Abstract
Purpose
Persistent acromioclavicular joint (ACJ) instability following high grade injuries causes significant symptoms. The importance of horizontal plane stability is increasingly recognised. There is little evidence of the ability of current implant methods to restore native ACJ stability in the vertical and horizontal planes. The purpose of this work was to measure the ability of three implant reconstructions to restore native ACJ stability.
Methods
Three groups of nine fresh-frozen shoulders each were mounted into a robotic testing system. The scapula was stationary and the robot displaced the clavicle to measure native anterior, posterior, superior and inferior (A, P, S, I) stability at 50 N force. The ACJ capsule, conoid and trapezoid ligaments were transected and the ACJ was reconstructed using one of three commercially available systems. Two systems (tape loop + screw and tape loop + button) wrapped a tape around the clavicle and coracoid, the third system (sutures + buttons) passed directly through tunnels in the clavicle and coracoid. The stabilities were remeasured. The data for A, P, S, I stability and ranges of A–P and S–I stability were analyzed by ANOVA and repeated-measures Student t tests with Bonferroni correction, to contrast each reconstruction stability versus the native ACJ data for that set of nine specimens, and examined contrasts among the reconstructions.
Results
All three reconstructions restored the range of A–P stability to that of the native ACJ. However, the coracoid loop devices shifted the clavicle anteriorly. For S–I stability, only the sutures + buttons reconstruction did not differ significantly from native ligament restraint.
Conclusions
Only the sutures + buttons reconstruction, that passed directly through tunnels in the clavicle and coracoid, restored all stability measures (A, P, S, I) to the native values, while the tape implants wrapped around the bones anteriorised the clavicle. These findings show differing abilities among reconstructions to restore native stability in horizontal and vertical planes. (300 words)
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Acromioclavicular joint (ACJ) injuries constitute 9% of shoulder girdle injuries [3]. The classic Rockwood classification correlates poorly with clinical symptoms, primarily due to its inability to accurately predict the anatomical structures injured in each case [7]. Clinical evaluation of ACJ stability has been based on anterior–posterior (AP) radiographs to measure superior displacement of the lateral clavicle [18], while it is difficult to measure horizontal displacement [1]. This is why methods of ACJ stabilisation have traditionally focused on preventing superior displacement of the clavicle [15, 20].
Many methods of ACJ stabilisation have been described, with no clear gold standard [1, 13, 17, 21]. Recently a more three dimensional understanding of the ACJ has developed, with a better appreciation of the restraints of horizontal displacements [8, 12, 22] and of the ligamentous structures [19]. Scheibel et al. [23] recognised persistent symptomatic horizontal plane instability in a high number of reconstructions that focused on coraco-clavicular (CC) ligament reconstruction alone. Whatever method is used for ACJ reconstruction it must be able to reproduce as closely as possible the native joint stability in both the vertical and horizontal planes. Jari et al. [9] measured superior and AP stability of the native and reconstructed ACJ using a robot. However, despite the availability of that test method, there has been little data published on the stability of ACJ reconstructions in both planes.
Several manufacturers have introduced artificial implants for ACJ stabilisation, with variations among their configurations of tapes and sutures and their fixation means, but there is little data available on their ability to stabilise the ACJ. Therefore a biomechanical study has been conducted to compare the potential of three commonly used reconstruction techniques based on artificial implants to restore the native stability (laxity) of the superior suspensory complex of the shoulder. This information should be clinically useful in guiding the choice and use of implants for ACJ reconstruction.
The aim of this work was to determine whether the three reconstruction methods could reproduce native ACJ stability in both vertical and horizontal planes. The null hypothesis was that there would not be significant differences among the three reconstruction methods, nor versus the native ACJ stability.
Materials and methods
Specimen preparation
Following Imperial College Healthcare Tissue Bank Research Ethics Committee approval R18045, 31 fresh-frozen cadaveric shoulders were obtained. Four were used in a pilot study, leaving 27 for analysis: 12 male and 15 female aged 55 ± 13 years. Each specimen comprised the scapula, clavicle, and proximal half-length of the humerus, with intact skin and soft tissues. Specimens were thawed at room temperature for 24 h prior to preparation.
The skin and overlying soft tissues were excised to expose the ACJ capsular ligaments, coracoid, clavicle, and trapezoid and conoid ligaments. The humeral shaft, scapular body and clavicle were exposed. The scapula inferior to the posterior spine, and the medial clavicle, were cleared of soft tissues. Twenty-five mm of the inferior corner of the scapula, and 20 mm of the medial clavicle were excised. The prepared medial end of the clavicle was secured in a cylindrical metal pot with polymethylmethacrylate bone cement. The inferior part of the scapular body was similarly secured in a rectangular metal mounting.
Robotic stability testing
The scapula mounting was fixed rigidly to the base of a robotic joint testing system (TX90, Staubli AG, Horgen, Switzerland). This system included a load cell which could measure in 6 degrees of freedom (DoF) (Omega 85, ATI Industrial Automation Co., Apex, NC). Previous work had validated this system to ± 0.11 mm in translation and ± 0.13° in rotation and in both native and prosthetic human joints [2, 14] using the method of Fujie [5]. The load cell was mounted onto the free end of the robot arm and the clavicle secured to it via a tubular mounting [12]. The ‘neutral’ datum point was established by moving the clavicle to minimise the forces and torques across the ACJ. The robot was initially operated in ‘force control’ to measure motion of the clavicle when a displacing force was imposed in a chosen direction. As the robot moved, it monitored any coupled forces that built up in other DoF and moved to minimise them. This led to a path of motion being defined while maintaining a constant rotational orientation. Force and displacement data were recorded continuously.
To measure the stability of the intact ACJ complex, the clavicle was moved in the AP and SI directions at a speed of less than 50 mm/min until a displacing force of 50 N was reached. The AP movement was in the horizontal plane and SI movement vertical. The orientation of the AP axis was defined by taking the coronal plane to bisect the angle between the plane of the scapula and long axis of the clavicle when the shoulder was viewed from the superior aspect. The pilot study on four shoulders confirmed that repeated displacements to the motion limits reached at 50 N displacing force with the ACJ intact did not cause irreversible elongation of the soft-tissues stabilising the ACJ; a similar study had used 70 N [9]. All tests were run three times in A, P, S and I directions.
The robot was switched off and the conoid and trapezoid ligaments and the ACJ capsule were transected with the clavicle held at the anatomical (neutral load) position. The specimens were separated into three groups of nine, which each received one of three reconstruction methods.
When the robot was switched on after the ACJ reconstruction, there was usually a load acting on the clavicle, due to tension in the reconstruction. The robot was then moved to a new neutral position. This movement was recorded, so that all data were related to the original position of the native ACJ. The robot then repeated the stability tests as before, moving the clavicle in 3 DoF of translation but without coupled rotations until the 50 N limit was reached. The displacement (mm) away from the neutral position was recorded. This was done three times for each specimen in A, P, S and I directions.
In all subsequent results, the stabilities presented are movement of the clavicle in relation to the fixed scapula.
Reconstruction methods
Tape loop plus screw fixation reconstruction
The Tape loop + screw fixation implant (Lockdown, Lockdown Surgical Inc., Chanhassen, MN, USA) was used according to the manufacturer’s recommended technique (Fig. 1). The conoid tubercle was identified and provisionally marked for more accurate placement. The lateral end of the clavicle was not excised as reduction was achievable without impingement in all cases. The base of the coracoid was exposed both medially and laterally to allow a hooked passing instrument (Lockdown Surgical Inc.) to be passed under it from medial to lateral. Once the tip was visualised, the length gauge was passed through. The metal tip of the length gauge was then passed through a loop in the gauge, to cinch it around the base at the apex of the ‘knuckle’ of the coracoid. The metal tip was then passed behind the clavicle at the level of the conoid tubercle and pulled over from posterior to anterior. A marking on the length gauge was read on the anterior cortex of the clavicle to determine the required implant length. The length gauge was passed back under the clavicle and uncinched, then used to pass the implant around the base of the coracoid and back over the clavicle in the described technique. Using the length gauge to assist in holding the reduction, a 2.5 mm AP tunnel was drilled, exiting lateral to the synthetic ligament on the posterior aspect of the clavicle. The length of the hole was measured and a 3.5 mm cortical bone screw was passed through the hard loop of the ligament and into the drilled hole to secure the ligament.
Tape loop plus button fixation reconstruction
The manufacturer’s technique for the Tape loop + button fixation implant (Infinity-Lock, Xiros Ltd., Leeds, UK) (Fig. 2) exposed the coracoid and a side-specific disposable hooked passing instrument (Xiros Ltd.) was passed underneath the coracoid from medial to lateral. A nitinol loop was passed through the hook and grasped a suture on the looped end of the synthetic ligament tape to pull it under the coracoid. The free ends of the implant were passed through the loop and the implant was cinched to the base of the coracoid, ensuring that it was at the apex of the ‘knuckle’ of the coracoid. A 4-mm hole was drilled through the clavicle at the level of the conoid tubercle, in the centre of the width of the clavicle and aiming at the base of the coracoid, using a guidewire and cannulated drill bit. The nitinol loop was passed through the hole to retrieve the free ends of the implant from below the clavicle. The ends of the implant were passed through a titanium alloy button and a single throw was made over it. One free end was passed over the posterior aspect of the clavicle and retrieved from underneath it, then the two ends were tied in a knot at the anterior aspect of the clavicle, avoiding a large bulk on top of the button (Fig. 2). Optional locking sutures in the knot were not placed during the biomechanical study.
Transosseous sutures with buttons reconstruction
The manufacturer’s standard technique was used for the transosseous sutures + buttons reconstruction (FiberTape sutures with Dog Bone buttons, Arthrex GmbH, Naples, FL, USA).
The reconstruction used two 3-mm drill holes, one at the base of the coracoid and another through the clavicle directly above it, positioned colinearly with the ACJ reduced. Two sutures with an attached titanium alloy button were shuttled up through the drill holes and then passed through another titanium alloy button above the clavicle (Fig. 3). The button below the coracoid was checked to ensure that it was snug to the bone prior to tightening the sutures and tying them in a knot on top of the button on the clavicle.
Statistical analysis
A power analysis based on published data [9] using G*Power v 3.1.9.7 showed that a change of laxity of 3 mm could be identified with 95% confidence and 82% power with eight specimens per test group.
The Shapiro–Wilk test confirmed that data were distributed normally with p = 0.05. For each of the three reconstruction methods the data were analyzed to calculate the change of stability from the native to the reconstructed state, with nine shoulders per group, in each test direction: A, P, S, I. Increased stability meant that the ACJ was less lax—overconstrained—post reconstruction, and vice-versa.
The ability of the three reconstructions to restore native stability was examined in each loading direction by one-way analysis of variance, with differences taken to be significant with p < 0.05. The data for each of the three reconstructions were examined for changes of stability from the native by two-way repeated-measures Student t tests, with p < 0.0167 with Bonferroni correction. The changes of stability from native values among the three reconstruction methods were then contrasted by unpaired two-way t tests, with p < 0.0167 with Bonferroni correction.
Results
Anterior stability
The ability of the three methods to restore native anterior stability differed significantly (p = 0.0180, Fig. 4). Post-testing found a significant difference of residual instability between the sutures plus buttons and tape loop plus screw methods (p = 0.0052). Both of the tape loop reconstruction methods that wrapped around the clavicle (with screw or button fixation) allowed residual anterior instability to persist (p < 0.0167), while the trans-osseous sutures method did not.
Posterior stability
The ability of the three reconstruction methods to restore native posterior stability differed significantly (p = 0.0312, Fig. 5). Although the tape loop plus screw fixation reconstruction caused overconstraint (p < 0.001), post-testing did not find significant differences among the reconstructions (n.s.).
Range of AP stability
The tape loop plus button procedure gave a range of AP stability (laxity) that did not differ significantly from the native ACJ. The anterior stability was significantly under-constrained, posterior stability tended to be overconstrained, so the mean AP position of the clavicle shifted 5 mm anteriorly.
The transosseous sutures plus buttons procedure gave range of AP stability, anterior stability and posterior stability that did not differ significantly from the native ACJ. The non-significant tendencies to underconstrain anteriorly and overconstrain posteriorly shifted the mean AP position 2 mm anteriorly.
The tape loop plus screw procedure overconstrained posterior displacement of the clavicle, and underconstrained anterior displacement. Although the range of AP stability was not found to differ significantly from the native stability, the mean AP position of the clavicle was shifted 7 mm anteriorly.
Superior stability
The superior stability of the three reconstruction methods differed significantly (p < 0.001, Fig. 6). The transosseous sutures plus buttons method restored native stability closer than the tape loop plus screw (p < 0.001) and the tape loop plus button (p = 0.0023) methods. The tape loop plus screw reconstruction allowed residual superior instability to persist (p = 0.0004) while the tape loop plus button did not (n.s.).
Inferior stability
The ability of the three reconstruction methods to restore native inferior stability did not differ significantly (n.s., Fig. 7). The reconstructions did not leave persisting residual instability (n.s.).
Range of superior-inferior stability
The tape loop plus button procedure was significantly less stable in S-I than the native ACJ (p = 0.0035). However, each of the superior and inferior stabilities was not significantly different to native stability (n.s.). These combined effects shifted the mean position of the clavicle 2 mm superiorly.
The transosseous sutures plus buttons procedure gave range of SI, superior, and inferior stabilities that did not differ significantly from the native ACJ. The tendencies to over-constrain superiorly and under-constrain inferiorly shifted the mean SI position of the clavicle inferiorly by 2 mm.
The tape loop plus screw procedure under-constrained superior displacement of the clavicle, and the inferior stability matched the native ACJ, so the reconstructed ACJ was significantly less stable in S-I than the native ACJ (p = 0.0014). The mean position of the range of SI displacements was elevated 4 mm after this reconstruction.
Discussion
The main finding of this study was that the three ACJ stabilisation methods had differing abilities to restore native ACJ stability in SI and AP directions, thus the null hypothesis was rejected. All three methods restored AP stability to native values, but the tapes which wrapped around the coracoid displaced the clavicle anteriorly from its native resting position, with resulting anterior instability and posterior overconstraint. The anterior and posterior stabilities and the range of AP stability were restored to native values by the transosseous sutures plus buttons construct. The native range of SI stability was restored by the transosseous sutures plus buttons construct, but the other two constructs left significant underconstraint. The superior laxity of the reconstructed ACJ did not differ significantly from the native laxity for the transosseous sutures plus buttons and tape loop plus button constructs.
Few biomechanical studies have compared implants for treatment of ACJ injuries despite their increasing availability. Ladermann et al. [11] compared ACJ and CC cerclage sutures with a suture tape plus bone anchors construct and a superior clavicle hook plate [24], finding that the cerclage sutures mimicked physiological behaviour more closely. Tulner et al. [25] reported that the tape loop plus screw fixation (Lockdown Surgical Inc) implant shifted the clavicle anteriorly, as in the present study, but this had no clinical importance. The superior displacement was similar to controls immediately post operatively but increased 4 mm by 6 months post-reconstruction. The present study found that this implant tended to overconstrain posterior displacement of the clavicle in relation to the acromion. This may be clinically beneficial if it resists anterior displacement of the scapula and collapse under load in vivo, but the reduced anterior/superior restraint when tested in the robot is a clinical surgical consideration when undertaking these procedures.
Reconstructions that require tunnels through the coracoid have led to fractures due to malposition or excessive sizing of the tunnels or possible attritional wear from polyethylene core suture material [3, 6, 16]. Fracture of the base of the coracoid has also been reported [10]. Techniques in the present study minimise coracoid fracture risk by requiring only 3 or 4 mm tunnels. There were no fractures of the coracoid, but this work was not a study of load to failure.
The transosseous sutures reconstruction gave the closest return to the native SI stability, with significant differences for the coracoid loop devices. This might be because it is difficult to remove all the excess slack and soft tissue interposition below the coracoid. In comparison, the trans-coracoid device is easier to tighten because the sutures pass directly between the two titanium buttons.
Clinical studies of the three implants have indicated similar patient satisfaction and complication rates. This should be taken into consideration along with the results of this biomechanical study when selecting an implant for this procedure.
The reconstructions were performed by four fellowship trained upper limb surgeons using the manufacturers’ techniques. However, the placement of drill holes and tensioning of knots was left to each surgeon’s individual judgment and were influenced by the anatomy of each specimen. This reflects real life circumstances during the use of these implants by different surgeons.
A limitation of the experiment was that it only measured stability at ‘time zero’, as though intra-operatively, and could not account for longer-term biological or mechanical effects which may allow return of laxity post-surgery [4]. The robot held the clavicle at its native position while the reconstructions were performed, after the soft tissues had been transected. This consistent test configuration avoided the need to remove/replace the specimen at each stage of surgery. A strength of this experimental design is the repeated-measures analysis of the reconstruction stability versus the native stability for each joint.
This work may be useful for the clinician in demonstrating the ability of the artificial constructs to restore native ACJ stability in both the vertical and horizontal planes. This study supports the use of the three commonly used synthetic AC joint ligament repair methods. The coracoid looped reconstructions in vitro have been shown to lead to anterior instability and posterior overconstraint. This may explain some previously reported anterior displacement of the clavicle in vivo which has not been shown to cause any clinical sequelae, and indeed may be beneficial in resisting physiological loads [25].
Conclusion
All three implants restored the native range of horizontal plane stability of the ACJ, but the tapes that looped around the coracoid shifted the resting position of the lateral clavicle anteriorly. The transosseous coracoid suture device also provided better superior stability than the tape loop devices, with the closest reproduction of the native ACJ stability.
References
Aliberti GM, Kraeutler MJ, Trogan JD, Mulcahey MK (2020) Horizontal instability of the acromioclavicular joint: a systematic review. Am J Sports Med 48:504–510
Athwal KK, El Daou H, Inderhaug E, Manning W, Davies AJ, Deehan DJ, Amis AA (2016) An in-vitro analysis of medial structures and a medial soft-tissue reconstruction in a constrained condylar total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 25:2646–2655
Dyrna F, de Oliveira CCT, Nowak M, Voss A, Obopilwe E, Braun S, Pauzenberger L, Imhoff AB, Mazzocca AD, Beitzel K (2018) Risk of fracture of the acromion depends on size and orientation of acromial bone tunnels when performing acromioclavicular reconstruction. Knee Surg Sports Traumatol Arthrosc 26:275–284
Frank RM, Cotter EJ, Leroux TS, Romeo AA (2019) Acromioclavicular joint injuries: evidence-based treatment. J Am Acad Orthop Surg 17(27):e775-788
Fujie H (1993) The use of robotics technology to study human joint kinematics: a new methodology. J Biomech Eng 115:211–217
Gerhardt DC, VanDerWerf JD, Rylander LS, McCarty EC (2011) Postoperative coracoid fracture after transcoracoid acromioclavicular joint reconstruction. J Shoulder Elbow Surg 20(5):e6-10
Granville-Chapman J, Torrance E, Rashid A, Funk L (2018) The Rockwood classification in acute acromioclavicular joint injury does not correlate with symptoms. J Orthop Surg. https://doi.org/10.1177/2309499018777886
Hislop P, Sakata K, Ackland DC, Gotmaker R, Evans MC (2017) Acromioclavicular joint stabilization. A biomechanical study of bidirectional stability and strength. Orthop J Sports Med. https://doi.org/10.1177/2325967119836751
Jari R, Costic RS, Rodosky MW, Debski RE (2004) Biomechanical function of surgical procedures for acromioclavicular joint dislocations. Arthroscopy 20:237–245
Jeon IH, Dewnany G, Hartley R, Neumann L, Wallace W (2007) Chronic acromioclavicular joint separation: the medium term results of coracoclavicular ligament reconstruction using braided polyester prosthetic ligament. Injury 38:1247–1253
Ladermann A, Gueorguiev B, Stimec B, Fasel J, Rothstick S, Hoffmeyer P (2013) Acromioclavicular joint reconstruction: a comparative biomechanical study of three techniques. J Shoulder Elbow Surg 22:171–178
Lee J, El-Daou H, Alkoheji M, Carlos A, Di Mascio L, Amis AA (2021) Ligamentous and capsular restraints to anterior-posterior and superior-inferior laxity of the acromioclavicular joint—a biomechanical study. J Shoulder Elbow Surg 30:1251–1256
Li X, Ma R, Bedi A, Dines D, Altchek D, Dines J (2014) Management of acromioclavicular joint injuries. J Bone Joint Surg Am 96:73–84
Lord BR, El Daou H, Zdanowicz U, Smigielski R, Amis AA (2019) The role of fibers within the tibial attachment of the anterior cruciate ligament in resisting tibial displacement. Arthroscopy 35:2101–2111
Maier D, Jaeger M, Reising K, Feucht M, Südkamp N, Izadpanah K (2016) Injury patterns of the acromioclavicular ligament complex in acute acromioclavicular joint dislocations: a cross-sectional fundamental study. BMC Musculoskelet Disord 17:385
Martetschlarger F, Saier T, Weigert A, Herbst E, Winkler M, Henschel J, Augat P, Imhoff A, Braun S (2016) Effect of coracoid drilling for acromioclavicular joint reconstruction techniques on coracoid fracture risk: a biomechanical study. J Arthroscopy 32:982–987
Mazzocca AD, Arciero RA, Bicos J (2007) Evaluation and treatment of acromioclavicular joint injuries. Am J Sports Med 35:316–329
Muccioli GMM, Manning C, Wright P, Grassi A, Zaffagnini S, Funk L (2016) Acromioclavicular joint reconstruction with the LARS ligament in professional versus non-professional athletes. Knee Surg Sports Traumatol Arthrosc 24:1961–1967
Nolte P-C, Ruzbarsky JJ, Midtgaard KS et al (2021) Quantitative and qualitative surgical anatomy of the acromioclavicular joint capsule and ligament. Am J Sports Med 49:1183–1191
Nuchtern JV, Sellenschloh K, Bishop N, Jauch S, Briem D, Hoffman M, Lehmann W, Pueschel K, Morlock M, Rueger JM, Großterlinden L (2013) Biomechanical evaluation of 3 stabilization methods on Acromioclavicular joint dislocation. Am J Sports Med 41:1387–1394
Phadke A, Bakti N, Bawale R, Singh B (2019) Current concepts in management of ACJ injuries. J Clin Orthop Trauma 10:480–485
Saccomanno MF, DeLeso C, Milano G (2014) Acromioclavicular joint instability: anatomy, biomechanics and evaluation. Joints 2:87–92
Scheibel M, Dröschel S, Gerhardt C, Kraus N (2011) Arthroscopically assisted stabilization of acute high-grade acromioclavicular joint separations. Am J Sports Med 39:1507–1516
Sim E, Schwarz N, Hocker K, Berzlanovich A (1995) Repair of complete acromioclavicular separations using the acromioclavicular-hook plate. Clin Orthop Relat Res 314:134–142
Tulner SA, Fowler A, Edwards K, Kumar V, Manning P, Geoghegan J, Wallace W (2019) Radiographic appearance of the distal clavicle in relation to the acromion after acromioclavicular joint reconstruction using a braided polyester mesh. J Clin Orthop Trauma 10:486–491
Acknowledgements
Human samples used in this research project were obtained from the MedCure tissue bank, Portland, OR, with permission from the Imperial College Healthcare Tissue Bank (ICHTB). ICHTB is supported by the National Institute for Health Research (NIHR) Biomedical Research Centre based at Imperial College Healthcare NHS Trust and Imperial College London. ICHTB is approved by Wales REC3 to release human material for research (17/WA/0161). The robotic joint testing system was funded by the Wellcome Trust and EPSRC supported Centre of excellence in application of technology to osteoarthritis at Imperial College London.
Funding
This study was funded by a grant from Xiros Ltd, paid to a research account of Imperial College London.
Author information
Authors and Affiliations
Contributions
MA: experimental work, manuscript review. HED: experimental work, robot controlling, data processing, manuscript review. JL: experimental work, manuscript review. AC: project planning, experimental work, manuscript review. LDM: project planning, grant application, experimental work, manuscript review. AAA: project planning, grant writing, management of experimental work, statistical analysis, manuscript editing.
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no conflicts of interest beyond the funding acknowledged.
Ethical approval
ICHTB Human Tissues Authority licence 12275, application R18045.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Alkoheji, M., El-Daou, H., Lee, J. et al. Acromioclavicular joint reconstruction implants have differing ability to restore horizontal and vertical plane stability. Knee Surg Sports Traumatol Arthrosc 29, 3902–3909 (2021). https://doi.org/10.1007/s00167-021-06700-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00167-021-06700-x