Abstract
Objectives
To investigate the use of a fast dynamic hyperpolarised 129Xe ventilation magnetic resonance imaging (DXeV-MRI) method for detecting and quantifying delayed ventilation in patients with chronic obstructive pulmonary disease (COPD).
Methods
Three male participants (age range 31–43) with healthy lungs and 15 patients (M/F = 12:3, age range = 48–73) with COPD (stages II–IV) underwent spirometry tests, quantitative chest computed tomography (QCT), and DXeV-MRI at 1.5-Tesla. Regional delayed ventilation was captured by measuring the temporal signal change in each lung region of interest (ROI) in comparison to that in the trachea. In addition to its qualitative assessment through visual inspection by a clinical radiologist, delayed ventilation was quantitatively captured by calculating a covariance measurement of the lung ROI and trachea signals, and quantified using both the time delay, and the difference between the integrated areas covered by the signal-time curves of the two signals.
Results
Regional temporal ventilation, consistent with the expected physiological changes across a free breathing cycle, was demonstrated with DXeV-MRI in all patients. Delayed ventilation was observed in 13 of the 15 COPD patients and involved variable lung ROIs. This was in contrast to the control group, where no delayed ventilation was demonstrated (p = 0.0173).
Conclusions
DXeV-MRI offers a non-invasive way of detecting and quantifying delayed ventilation in patients with COPD, and provides physiological information on regional pulmonary function during a full breathing cycle.
Key Points
• Dynamic xenon MRI allows for the non-invasive detection and measurement of delayed ventilation in COPD patients.
• Dynamic xenon MRI during a free breathing cycle can provide unique information about pulmonary physiology and pulmonary disease pathophysiology.
• With further validation, dynamic xenon MRI could offer a non-invasive way of measuring collateral ventilation which can then be used to guide lung volume reduction therapy (LVRT) for certain COPD patients.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
First described by Van Allen et al [1] in 1931, pulmonary collateral ventilation refers to the phenomenon of ventilation of alveolar structures through passages or channels other than the normal airways [2]. It is known to be associated with emphysema in patients with chronic obstructive pulmonary disease (COPD) where it is believed that pathological distal airspace dilatation and wall destruction lead to the formation of alternative channels of airflow [3]. However, no clear relationship has been established between the presence and extent of collateral ventilation and COPD disease severity [3].
In order to determine the likelihood of treatment success when considering patients for treatment options targeting pulmonary hyper-expansion in severe COPD, such as endobronchial valve placement or surgical lung resection, the assessment of collateral ventilation is essential [4]. In current clinical practice, COPD patients can be assessed for the presence of collateral ventilation either indirectly with computed tomography (CT) or directly via a commercially available bronchoscopic system (Chartis® system, Pulmonx Inc.) [5, 6]. There is a need for alternative non-invasive functional imaging methods to assess patient lungs for collateral ventilation, because bronchoscopic systems are invasive, and chest CT provides limited functional pulmonary information and involves ionising radiation.
133Xe scintigraphy [7] and xenon-enhanced dynamic dual-energy CT [8, 9] have previously been employed to obtain images of collateral ventilation but incur exposure to ionising radiation. Hyperpolarised gas imaging has emerged as a valuable technique for evaluating ventilation [10,11,12], offering the advantage of enabling the assessment of gas diffusion to comprehensively evaluate both emphysema distribution and collateral ventilation [13]. Hyperpolarised 3He MRI has successfully demonstrated collateral ventilation directly [14], and long-range diffusion measurements with 3He may indicate that collateral ventilation is taking place [14].
Hyperpolarised 3He MRI has been more extensively studied than 129Xe MRI due to its stronger magnetic moment, resulting in a higher spatial resolution and signal-to-noise ratio than 129Xe [15]. However, the use of 129Xe is gaining popularity in recent years due to its abundant natural supply and lower cost [16]. Recent breakthroughs in 129Xe production have allowed it to achieve levels of polarisation similar to those with 3He [17]. Positive correlations have been established between functional ventilation information acquired using hyperpolarised 3He and hyperpolarised 129Xe MRI [18]. The higher density of 129Xe makes it more sensitive to capturing ventilation defects [19]. Its greater solubility and larger chemical shift enable the acquisition of additional functional information about gas exchange [20,21,22]. Both gases, at the imaging doses prescribed, are well-tolerated by both healthy individuals and patients with lung diseases [23,24,25]. More recently, hyperpolarised 129Xe has been shown to be a promising tool for measuring lobar ventilation [26, 27], and for identifying temporal changes in the arrival of xenon gas to different parts of the lung [28, 29].
Most existing studies [14, 30] on the use of hyperpolarised gas MRI for characterising delayed ventilation in COPD patients are based on time-resolved static MR sequences during a single breath-hold. However, relatively slowly filling ventilation defects can be challenging to assess in one breath-hold. Multi-breath imaging techniques have been proposed and tested [31, 32], where during each breath, a fraction of the hyperpolarised gas is replaced by newly arrived gas, and as new scans are acquired at each subsequent breath-hold, a volume fractional ventilation measure is calculated and used to assess air trapping and delayed ventilation. However, these methods are susceptible to lung movements, which give rise to localisation errors, as well as signal decays due to gas relaxation between scan acquisitions.
The static nature of the aforementioned methods makes them incapable of capturing the dynamics of pulmonary function during different phases of a breathing cycle, which can lead to a failure in identifying lung areas with significant ventilation problems, such as the ones affected by collateral ventilation and pulmonary air leaks. An understanding of the dynamic ventilation process during a complete breathing cycle could offer unique insight into pulmonary physiology and pathophysiology in disease states. Finally, the length of breath-hold required for high quality scans in static hyperpolarised gas MRI often proves challenging for patients with severely diseased lungs.
In addition to an earlier 3He-based work [33] supporting the idea of dynamic ventilation MRI during free breathing, two more recent pilot studies [34, 35] using dynamic 129Xe MRI have reported promising physiologically compatible results in gas phantoms, healthy volunteers and a limited number of COPD patients, prompting further research into this topic. In this paper, a fast dynamic hyperpolarised 129Xe ventilation magnetic resonance imaging (DXeV-MRI) method is presented, for assessing delayed ventilation in patients with COPD.
Methods and materials
The study was approved by the UK National Research Ethics Service South Central Committee (Berkshire UK, REC 11/SC/0487 and REC 11/SC/0488). Written informed consent was obtained from all study participants.
Three male participants (aged 31, 34 and 43, respectively) with healthy lungs were imaged between September 2017 and May 2018. Fifteen patients (M/F = 12:3, age range 48–73) diagnosed with COPD were prospectively enrolled from a tertiary referral centre and imaged between February 2016 and August 2017. The study recruitment criteria are given in the appendices.
Participants in both cohorts underwent a spirometry tests, DXeV-MRI at 1.5-Tesla and quantitative computed tomography (QCT) of the chest at a single time point. In the COPD cohort, study measures were completed during disease stability, defined as following a 2-week window without COPD exacerbation or change in medications.
A summary of demographics for the patients with COPD included in the study is given in Table 1.
129Xe polarisation and delivery
Isotropically enriched 129Xe gas (86% 129Xe, Spectra Gases Inc.) was polarised to 10–15% by rubidium vapour spin-exchange optical pumping (SEOP), and cryogenically accumulated in 1-L doses using a commercial polariser (Model 9300, Polarean). Polarisation was measured using a commercial polarisation measurement station (Model 2881, Polarean).
Hyperpolarised 129Xe was thawed into a Tedlar® bag (Jensen Inert Products) and administered within 10 minutes of production to participants who were lying supine in the MRI scanner. Participants were given standard breathing instructions prior to image acquisition and asked to perform practice breaths. They are instructed to exhale to functional residual capacity (FRC) and then inhale the full 1-L gas content within 3 seconds during the inhalation period, followed by a 5-second breath-hold, then exhale the gas over 3 seconds. The total scanning time was 20 seconds, including a baseline period of 2 seconds before the initial inhalation and a flush interval of 5 seconds after the end of exhalation. Flush interval is the time interval between the exhalation of xenon gas and end of the scan, when the signal intensity drops significantly due to the reduced concentration of 129Xe and incoming oxygen.
1H-MRI
Both 1H-MRI and DXeV were performed on a 1.5-Tesla Signa HDx whole-body MR system (GE Healthcare). For 1H-MRI, plane localiser images were acquired using double-inversion-recovery black-blood imaging sequence with the following scanning parameters: number of slices: 13, slice thickness: 15 mm, imaging plane: coronal, and image reconstruction size: 128 × 128 pixels.
DXeV-MRI
For DXeV-MRI, participants were fitted with a flexible twin Helmholtz quadrature transmit-receive coil (Clinical MR Solutions) tuned to the 129Xe Larmor frequency for 1.5 Tesla (~ 17.7 MHz).
The scans were acquired using a two-interleaved spiral k-space sampling approach, due to its superior temporal resolution, which has been demonstrated previously [35]. The following pulse sequence parameters were used: number of slices: 13, slice thickness: 15 mm, imaging plane: coronal, and image reconstruction size: 128 × 128 pixels. A total of 32 volume images were acquired during the 20-second scanning time, with each volume image taking 625 milliseconds. The total table time for each HP 129Xe-MRI scan is about 5 minutes.
Spirometry tests
Spirometry tests were completed on all participants using a Compact plus flowmeter pulmonary function testing station (Hypairm Medisoft Group) and included forced expiratory volume in 1 second (FEV1) and FEV1/forced vital capacity (FVC). Recordings were compared to predicted values from standard published data for the given patient population.
Quantitative CT
QCT was performed on a 16-slice GE Discovery 670 scanner (GE Healthcare). Images were acquired with a 1.25-mm slice thickness during suspended tidal inspiration following inhalation of 1 L oxygen via a Tedlar® bag from FRC to ensure lung volumes were as similar to HP 129Xe-MRI as possible. All participants had received breath-hold training beforehand to ensure scan reproducibility.
Image data analyses
Regions of interest (ROIs) were first manually delineated in the trachea, and the upper and lower parts of each lung by a clinical radiologist with 2 years of thoracic radiology experience, on every coronal slice (see Fig. 1a–d), based on the presence of oblique fissures on QCT and anatomical knowledge of the thoracic structures from the 1H-MRI scan. The lung ROIs were labelled as left upper, left lower, right upper and right lower lungs. Hyperpolarised 129Xe signal decay occurred to the flip angle and longitudinal decay of hyperpolarised 129Xe (T1 relaxation) during the acquisition of time-series images [35]. Decay of polarisation was accounted for through normalisation of images at each time point using a well-ventilated area in the trachea to address variations in scan conditions between subjects. To ensure consistency in normalisation, the final two timesteps (frames 31 and 32) were omitted in the analysis, because the tracheal signal is not always present in those frames for all study patients. All signals were also normalised to the maximum mean value measured during the breathing cycle, for the ease of comparison.
Pulmonary Toolkit (PTK), an open-source software, was used to analyse QCT. The percentage emphysema on a global and lobar basis was determined, in the latter case employing a fissure-based pulmonary lobar segmentation supported by software. The extent of emphysema was quantified as percentage area with an attenuation value of less than 950 Hounsfield Unit, given as median and 25th and 75th percentiles (%LAA).
Collateral ventilation manifests as delayed filling of ventilation defects on imaging. However, delayed filling of ventilation defects can also be attributed to partial obstruction, air trapping or narrowing of peripheral airways [14]. By definition, a signal can only be classified as collateral ventilation if the observed delayed ventilation originates from corresponding reduction in expected ventilation in an adjacent pulmonary airspace [36]. In this study, delayed ventilation was defined as a detectable increase or persistent high signal over baseline in a particular lung ROI that is not seen elsewhere in the ventilated lung. We would refer to any observed delayed filling as simply delayed ventilation and reserve the term collateral ventilation for when this delayed filling can be further validated using gold standard methods such as Chartis®.
For delayed ventilation detection, the covariance of signals was computed using the following equation (Eq. 1):
where X and Y are the normalised temporal signals measured in the trachea and lung ROI, respectively; xi and yi are the signals in the trachea and lung ROI at timepoint i, respectively; μx and μy are their expected means; and n is the number of time steps during image acquisition. A value of 0.97 was chosen as the detection threshold, based on the calculated covariance of the control cohort, where visually there is minimal lobar collateral/delayed ventilation, in keeping with physiological expectation [36]. Cases with a covariance of less 0.97 were classed as having demonstrated delayed ventilation; all such cases were confirmed by visual inspection of the temporal signal maps by a clinical radiologist.
The amount of time delay was characterised by the temporal difference between the maximal ventilation signal in the trachea and that in the lung ROI. The tracheal 129Xe signal was normalised for each patient, since it is little affected by COPD and is most representative of true hyperpolarised MRI signal decay. In order for delayed ventilation to be observed, a minimum delay of 1 timestep (0.6 second) is required. This timescale is in keeping with that published in relevant literature [14].
In cases where delayed ventilation was observed, it was quantified by computing the difference in the integrated areas of the signal-time curve of the trachea and that of the lung ROI in question, which is referred to simply as the signal-time product difference in this paper. This concept is illustrated in Fig. 2.
A clinical radiologist with 2 years of thoracic radiology experience reviewed the DXeV-MRI and corresponding QCT coronal images visually for delayed ventilation and any discernable structural differences, such as incomplete fissures, that might have contributed to it.
Image data processing and statistical analyses were performed using in-house software developed in a MATLAB® environment (MathWorks). A p value < 0.05 was considered to be statistically significant. Spearman’s coefficient was computed to assess the correlations of spirometry test measurements, %LAA, trachea-lung ROI covariance, time delay and the quantity of delayed ventilation (signal-time product difference).
Results
To numerically capture and measure delayed ventilation, covariance was computed using Eq. 1. Results from the COPD cohort are given in Fig. 3a–d. Thirteen of the 15 COPD participants demonstrated some form of delayed ventilation, present in all four lung ROIs in six, three in two, two in three and one in the remaining two participants. By contrast, none of the participants in the control group demonstrated any delayed ventilation, both numerically and on visual inspection. The average covariance of the normal cohort is statistically significantly different from that of the COPD group (p = 0.0173).
For those where there was detectable delayed ventilation, the time delay and quantity of signal delay were calculated, the latter measured in signal-time product difference. The results are given in Fig. 3e and f, respectively. There were cases where a zero time delay was measured despite notable delayed ventilation; this is because of slower regional signal decay in the lung ROI in comparison to that in the trachea, rather than there being an actual temporal delay in its maximal signal.
Various correlation analysis results are given in Table 2. There was no statistically significant correlation with spirometry tests: between the covariance measure and %predicted FEV1 (p = 0.12), with delayed time (p = 0.84), nor quantity of delayed ventilation (p = 0.75). No correlation of statistical significance was found between %LAA and % predicted FEV1 (p = 0.57), which is in accordance with results reported in earlier literature [26]. However, statistically significant correlations between the trachea-lung ROI covariance and %LAA (R = − 0.40, p = 0.02) and time delayed (R = 0.34, p = 0.05) were established, both on a regional basis. No such correlation is, however, found with the quantity of delayed ventilation (p = 0.81).
The CT and DXeV-MRI data from a control cohort participant with healthy lungs and a COPD patient (Patient 1) are given in Fig. 4a and b, respectively. The corresponding fused end-inspiratory DXeV-MRI and 1H-MRI for the same participants are presented in Fig. 4d and f, and the corresponding CT data are given in Fig. 4c and e, respectively. The healthy participant did not demonstrate any emphysematous change on CT (Fig. 4c), nor any ventilation defects on DXeV-MRI (Fig. 4d). The %LAA for the COPD patient, on the other hand, was measured to be 21.9% and 26.3% for left and right lungs, respectively, and is confirmed by the presence of ventilation defects in those lungs (Fig. 4f). For this patient, the covariance measurements of the left upper, left lower, right upper and right lower lungs compared to the trachea are 0.69, 0.88, 0.85 and 0.88, respectively. Their corresponding time delays are 3.6, 3.6, 4.2 and 3.6 seconds, respectively, and the quantities of delayed ventilation (signal-time product difference) are 0.47, 1.02, 1.31 and 1.32, respectively.
Figure 5 shows the temporal ventilation signals in the trachea and lung ROIs for a control participant, a COPD patient with no detectable delayed ventilation (Patient 6), and a COPD patient with delayed ventilation in the right upper lung (Patient 3). The presence of delayed ventilation is clearly demonstrated in the second COPD participant. Figure 6 shows the anterior to posterior coronal temporal signals of a COPD patient with delayed ventilation in both lungs (Patient 14). The trachea signal presented in this case is the average signal computed from several coronal slices, as in the cases presented in Fig. 5. This is because due to tracheal anatomy, there is only signal on the central slices, precluding a coronal slice by slice comparison with the coronal lung ROI signals. There was no clear discrepancy between coronal slices in capturing the delayed ventilation. The corresponding DVeX-MRI images are given in Figs. 5 and 6, truncated for the purpose of illustrating key features in the dynamic sequence.
Discussion
The utility of DXeV-MRI for assessing delayed ventilation is supported by our experimental findings, where a quantifiable delayed ventilation was observed in at least one lung ROI in 87% of the COPD cohort and none in the control group (p = 0.0173), although the difference in group sizes should be noted when interpreting this result. The time delay was measured in the order of seconds with respect to the trachea signal, and the amount of delayed ventilation was quantified using signal-time product difference. Amongst the ROIs where delayed ventilation is detected, the mean time delay was 2.0 seconds (95% CI [1.3, 2.7]), and the mean signal-time product difference was 0.93 (95% CI [0.72, 1.14]). These results demonstrate a level of consistency of the degree of delayed ventilation in cases where it is observed. All cases where delayed ventilation was observed were also confirmed visually on the DXeV images by a clinical radiologist observer.
The prevalence of observed delayed ventilation is consistent with that published in earlier literature using hyperpolarised 3He time-resolved breath-hold MRI [14]. Our findings have established statistically significant correlations between the QCT-derived %LAA and the trachea-ROI covariance, and time delay, computed on a regional basis. This provided evidence that the presence and extent of delayed ventilation measured using our method correlated with the degree of emphysema, as measured on QCT. The ability to assess delayed ventilation may enable the assessment of collateral ventilation, pending further research work and clinical validation, to guide lung volume reduction therapy (LVRT) for certain COPD patients, achieved either surgically by lung resection or bronchoscopically by placing an endobronchial valve. Pulmonary lobes that are most likely to benefit from LVRT are those least affected by collateral ventilation [36]. The short table time of our HP 129Xe-MRI scan, in most cases under 5 minutes, would support its use in a wide range of applications without adding substantially to the existing clinical workload.
There are several limitations to our study. First, we have not compared our results to those from Chartis®, the current gold standard method of measuring collateral ventilation. It would be desirable as future work if we make this comparison to further validate our findings.
In this study, adjustment for the variation in the local oxygen concentration was not included in the analysis. T1 may vary between 26 to 33 seconds between different ROIs of the same participant, as we discussed in our previous work [28]. Considering the time delay was calculated over a time period of around 10 seconds, this T1 variation might result in up to 10% variation in signal-time product difference and/or time delay.
Finally, manual delineation of the lung ROIs was performed based on anatomical knowledge of the pulmonary lobes and observed fissures. On coronal DXeV slices, especially where there is a heavy presence of ventilation defects, lobar fissural identification may be challenging, in which case this information had to be extrapolated from neighbouring slices where the fissure was more visible. For this reason, the ROIs were labelled as lung areas, rather than actual pulmonary lobes. To perform a full lobar analysis, visualisation of the horizontal fissures on DXeV would be required, which, from our experience, is not possible on most DVeX-MRI scans involving COPD patients. In our past work with static 129Xe-MRI, lobar analysis has been achieved through the use of advanced machine learning tools to co-register MRI data to QCT, the latter of which is far superior in showing the pulmonary fissures. In that work, 1H-MRI was used as an intermediary, which provided key MRI anatomical information to facilitate inter-modal registration. However, this was not feasible in this dynamic sequence study, as a separate 1H-MRI would be needed for each time point in the breathing cycle. One possible solution to this could involve ventilation map modelling using CT and spirometry test data.
In summary, this study has demonstrated the use of DXeV-MRI as a non-invasive way of detecting and quantifying delayed ventilation in patients with COPD, and providing physiological information on regional pulmonary function during a full breathing cycle.
Abbreviations
- %LAA:
-
Percentage of low attenuations areas on CT in emphysema
- 129Xe:
-
Xenon isotope 129
- 133Xe:
-
Xenon isotope 133
- 3He:
-
Helium isotope 3
- CI:
-
Confidence interval
- COPD:
-
Chronic obstructive pulmonary disease
- CT:
-
Computed tomography
- DV:
-
Delayed ventilation
- DXeV:
-
Dynamic hyperpolarised xenon-129 ventilation sequence
- FEV1:
-
Forced expiratory volume in one second
- FRC:
-
Functional residual capacity
- FVC:
-
Forced vital capacity
- LVRT:
-
Lung volume reduction therapy
- MRI:
-
Magnetic resonance imaging
- PTK:
-
Pulmonary Toolkit
- QCT:
-
Quantitative computed tomography
- ROI:
-
Region of interest
- SEOP:
-
Spin-exchange optical pumping
References
Van Allen C, Lindskog G, Richter H (1931) Collateral respiration. Transfer of air collaterally between pulmonary lobules. J Clin Invest 10:559–590
Terry PB, Traystman RJ, Newball HH, Batra G, Menkes HA (1978) Collateral ventilation in man. N Engl J Med 298:10–15
Gompelmann D, Eberhardt R, Herth FJF (2013) Collateral ventilation. Respiration 86:515–520
Klooster K, ten Hacken NH, Hartman JE, Kerstjens HA, van Rikxoort EM, Slebos DJ (2015) Endobronchial valves for emphysema without interlobar collateral ventilation. N Engl J Med 373:2325–2335
Gompelmann D, Eberhardt R, Michaud G, Ernst A, Herth FJ (2010) Predicting atelectasis by assessment of collateral ventilation prior to endobronchial lung volume reduction: a feasibility study. Respiration 80:419–425
Yin Y, Schuhmann M, Gompelmann D et al (2016) CT surrogates of interlobar collateral ventilation: comparison against Chartis system measurements. Eur Respir J 48:OA473
Salanitri J, Kalff V, Kelly M et al (2005) 133Xenon ventilation scintigraphy applied to bronchoscopic lung volume reduction techniques for emphysema: relevance of interlobar collaterals. Intern Med J 35:97–103
Goo HW, Yang DH, Kim N et al (2011) Collateral ventilation to congenital hyperlucent lung lesions assessed on xenon-enhanced dynamic dual-energy CT: an initial experience. Korean J Radiol 12:25–33
Goo HW, Chae EJ, Seo JB, Hong SJ (2008) Xenon ventilation CT using a dual-source dual-energy technique: dynamic ventilation abnormality in a child with bronchial atresia. Pediatr Radiol 38:1113–1116
Adamson EB, Ludwig KD, Mummy DG, Fain SB (2017) Magnetic resonance imaging with hyperpolarized agents: methods and applications. Phys Med Biol 62:R81–R123
Flors L, Altes TA, Mugler JP 3rd et al (2015) New insights into lung diseases using hyperpolarized gas MRI. Radiologia 57:303–313
Mugler JP 3rd, Altes TA (2013) Hyperpolarized 129Xe MRI of the human lung. J Magn Reson Imaging 37:313–331
Marshall H, Collier GJ, Johns CS et al (2019) Imaging collateral ventilation in patients with advanced chronic obstructive pulmonary disease: relative sensitivity of 3 he and 129 Xe MRI. J Magn Reson Imaging 49:1195–1197
Marshall H, Deppe MH, Parra-Robles J et al (2012) Direct visualisation of collateral ventilation in COPD with hyperpolarised gas MRI. Thorax 67:613–617
Kern AL, Vogel-Claussen J (2018) Hyperpolarized gas MRI in pulmonology. Br J Radiol 91:1–16
Barskiy DA, Coffey AM, Nikolaou P et al (2017) NMR hyperpolarization techniques of gases. Chemistry 23:725–751
Ruset IC, Ketel S, Hersman FW (2006) Optical pumping system design for large production of hyperpolarized Xe129. Phys Rev Lett 96:053002
Stewart NJ, Chan HF, Hughes PJC et al (2018) Comparison of 3He and 129Xe MRI for evaluation of lung microstructure and ventilation at 1.5T. J Magn Reson Imaging 48:632–642
Svenningsen S, Kirby M, Starr D et al (2013) Hyperpolarized 3He and 129Xe MRI: differences in asthma before bronchodilation. J Magn Reson Imaging 38:1521–1530
Roos JE, McAdams HP, Kaushik SS, Driehuys B (2015) Hyperpolarized gas MR imaging: technique and applications. Magn Reson Imaging Clin N Am 23:217–229
Doganay O, Chen M, Matin T et al (2019) Magnetic resonance imaging of the time course of hyperpolarized 129 Xe gas exchange in the human lungs and heart. Eur Radiol 29:2283–2292
Cleveland ZI, Cofer GP, Metz G et al (2010) Hyperpolarized Xe MRimaging of alveolar gas uptake in humans. PLoS One 5:e12192
van Beek EJ, Wild JM, Kauczor HU, Schreiber W, Mugler JP 3rd, de Lange EE (2004) Functional MRI of the lung using hyperpolarized 3-helium gas. J Magn Reson Imaging 20:540–554
Shukla Y, Wheatley A, Kirby M et al (2012) Hyperpolarized 129Xe magnetic resonance imaging. Tolerability in healthy volunteers and subjects with pulmonary disease. Acad Radiol 19:941–951
Driehuys B, Martinez-Jimenez S, Cleveland ZI et al (2011) Chronic obstructive pulmonary disease: safety and tolerability of hyperpolarized 129 Xe MR imaging in healthy volunteers and patients. Radiology 262:279–289
Matin TN, Rahman N, Nickol AH et al (2017) Chronic obstructive pulmonary disease: lobar analysis with hyperpolarized 129 Xe MR imaging. Radiology 282:857–868
Pike D, Mohan S, Ma W et al (2015) Pulmonary imaging abnormalities in an adult case of congenital lobar emphysema. J Radiol Case Rep 9:9–15
Doganay O, Matin T, Chen M, et al (2018) Time-series hyperpolarized xenon-129 MRI of lobar lung ventilation of COPD in comparison to V/Q-SPECT/CT and CT. Eur Radiol, 10.1007/s00330-018-5888-y
Xiao S, Deng H, Duan C et al (2019) Highly and adaptively undersampling pattern for pulmonary hyperpolarized 129 Xe dynamic MRI. IEEE Trans Med Imaging 38:1240–1250
Marshall H, Parra-Robles J, Deppe MH et al (2014) 3He pO2 mapping is limited by delayed-ventilation and diffusion in chronic obstructive pulmonary disease. Magn Reson Med 71:1172–1178
Deninger AJ, Månsson S, Petersson JS et al (2002) Quantitative measurement of regional lung ventilation using 3He MRI. Magn Reson Med 48:223–332
Horn FC, Deppe MH, Marshall H et al (2014) Quantification of regional fractional ventilation in human subjects bymeasurement of hyperpolarized 3He washout with 2D and 3D MRI. J Appl Physiol (1985) 116:129–139
Holmes JH, Korosec FR, Du J et al (2007) Imaging of lung ventilation and respiratory dynamics in a single ventilation cycle using hyperpolarized he-3 MRI. J Magn Reson Imaging 26:630–636
Xiao S, Deng H, Duan C et al (2018) Considering low-rank, sparse and gas-inflow effects constraints for accelerated pulmonary dynamic hyperpolarized 129Xe MRI. J Magn Reson 290:29–37
Doganay O, Matin TN, Mcintyre A et al (2018) Fast dynamic ventilation MRI of hyperpolarized 129Xe using spiral imaging. Magn Reson Med 79:2597–2606
Cetti E, Moore A, Geddes DM (2006) Collateral ventilation. Thorax 61:371–373
Acknowledgements
The authors would like to thank Claire Trumper, Jennifer Lee and Kenneth Jacob for their role in xenon gas production and delivery, Andrea Byles and Tara Harris for their help with patient recruitment and supervision, and Radiology and Respiratory Physiology Department staff for their assistance in performing the study scans and lung function tests.
Funding
This study has received funding from the National Institute for Health Research (NIHR) Oxford Biomedical Research Centre Programme and clinical trial sponsorship by the Oxford University Hospitals NHS Foundation Trust.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Fergus Gleeson.
Conflict of interest
The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was obtained from all subjects (patients) in this study.
Ethical approval
Institutional Review Board approval was obtained. The study was approved by the UK National Research Ethics Service South Central Committee (Berkshire, UK, REC 11/SC/0487 and REC 11/SC/0488).
Study subjects or cohorts overlap
Some study subjects or cohorts have been previously reported in:
1. Matin TN, Rahman N, Nickol AH, et al (2017) Chronic obstructive pulmonary disease: lobar analysis with hyperpolarized 129 Xe MR imaging. Radiology 282:857–868.
2. Doganay O, Matin T, Chen M, et al (2018) Time-series hyperpolarized xenon-129 MRI of lobar lung ventilation of COPD in comparison to V/Q-SPECT/CT and CT. Eur Radiol. DOI: https://doi.org/10.1007/s00330-018-5888-y.
3. Doganay O, Matin TN, Mcintyre A, et al (2018) Fast dynamic ventilation MRI of hyperpolarized 129 Xe using spiral imaging. Magn Reson Med 79:2597–2606.
Methodology
• prospective
• experimental
• performed at one institution
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Chen, M., Doganay, O., Matin, T. et al. Delayed ventilation assessment using fast dynamic hyperpolarised Xenon-129 magnetic resonance imaging. Eur Radiol 30, 1145–1155 (2020). https://doi.org/10.1007/s00330-019-06415-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-019-06415-1