Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 °C to + 20 ° C

  • Published:
International Journal of Thermophysics Aims and scope Submit manuscript

Abstract

The first European humidity key comparison EURAMET-T.K6 was completed in 2008, and it covered the dew-point temperature range from −50 °C to + 20 °C. Both LPM and MIKES participated in the comparison, but a new low dew-point generator was introduced at LPM as a result of progress in the EUROMET P912 project. To extend the range of available comparison evidence down to −70 °C and to study the validity of improved uncertainties of LPM, a bilateral comparison was carried out between LPM and MIKES in 2009–2010. The applied comparison procedure was similar to that applied in EURAMET-T.K6. However, only one transfer standard was used instead of two units and the measurement point −70 °C was added in the measurement scheme. The results show that the bilateral equivalence between LPM and MIKES is between (0.00 ± 0.06) °C and (0.02 ± 0.08) °C in the range from −50 °C to + 20 °C and (0.01 ± 0.10) °C at −70 °C. Using MIKES results as the link to the EURAMET.T-K6, it is shown that the difference between the results obtained with the new LPM dew-point temperature standard and the EURAMET Comparison Reference Values is between (−0.02 ± 0.08) °C at 20 °C and (+ 0.02 ± 0.07) ° C at −50 °C.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. M. Heinonen, M. Anagnostou, S. Bell, M. Stevens, R. Benyon, R.A. Bergerud, J. Bojkovski, R. Bosma, J. Nielsen, N. Böse, P. Cromwell, A. Kartal Dogan, S. Aytekin, A. Uytun, V. Fernicola, K. Flakiewicz, B. Blanquart, D. Hudoklin, P. Jacobson, A. Kentved, I. Lóio, G. Mamontov, A. Masarykova, H. Mitter, R. Mnguni, J. Otych, A. Steiner, N. Szilágyi Zsófia, D. Zvizdic, Int. J. Thermophys. doi:10.1007/s10765-011-0950-x

  2. D. Zvizdic, M. Heinonen, D. Sestan, submitted to Int. J. Thermophys.

  3. M. Heinonen, in Proceedings of TEMPMEKO 2001, 8th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by B. Fellmuth, J. Seidel, G. Scholz (VDE Verlag, Berlin, 2002), pp. 217–222

  4. M. Heinonen, L. Uusipaikka, in Proceedings of TEMPMEKO 2004, 9th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by D. Zvizdić, L.G. Bermanec, T. Veliki, T. Stašić (FSB/LPM, Zagreb, 2004), pp. 669–675

  5. M. Heinonen, in Proceedings of 5th International Symposium on Humidity and Moisture—ISHM 2006 (CD-ROM). ISBN: 978-85-86920-04-2 (INMETRO, 2007)

  6. Heinonen M., Vilbaste M.: Int. J. Thermophys. 29, 1589 (2008)

    Article  ADS  Google Scholar 

  7. IEC 60751 (1995-07) (ISO, Geneva, 1995)

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Martti Heinonen.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Heinonen, M., Zvizdic, D. & Sestan, D. Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 °C to + 20 ° C. Int J Thermophys 33, 1451–1457 (2012). https://doi.org/10.1007/s10765-011-1105-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10765-011-1105-9

Keywords

Navigation