Abstract.
At the IRIS facility a rather precise method of the target-ion source unit ionization efficiency measurement has been developed. The method exploits an off-line mass-separator for the implantation of the ion beams of selected stable isotopes into a tantalum foil placed inside a Faraday cup in the focal plane of the mass-separator. After the implantation of the required amount of the investigated species, tantalum foil has been inserted into the volume of the target-ion source unit prepared for the on-line utilization at the IRIS on-line separator. The first tests have ensured the ionization efficiency values (90±10)% for Rb and (85±10)% for Cs in the empty combined target-ion source unit, which was used as a reference one. For the target-ion source unit with UC target material inside the measured value of the ionization efficiency of Rb was (52±20)%.
Similar content being viewed by others
References
V.N. Panteleev, A.E. Barzakh, D.V. Fedorov, A.M. Ionan, K.A. Mezilev, F.V. Moroz, S.Yu. Orlov, Yu.M. Volkov, A. Andrighetto, G. Lhersonneau, V. Rizzi, L.B. Tecchio, M. Dubois, G. Gaubert, P. Jardin, N. Lecesne, R. Leroy, J.Y. Pacquet, M.G. Saint Laurent, A.C.C. Villari, O. Bajeat, S. Essabaa, C. Lau, M. Menna et al., Rev. Sci. Instrum. 77, 03A705 (2006)
G.D. Alkhasov, Yu.S. Blinnikov, H. Huhnermann, S.Yu. Orlov, V.N. Panteleev, A.G. Polyakov, D.M. Seliverstov, Yu.M. Volkov, H. Wagner, PNPI Research Report 1994–1995 (Russian Academy of Sciences, Petersburg Nuclear Physics Institute, Gatchina, 1996), p. 83
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Panteleev, V., Barzakh, A., Fedorov, D. et al. Target-ion source unit ionization efficiency measurement by a method of stable ion beam implantation. Eur. Phys. J. Spec. Top. 150, 301–302 (2007). https://doi.org/10.1140/epjst/e2007-00329-x
Issue Date:
DOI: https://doi.org/10.1140/epjst/e2007-00329-x