Abstract
Prosthesis-patient mismatch (PPM) occurs when the effective orifice area of the prosthesis is too small in relation to the patient's body surface area. There are few data available on the frequency and prognostic impact of PPM after transcatheter aortic valve implantation (TAVI). Our aim was to determine the prevalence of PPM and to investigate its association with medium-term clinical course of patients undergoing TAVI. We included 185 patients undergoing TAVI (79 ± 5 years, 49% male, 98% CoreValve) between April-2008 and December-2014. The effective orifice area (EOA) was determined by transthoracic echocardiography prior and after the procedure. We defined PPM as indexed EOA ≤ 0.85 cm2/m2 (severe PPM if ≤ 0.65 cm2/m2). All cause death, stroke and hospitalization for heart failure were considered as major clinical events. 45 patients (24%) showed PPM (severe 11 patients, 6%). PPM was associated with a higher EuroSCORE (OR 1.06, IC 95% 1.01–1.12, p = 0.03), body surface area ≥ 1.72 m2 (OR 3.58, IC 95% 1.30–9.87, p = 0.01) and small aortic annulus (OR 0.73, IC 95% 0.55–0.92, p = 0.03); and severe PPM with small prostheses size (OR 17.79, IC 95% 1.87–169.78, p = 0.012). The mean event-free survival was 34 ± 26 months. Patients with severe PPM showed lower rates of event free survival than the rest of the series (52% vs. 84%, p = 0.04) at 34 months follow up. In our series, PPM was present in a quarter of the patients after TAVI. Higher EuroSCORE, smaller prosthesis size, larger body surface area and smaller aortic annulus diameter were associated with PPM. Severe PPM was an independent factor associated with major events at medium-term follow up.
Similar content being viewed by others
References
Grube E, Schuler G, Buellesfeld L, Gerckens U, Linke A, Zickmann B et al (2007) Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding Core-Valve Prosthesis. J Am Coll Cardiol 50:69–76
Tamburino C, Capodanno D, Ramondo A, Petronio AS, Ettori F, Santoro G et al (2011) Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients. Circulation 123:299–308
Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG et al (2010) Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med 363:1597–1607
Smith C, Leon M, Mack M, Miller D, Moses J, Svensson L et al (2011) Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med 364(23):2187–2198
Kodali S, Williams M, Smith S, Lars G, Webb JG, Makkar RR et al (2012) Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med 366:1686–1695
Hahn R, Pibarot P, Stewart W, Weissman NJ, Gopalakrishnan D, Keane MG et al (2013) Comparison of transcatheter and surgical aortic valve replacement in severe aortic stenosis a longitudinal study of echocardiography parameters in Cohort A of the PARTNER trial (placement of aortic transcatheter valves). JACC 61(25):2514–21
Holmes DR, Mack MJ, Kaul S, Agnihotri A, Alexander KP, Bailey SR et al (2012) 2012 ACCF / AATS / SCAI / STS expert consensus document on transcatheter aortic valve replacement. Cathet Cardiovasc Interv 79:1023–1082
Avanzas P, Pascual I, Muñoz-García AJ, Segura J, Alonso-briales JH, Suárez de Lezo J et al (2017) Seguimiento a largo plazo de pacientes con estenosis aórtica grave tratados con prótesis autoexpandible. Rev Esp Cardiol 70(4):247–253
Lancellotti P, Pibarot P, Chambers J et al (2016) Recommendations for the imaging assessment of prosthetic heart valves: a report from the European Association of Cardiovascular Imaging endorsed by the Chinese Society of Echocardiography, the Inter-American Society of Echocardiography, and the Brazilian Department of Cardiovascular Imaging. Eur Heart J Cardiovasc Imaging 17:589–590
Kappetein AP, Head SJ, Genereux P, Piazza N, Van Mieghem NM, Blackstone EH et al (2012) Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document. Eur J Cardiothorac Surg 42:S45–S60
Dumesnil J, Pibarot P (2011) Prosthesis-patient mismatch: an update. Curr Cardiol Rep 13(3):250–257
Takagi H, Yamamoto H, Iwata K, Goto SN, Umemoto T (2012) A meta-analysis of effects of prosthesis-patient mismatch after aortic valve replacement on late mortality. Int J Cardiol 159(2):150–154
Chen J, Lin Y, Kang B, Wang Z (2014) Indexed effective orifice area is a significant predictor of higher mid- and long-term mortality rates following aortic valve replacement in patients with prosthesis-patient mismatch. Eur J Cardio-thoracic Surg 45(2):234–240
Takagi H, Umemoto T (2016) Prosthesis-patient mismatch after transcatheter aortic valve implantation. Ann Thorac Surg 101(3):872–880
Pibarot P, Weissman NJ, Stewart WJ, Hahn RT, Lindman BR, McAndrew T et al (2014) Incidence and sequelae of prosthesis-patient mismatch in transcatheter versus surgical valve replacement in high-risk patients with severe aortic stenosis: a PARTNER trial cohort-a analysis. J Am Coll Cardiol 64(13):1323–1334
Muñoz-García AJ, Muñoz-García M, Carrasco-Chinchilla F, Molina-Mora MJ, Rodríguez-Bailón I, Domínguez-Franco AJ et al (2013) Incidence and clinical outcome of prosthesis-patient mismatch after transcatheter aortic valve implantation with the CoreValve prosthesis. Int J Cardiol 167(3):1074–1076
Zorn GL, Little SH, Tadros P, Deeb GM, Gleason TG, Heiser J et al (2016) Prosthesis-patient mismatch in high-risk patients with severe aortic stenosis: a randomized trial of a self-expanding prosthesis. J Thorac Cardiovasc Surg 151(4):1014–1023
da Silva C, Sahlen A, Winter R, Bäck M, Rück A, Settergren M et al (2014) Prosthesis-patient mismatch after transcatheter aortic valve implantation: impact of 2D-transthoracic echocardiography versus 3D-transesophageal echocardiography. Int J Cardiovasc Imaging 30:1549–1557
Zoghbi WA, Chambers JB, Dumesnil JG, Foster E, Gottdiener JS, Grayburn PA et al (2009) Recommendations for evaluation of prosthetic valves with echocardiography and doppler ultrasound. A report from the American Society of Echocardiography’s guidelines and standards committee and the task force on prosthetic valves, developed in conjunction. J Am Soc Echocardiogr 22(9):975–1014
Quiñones M, Douglas P, Foster E, Gorcsan J, Lewis J, Pearlman A et al (2003) ACC/AHA clinical competence statement on echocardiography: a report of the American College of Cardiology/American Heart Association/American College of Physicians–American Society of Internal Medicine. Task force on clinical competence. JACC 41:687–708
Evangelista A, María Á, Gómez A, Durán RM, Yagüela MM, María J et al (2000) Guías de práctica clínica de la Sociedad Española de Cardiología en ecocardiografía. Rev Esp Cardiol 53(5):663–683
Lang RM, Bierig M, Devereux RB, Flachskampf FA, Foster E, Pellikka PA et al (2006) Recommendations for chamber quantification. Eur J Echocardiogr 7(2):79–108
Dubois D, Dubois E (1916) A formula to estimate the approximate surface area of height and weight be known. Arch Intern Med 17:863–871
Mesa D, Suárez de Lezo J, Alvarez-Ossorio M, Ruiz Ortiz M, Delgado Ortega M, León del Pino M et al (2011) Measurement of aortic valve annulus using different cardiac imaging techniques in transcatheter aortic valve implantation: agreement with finally implanted prosthesis size. Echocardiography 28:388–396
León C, Suárez de Lezo J, Mesa D, Pan M, Ruiz M, Delgado M et al (2011) Early development of leaks in the CoreValve percutaneous aortic valve prosthesis: echocardiographic assessment. Rev Esp Cardiol 64:67–70
Cejudo Díaz del Campo L, Mesa Rubio D, Ruiz Ortiz M, Delgado Ortega M, Pan M, Ojeda Pineda S et al (2012) Usefulness of transesophageal echocardiography during implantation of the CoreValve percutaneous aortic prosthesis: Influence of the learning curve. Catheter Cardiovasc Interv 80:964–971
Avanzas P, Munoz-Garcia A, Segura J, Pan M, Alonso-Briales J, Lozano I et al (2010) Percutaneous Implantation of the CoreValve(registered trademark) self-expanding aortic valve prosthesis in patients with severe aortic stenosis: early experience in Spain. Rev Esp Cardiol 63(2):141–148
Rodés-cabau J, Pibarot P, Suri RM, Kodali S, Thourani VH, Szeto WY et al (2014) Impact of aortic annulus size on valve hemodynamics and clinical outcomes after transcatheter and surgical aortic valve replacement insights from the PARTNER trial. Circ Cardiovasc Interv 7:701–711
Pibarot P, Clavel M, Dahou A (2015) Patient and procedure selection for the prevention of prosthesis-patient mismatch following aortic valve replacement. EuroIntervention 11:106–109
Lebera A, Eichingerc W, Rieberb J, Lieberc M, Schlegerc S, Ebersbergerb U et al (2013) MSCT guided sizing of the Edwards Sapien XT TAVI device: Impact of different degrees of oversizing on clinical outcome. Int J Cardiol 168(3):2658–2664
Samim M, Stella PR, Agostoni P, Kluin J, Ramjankhan F, Sieswerda G et al (2013) A prospective “oversizing” strategy of the Edwards SAPIEN bioprosthesis: results and impact on aortic regurgitation. J Thorac Cardiovasc Surg 145(2):398–405
Dayan V, Soca G, Stanham R, Lorenzo A, Ferreiro A (2015) Is patient–prosthesis mismatch a predictor of survival or a surrogate marker of co-morbidities in cardiac surgery? Int J Cardiol 190:389–392
Tzikas A, Piazza N, Geleijnse ML, Van Mieghem N, Nuis RJ, Schultz C et al (2010) Prosthesis-patient mismatch after transcatheter aortic valve implantation with the medtronic corevalve system in patients with aortic stenosis. Am J Cardiol 106(2):255–260
Nishimura R, Otto C, Bonow R, Mack M, Carabello B, McLeod C et al (2017) 2017 AHA/ACC focused update of the 2014 AHA/ACC guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association task force on clinical practice guidelines. Circulation 135(21):1–123
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors whose names are listed in the first page report no affiliation nor involvement in an organization or entity with a financial or non-financial interest in the subject matter or materials discussed in this manuscript.
Rights and permissions
About this article
Cite this article
León del Pino, M.C., Ruíz Ortiz, M., Delgado Ortega, M. et al. Prosthesis-patient mismatch after transcatheter aortic valve replacement: prevalence and medium term prognostic impact. Int J Cardiovasc Imaging 35, 827–836 (2019). https://doi.org/10.1007/s10554-018-01519-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10554-018-01519-z