Corneal power measurement with a new aberrometer/corneal topographer in eyes after small incision lenticule extraction for myopia

  • Original Paper
  • Published:
International Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To assess corneal power measurements obtained by the OPD SCAN III Topographer in eyes with prior myopic small incision lenticule extraction (SMILE) surgery.

Methods

Sixty untreated myopic eyes of sixty subjects and forty previous myopic SMILE surgery eyes of forty subjects were consecutively enrolled in the present study. Manifest refraction, OPD SCAN III and Pentacam HR were performed. Keratometric measurements assessed by OPD SCAN III—simulated keratometry, average pupil power and effective central corneal power (ECCP) were compared with mean keratometry (Km) obtained by Pentacam HR in the untreated group and the clinical history method (CHM) in the treated group.

Results

In the untreated group, no statistically significant differences were revealed between all corneal power measurements obtained with OPD SCAN III and Km. In the treated group, all the corneal power measurements were statistically different from the CHM except for the Haigis method and the Shammas method, while ECCP had a statistically but not clinically significant overestimation of 0.42 D with 95% limit of agreement (LOA) of − 0.81 D to 1.64 D. The three modified ECCP had better prediction performance with narrower 95% of LOA lying in (− 1.20, 1.20 D) (− 1.22, 1.23 D) and (− 0.90, 1.00 D), respectively.

Conclusions

The ECCP provided with OPD SCAN III could be used as an alternative option for the CHM after specific modifications in eyes with previous myopic SMILE surgery when the preoperative data are unavailable considering the narrowest agreement between the modified ECCP and the CHM. Otherwise, caution must be raised considering the wide LOA.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Seitz B, Langenbucher A (2000) Intraocular lens power calculation in eyes after corneal refractive surgery. J Refract Surg 16:349–361

    CAS  PubMed  Google Scholar 

  2. Hoffer KJ (1995) Intraocular lens power calculation for eyes after refractive keratotomy. J Refract Surg 11:490–493

    CAS  PubMed  Google Scholar 

  3. Haigis W (2003) Corneal power after refractive surgery for myopia: contact lens method. J Cataract Refract Surg 29:1397–1411

    Article  Google Scholar 

  4. Wang L, Booth MA, Koch DD (2004) Comparison of intraocular lens power calculation methods in eyes that have undergone LASIK. Ophthalmology 111:1825–1831

    Article  Google Scholar 

  5. Shammas HJ, Shammas MC, Garabet A, Kim JH, Shammas A, LaBree L (2003) Correcting the corneal power measurements for intraocular lens power calculations after myopic laser in situ keratomileusis. Am J Ophthalmol 136:426–432

    Article  Google Scholar 

  6. Wang L, Hill WE, Koch DD (2010) Evaluation of intraocular lens power prediction methods using the American Society of Cataract and Refractive Surgeons post-keratorefractive intraocular lens power calculator. J Cataract Refract Surg 36:1466–1473

    Article  Google Scholar 

  7. Pedersen IB, Ivarsen A, Hjortdal J (2015) Three-year results of small incision lenticule extraction for high myopia: refractive outcomes and aberrations. J Refract Surg 31:719–724

    Article  Google Scholar 

  8. Hansen RS, Lyhne N, Grauslund J, Vestergaard AH (2016) Small-incision lenticule extraction (SMILE): outcomes of 722 eyes treated for myopia and myopic astigmatism. Graefes Arch Clin Exp Ophthalmol 254:399–405

    Article  Google Scholar 

  9. Gyldenkerne A, Ivarsen A, Hjortdal JO (2015) Comparison of corneal shape changes and aberrations induced by FS-LASIK and SMILE for myopia. J Refract Surg 31:223–229

    PubMed  Google Scholar 

  10. Qian Y, Huang J, Zhou X, Hanna RB (2015) Corneal power distribution and functional optical zone following small incision lenticule extraction for myopia. J Refract Surg 31:532–538

    Article  Google Scholar 

  11. McAlinden C, Khadka J, Pesudovs K (2011) A comprehensive evaluation of the precision (repeatability and reproducibility) of the Oculus Pentacam HR. Invest Ophthalmol Vis Sci 52:7731–7737

    Article  Google Scholar 

  12. Guilbert E, Saad A, Elluard M, Grise-Dulac A, Rouger H, Gatinel D (2016) Repeatability of keratometry measurements obtained with three topographers in keratoconic and normal corneas. J Refract Surg 32:187–192

    Article  Google Scholar 

  13. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310

    Article  CAS  Google Scholar 

  14. Gale RP, Saldana M, Johnston RL, Zuberbuhler B, McKibbin M (2009) Benchmark standards for refractive outcomes after NHS cataract surgery. Eye (Lond) 23(1):149–152

    Article  CAS  Google Scholar 

  15. Savini G, Barboni P, Carbonelli M, Ducoli P, Hoffer KJ (2015) Intraocular lens power calculation after myopic excimer laser surgery: selecting the best method using available clinical data. J Cataract Refract Surg 41:1880–1888

    Article  Google Scholar 

  16. Pan C, Hua Y, Huang J, Tan W, Lu W, Wang Q (2016) Corneal power measurement with the dual Scheimpflug-placido topographer after myopic excimer laser surgery. J Refract Surg 32:182–186

    Article  Google Scholar 

  17. Savini G, Barboni P, Profazio V, Zanini M, Hoffer KJ (2008) Corneal power measurements with the Pentacam Scheimpflug camera after myopic excimer laser surgery. J Cataract Refract Surg 34:809–813

    Article  Google Scholar 

  18. Wei P, Wang Y, Chan TCY, Ng ALK, Cheng GPM, Jhanji V (2017) Determining total corneal power after small-incision lenticule extraction in myopic eyes. J Cataract Refract Surg 43:1450–1457

    Article  Google Scholar 

  19. Savini G, Barboni P, Zanini M (2007) Correlation between attempted correction and keratometric refractive index of the cornea after myopic excimer laser surgery. J Refract Surg 23:461–466

    Article  Google Scholar 

  20. Holladay JT, Hill WE, Steinmueller A (2009) Corneal power measurements using scheimpflug imaging in eyes with prior corneal refractive surgery. J Refract Surg 25:862–868

    Article  Google Scholar 

  21. Falavarjani KG, Hashemi M, Joshaghani M, Azadi P, Ghaempanah MJ, Aghai GH (2010) Determining corneal power using Pentacam after myopic photorefractive keratectomy. Clin Exp Ophthalmol 38:341–345

    Article  Google Scholar 

  22. Qian Y, Liu Y, Zhou X, Naidu RK (2017) Comparison of corneal power and astigmatism between simulated keratometry, true net power, and total corneal refractive power before and after SMILE surgery. J Ophthalmol 2017:9659481

    PubMed  PubMed Central  Google Scholar 

  23. Oh JH, Kim SH, Chuck RS, Park CY (2014) Evaluation of the Pentacam ray tracing method for the measurement of central corneal power after myopic photorefractive keratectomy. Cornea 33:261–265

    Article  Google Scholar 

  24. Naeser K, Savini G, Bregnhoj JF (2016) Corneal powers measured with a rotating Scheimpflug camera. Br J Ophthalmol 100:1196–1200

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Xiaohua Lei.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pan, C., Tan, W., Hua, Y. et al. Corneal power measurement with a new aberrometer/corneal topographer in eyes after small incision lenticule extraction for myopia. Int Ophthalmol 39, 2815–2824 (2019). https://doi.org/10.1007/s10792-019-01128-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10792-019-01128-2

Keywords

Navigation