Introduction

Currently, standard of care in the treatment of locally advanced cervical cancer is radiochemotherapy including brachytherapy [1]. During the past two decades, the performance of brachytherapy has been substantially improved by the integration of MRI into treatment planning and by the development of combined intracavitary/interstitial applicators. Use of MRI enables precise visualization of an adaptive target volume, whereas combined intracavitary/interstitial applicator systems allow the dose to be shaped accordingly. The main advantage of these applicators is that in addition to the standard intracavitary applicator (e.g., tandem/ring or tandem/ovoid), interstitial needles can be inserted through predefined holes in the ring or ovoid into more lateral parts of the cervix and/or parametria. However, the optimal workflow and in particular the most adequate utilization of the numerous needle positions for an individual patient remain unclear. The standard procedure is to choose the positions based on clinical estimation guided by gynecological examination and any prebrachytherapy MRI scans, if available. As the needles are directly linked to the intracavitary applicator, correct positioning of the tandem is of utmost importance. Nevertheless, even with optimal preparation such as full MRI-based preplanning [2] with use of 3D-printed applicators, uncertainty remains intraoperatively regarding whether the needles have actually reached the intended locations. Therefore, there has recently been increasing interest in the intraoperative use of transabdominal [3] and transrectal ultrasound (TRUS) in gynecological brachytherapy

TRUS has been proven to be noninferior to MRI in terms of tumor and brachytherapy target volume assessment for cervical cancer, albeit until now with major limitations in applicator depiction [4,5,6,7]. While real-time needle visualization and guidance is standard of care in prostate cancer, it is still in its infancy for gynecological brachytherapy. The aim of the current study is to analyze visibility and spatial differences of interstitial needles in TRUS in comparison to MRI in patients treated with magnetic resonance image-guided adaptive brachytherapy (MR-IGABT) for cervical cancer with combined intracavitary/interstitial applicators and to predict the expected treatment plan quality by TRUS.

Materials and methods

Patients and treatment

This is a prospective single-arm cohort study which was approved by the institutional ethics committee. Patients were eligible if they had (1) biopsy-proven cervical cancer with a local FIGO stage IB-IVA, (2) treatment with MR-IGABT, (3) utilization of a combined intracavitary/interstitial approach, and (4) written informed consent. The overall treatment comprised pelvic external beam radiotherapy (45 Gy in 25 fractions of 1.8 Gy) with concomitant chemotherapy (cisplatin 40 mg/m2 body surface for up to five weekly cycles) followed by MR-IGABT. Brachytherapy was delivered at the end of treatment typically over four fractions administered within two applications aiming at a total dose of > 90 Gy EQD2 (equieffective dose, reference dose 2 Gy per fraction, linear quadratic model, α/β = 10 Gy) to 90% (D90) of CTVHR. Venezia-type applicators (Elekta AB, Stockholm, Sweden) with commercially available interstitial obturator-bound plastic needles were used for target coverage and dose delivery. The obturators consisted of stainless steel and tungsten. Interstitial needles were inserted through predefined template positions within the applicator ring in a straight and/or oblique orientation as well as “free-hand” without attachment to the applicator. Straight insertion is defined as parallel to the tandem in the axis of the uterus. Oblique insertion is defined as diagonally from the ring directed to the parametrial tissue with an angle of 20° to the tandem. Free-hand needles were typically implanted either laterally to the ring into the middle/distal parametrium for regions not reachable with predefined needle positions or anteriorly to the ring in the periurethral area.

Transrectal ultrasound

Free-hand ultrasound examination without use of a stepper unit was performed intraoperatively by a single investigator (= the treating physician) using a biplane transrectal probe (bk3000; BK Medical, Burlington, MA, USA) while the patients were in lithotomy position and under anesthesia following a dedicated in-house standard protocol. The high-resolution biplane transrectal probe (E14CL4b) has an image field sector of 138° in transverse view with a frequency range of 6–12 MHz in the herein applied scanning mode (B-mode) and a focal range of 3–60 mm. Our standard settings include a frequency range of 6 MHz and a focal range of 60 mm. The in-house standard protocol includes a first TRUS scan before applicator insertion for assessment of anatomy and expected target volume. Consecutive scans were then performed during or after tandem insertion and for each inserted needle separately (starting with the most anterior needles to avoid artefacts). The visualized target volume and the position of the first needle then determined the choice for the position of the subsequent ones. For predefined standard positions (straight or oblique) within the applicator, needle insertion and scanning were usually done sequentially (i.e., insertion of the needle for approximately one cm without TRUS – TRUS needle position check – if acceptable, further insertion until intended depth offline; if not acceptable, selection of a presumably more appropriate position) and only in critical situations (e.g., large vessels or bowel in direct proximity) or for free-hand needles simultaneously (i.e., insertion of needle below mucosa–TRUS position check and further insertion with online TRUS visualization). A final scan was done as soon as a satisfactory implant was reached to evaluate the overall implant geometry and to take representative screenshots. A workflow describing intraoperative procedures is also given in Fig. 1. CTVHR, applicator, and interstitial needles (always with the obturator in place) were visualized on both transverse and sagittal images. Only transverse images were used for image evaluation. An example of a TRUS image is given in Fig. 2.

Fig. 1
figure 1

Process workflow describing intraoperative procedures. Steps involving transrectal ultrasound (TRUS) are colored in dark blue, steps without TRUS are colored in light blue

Fig. 2
figure 2

Example of a cervical cancer patient with the applicator as well as four straight and two oblique needles in place. The T2-weighted axial MRI in a and d, T1-weighted axial MRI in the b and e, and the corresponding axial transrectal ultrasound image in c and f. Tandem (red), straight needles (green), and oblique needles (blue) are marked in df

MRI

MRI was performed using an Ingenia 1.5T machine (Philips, Amsterdam, the Netherlands) according to the Gyn Groupe Européen de Curiethérapie and the European Society for Radiotherapy and Oncology (GEC-ESTRO) recommendations for imaging in IGABT [8]. T2-weighted axial and para-axial images were obtained covering at least the whole uterine fundus to the caudal border of the pubic bone including any vaginal tumor extension with the applicator in place for image analysis and treatment planning. Slice thickness was 3 mm. An example of MRI is given in Fig. 2.

Image analysis

The primary aim of this analysis was the prospective assessment of quantitative and qualitative needle visibility on TRUS images in comparison to MRI. Therefore, the TRUS transducer was manually moved along the longitudinal axis of the uterus to locate the maximum axial diameter of the CTVHR and to perform the intended measurements at this height. All image datasets were evaluated in transverse view in the axis of the uterus at the level of the largest CTVHR diameter. At this level, transverse images were analyzed in regard of (1) needle visibility (yes vs. no); (2) distance of each needle to the tandem in millimeters; (3) distance of each needle to the outer border of the CTVHR in millimeters; and (4) qualitative visibility, which was scored as follows: 0 = no visibility; 1 = poor discrimination, needle surface blurred; 2 = fair discrimination, needle surface indistinct; 3 = excellent discrimination, needle surface distinct. Each available needle position was systematically labelled for exact identification and subsequent comparison. The distance of each needle to the tandem and to the outer boarder of the CTVHR was then measured on the respective MRI in the same manner as described above for comparison. Furthermore, the expected treatment plan quality based on implant geometry visualized by TRUS imaging was rated with the following scoring system: 1 = excellent (CTVHR and OAR soft constraints expected to be met), 2 = sufficient (CTVHR or OAR soft constraints expected to be violated), 3 = poor (CTVHR and OAR soft constraints expected to be violated), 4 = insufficient (CTVHR and/or OAR hard constraints expected to be violated), and compared to the respective final clinically approved MRI-based treatment plan. The final MRI-based treatment plan quality was rated with the following scoring system: 1 = excellent (CTVHR and OAR soft constraints were met), 2 = sufficient (CTVHR or OAR soft constraints were violated), 3 = poor (CTVHR and OAR soft constraints were violated), 4 = insufficient (CTVHR and/or OAR hard constraints were violated). The underlying soft and hard constraints were based on the EMBRACE II study protocol [9].

TRUS and MRI evaluation and measurements were performed by three radiation oncologists with more than 5 years of experience. To reduce a potential bias, all TRUS images were evaluated intraoperatively or directly after, without knowledge of subsequent MRI findings. The results of the measurements were promptly (during or immediately after the procedure) documented by the examiners using the case report form (CRF; see supplementary material).

Statistical analysis

Descriptive statistics using SPSS statistics (version 24, IBM, Armonk, NY, USA) were done to calculate maximum (max), minimum (min), median, mean ± standard deviation (SD) for distance and visibility score for all needles as well as for straight, oblique, and free-hand needles separately. Paired t‑tests were performed to test for statistically significant differences between MRI and TRUS. Significance is defined as p < 0.05 after two-sided testing.

Results

Between 05.2022 and 12.2022, 23 patients with 41 applications and a total of 230 needles (186 straight through the ring, 34 oblique through the ring, 10 free-hand insertions) were included in this study. The Tumor/Node/Metastasis (TNM) T stages consisted of T1b1 (n = 1), T2b (n = 18), T3b (n = 2), and T4 (n = 2, both with urinary bladder infiltration). The treated CTVHR at the time of brachytherapy had a mean ± SD volume of 33.2 cm3 ± 12.9 cm3 (range 11–63 cm3, median 31 cm3). The median number of implanted needles per patient was n = 6 (range 3–12), a median of 5 needles (range 2–9) per patient were used for treatment. In total, 197/230 needles (86%) were used for treatment.

Overall, 228/230 needles (99.1%) were visible. Mean visibility score ± SD was 2.5 ± 0.7 for all visible needles. The respective results for straight, oblique, and free-hand needles did not differ and are shown in Table 1. The two non-visible needles (both free-hand inserted) were masked by artefacts or outside the field of view.

Table 1 Max, min, median, mean ± SD for visibility score, distance to tandem in millimeters on transrectal ultrasound (TRUS) at maximum target diameter, distance to tandem in millimeters on MRI at maximum target diameter, and difference between TRUS and MRI in millimeters at maximum target diameter; displayed for all visible needles as well as for visible straight, oblique, and “free-hand” needles separately

Max, min, mean ± SD distance of the visible needles to tandem was 35 mm, 4 mm, 17.8 mm ± 4.0 mm on MRI and 34 mm, 6 mm, 17.6 mm ± 3.7 mm on TRUS, respectively. Max and mean ± SD difference between MRI and TRUS were 8 mm and −0.1 mm ± 1.6 mm, respectively, which was not significant (p = 0.39). There was also no significant difference regarding straight (p = 0.41), oblique (p = 0.15), and free-hand needles (p = 0.09) separately. The respective results are summarized in Table 1. A difference of more than 3 mm between TRUS and MRI was found in 8 of 228 needles (3.5%).

The distance between tandem and the outer border of the CTVHR at its widest point showed no significant difference between TRUS and MRI (mean ± SD of 1.9 mm ± 2.2 mm, p = 0.26). Straight needles had a mean distance of 5.3 mm ± 5.5 mm to the outer border of the CTVHR on TRUS, whereas oblique needles showed a mean distance of 3.0 mm ± 2.7 mm.

The expected plan quality based on TRUS imaging and actual plan quality after MRI-based planning are compared in Table 2 and showed concordance in 28/41 implants (68%). Seven implants were underestimated by TRUS and six overestimated.

Table 2 Comparison of expected implant quality based on transrectal ultrasound (TRUS) images and actual implant quality after MRI-based planning

Discussion

This study prospectively investigated quantitative and qualitative needle visibility on TRUS in comparison to MRI in the treatment of locally advanced cervical cancer with combined intracavitary/interstitial MR-IGABT utilizing Venezia-type applicators as part of curative radiochemotherapy. Almost all needles (99%) were identifiable on TRUS, no matter which route of insertion (straight or oblique) was used. This contrasts with a previously published retrospective analysis [10] in which 87% of straight needles and only 51% of oblique needles could be detected on TRUS. The better detectability in the prospective study might be explained by the use of a different ultrasound device with slightly better image quality but, above all, by a different approach used for image acquisition. The TRUS probe was used free-handedly within the prospective study, thereby enabling the investigators to actively search and follow the needles in order to accurately adapt to the individual patient anatomy, while in the previous study, the TRUS probe was fixated by a stepper unit with a limited degree of freedom, causing more artefacts by, e.g., air bubbles. Furthermore, the distal parts of the parametrial space were partly outside the field of view with use of the stepper unit. Not only quantitative but also qualitative needle visibility improved from the retrospective (mean visibility score 1.4 ± 0.5 SD) to prospective analysis (mean visibility score 2.5 ± 0.7 SD). This might again be explained by the free-handed use of the TRUS probe, which allowed a closer proximity between probe and needle. Another important aspect could be found in the different needle material (titanium in the retrospective study vs. plastic [+ obturator] in the prospective study), which might impact on the extent of acoustic shadowing behind the needle, particularly in cases that require a higher number of needles for target coverage. The difference in distance between needles and tandem comparing TRUS and MRI is slightly better when comparing the current prospective assessment (mean ± SD: −0.1 ± 1.6 mm, 97% within a difference of 3 mm) to the retrospective study (mean ± SD: −0.3 ± 2.6 mm, 89% within a difference of 3 mm). Similarly, Rodgers et al. reported on visibility of all needles and a point difference of less than 3 mm in 78% of needles when using transvaginal ultrasound [11]. Comparing those differences to proposed uncertainties of less than 2 mm in quality assurance phantom tests [12], the clinical in vivo situation does not exceed this value by more than 1 mm in most cases.

Our findings show that the spatial relation of visible needles to the applicator is depicted equivalently on TRUS and MRI, which strongly suggests using TRUS for real-time needle guidance. Benefits of this approach are prompt guidance, evaluation, and confirmation of needle positions directly in the operating room (OR) leading to improved implant quality before the patient is transferred to the MRI facility. As a consequence, additional workload, time, and costs might be saved by avoiding repetitive transfers between OR and MRI.

The relevance of TRUS for IGABT is also reflected in the congruence of estimated implant quality based on TRUS imaging compared to actual implant quality after MRI-based treatment planning with regard to OAR and target dose constraints. In detail, three investigators with more than 5 years of experience in cervical cancer brachytherapy rated and, if necessary, adapted the implant in real time using TRUS imaging before MRI was performed. This led to excellent or sufficient implant quality in most cases (39/41; 95%), with only two patients having poor and insufficient treatment plans. The patient with expected and actual poor plan quality had a violated soft constraint of the CTVHR (7.4 Gy at 90% of CTVHR) at the first implant due to a myoma blocking a needle from further insertion, which would have led to a projected summed dose of 87 Gy EQD210Gy at 90% of the CTVHR in total. The second implant was rotated to a different angle, enabling avoidance of the myoma and thereby leading to a higher D90 of the CTVHR (summed dose of 90.1 Gy EQD210Gy). However, the soft constraint of the bowel remained violated (summed D2cm3 of 74.6 Gy EQD23Gy) due to proximity of a bowel loop to the CTVHR. In the patient with expected and actual insufficient plan quality, the hard constraint of the urinary bladder (summed D2cm3 of 92 Gy EQD23Gy) was violated due to urinary bladder infiltration. This was accepted in favor of sufficient target coverage. Both patients are currently without evidence of disease and without symptoms >G2 at 10 and 8 months post brachytherapy, respectively.

In prostate cancer brachytherapy, TRUS-based needle reconstruction and treatment planning is already possible [12,13,14,15]. In contrast, TRUS is suitable for target visualization and tandem and needle depiction in gynecological brachytherapy, but TRUS-only planning is not possible at the moment due to incomplete depiction of the surrounding OARs and uncertainties in applicator reconstruction. A possibility to overcome these limitations would be computed tomography (CT) (for OAR delineation) and TRUS (for target delineation) image fusion with automized applicator reconstruction by applicator tracking. Although this workflow was successfully tested with use of optical tracking, there were various disadvantages, with optical tracking limiting broad clinical implementation [7, 16]. Electromagnetic tracking (EMT) has been proven valuable to display, confirm, and follow needle positions in breast and prostate brachytherapy [17,18,19,20,21,22], and might also offer an attractive solution for cervical cancer brachytherapy. A prospective study to investigate EMT tracking for cervical cancer brachytherapy is currently under preparation in our department.

Strengths of this prospective investigation can be found in the number of patients and the high number of used needles (n = 230) in particular. To the best of our knowledge, this is the first study to investigate the visibility of different routes of needle insertion prospectively on TRUS. A new aspect is the estimation of implant quality (in the sense of “usability”), which might help to ease the whole treatment process. Lastly, the close local and temporal proximity between application of TRUS and MRI minimized the risk of needle or applicator movement between the two imaging modalities. Limitations might be found in the use of only a single applicator type and in the limited number and thereby limited data on free-hand needles.

Conclusion

In this cohort, use of TRUS for real-time interstitial needle guidance resulted in high-quality implants due to excellent visibility of needles during insertion, thereby rendering TRUS-guided brachytherapy a promising modality to pursue and develop for gynecological indications.