Purpose
The O-arm-based navigation increases the accuracy of pedicle screw positioning and offers the possibility of performing a 3D scan before wound closure. However, repeating the 3D scan exposes the patient to additional radiation. We combined O-arm navigation with pedicle screw (PS) stimulation followed by a 3D scan to evaluate their accuracy and aimed for the creation of a protocol that maximizes the safety and minimizes radiation.
Methods
Patients had pedicle screws insertion using O-arm spinal navigation, then had PS triggered electromyography (EMG), and finally a 3D scan to evaluate the accuracy of screw position.
Results
447 screws were inserted in 71 patients. In 10 patients, 11 screws needed repositioning. Comparing results of PS triggered EMG responses to the 3D scan, we found: (a) negative stimulation response with negative 3D scan findings, corresponding to 432 acceptable screw position (96.6 %) in 58 patients (81.7 %). In these cases, the redo 3D scan could be avoided. (b) Positive stimulation response with positive 3D scan findings, corresponding to 7 unacceptable screw position (1.5 %) in 6 patients (8.4 %). In these cases, PS stimulation detected malpositioned screws that would be missed without a redo 3D scan.
Conclusion
We propose a protocol of routinely performing PS stimulation after screw insertion using spinal navigation. In case of positive stimulation, a 3D scan must be performed to rule out a probable screw mal position (6 patients 8.4 %). However, in case of negative stimulation, redo 3D scan can be avoided in 81.7 % of patients.
Similar content being viewed by others
References
Abdullah KG, Bishop FS, Lubelski D, Steinmetz MP, Benzel EC, Mroz TE (2012) Radiation exposure to the spine surgeon in lumbar and thoracolumbar fusions with the use of an intraoperative computed tomographic 3-dimensional imaging system. Spine (Phila Pa 1976) 37(17):1074–1078
Alemo S, Sayadipour A (2010) Role of intraoperative neurophysiologic monitoring in lumbosacral spine fusion and instrumentation: a retrospective study. World Neurosurg 73(1):72–76
Attar A, Ugur HA, Uz A, Tekdemir I, Genec Y, Egemen N (2001) Lumbar pedicle: surgical anatomic evaluation and relationships. Eur Spine J 10(1):10–15
Bose B, Wierzbowski LR, Sestokas AK (2002) Neurophysiologic monitoring of spinal nerve root function during instrumented posterior lumbar spine surgery. Spine (Phila Pa 1976) 27(13):1444–1450
Castro Castro J, Rodino Padín J, Pinzón Millán A, Agulleiro Díaz JP, Villa Fernández JM, Al Pastor Zapata (2013) Posterior lumbar fusion using the O-arm surgical imaging system: initial experience. Neurocirugia (Astur) 24(1):1–8
Castro WH, Halm H, Jerosch J, Malms J, Steinbeck J, Blasius S (1996) Accuracy of pedicle screw placement in lumbar vertebrae. Spine (Phila Pa 1976) 21(11):1320–1324
Costa F, Villa T, Anasetti F, Tomei M, Ortolina A, Cardia A et al (2013) Primary stability of pedicle screws depends on the screw positioning and alignment. Spine J 13(12):1934–1939
Donohue ML, Swaminathan V, Gilbert JL, Fox CW, Smale J, Moquin RR et al (2012) Intraoperative neuromonitoring: can the results of direct stimulation of titanium-alloy pedicle screws in the thoracic spine be trusted? J Clin Neurophysiol 29(6):502–508
Gelalis ID, Paschos NK, Pakos EE, Politis AN, Arnaoutoglou CM, Karageorgos AC et al (2012) Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques. Eur Spine J 21(2):247–255
Isley MR, Zhang XF, Balzer JR, Leppanen RE (2012) Current trends in pedicle screw stimulation techniques: lumbosacral, thoracic, and cervical levels. Neurodiagn J 52(2):100–175
Kulik G, Pralong E, McManus J, Debatisse D, Schizas C (2013) A CT-based study investigating the relationship between pedicle screw placement and stimulation threshold of compound muscle action potentials measured by intraoperative neurophysiological monitoring. Eur Spine J 22(9):2062–2068
Ling JM, Dinesh SK, Pang BC, Chen MW, Lim HL, Louange DT et al (2014) Routine spinal navigation for thoraco-lumbar pedicle screw insertion using the O-arm three-dimensional imaging system improves placement accuracy. J Clin Neurosci 21(3):493–498
Luther N, Iorgulescu JB, Geannette C, Gebhard H, Saleh T, Tsiouris AJ et al (2013) Comparison of navigated versus non-navigated pedicle screw placement in 260 patients and 1434 screws: screw accuracy, screw size, and the complexity of surgery. J Spinal Disord Tech [Epub ahead of print]
Mason A, Paulsen R, Babuska JM, Rajpal S (2014) The accuracy of pedicle screw placement using intraoperative image guidance systems. J Neurosurg Spine 20(2):196–203
Mavrogenis AF, Papagelopoulos PJ, Korres DS, Papadopoulos K, Sakas DE, Pneumaticos S (2009) Accuracy of pedicle screw placement using intraoperative neurophysiological monitoring and computed tomography. J Long Term Eff Med Implants 19(1):41–48
Montes E, De Blas G, Regidor I, Barrios C, Burgos J, Hevia E et al (2012) Electromyographic thresholds after thoracic screw stimulation depend on the distance of the screw from the spinal cord and not on pedicle cortex integrity. Spine J 12(2):127–132
Parker SL, Amin AG, Farber SH, McGirt MJ, Sciubba DM, Wolinsky JP et al (2011) Ability of electromyographic monitoring to determine the presence of malpositioned pedicle screws in the lumbosacral spine: analysis of 2450 consecutively placed screws. J Neurosurg Spine 15(2):130–135
Resnick DK, Choudhri TF, Dailey AT, Groff MW, Khoo L, Matz PG et al (2005) American Association of Neurological Surgeons/Congress of Neurological Surgeons Guidelines for the performance of fusion procedures for degenerative disease of the lumbar spine. Part 15: electrophysiological monitoring and lumbar fusion. J Neurosurg Spine 2(6):725–732
Samdani AF, Tantorski M, Cahill PJ, Ranade A, Koch S, Clements DH et al (2011) Triggered electromyography for placement of thoracic pedicle screws: is it reliable? Eur Spine J 20(6):869–874
Shin MH, Hur JW, Ryu KS, Park CK (2013) Prospective Comparison Study between the Fluoroscopy Guided and Navigation Coupled with O-arm® Guided Pedicle Screw Placement in the Thoracic and Lumbosacral Spines. J Spinal Disord Tech [Epub ahead of print]
Stauff MP, Freedman BA, Kim JH, Hamasaki T, Yoon ST, Hutton WC (2014) The effect of pedicle screw redirection after lateral wall breach—a biomechanical study using human lumbar vertebrae. Spine J 14(1):98–103
Laine T, Lund T, Ylikoski M, Lohikoski J, Schlenzka D (2000) Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patients. Eur Spine J 9(3):235–240
Tabaraee E, Gibson AG, Karahalios DG, Potts EA, Mobasser JP, Burch S (2013) Intraoperative cone beam-computed tomography with navigation (O-ARM) versus conventional fluoroscopy (C-ARM): a cadaveric study comparing accuracy, efficiency, and safety for spinal instrumentation. Spine (Phila Pa 1976) 38(22):1953–1958
Van de Kelft E, Costa F, Van der Planken D, Schils F (2012) A prospective multicenter registry on the accuracy of pedicle screw placement in the thoracic, lumbar, and sacral levels with the use of the O-arm imaging system and StealthStation Navigation. Spine (Phila Pa 1976) 37(25):1580–1587
Waschke A, Walter J, Duenisch P, Reichart R, Kalff R, Ewald C (2013) CT-navigation versus fluoroscopy-guided placement of pedicle screws at the thoracolumbar spine: single center experience of 4500 screws. Eur Spine J 22(3):654–660
Conflict of interest
The authors attest having no conflict of interest related to the article.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kassis, S.Z., Abukwedar, L.K., Msaddi, A.K. et al. Combining pedicle screw stimulation with spinal navigation, a protocol to maximize the safety of neural elements and minimize radiation exposure in thoracolumbar spine instrumentation. Eur Spine J 25, 1724–1728 (2016). https://doi.org/10.1007/s00586-015-3973-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00586-015-3973-3