Abstract
We experienced a case of ventricular assist with both a pulsatile-flow and a continuous-flow pump in a pediatric patient, and herein report the clinical course and characteristics of the pumps. A 6-year-old female was diagnosed with fulminant myocarditis and transferred to our hospital for mechanical support. After 12 days of extracorporeal membrane oxygenation, we implanted a left ventricular assist device (LVAD) and a right ventricular assist device (RVAD) using centrifugal Gyro pumps with a membrane oxygenator in a paracorporeal fashion. The membrane oxygenator was removed on postoperative day (POD) 4, and the patient was weaned from the respirator on POD 6. The LVAD was exchanged on POD 13 and 17, and the RVAD was exchanged on POD 14 because of thrombus formation inside the pumps. The RVAD was removed on POD 25. On POD 32, the patient experienced cerebral infarction and the centrifugal Gyro pump was switched to an extracorporeal pulsatile pump. No thromboembolic event occurred after pump conversion, although continuous administration of vasodilators was required to avoid hypertension. She underwent successfully heart transplantation in the USA after 8 months of ventricular support. A centrifugal pump is considered useful for pediatric patients, as pump flow and blood pressure can be relatively easily controlled in the postoperative acute phase compared with the pulsatile pump. However, special care should be taken to monitor for thrombus formation when support length becomes longer than 13 days, and a switch to a pulsatile pump should be considered once the hemodynamic status stabilizes.
Similar content being viewed by others
References
Homma A, Taenaka Y, Tatsumi E, et al. Development of a compact wearable pneumatic drive unit for a ventricular assist device. J Artif Organs. 2008;11:182–90.
Inoue T, Kitamura T, Torii S, et al. Five-week use of a monopivot centrifugal blood pump as a right ventricular assist device in severe dilated cardiomyopathy. J Artif Organs. 2014;17:95–8.
Takano T, Nakata K, Yoshikawa M, et al. Development of an antithrombogenic and antitraumatic blood pump: the Gyro C1E3. ASAIO J. 2000;46:123–7.
Martin J, Sarai K, Schindler M, van de Loo A, Yoshitake M, Beyersdorf F. MEDOS HIA-VAD biventricular assist device for bridge to recovery in fulminant myocarditis. Ann Thorac Surg. 1997;63:1145–6.
Wei-Che C, Marvin JS, Danny B. Thrombus formation patterns in the HeartMate II ventricular assist device: clinical observations can be predicted by numerical simulations. ASAIO J. 2014;60:237–40.
Author information
Authors and Affiliations
Corresponding authors
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Rights and permissions
About this article
Cite this article
Kehara, H., Takano, T., Terasaki, T. et al. Biventricular support using a centrifugal pump in a 6 year old with fulminant myocarditis. J Artif Organs 20, 166–169 (2017). https://doi.org/10.1007/s10047-016-0939-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-016-0939-0