Abstract
Purpose
The off-label use of flow diverting stents (FDS) for treating cerebral aneurysms in small distal vessels is increasing in clinical practice with encouraging results; however, data directly addressing the parent vessel size are still scarce. Our aim was to evaluate the safety and efficacy of FDS placement in anterior and posterior circulation aneurysms with parent arteries ≤ 2 mm in a real-world representative setting.
Methods
We retrospectively reviewed patients treated with FDS at the three participating university hospitals between 2009 and 2021. The inclusion criteria were the placement of at least one FDS in a parent vessel with a maximum diameter of 2 mm or less. The primary clinical safety endpoint was the absence of death, major or minor symptomatic stroke, transient ischemic attack and procedure-related intracranial hemorrhage. Clinical outcome was assessed using the modified Rankin Scale (mRS) score at the follow-up visit at 1 year. The primary efficacy endpoint was defined as complete and near-complete occlusion at the 1‑year follow-up.
Results
We identified 55 patients harboring 56 aneurysms. The primary clinical safety endpoint was obtained in 93% of cases. The hemorrhagic and thromboembolic complication rates were 4% and 9%, respectively. No patient died or had a relevant discrepancy (> 1 point) at the pre-mRS and post-mRS. The primary efficacy endpoint was reached in 80% of cases.
Conclusion
The use of FDS in vessels < 2 mm is technically feasible with good aneurysm occlusion rates and an acceptable safety profile. Nevertheless, it is essential to be aware of the of the main complications associated when operating in small diameter vessels.
Similar content being viewed by others
References
Arrese I, Sarabia R, Pintado R, Delgado-Rodriguez M. Flow-diverter devices for intracranial aneurysms: systematic review and meta-analysis. Neurosurgery. 2013;73:193–9; discussion 199–200.
Briganti F, Leone G, Marseglia M, Mariniello G, Caranci F, Brunetti A, Maiuri F. Endovascular treatment of cerebral aneurysms using flow-diverter devices: A systematic review. Neuroradiol J. 2015;28:365–75.
Yan Y, Zhu D, Tang H, Huang Q. Safety and Efficacy of Flow Diverter Treatment for Aneurysm in Small Cerebral Vessels: A Systematic Review and Meta-Analysis. World Neurosurg. 2018;115:54–64.
O’Kelly CJ, Krings T, Fiorella D, Marotta TR. A novel grading scale for the angiographic assessment of intracranial aneurysms treated using flow diverting stents. Interv Neuroradiol. 2010;16:133–7.
Lv X, Yang H, Liu P, Li Y. Flow-diverter devices in the treatment of intracranial aneurysms: A meta-analysis and systematic review. Neuroradiol J. 2016;29:66–71.
Dinc H, Saatci I, Oguz S, Baltacioglu F, Yildiz A, Donmez H, Belet U, Onal B, Andic C, Koc O, Kocak O, Koroglu M, Cetin M, Cekirge HS. Long-term clinical and angiographic follow-up results of the dual-layer flow diverter device (FRED) for the treatment of intracranial aneurysms in a multicenter study. Neuroradiology. 2021;63:943–52.
Quilici L, Boccardi E; FRED Italian Registry Group. FRED Italian Registry: a multicenter experience with the flow re-direction endoluminal device for intracranial aneurysms. J Neurosurg. 2019. https://doi.org/10.3171/2019.1.JNS183005. Epub ahead of print.
Becske T, Brinjikji W, Potts MB, Kallmes DF, Shapiro M, Moran CJ, Levy EI, McDougall CG, Szikora I, Lanzino G, Woo HH, Lopes DK, Siddiqui AH, Albuquerque FC, Fiorella DJ, Saatci I, Cekirge SH, Berez AL, Cher DJ, Berentei Z, Marosfoi M, Nelson PK. Long-Term Clinical and Angiographic Outcomes Following Pipeline Embolization Device Treatment of Complex Internal Carotid Artery Aneurysms: Five-Year Results of the Pipeline for Uncoilable or Failed Aneurysms Trial. Neurosurgery. 2017;80:40–8.
Cimflova P, Özlük E, Korkmazer B, Ahmadov R, Akpek E, Kizilkilic O, Islak C, Kocer N. Long-term safety and efficacy of distal aneurysm treatment with flow diversion in the M2 segment of the middle cerebral artery and beyond. J Neurointerv Surg. 2021;13:631–6.
Dabus G, Grossberg JA, Cawley CM, Dion JE, Puri AS, Wakhloo AK, Gonsales D, Aguilar-Salinas P, Sauvageau E, Linfante I, Hanel RA. Treatment of complex anterior cerebral artery aneurysms with Pipeline flow diversion: mid-term results. J Neurointerv Surg. 2017;9:147–51.
De Macedo Rodrigues K, Kühn AL, Tamura T, Dabus G, Kan P, Marosfoi MG, Lozano JD, Perras M, Brooks C, Howk MC, Hou SY, Rex DE, Massari F, Gounis MJ, Wakhloo AK, Puri AS. Pipeline Embolization Device for Pericallosal Artery Aneurysms: A Retrospective Single Center Safety and Efficacy Study. Oper Neurosurg (Hagerstown). 2018;14:351–8.
Lin N, Lanzino G, Lopes DK, Arthur AS, Ogilvy CS, Ecker RD, Dumont TM, Turner RD 4th, Gooch MR, Boulos AS, Kan P, Snyder KV, Levy EI, Siddiqui AH. Treatment of Distal Anterior Circulation Aneurysms With the Pipeline Embolization Device: A US Multicenter Experience. Neurosurgery. 2016;79:14–22.
Martínez-Galdámez M, Romance A, Vega P, Vega A, Caniego JL, Paul L, Linfante I, Dabus G. Pipeline endovascular device for the treatment of intracranial aneurysms at the level of the circle of Willis and beyond: multicenter experience. J Neurointerv Surg. 2015;7:816–23.
Möhlenbruch MA, Kizilkilic O, Killer-Oberpfalzer M, Baltacioglu F, Islak C, Bendszus M, Cekirge S, Saatci I, Kocer N. Multicenter Experience with FRED Jr Flow Re-Direction Endoluminal Device for Intracranial Aneurysms in Small Arteries. AJNR Am J Neuroradiol. 2017;38:1959–65.
Nossek E, Zumofen DW, Setton A, Potts MB, Raz E, Shapiro M, Riina HA, De Miquel MA, Chalif DJ, Nelson PK. Treatment of distal anterior cerebral artery aneurysms with the Pipeline Embolization Device. J Clin Neurosci. 2017;35:133–8.
Jesser J, Alberalar ND, Kizilkilic O, Saatci I, Baltacioglu F, Özlük E, Killer-Oberpfalzer M, Vollherbst DF, Islak C, Cekirge SH, Bendszus M, Möhlenbruch M, Koçer N. Safety and Efficacy of the FRED Jr Flow Re-Direction Endoluminal Device for Intracranial Aneurysms: Retrospective Multicenter Experience With Emphasis on Midterm Results. Front Neurol. 2021;12:722183.
Cagnazzo F, Perrini P, Dargazanli C, Lefevre PH, Gascou G, Morganti R, di Carlo D, Derraz I, Riquelme C, Bonafe A, Costalat V. Treatment of Unruptured Distal Anterior Circulation Aneurysms with Flow-Diverter Stents: A Meta-Analysis. AJNR Am J Neuroradiol. 2019;40:687–93.
Bender MT, Zarrin DA, Campos JK, Lin LM, Huang J, Caplan JM, Tamargo RJ, Colby GP, Coon AL. Tiny Pipes: 67 Cases of Flow Diversion for Aneurysms in Distal Vessels Measuring Less Than 2.0 mm. World Neurosurg. 2019;127:e193–201.
Puri AS, Massari F, Asai T, Marosfoi M, Kan P, Hou SY, Howk M, Perras M, Brooks C, Clarencon F, Gounis MJ, Wakhloo AK. Safety, efficacy, and short-term follow-up of the use of Pipeline Embolization Device in small (<2.5 mm) cerebral vessels for aneurysm treatment: single institution experience. Neuroradiology. 2016;58:267–75.
Lall RR, Crobeddu E, Lanzino G, Cloft HJ, Kallmes DF. Acute branch occlusion after Pipeline embolization of intracranial aneurysms. J Clin Neurosci. 2014;21:668–72.
Hohenstatt S, Arrichiello A, Conte G, Craparo G, Caranci F, Angileri A, Levi D, Carrafiello G, Paolucci A. Branch vessel occlusion in aneurysm treatment with flow diverter stent. Acta Biomed. 2020;91:e2020003.
Ravindran K, Casabella AM, Cebral J, Brinjikji W, Kallmes DF, Kadirvel R. Mechanism of Action and Biology of Flow Diverters in the Treatment of Intracranial Aneurysms. Neurosurgery. 2020;86(Suppl 1):S13–9.
Saleme S, Iosif C, Ponomarjova S, Mendes G, Camilleri Y, Caire F, Boncoeur MP, Mounayer C. Flow-diverting stents for intracranial bifurcation aneurysm treatment. Neurosurgery. 2014;75:623–31; quiz 631.
Dabhi N, Sarathy D, Snyder MH, Kellogg RT, Park MS. Flow Diverter Devices for Treatment of Intracranial Aneurysms in Small Parent Vessels-A Systematic Review of Literature. World Neurosurg. 2022;162:183–94.e7.
Brinjikji W, Lanzino G, Cloft HJ, Siddiqui AH, Hanel RA, Kallmes DF. Platelet Testing is Associated with Worse Clinical Outcomes for Patients Treated with the Pipeline Embolization Device. AJNR Am J Neuroradiol. 2015;36:2090–5.
Delgado Almandoz JE, Crandall BM, Scholz JM, Fease JL, Anderson RE, Kadkhodayan Y, Tubman DE. Last-recorded P2Y12 reaction units value is strongly associated with thromboembolic and hemorrhagic complications occurring up to 6 months after treatment in patients with cerebral aneurysms treated with the pipeline embolization device. AJNR Am J Neuroradiol. 2014;35:128–35.
Adeeb N, Griessenauer CJ, Foreman PM, Moore JM, Shallwani H, Motiei-Langroudi R, Alturki A, Siddiqui AH, Levy EI, Harrigan MR, Ogilvy CS, Thomas AJ. Use of Platelet Function Testing Before Pipeline Embolization Device Placement: A Multicenter Cohort Study. Stroke. 2017;48:1322–30.
Cagnazzo F, Cappucci M, Dargazanli C, Lefevre PH, Gascou G, Riquelme C, Bonafe A, Costalat V. Treatment of Distal Anterior Cerebral Artery Aneurysms with Flow-Diverter Stents: A Single-Center Experience. AJNR Am J Neuroradiol. 2018;39:1100–6.
Funding
There was no financial or material support for this study.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
S. Hohenstatt, S.L. Vinci, D.F. Vollherbst, A. Tessitore, N. Schmitt, A. Pitrone, A.A. Caragliano, M. Velo, M.A. Möhlenbruch and A. Paolucci declare that they have no competing interests.
Supplementary Information
62_2022_1187_MOESM1_ESM.docx
Table 1. Detailed aneurysm and procedure related characteristics. F female, M male, FDS flow diverter stent, ACA anterior cerebral artery, MCA middle cerebral artery, PCA posterior cerebral artery, VA vertebral artery, AcomA anterior communicating artery, SAH subarachnoid hemorrhage, IPH intraparenchymal hemorrhage
Rights and permissions
About this article
Cite this article
Hohenstatt, S., Vinci, S.L., Vollherbst, D.F. et al. Flow Diverting Stents in Cerebral Small Caliber Vessels (< 2 mm) for Aneurysm Treatment. Clin Neuroradiol 33, 99–105 (2023). https://doi.org/10.1007/s00062-022-01187-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00062-022-01187-6