Abstract
Purpose
This study compared the results of the new Sysmex PA-100 AST System, a point-of-care analyser, with routine microbiology for the detection of urinary tract infections (UTI) and performance of antimicrobial susceptibility tests (AST) directly from urine.
Methods
Native urine samples from 278 female patients with suspected uncomplicated UTI were tested in the Sysmex PA-100 and with reference methods of routine microbiology: urine culture for bacteriuria and disc diffusion for AST.
Results
The analyser delivered bacteriuria results in 15 min and AST results within 45 min. Sensitivity and specificity for detection of microbiologically confirmed bacteriuria were 84.0% (89/106; 95% CI: 75.6–90.4%) and 99.4% (155/156; 95% CI: 96.5–100%), respectively, for bacterial species within the analyser specifications. These are Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Enterococcus faecalis and Staphylococcus saprophyticus, which are common species causing uncomplicated UTI. Overall categorical agreement (OCA) for AST results for the five antimicrobials tested in the Sysmex PA-100 (amoxicillin/clavulanic acid, ciprofloxacin, fosfomycin, nitrofurantoin and trimethoprim) ranged from 85.4% (70/82; 95%CI: 75.9–92.2%) for ciprofloxacin to 96.4% (81/84; 95% CI: 89.9–99.3%) for trimethoprim. The Sysmex PA-100 provided an optimal treatment recommendation in 218/278 cases (78.4%), against 162/278 (58.3%) of clinical decisions.
Conclusion
This first clinical evaluation of the Sysmex PA-100 in a near-patient setting demonstrated that the analyser delivers phenotypic AST results within 45 min, which could enable rapid initiation of the correct targeted treatment with no further adjustment needed. The Sysmex PA-100 has the potential to significantly reduce ineffective or unnecessary antibiotic prescription in patients with UTI symptoms.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The World Health Organisation (WHO) has raised concern regarding the overuse or use of inappropriate antimicrobials [1, 2]. In Europe 80–90% of antibiotics are dispensed in primary care settings [3] and in the United States at least 28% of the antibiotic courses prescribed each year are unnecessary [4]. This facilitates and escalates the spread of antimicrobial resistance (AMR), which is one of the greatest challenges to global public health and threatens our ability to treat common infections [1]. This prompted the World Health Assembly to adopt a multi-faceted global action plan with 5 main strategic objectives, including optimising the use of antimicrobial agents and investing in new diagnostic tools [5].
Urinary tract infections (UTIs) are among the most common bacterial infections with more than 400 million cases worldwide in 2019, imposing a significant burden on global health systems [6]. These patients usually present at a general practice or outpatient setting. The treating clinician does not have immediate knowledge of the causative organism since antimicrobial susceptibility tests (AST) in a clinical microbiology laboratory require 2–3 days, or in some instances, urine samples are not sent for laboratory analysis. Therefore, empirical treatment with a broad-spectrum antimicrobial is typically initiated, which may be ineffective. Once urine culture results are available, treatment is rarely adjusted since the patient may have recovered or is not present anymore. These sub-optimal empirical treatment situations create the opportunity for both treatment failure and/or an increase in drug-resistant uropathogens within the population [7, 8].
To optimise targeted antibiotic treatment for patients with UTI, the uropathogen must be identified and AST performed. Culture-based microbiological diagnostic methods are labour intensive, time-consuming and require trained personnel. These drawbacks have prompted the development of several laboratory and point-of-care (POC) approaches to supplement the culture-based methods [9, 10] and to optimize immediate treatment decisions. Genotypic AST methods detect the presence of specific resistance genes and produce results faster than conventional microbiology, but they are expensive and require knowledge of the target genes. Therefore, phenotypic AST methods remain the gold standard since the actual growth of the microorganism is monitored while in the presence of the antimicrobial agent [11]. Efforts have been made to decrease the time to obtain a definitive AST result, specifically by utilising microfluidic or nanofluidic technologies [12,13,14,15]. However, a gap still exists for accurate, cost-effective and reliable POC tests implementing this technology, which could rapidly detect the uropathogen and provide physicians with a susceptibility profile. The latter might then be used for immediate targeted treatment, avoiding the need for empirical antimicrobial use and de-escalation.
This study evaluated the diagnostic accuracy and clinical utility of the novel automated POC Sysmex PA-100 AST System (PA-100). It is designed for use in uncomplicated UTI. It detects bacteriuria in 15 min using fresh native urine and, in positive samples, performs phenotypic AST in an additional 15–30 min by automated phase-contrast monitoring of bacterial growth in the presence of an antimicrobial in Mueller–Hinton broth. This is the first time this technology has been assessed in a POC clinical setting for uncomplicated UTI.
Methods
Study design and patient population
This prospective, non-interventional study was conducted at Fundació Puigvert (FP) and Hospital de la Santa Creu i Sant Pau (HSP), Barcelona. Criteria for enrolment were female, non-pregnant patients, > 18 years old, presenting with current UTI symptoms in the urinary tract that had been present for < 7 days, with no antibiotic treatment within the previous 3 days, and who provided a clean-catch mid-stream urine sample. Patients with complicated UTI (e.g., anatomic abnormalities, renal disease, etc.) were excluded.
Informed consent was obtained from each patient. The study was conducted according to the Declaration of Helsinki and approved by the Ethics Committees of FP (13/07/2022) and HSP (07/09/2022). The study had no impact on the normal diagnostic and therapeutic procedures since clinicians were not informed of the PA-100 results during the study. Clinical information was obtained from personal interviews and electronic patient records.
Sample size calculation
The expected sensitivity and specificity for detection of bacteriuria was 90% for both parameters. Assuming a prevalence of 0.6, significance level of 0.05, a precision of 0.05 and a drop-out of 10%, the calculated sample size for detection of bacteriuria was 385 patients.
The expected overall categorical agreement (OCA) for the AST was > 90% [16]. For a precision of 10%, the number of resistant strains to be collected was < 100 samples for all antimicrobials.
Urine collection
Clean-catch midstream urine samples were collected and divided into 2 aliquots. In two cases samples were obtained from an in- and out-catheter. The first aliquot was processed within 30 min on the PA-100 (Sysmex Astrego, Uppsala, Sweden). The second aliquot was stored at 4 °C for < 4 h before reference method analysis.
Test methods
PA-100
The fresh native urine sample was mixed and 400 µl pipetted into a PA-100 cartridge, which was inserted into the instrument.
The PA-100 automatically processes the sample after cartridge insertion. During the analysis, the urine sample flows through a filter into a nanofluidic chip. Bacterial cells, if present, are trapped in individual channels, whereas eukaryotic cells and other urine components, larger than bacteria, are filtered and do not enter the nanofluidic chip. After this loading step, Mueller Hinton broth II flows through reservoirs where dried antibiotics are kept. The growth medium, now containing a constant amount of antibiotics, flows into the channels where bacteria are trapped. All substances are prefilled in the cartridge. The incubation starts and the flow is kept constant throughout the analysis. The process is controlled by pressure regulation in the cartridge.
The nanofluidic chip contains approximately 11,000 nanochannels, grouped into 32 sections. Each section, containing a few hundred individual channels, is exposed to a different antibiotic and concentration of antibiotic. These antibiotic concentrations are used to calculate the resistance profile, according to a proprietary algorithm, which has been calibrated against the EUCAST Clinical Breakpoint tables [17].
During the analysis, the cartridge is moved by a mechanical system, which positions the nanofluidic chip in the field of view of a phase contrast microscope. Images are taken of all 32 sections in approximately 1-min intervals. A proprietary image analysis algorithm creates videos of the recorded images and calculates the growth rate of each of the sections. This growth rate is compared to the control section (no exposure to antibiotics).
The system detects the presence of bacteriuria (≥ 50,000 CFU/mL) and displays AST results as S (susceptible), I (susceptible, increased exposure), R (resistant) or not applicable, for amoxicillin/clavulanic acid, ciprofloxacin, fosfomycin, nitrofurantoin and trimethoprim, according to EUCAST [17]. A positive bacteriuria is stated without informing the user about the presumptive species identified by the analyzer and used to generate the AST result. The user receives a positive / negative notification only. Other results are possible: E (technical error) and LG (low growth: growth rate of the bacteria in the sample too low for accurate rapid AST results). The cartridge is designed for detection of the most frequent species causing uncomplicated UTI: Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Enterococcus faecalis and Staphylococcus saprophyticus. An internal, non-disclosed proprietary algorithm is used to classify the bacteria into three different groups (Enterobacterales, Staphylococcus and Enterococcus) and to apply the relevant EUCAST breakpoints. Other non-fastidious bacteria might also be detected and displayed as positive. As the algorithm is not trained on those species, there is a chance that incorrect EUCAST breakpoints may be applied to bacteria uncommon in UTI potentially impacting those strains with resistance close to the breakpoint.
Furthermore, as S. saprophyticus and Enterococcus do not feature a breakpoint for oral fosfomycin, the analyzer will display “NA” for “not applicable” in the fosfomycin AST result if the algorithm concludes that the detected bacteria are members of the Staphylococcus or Enterococcus genera.
A more detailed description of the technology can be found elsewhere [14, 18,19,20,21].
Urinalysis
Urine cell counting was performed on an automated UN series Urinalysis flow cytometer analyser (Sysmex, Kobe, Japan).
Culture methods and analysis
Urine culture was initiated within 4 h of collection using 10 µL sterile disposable loops in Columbia Blood Agar (BA) (BD, Franklin Lakes, NJ, USA) and UTI urine chromogenic agar (ThermoFisher, Waltham, MA, USA) using the streaking technique. Also, 1 µL of urine was plated on an additional UTI chromogenic agar plate using the quantitative technique. Plates were incubated at 35 °C in air (UTI) or under 7% CO2 (BA) and read overnight and at 48 h.
The number of colonies on the quantitative agar plate was recorded as CFU/mL. Clinical-microbiological evaluation was made according to current standards [22]. Any clinically significant isolate was identified by VITEK2 (bioMérieux, Marcy-l’Étoile, France) or MALDI-TOF (Bruker, Billerica, MA, USA). Non-positive samples were classified as negative or contaminated by vaginal microbiota or another cause.
Samples were analysed for the presence of antimicrobials with a Bacillus licheniformis strain [23]. Any inhibition was recorded as probable presence of an antimicrobial in the sample.
Antimicrobial susceptibility tests
Susceptibility tests were performed by disc-diffusion on agar for all 5 antimicrobials [17] and interpreted according to EUCAST 2022 for uncomplicated UTI when available. Fosfomycin breakpoints for Enterococci are not provided by EUCAST and CLSI 2020 criteria were used instead [24]. For Enterococci trimethoprim susceptibility, a disc-diffusion ECOFF (Epidemiological cut-off value) diameter of 21 mm was used (EUCAST) and ampicillin susceptibility was used to test amoxicillin/clavulanic acid PA-100 results. MIC amoxicillin/clavulanic acid for Enterococci was done for any strain of this species resistant to this antimicrobial. Cefoxitin-deduced oxacillin susceptibility test results were used for amoxicillin/clavulanic acid in the Staphylococci disc-diffusion method (EUCAST). In addition, broth microdilution was performed retrospectively on frozen isolates for amoxicillin/clavulanic acid, ciprofloxacin, nitrofurantoin and trimethoprim using Sensititre (ThermoFisher) and agar dilution for fosfomycin (Liofilchem, Roseto degli Abruzzi, Italy). Amoxicillin/clavulanic acid breakpoint for Enterobacterales was the 2022 EUCAST value for uncomplicated UTI (16 mm diameter for a 20–10 µg disc). Fosfomycin and nitrofurantoin disc-diffusion results for all Enterobacterales were interpreted according to the breakpoints given by EUCAST for E. coli uncomplicated UTI.
Detection of bacteriuria by PA-100 in clinical samples
Analytical performance of the PA-100 was calculated twice. One analysis was aligned to the specifications in the instructions for use (IFU) of the device (true positive defined as PA-100 positive and growth of E. coli, K. pneumoniae, P. mirabilis, E. faecalis or S. saprophyticus strain in culture; true negative defined as PA-100 negative and reference negative; detailed criteria in supplementary Table 1). The second analysis assessed the clinical performance of the PA-100 for all encountered species in samples considered positive by the microbiologist [22]. Samples with a positive Bacillus licheniformis inhibitor test would have been excluded, however, no sample showed inhibition [23].
Potential interference by host cells
Host cells were quantified using an UF series automated urine flow cytometer. A logistic regression was used to investigate the impact of the presence of host cells on a false negative bacteriuria result. The analysis focused on false negative results as this study only identified one false positive bacteriuria result by the PA-100.
Comparison of PA-100 AST results with routine microbiology on clinical samples
PA-100 AST results were compared with disc-diffusion. For discrepancies, broth microdilution was used as adjudicator for amoxicillin/clavulanic acid, ciprofloxacin, nitrofurantoin, trimethoprim, and agar dilution for fosfomycin. Results were analysed twice, once for samples containing of E. coli, K. pneumoniae, P. mirabilis, S. saprophyticus or E. faecalis and once for all samples (detailed criteria in supplementary Table 1). PA-100 AST results with an NA, LG or E message were excluded. Analysis was done according to ISO20776-2:2021 [25], in addition, OCA was reported.
Comparison with current clinical routine treatment
To assess the potential clinical usefulness of the PA-100, the treatment decisions made for enrolled patients were recorded. If the clinician prescribed an antibiotic other than nitrofurantoin, fosfomycin, amoxicillin/clavulanic acid, cotrimoxazole, trimethoprim or ciprofloxacin, the results of routine microbiology were utilized to assess potential resistance.
The priority list of antimicrobials for the hypothetical outcome was based on the assumption that the most frequent pathogen would be E. coli and that the analytical AST performance in a clinical setting was known. Antimicrobial side effects were also considered. If an antibiotic was reported as R, E or LG, the next antibiotic in the priority list was assessed. The order of antibiotics considered was nitrofurantoin > fosfomycin > trimethroprim > amoxicillin/clavulanic acid > ciprofloxacin.
Statistical analysis
Analysis was performed using MedCalc statistical analysis software version 20.104 (Medcalc Software, Ostend, Belgium). Comparison of proportions was done by N-1 chi-squared test. When comparing proportions, the confidence interval (CI) was calculated according to the recommended method [26]. CIs for sensitivity, specificity and accuracy were “exact” Clopper-Pearson CI. CIs for the likelihood ratios were calculated using the “Log method”.
Results
Study population
A total of 307 patients were enrolled. Twenty-nine patients with complicated UTI were excluded retrospectively. The final study population consisted of 278 patients, 199 enrolled at FP between September 2022 and March 2023, and 79 enrolled at HSP between September 2022 and January 2023. Patient characteristics are summarized in Table 1.
Detection of bacteriuria
The performance of the PA-100 was compared to the microbiological reference (Tables 2 and 3). Under IFU specifications bacteriuria detection by the PA-100 in positive urine cultures was 89 out of 106 (sensitivity 84.0%, CI:75.6% to 90.4%). Of 94 samples positive for Gram-negative bacteria 85 were detected, while 4 out of 8 Gram-positive bacteria were detected. Culture negative urines, confirmed as negative by the PA-100, were 155 out of 156 (specificity 99.4%, CI: 96.5% to 99.8%). Overall accuracy was 93.1% (89.1% to 95.9%).
Considering all samples judged as positive by the microbiologist irrespective of species or count (Tables 2 and 3), the PA-100 sensitivity was 65.7% (94/143, CI: 57.3% to 73.5%). This lower sensitivity was associated with a positive culture at concentrations < 50,000 CFU/mL (21 cases) or a positive culture for species outside the analyser specifications (1 Citrobacter koseri, 1 Klebsiella oxytoca, 2 Pseudomonas aeruginosa, 3 Streptococcus anginosus, 2 Streptococcus agalactiae and 1 Staphylococcus warneri). Of 122 samples with Gram-negative bacteria, 87 were detected, while 7 out of 20 samples with Gram-positive bacteria were detected. Specificity was 99.3% (134 of 135, CI 95.9% to 100%) and accuracy 82.0% (77.0% to 86.3%).
Potential interference by host cells
Fragments of host blood cells or epithelial cells may be present in urine samples, and these could enter or block the nanofluidic channels, preventing the entrapment of bacteria. When tested with a logistic regression model, red blood cell, white blood cell and epithelial cell counts in the urine were not associated with false negative results (p values were 0.1146, 0.7760 and 0.2303, respectively).
Performance of PA-100 antimicrobial susceptibility test results
Performance of the PA-100 AST on fresh urine samples within IFU specifications is described in Table 4. Similar results were obtained for the performance of the PA-100 AST on all fresh urine samples from the study (Table 5). OCA over all antibiotics was 94.6% (CI 91.9%-96.6%) and 94.0% (CI 91.2%-96.1%), respectively.
Performance characteristics of the AST are stated according to ISO 20776–2 2021. In addition, we investigated the frequency of minor errors (mE), major errors (ME) and very major errors (VME). For amoxicillin/clavulanic acid the frequency of ME was 6.5% (2.1 – 14.5) and of VME was 3.9% (0.8 – 11). For ciprofloxacin the frequency of mE was 12.2% (6 – 21.3), of ME was 3.7% (0.8 – 10.4) and of VME was 1.2% (0 – 6.6). For fosfomycin the frequency of ME was 3% (0.4 – 10.4) and of VME was 3% (0.4 – 10.4). For nitrofurantoin the frequency of ME was 3.8% (0.8 – 10.6) and of VME was 1.3% (0—6.8). For trimethoprim the frequency of ME was 2.4% (0.3 – 8.4) and of VME was 1.2% (0 – 6.5).
Microdilution results in agreement with PA-100 results changed 0 S and 11 R results for amoxicillin/clavulanic acid achieved by disc diffusion, 0 S and 1 R results for nitrofurantoin, 0 S and 0 R results for trimethoprim, 0 S, 1 R and 0 I results for ciprofloxacin, 0 S and 6 R results for fosfomycin.
Oral fosfomycin breakpoints are not provided by EUCAST for S. saprophyticus and Enterococcus. Out of three total Enterococcus infections two were below the cut-off of 50,000 CFU/mL. The sample with more than 50,000 CFU/mL and one of the samples with less were detected by the system and correctly flagged NA for fosfomycin. Out of 7 S. saprophyticus infections 7 were above the cut-off of 50,000 CFU/ml. Two of these samples were detected by the PA-100 as bacteriuria positive. One sample was correctly flagged as NA and one sample resulted in no AST result due to LG being displayed. In conclusion, no sample received an incorrect S or R result for fosfomycin.
In general, NA for amoxicillin/clavulanic acid was flagged in one case each of Aerococcus urinae, Aerococcus viridan, Enterococcus faecalis, Escherichia coli, Klebsiella pneumoniae, Staphylococcus epidermis and Staphylococcus saphrophyticus. NA for fosfomycin was flagged in one case each of Aerococcus urinae, Aerococcus viridans group, Escherichia coli, Klebsiella pneumoniae, Lactococcus garvieae, Staphylococcus epidermidis and Staphylococcus saphrophyticus, and two cases of Enterococcus faecalis. NA for trimethroprim was flagged in one case of Enterococcus faecalis and one case of Lactococcus garvieae.
The frequency of invalid results due to LG, despite a positive result for bacteriuria, or technical errors, was 8.9% for amoxicillin/clavulanic acid, 7.8% for ciprofloxacin, 19.1% for fosfomycin, 10.1% for nitrofurantoin and 4.5% for trimethoprim.
Comparison of PA-100 results with current best practice
To provide context to the analytical performance of the PA-100, actionable recommendations of the analyser, as well as clinical treatment decisions, were compared with clinical microbiology results as a reference (Table 6). Clinicians did not have access to the PA-100 results but based their decisions on current clinical best practice. In reference bacteriuria-negative patients, PA-100 recommendation not to treat was significantly more frequent than actual clinical decisions (99.3% vs 49.6%; p < 0.0001). In reference bacteriuria-positive patients, PA-100 recommendation not to treat was also significantly more frequent than actual clinical decisions (34.3% vs 15.4%; p = 0.0002). Importantly, however, considering the microbiological resistance profile of the isolated bacteria, the PA-100 recommendation for an inappropriate antibiotic was significantly lower than actual clinical decisions (2.1% vs 18.2%; p < 0.0001).
Overall frequency of optimal treatment defined as a) negative urine culture, patients not treated and b) positive urine culture, patients treated with an antimicrobial to which the pathogen was susceptible according to the final routine culture report, was significantly higher for the PA-100 recommendation compared to the clinical decision (78.4% vs 58.3%; p < 0.0001).
Discussion
This study documents for the first time the performance of the PA-100 in a clinical setting using patient urine samples. Previous publications were limited to isolates [14, 18]. The PA-100 is not intended as a substitute for a microbiology laboratory, but as a diagnostic tool to improve the standard of care in near-patient settings. The instrument detects bacteriuria and performs phenotypic AST by automated phase-contrast microscopy monitoring. The system is easy to use, and interpretation of results does not require knowledge of clinical microbiology. The PA-100 uses fresh urine samples within 30 min of collection, which is commensurate with its deployment in a POC setting where rapid results are required. In addition, no pretreatment of urine is necessary.
The PA-100 detected bacteriuria and provided an AST profile under IFU specifications with 84.0% sensitivity and 99.4% specificity. For positive urine cultures outside IFU specifications sensitivity was lower (65.7%), but specificity remained high (99.3%). Challenges were encountered in detecting Gram-positive strains (5 out of 7 S. saprophyticus, 2 out of 4 E. faecalis and 3 out of 3 S. anginosus were missed). Furthermore, the PA-100 detects bacteriuria when samples exhibit > 50,000 CFU/mL, whereas the reference criteria set by the microbiology laboratory classified samples as positive at lower levels of bacteriuria, according to current clinical practice [27].
Overall categorical agreement for the five antimicrobials evaluated in urine samples under IFU specifications was satisfactory for its POC design (94.6%, CI: 91.9%-96.6%). While sensitivity of the AST results appeared to be low (ranging from 75% to 96.3%), the VITEK2 had a similar performance when compared to our microbiological reference (data not shown). This is in line with the performance of other automated AST systems [28]. In a clinical setting the PA-100 uses native urine samples, as opposed to isolates, making it more challenging to provide a correct result. It is important to highlight that the NPV for resistance in our cohort was > 93% for all antibiotics tested, which indicates that a susceptible test result could reliably be used for treatment decisions. Also, the very good NPV is consistent with its use in uncomplicated UTI, where AMR is typically low, and it would even improve in settings where AMR is not as frequent as in this study population.
As described in the methods, in samples where bacteria like Pseudomonas, Aerococcus, or Lactococcus were detected as positive by the system, an incorrect AST reference might be used by the PA-100. Furthermore, the PA-100 does not distinguish between different species of Enterobacterales and thus E. coli fosfomycin and nitrofurantoin breakpoints are always used, irrespective of the final microorganism identification. Since the system does not display which microorganism is assumed, there is also no leverage to immediately reflect on this. This problem of course only applies to strains with resistance features close to their respective breakpoints, but it is a potential drawback of the system.
The PA-100 markedly outperformed current clinical decisions. Overall, the frequency of optimally treated patients, i.e., non-infected patients that were not exposed to antimicrobials and infected patients that were treated with an effective antimicrobial, could have improved significantly from 58.3% to 78.4% (p < 0.0001) if the physician had had access to the PA-100 AST results during consultation. It is important to note that this comparison is hypothetical, and that the actual clinical impact in a healthcare setting still needs to be demonstrated.
Despite sensitivity not being high enough to reliably rule out infection (NPV of 73.2% in this study), the excellent specificity of the PA-100 and the rapid availability of accurate diagnostic and AST data would enable physicians attending to patients with UTI to implement correct targeted treatment in positive cases without delay, avoiding the overuse of broad-spectrum antimicrobials and the subsequent need to de-escalate. In this study 18.2% of the antimicrobials prescribed empirically in bacteriuria-positive patients were ineffective due to bacterial resistance, and utilising PA-100 AST guidance would have decreased this to 2.1% (p < 0.0001). This is especially important in uncomplicated UTI, where treatment adjustment is seldom done, making this disease an important source of resistant strains and exacerbating possible overtreatment by physicians (36.0%, 68/189 of all antimicrobials dispensed in this study were for culture-negative patients).
A limitation of the PA-100 was the relatively high frequency of samples with bacterial growth too low to perform an AST for one of the antimicrobials in the cartridge. However, switching to a different antimicrobial for which an AST was reported, was an obvious solution for this problem, and results not available for all antimicrobials only occurred in five positive PA-100 samples.
Follow-up interventional studies in which the antimicrobials are prioritised depending on the local prevalence of resistant bacteria will be useful to establish the actual clinical impact of the PA-100. It would also be informative to repeat this study in a setting with low prevalence of resistance and strict antibiotic prescription behaviour, as in such settings the PA-100 could potentially detect otherwise untreated bacteriuria.
In conclusion, the PA-100 has implications for clinical practice since it provides bacteriuria and phenotypic AST results within 45 min to guide physicians in quickly making informed decisions and thus significantly improve treatment of patients with UTI. It also has the potential to reduce the inappropriate use of antimicrobials and, hence, overall AMR frequency in the population.
Data Availability
The datasets generated during and/or analysed during the current study are not publicly available due intellectual property but are available from the corresponding author on reasonable request.
References
World Health Organization (2015) Worldwide country situation analysis: response to antimicrobial resistance: summary. https://apps.who.int/iris/handle/10665/163473. Accessed 16 November 2023
World Health Organization (2021) Global Antimicrobial Resistance and Use Surveillance System (GLASS) Report. https://www.who.int/publications/i/item/9789240027336. Accessed 16 November 2023
Sijbom M, Büchner FL, Saadah NH, Numans ME, de Boer MGJ (2023) Determinants of inappropriate antibiotic prescription in primary care in developed countries with general practitioners as gatekeepers: a systematic review and construction of a framework. BMJ Open 13:e065006. https://doi.org/10.1136/bmjopen-2022-065006
Centers for Disease Control and Prevention (CDC) (2022) Antibiotic Use in the United States, 2022 Update: Progress and Opportunities.https://www.cdc.gov/antibiotic-use/stewardship-report/current.html. Accessed 16 November 2023
World Health Organization (2015) Global action plan on antimicrobial resistance. https://www.who.int/publications/i/item/9789241509763. Accessed 16 November 2023
Zeng Z, Zhan J, Zhang K, Chen H, Cheng S (2022) Global, regional, and national burden of urinary tract infections from 1990 to 2019: an analysis of the global burden of disease study 2019. World J Urol 40(3):755–763. https://doi.org/10.1007/s00345-021-03913-0
Kornfält Isberg H, Melander E, Hedin K, Mölstad S, Beckman A (2019) Uncomplicated urinary tract infections in Swedish primary care; etiology, resistance and treatment. BCM Infect Dis 19:155. https://doi.org/10.1186/s12879-019-3785-x
Mo Y (2022) Rapid Diagnostics for Antibiotic Resistance: Urgent Need for Strong Clinical Evidence. Clin Infect Dis 75(12):2076–2078. https://doi.org/10.1093/cid/ciac358
Davenport M, Mach KE, Shortliffe LMD, Banaei N, Wang TH, Liao JC (2017) New and developing diagnostic technologies for urinary tract infections. Nat Rev Urol 14(5):296–310. https://doi.org/10.1038/nrurol.2017.20
Santos M, Mariz M, Tiago I, Martins J, Alarico S, Ferreira P (2022) A review on urinary tract infections diagnostic methods: Laboratory-based and point-of-care approaches. J Pharm Biomed Anal 219:114889. https://doi.org/10.1016/j.jpba.2022.114889
Smith KP, Kirby JE (2019) Rapid Susceptibility Testing Methods. Clin Lab Med 39(3):333–344. https://doi.org/10.1016/j.cll.2019.04.001
Murray C, Adeyiga O, Owsley K, Di Carlo D (2015) Research highlights: Microfluidic analysis of antimicrobial susceptibility. Lab Chip 15:1226–1229. https://doi.org/10.1039/c5lc90017d
Campbell J, McBeth C, Kalashnikov M, Boardman AK, Sharon A, Sauer-Budge AF (2016) Microfluidic advances in phenotypic antibiotic susceptibility testing. Biomed Microdevices 18(6):103. https://doi.org/10.1007/s10544-016-0121-8
Baltekin Ö, Boucharin A, Tano E, Andersson DI, Elf J (2017) Antibiotic susceptibility testing in less than 30 min using direct single-cell imaging. PNAS 144(34):9170–9175. https://doi.org/10.1073/pnas.1708558114
Reece A, Xia B, Jiang Z, Noren B, McBride R, Oakey J (2016) Microfluidic techniques for high throughput single cell analysis. Curr Opin Biotechnol 40:90–96. https://doi.org/10.1016/j.copbio.2016.02.015
Clark RB, Lewinski MA, Loeffelholz MJ, Tibbetts RJ (2009) CUMITECH 31A, Verification and Validation of Procedures in the Clinical Microbiology Laboratory. Washington DC: Coordinating ed. SE Sharp, ASM Press
European Society of Clinical Microbiology and Infections. EUCAST (2023) https://www.eucast.org/. Accessed 16 November 2023
Baltekin Ö. et. al. (2022) Reproducibility of fully automated AST for direct near patient testing. (Poster Presentation). In ECCMID 2022, Lisbon. Astrego Diagnostics AB
Ö Baltekin ATA Johnsson AYW Wong K Nilsson B Mert L Söderberg E Wistrand-Yuen V Özenci 2021 Evaluation of an ultra-rapid antibiotic susceptibility testing method on positive blood cultures with Escherichia coli MedRxiv https://doi.org/10.1101/2021.12.14.21267046
Elf J, Öhman O, Lovmar M, Baltekin Ö. (2018) Microfluidic device for cell characterization. WO2019/088904 A1 (Patent application)
Tomlinson E, Jones HE, James R, Cooper C, Stokes C, Begum S, Watson J, Hay AD, Ward M, Thom H, Whiting P (2023) Clinical effectiveness of point of care tests for diagnosing urinary tract infection: a systematic review. Clin Microbiol Infec. https://doi.org/10.1016/j.cmi.2023.10.005
Leber AL. (2022) Urine Cultures. In: Leber AL (ed) Clinical Microbiology Procedures Handbook. 4th edn. Washington DC: ASM Press; Chapter 3.12.
Liu YC, Huang WK, Huang TS, Kunin CM (1999) Detection of Antimicrobial Activity in Urine for Epidemiologic Studies of Antibiotic Use. J Clin Epidemiol 52:539–545. https://doi.org/10.1016/S0895-4356(99)00027-X
Clinical and Laboratory Standards Institute (2020) M100 Performance Standards for Antimicrobial Susceptibility Testing, 30th Edition
International Organization for Standardization(2021) ISO 20776–2:2021. Clinical laboratory testing and in vitro diagnostic test systems. Geneva: ISO
Altman DG (2000) Statistics in medical journal: some recent trends. Stat Med 19(23):3275–3289. https://doi.org/10.1002/1097-0258(20001215)19:23%3c3275::AID-SIM626%3e3.0.CO;2-M
McCarter Y, Burd EM, Hall GS, Zervos M (2009) CUMITECH 2C: Laboratory Diagnosis of Urinary Tract Infections, American Society for Microbiology. Coordinating ed. Sharp SE.
Soares A, Pestel-Caron M, Leysour de Rohello F, Bourgoin G, Boyer S, Caron F (2020) Area of technical uncertainty for susceptibility testing of amoxicillin/clavulanate against Escherichia coli: analysis of automated system, Etest and disk diffusion methods compared to the broth microdilution reference. Clin Microbiol Infec 1685:e1-1685.e6. https://doi.org/10.1016/j.cmi.2020.02.038
Acknowledgements
Theresa L Coetzer, Sysmex, Hamburg, Germany for text review
Sukayna Ferdi for recruiting patients and technical work in Hospital de la Santa Creu i Sant Pau
Emergency and Laboratory Departments of Fundació Puigvert and Hospital de la Santa Creu i Sant Pau
Funding
This work was supported by Sysmex Europe SE, Norderstedt, Germany.
Author information
Authors and Affiliations
Contributions
Carles Alonso-Tarrés participated in the design, development and analysis of the study, as well as writing of the manuscript.
Carla Benjumea Moreno participated in the design, development and analysis of the study, as well as writing of the manuscript.
Carles Alonso‑Tarrés and Carla Benjumea Moreno contributed equally to this paper and are co-first authors.
Ferran Navarro gave key ideas to the study, participated in the analysis and revised the final manuscript.
Aline C. Habison participated in the design, analysis and drafting of the study.
Elisenda Gonzàlez-Bertran participated in the acquisition of data, technical work and made intellectual contributions.
Francisco Blanco participated in the design, analysis and drafting of the study.
Jaume Borràs participated recruiting patients and gave some key intellectual ideas for the study.
Montserrat Garrigó participated recruiting patients and intellectual content.
Jarob Saker participated in the design, development, analysis of the study, as well as writing of the manuscript.
Corresponding author
Ethics declarations
Ethics approval
This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Ethics Committees of FP (13/07/2022) and HSP (07/09/2022). Informed consent was obtained from each patient.
Competing interests
Sysmex paid the institution to elaborate the study. Carles Alonso-Tarrés works at Fundació Puigvert. Sysmex paid the institution and to Carles Alonso-Tarrés to elaborate the study. Dr. Alonso-Tarrés has also received payments from Sysmex by his participation in an Advisory Board on Urinary tract Infections and for attending a congress. Benjumea Moreno, Carla works at Fundació Puigvert. Syxmex paid the institution and to Carla Benjumea to elaborate the study. Aline Habison is an employee of Sysmex Europe SE in the department of New Business Development. Elisenda Gonzàlez-Bertran worked at Fundació Puigvert to participate in the study. Sysmex paid to Fundació Puigvert and Ms. Gonzàlez-Bertran to elaborate the study. Francisco Blanco is an employee of Sysmex Europe SE in the department of Primary Care. Jarob Saker is an employee of Sysmex Europe SE in the department of Medical Science. Ferran Navarro, Jaume Borràs and Montserrat Garrigó had no conflict of interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Alonso-Tarrés, C., Benjumea Moreno, C., Navarro, F. et al. Bacteriuria and phenotypic antimicrobial susceptibility testing in 45 min by point-of-care Sysmex PA-100 System: first clinical evaluation. Eur J Clin Microbiol Infect Dis (2024). https://doi.org/10.1007/s10096-024-04862-3
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10096-024-04862-3