the DLP materials had relatively low deformation values in all printing directions. In contrast, the SLS1 and SLS2 materials had large deformation values in all printing directions at maximum stress and rupture stress values. These materials are among the most ductile ones in the selected population of specimens. As it was pointed out before, these materials did not break at maximum deformation in some tests. It is worth highlighting the behavior of Polyjet. The maximum strain reached with this material was very high in the X and Y printing direction. However, the maximum allowable strain in the Z direction was extremely low.

Table 9 Deformation at maximum stress (− M) and maximum strain (− R)

It can be seen that some materials seem to have a brittle behaviour, while others show a more ductile behaviour. In ductile materials, there is a significant deformation before rupturing, which alerts of an imminent fracture. This way, ductile materials can prevent pieces from breaking unexpectedly in the mouth. On the other hand, fragile materials break suddenly with very little elastic deformation and, therefore, without prior warning.

A further study of the test results can contribute to gaining a better insight into these materials. Thus, an analysis and a discussion of the results obtained from the conducted tensile and bending tests are carried out next.

Figures 11 and 12 show the maximum allowable stress vs flexural modulus and vs the stress at the maximum strain, respectively, in the bending tests in the three printing directions. Average values are marked with a ‘ + ’ symbol. In brittle materials, both values are quite similar and the average value is located close to the diagonal of the box. On the contrary, the further away from the average value from the diagonal is, the more ductile the behaviour of the material is. Furthermore, the length of the axes of the coloured ellipses represents the standard deviations of the measured values.

Figure 11
figure 11

Maximum allowable stress versus flexural modulus

Figure 12
figure 12

Maximum allowable stress versus stress at maximum strain

It can be seen that the highest maximum allowable stresses were reached with SLA2 and PMMA materials while the lowest values were reported with SLS1 specimens. SLS1, SLS2, Polyjet, and PMMA specimens showed highly repetitive results. According to these low standard deviation values, devices manufactured with this material will show a homogenous performance. On the contrary, SLA2 and DLP2 showed a higher dispersion of the results in the bending tests carried out in the X and Y direction. In these cases, an appropriate design of the devices taking into account the expected loads on the manufactured parts might reduce the effects of bending stresses.

Next, Fig. 13 shows strain at yield point vs maximum strain in the bending tests. Average values above the diagonal indicate a ductile behaviour of the materials.

Figure 13
figure 13

Yield strain versus maximum strain

It can be observed that Polyjet in the X and Y directions and SLS2 and SLS1 in all cases showed great deformation levels, being the most ductile materials. On the contrary, DLP materials showed lower admissible strain rates. In most cases, the dispersion of the yield strain values was low. However, high standard deviation values were observed for the maximum strain values. This way, SLA1 and DLP3 showed great dispersion in the measured values.

Materials such as PMMA, SLA2 and DLP2 have intermediate mechanical behaviour compared to the other materials, making them preferable to manufacture dental devices. Results show heterogeneous behaviour in the tested biocompatible materials.

Regarding the influence of printing direction, Table 10 shows the relative change in the percentage of Maximum Allowable Stress (MAS) measured in the Y and Z directions with respect to the values obtained in the X direction in the bending tests.

Table 10 Comparison of maximum allowable stresse s

Results suggest that the printing direction influences the results obtained in some cases. This way, specimens manufactured with Polyjet and SLS1 showed a degraded performance in the Z direction in traction tests. On the contrary, DLP specimens printed in the Z direction provided better results. Results obtained with SLS2, PMMA and SLA showed that the mechanical behaviour of the tested specimens was homogeneous regardless of the printing direction.

Table 11 shows that the estimated flexural modulus (FM) obtained from the tests is significantly affected by the printing directions. In some cases, the flexural modulus in the X direction was the highest one, such as Polyjet and SLS1. On the other hand, in DLP1, DLP3 and SLA1, the highest value of the flexural modulus was measured in the Z direction, being more than 10% higher compared to the other two directions.

Table 11 Comparison of flexural modulus values

It can be observed that, in some cases, the printing direction significantly changes the mechanical properties of the specimens. This fact might be due to a poor connection between the printing layers. Therefore, if a dental splint made of Polijet or SLS1 materials is printed in the Z-direction, the forces required for removal of the splint may cause it to break easily due to retention of the device in the dental arches. In consequence, it is recommended to print the devices in a direction in which the device is not strained to avoid this problem. However, if there are forces in all directions, the design must be carried out taking into account the manufacturing direction.

Finally, although the material is the same in some AM technologies, “acrylic resin,” the manufacturers use different percentages of compositions according to the technology used. Therefore, the results obtained in this paper are related to the materials provided by the printer manufacturers. However, the study of the influence of material composition is out of the scope of this paper.

In summary, when designing an intraoral device, it is advisable to take into account the efforts to which it will be subjected and their directions, as well as the deformations allowed for proper operation. With this information, the appropriate manufacturing technology and material can be selected.

Conclusions

This paper focuses on the influence of the manufacturing process and printing direction on the properties of parts manufactured with a wide range of biocompatible materials used in dentistry. Specimens of these materials have been characterized by conducting tensile and bending tests.

From the analysis of the obtained results, it can be concluded that:

  • Specimens manufactured with Stereolithography (SLA) and milling (polymethyl methacrylate PMMA) achieved high maximum allowable and rupture stress values. Particularly, SLA2 showed a high flexural modulus value in all printing directions.

  • Polyjet printing and Selective Laser Sintering (SLS) technologies led to force-deformation curves with low maximum stress and high deformation values.

  • The Digital Light Processing (DLP) technology showed intermediate and homogeneous performance compared with the other technologies regardless of the material used.

  • Printing direction significantly influences the mechanical properties of the manufactured specimens in some cases. This way, specimens manufactured with Polyjet and SLS1 showed a degraded performance in the Z direction. On the contrary, the performance of SLS2, PMMA, and SLA was similar in all printing directions.

Lastly, the final selected combination of biocompatible material and printing technology should take into account the value and direction of the expected loads on the manufactured devices. In general, the best materials for this type of application should have high toughness with a good balance between strength and elasticity, such as in the case of SLS2 in all printing directions and DLP3 and SLA2 in the X-direction. It should be noted that a dental device might not work properly when a permanent deformation has been caused. This work will contribute to gaining insights into the best possible combination of printing technology, printing direction, and biocompatible material for the manufacture of dental devices. Furthermore, cost and printing times should also be evaluated to select the most suitable combination of factors. Future works will focus on evaluating the change in the properties of the printed parts over time depending on the materials used and the environmental conditions in which they are used. This work will serve as a baseline to be compared with the results obtained in long-term testing. Furthermore, SEM analysis on some of the best-performing materials for analysing the type of failure, type of deformations and to analyse the microstructure of the specimens will also be tackled in future papers.