Document Reviewers: Serge Boveda, MD, PhD; Michael R. Gold, MD, PhD, FHRS; Roberto Keegan, MD; Valentina Kutyifa, MD, PhD, FHRS, FESC, FACC; Chu-Pak Lau, MD, FHRS, CCDS; Mark A. McGuire, MBBS, PhD; Siva K. Mulpuru, MD, FHRS, CCDS; David J. Slotwiner, MD, FHRS; William Uribe, MD, MBA, FHRS.

TABLE OF CONTENTS

Abstract………………………………………In this issue

Manufacturer-Specific Translation of ICD Programming Recommendations: Abbott (Formerly St. Jude Medical)………………………………………....In this issue

Manufacturer-Specific Translation of ICD Programming Recommendations: BIOTRONIK……………….In this issue

Manufacturer-Specific Translation of ICD Programming Recommendations: Boston Scientific……………In this issue

Manufacturer-Specific Translation of ICD Programming Recommendations: Medtronic…………………..In this issue

Manufacturer-Specific Translation of ICD Programming Recommendations: MicroPort CRM (Formerly LivaNova and Sorin Group)………………………………...In this issue

Appendix 1 Author disclosure table………….In this issue

Appendix 2 Reviewer disclosure table………..In this issue

1 Manufacturer-specific translation of ICD programming recommendations

The manufacturer-specific programming settings/choices set forth below are based on a compilation of clinical expertise and clinical trial data as reported in the 2015 HRS/EHRA/APHRS/SOLAECE Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing, of which this Appendix B is a part. These recommended settings/choices represent a diligent and good faith effort on the part of the writing committee to translate the consensus statement recommendations to device settings/choices for the four specified clinical issues/implantable cardioverter-defibrillator (ICD) therapies where the writing committee considered that there was sufficient consensus and supporting data to make recommendations intended to improve the safety, morbidity, and mortality profile of patients with these clinical issues/ICD therapies. They are the recommendations of the writing committee only. They do not represent the position or recommendations of HRS, EHRA, LAHRS (formerly SOLAECE), or APHRS, nor are they the manufacturer’s nominal settings or the precise programming tested during clinical trials of these devices, nor are they necessarily the settings/choices that would be recommended by the manufacturer. These recommended settings/choices are not applicable in all circumstances. As stated in the Introduction to the consensus statement, “The care of individual patients must be provided in context of their specific clinical condition and the data available on that patient.” Each treating physician must carefully consider the circumstances of their individual patient and determine whether these recommended settings/choices are appropriate to their patient’s circumstances.

1.1 Abbott (Formerly St. Jude Medical) *Settings that are not nominal are marked with an asterisk

figure a

2 Manufacturer-specific translation of ICD programming recommendations

The manufacturer-specific programming settings/choices set forth below are based on a compilation of clinical expertise and clinical trial data as reported in the 2015 HRS/EHRA/APHRS/SOLAECE Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing, of which this Appendix B is a part. These recommended settings/choices represent a diligent and good faith effort on the part of the writing committee to translate the consensus statement recommendations to device settings/choices for the four specified clinical issues/implantable cardioverter-defibrillator (ICD) therapies where the writing committee considered that there was sufficient consensus and supporting data to make recommendations intended to improve the safety, morbidity, and mortality profile of patients with these clinical issues/ICD therapies. They are the recommendations of the writing committee only. They do not represent the position or recommendations of HRS, EHRA, LAHRS (formerly SOLAECE), or APHRS, nor are they the manufacturer’s nominal settings or the precise programming tested during clinical trials of these devices, nor are they necessarily the settings/choices that would be recommended by the manufacturer. These recommended settings/choices are not applicable in all circumstances. As stated in the Introduction to the consensus statement, “The care of individual patients must be provided in context of their specific clinical condition and the data available on that patient.” Each treating physician must carefully consider the circumstances of their individual patient and determine whether these recommended settings/choices are appropriate to their patient’s circumstances.

2.1 BIOTRONIK *Settings that are not nominal are marked with an asterisk

figure b

3 Manufacturer-specific translation of ICD programming recommendations

The manufacturer-specific programming settings/choices set forth below are based on a compilation of clinical expertise and clinical trial data as reported in the 2015 HRS/EHRA/APHRS/SOLAECE Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing, of which this Appendix B is a part. These recommended settings/choices represent a diligent and good faith effort on the part of the writing committee to translate the consensus statement recommendations to device settings/choices for the four specified clinical issues/implantable cardioverter-defibrillator (ICD) therapies where the writing committee considered that there was sufficient consensus and supporting data to make recommendations intended to improve the safety, morbidity, and mortality profile of patients with these clinical issues/ICD therapies. They are the recommendations of the writing committee only. They do not represent the position or recommendations of HRS, EHRA, LAHRS (formerly SOLAECE), or APHRS, nor are they the manufacturer’s nominal settings or the precise programming tested during clinical trials of these devices, nor are they necessarily the settings/choices that would be recommended by the manufacturer. These recommended settings/choices are not applicable in all circumstances. As stated in the Introduction to the consensus statement, “The care of individual patients must be provided in context of their specific clinical condition and the data available on that patient.” Each treating physician must carefully consider the circumstances of their individual patient and determine whether these recommended settings/choices are appropriate to their patient’s circumstances.

3.1 Boston Scientific *Settings that are not nominal are marked with an asterisk

figure c

4 Manufacturer-specific translation of ICD programming recommendations

The manufacturer-specific programming settings/choices set forth below are based on a compilation of clinical expertise and clinical trial data as reported in the 2015 HRS/EHRA/APHRS/SOLAECE Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing, of which this Appendix B is a part. These recommended settings/choices represent a diligent and good faith effort on the part of the writing committee to translate the consensus statement recommendations to device settings/choices for the four specified clinical issues/implantable cardioverter-defibrillator (ICD) therapies where the writing committee considered that there was sufficient consensus and supporting data to make recommendations intended to improve the safety, morbidity, and mortality profile of patients with these clinical issues/ICD therapies. They are the recommendations of the writing committee only. They do not represent the position or recommendations of HRS, EHRA, LAHRS (formerly SOLAECE), or APHRS, nor are they the manufacturer’s nominal settings or the precise programming tested during clinical trials of these devices, nor are they necessarily the settings/choices that would be recommended by the manufacturer. These recommended settings/choices are not applicable in all circumstances. As stated in the Introduction to the consensus statement, “The care of individual patients must be provided in context of their specific clinical condition and the data available on that patient.” Each treating physician must carefully consider the circumstances of their individual patient and determine whether these recommended settings/choices are appropriate to their patient’s circumstances.

4.1 Medtronic *Settings that are not nominal are marked with an asterisk

figure d

5 Manufacturer-specific translation of ICD programming recommendations

The manufacturer-specific programming settings/choices set forth below are based on a compilation of clinical expertise and clinical trial data as reported in the 2015 HRS/EHRA/APHRS/SOLAECE Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing, of which this Appendix B is a part. These recommended settings/choices represent a diligent and good faith effort on the part of the writing committee to translate the consensus statement recommendations to device settings/choices for the four specified clinical issues/implantable cardioverter-defibrillator (ICD) therapies where the writing committee considered that there was sufficient consensus and supporting data to make recommendations intended to improve the safety, morbidity, and mortality profile of patients with these clinical issues/ICD therapies. They are the recommendations of the writing committee only. They do not represent the position or recommendations of HRS, EHRA, LAHRS (formerly SOLAECE), or APHRS, nor are they the manufacturer’s nominal settings or the precise programming tested during clinical trials of these devices, nor are they necessarily the settings/choices that would be recommended by the manufacturer. These recommended settings/choices are not applicable in all circumstances. As stated in the Introduction to the consensus statement, “The care of individual patients must be provided in context of their specific clinical condition and the data available on that patient.” Each treating physician must carefully consider the circumstances of their individual patient and determine whether these recommended settings/choices are appropriate to their patient’s circumstances.

5.1 MicroPort CRM (Formerly LivaNova and Sorin Group) *Settings that are not nominal are marked with an asterisk

figure e