Introduction

Since the publication of the first studies in early 2015 [1] and the meta-analysis of the “Big Five” by Goyal et al. in 2016 [2], mechanical thrombectomy (MT) has become an established procedure for the treatment of acute ischemic stroke. The DAWN and DEFUSE‑3 trials (2018) also demonstrated that MT is superior to intravenous thrombolysis alone, even in patients up to 24 h after symptom onset [3, 4]. Consequently, the number of MTs has increased dramatically in recent years, and further growth is expected in the future [5].

Thus, there is growing interest in monitoring and reducing the resulting radiation exposure for patients and staff during these interventions [6]. In Germany, diagnostic reference levels (DRL) are required by law to limit the patient exposure during radiological procedures (§125 Radiation Protection Ordinance (StrlSchV) and §185 Radiation Protection Act (StrlSchG)). The DRLs were defined by the Federal Office for Radiation Protection (BfS) following the recommendations of ICRP (International Commission on Radiological Protection) 135. The DRLs refer to the mean value of the corresponding parameter over 10–20 (conventional radiography and computer tomography, CT) or 20–30 (fluoroscopy and interventional radiology) studies of the same type [7]. Thus, the DRL may be exceeded in single cases; however, the DRL should not be exceeded by the average value of a study group. The data volume and diversity of centers providing data are crucial for the significance of DRL. For neuroradiological interventions such as MT, the DeGIR/DGNR registry, which was started in 2012, provides the largest dataset and is a representative sample for MT in Germany [8]. The DeGIR/DGNR data were last evaluated in 2018 by Schegerer et al., in which a general DRL of 18,000 cGy∙cm2 for the dose area product (DAP) was established for thrombus aspiration [9]. By the end of 2022, the BfS updated the DRL and a generally applicable DRL of 14,000 cGy∙cm2 was established for the endovascular treatment of acute stroke [10].

In conventional radiography and CT examinations, the patient dose primarily depends on the patient’s body mass index (BMI) and device settings. The prediction of patient exposure to radiological interventions is considerably more complex, which can lead to large deviations to the DLR in individual cases. Studies have linked these dose deviations to the number of passages [11, 12], occlusion localization [12], and technical approaches [13] during MT; however, owing to the limited number of patients, the results reflect only local centers and a small number of cases. The DeGIR/DGNR data contain a significantly larger number of cases from independent centers and are thus better suited for representing real-world situations when analyzing possible dose dependencies [8].

This study aimed to evaluate the DeGIR/DGNR registry data from 2019–2021 with respect to radiation exposure during MT (expressed by the DAP), including the influence of technical and clinical parameters.

Methods

DeGIR/DGNR Registry Data

The DeGIR/DGNR registry data for neurorecanalization (module E, stroke therapy) from 2019–2021 were retrospectively analyzed to evaluate patient exposure during MT.

In addition to DAP and fluoroscopy time (FT) as dose-specific parameters, the DeGIR/DGNR registry provides various demographic and clinical parameters. Table 1 summarizes the parameters used to evaluate possible dose dependence.

Table 1 Entries of the DeGIR/DGNR registry data used for the evaluation

Data Filtration

Participation in the DeGIR/DGNR registry is voluntary, and data can be entered via a web portal (Samedi GmbH, Berlin, Germany). The use of free text fields, for example, for the number of passages, FT or DAP, combined with the large volume of data inevitably leads to incorrect entries. Therefore, non-plausible entries were filtered using a Python data analysis library (pandas) prior to the analysis. The exclusion criteria were incomplete dose data, identical entries of DAP and FT for two consecutive studies, the amount of reported FT exceeding the amount of reported DAP, and unusually high or low DAP values associated with a unit atypical for the institution. The data filtering process is shown as a flowchart in Fig. 1.

Fig. 1
figure 1

Flowchart for the prefiltering of the DeGIR/DGNR registry data

Statistical Methods and Presentation of Results

Descriptive statistics were used to evaluate the results. The mean value (MV) and empirical standard deviation (SD) were calculated to describe the registry data. According to ICRP 135, the results of the dose distributions were described by specifying the median (Q0.5) and interquartile range (IQR = Q0.25–Q0.75). While the median and the first quartile (Q0.25) serve as the desired optimization values, the third quartile (Q0.75) can be used as a benchmark for establishing diagnostic reference levels [6].

Mood’s median test was used to evaluate significant differences in the medians of more than two sample groups. Dunn’s test was performed to compare individual samples. The general significance level was set at p < 0.05. Statistical analyses were performed using OriginPro, version 2022 (OriginLab Corporation, Northampton, MA, USA).

Results

After applying the filtering criteria (Fig. 1), a total of 59,345 evaluable records remained in Module E of the DeGIR/DGNR registry for 2019–2021. The data were distributed among the following indications: intracranial stenosis (1092), vasospasm therapy (5391), carotid stenting (9817) and stroke treatment (43,045). The details of the study population within the stroke treatment subset are presented in Table 2.

Table 2 Study population after application of the filter criteria

Analysis of the DeGIR/DGNR registry data for 2019–2021 showed a median of 7337.5 cGy∙cm2 and an IQR of 4064–12,263 cGy∙cm2 for DAP in MT. The former DRL of 18,000 cGy∙cm2 [9], which was valid for the study period, was satisfied in 88% of the cases considered. The DAP exceeded 50,000 cGy∙cm2 in 425 (1%) and 100,000 cGy∙cm2 in 93 (0.2%) of the cases. In 18 of the 180 participating centers, the third quartile exceeded the former DRL of 18,000 cGy∙cm2.

Recanalization Success

Figure 2a shows the distribution of successful recanalization, defined as an mTICI score of ≥ 2b [17]. Thus, 89.7% and 10.3% of the analyzed MTs were successful and unsuccessful, respectively.

Fig. 2
figure 2

a Recanalization rate for MT. b Boxplots of DFP distributions as a function of recanalization success. Moodʼs median test: p < 0.001. Specification of Q25, Q50 and Q75 at the box plots

Figure 2b shows the DAP distribution as a function of recanalization success. Although no significant differences in patient exposure were observed for mTICI scores of 0–2a (p = 0.19), patient exposure decreased with increasing mTICI score of more than 2b (p < 0.001). Compared to unsuccessful MT (mTICI < 2b), the DAP was on average approximately 30% lower in patients with successful MT (mTICI ≥ 2b). The key parameters of the distributions are summarized in Table 3, rows 1–2.

Table 3 Key parameters of the evaluated DAP distributions. MT includes both MT alone and MT + LIT, and stent retrieval includes stent retrieval alone and stent retrieval + aspiration

Technical Approach

Figure 3a shows the frequency of recanalization techniques used in MT. A distinction was made between stent retrieval (72.1%) and aspiration only (19.3%). For stent retrieval, cases with stent retrieval alone (5805) and the combination technique of stent retrieval + aspiration (25,221) were considered.

Fig. 3
figure 3

a Frequency distributions of the recanalization procedures used in MT. b Boxplots of DAP distributions as a function of the recanalization technique. Mood’s median test: p < 0.001, Dunn’s test: p < 0.001 (1), p < 0.001 (2). MT mechanical thrombectomy. Specification of Q25, Q50 and Q75 at the box plots

Figure 3b shows the dose distributions for DAP as a function of the recanalization techniques considered. The key parameters of the distributions are summarized in Table 3, rows 3–5. Radiation exposure was significantly higher when using stent retrieval compared to that using aspiration alone (p < 0.001). No significant difference was found between the distribution of stent retrieval alone and aspiration + stent retrieval. When recanalization was achieved solely by local intra-arterial thrombolysis (LIT), the median DAP was only 5700 cGy·cm2 (IQR = 3051–9786 cGy·cm2) and thus significantly lower than MT (p < 0.001).

Additional Stenting (Extracranial/Intracranial)

Figure 4a shows the frequency distribution of additional procedures during MT. Extracranial stenting during MT (tandem occlusions), intracranial stenting, and both extracranial and intracranial stenting were required in 10.2%, 3.7%, and 0.4% of the cases, respectively.

Fig. 4
figure 4

a Frequency distribution of additional stenting in MT. b Boxplots of DAP distributions as a function of additional stenting. Mood’s median test: p < 0.001, Dunn’s test: p < 0.001 (1), p < 0.001 (2), p < 0.06 (3), p = 0.001 (4), p < 0.001 (5), p < 0.001 (6). Specification of Q25, Q50, and Q75 at the box plots

Figure 4b shows the resulting dose distributions for the DAP. The key parameters of the distributions are summarized in Table 3, rows 6–9. The need for intracranial and/or extracranial stenting significantly increases the radiation dose, and the highest values are observed with intracranial stenting and the combination of intracranial and extracranial stenting (p < 0.001).

Occlusion Localization

Figure 5a shows the frequency distribution of the reported occlusion localizations. Most occlusions (87%) involved the anterior circulation (48% in the middle cerebral artery (MCA), 39% in other vessels), 10.5% involved the posterior circulation, and 2.4% spread over multiple vascular territories.

Fig. 5
figure 5

a Frequency distributions of the reported occlusion localizations at MT. b Boxplots of DAP distributions as a function of occlusion location. Mood’s median test: p < 0.001, Dunn’s test: p < 0.001 (1), p = 0.07 (2), p < 0.001 (3), p < 0.001 (4), p < 0.001 (5), p < 0.001 (6). Specification of Q25, Q50, and Q75 at the box plots

Figure 5b shows the dose distributions for DAP as a function of the occlusion location. The key parameters of the distributions are summarized in Table 3, rows 10–13. The lowest and highest DAP values were observed in occlusions of the MCA (anterior circulation) and multiple vascular territories, respectively. The impact of the location of the occlusion on DAP was statistically significant (p < 0.001).

Number of Passages

Figure 6a shows the frequency distribution of the required thrombectomy passages. Most interventions (79.8%) were completed after a maximum of three passages. More than seven passages were required in only 2% of the cases. The maximum number of documented passages is 20.

Fig. 6
figure 6

a Frequency distributions of the required recanalization passages during MT. b Boxplots of DAP distributions as a function of the number of passages required. Mood’s median test: p < 0.001, Dunn’s test: p < 0.001 (1), p < 0.001 (2), p < 0.001 (3). Specification of Q25, Q50, and Q75 at the box plots

Figure 6b shows the dose distributions for the DAP as a function of the required number of passages. The key parameters of the distributions are summarized in Table 3, rows 14–17. Four groups were formed to improve the statistics. Group boundaries were defined such that the median DAP increased by approximately 50% from 1 group to the next. The correlation between increased number of passages and radiation exposure was statistically significant (p < 0.001).

Case Volume per Center

To investigate the influence of the performing center on patient dose, registry data were divided into three case volume groups according to the number of cases per center in the analysis period of 3 years: ≤ 150 (1), 151–450 (2), ≥ 451 (3).

Figure 7a shows the number of centers, and Fig. 7b shows the number of cases entered per case volume group.

Fig. 7
figure 7

a Number of centers per case volume group (period: 3 years). b Number of registered data per case volume group (period: 3 years). c Boxplots of DAP distributions as a function of case volume per center p < 0.001, Dunn’s test: p = 0.3 (1), p < 0.001 (2). Specification of Q25, Q50, and Q75 at the box plots

Figure 7c shows the dose distributions for DAP depending on the case volume group. The key parameters of the distributions are summarized in Table 3, rows 18–20. Although no significant dose difference was detected between case volume groups 1 and 2 (p = 0.3), the average patient exposure in group 3 was significantly higher, by approximately 30% (p < 0.001).

Discussion

To date, this is the largest study on patient dose in interventional stroke treatment. Our data showed dose dependences on the recanalization technique, additional stenting, occlusion localization, number of passages, number of cases per center, and recanalization success.

Table 4 summarizes the current comparative studies on patient exposure during MT. With the exception of the study by Klepanec et al. [14], in which only monoplanar interventions were considered, patient exposure in previous studies were higher than that observed in this study. The exact fractions of monoplanar or biplanar interventions used in this study are unknown; however, it can be assumed that most interventions were performed on biplane angio suites.

Table 4 Comparative studies of patient exposure during MT

In the largest comparative study by Schegerer et al. [9], MT was only one of several X‑ray applications evaluated, all data > 54,000 cGy∙cm2 and < 1800 cGy∙cm2 were excluded. If these criteria were applied to the DeGIR/DGNR data presented here, the median and IQR would increase to 7750 cGy∙cm2 and 4597–12,519 cGy∙cm2, respectively; however, they would remain below the values reported by Schegerer et al.

The decreasing trend observed with respect to patient exposure during MT may be due to the wider availability of new device technologies with automatic exposure control (AEC) and increased experience in performing MT.

Recanalization Success

The recanalization rate of 89.7% (mTICI score of ≥ 2b) observed here is consistent with the results of the current literature, which report technical success rates of 87–90% [19, 20].

Farah et al. [15] and Klepanec et al. [14] reported significantly lower patient exposure in successful versus unsuccessful recanalization. Unsuccessful interventions were more often associated with an increased number of thrombectomy passages, averaging 3.8 (mTICI < 2b) vs. 2.3 (mTICI ≥ 2b). The lowest dose values were achieved only in rapid and successful interventions, with an mTICI of 3.

Technical Approach

If MT is performed by stent retrieval, the patient exposure is approximately 50% higher than that by aspiration alone. As stent retrieval was used in most cases (72.1%), the dose distribution of MT was primarily characterized by stent retrieval. These results are consistent with those obtained by Weyland et al., in which a dose difference of approximately 43% was observed between stent retrieval and aspiration [13].

Because most examinations are performed using stent retrieval, the official DRL should also be based on this group. Therefore, a DRL of 14,000 cGy·cm2 is recommended. This also corresponds to the current DRL in Germany [10]. If the occlusion can be removed by aspiration alone, the data indicate that a lower dose of exposure can be expected.

Additional Stenting (Extracranial/Intracranial)

When MT was performed with extracranial stenting (tandem occlusions), an increased patient exposure of almost 50% occurred on average compared with MT without additional stenting because of the higher complexity of the procedure. These results are consistent with those of Peter et al. [12], who observed dose dependence between tandem occlusions and anterior or posterior circulation occlusions alone. When intracranial stenting or a combination of extracranial and intracranial stenting, was necessary, the data showed a dose increase of 100–150% on average; however, in these cases, only a limited amount of data must be considered (see Fig. 4a).

Occlusion Localization

The mean DAP for occlusions in multiple vascular territories was approximately 20% higher than that for occlusions in anterior or posterior circulation alone. If the occlusions of the anterior circulation were confined to the MCA, the mean DAP was 26% lower. No significant dose difference was observed between the occlusions of the anterior (ACI + ACA/MCA) and posterior circulations (p = 0.07), which is consistent with the results of Peter et al. [12] and Farah et al. [15].

Number of Passages

Overall, the dose clearly increased with the number of passages, which was in agreement with previous studies [11, 12]. Compared to a single thrombectomy passage, the patient exposure doubled after 4 passages and tripled after approximately 8–9 passages.

Case Volume per Center

A study by Weyland et al. demonstrated that the learning curve is the largest for the first 25 examinations, and that dose-dependence on examiner experience is hardly detectable after more than 25 examinations [16]. No dose-dependence on the number of cases was observed between groups 1 and 2 for the grouping chosen here. Furthermore, regarding the recanalization rate (mTCI ≥ 2b), no difference was observed between groups 1 (90.0%), 2 (89.6%), and 3 (89.8%).

However, it is surprising that patient exposure was significantly higher in group 3 at the most experienced centers. This may be attributed to the increased incidence of complicated cases and/or the increased training of residents in these high-volume centers.

Overall, the results demonstrate a consistently high level of quality in MT performance provided by various centers.

Limitations

Entry into the DeGIR/DGNR registry is voluntary, which implies that not all hospitals participate and some entries may be incomplete or incorrect. Therefore, the validity of the data depends on the correct choice of suitable filter criteria, which can increase the bias effect. Documentation of dose data in the DeGIR/DGNR registry is limited to DAP and FT and is thus severely restricted. Previous studies have demonstrated that the disclosure of skin doses or at least the reference air kerma is necessary for the estimation of deterministic radiation damage, especially for interventional procedures such as MT [21].

Conclusion

This study represents the largest dose evaluation for thrombectomy patients based on DeGIR/DGNR registry data. The results demonstrated a decreasing trend in patient exposure during MT in Germany. Based on the available data, the current DRL of 14,000 cGy·cm2 was deemed appropriate.

The recanalization technique, occlusion localization, number of required passages, and recanalization success affect radiation exposure. These factors can be used to identify the cause of excess DRL and optimize the treatment workflow.