Abstract
Objective
This study aimed to prospectively assess the visibility of interstitial needles on transrectal ultrasound (TRUS) in cervical cancer brachytherapy patients and evaluate its impact on implant and treatment plan quality.
Material and methods
TRUS was utilized during and after applicator insertion, with each needle’s visibility documented through axial images at the high-risk clinical target volume’s largest diameter. Needle visibility on TRUS was scored from 0 (no visibility) to 3 (excellent discrimination, margins distinct). Quantitative assessment involved measuring the distance between tandem and each needle on TRUS and comparing it to respective magnetic resonance imaging (MRI) measurements. Expected treatment plan quality based on TRUS images was rated from 1 (meeting all planning objectives) to 4 (violation of High-risk clinical target volume (CTVHR) and/or organ at risk (OAR) hard constraints) and compared to the final MRI-based plan.
Results
Analysis included 23 patients with local FIGO stage IB2-IVA, comprising 41 applications with a total of 230 needles. A high visibility rate of 99.1% (228/230 needles) was observed, with a mean visibility score of 2.5 ± 0.7 for visible needles. The maximum and mean difference between MRI and TRUS measurements were 8 mm and –0.1 ± 1.6 mm, respectively, with > 3 mm discrepancies in 3.5% of needles. Expected treatment plan quality after TRUS assessment exactly aligned with the final MRI plan in 28 out of 41 applications with only minor deviations in all other cases.
Conclusion
Real-time TRUS-guided interstitial needle placement yielded high-quality implants, thanks to excellent needle visibility during insertion. This supports the potential of TRUS-guided brachytherapy as a promising modality for gynecological indications.
Similar content being viewed by others
Introduction
Currently, standard of care in the treatment of locally advanced cervical cancer is radiochemotherapy including brachytherapy [1]. During the past two decades, the performance of brachytherapy has been substantially improved by the integration of MRI into treatment planning and by the development of combined intracavitary/interstitial applicators. Use of MRI enables precise visualization of an adaptive target volume, whereas combined intracavitary/interstitial applicator systems allow the dose to be shaped accordingly. The main advantage of these applicators is that in addition to the standard intracavitary applicator (e.g., tandem/ring or tandem/ovoid), interstitial needles can be inserted through predefined holes in the ring or ovoid into more lateral parts of the cervix and/or parametria. However, the optimal workflow and in particular the most adequate utilization of the numerous needle positions for an individual patient remain unclear. The standard procedure is to choose the positions based on clinical estimation guided by gynecological examination and any prebrachytherapy MRI scans, if available. As the needles are directly linked to the intracavitary applicator, correct positioning of the tandem is of utmost importance. Nevertheless, even with optimal preparation such as full MRI-based preplanning [2] with use of 3D-printed applicators, uncertainty remains intraoperatively regarding whether the needles have actually reached the intended locations. Therefore, there has recently been increasing interest in the intraoperative use of transabdominal [3] and transrectal ultrasound (TRUS) in gynecological brachytherapy
TRUS has been proven to be noninferior to MRI in terms of tumor and brachytherapy target volume assessment for cervical cancer, albeit until now with major limitations in applicator depiction [4,5,6,7]. While real-time needle visualization and guidance is standard of care in prostate cancer, it is still in its infancy for gynecological brachytherapy. The aim of the current study is to analyze visibility and spatial differences of interstitial needles in TRUS in comparison to MRI in patients treated with magnetic resonance image-guided adaptive brachytherapy (MR-IGABT) for cervical cancer with combined intracavitary/interstitial applicators and to predict the expected treatment plan quality by TRUS.
Materials and methods
Patients and treatment
This is a prospective single-arm cohort study which was approved by the institutional ethics committee. Patients were eligible if they had (1) biopsy-proven cervical cancer with a local FIGO stage IB-IVA, (2) treatment with MR-IGABT, (3) utilization of a combined intracavitary/interstitial approach, and (4) written informed consent. The overall treatment comprised pelvic external beam radiotherapy (45 Gy in 25 fractions of 1.8 Gy) with concomitant chemotherapy (cisplatin 40 mg/m2 body surface for up to five weekly cycles) followed by MR-IGABT. Brachytherapy was delivered at the end of treatment typically over four fractions administered within two applications aiming at a total dose of > 90 Gy EQD2 (equieffective dose, reference dose 2 Gy per fraction, linear quadratic model, α/β = 10 Gy) to 90% (D90) of CTVHR. Venezia-type applicators (Elekta AB, Stockholm, Sweden) with commercially available interstitial obturator-bound plastic needles were used for target coverage and dose delivery. The obturators consisted of stainless steel and tungsten. Interstitial needles were inserted through predefined template positions within the applicator ring in a straight and/or oblique orientation as well as “free-hand” without attachment to the applicator. Straight insertion is defined as parallel to the tandem in the axis of the uterus. Oblique insertion is defined as diagonally from the ring directed to the parametrial tissue with an angle of 20° to the tandem. Free-hand needles were typically implanted either laterally to the ring into the middle/distal parametrium for regions not reachable with predefined needle positions or anteriorly to the ring in the periurethral area.
Transrectal ultrasound
Free-hand ultrasound examination without use of a stepper unit was performed intraoperatively by a single investigator (= the treating physician) using a biplane transrectal probe (bk3000; BK Medical, Burlington, MA, USA) while the patients were in lithotomy position and under anesthesia following a dedicated in-house standard protocol. The high-resolution biplane transrectal probe (E14CL4b) has an image field sector of 138° in transverse view with a frequency range of 6–12 MHz in the herein applied scanning mode (B-mode) and a focal range of 3–60 mm. Our standard settings include a frequency range of 6 MHz and a focal range of 60 mm. The in-house standard protocol includes a first TRUS scan before applicator insertion for assessment of anatomy and expected target volume. Consecutive scans were then performed during or after tandem insertion and for each inserted needle separately (starting with the most anterior needles to avoid artefacts). The visualized target volume and the position of the first needle then determined the choice for the position of the subsequent ones. For predefined standard positions (straight or oblique) within the applicator, needle insertion and scanning were usually done sequentially (i.e., insertion of the needle for approximately one cm without TRUS – TRUS needle position check – if acceptable, further insertion until intended depth offline; if not acceptable, selection of a presumably more appropriate position) and only in critical situations (e.g., large vessels or bowel in direct proximity) or for free-hand needles simultaneously (i.e., insertion of needle below mucosa–TRUS position check and further insertion with online TRUS visualization). A final scan was done as soon as a satisfactory implant was reached to evaluate the overall implant geometry and to take representative screenshots. A workflow describing intraoperative procedures is also given in Fig. 1. CTVHR, applicator, and interstitial needles (always with the obturator in place) were visualized on both transverse and sagittal images. Only transverse images were used for image evaluation. An example of a TRUS image is given in Fig. 2.
MRI
MRI was performed using an Ingenia 1.5T machine (Philips, Amsterdam, the Netherlands) according to the Gyn Groupe Européen de Curiethérapie and the European Society for Radiotherapy and Oncology (GEC-ESTRO) recommendations for imaging in IGABT [8]. T2-weighted axial and para-axial images were obtained covering at least the whole uterine fundus to the caudal border of the pubic bone including any vaginal tumor extension with the applicator in place for image analysis and treatment planning. Slice thickness was 3 mm. An example of MRI is given in Fig. 2.
Image analysis
The primary aim of this analysis was the prospective assessment of quantitative and qualitative needle visibility on TRUS images in comparison to MRI. Therefore, the TRUS transducer was manually moved along the longitudinal axis of the uterus to locate the maximum axial diameter of the CTVHR and to perform the intended measurements at this height. All image datasets were evaluated in transverse view in the axis of the uterus at the level of the largest CTVHR diameter. At this level, transverse images were analyzed in regard of (1) needle visibility (yes vs. no); (2) distance of each needle to the tandem in millimeters; (3) distance of each needle to the outer border of the CTVHR in millimeters; and (4) qualitative visibility, which was scored as follows: 0 = no visibility; 1 = poor discrimination, needle surface blurred; 2 = fair discrimination, needle surface indistinct; 3 = excellent discrimination, needle surface distinct. Each available needle position was systematically labelled for exact identification and subsequent comparison. The distance of each needle to the tandem and to the outer boarder of the CTVHR was then measured on the respective MRI in the same manner as described above for comparison. Furthermore, the expected treatment plan quality based on implant geometry visualized by TRUS imaging was rated with the following scoring system: 1 = excellent (CTVHR and OAR soft constraints expected to be met), 2 = sufficient (CTVHR or OAR soft constraints expected to be violated), 3 = poor (CTVHR and OAR soft constraints expected to be violated), 4 = insufficient (CTVHR and/or OAR hard constraints expected to be violated), and compared to the respective final clinically approved MRI-based treatment plan. The final MRI-based treatment plan quality was rated with the following scoring system: 1 = excellent (CTVHR and OAR soft constraints were met), 2 = sufficient (CTVHR or OAR soft constraints were violated), 3 = poor (CTVHR and OAR soft constraints were violated), 4 = insufficient (CTVHR and/or OAR hard constraints were violated). The underlying soft and hard constraints were based on the EMBRACE II study protocol [9].
TRUS and MRI evaluation and measurements were performed by three radiation oncologists with more than 5 years of experience. To reduce a potential bias, all TRUS images were evaluated intraoperatively or directly after, without knowledge of subsequent MRI findings. The results of the measurements were promptly (during or immediately after the procedure) documented by the examiners using the case report form (CRF; see supplementary material).
Statistical analysis
Descriptive statistics using SPSS statistics (version 24, IBM, Armonk, NY, USA) were done to calculate maximum (max), minimum (min), median, mean ± standard deviation (SD) for distance and visibility score for all needles as well as for straight, oblique, and free-hand needles separately. Paired t‑tests were performed to test for statistically significant differences between MRI and TRUS. Significance is defined as p < 0.05 after two-sided testing.
Results
Between 05.2022 and 12.2022, 23 patients with 41 applications and a total of 230 needles (186 straight through the ring, 34 oblique through the ring, 10 free-hand insertions) were included in this study. The Tumor/Node/Metastasis (TNM) T stages consisted of T1b1 (n = 1), T2b (n = 18), T3b (n = 2), and T4 (n = 2, both with urinary bladder infiltration). The treated CTVHR at the time of brachytherapy had a mean ± SD volume of 33.2 cm3 ± 12.9 cm3 (range 11–63 cm3, median 31 cm3). The median number of implanted needles per patient was n = 6 (range 3–12), a median of 5 needles (range 2–9) per patient were used for treatment. In total, 197/230 needles (86%) were used for treatment.
Overall, 228/230 needles (99.1%) were visible. Mean visibility score ± SD was 2.5 ± 0.7 for all visible needles. The respective results for straight, oblique, and free-hand needles did not differ and are shown in Table 1. The two non-visible needles (both free-hand inserted) were masked by artefacts or outside the field of view.
Max, min, mean ± SD distance of the visible needles to tandem was 35 mm, 4 mm, 17.8 mm ± 4.0 mm on MRI and 34 mm, 6 mm, 17.6 mm ± 3.7 mm on TRUS, respectively. Max and mean ± SD difference between MRI and TRUS were 8 mm and −0.1 mm ± 1.6 mm, respectively, which was not significant (p = 0.39). There was also no significant difference regarding straight (p = 0.41), oblique (p = 0.15), and free-hand needles (p = 0.09) separately. The respective results are summarized in Table 1. A difference of more than 3 mm between TRUS and MRI was found in 8 of 228 needles (3.5%).
The distance between tandem and the outer border of the CTVHR at its widest point showed no significant difference between TRUS and MRI (mean ± SD of 1.9 mm ± 2.2 mm, p = 0.26). Straight needles had a mean distance of 5.3 mm ± 5.5 mm to the outer border of the CTVHR on TRUS, whereas oblique needles showed a mean distance of 3.0 mm ± 2.7 mm.
The expected plan quality based on TRUS imaging and actual plan quality after MRI-based planning are compared in Table 2 and showed concordance in 28/41 implants (68%). Seven implants were underestimated by TRUS and six overestimated.
Discussion
This study prospectively investigated quantitative and qualitative needle visibility on TRUS in comparison to MRI in the treatment of locally advanced cervical cancer with combined intracavitary/interstitial MR-IGABT utilizing Venezia-type applicators as part of curative radiochemotherapy. Almost all needles (99%) were identifiable on TRUS, no matter which route of insertion (straight or oblique) was used. This contrasts with a previously published retrospective analysis [10] in which 87% of straight needles and only 51% of oblique needles could be detected on TRUS. The better detectability in the prospective study might be explained by the use of a different ultrasound device with slightly better image quality but, above all, by a different approach used for image acquisition. The TRUS probe was used free-handedly within the prospective study, thereby enabling the investigators to actively search and follow the needles in order to accurately adapt to the individual patient anatomy, while in the previous study, the TRUS probe was fixated by a stepper unit with a limited degree of freedom, causing more artefacts by, e.g., air bubbles. Furthermore, the distal parts of the parametrial space were partly outside the field of view with use of the stepper unit. Not only quantitative but also qualitative needle visibility improved from the retrospective (mean visibility score 1.4 ± 0.5 SD) to prospective analysis (mean visibility score 2.5 ± 0.7 SD). This might again be explained by the free-handed use of the TRUS probe, which allowed a closer proximity between probe and needle. Another important aspect could be found in the different needle material (titanium in the retrospective study vs. plastic [+ obturator] in the prospective study), which might impact on the extent of acoustic shadowing behind the needle, particularly in cases that require a higher number of needles for target coverage. The difference in distance between needles and tandem comparing TRUS and MRI is slightly better when comparing the current prospective assessment (mean ± SD: −0.1 ± 1.6 mm, 97% within a difference of 3 mm) to the retrospective study (mean ± SD: −0.3 ± 2.6 mm, 89% within a difference of 3 mm). Similarly, Rodgers et al. reported on visibility of all needles and a point difference of less than 3 mm in 78% of needles when using transvaginal ultrasound [11]. Comparing those differences to proposed uncertainties of less than 2 mm in quality assurance phantom tests [12], the clinical in vivo situation does not exceed this value by more than 1 mm in most cases.
Our findings show that the spatial relation of visible needles to the applicator is depicted equivalently on TRUS and MRI, which strongly suggests using TRUS for real-time needle guidance. Benefits of this approach are prompt guidance, evaluation, and confirmation of needle positions directly in the operating room (OR) leading to improved implant quality before the patient is transferred to the MRI facility. As a consequence, additional workload, time, and costs might be saved by avoiding repetitive transfers between OR and MRI.
The relevance of TRUS for IGABT is also reflected in the congruence of estimated implant quality based on TRUS imaging compared to actual implant quality after MRI-based treatment planning with regard to OAR and target dose constraints. In detail, three investigators with more than 5 years of experience in cervical cancer brachytherapy rated and, if necessary, adapted the implant in real time using TRUS imaging before MRI was performed. This led to excellent or sufficient implant quality in most cases (39/41; 95%), with only two patients having poor and insufficient treatment plans. The patient with expected and actual poor plan quality had a violated soft constraint of the CTVHR (7.4 Gy at 90% of CTVHR) at the first implant due to a myoma blocking a needle from further insertion, which would have led to a projected summed dose of 87 Gy EQD210Gy at 90% of the CTVHR in total. The second implant was rotated to a different angle, enabling avoidance of the myoma and thereby leading to a higher D90 of the CTVHR (summed dose of 90.1 Gy EQD210Gy). However, the soft constraint of the bowel remained violated (summed D2cm3 of 74.6 Gy EQD23Gy) due to proximity of a bowel loop to the CTVHR. In the patient with expected and actual insufficient plan quality, the hard constraint of the urinary bladder (summed D2cm3 of 92 Gy EQD23Gy) was violated due to urinary bladder infiltration. This was accepted in favor of sufficient target coverage. Both patients are currently without evidence of disease and without symptoms >G2 at 10 and 8 months post brachytherapy, respectively.
In prostate cancer brachytherapy, TRUS-based needle reconstruction and treatment planning is already possible [12,13,14,15]. In contrast, TRUS is suitable for target visualization and tandem and needle depiction in gynecological brachytherapy, but TRUS-only planning is not possible at the moment due to incomplete depiction of the surrounding OARs and uncertainties in applicator reconstruction. A possibility to overcome these limitations would be computed tomography (CT) (for OAR delineation) and TRUS (for target delineation) image fusion with automized applicator reconstruction by applicator tracking. Although this workflow was successfully tested with use of optical tracking, there were various disadvantages, with optical tracking limiting broad clinical implementation [7, 16]. Electromagnetic tracking (EMT) has been proven valuable to display, confirm, and follow needle positions in breast and prostate brachytherapy [17,18,19,20,21,22], and might also offer an attractive solution for cervical cancer brachytherapy. A prospective study to investigate EMT tracking for cervical cancer brachytherapy is currently under preparation in our department.
Strengths of this prospective investigation can be found in the number of patients and the high number of used needles (n = 230) in particular. To the best of our knowledge, this is the first study to investigate the visibility of different routes of needle insertion prospectively on TRUS. A new aspect is the estimation of implant quality (in the sense of “usability”), which might help to ease the whole treatment process. Lastly, the close local and temporal proximity between application of TRUS and MRI minimized the risk of needle or applicator movement between the two imaging modalities. Limitations might be found in the use of only a single applicator type and in the limited number and thereby limited data on free-hand needles.
Conclusion
In this cohort, use of TRUS for real-time interstitial needle guidance resulted in high-quality implants due to excellent visibility of needles during insertion, thereby rendering TRUS-guided brachytherapy a promising modality to pursue and develop for gynecological indications.
Abbreviations
- CRF:
-
Case report form
- CTVHR :
-
High-risk clinical target volume
- EQD2:
-
Equieffective dose, reference dose 2 Gy per fraction
- GEC-ESTRO:
-
Groupe Européen de Curiethérapie and the European Society for Radiotherapy and Oncology
- Max:
-
Maximum
- Min:
-
Minimum
- MRI:
-
Magnetic resonance imaging
- MR-IGABT:
-
Magnetic resonance image-guided adaptive brachytherapy
- OAR(s):
-
Organ(s) at risk
- OR:
-
Operating room
- SD:
-
Standard deviation
- TRUS:
-
Transrectal ultrasound
References
Cibula D et al (2023) ESGO/ESTRO/ESP Guidelines for the management of patients with cervical cancer—Update 2023. Int J Gynecol Cancer 1;33(5):649–666 (May)
Fokdal L et al (2013) Clinical feasibility of combined intracavitary/interstitial brachytherapy in locally advanced cervical cancer employing MRI with a tandem/ring applicator in situ and virtual preplanning of the interstitial component. Radiother Oncol 107(1):63–68 (Apr)
Van Dyk S et al (2009) Conformal brachytherapy planning for cervical cancer using transabdominal ultrasound. Int J Radiat Oncol Biol Phys 1;75(1):64–70 (Sep)
Epstein E et al (2013) Early-stage cervical cancer: Tumor delineation by magnetic resonance imaging and ultrasound—A European multicenter trial. Gynecol Oncol 128(3):449–453 (Mar)
Schmid MP et al (2013) Feasibility of transrectal ultrasonography for assessment of cervical cancer. Strahlenther Onkol 189(2):123–128 (Feb)
Schmid MP et al (2016) Transrectal ultrasound for image-guided adaptive brachytherapy in cervix cancer—An alternative to MRI for target definition? Radiother Oncol 120(3):467–472 (Sep)
Smet S et al (2020) Hybrid TRUS/CT with optical tracking for target delineation in image-guided adaptive brachytherapy for cervical cancer. Strahlenther Onkol 196(11):983–992 (Nov)
Dimopoulos JC et al (2012) Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (IV): Basic principles and parameters for MR imaging within the frame of image based adaptive cervix cancer brachytherapy. Radiother Oncol 103(1):113–122 (Apr)
Pötter R., et al.; The EMBRACE II study: The outcome and prospect of two decades of evolution within the GEC-ESTRO GYN working group and the EMBRACE studies. Clin Transl Radiat Oncol. 2018 Jan 11;9:48–60.
Knoth J., et al.; Toward 3D-TRUS image-guided interstitial brachytherapy for cervical cancer. Brachytherapy. 2022 Mar-Apr;21(2):186–192.
Rodgers JR et al (2019) Intraoperative 360-deg three-dimensional transvaginal ultrasound during needle insertions for high-dose-rate transperineal interstitial gynecologic brachytherapy of vaginal tumors. J Med Imaging (bellingham) 6(2):25001 (Apr)
Siebert FA et al (2020) GEC-ESTRO/ACROP recommendations for quality assurance of ultrasound imaging in brachytherapy. Radiother Oncol 148:51–56 (Jul)
Batchelar D., et al.; Validation study of ultrasound-based high-dose-rate prostate brachytherapy planning compared with CT-based planning. Brachytherapy. 2014 Jan-Feb;13(1):75–9.
Zheng D., et al.; A novel method for accurate needle-tip identification in trans-rectal ultrasound-based high-dose-rate prostate brachytherapy. Brachytherapy. 2011 Nov-Dec;10(6):466–73.
Siebert FA et al (2009) Imaging of implant needles for real-time HDR-brachytherapy prostate treatment using biplane ultrasound transducers. Med Phys 36(8):3406–3412 (Aug)
Nesvacil N., et al.; Combining transrectal ultrasound and CT for image-guided adaptive brachytherapy of cervical cancer: Proof of concept. Brachytherapy. 2016 Nov–Dec;15(6):839–844.
Poulin E et al (2015) Fast, automatic, and accurate catheter reconstruction in HDR brachytherapy using an electromagnetic 3D tracking system. Med Phys 42(3):1227–1232 (Mar)
Yang Z et al (2020) Verification of needle guidance accuracy in pelvic phantom using registered ultrasound and MRI images for intracavitary/interstitial gynecologic brachytherapy. J Contemp Brachytherapy 12(2):147–159 (Apr)
Dehghan E et al (2018) EM-enhanced US-based seed detection for prostate brachytherapy. Med Phys 45(6):2357–2368 (Jun)
Bharat S et al (2014) Electromagnetic tracking for catheter reconstruction in ultrasound-guided high-dose-rate brachytherapy of the prostate. Brachytherapy Nov-dec 13(6):640–650
Boutaleb S et al (2015) Performance and suitability assessment of a real-time 3D electromagnetic needle tracking system for interstitial brachytherapy. J Contemp Brachytherapy 7(4):280–289 (Aug)
Beaulieu L et al (2018) Real-time electromagnetic tracking based treatment platform for high-dose-rate prostate brachytherapy: Clinical workflows and end-to-end validation. Brachytherapy. Jan 17(1):103–110
Funding
The work of ACMIT (represented by G. Kronreif) was cofinanced by funding in the framework of the Austrian FFG COMET program. The patent “Methods and systems for brachytherapy planning based on imaging data,” 2017, United States Patent and Trademark Office (patent number US20170120072) was issued. The funding sources were not involved in planning, conduction, interpretation, or writing of this research work.
Funding
Open access funding provided by Medical University of Vienna.
Author information
Authors and Affiliations
Contributions
All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Conflict of interest
J. Knoth, A. Sturdza, N. Nesvacil, C. Kirisits, and M.P. Schmid received honoraria for educational events from Elekta AB, Sweden. N. Nesvacil was supported by the Austrian Science Fund (FWF), project KLI695-B33. G. Kronreif represents the work of ACMIT. A. Zaharie, V. Dick, and J. Widder declare that they have no competing interests.
Ethical standards
All procedures performed in studies involving human participants or on human tissue were in accordance with the ethical standards of the institutional and/or national research committee and with the 1975 Helsinki declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Knoth, J., Sturdza, A., Zaharie, A. et al. Transrectal ultrasound for intraoperative interstitial needle guidance in cervical cancer brachytherapy. Strahlenther Onkol (2024). https://doi.org/10.1007/s00066-024-02207-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s00066-024-02207-9