Abstract
A high-resolution, low offset delta-sigma analog to digital converter is presented for detecting photoplethysmography (PPG) signals, which oscillates in 1/f noise frequency band between 0.1 and 10 Hz. A delta-sigma analog-to-digital converter (DS ADC) is choosen for PPG application because it has wide dynamic range, high-resolution, and can be integrated with chopper-based operational amplifier. a chopper-based operational amplifier converts 1/f noise and dc offset to high-frequency noise, which acts as a virtual high-pass filter. Implementing a lowpass filter using a decimation filter can filter the high-frequency signal. Thus, the circuit becomes a new bandpass filter. A second-order continuous-time delta-sigma modulator is used in the DS ADC circuits. TSMC 180 nm technology was used to design and simulate the delta-sigma modulator circuits. The active area of the continuous-time delta-sigma modulator is 746 µm × 399 µm and was built using TSMC 180 nm technology. It has a bandwidth of 10 Hz and 512 oversampling ratios. The SFDR of the circuit is greater than 70 dB. The delta-sigma modulator consumes 35.61 µW of power. To ensure the circuitry works in a variety of scenarios, the simulation is run in three different types of corners, SS, TT, and FF, successfully validating the designed performance.
Similar content being viewed by others
References
Ahmad S, Chen S, Soueidan K et al (2012) Electrocardiogram-assisted blood pressure estimation. IEEE Trans Biomed Eng 59:608–618. https://doi.org/10.1109/TBME.2011.2180019
Chen H-L, Chen P-S, Chiang J-S (2009) A low-offset low-noise sigma-delta modulator with pseudorandom chopper-stabilization technique. IEEE Trans Circuits Syst I 56:2533–2543. https://doi.org/10.1109/TCSI.2009.2018922
Enz CC, Vittoz EA, Krummenacher F (1987) A CMOS chopper amplifier. IEEE J Solid-State Circuits 22:335–342. https://doi.org/10.1109/JSSC.1987.1052730
Forouzanfar M, Ahmad S, Batkin I et al (2013) Coefficient-free blood pressure estimation based on pulse transit time-cuff pressure dependence. IEEE Trans Biomed Eng 60:1814–1824. https://doi.org/10.1109/TBME.2013.2243148
Heart Foundation (2021) Key Statistics: Cardiovascular Disease, Statistics and Information on Cardiovascular disease in Australia. The Heart Foundation. https://www.heartfoundation.org.au/activities-finding-or-opinion/key-stats-cardiovascular-disease. Accessed 31 Oct 2021
Hogenauer E (1981) An economical class of digital filters for decimation and interpolation. IEEE Trans Acoust 29:155–162. https://doi.org/10.1109/TASSP.1981.1163535
Huijsing J (2017). Operational Amplifiers: Theory and Design 3rd Edition. Springer, Switzerland https://doi.org/10.1007/978-3-319-28127-8
Kao Y-H, Chao PC-P, Wey C-L (2018) Towards maximizing the sensing accuracy of an cuffless, optical blood pressure sensor using a high-order front-end filter. Microsyst Technol 24:1–10. https://doi.org/10.1007/s00542-018-3877-3
Kao Y-H, Chao PC-P, Wey C-L (2019) Design and validation of a new PPG module to acquire high-quality physiological signals for high-accuracy biomedical sensing. IEEE J Select Topics Quantum Electron 25:1–10. https://doi.org/10.1109/JSTQE.2018.2871604
Klumperink EAM, Gierkink SLJ, Van Der Wel AP, Nauta B (2000) Reducing MOSFET 1=f Noise and Power Consumption by Switched Biasing. IEEE J Solid-State Circuits 35:994–1001. https://doi.org/10.1109/4.848208
Konijnenburg M, Stanzione S, Yan L et al (2016) A Multi(bio)sensor acquisition system with integrated processor, power management, 8×8 LED drivers, and simultaneously synchronized ECG, BIO-Z, GSR, and Two PPG readouts. IEEE J Solid-State Circuits 51:2584–2595. https://doi.org/10.1109/JSSC.2016.2605660
Lee Y, Lee H, Jang J et al (2017) Sticker-type hybrid photoplethysmogram monitoring system integrating cmos ic with organic optical sensors. IEEE J Emerg Sel Topics Circuits Syst 7:50–59. https://doi.org/10.1109/JETCAS.2016.2630301
Lee H, Kim E, Lee Y, et al (2018) Toward all-day wearable health monitoring: An ultralow-power, reflective organic pulse oximetry sensing patch. Sci Adv 4:eaas9530. https://doi.org/10.1126/sciadv.aas9530
Marefat F, Erfani R, Kilgore KL, Mohseni P (2020a) A 280 μW, 108 dB DR PPG-readout IC with reconfigurable, 2nd-order, incremental ΔΣM front-end for direct light-to-digital conversion. IEEE Trans Biomed Circuits Syst 14:1183–1194. https://doi.org/10.1109/TBCAS.2020.3038046
Marefat F, Erfani R, Mohseni P (2020b) A 1-V 8.1-µW PPG-recording front-end with > 92-dB DR using light-to-digital conversion with signal-aware DC subtraction and ambient light removal. IEEE Solid-State Circuits Lett 3:17–20. https://doi.org/10.1109/LSSC.2019.2957261
Pandey RK, Pribadi EF, Chao PC-P (2019) A new adaptive readout system for a new OLED OPD flexible patch PPG sensor. 2019 IEEE SENSORS. IEEE, pp 1–4. https://doi.org/10.1109/SENSORS43011.2019.8956825
Pandey RK, Chao PCP (2021) External temperature sensor assisted a new low power photoplethysmography readout system for accurate measurement of the bio-signs. Microsyst Technol 27:2315–2343. https://doi.org/10.1007/s00542-020-05106-y
Pandey RK, Chao PCP (2022) A Dual-Channel PPG Readout System With Motion Tolerant Adaptability for OLED-OPD Sensors. IEEE Trans Biomed Circuits Syst 16(1):36–51. https://doi.org/10.1109/TBCAS.2021.3138996
Pribadi EF, Pandey RK, Chao PC-P (2020) Optimizing a novel PPG sensor patch via optical simulations towards accurate heart rates. Microsyst Technol 26:3409–3420. https://doi.org/10.1007/s00542-020-04895-6
Pribadi EF, Pandey RK, Chao PC-P (2021) Design and implementation of a new light to digital converter for the PPG sensor. Microsyst Technol 27:2461–2472. https://doi.org/10.1007/s00542-020-05154-4
Priyanka KNG, Chao PC-P, Tu T-Y, et al (2018) Estimating blood pressure via artificial neural networks based on measured photoplethysmography waveforms. 2018 IEEE SENSORS. IEEE, pp 1–4. https://doi.org/10.1109/ICSENS.2018.8589796
Vercaemer D, Raman J, Rombouts P (2017) Passive loop filter assistance for ctsdms. IEEE Trans Circuits Syst II 64:1157–1161. https://doi.org/10.1109/TCSII.2016.2633002
Virani SS, Alonso A, Benjamin EJ et al (2020) Heart disease and stroke statistics-2020 update: a report from the american heart association. Circulation 141:e139–e596. https://doi.org/10.1161/CIR.0000000000000757
Yoon Y, Duan Q, Yeo J et al (2016) A delta-sigma modulator for low-power analog front ends in biomedical instrumentation. IEEE Trans Instrum Meas 65:1530–1539. https://doi.org/10.1109/TIM.2016.2534358
Acknowledgements
This work is supported by the Minister of Science and Technology under 109-2221-E-009-163, 109-2622-8-009-018-TE1, 110-2622-8-009-011-TE1, 110-2223-E-A49-001-, 109-2622-E-009-027-, and 110-2218-E-A49-020-MBK. And the authors would like to acknowledge chip fabrication support provided by Taiwan Semiconductor Research Institute (TSRI), Taiwan, R. O. C.
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Pribadi, E.F., Pandey, R.K. & Chao, P.CP. A new delta-sigma analog to digital converter with high-resolution and low offset for detecting photoplethysmography signal. Microsyst Technol 28, 2369–2379 (2022). https://doi.org/10.1007/s00542-022-05360-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00542-022-05360-2