Characterization of Luminance and Color Properties of 6-MP Wide-Screen Displays

  • Published:
Journal of Digital Imaging Aims and scope Submit manuscript

Abstract

Luminance and color performance are routinely evaluated as part of acceptance testing of displays used in diagnostic radiology. Previous work has indicated that as some diagnostic liquid crystal displays (LCDs) increase in backlight hours (BLH), the luminance measured with an external luminance meter exceeds the luminance reported by the manufacturer’s built-in meter. The purposes of this work were as follows: first, to characterize several luminance and color performance characteristics for 23 Barco Coronis Fusion 6-MP MDCC 6230 color displays and, second, to provide initial data for a longitudinal study evaluating changes in luminance and color performance as BLH increase. Grayscale display conformance and maximum luminance were evaluated using a calibrated luminance meter and AAPM Task Group 18 test patterns, and agreement between target and measured luminance was calculated. Luminance uniformity was evaluated by calculating maximum luminance deviation. Color point and color uniformity were evaluated using a spectrophotometer, and the radial color distances between the corners and center of the display were calculated. Above 3 cd/m2, there was good agreement between the target and measured luminance. At the maximum luminance, the mean difference was less than 1 %. The mean maximum luminance deviation for these displays was 10.40 ± 2.38 %. Color point was observed to be very consistent between displays with mean values of u′ and v′ of 0.187 ± 0.002 and 0.474 ± 0.004, respectively. Among all displays, maximum radial color distance had a mean value of 0.003 ± 0.001. These data provide a baseline for the acceptance of future displays as well as for longitudinal studies of luminance and color performance.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Silosky M, Marsh RM: Constancy of built-in luminance meter measurements in diagnostic displays. Med Phys 40(12):121902, 2013

    Article  CAS  PubMed  Google Scholar 

  2. Smith T, Guild J: The C.I.E. colorimetric standards and their use. Trans Opt Soc 33(3):73–134, 1931

    Article  Google Scholar 

  3. American Association of Physicists in Medicine, “Assessment of display performance for medical imaging systems,” AAPM On-line Report No. 3 (AAPM, College Park, MD, 2005), see www.aapm.org

  4. Harrington RE: Effect of color temperature on apparent brightness. J Opt Soc Am 44(2):113–116, 1954

    Article  CAS  PubMed  Google Scholar 

  5. Walz-Flannigan A, Stekel S, Weber H, Lanners D, Jonsgaard R, Peterson T, Langer SG: Aging and Quality Control of Color LCDs for Radiologic Imaging. J Digital Imaging 24(5):828–832, 2011

    Article  CAS  Google Scholar 

  6. Pauli H: Proposed extension of the CIE recommendation on ‘uniform color spaces, color difference equations, and metric color terms. J Opt Soc Am 66(8):866–867, 1976

    Article  Google Scholar 

  7. Samei E: Technical and psychophysical considerations for digital mammographic displays. Radiographics 25(2):491–501, 2005

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michael Silosky.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Silosky, M., Marsh, R.M. Characterization of Luminance and Color Properties of 6-MP Wide-Screen Displays. J Digit Imaging 29, 7–13 (2016). https://doi.org/10.1007/s10278-015-9811-7

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10278-015-9811-7

Keywords

Navigation