Flow Diverting Stents in Cerebral Small Caliber Vessels (< 2 mm) for Aneurysm Treatment

A Three Center Retrospective Study

  • Original Article
  • Published:
Clinical Neuroradiology Aims and scope Submit manuscript

Abstract

Purpose

The off-label use of flow diverting stents (FDS) for treating cerebral aneurysms in small distal vessels is increasing in clinical practice with encouraging results; however, data directly addressing the parent vessel size are still scarce. Our aim was to evaluate the safety and efficacy of FDS placement in anterior and posterior circulation aneurysms with parent arteries ≤ 2 mm in a real-world representative setting.

Methods

We retrospectively reviewed patients treated with FDS at the three participating university hospitals between 2009 and 2021. The inclusion criteria were the placement of at least one FDS in a parent vessel with a maximum diameter of 2 mm or less. The primary clinical safety endpoint was the absence of death, major or minor symptomatic stroke, transient ischemic attack and procedure-related intracranial hemorrhage. Clinical outcome was assessed using the modified Rankin Scale (mRS) score at the follow-up visit at 1 year. The primary efficacy endpoint was defined as complete and near-complete occlusion at the 1‑year follow-up.

Results

We identified 55 patients harboring 56 aneurysms. The primary clinical safety endpoint was obtained in 93% of cases. The hemorrhagic and thromboembolic complication rates were 4% and 9%, respectively. No patient died or had a relevant discrepancy (> 1 point) at the pre-mRS and post-mRS. The primary efficacy endpoint was reached in 80% of cases.

Conclusion

The use of FDS in vessels < 2 mm is technically feasible with good aneurysm occlusion rates and an acceptable safety profile. Nevertheless, it is essential to be aware of the of the main complications associated when operating in small diameter vessels.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Arrese I, Sarabia R, Pintado R, Delgado-Rodriguez M. Flow-diverter devices for intracranial aneurysms: systematic review and meta-analysis. Neurosurgery. 2013;73:193–9; discussion 199–200.

    Article  PubMed  Google Scholar 

  2. Briganti F, Leone G, Marseglia M, Mariniello G, Caranci F, Brunetti A, Maiuri F. Endovascular treatment of cerebral aneurysms using flow-diverter devices: A systematic review. Neuroradiol J. 2015;28:365–75.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Yan Y, Zhu D, Tang H, Huang Q. Safety and Efficacy of Flow Diverter Treatment for Aneurysm in Small Cerebral Vessels: A Systematic Review and Meta-Analysis. World Neurosurg. 2018;115:54–64.

    Article  PubMed  Google Scholar 

  4. O’Kelly CJ, Krings T, Fiorella D, Marotta TR. A novel grading scale for the angiographic assessment of intracranial aneurysms treated using flow diverting stents. Interv Neuroradiol. 2010;16:133–7.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Lv X, Yang H, Liu P, Li Y. Flow-diverter devices in the treatment of intracranial aneurysms: A meta-analysis and systematic review. Neuroradiol J. 2016;29:66–71.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Dinc H, Saatci I, Oguz S, Baltacioglu F, Yildiz A, Donmez H, Belet U, Onal B, Andic C, Koc O, Kocak O, Koroglu M, Cetin M, Cekirge HS. Long-term clinical and angiographic follow-up results of the dual-layer flow diverter device (FRED) for the treatment of intracranial aneurysms in a multicenter study. Neuroradiology. 2021;63:943–52.

    Article  PubMed  Google Scholar 

  7. Quilici L, Boccardi E; FRED Italian Registry Group. FRED Italian Registry: a multicenter experience with the flow re-direction endoluminal device for intracranial aneurysms. J Neurosurg. 2019. https://doi.org/10.3171/2019.1.JNS183005. Epub ahead of print.

    Article  PubMed  Google Scholar 

  8. Becske T, Brinjikji W, Potts MB, Kallmes DF, Shapiro M, Moran CJ, Levy EI, McDougall CG, Szikora I, Lanzino G, Woo HH, Lopes DK, Siddiqui AH, Albuquerque FC, Fiorella DJ, Saatci I, Cekirge SH, Berez AL, Cher DJ, Berentei Z, Marosfoi M, Nelson PK. Long-Term Clinical and Angiographic Outcomes Following Pipeline Embolization Device Treatment of Complex Internal Carotid Artery Aneurysms: Five-Year Results of the Pipeline for Uncoilable or Failed Aneurysms Trial. Neurosurgery. 2017;80:40–8.

    Article  PubMed  Google Scholar 

  9. Cimflova P, Özlük E, Korkmazer B, Ahmadov R, Akpek E, Kizilkilic O, Islak C, Kocer N. Long-term safety and efficacy of distal aneurysm treatment with flow diversion in the M2 segment of the middle cerebral artery and beyond. J Neurointerv Surg. 2021;13:631–6.

    Article  PubMed  Google Scholar 

  10. Dabus G, Grossberg JA, Cawley CM, Dion JE, Puri AS, Wakhloo AK, Gonsales D, Aguilar-Salinas P, Sauvageau E, Linfante I, Hanel RA. Treatment of complex anterior cerebral artery aneurysms with Pipeline flow diversion: mid-term results. J Neurointerv Surg. 2017;9:147–51.

    Article  PubMed  Google Scholar 

  11. De Macedo Rodrigues K, Kühn AL, Tamura T, Dabus G, Kan P, Marosfoi MG, Lozano JD, Perras M, Brooks C, Howk MC, Hou SY, Rex DE, Massari F, Gounis MJ, Wakhloo AK, Puri AS. Pipeline Embolization Device for Pericallosal Artery Aneurysms: A Retrospective Single Center Safety and Efficacy Study. Oper Neurosurg (Hagerstown). 2018;14:351–8.

    Article  PubMed  Google Scholar 

  12. Lin N, Lanzino G, Lopes DK, Arthur AS, Ogilvy CS, Ecker RD, Dumont TM, Turner RD 4th, Gooch MR, Boulos AS, Kan P, Snyder KV, Levy EI, Siddiqui AH. Treatment of Distal Anterior Circulation Aneurysms With the Pipeline Embolization Device: A US Multicenter Experience. Neurosurgery. 2016;79:14–22.

    Article  PubMed  Google Scholar 

  13. Martínez-Galdámez M, Romance A, Vega P, Vega A, Caniego JL, Paul L, Linfante I, Dabus G. Pipeline endovascular device for the treatment of intracranial aneurysms at the level of the circle of Willis and beyond: multicenter experience. J Neurointerv Surg. 2015;7:816–23.

    Article  CAS  PubMed  Google Scholar 

  14. Möhlenbruch MA, Kizilkilic O, Killer-Oberpfalzer M, Baltacioglu F, Islak C, Bendszus M, Cekirge S, Saatci I, Kocer N. Multicenter Experience with FRED Jr Flow Re-Direction Endoluminal Device for Intracranial Aneurysms in Small Arteries. AJNR Am J Neuroradiol. 2017;38:1959–65.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Nossek E, Zumofen DW, Setton A, Potts MB, Raz E, Shapiro M, Riina HA, De Miquel MA, Chalif DJ, Nelson PK. Treatment of distal anterior cerebral artery aneurysms with the Pipeline Embolization Device. J Clin Neurosci. 2017;35:133–8.

    Article  PubMed  Google Scholar 

  16. Jesser J, Alberalar ND, Kizilkilic O, Saatci I, Baltacioglu F, Özlük E, Killer-Oberpfalzer M, Vollherbst DF, Islak C, Cekirge SH, Bendszus M, Möhlenbruch M, Koçer N. Safety and Efficacy of the FRED Jr Flow Re-Direction Endoluminal Device for Intracranial Aneurysms: Retrospective Multicenter Experience With Emphasis on Midterm Results. Front Neurol. 2021;12:722183.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Cagnazzo F, Perrini P, Dargazanli C, Lefevre PH, Gascou G, Morganti R, di Carlo D, Derraz I, Riquelme C, Bonafe A, Costalat V. Treatment of Unruptured Distal Anterior Circulation Aneurysms with Flow-Diverter Stents: A Meta-Analysis. AJNR Am J Neuroradiol. 2019;40:687–93.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Bender MT, Zarrin DA, Campos JK, Lin LM, Huang J, Caplan JM, Tamargo RJ, Colby GP, Coon AL. Tiny Pipes: 67 Cases of Flow Diversion for Aneurysms in Distal Vessels Measuring Less Than 2.0 mm. World Neurosurg. 2019;127:e193–201.

    Article  PubMed  Google Scholar 

  19. Puri AS, Massari F, Asai T, Marosfoi M, Kan P, Hou SY, Howk M, Perras M, Brooks C, Clarencon F, Gounis MJ, Wakhloo AK. Safety, efficacy, and short-term follow-up of the use of Pipeline Embolization Device in small (<2.5 mm) cerebral vessels for aneurysm treatment: single institution experience. Neuroradiology. 2016;58:267–75.

    Article  PubMed  Google Scholar 

  20. Lall RR, Crobeddu E, Lanzino G, Cloft HJ, Kallmes DF. Acute branch occlusion after Pipeline embolization of intracranial aneurysms. J Clin Neurosci. 2014;21:668–72.

    Article  PubMed  Google Scholar 

  21. Hohenstatt S, Arrichiello A, Conte G, Craparo G, Caranci F, Angileri A, Levi D, Carrafiello G, Paolucci A. Branch vessel occlusion in aneurysm treatment with flow diverter stent. Acta Biomed. 2020;91:e2020003.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Ravindran K, Casabella AM, Cebral J, Brinjikji W, Kallmes DF, Kadirvel R. Mechanism of Action and Biology of Flow Diverters in the Treatment of Intracranial Aneurysms. Neurosurgery. 2020;86(Suppl 1):S13–9.

    Article  Google Scholar 

  23. Saleme S, Iosif C, Ponomarjova S, Mendes G, Camilleri Y, Caire F, Boncoeur MP, Mounayer C. Flow-diverting stents for intracranial bifurcation aneurysm treatment. Neurosurgery. 2014;75:623–31; quiz 631.

    Article  PubMed  Google Scholar 

  24. Dabhi N, Sarathy D, Snyder MH, Kellogg RT, Park MS. Flow Diverter Devices for Treatment of Intracranial Aneurysms in Small Parent Vessels-A Systematic Review of Literature. World Neurosurg. 2022;162:183–94.e7.

    Article  PubMed  Google Scholar 

  25. Brinjikji W, Lanzino G, Cloft HJ, Siddiqui AH, Hanel RA, Kallmes DF. Platelet Testing is Associated with Worse Clinical Outcomes for Patients Treated with the Pipeline Embolization Device. AJNR Am J Neuroradiol. 2015;36:2090–5.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Delgado Almandoz JE, Crandall BM, Scholz JM, Fease JL, Anderson RE, Kadkhodayan Y, Tubman DE. Last-recorded P2Y12 reaction units value is strongly associated with thromboembolic and hemorrhagic complications occurring up to 6 months after treatment in patients with cerebral aneurysms treated with the pipeline embolization device. AJNR Am J Neuroradiol. 2014;35:128–35.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Adeeb N, Griessenauer CJ, Foreman PM, Moore JM, Shallwani H, Motiei-Langroudi R, Alturki A, Siddiqui AH, Levy EI, Harrigan MR, Ogilvy CS, Thomas AJ. Use of Platelet Function Testing Before Pipeline Embolization Device Placement: A Multicenter Cohort Study. Stroke. 2017;48:1322–30.

    Article  PubMed  Google Scholar 

  28. Cagnazzo F, Cappucci M, Dargazanli C, Lefevre PH, Gascou G, Riquelme C, Bonafe A, Costalat V. Treatment of Distal Anterior Cerebral Artery Aneurysms with Flow-Diverter Stents: A Single-Center Experience. AJNR Am J Neuroradiol. 2018;39:1100–6.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Funding

There was no financial or material support for this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Agostino Tessitore.

Ethics declarations

Conflict of interest

S. Hohenstatt, S.L. Vinci, D.F. Vollherbst, A. Tessitore, N. Schmitt, A. Pitrone, A.A. Caragliano, M. Velo, M.A. Möhlenbruch and A. Paolucci declare that they have no competing interests.

Supplementary Information

62_2022_1187_MOESM1_ESM.docx

Table 1. Detailed aneurysm and procedure related characteristics. F female, M male, FDS flow diverter stent, ACA anterior cerebral artery, MCA middle cerebral artery, PCA posterior cerebral artery, VA vertebral artery, AcomA anterior communicating artery, SAH subarachnoid hemorrhage, IPH intraparenchymal hemorrhage

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hohenstatt, S., Vinci, S.L., Vollherbst, D.F. et al. Flow Diverting Stents in Cerebral Small Caliber Vessels (< 2 mm) for Aneurysm Treatment. Clin Neuroradiol 33, 99–105 (2023). https://doi.org/10.1007/s00062-022-01187-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00062-022-01187-6

Keywords

Navigation