Predictability of intraocular lens power calculation in eyes after phototherapeutic keratectomy

  • Clinical Investigation
  • Published:
Japanese Journal of Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To compare the predictability of intraocular lens (IOL) power calculation using several corneal power measurements in eyes that underwent phototherapeutic keratectomy (PTK).

Study design

Retrospective case series.

Methods

We reviewed the clinical charts of 42 eyes of 25 consecutive patients who underwent cataract surgery after PTK for granular corneal dystrophy or band keratopathy. IOL power calculations were performed using the SRK/T formula with four corneal power measurements [automated keratometry (AK) measured with a partial coherence interferometer, simulated keratometry (Sim K), true net power (TNP), and total corneal refractive power (TCRP) measured with a rotating Scheimpflug camera]; we determined the prediction error, absolute error, and percentage within ± 1.0 D of the targeted refraction, 1 month postoperatively.

Results

The prediction error in the TCRP group was significantly better than those in the AK, Sim K, and TNP groups. The absolute error was also significantly better than those in the AK and Sim K groups, but not significantly different from that in the TNP group. The percentages of within ± 0.5 and 1.0 D in the TCRP group were significantly higher than those in the AK and Sim K groups, but not significantly different from that in the TNP group.

Conclusions

The TCRP provides the highest predictability of IOL power calculation in post-PTK eyes. This result suggests that the use of the TCRP, rather than of conventional anterior keratometry, may be clinically helpful for improving the refractive accuracy of post-PTK eyes.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Rapuano CJ. Excimer laser phototherapeutic keratectomy. Int Ophthalmol Clin. 1996;36:127–36.

    Article  CAS  Google Scholar 

  2. Amano S, Oshika T, Tazawa Y, Tsuru T. Long-term follow-up of excimer laser phototherapeutic keratectomy. Jpn J Ophthalmol. 1999;43:513–6.

    Article  CAS  Google Scholar 

  3. Fagerholm P. Phototherapeutic keratectomy: 12 years of experience. Acta Ophthalmol Scand. 2003;818:19–32.

    Article  Google Scholar 

  4. Amm M, Duncker GI. Refractive changes after phototherapeutic keratectomy. J Cataract Refract Surg. 1997;23:839–44.

    Article  CAS  Google Scholar 

  5. Dogru M, Katakami C, Yamanaka A. Refractive changes after excimer laser phototherapeutic keratectomy. J Cataract Refract Surg. 2001;27:686–92.

    Article  CAS  Google Scholar 

  6. Jung SH, Han KE, Sgrignoli B, Kim TI, Lee HK, Kim EK. Intraocular lens power calculations for cataract surgery after phototherapeutic keratectomy in granular corneal dystrophy type 2. J Refract Surg. 2012;28:714–24.

    Article  Google Scholar 

  7. Yaguchi Y, Negishi K, Saiki M, Torii H, Tsubota K. Comparison of the accuracy of intraocular lens power calculations for cataract surgery in eyes after phototherapeutic keratectomy. Jpn J Ophthalmol. 2016;60:365–72.

    Article  Google Scholar 

  8. Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333–40.

    Article  CAS  Google Scholar 

  9. Savini G, Barboni P, Carbonelli M, Hoffer KJ. Comparison of methods to measure corneal power for intraocular lens power calculation using a rotating Scheimpflug camera. J Cataract Refract Surg. 2013;39:598–604.

    Article  Google Scholar 

  10. Kane JX, Van Heerden A, Atik A, Petsoglou C. Intraocular lens power formula accuracy: comparison of 7 formulas. J Cataract Refract Surg. 2016;42:1490–500.

    Article  Google Scholar 

  11. Reitblat O, Levy A, Kleinmann G, Lerman TT, Assia EI. Intraocular lens power calculation for eyes with high and low average keratometry readings: comparison between various formulas. J Cataract Refract Surg. 2017;9:1149–56.

    Article  Google Scholar 

Download references

Funding

The authors have no financial or proprietary interests in the materials presented herein.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kazutaka Kamiya.

Ethics declarations

Conflict of interest

R Yoneyama, none; K. Kamiya, Honorarium (STAAR Surgical, Carl Zeiss Meditec AG, Santen, Senjyu, AMO, Kowa, Otsuka); K. Iijima, None; M. Takahashi, None; N. Shoji, Honorarium (TOMEY, RE Medical, Kowa Pharmaceutical, Otsuka, Santen, Alcon, Senju, Pfizer), Grant (TOMEY, Kowa Pharmaceutical, Otsuka, Santen, Chuo Sangio, HOYA, Alcon, Alcon Pharmaceuticals, Senju, Ogura, Pfizer).

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Corresponding author: Kazutaka Kamiya

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yoneyama, R., Kamiya, K., Iijima, K. et al. Predictability of intraocular lens power calculation in eyes after phototherapeutic keratectomy. Jpn J Ophthalmol 64, 62–67 (2020). https://doi.org/10.1007/s10384-019-00699-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-019-00699-y

Keywords

Navigation