Abstract
Purpose
Sumitomo Bakelite has developed a steerable microcatheter with a mechanism on the handle for changing the direction of the catheter tip. Experiments were conducted in an animal model to evaluate the practicality of the catheter with the goal of introduction into clinical use.
Materials and Methods
The steerable microcatheter has the following specifications: external diameter 2.4F/2.9F (distal/proximal portion); internal diameter 0.021 inches; effective length 125 cm; and it is fitted with a dial on the proximal portion for moving the tip. Five interventional radiologists evaluated the microcatheter in animal experiments using piglets. The test catheter was compared with commercial straight microcatheters based on procedure time required to insert the catheter into two prescribed patterns of arterial branching (P-1: hepatic artery, P-2: iliac artery), fluoroscopy time, amount of contrast agent used, and frequency of guidewire use. Furthermore, the renal arteries into which each catheter was inserted were excised and examined for the extent of vascular damage.
Results
Compared to the commercial products, the steerable microcatheter used without a guidewire required shorter procedure time (43.0/75.2% reductions in P-1/P-2), fluoroscopy time tended to be shorter (5.1/74.1% reductions), and the amount of contrast agent used tended to be lower (38.2/52.3% reductions). No damage was seen to the luminal surface of the renal arteries traversed by either microcatheter. During the experiment, there were no problems or defects with either microcatheter while intravascular procedure.
Conclusion
These animal experiments confirmed that the steerable microcatheter had the functionality required for its development as a commercial medical device.
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs00270-017-1789-8/MediaObjects/270_2017_1789_Fig1_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs00270-017-1789-8/MediaObjects/270_2017_1789_Fig2_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs00270-017-1789-8/MediaObjects/270_2017_1789_Fig3_HTML.jpg)
Similar content being viewed by others
References
Llovet JM, Real MI, Montaña X, et al. Arterial embolisation or chemoembolisation versus symptomatic treatment in patients with unresectable hepatocellular carcinoma: a randomised controlled trial. Lancet. 2002;359(9319):1734–9.
Lo CM, Ngan H, Tso WK, et al. Randomized controlled trial of transarterial lipiodol chemoembolization for unresectable hepatocellular carcinoma. Hepatology. 2002;35(5):1164–71.
Llovet JM, Bruix J. Systematic review of randomized trials for unresectable hepatocellular carcinoma: chemoembolization improves survival. Hepatology. 2003;37(2):429–42.
Matsui O, Kadoya M, Yoshikawa J, et al. Small hepatocellular carcinoma: treatment with subsegmental transcatheter arterial embolization. Radiology. 1993;188(1):79–83.
Miyayama S, Matsui O, Yamashiro M, et al. Ultraselective transcatheter arterial chemoembolization with a 2-f tip microcatheter for small hepatocellular carcinomas: relationship between local tumor recurrence and visualization of the portal vein with iodized oil. J Vasc Interv Radiol. 2007;18(3):365–76.
Yamakado K, Miyayama S, Hirota S, et al. Hepatic arterial embolization for unresectable hepatocellular carcinomas: do technical factors affect prognosis? Jpn J Radiol. 2012;30(7):560–6.
de Baere T, Arai Y, Lencioni R, et al. Treatment of liver tumors with lipiodol TACE: technical recommendations from experts opinion. Cardiovasc Interv Radiol. 2016;39(3):334–43.
Homma A, Oridate N, Suzuki F, et al. Superselective high-dose cisplatin infusion with concomitant radiotherapy in patients with advanced cancer of the nasal cavity and paranasal sinuses: a single institution experience. Cancer. 2009;115(20):4705–14.
Lange M, Bültel H, Weglage H, Löffeld P, Wichter T. Using a steerable guiding sheath to implant an AMPLATZER™Amulet™ Left Atrial Appendage Occluder for prevention of thromboembolic stroke. Int J Cardiol. 2016;221:466–7.
Gross DM, Cox MA, Denson SB, Ferguson L. Unique use of a tip-deflecting guide wire in removing a catheter embolus from an infant. Pediatr Cardiol. 1987;8(2):117–9.
Soyama T, Yoshida D, Sakuhara Y, Morita R, Abo D, Kudo K. The steerable microcatheter: a new device for selective catheterisation. Cardiovasc Interv Radiol. 2017;40(6):947–52.
Ikoma A, Nakai M, Kamisako A, Tanaka R, Sonomura T. A guide for interventional radiology experiments using swine. Jpn J Interv Radiol. 2016;31(1):26–33.
Rose SC, Halstead GD, Narsinh KH. Pressure-directed embolization of hepatic arteries in a porcine model using a temporary occlusion balloon microcatheter: proof of concept. Cardiovasc Interv Radiol. 2017. doi:10.1007/s00270-017-1753-7.
Acknowledgements
This study was supported by Sumitomo Bakelite Co., Ltd..
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
All authors declare that they received financial assistance from Sumitomo Bakelite Co., Ltd. in this experiment.
Rights and permissions
About this article
Cite this article
Inaba, Y., Arai, Y., Sone, M. et al. Experiments for the Development of a Steerable Microcatheter. Cardiovasc Intervent Radiol 40, 1921–1926 (2017). https://doi.org/10.1007/s00270-017-1789-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-017-1789-8