Since its introduction, cone beam CT (CBCT) has been mainly utilized for image-guided radiotherapy (IGRT) [14]. CBCT images generated prior to treatment can also be used for adaptive radiotherapy (ART) [5, 6] in various ways, ranging from using the CBCT to estimate the dosimetric effect of patient weight loss, CBCT assessment in order to choose a “plan-of-the-day,” to dose recalculation and treatment plan generation on CBCT images [711]. Commercially available treatment planning systems (TPSs) are now able to implement ART strategies.

CBCT image quality is dependent on acquisition parameters and, compared with diagnostic-quality planning CT images, exhibit increased artifacts and poorer image contrast owing to increased scatter, which itself is dependent on the size of the scanned object/patient [1214]. CBCT images (CBCTs) generated using XVI on Elekta treatment units are not in true CT numbers (CTNs) and, therefore, in their original form, such CBCTs cannot be used directly for accurate dose computation [12]. In order to be able to perform CBCT-based ART, it is important that dose can be calculated accurately (i.e., consistent with that calculated on diagnostic CT images) on the CBCT image; various methods have been suggested for CTN adjustment in order to achieve this [12, 1521].

Strategies for CTN adjustment include calibration of the CBCT image voxel values with physical density following CBCT acquisitions of phantoms with inserts of differing densities [1519]. Richter et al. [17] proposed a method based on average CBCT values for separate treatment sites in order to generate population-specific conversion curves for brain, head and neck (H&N), thorax, and prostate treatment sites. Apart from taking a large amount of time to define, conversion curves are subject to errors caused by image artifacts and patient variability, resulting in differences of 5 % and larger between doses calculated on CBCTs compared with CT-based calculation [12, 17]. Patient-specific conversion curves have also been investigated [20]. However, these methods are still prone to dosimetric errors resulting from CBCT artifacts.

Poludniowski et al. have shown that differences, between the doses on CBCT and CT, of less than 2.5 % can be achieved when the CBCT has been reconstructed after scatter correction of the individual projections [20, 22]. The implementation of such an approach can be slow and difficult to introduce into a clinical workflow. Commercially available systems that generate CBCTs in CTNs by using sophisticated scatter-correction algorithms are starting to become commercially available in recent software releases. However, the accuracy of dose calculation on such images has yet to be validated.

Relatively unsophisticated methods, such as density overrides of regions of interest, can be used to populate CBCTs with density values that result in acceptable dose calculation accuracy [12, 20, 21]. Such strategies can be fast to implement and can easily be adapted into clinical workflows. These techniques are not as sensitive to the problems, such as image artifacts, that the more sophisticated methods struggle to cope with.

Unlike other studies presented for density assignment methods [12, 21] and other approaches [1520, 23], we use radiotherapy planning CTs (PCTCBCT) acquired on the same day as the CBCT as the ground truth for dose calculation. Thus, we have selected patients with minimal anatomical differences between CBCTs and planning CTs, eliminating the need for, and uncertainties associated with, deformable image registration (DIR) [2427]. Using the original planning CT (PCTorig), which may have been captured weeks before the CBCT, as the ground truth for dose calculation is unsuitable in the context of ART because establishing an estimate of the difference in dose between that calculated on the PCTorig and CBCT is itself a subject of importance. We have investigated three anatomical sites: H&N, pelvis (prostate and bladder), and lung.

The aim of this work was to assess the CBCT dose calculation accuracy for density override approaches and to compare this with the physics-based scatter correction of Poludniowski [20, 22]. We evaluated the accuracy of the RayStation TPS (V3.99, RaySearch Laboratories, Stockholm, Sweden) density override approach and compared it with more simple density override methods investigated in the literature [12]. We aimed to establish, for different anatomical locations, the methods that were best in terms of CBCT dose calculation accuracy. The methods for CTN adjustment investigated in this work are transferable to dose calculation on magnetic resonance (MR) images and could, therefore, be of interest to MR-based ART [28].

Patients and methods

We selected patients who had a CBCT and diagnostic CT (both with the same immobilization and/or support devices) on the same day. The study consisted of four H&N cases, four pelvis cases (two prostate and two bladder treatment sites), and three lung cases (two left upper lobe GTVs, one right upper lobe GTV). The RayStation TPS was used for dose calculations, using a collapsed cone algorithm. The median time difference between the CBCT and PCTCBCT was 59 min (range: 17–215 min).

CBCT dose calculation

The CBCTs were acquired with Elekta XVI-V4.5 (XVI Elekta, Stockholm, Sweden); the scanning parameters are displayed in Table 1. To calculate dose on the CBCT images, an external patient contour was defined. First, an external contour of the CBCT image was defined using the TPS automated algorithm that uses a thresholding technique based on the image data. Small holes were then removed from the threshold-generated volume and finally contours not associated with the largest object were deleted. Next, a field-of-view (FOV) region of interest (ROI) was defined for the CBCT image set (see Table 1 for FOV parameters). Finally, with a priori knowledge of the planning CT scan, an automated algorithm (available in the TPS) updated the external contour of the CBCT (where necessary) to extend beyond the CBCT FOV. A missing tissue ROI (with density = 1 g/cm3) is created and is defined as the newly generated external ROI exterior to the FOV ROI. Figure 1 (top left) displays the FOV, external, and missing tissue ROIs for an H&N CBCT image as a red contour, green contour, and gray color wash, respectively.

Fig. 1
figure 1

The RayStation treatment planning system (TPS) cone beam computed tomography (CBCT) CT number (CTN) adjustment method. Top row from left: Sagittal slice of a CBCT image of an H&N cancer patient viewed within the TPS; the CBCT after density assignment by the TPS (regions assigned as bone are shown as yellow, for example); CTN-density table generated for the CBCT image. Bottom row, left: Typical CBCT acquisition of a patient with a tumor in their right lung and (right) the same CBCT image but with the field of view (red contour), external (green contour), and left lung (orange contour) regions of interest displayed

Table 1 CBCT scanning parameters for the three anatomical sites investigated in this studya

CTN adjustment using CBCT reconstruction

Following Poludniowski [20, 22, 29], CBCT reconstructions (CBCTr) were made after scatter correction. The software developed in-house explicitly simulates and corrects for scatter in the projection images using the patients PCTorig as a prior [20, 22]. A patient-independent look-up table was applied to the reconstructed images in order to convert the reconstructed CTNs into physical density [29].

CTN adjustment using density overrides

The “water only” (W) method, where all tissue was assigned as water (1 g/cm3), was investigated in all three treatment sites. For pelvis and lung cases, “water-and-bone” (WB) and “water-and-lung” (WL) techniques were also investigated, respectively. For WB, the CBCT voxels were assigned as either water or bone. Bone ROIs were generated for each CBCT image using a segmentation algorithm based on gray value thresholds on the PCTorig data and copying the resulting ROIs onto the CBCT image. The bone ROIs were edited if necessary, to ensure they were appropriate with reference to the CBCT image. The PCTorig bone ROIs were additionally used to establish an appropriate density to assign to the CBCT bone ROI. The average CTN of bone ROIs on PCTorig was 1,400, which was similar to previously reported data [30] and equivalent to 1.35 g/cm3. This density was used for all cases included in this study. In the WL technique, the patient was assumed to be made up of either water or lung tissue (0.2 g/cm3), where the density of lung was calculated using similar methods as those used to determine the pelvic bone density.

More sophisticated density override techniques were investigated using the CBCT density-assignment tools available in the TPS where six different set densities (air, lung, adipose tissue, connective tissue, cartilage/bone, and higher density for prosthesis) are assigned to the CBCT image. This is achieved by binning the CBCT image histogram into six density levels. The CBCT image thresholds are individual to each patient whereas the assigned density values are fixed (Fig. 1). An automatic algorithm (RSauto) extracts the various density thresholds by approximating the CBCT image histogram with two normal distributions which were then interrogated to determine the threshold values. The physical densities applied by the TPS to the various tissue types were: air = 0.00121 g/cm3, lung = 0.26 g/cm3, adipose = 0.95 g/cm3, connective tissue = 1.05 g/cm3, cartilage/bone = 1.6 g/cm3, and other (such as prosthetic hip) = 3 g/cm3. The consideration of the other override approaches (W, WL, WB) provides a useful benchmark and understanding for the more complex RSauto methodology. Table 2 outlines the CTN adjustment techniques investigated in this study.

Table 2 The different CTN adjustment methods investigated in this study

Limited FOV for lung CBCTs

At our institution, the FOV for CBCT acquisitions for lung patients does not encompass the entire patient contour (Table 1, Fig. 1). When performing CTN adjustment using CBCTr or RSauto, it may, therefore, be necessary to perform an additional density override to the contralateral lung. As well as applying a density override of 1 g/cm3 to the missing tissue ROI (region shaded turquoise in Fig. 1), an additional density override of 0.2 g/cm3 was applied to the left lung outside of the FOV (shaded orange in Fig. 1). If CBCT scans of the lung were to be routinely used for dose calculation, we would recommend using either a medium or large FOV to ensure the entire patient contour at the superior/inferior extent of the tumor was encompassed. This contralateral lung density issue has only a small effect on treatments where all beams enter ipsilaterally. However, for treatments where one (or more) beams entered contralaterally, the lung ROI from the planning CT scan was copied to the CBCT image and a density override was applied (Fig. 1).

Data analysis

Dose statistics and dose-difference maps were generated using the TPS. Dose–volume histogram (DVH) analysis was performed on ROIs that were propagated to the CBCT from the PCTCBCT scan.

Results

Differences in dose-volume statistics between that calculated on the CBCT and the ground truth, for all three treatment sites, are shown in Table 3 and Fig. 2.

Fig. 2
figure 2

Dose difference maps for between doses calculated on CBCT images and the ground truth (PCTCBCT). From top row to bottom: (1) a pelvic case with anterior-posterior distance (DAP) = 23 cm; (2) a pelvic case with DAP = 32 cm; (3) an H&N case; and (4) a lung case. All dose difference maps are presented as a percentage of the prescribed dose

Table 3 Dose difference statistics of target and organs at risk as calculated on the CBCT compared with the ground truth

Internal anatomy variability of a prostate case is illustrated in Fig. 3, demonstrating that the use of PCTCBCT as the ground truth is, in some instances, essential for meaningful DVH dosimetric comparison for CBCT dose calculation accuracy. Previous studies [12, 1521, 23] of CBCT dose calculation accuracy have either used rigid (which may have been acquired weeks prior to the CBCT and therefore exhibit substantial anatomical differences) or deformable image registrations (and their associated uncertainty) between the original planning CT and CBCT.

Fig. 3
figure 3

Left: Sagittal images of a prostate RT treatment plan with dose calculated on (top) the PCTorig; (middle) PCTCBCT; and (bottom) the CBCT. The rectum is shown as an orange contour and the dose is shown in color wash relative to the prescribed dose. Right: dose–volume histogram of the rectum for the dose calculated on the PCTorig scan (orange line); the PCTCBCT (blue line); and the CBCT (red dashed line). The RSauto method was used for CTN adjustment of the CBCT image. The rectum OAR was similar in size and shape on the CBCT and the PCTCBCT but was very different to that seen on the PCTorig scan. PCT planning computed tomography, CBCT cone beam CT, CTN CT number, OAR organ at risk

Pelvis cases

RSauto tended to underestimate the proportion of lower-density tissue present in larger patients (Fig. 4), corresponding to an underestimation of the dose. This underestimation has been observed in a larger cohort of pelvic cases that were not used in this dose validation study as they did not have a CBCT and CT acquired on the same day. However, for patients with anterior-posterior distance (DAP) < 25 cm, RSauto resulted in the most accurate dose distributions of all CTN adjustment methods with an average of 0.7 % absolute difference in dose when compared with the ground truth. Furthermore, for patients with DAP < 25 cm, the WB method worked well with the assignment of water (1 g/cm3) to all non-bone tissue, being adequate in such cases as these patients commonly have similar amounts of lower-density (0.95 g/cm3) and connective (1.05 g/cm3) tissue. However, for patients with higher proportions of adipose tissue, the WB method did not perform as well. The W method performed better than WB for patients with DAP > 25 cm, with the additional adipose tissue being balanced out by the higher-density bone. The CBCTr method was unable to accurately reconstruct higher-density material for the larger patient, resulting in a similar effect to that observed with the W method for patients with DAP < 25 cm. For this treatment site, the RSauto method produced the best results when compared with the ground truth, with an average absolute difference of 0.7 % (range: 0.1–2.5 %) when only considering patients with DAP < 25 cm. For CBCT images acquired around the pelvis of patients with DAP < 25 cm, we therefore recommend using the RSauto method for CTN adjustment. However, the WB and W methods worked with sufficient accuracy [20] for patients with DAP < 25 cm and > 25 cm, respectively.

Fig. 4
figure 4

Sagittal (top row) and coronal slices (bottom row) of a pelvic case with a high proportion of adipose tissue. From left to right: the PCTCBCT with tissue density < 0.95 g/cm3 colored purple, and CBCT with RSauto CTN adjustment. In the CBCT images purple, turquoise, and yellow represent adipose (0.95 g/cm3), connective tissue (1.05 g/cm3), and bone (1.6 g/cm3), respectively. PCT planning computed tomography, CBCT cone beam CT, CTN CT number

H&N cases

All three methods investigated for CTN adjustment for H&N cases resulted in acceptable [20] average absolute dose deviations of ≤ 1.0 % when compared with the ground truth, with CBCTr (0.5 %) and RSauto (0.6 %) performing best. For CBCT images acquired of the H&N, on the basis of the results presented in Table 3, its integration within a TPS, and its ease and speed of implementation, we recommend using the RSauto method for CTN adjustment. Larger dose differences were observed in the inferior part of the target at the level of the shoulders. Similar to the methods employed for the contralateral lung ROIs (Fig. 1), this dose disagreement may be improved by generating additional bone ROIs for the shoulders to enable more accurate dose calculation on the CBCT data set.

Lung cases

Assuming that density overrides external to the FOV ROI were applied as per Fig. 1, the WL and RSauto methods generated dose distributions most similar to PCTCBCT calculations, with average absolute dose differences of 1.3 and 1.4 %, respectively (Table 3). When the RSauto or CBCTr methods were applied to VMAT treatments without applying additional contralateral lung density overrides, the average absolute dose difference increased to over 3.0 %. The W method performed poorly for lung cases as, unlike the other methods investigated for this site, it did not account for the lower-density lung tissue present. On the basis of the results presented in Table 3, for CBCT images acquired of the lung, we recommend using either the RSauto or WL method for CTN adjustment. This study did not investigate the dependence of CBCT-based dose calculations on lung intra-fraction motion.

Discussion

For the sites investigated in this study, the differences between the doses calculated on the CBCT compared with the diagnostic CT, when using RSauto, were similar or better than other reported results. Fotina et al. demonstrated that density override techniques applied to CBCT scans of the prostate resulted in dosimetric differences below 2 %, whereas the use of conversion curves to calculate dose on CBCT images has been shown to result in differences of up to 5 % [12]. However, Richter et al. demonstrated that conversion curves specific to anatomical sites can result in comparable dose calculation accuracy with differences below 2 % [17].

Owing to weight loss during radiotherapy for H&N cancer [7, 24, 31, 32], for the H&N cases it was necessary to use the PCTCBCT as the input to the algorithm within the TPS used to define the CBCT body contour. When the PCTorig was used, the large differences in the body contour between the PCTCBCT and CBCT meant dose differences between that calculated on the CBCT and the ground truth could not be assessed with confidence. The use of the PCTCBCT data as a priori knowledge for this validation study reduced the uncertainty of the shape of the CBCT body contour, resulting in more accurate estimations of CBCT dose. If CBCT scans of the H&N were to be routinely used for dose calculation, we would recommend using either a medium or large FOV to ensure the entire patient contour at the superior/inferior extent of the tumor was imaged in order to generate the entire patient contour on the CBCT image. Cupping artifacts can have a large effect on CBCT dose calculations, and the degree to which they affect the dose depends on the CTN adjustment method used. CBCT-density look-up table techniques are the most sensitive in this context. However, CBCT dose distributions generated using density override methods, such as those discussed in this study, are also affected by image quality with increasing sensitivity with increasing number of density bins. However, the results indicate that binning the CBCT voxels into six densities is an appropriate methodology that results in dose distributions that are sufficiently accurate. As well as automatically defining the density threshold levels, the TPS allows them to be manually defined by the user. Especially important for larger-sized patients, this approach could be used to ensure an appropriate proportion is assigned as lower-density adipose tissue. However, tissue assignment as a manual process is subject to error, which was not investigated in this study but will be explored in future work. Although this study involved a small number of cases in each of the three treatment sites, we believe that we have identified patients with minimal anatomic differences between the CBCT and PCTCBCT, enabling the accuracy of CBCT dose calculation to be assessed and for the RSauto method to be validated.

Conclusion

We investigated various methods, including RSauto that has not been assessed before, for CTN adjustment for CBCT dose calculations. RSauto methods were validated, resulting in CBCT dose calculations for lung patients, for patients treated in the pelvis with DAP < 25 cm, and for H&N cases, which were similar to those calculated on diagnostic CT images. Its implementation into a commercially available TPS makes the RSauto approach the fastest and easiest-to-use of all methods investigated in this study and it can easily be performed as part of a typical clinical workflow. The differences reported in this study between the doses calculated on the CBCT compared with the ground truth, when using density overrides, were similar to other reported results [12] and better than those that use CBCT-density look-up table methods [12]. However, in this study we used a planning CT acquired on the same day as the CBCT as the ground truth in order to minimize the uncertainties often present in studies of this kind. Although the RayStation TPS was used for this work, the methods described could be replicated in other treatment planning systems. Furthermore, the density override methods for CTN adjustment that were investigated in this work are applicable and transferable to dose calculation on segmented MR images.