Abstract
This study was designed to assess the accuracy of end-tidalPco 2 and transcutaneousPco 2 as measurements of arterialPco 2 in extubated, spontaneously breathing patients recovering from general anesthesia. In 30 patients, measurement of arterial transcutaneous, and end-tidalPco 2 were taken simultaneously with body temperature approximately every 15 minutes over a 2-hour period. ArterialPco 2 values were corrected for body temperature. Values for Paco 2 were compared with those forPetCO2 and Psco 2 by linear regression analysis and by calculation of bias ± precision. Thirty-six percent of the capnogram tracings obtained did not develop a plateau phase. We found poor correlation between end-tidal and arterialPco 2 regardless of the shape of the capnogram tracing, as well as poor correlation between transcutaneous and arterialPco 2. Although the measurements of bias and precision of noninvasivePco 2 monitors in this population are comparable to studies in other populations, we advise caution in relying on the routine use ofPetCO2 or Psco 2 for the noninvasive assessment of respiratory depression in extubated, spontaneously breathing patients recovering from general anesthesia.
Similar content being viewed by others
References
Knill RL, Gelb AW. Ventilatory response to hypoxia and hypercapnia during halothane sedation and anesthesia in man. Anesthesiology 1978;49:244–251
Nunn JF. Applied respiratory physiology. London: Butterworth, 1987:357–366
Monaco F, McQuitty JC, Nickerson BG. Calibration of a heated transcutaneous carbon dioxide electrode to reflect arterial carbon dioxide. Am Rev Respir Dis 1983;127:322–324
Bergman NA. Temperature coefficients for PCO2 and PO2 in bloods with varying acid-base status.J Appl Physiol 1968;24:225–228
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–310
Tremper KK, Shoemaker WC, Shippy CR, et al. Transcutaneous PCO2 monitoring on adult patients in the ICU and the operating room. Crit Care Med 1981;9:752–755
Brambilla I, Micallef E, Sacerdoti C, et al. Value of nocturnal monitoring of transcutaneous O2 and CO2 pressures in adults with respiratory failure. Respiration 1985;48:81–90
Rafferty TD, Marrero O, Nardi D, et al. Relationship between transcutaneous and arterial carbon dioxide tension in adult patients anesthetized with nitrous oxidefentanyl and nitrous oxide-enflurane. Anesth Analg 1981;60:504–507
Brunner JX, Westenskow DR. How the rise time of carbon dioxide analyzers influences the accuracy of carbon dioxide measurements. Br J Anaesth 1988;61:628–638
Shankar KB, Moseley H, Kumar Y. Negative arterial to end-tidal gradients. Can J Anaesth 1991;38:260–261
Phan CQ, Tremper KK, Lee SE, et al. Non-invasive monitoring of carbon dioxide: A comparison of the partial pressure of transcutaneous and end-tidal carbon dioxide with the partial pressure of arterial carbon dioxide. J Clin Monit 1987;3:149–154
Fletcher R, Jonson B, Cumming G, et al. The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth 1981;53:77–87
Hoffman RA, Krieger BP, Kramer MR, et al. End-tidal carbon dioxide in critically ill patients during changes in mechanical ventilation. Am Rev Respir Dis 1989;140:1265–1268
Jones NL, Robertson DG, Kane JW. Differences between end-tidal and arterial CO2 in exercise. J Appl Physiol 1979;47:954–960
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Kavanagh, B.P., Sandler, A.N., Turner, K.E. et al. Use of end-tidalPco 2 and transcutaneousPco 2 as noninvasive measurement of arterialPco 2 in extubated patients recovering from general anesthesia. J Clin Monitor Comput 8, 226–230 (1992). https://doi.org/10.1007/BF01616780
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01616780