Introduction

Osteoporotic fractures are a burden for the public health costing 37 billion Euros in EU and 16 billion Dollars in USA every year.6,24 Hip fractures occur when the hip load generated at the hip while falling exceeds its load bearing capacity, or strength.5 Hip strength is a function of femur morphology, bone mineral density (BMD) and bone architecture43,45,47 whereas the hip load experienced during a fall relates to the dynamics of the fall, the body size and shape, the stiffness of the hip, of its surrounding tissues and of the floor.4,46,49 Yet, the interdependent role of hip strength and fall dynamics in osteoporotic hip fractures is unclear.

According to the World Health Organization (WHO), the diagnosis of osteoporosis and the consequent clinical indication for treatment of patients at risk of fracture is typically based on Dual-Energy X-ray Absorptiometry (DXA) measurements of areal Bone Mineral Density (aBMD), which represents a surrogate of hip strength.8,26,27,28,51 However, the diagnosis of osteoporosis has shown 59% sensitivity and 75% specificity in identifying fracture from non-fracture cases,8 motivating research into advanced diagnostic methods. For example, finite-element calculation of hip strength, taking in input the 3D morphology and BMD distribution of the hip from computed tomography (CT) images, showed 6–7% improved accuracy over corresponding aBMD measurements alone.50 Other authors developed multivariate statistical models using a variety of parameters, such as BMD, age, history of fragility fracture, premature menopause, oral corticosteroid use and ethnicity, showing 83% sensitivity and 54% specificity.8,26 Therefore, a large fraction of people at risk of hip fracture are currently not getting preventive fracture treatment, while several other healthy individuals are being treated despite their moderate risk of hip fracture.

Biomechanical studies of hip fracture have most often focused either on hip strength or fall dynamics, separately. Several hip fracture studies used relatively low loading rates (impact speed ranging from 15 to 50 mm/s, with test performed in material testing machines)13,52 or high-energy dynamic tests (impact speed approximately 114 mm/s, with a drop tower driven by gravity acceleration)16,19,23 showing that hip strength is a function of BMD, morphology and loading rate. Some authors suggested that the highly variable hip bone microstructure observed in high-resolution CT images of osteoporotic femoral heads may play an important role in determining hip strength.2 Other studies4,32 focused specifically on the fall dynamics, showing that the hip load at touch down is a function of body anthropometry (height and weight), stiffness of the hip, the nearby soft-tissues, the pelvis and the floor. Yet, to the best of the authors’ knowledge, only a few studies have replicated hip fractures in vitro by accounting for variations of body fall dynamics and compliance of the soft tissues surrounding the hip,16,17,18,19 showing that the compliance of the soft tissues surrounding the hip is an important co-factor for the risk of fracture of the hip.19

Here, we hypothesize that variations of body anthropometry, conjointly with the bone strength, determine the risk of hip fracture. As such, we aimed at comparing, in a cohort of donors, the variation in impact energy due to differences in body anthropometry, to the variation of hip fracture energy during a simulated sideways fall. A novel drop-tower experiment was developed for varying the impact energy according to the donor anthropometry. Ten femurs from elderly donors spanning a large range of hip strength, body height, weight and Body Mass Index (BMI) were tested. The impact force, displacement and strain at the superior femoral neck were determined. Variations of impact energy were compared to variations in fracture energy, fracture load and to factors affecting hip strength such as aBMD, bone morphology and bone spatial distribution. Finally, the classification of osteoporosis based on aBMD (T-score) was compared between fracture and non-fracture cases in the experiment. False-positive and false-negative cases were analysed and discussed.

Material and Methods

Femur Specimens

Ten fresh-frozen human cadaver femurs, without the soft-tissues, were obtained from elderly Caucasian donors through a body donation program in USA (Science Care, Phoenix, USA), who collected written informed consent. Exclusion criteria included donors younger than 45, bone cancer and interruption of the normal ambulation for longer than one year before death. Both genders were accepted. The range of body height, weight and age at death were, respectively, 142–185 cm, 32–136 kg and 56–91 years (Table 1). Specimens were stored at – 20 °C at the Biomechanics and Implants Laboratory of Flinders University (Tonsley, South Australia, Australia) ensuring the mechanical properties of the specimens were properly maintained.29 Ethics clearance was obtained from the institutional Social and Behavioural Research Ethics Committee (SBREC 6380).

Table 1 Donor’s anthropometry and femoral dimensions.

The diaphysis of each femur was cut at 180 mm from the top of femoral head and the soft tissues were removed. The diaphysis was abducted by 8° and potted 55 mm deep in a cylindrical aluminium cup using dental cement (Vertex Self Curing, Vertex-Dental, The Netherlands). Two polished aluminium hemi-spherical caps were fixed to the femoral head and the greater trochanter using a 5 mm-thick layer of dental cement to distribute the load on the bone, prevent local crushing and minimize friction (Fig. 1). The specimens and the hammer were painted using a white-on-black speckle pattern.42 A white water-based homogeneous paint was sprayed as a background over the entire proximal femur and the hammer. A black water-based paint was used to create random speckles 4–15 pixels in size (1 pixel = ca. 0.05 mm), so that the speckle size was smaller than the facet size for the Digital Image Correlation (DIC) analysis (15–19 pixels).42

Figure 1
figure 1

On the left-hand side, the drop tower machine. The aluminium plate and the distal hinge joint (a), the low-friction xy table (b), the hammer equipped with the impact load cell (c) are displayed. On the right-hand side, a schematic representation of the process used for determining the drop height and mass including the jack-knife model (top), the energy balance at the hip (centre) and the schematic representation of the simulated impact on the femur (bottom).

Determination of Impact Energy, Drop Mass and Height

The drop tower mass and height were determined using the donors anthropometry and thickness of the soft-tissues pad surrounding the hip.7,16 The body was simulated as a two-link 45° jack-knife model. The total kinetic energy at touchdown was modelled as a function of the portion (seven-twentieth) of the body weight acting on the hip and the body height according to the equation by van den Kroonenberg et al.49:

$$E_{\text{b}} = \frac{1}{2} \times \left( {\frac{7}{20}{\text{BW}}} \right) \times \left( {2.72 \times \sqrt h } \right)^{2}$$
(1)

where BW is the patient’s body weight expressed in kilograms, h is the body height expressed in meters and Eb is the total energy of the impact expressed in Joules. The energy absorbed by the soft-tissue and pelvis (Ea) was determined as a function of the soft-tissue thickness (STT) by using the relationship by Robinovitch et al.46:

$$E_{\text{a}} = \frac{{\left( { - 3.4 + 1.69 \times {STT}} \right) \times E_{\text{b}} }}{140}$$
(2)

The soft-tissue thickness was estimated using the donors’ BMI and a regression equation based on a cohort of men and women (STT = 3.4 × BMI − 44.77) within 19–29 BMI range.7 Since the fall energy absorbed by the soft-tissue Ea necessarily tends to Eb as BMI increases, we assumed a fixed value of the energy absorbed by the soft tissues for BMI greater than 29. The energy taken by the femur (Ef) was then the difference between the total energy generated during the fall (Eb) and the energy absorbed by the soft tissue (Ea). The drop height of the hammer was calculated as follows:

$$H = \frac{{M/E_{\text{f}} }}{g}$$
(3)

where H is the dropping height of the hammer in meters, g the gravity acceleration (9.81 m/s2) and M the drop mass (7/20 BW).

The Drop-Tower Experiment

The drop-tower machine (Fig. 1) consisted of a base aluminium plate, a vertical beam and a slider mounted on low-friction ball-bearing rails. The hammer hosted a modular mass (5–50 kg) and was equipped with a uniaxial piezoelectric impact load cell (KPA MNC, Kelba, Australia). The load cell was plugged to a data logger (cDAQ-9188, National Instruments, USA). The specimen was fixed to the base of the drop tower machine through a hinge, which allowed the specimen to rotate freely in the quasi-frontal plane. The specimen diaphysis was abducted by 10° and internally rotated by 15° with respect to the loading axis, mimicking the average femoral pose during a fall on a side.1 The position of the distal constraint was adjusted to align the longitudinal axis of the impact load cell (hammer) to the greater trochanter. The cap of the femoral head was placed in the centre of a low-friction horizontal xy table.

The impact force on the greater trochanter was recorded at 20 kHz and low pass filtered using a zero lag 3rd-order recursive Butterworth filter (3 kHz cut-off frequency).20 The hammer position and the superior femoral neck surface were video-recorded at 20,000 fps using two high-speed cameras (Phantom v1212, Vision Research, USA). The drop mass was unfastened manually. The load data and DIC acquisitions were triggered using a photoelectric sensor placed immediately above the impact region. The displacement of the hammer and the strain on the superior surface of the femoral neck were measured from the recorded high-speed cameras images with a three-dimensional DIC algorithm (Istra4D, Dantec Dynamics, Denmark). Before each test, DIC calibration was performed using a calibration target provided by the manufacturer. The DIC procedure was optimized and validated earlier, providing a spatial resolution of 3 mm and a strain random error below 200 microstrain, using a 15–19 pixels facet size, 5 pixels grid spacing, a 5 × 5 contour smoothing and a validity quote of 55%.40,41

The peak force and the force at crack opening were obtained from the load cell measurements. The impact start time was defined as the point when the impact force raised over 10 N and the impact end time when the force profile reached the first minimum value after decreasing the force below 10% of the max force. The time at crack opening was visually identified from the high-speed video. The position of the hammer was tracked by the DIC, using the hammer position at touchdown (start time) as reference for the displacement measurement. The energy absorbed by the specimen was calculated as the integral of the force over the displacement. The total energy and the energy at the time of crack opening were calculated (Fig. 2).

Figure 2
figure 2

A representative load-displacement curve (specimen #2). The force at crack opening, the peak force, the total fracture energy (sum of the orange and striped portions) and the energy at crack opening (stripped portion) are displayed.

Clinical Classification of Fracture Risk

The hip aBMD and the corresponding T-score values were estimated from CT scans. In summary, specimens and a densitometry calibration phantom (QCT Pro, Mindways, USA) were scanned using a CT scanner (Optima CT660, General Electric, USA); the phantom comprised five dipotassium hydrogen phosphate (K2HPO4) samples (equivalent density range 58.88–157.13 mg cm−3). The grey-levels in the CT images were calibrated to equivalent voxel-by-voxel bone density levels in the phantom. The hip bone mass was calculated from the calibrated CT images, then projected on the frontal plane. The aBMD was calculated using the bone mass and the area of the hip projected in the frontal plane3 and then converted into equivalent T-score values using the regression equations by Khoo et al.31 Specimens were categorized into normal (T-score > − 1), osteopenic (− 1 > T-score > − 2.5) and osteoporotic (T-score < − 2.5).

Microstructural Imaging

The microarchitecture of the specimens was imaged using a high-resolution synchrotron-light micro-CT (SR-microCT, Australian Synchrotron, Clayton VIC, Australia) using an earlier ad-hoc imaging protocol.36 Briefly, the X-ray projection images were acquired using a dedicated detector (2560 × 2160 pixels) and off-axis scanning mode, resulting in scanned and reconstructed field of view equal to 145.7 × 145.7 × 131.4 mm at an isotropic voxel size of 0.03 mm.

Analysis of the Data

The ability of the new protocol to reproduce clinically relevant hip fractures was assessed, by comparing (1) the fracture pattern to the AO/OTA fracture classification38 and (2) the fracture mechanics (time to fracture, force at crack opening, peak force, principal tensile and compressive strain maps) to corresponding published values.12,17,23,52 We also evaluated tensile, compressive and shear strain maps at the location of the fracture onset in the high-speed video, to ascertain whether shear maps are a better indicator of fracture onset location than principal tensile and compressive strain as observed in an earlier study on fracture of metastatic vertebrae.39

As we hypothesize that variations of body anthropometry, conjointly with the bone strength, determine the risk of hip fracture, we expect the variation of the energy delivered to the femur while falling and the variation of the hip fracture energy to overlap. Over the ten specimens, the range and coefficients of variation (standard deviation/average) of the total impact energy due to changes in body anthropometry and of the hip fracture energy were compared.

The normality of the distributions was verified using a Shapiro-Wilk test. Pearson’s correlation was used to identify relationships between T-score, total impact energy, energy delivered to the femur, hip fracture energy and fracture load (Prism 8, GraphPad Software, USA). Statistical significance was set to 0.05.

Specimens at high and low risk of facture were categorized using the threshold T-score equal to − 2.5, below which patients are considered at high risk of hip fracture according to the current WHO guidelines. False-positive and false-negative cases were analysed by comparing the impact energy and T-score. The effect of hip morphology on fracture was assessed by comparing the lever arm between the two points of contact on the femoral head and the greater trochanter, using quasi-frontal projection images of the micro-CT scans. The presence of large bone microstructural defects, such as large bone voids,22 at the fracture onset location, was ascertained in the microstructural images by visualizing 5 mm thick quasi-frontal micro-CT slices through the fracture onset location using Drishti software.33

Results

Seven femurs out of ten fractured and displayed pertrochanteric and subcapital fracture patterns, according to the AO/OTA fracture classification.38 The peak force was 2984 ± 913 N (average ± standard deviation) (Table 2). The time to fracture onset and to peak force were 2.7 ± 0.9 ms and 4.1 ± 5.1 ms, respectively. Fractured specimens showed local tensile strains on the neck surface at the fracture onset location and compressive strains over a large portion of the superior neck exceeding 3% (Fig. 3). Of the specimens that fractured, the strain map at the location of fracture onset was available for three specimens (#1, #5, #8), and the maximum shear strain gradient was visually closer and more localized around the fracture onset location than the principal strains. For the non-fractured specimens, strain reached 0.3 – 0.5% both in tension and compression.

Table 2 Comparison between the classifications of fracture cases based on aBMD (T-score) and the observed fractures types.
Figure 3
figure 3

High-speed camera images with strain maps obtained by applying DIC. The fracture onset location (red cross, first column) and the crack pattern (orange dotted lines), the principal tensile (second column), the principal compressive (third column) and the shear strain prior to crack opening (fourth column). The images of specimen #6 and the strain maps for specimen #6 and #7 are not available.

The soft tissue absorbed the majority (122 ± 96 J) of the total impact energy (172 ± 75 J) reducing the energy delivered to the femur to 50 ± 34 J. The energy taken by the femur had a coefficient of variation of 69% (34 J/50 J), which is comparable to the coefficient of variation of the T-score in the same cohort (CV = 65%). For the seven fractured specimens, the energy taken by the femur at crack opening was 3.1 ± 3.2 J (CV = 103%) and the energy to fracture the femur was 18 ± 11 J (CV = 61%). For the three specimens that did not fracture, the energy taken at the time of peak force was 14 − 30 J. The T-score was significantly and positively correlated to the total energy (r = 0.65, p-value = 0.04). A positive trend was observed between the T-score and the energy absorbed by the soft tissues (r = 0.55, p-value = 0.10) while a negative trend between the T-score and the fracture energy (r = − 0.51, p-value = 0.30). No correlation was found between the T-score and the energy delivered to the femur (r = − 0.13, p-value = 0.72).

The aBMD in the specimens spanned the lower range for the Caucasian population (T-score = -2.3 ± 1.5). Specimens included six osteoporotic, three osteopenic and one normal femur. The presence of osteoporosis correctly classified fracture from non-fracture for 7 of the 10 cases. One normal specimen (#1, T-score = 0.77) did fracture (false-negative) under a 14 J impact energy (Eb = 244 J; Ea = 230 J; height = 163 cm; weight = 116 kg; BMI = 44). One osteopenic specimen (#2, T-score = − 1.03) fractured (false-negative) using a nominal impact energy equal to 72 J (Eb = 168 J; Ea = 96 J; height = 168 cm; weight = 78 kg; BMI = 28). One osteoporotic specimen (#9, T-score = − 3.59) did not fracture (false-positive) under 11 J impact energy (Eb = 174 J; Ea = 163 J; height = 152 cm; weight = 88 kg; BMI = 38). There was a variable distance between the two contact points on the femoral head with the xy table and the greater trochanter with the load cell across the false-negative and false-positive cases (Fig. 4). Specimen #1, which had a normal aBMD (T-score = 0.77), displayed a large subcortical void at fracture onset location. No evident signs of major subcortical voids were observed in specimens #2 and #9.

Figure 4
figure 4

The femur microarchitecture visualized by micro-CT (0.03 mm voxel size) shown as a 5 mm thick quasi-on a quasi-frontal cross section for the false-negative (#1 and #2) and false-positive (#9) cases. The location of the slice is displayed (yellow shaded region) superimposed to the high-speed images for DIC taken during the drop-tower experiment, together with the cortical fracture onset (red cross) and pattern (orange dotted lines). The loading condition of the impact is also schematically displayed in the microarchitectural images. The l indicates the lever arm between the impact load direction and the reaction direction, #2 had a larger lever arm compared to #1 and #9.

Discussion

The aim of this study was to develop a drop-tower testing protocol for replicating the dynamics of hip fracture in elderly adults to test the hypothesis that variations in body anthropometry and bone strength determine the risk of hip fracture. The variation of the hip impact energy caused by variation of body anthropometry was compared to that of hip fracture energy. The experimental fracture and non-fracture cases were analysed with respect to the classification of fracture and non-fracture cases using the T-score threshold (T < − 2.5, osteoporosis). We found that (a) the novel drop-tower protocol produces clinically relevant hip fractures; (b) the variation across individuals of impact energy at the hip while falling on a side overlaps with that of the hip fracture energy; and (c) variation in impact energy, bone morphology and the presence of major subcortical bone voids can limit the ability of the classification of fracture and non-fracture cases using only the T-score definition of osteoporosis. As such, body anthropometry contributes to determine the risk of hip fracture in osteoporosis. Nevertheless, bone morphology and major subcortical bone voids determine fracture load (strength) and energy, and may also co-contribute to the low sensitivity and specificity of the current clinical practice for diagnosing patients at high risk of hip fracture.

The hip fracture patterns were consistent with the AO/ATA hip fracture classification,38 and the superficial neck strains and timing in the experiment were consistent with earlier reports,17,23 confirming a realistic fracture dynamic in the experiment. Interestingly, the shear strain map, when available, better indicated the location of fracture onset than principal strain maps, possibly indicating a different failure mechanism48 than the opening in tension or crushing in compression most often used for fracture prediction.12,37 The fracture load (range 1230–3976 N) was in the lower range of that reported in the literature (1170–7601 N),12,17,23,52 likely because the specimens in the present study were in the lower range of aBMD distribution in the elderly population (aBMD: 0.4–1.0 mg/cm2) compared to those reported in the aforementioned studies (aBMD: 0.4–1.85 mg/cm2).12,17,23,52 Therefore, the present results apply to the part of the population most at risk of hip fracture.

The range of energy delivered to the femur (11–95 J) overlapped with that of the hip fracture energy (8.6–39 J). Variation of body anthropometry (body weight, height and BMI) caused a coefficient of variation of impact energy and load over the sample (CV = 61–103%) comparable to that of aBMD (CV = 61%). Therefore, variation of body inertia, determined by body anthropometry, is an important co-factor in determining the risk of hip fracture in elderly individuals. This is in agreement with earlier findings obtained with a different setup and protocol,17 hence showing the limit of fracture prediction based on aBMD alone. We also found that the majority of the fall energy can be absorbed by the soft-tissues and that the total fall energy and the energy absorbed by the soft-tissues are correlated with the aBMD, as an epidemiological study has shown.34 This result is in agreement with the known negative correlation between BMI and risk of fracture,14,21 the notion that the soft-tissue thickness plays a fundamental role in establishing the risk of fracture4,7,45 absorbing the largest part of the energy in the hip region16 and the effectiveness of hip protectors in preventing femur fractures.9,25 However, the energy delivered to the femur was not correlated to the T-score. As such, T-score and body anthropometry are independent factors that, conjointly, determine the risk of hip fracture.

There was a surprising agreement between the ability of the T-score to predict fracture in the present experiment (70% accuracy) and that reported in the clinical literature (59–75%).8,11,26,27,28,51 Among the three cases not correctly classified using the definition of osteoporosis, we observed a highly variable impact energy. Specimen (#1), showing normal aBMD, fractured under a low-energy impact (14 J). An osteopenic specimen (#2), also not considered at high risk of hip fracture according to the definition of osteoporosis, fractured but after being subjected to a high-energy impact (72 J); that specimen showed no major subcortical voids. The osteoporotic specimen (#9), which was expected to fracture under a low-energy impact (11 J), did not fracture. We also observed a major subcortical void (#1) and a variable moment arm (#1, #2, #9) between the impact force on the trochanter and the reaction force on the femoral head. Thus, these findings support the notion that microstructural weak points22,43 and hip morphology can influence the fracture risk.15,35 Therefore, it appears that multi-factorial models accounting for body anthropometry, bone mass, bone spatial distribution and morphology may help improving current classification techniques in osteoporosis.

One limitation of the present study is the relatively low number of specimens (n = 10), which may limit the generality of our findings. For example, the small size prevents from analysing differences in hip fracture dynamics due to donor gender, age, bone morphology and microstructure. More research is necessary to address this point. It is important to note that in the present study, the soft-tissue thicknesses were estimated using a regression equation, possibly raising concerns about how well the soft-tissue thickness was represented for each donor. However, studying the individual risk of fracture in each single donor was not an aim of this study. The present study provides the first experimental evidence of the effect of body anthropometry on hip fracture mechanics in a cohort of elderly people spanning the lower range of bone density and exhibiting a large variation in body anthropometry. We also show that the fracture and non-fracture cases that were misclassified by the definition of osteoporosis do show specific energy, bone morphological and microstructural features, hence, supporting earlier (and future) studies investigating these aspects. Another limitation resides in the single representative loading configuration.30 The effect of body anthropometry on the risk of hip fracture reported here may vary for different falling scenarios. Finally, the propensity to fall, which is a function of neuromotor health condition44 and the related potential protective effect of muscle contraction,10,16 determines the likelihood that a fall event or fracture will occur. The present study analysed the effect of hip strength, in relation to its morphology, bone density and the presence of large microstructural defects, such as large bone voids (as seen in microcomputed tomography), and fall dynamics (as a function of body anthropometry) independently from the propensity to fall. Further retrospective and prospective studies are necessary to demonstrate the validity of the conjoint use of body anthropometry and hip strength for fragility prediction in clinics.

In conclusion, body anthropometry is an independent co-factor in hip fracture dynamics and may help improving fracture prediction in osteoporosis.