Introduction

Coronary computed tomography angiography (CCTA) has emerged as the clinical diagnosis standard for coronary artery disease (CAD) in patients with stable chest pain [1]. Additionally, CCTA is recommended by the European Society of Cardiology (ESC) for the screening of chronic coronary syndrome in patients with an intermediate risk of CAD [2,3,4]. This non-invasive imaging modality has gained considerable popularity due to its very good sensitivity and negative predictive value (85–91% and 83–91%, respectively) for detecting coronary plaque and ruling out significant stenosis [5,6,7]. Moreover, emerging technologies in CCTA have the potential to facilitate personalized risk assessment through plaque characterization and cardiovascular metrics [8]. Bridging the gap between these advancements and clinical application warrants careful consideration of the drawbacks of iodinated contrast medium (CM) administration against the benefit of CCTA. Although the incidence of post-contrast acute kidney injury (PC-AKI) in patients is controversial, patients with CAD often exhibit risk factors for PC-AKI, such as estimated glomerular filtration rate (eGFR) < 30 ml/min/1.73 m2, severe heart disease, diabetes mellitus, dehydration, and use of nephrotoxic medication [9,10,11]. Additionally, CM contributes to an increasing worldwide environmental risk by contamination of drinking water sources [12]. Decreasing CM dose combines goals for reducing patient risk, costs, and the ecological footprint associated with pharmaceutical waste. Therefore, the most recent guidelines of the Society of Cardiovascular Computed Tomography (SCCT) recommend reduced tube voltage acquisitions or dual-energy acquisitions with virtual mono-energetic image (VMI) reconstructions for CCTA examinations [13].

One of the available dual-energy or spectral imaging techniques is detector-based dual-layer CT (DLCT), which enables VMI reconstructions across a wide energy spectrum ranging from 40 to 200 kiloelectron volt (keV). Attenuation of iodinated CM increases at low energy levels (< 70 keV), approaching the k-edge of iodine at 33.2 keV, and thereby improving vascular enhancement through the photoelectric effect. Van Hamersvelt et al. demonstrated a potential reduction in CM dose by 40–60% at maintained image quality through the utilization of low VMI reconstructions in-vitro [14]. This has been verified by several other phantom and clinical studies [15,16,17,18,19].

By lowering the tube voltage from 140 to 120 kilovolt peak (kVp), vascular contrast enhancement is increased by 20–25% [20, 21]. Besides facilitating CM reduction, the use of lower tube voltages also results in a reduction of radiation dose [14, 19, 22]. However, obtaining diagnostic images in obese patients may be challenging, as the presence of increased adipose tissue leads to higher photon attenuation, adversely affecting contrast-to-noise ratios (CNR) and noise levels [23]. Therefore, patient size must be taken into careful consideration when considering optimal VMI levels and tube voltages for diagnostic cardiac imaging.

Similarly, elevated heart rate (HR) is a known cause of motion artifacts and reduced image quality in CCTA. Despite the potential benefits of CM dose reduction and the impact of patient characteristics on image quality, to the best of our knowledge, no study has systematically evaluated the combined impact of patient size, tube voltage, and HR on CM dose reduction for VMI reconstructions [16]. These findings may guide clinicians in tailoring imaging protocols based on indivdual patient characteristics, thereby optimizing both diagnostic accuracy and patient safety.

Therefore, our aim was to systematically assess VMI levels that achieve comparable CNR for DLCT at 50% CM-dose at different tube voltages, simulated patient sizes, and HR compared to the reference protocol (120 kVp, conventional reconstruction, 100% CM dose).

Materials and methods

Phantom design and setup

An anthropomorphic thorax phantom (QRM-Thorax, PTW GmbH) consisting of a spine insert, artificial lungs, and soft tissue-specific material was used for this study. Three patient sizes were simulated using two fatty tissue-specific extension rings (QRM-Extension Ring M and L, PTW GmbH) to increase the outer diameter of the phantom model from 300 × 200 mm with no ring (small patient size) to 350 × 250 mm with a medium ring (average patient size) and 400 × 300 mm with a large ring (large patient size), Fig. 1a [24].

Fig. 1
figure 1

Overview anthropomorphic thorax phantom. a) An anthropomorphic thorax phantom was used to simulate three patient sizes using various extension rings of fat−equivalent material (1). The phantom included a fillable water compartment (2) with a hollow artificial coronary artery (3) filled with iodinated contrast material at different dilutions. A computer−controlled motion unit (4) moved the artificial artery at various velocities along the direction indicated by the yellow arrow. The velocities correspond to different heart rates, and the image acquisition was synchronized to the robotic arm movement cycle by using a simulated electrocardiogram (ECG) signal output (5) of the motion unit as ECG input for the CT. b) The artificial artery contains solid water sections (red), and five hydroxyapatite calcifications (grey). All measurements were conducted in the middle of the solid water region between the two leftmost calcification rings (blue line). At the center of the artificial artery, a lumen of 5 mm in diameter was present throughout the phantom, which was filled with two contrast media concentrations

At the center of the thorax phantom, we positioned a water compartment containing a hollow artificial coronary artery of 5 mm diameter, Fig. 1b. The lumen of this artificial artery was filled with iodinated CM (Omnipaque 350; GE Healthcare) at two different dosages, resulting in an approximate attenuation of 400 and 200 Hounsfield units (HU) at 100% and 50% CM dose at 120 kVp and a small patient size, respectively [25,26,27]. A computer-controlled motion unit (QRM-Sim2D, PTW GmbH) was used to move the artificial coronary artery within the water-filled compartment in the horizontal plane perpendicular to the scan direction at constant velocities of 0, 10, 20, and 30 mm/s, which are average in-vivo coronary movement velocities at HR of 0, < 60, 60–75, and > 75 beats per minute (bpm), respectively [28]. To ensure linear motion, image acquisition was synchronized to the robotic arm movement cycle by using a simulated electrocardiogram (ECG) signal output of the motion unit as ECG input for the CT.

Data acquisition and reconstruction

All data was acquired on a DLCT (Spectral CT7500, Philips Healthcare) using a clinical protocol for CCTA, Table 1. In line with vendor recommendations, a sequential scan mode was used for HR of 0, < 60, and 60–75 bpm, while a low pitch helical acquisition with retrospective gating was performed for > 75 bpm. Each axial data acquisition covered a 360° angular range and offered a z-coverage of the reconstruction volume just under the 8 cm detector width. Image reconstruction was performed using an aperture-weighted cone beam approach with a slice thickness of 0.9 mm and an increment of 0.45 mm. For each combination of HR and patient size, two parameters were adjusted: tube voltage (120 and 140 kVp) and CM dose (100% and 50%). For each of these combinations of HR and patient size, the conventional acquisition at 120 kVp with 100% CM dose was defined as the reference. All scans with 50% CM dose were defined as experimental to improve readibility. The phantom was scanned five times for each combination of CM dose, tube voltage, HR, and patient size, with a small translation and rotation of the phantom between each scan. In addition to the conventional reconstruction, VMI levels were reconstructed from 40 to 70 keV at intervals of 5 keV.

Table 1 Acquisitions and reconstruction parameters

Data analysis

To assess the added value of reduced VMI levels, CNR was assessed from all reconstructions using a validated Python script (version 3.8) [29]. Measurements were exclusively confined to the solid water section of the phantom, specifically the most centrally located slice between the two leftmost and least dense calcification rings, Fig. 1b. The mean and standard deviation (SD) was calculated in a ROI at the center of the lumen, Fig. 2. A simple threshold was used to discriminate background material (water) from CM in the lumen. The value of the threshold was the mean of the background signal plus 3.5 times the SD.

CNR was defined as:

$$\text{C}\text{N}\text{R}=\frac{{\text{C}\text{T}}_{\text{l}\text{u}\text{m}\text{e}\text{n}}-{\text{C}\text{T}}_{\text{b}\text{a}\text{c}\text{k}\text{g}\text{r}\text{o}\text{u}\text{n}\text{d}}}{{{\upsigma }}_{\text{b}\text{a}\text{c}\text{k}\text{g}\text{r}\text{o}\text{u}\text{n}\text{d}}}$$
Fig. 2
figure 2

Summary of Image analysis. The threshold application on the original data (left, window setting: 750/90 HU (W/L)) is used to discriminate between background material (water) and contrast material (CM) in the lumen using a fully automatically calculated threshold based on the mean and standard deviation (SD) of the background signal (threshold = mean background + 3.5 times SD background) (middle). For each heart rate, the threshold in Hounsfield Units (HU) is indicated. The resulting contrast−to−noise ratio was based on the attenuation difference between the background and lumen, whereby the lumen attenuation was calculated within 60% of the total lumen (indicated in black in the rightmost column), to omit any edge or partial volume effects. Abbreviations: bpm: beats per minute, CM: contrast medium, HU: Hounsfield units, SD: standard deviation

with CTlumen the mean CT number (in HU) of the coronary artery, CTbackground the mean CT number of the background material, and σbackground the standard deviation of the mean HU of the background. The lumen was defined as a central ROI of 60% of the total lumen area to omit any edge effects, Fig. 2. For all acquisitions, CNR with 95% confidence intervals (CI) were calculated from the five repetitions for each patient size, HR, tube voltage, and CM dose. For each combination of patient size and HR, CNR was compared to the reference. Non-overlapping 95% CI of CNR were deemed significantly different. For each patient size, noise was compared to the reference (conventional reconstruction at 120 kVp, 100% CM dose, and at 0 bpm). Statistical analysis was performed with SPSS v26.0 (IBM SPSS version 26.0, IBM corporation).

Results

At the reference, the mean luminal CM attenuation [95% CI] was 411 [400, 423], 393 [374, 413], and 318 [281, 356] HU for small, medium, and large phantoms, respectively. At 50% CM, the corresponding luminal attenuation was 192 [166, 219], 201 [194, 207], and 193 [185, 202] HU. Attenuation, noise and CNR for all combinations of CM doses, tube voltages, VMI levels, patient sizes and HR are shown in Online Resource 1–4. Volumetric CT dose index volume (CTDIvol) values are listed in Table 1.

Attenuation

With decreasing VMI levels, attenuation increased for all patient sizes, HR, CM dilutions and tube voltage. For all experimental acquisitions (50% CM dose) at least one VMI level showed comparable attenuation to the reference. At 100% CM dose, a VMI at 60 keV showed the least difference in attenuation compared to conventional acquisition (Fig. 3).

Fig. 3
figure 3

Representative images of 100% and 50% contrast dose. The conventional (Conv.) image at 120 kVp and four virtual monoenergetic image (VMI) levels (70/90 HU (W/L)) are provided for a static lumen and large patient size. The reference result is indicated with an asterisk*. For all reconstructions, the resulting contrast−to−noise−ratio) is given in the right lower corner. Abbreviations: VMI: virtual monoenergetic image, Conv.: conventional, keV: kiloelectron volt, kVp: kilovolt peak

Noise

The mean noise level of the reference was 12 [11, 12], 15 [15, 16] and 19 [19, 20] HU for small, medium, and large patients, respectively. Mean noise was lower at 140 kVp and all VMI levels (Online Resource 1–4). VMI showed increasing degrees of noise with decreasing VMI levels for all patient sizes and HR, and for both contrast dosages and tube voltages. Experimental acquisitions at 40 keV showed higher mean noise than the reference, with experimental VMI at 65 keV possessing the most comparable noise levels to the reference.

Contrast-to-noise ratio

VMI showed increasing CNR with decreasing VMI levels for all patient sizes and HR, and for both contrast dosages and tube voltages (Figs. 4, 5 and 6). For all experimental acquisitions (50% CM dose), at least one VMI level showed comparable CNR to the reference, ranging from 40 to 65 keV, Table 2. VMI from 140 kVp tended to show slightly higher CNR than from 120 kVp, but not for large patients or HR > 75 bpm, although the differences were not significant (Figs. 4, 5 and 6). At 100% CM dose, VMI at 60 keV showed the least difference in CNR compared to the reference. For acquisitions < 60 bpm at 120 kVp with 50% CM dose, comparable CNR were found at VMI levels of 40, 40–45 and 40 keV for small, medium, and large patients, respectively (Table 2). As the HR increased to 60–75 bpm, those VMI levels shifted to 40–45 for small patients, 40–45 for medium patients, and 40–50 keV for large patients. Similarly, for HR > 75 bpm, comparable CNR were found at VMI levels of 40–60, 40–60, and 40–55 keV for small, medium, and large patients, respectively.

Fig. 4
figure 4

Graph of contrast-to-noise ratios (CNR) for a small patient size. Data are presented as mean and 95% confidence interval. The reference (conventional reconstruction, 120 kVp, 100% contrast dose) confidence interval is extrapolated with dotted line, to assess significant differences for all other combinations of tube voltage, contrast dose, and reconstruction. Results are shown for simulated heart rates of 0 (top left), <60 (top right), 60–75 (bottom left), and >75 (bottom right) beats per minute. Abbreviations: bpm: beats per minute, keV: kiloelectron volt, kVp: kilovolt peak

Fig. 5
figure 5

Contrast-to-noise ratios for an average patient size. Data are presented as mean and 95% confidence interval. The reference (conventional reconstruction, 120 kVp, 100% contrast dose) confidence interval is extrapolated with dotted line, to assess significant differences for all other combinations of tube voltage, contrast dose, and reconstruction. Results are shown for heart rates of 0 (top left), <60 (top right), 60–75 (bottom left), and >75 (bottom right) beats per minute. Abbreviations: bpm: beats per minute, keV: kiloelectron volt, kVp: kilovolt peak

Fig. 6
figure 6

Contrast-to-noise ratios for a large patient size. Data are presented as mean and 95% confidence interval. The reference (conventional reconstruction, 120 kVp, 100% contrast dose) confidence interval is extrapolated with dotted line, to assess significant differences for all other combinations of tube voltage, contrast dose, and reconstruction. Results are shown for heart rates of 0 (top left), <60 (top right), 60–75 (bottom left), and >75 (bottom right) beats per minute. Abbreviations: mAs: milliampere−seconds, bpm: beats per minute, keV: kiloelectron volt, kVp: kilovolt peak

Table 2 keV equivalent (range) compared to the heart rate category specific conventional reference (120 kVp; 100% contrast medium) per stratification

Discussion

This phantom study assessed the CNR in a reduced CM dose dual-energy CCTA protocol, considering different simulated patient sizes, VMI levels, kVp settings, and HR. Notably, a VMI level of 40 keV facilitated CM dose savings of up to 50% with comparable CNR for all combinations of the evaluated HR and patient sizes.

Additionally, we observed a consistent increase in CNR with decreasing VMI levels for all acquisitions. This indicates that the attenuation of iodine increases relatively more than background noise. However, it is essential to recognize that this effect may vary between different scanner vendors, necessitating protocol optimization for individual scanners [14].

Our findings align well with previous clinical and phantom studies that demonstrated similar CM reduction in dual-energy CCTA while maintaining sustained CNR [16, 30,31,32,33]. Oda et al. studied the effects of VMI using a DLCT to reduce the CM dose by 50% in 60 patients referred for CCTA [30]. Notably, 50% of the included patients had renal insufficiency. The authors reported a significant increase in mean coronary attenuation and visual scores, while no significant differences were noted in other aspects of visual image quality at a VMI of 50 keV when compared to conventional 120 kVp. Similarly, Raju et al. included 102 patients who were randomized to conventional CCTA (n = 53) and received 80 ml CM (120 kVp if BMI > 30, 100 kVp if BMI < 30), or fast kVp-switching dual-energy computed tomography (DECT) (n = 49) using 35 ml CM and 60 keV VMI reconstruction. The study reported a comparable signal-to-noise ratio, CNR, and rate of diagnostic interpretability, but inferior image quality Likert scores of the DECT cohort [33]. It is unclear if the > 50% contrast reduction combined with a relatively high VMI of 60 keV might have affected these inferior image quality scores. To the best of our knowledge, no previous clinical studies have stratified patients based on patient size or HR, and no phantom studies have incorporated a moving coronary artery to simulate different HR. Thus, our phantom study provides novel insights, suggesting that dual-energy CCTA with half CM dose at low VMI (40 keV) reconstruction show comparable image quality to conventional CCTA in most settings. In clinical practice, the potential for CM reduction is limited in some patient groups as a VMI level of 70 keV or higher is considered optimal for the evaluation of plaques and late enhancement, or in the precense of stents and implants. Specifically, our study did not incorporate cardiovascular disease, and, therefore, whether 40 keV provides sufficient image quality in intermediate risk patients with low calcium scores, requires further investigation.

Dual-energy acquisitions at higher tube voltages are expected to increase penetration of high-energy photons into the lower detector layer and improve spectral separation and sensitivity [34, 35]. Although few studies have compared dual-energy CCTA at different tube voltages, acquisitions at 140 kVp have demonstrated higher agreement with invasive coronary angiograms [36], albeit at the cost of a higher radiation dose. Similarly, a reduced 100 kVp CCTA protocol has been proposed to improve iodine attenuation, enabling both CM dose and radiation dose reduction [37]. The PROTECTION VI study reported a median radiation dose reduction of 50%, and of CM volume by 13% for tube voltages of 90–100 kVp when compared to conventional 120 kVp imaging. However, at the time of data acquisition of this study, our scanner could not acquire spectral data at 100 kVp. Consequently, further research initiatives are needed to evaluate the effects of additional tube voltage reduction.

Our study presents a potential for substantially reducing CM dose, thereby potentially mitigating health impact on patients. Moreover, given that via urine excretion CM enters the ecosystem, elevated CM levels have been detected in drinking water in several countries, therefore moderating CM usage is becoming ever more relevant [12]. Although CM are metabolically stable, they have shown to be degradable, reactive, and unstable under normal environmental conditions, posing potentially unwanted and unexpected ecological effects [38].

Several limitations may have influenced the results obtained in this study. First, the linear two-dimensional motion of the phantom in this study does not completely simulate a clinical situation, as in-vivo movements are three dimensional, and much more complex. In addition, the low VMI reconstructions used in this study might also affect subjective image quality by introducing blooming artifacts, for example in segments with calcifications. As the current study utilized no cardiovascular disease, future studies should be leveraged in clinical practice. Second, HR were translated and approximated to motion velocities based on one single study [28]. However, the broad HR ranges used in this study warrant this limitation. Third, due to inherent vendor-specific differences, our results may not be valid for other scanners. Finally, it is noteworthy that, lowering VMI levels may introduce artifacts, thus the CNR reported in this study might not be fully representative for the image quality as a whole in a clinical setting and clinical testing will be required before implementation of low contrast protocols.

Conclusion

This comprehensive phantom study indicated that a detector-based DLCT image reconstruction at 40 keV may facilitate 50% CM dose reduction for various patient sizes and HR with equivalent CNR compared to conventional CCTA at 100% CM dose, although clinical validation is needed.