Keywords

1 Introduction

Ti and its alloys (along with stainless steels and Co-Cr alloys) are recognized as primary metallic biomaterials because of their excellent balance of strength and ductility, along with their high corrosion resistance when forming passivation films, low allergenic nature, and bone compatibility (osseointegration) [1, 2]. However, biofunctions of metallic biomaterials are insufficient compared to those of ceramic and polymer biomaterials. Some ceramics have chemical compositions close to that of hard tissue and exhibit bioactivity and bioresorbability, while polymers can polymerize to form functional groups and segments for biomedical applications. Therefore, surface modification is necessary to improve the biofunctionality of Ti and its alloys [3, 4]. Metallic biomaterial microstructures can generally be controlled to exhibit desired mechanical properties seen in the bulk material, an advantage they hold over ceramic and polymer biomaterials. Surface modification can in turn impart biofunctionality without impacting the bulk properties.

The different processes used for surface modification of metallic biomaterials are summarized in Fig. 2.1 [5]. These techniques are classified into two groups of wet and dry processes. Wet processes use aqueous solution for steps such as immersion and electrifying, while dry processes are based on the reactions of the surface with gases, ions, and plasma. Surface modification can control the morphology, composition, and/or phase of a given material. Controlling the surface morphology allows for high mechanical bonding between metallic biomaterials and hard tissues (mechanical anchoring) and for control of biological reactions of cells and proteins on the modified surface. Acid etching, blasting, laser treatment, additive manufacturing (AM), and porous treatments using beads, Ti plasma spraying, and fiber mesh coating are various examples of these processes.

Fig. 2.1
figure 1

Surface modification processes of metallic biomaterials

Controlling the composition/phase of a surface can be conducted through physical, chemical, and biological processes. Since the primary application of metallic biomaterials is for hard-tissue substitution, bioceramics with excellent bone compatibility can be utilized as agents for surface composition/phase control. Processes such as thermal spraying, vapor deposition, electrochemical treatment, and immersion in simulated body fluid (SBF) have been used to coat apatite, which is an analogue for the inorganic components of hard tissues. In addition, TiO2, calcium phosphate, sodium titanate, and CaTiO3 coatings formed by anodic oxidation, gas treatment, NaOH treatment and heating, ion implantation, and chemical conversion can improve bone compatibility of implant substrates under biological conditions. Anodic oxidation changes both the chemical composition and roughness of a surface. Hybrid coatings of ceramics and polymers are candidates for further improvements in the biofunctionality of metallic biomaterials. To this end, hybrids such as calcium phosphate+collagen and calcium phosphate+bone morphogenetic protein have been seen in recent study [6–8].

In this chapter, the formation and evaluation of ceramic coatings of Ti and its alloys are described with a focus on their antibacterial potential. Radio-frequency (RF) magnetron sputtering and thermal oxidation are the main processes discussed here for controlling composition/phase of surfaces, which are classified as dry processes. Surgical site infection (SSI) related to implants occurs at rates of 2–30 % depending on the surgical site and the procedure, and SSI is even more frequent after revision surgery [9, 10]. Several approaches are known to lower the risk of SSI, such as the use of antibiotics. However, prolonged use of antibiotics at higher doses can lead to systematic drug resistance and local toxicity [11]. The use of ceramic coatings with antibacterial activity is a powerful method to mitigate SSI.

2 Antibacterial Properties of Ag-Containing Amorphous Calcium Phosphate Coating Films

Ag, Cu, and Zn ions are known antibacterial agents. Ag ions are effective against many types of bacteria, are less likely to cause bacterial resistance, and express antibacterial activity at lower concentrations than Cu and Zn ions [12–14]. Here, amorphous calcium phosphate (ACP) coating films on Ti and its alloys are fabricated by RF magnetron sputtering and evaluated in vivo and in vitro [15–17]. Coating of implants with Ag-containing ACP is one possible technique for preventing infection; ACP possesses bioresorbability under biological conditions, which causes the continuous release of Ag ions from such Ag-containing films, creating antibacterial activity for a desired duration. The addition of Ag to HAp (hydroxyapatite, Ca10(PO4)6(OH)2) has been studied previously [18–21], but HAp is less resorbable than ACP. Both antibacterial activity and improved bone forming are expected from implants coated with Ag-containing ACP.

Ag-containing β-TCP (β-type tricalcium phosphate, Ca3P2O8) sintered body was fabricated by hot-pressing Ag and β-TCP raw powders as targets in RF magnetron sputtering. The Ag content in Ag-containing β-TCP targets was 0, 15, and 30 mass% as listed in Table 2.1. Plates (10 × 10 × 1 mm) of mirror-polished commercially pure (CP) Ti (grade 2) and blasted Ti-6mass%-Al-4mass%V (Ti-6Al-4V) alloy were used as substrates. The thickness of the coating films for analysis and evaluation of antibacterial properties was fixed at 0.5 μm, as controlled by adjusting the deposition time.

Table 2.1 Notation and composition of sputtering targets (mass%)

Figure 2.2 shows X-ray diffraction (XRD) patterns of the coating films formed on mirror-polished CP Ti substrates using a 15mass%-Ag-containing TCP (15AgTCP) target. A broad peak is present at a 2θ value of 25–35°, which is a characteristic of ACP coating films [15]. Cross-sectional scanning electron microscopy (SEM) images of coating films on mirror-polished CP Ti substrates are present in Fig. 2.3. The coating films are dense and uniform, with good adhesion to the substrates. The deposition rates were 0.02–0.1 nm⋅s−1, depending on the target composition and RF power.

Fig. 2.2
figure 2

XRD patterns of coating films fabricated on mirror-polished CP Ti substrates at RF powers of 50, 100, and 150 W using a 15AgTCP target

Fig. 2.3
figure 3

Cross-sectional SEM image of coating films fabricated on mirror-polished CP Ti substrates using (a) 0AgTCP, (b) 15AgTCP, and (c) 30AgTCP targets

The Ag contents of the coating films formed using 15AgTCP and 30AgTCP targets were 2 and 15 mass%, respectively, lower than those of the targets. Lower Ag contents compared to target compositions have been previously reported in Ag-containing HAp coating films [22]. These discrepancies may be linked to differences in the ionization rates between elements during sputtering.

E. coli and S. aureus, Gram-negative and Gram-positive bacteria, respectively, were used for antibacterial testing. Ag-containing ACP coating films on blasted Ti-6Al-4V substrates were shaking cultured at 200 rpm in 1/500 nutrient broth (NB) solution with an initial bacterial concentration of 1 × 107 CFU (colony-forming unit)·mL−1 for E. coli and 1 × 105 CFU·mL−1 for S. aureus. Solution temperature was maintained at 310 K and incubation was carried out for 10.8 and 86.4 ks. Afterward, the numbers of viable bacteria colonies were counted using a smear-plate culture method.

Figure 2.4 shows the relationship between the incubation time and the number of viable bacteria for coating films formed on blasted Ti-6Al-4V substrates using a 15AgTCP target. After incubation for 10.8 ks, the number of viable E. coli was less than 1, which was plotted as 100 CFU. The number of viable S. aureus decreased after incubation for 10.8 ks and became less than 1 after incubation for 86.4 ks.

Fig. 2.4
figure 4

Number of viable (a) E. coli and (b) S. aureus after incubation with Ag-containing ACP coating films fabricated on blasted Ti-6Al-4V substrates using a 15AgTCP target

The amounts of Ca, P-related, and Ag ions after immersion of Ag-containing ACP coating films in the 1/500 NB solution with shaking at 200 rpm are depicted in Fig. 2.5. Ag ions elute from the Ag-containing ACP coating films and exhibit antibacterial activity. The Ag ion concentration was almost constant over all measured immersion times, caused by the formation of AgCl between the eluted Ag+ and Cl− present in the 1/500 NB solution.

Fig. 2.5
figure 5

Amounts of Ca, P-related, and Ag ions in 1/500 NB solution after immersion of Ag-containing ACP coating films fabricated on blasted Ti-6Al-4V using a 15AgTCP target

3 Photocatalytic Activity of TiO2 Layers Formed by Two-Step Thermal Oxidation

3.1 UV (Ultraviolet) Response

TiO2 layers have been used as surface coatings of Ti implants [23], as TiO2 exhibits photocatalytic activity under UV-light irradiation, such as photoinduced superhydrophilicity and photodegradation of organic compounds [24]. The photoinduced superhydrophilicity of TiO2 layers on Ti improves bone compatibility [25, 26], and photodegradation contributes to the antibacterial activity of Ti implants by killing nearby bacteria through radical formation [27, 28]. Rutile is a thermodynamically stable phase of TiO2 with a band gap of 3.0 eV, while anatase is a metastable phase of TiO2 having a band gap of 3.2 eV. The lifetimes of these electronic excitations are an order of magnitude larger for anatase as compared to rutile [29], and as such the higher photocatalytic activity is expected in anatase.

The present authors presented a two-step thermal oxidation process for the fabrication of anatase-rich TiO2 layer on Ti and its alloys [30–32], consisting of initial treatment under CO-containing atmosphere and subsequent treatment in air, and evaluated the photocatalytic activity of anatase-containing TiO2 layers on CP Ti and Ti-25mass%Mo and Ti-25mass%Nb alloys [32]. The anatase fraction in the TiO2 layer can be controlled during this process by controlling the temperature of the second step. By using TiO2 layers with various anatase fractions, the degradation rate of methylene blue (MB) under UV-light irradiation was measured. Illumination was performed using a UV lamp with a central wavelength of 351 nm. The UV-light intensity was 1.0 mW⋅cm−2 at the surface of the specimen. The effects of anatase fraction on the MB degradation rates are summarized in Fig. 2.6 [32], where the maximum rate was obtained at an anatase fraction of approximately 0.8 for both CP Ti and its alloys. Su et al. [33] measured the MB degradation rates of TiO2 films on CP Ti formed by anodic oxidation, where the grain sizes, surface areas, and crystallinity of the TiO2 films were fixed, but the anatase fraction was varied. They reported a maximum decomposition rate at an anatase fraction of approximately 0.6 and indicated that electrons that are photoexcited in rutile can migrate to the conduction band of anatase to leave electron holes in the rutile; as a result, recombination is effectively suppressed in anatase-rich TiO2 layers. In our case, the second step temperature was altered to control the resulting anatase fraction, but the thickness and surface roughness were not controlled. Nevertheless, a similar dependence of decomposition rate upon the anatase fraction was obtained compared to Su et al.; these anatase-rich TiO2 layers show excellent photodegradation of organic compounds.

Fig. 2.6
figure 6

Effects of the anatase fraction in the TiO2 layers on the rate constant of methylene blue degradation under UV-light irradiation for CP Ti and the Ti-25mass%Mo and Ti-25mass%Nb alloys [Reprinted from Ref. [32] with permission from Elsevier]

3.2 Visible-Light Response

The visible-light-responsive photocatalytic activity of implants is useful for antibacterial activity during operation and in the reactivation of implants. Because SSI is caused by endogenous flora in the patient’s skin, mucous membranes, or hollow viscera, SSI cannot simply be prevented by sterilization before operation [34]. In order to express the photocatalytic activity of TiO2, UV-light irradiation is generally required, which is harmful to the human body. Therefore, visible-light response is more desirable for preventing infection during operation. Reactivation of implants is required during usage for some periods in a human body. While UV-light irradiation is known to reactivate Ti implant surfaces [35], reactivation using visible-light irradiation is more favorable for the human body.

Au solutes in TiO2 [36] and the addition of metallic Au nanoparticles to TiO2 [36, 37] can lead to visible-light response. Therefore, two-step thermal oxidation was applied to a Ti-Au alloy in order to fabricate Au-containing TiO2 layers with visible-light-responsive photocatalytic activity. Ti-4.2at%Au alloy was employed as a substrate material. The first step treatment was conducted under an Ar-1%CO atmosphere at 1073 K for 3.6 ks, and the subsequent second step treatment was conducted in air at 673 K for 10.8 ks. An anatase-rich TiO2 layer was obtained on the Ti-4.2at%Au alloy substrate after this process. Figure 2.7 depicts the cross-sectional microstructure of the anatase-rich TiO2 layer as observed via transmission electron microscopy (TEM). Au nanoparticles can be seen here as small black particles approximately 5 nm in diameter.

Fig. 2.7
figure 7

Cross-sectional TEM image of Ti-4.2at%Au alloy after two-step thermal oxidation

The photocatalytic activity of the Au-containing TiO2 layer was evaluated by self-cleaning tests (JIS R 1753: 2013) based on the decomposition of a layer of steric acid on the Au-containing TiO2 layer under visible-light irradiation, where the steric acid layer was prepared by dip coating. Substrates with this steric acid layer were irradiated with visible light with an intensity of 10 mW·cm−2 at the surface of the steric acid layer, for as long as 86.4 ks. The light source was a Xe lamp with a UV filter to produce light of λ >400 nm. Changes in the water contact angles under visible-light irradiation were measured to determine the extent of steric acid photodegradation, and the TiO2 layer expressed visible-light-responsive photocatalytic activity.

Figure 2.8 shows the changes in water contact angles on as-polished and two-step (i.e., after two-step thermal oxidation) CP Ti and Ti-4.2at%Au substrates under visible-light irradiation [38]. The water contact angles on the as-polished specimens remained almost constant, indicating no visible-light response. On the other hand, the two-step specimens showed decreased water contact angles. For the two-step Ti-4.2at%Au alloy, the water contact angle decreased from 80 to 10°. Silva et al. [37] reported that electrons in Au nanoparticles (1.87–6.40 nm in diameter) attached on TiO2 particles were excited by surface plasmon resonance and migrated to TiO2 particles, producing H2 by the photoreduction of H2O. The Au particle size recommended in their study is close to that of our study (Fig. 2.7). One possible mechanism of visible-light response in the Au-containing TiO2 layer formed by two-step thermal oxidation involves the decrease in band gap due to Au solutes in the TiO2 layer [36] and/or the surface plasmon resonance of the Au nanoparticles in the TiO2 layer [37].

Fig. 2.8
figure 8

Measured water contact angles on the as-polished and two-step CP Ti and Ti-4.2at%Au alloy substrates during visible-light irradiation

4 Summary

The surface modification of Ti and its alloys using ceramic coatings and their antibacterial activity were described in this chapter. Both Ag-containing ACP coating films and Au-containing TiO2 layers, fabricated by RF magnetron sputtering and two-step thermal oxidation, respectively, indicated antibacterial activity. Control over Ag ion elution from Ag-containing ACP coating films and the improved visible-light response of Au-containing TiO2 layers will be discussed in future publications.

Ti and its alloys are expected to see continued use as metallic biomaterials for implants. Figure 2.9 shows factors that must be considered when developing ceramic coatings on Ti and its alloy substrates for biomedical applications [5]. The interface between a coating film and the human body requires designed properties such as hard/soft tissue and blood compatibilities, antibacterial activity, wear and corrosion resistance, resorbability, and suppression/enhancement of protein and cell attachment. These properties are affected by surface morphology, phase, composition, and orientation of the coating films. High-strength and durable bonding between a coating film and substrates, which are affected by physical and chemical properties of the substrates and the morphology at the interface, are also crucial. Therefore, the mechanical properties and thickness of coating films, as well as the reaction between coating film and substrates, must be taken into consideration.

Fig. 2.9
figure 9

Factors in biomedical ceramic coatings on Ti and its alloy substrates