Optimal ablation settings of TactiFlex SE laser-cut irrigated-tip catheter: comparison with ThermoCool SmartTouch SurroundFlow porous irrigated-tip catheter

  • ORIGINAL RESEARCH
  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Background

The TactiFlex SE catheter (TFSE, Abbott) with a contact force (CF) sensor and a laser-cut irrigated-tip has recently become available but lacks a lesion quality marker. This study aimed to explore distinctions in lesion characteristics between the TFSE and the ThermoCool SmartTouch SurroundFlow catheter (STSF, Biosense Webster), which utilizes a porous irrigated tip, and to assess the most effective application settings for the TFSE.

Methods

Lesions were generated using varying settings of radiofrequency power (30–50 W), CF (10–20 g), application duration (10–40 s), and catheter orientation (perpendicular or parallel) in an ex vivo porcine model. Comparative analysis between the TFSE and STSF was conducted for lesion characteristics and incidence of steam pops using predictive models in regression analyses.

Results

Among 720 applications, the TFSE exhibited a significantly lower incidence of steam pops compared to the STSF (0.6% vs. 36.8%, P < 0.001). Moreover, coefficients of determination (R2) for the TFSE were higher than those for the STSF concerning lesion depth (0.710 vs. 0.541) and volume (0.723 vs. 0.618). The lesion size generated with the TFSE was notably smaller than that with the STSF under identical application settings. Additionally, to achieve a lesion depth ≥ 4.0 mm, the TFSE required an application duration 8–12 s longer than the STSF under similar settings.

Conclusions

The TFSE demonstrated a lower incidence of steam pops and superior predictability in lesion size compared to the STSF. However, the TFSE necessitated a longer application duration than the STSF to achieve an adequate lesion size.

Graphical Abstract

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

Data Availability

Data can be made available upon reasonable request.

References

  1. Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14:e275–444.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Nogami A, Kurita T, Abe H, Ando K, Ishikawa T, et al. JCS/JHRS 2019 guideline on non-pharmacotherapy of cardiac arrhythmias. J Arrhythm. 2021;37(4):709–870.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Ariyarathna N, Kumar S, Thomas SP, Stevenson WG, Michaud GF. Role of contact force sensing in catheter ablation of cardiac arrhythmias: evolution or history repeating itself? JACC Clin Electrophisiol. 2018;4:707–23.

    Article  Google Scholar 

  4. Lakkireddy D, Reddy YM, Atkins D, Rajasingh J, Kanmanthareddy A, et al. Effect of atrial fibrillation ablation on gastric motility: the atrial fibrillation gut study. Circ Arrhythm Electrophysiol. 2015;8:531–6.

    Article  PubMed  Google Scholar 

  5. Pappone C, Oral H, Santinelli V, Vicedomini G, Lang CC, Manguso F, et al. Atrio-esophageal fistula as a complication of percutaneous transcatheter ablation of atrial fibrillation. Circulation. 2004;109:2724–6.

    Article  PubMed  Google Scholar 

  6. Nakagawa H, Ikeda A, Govari A, Papaioannou T, Constantine G, Bar-Tal M. Prospective study using a new formula incorporating contact force, radiofrequency power and application time (Force-Power-Time Index) for quantifying lesion formation to guide long continuous atrial lesions in the beating canine heart. Circulation. 2013;128:A12104 (abstr).

  7. Hussein A, Das M, Riva S, Morgan M, Ronayne C, et al. Use of ablation index-guided ablation results in high rate of durable pulmonary vein isolation and freedom from arrhythmia in persistent atrial fibrillation patients: the praise study results. Circ Arrhythm Electrophysiol. 2018;11:e006576.

    Article  PubMed  Google Scholar 

  8. Matsumoto K, Kawano D, Sasaki W, Tanaka N, Narita M, et al. Detailed investigation of the lesion formation with a novel contact force sensing catheter with a mesh-shaped irrigation tip. J Arrhythm. 2023;39(2):166–74.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Calzolari V, De Mattia L, Indiani S, Crosato M, Furlanetto A, et al. In vitro validation of the lesion size index to predict lesion width and depth after irrigated radiofrequency ablation in a porcine model. JACC Clin Electrophisiol. 2017;10:1126–35.

    Article  Google Scholar 

  10. Moreno J, Quintanilla JG, Molina-Morua R, Garcia-Torrent MJ, Angulo-Hernandez MJ, et al. Morphological and thermodynamic comparison of the lesions created by 4 open-irrigated catheters in 2 experimental models. J Cardiovasc Electrophysiol. 2014;25(12):1391–9.

    Article  PubMed  Google Scholar 

  11. Wittkampf FHM, Nakagawa H, Foresti S, Aoyama H, Jackman WM. Saline-irrigated radiofrequency ablation electrode with external cooling. J Cardiovasc Electrophysiol. 2005;16(3):323–8.

    Article  PubMed  Google Scholar 

  12. Barkagan M, Leshem E, Rottmann M, Sroubek J, Shapira-Daniels A, Anter E. Expandable lattice electrode ablation catheter: a novel radiofrequency platform allowing high current at low density for rapid, titratable, and durable lesions. Circ Arrhythm Electrophysiol. 2019;12:e007090.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Leshem E, Zilberman I, Barkagan M, Shapira-Daniels A, Sroubek J, et al. Temperature-controlled radiofrequency ablation using irrigated catheters: maximizing ventricular lesion dimensions while reducing steam-pop formation. JACC Clin Electrophysiol. 2020;6(1):83–93.

    Article  PubMed  Google Scholar 

  14. Kumar S, Romero J, Stevenson WG, et al. Impact of lowering irrigation flow rate on atrial lesion formation in thin atrial tissue. J Am Coll Cardiol EP. 2017;3:1114–25.

    Google Scholar 

  15. Ikenouchi T, Takigawa M, Goya M, Martin CA, Yamamoto T, et al. Comparison of lesion characteristics using temperature-flow-controlled versus conventional power-controlled ablation with fixed ablation index. J Cardiovasc Electrophysiol. 2023;34(4):908–17.

    Article  PubMed  Google Scholar 

  16. Teres C, Soto-Iglesias D, Penela D, Jáuregui B, Ordoñez A, et al. Personalized paroxysmal atrial fibrillation ablation by tailoring ablation index to the left atrial wall thickness: the ‘Ablate by-LAW’ single-centre study-a pilot study. Europace. 2022;24(3):390–9.

    Article  PubMed  Google Scholar 

  17. Whitaker J, Karády J, Karim R, Tobon-Gomez C, Fastl T, et al. Standardised computed tomographic assessment of left atrial morphology and tissue thickness in humans. Int J Cardiol Heart Vasc. 2020;24(32):100694.

    Google Scholar 

  18. Platonov PG, Ivanov V, Ho SY, Mitrofanova L. Left atrial posterior wall thickness in patients with and without atrial fibrillation: data from 298 consecutive autopsies. J Cardiovasc Electrophysiol. 2008;19(7):689–92.

    Article  PubMed  Google Scholar 

  19. Hoffmann P, Diaz Ramirez I, Baldenhofer G, Stangl K, Mont L, et al. Randomized study defining the optimum target interlesion distance in ablation index-guilded atrial fibrillation ablation. Europace. 2022;22:1480–6.

    Article  Google Scholar 

  20. Taghji P, El Haddad M, Phlips T, Wolf M, Knecht S, et al. Evaluation of a strategy aiming to enclose the pulmonary veins with contiguous and optimized radiofrequency lesions in paroxysmal atrial fibrillation: a pilot study. JACC Clin Electrophysiol. 2018;4(1):99–108.

    Article  PubMed  Google Scholar 

  21. Chen S, Schmidt B, Bordignon S, Urbanek L, Tohoku S, et al. Ablation Index-guiled 50W ablation for pulmonary vein isolation in patients with atrial fibrillation: procedural data, lesion analysis, and initial results from the FAFA AI Hight Power Study. J Cardiovasc Electroophysiol. 2019;30(12):2724–31.

    Article  Google Scholar 

Download references

Acknowledgements

We thank the doctors, clinical engineers, and staff at the Cardiology Department of Chukyo Hospital for their excellent support during this study.

Funding

The authors did not receive support from any organization for the submitted work.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Satoshi Yanagisawa.

Ethics declarations

Ethical approval

This study was approved by the local institutional ethics committee of Chukyo Hospital.

Informed consent

Not applicable in this ex vivo study.

Conflict of interest

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Below is the link to the electronic supplementary material.

Supplementary file1 (DOCX 334 KB)

Rights and permissions

Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Suga, K., Kato, H., Akita, S. et al. Optimal ablation settings of TactiFlex SE laser-cut irrigated-tip catheter: comparison with ThermoCool SmartTouch SurroundFlow porous irrigated-tip catheter. J Interv Card Electrophysiol (2024). https://doi.org/10.1007/s10840-024-01797-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1007/s10840-024-01797-9

Keywords

Navigation