Skip to main content
Log in

Single-catheter approach to pulmonary vein reisolation in selected patients

Data from a prospective registry

Neuer Ansatz zur Pulmonalvenenreisolation bei ausgewählten Patienten: Mapping und Ablation mit einem einzelnen zirkulären Katheter

Daten eines prospektiven Registers

  • Original articles
  • Published:
Herz Aims and scope Submit manuscript

Abstract

Background

In addition to different types of single-tip ablation catheters for pulmonary vein (PV) reisolation, a newly developed circular mapping and ablation catheter (nMARQ®) has been available since 2013 and is currently used only in initial PV isolation procedures. In this prospective registry we present feasibility and efficacy data for PV reisolation procedures with a single-catheter approach (nMARQ®) compared with a standard approach using a single-tip ablation catheter and a circular mapping catheter.

Methods

We included 35 carefully selected patients in this prospective registry and assigned them in a 2:1 ratio to undergo either PV reisolation with a single-tip ablation catheter together with a steerable circular mapping catheter (group 1) or with the nMARQ®catheter only (group 2). The recurrence rate was calculated for atrial tachyarrhythmias with a duration of > 30 s during a mean follow-up of 12.7 months.

Results

Reisolation of all PVs was achieved in all patients of both groups. In group 2, all gaps could be correctly identified with the nMARQ® catheter. PV isolation was clearly visible on the nMARQ® catheter in all targeted veins. With the nMARQ® catheter the ablation time decreased significantly (6.3 ± 3.0 vs. 18.6 ± 13.9 min, p < 0.05). The recurrence rate of atrial fibrillation did not differ significantly between the two groups (37.5 vs. 45.5 %, p = 0.66).

Conclusion

In selected patients, a complete PV reisolation procedure is feasible with a singular circular mapping and ablation catheter. The 12-months success rate is comparable to a classic approach with a combination of a single-tip ablation catheter and a circular mapping catheter.

Zusammenfassung

Hintergrund

Neben den verschiedenen Arten von Ablationskathetern mit einer Spitze zur Pulmonalvenenreisolation (Re-PVI) ist seit 2013 ein neu entwickelter zirkulärer Mapping- und Ablationskathether (nMARQ®) verfügbar, der derzeit nur bei initialen PVI-Eingriffen verwendet wird. In diesem prospektiven Register werden Daten zur Machbarkeit und Wirksamkeit von Re-PVI-Eingriffen mit einem Einzelkatheterverfahren (unter Verwendung des nMARQ®) mit einem Standardansatz unter Verwendung eines Ablationskatheters mit einer Spitze und eines zirkulären Mapping-Katheters verglichen.

Methoden

In das prospektive Register wurden 35 sorgfältig ausgewählte Patienten aufgenommen und im Verhältnis von 2:1 entweder der Re-PVI mit einer Kombination aus einem Ablationskatheter mit einer Spitze und einem steuerbaren zirkulären Mapping-Katheter (Gruppe 1) oder allein mit dem nMARQ®-Katheter (Gruppe 2) zugewiesen. Dann wurde die Rezidivrate für atriale Tachyarrhythmien mit einer Dauer von > 30 s während einer mittleren Nachbeobachtungszeit von 12,7 Monaten ermittelt.

Ergebnisse

Es wurde bei allen Patienten in beiden Gruppen eine Reisolation aller Pulmonalvenen (PV) erzielt. In Gruppe 2 wurden alle Öffnungen korrekt mit dem nMARQ®-Katheter identifiziert. Die PVI war mit dem nMARQ®-Katheter in allen behandelten Venen deutlich sichtbar. Dabei nahm die Ablationsdauer mit dem nMARQ®-Katheter signifikant ab (6,3 ± 3,0 vs. 18,6 ± 13,9 min; p < 0,05). Die Rezidivrate für ein Vorhofflimmern unterschied sich nicht signifikant zwischen den beiden Gruppen (37,5 vs. 45,5 %; p = 0,66).

Schlussfolgerung

Bei ausgewählten Patienten ist eine vollständige Re-PVI mit einem einzelnen zirkulären Mapping- und Ablationskatheter durchführbar. Die 12-Monats-Erfolgsrate ist vergleichbar mit der des klassischen Ansatzes, bei dem die Kombination aus einem Ablationskatheter mit einer Spitze und einem zirkulären Mapping-Katheter verwendet wird.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Mahida S, Hooks DA, Nentwich K, Ng GA, Grimaldi M, Shin DI, Derval N, Sacher F, Berte B, Yamashita S, Denis A, Hocini M, Deneke T, Haissaguerre M, Jais P (2015) nMARQ ablation for atrial fibrillation: results from a multicentre study. J Cardiovasc Electrophysiol 26:724–729

    Article  PubMed  Google Scholar 

  2. Arroja JD, Zimmermann M (2015) Phrenic nerve lesion: a potential complication of the nMARQ ablation technique. Int J Cardiol 180:91–92

    Article  PubMed  Google Scholar 

  3. Rillig A, Lin T, Burchard A, Kamioka M, Heeger C, Makimoto H, Metzner A, Wissner E, Wohlmuth P, Ouyang F, Kuck KH, Tilz RR (2015) Modified energy settings are mandatory to minimize oesophageal injury using the novel multipolar irrigated radiofrequency ablation catheter for pulmonary vein isolation. Europace 17(3):396–402

    Article  PubMed  Google Scholar 

  4. Rosso R, Halkin A, Michowitz Y, Belhassen B, Glick A, Viskin S (2014) Radiofrequency ablation of paroxysmal atrial fibrillation with the new irrigated multipolar nMARQ ablation catheter: verification of intracardiac signals with a second circular mapping catheter. Heart Rhythm 11(4):559–565

    Article  PubMed  Google Scholar 

  5. Zellerhoff S, Daly M, Lim HS, Denis A, Komatsu Y, Jesel L, Derval N, Sacher F, Cochet H, Knecht S, Yiem S, Hocini M, Haïssaguerre M, Jaïs P (2014) Pulmonary vein isolation using a circular, open irrigated mapping and ablation catheter (nMARQ): a report on feasibility and efficacy. Europace 16(9):1296–1303

    Article  PubMed  Google Scholar 

  6. Scaglione M, Caponi D, Anselmino M, Di Clemente F, Blandino A, Ferraris F, Di Donna P, Ebrille E, Halimi F, Leclercq JF, Iunco C, Vaudagna C, Cesarani F, Gaita F (2014) Pulmonary vein isolation with a new multipolar irrigated radiofrequency ablation catheter (nMARQ™): feasibility, acute and short-term efficacy, safety, and impact on postablation silent cerebral ischemia. J Cardiovasc Electrophysiol 25(12):1299–1305

    PubMed  Google Scholar 

  7. Shin DI, Kirmanoglou K, Eickholt C, Schmidt J, Clasen L, Butzbach B, Rassaf T, Merx M, Kelm M, Meyer C (2014) Initial results of using a novel irrigated multielectrode mapping and ablation catheter for pulmonary vein isolation. Heart Rhythm 11(3):375–383

    Article  PubMed  Google Scholar 

  8. Deneke T, Schade A, Müller P, Schmitt R, Christopoulos G, Krug J, Szöllösi G, Mügge A, Kerber S, Nentwich K (2014) Acute safety and efficacy of a novel multipolar irrigated radiofrequency ablation catheter for pulmonary vein isolation. J Cardiovasc Electrophysiol 25(4):339–345

    Article  PubMed  Google Scholar 

  9. Stabile G, De Ruvo E, Grimaldi M, Rovaris G, Soldati E, Anselmino M, Solimene F, Iuliano A, Sciarra L, Schillaci V, Bongiorni MG, Gaita F, Calò L (2015) Safety and efficacy of pulmonary vein isolation using a circular, open-irrigated mapping and ablation catheter: a multicenter registry. Heart Rhythm. doi: 10.1016/j.hrthm.2015.05.025. (pii: S1547-5271(15)00628-1)

    Google Scholar 

  10. Lauschke J, Schneider R, Bänsch D (2015) Mapping and ablation of a left atrial macro reentrant tachycardia originating from the left superior pulmonary vein. Herz 40(Suppl 2):137–139

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to J. Lauschke.

Ethics declarations

Conflict of interest

J. Lauschke, R. Schneider, J. Wissmann, T. Tischer, and D. Bänsch state that there are no conflicts of interest.

Additional information

J. Lauschke and R. Schneider contributed equally to this manuscript.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lauschke, J., Schneider, R., Wissmann, J. et al. Single-catheter approach to pulmonary vein reisolation in selected patients. Herz 41, 625–629 (2016). https://doi.org/10.1007/s00059-016-4402-z

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00059-016-4402-z

Keywords

Schlüsselwörter

Navigation