Abstract
Object
In this study, the feasibility of in vivo proton magnetic resonance spectroscopic imaging (1H MRSI) of the healthy human brain at a field strength of 9.4 T, using conventional acquisition techniques, is examined and the initial experience is summarized.
Materials and methods
MRSI measurements were performed on a 9.4 T MR scanner (Siemens, Erlangen, Germany) equipped with head-only gradient insert (AC84, Siemens) and custom-developed, 8-channel transmit/24-channel receive, and 16-channel transmit/31-channel receive coils. Spectra were acquired from the superior part of the human brain with a modified STEAM sequence. Spectral quantification was done with LCModel software.
Results
Reasonable quality and signal-to-noise ratio of the acquired spectra allowed reliable quantification of 12 metabolites (Cramer-Rao lower bounds < 20 %), some of which may be difficult to quantify at field strengths below 7 T due to overlapping resonances or low concentrations.
Conclusion
While further developments are necessary to minimize chemical shift displacement and homogeneity of the transmit field, it is demonstrated that in vivo 1H MRSI at a field strength of 9.4 T is possible. However, further studies applying up-to-date techniques to overcome high-field specific problems are needed in order to assess the potential gain in sensitivity that may be offered by MRSI at 9.4 T.
Similar content being viewed by others
References
Pohmann R, Shajan G, Balla DZ (2011) Contrast at ultra-high field: relaxation times, magnetization transfer and phase in the rat brain at 16.4 T. Magn Reson Med 66:1572–1581
Tkac I, Gruetter R (2005) Methodology of 1H NMR spectroscopy of the human brain at very high magnetic fields. Appl Magn Reson 29(1):139–157
Marjanska M, Auerbach EJ, Valabregue R, de Moortele Van, Adriany G, Garwood M (2012) Localized 1H NMR spectroscopy in different regions of human brain in vivo at 7 T: T2 relaxation times and concentrations of cerebral metabolites. NMR Biomed 25:332–339
Henning A, Fuchs A, Murdoch JB, Boesinger P (2009) Slice-selective FID acquisition, localized by outer volume suppression (FIDLOVS) for 1H-MRSI of the human brain at 7 T with minimal signal loss. NMR Biomed 22:683–696
Tkac I, Oez G, Adriany G, Ugurbil K, Gruetter R (2009) In-vivo 1H NMR spectroscopy of the human brain at high magnetic fields: metabolite quantification at 4T vs. 7T. Magn Reson Med 62:868–879
Mekle R, Mlynarik V, Gambarota G, Hergt M, Krueger G, Gruetter R (2009) MR spectroscopy of the human brain with enhanced signal intensity using ultrashort echo times on a clinical platform at 3T and 7T. Magn Reson Med 61:1279–1285
Yang S, Hu J, Kou Z, Yang Y (2008) Spectral simplification for resolved glutamate and glutamine measurement using a standard STEAM sequence with optimized timing parameters at 3, 4, 4.7, 7 and 9.4 T. Magn Reson Med 59:236–244
Choi C, Dimitrov I, Douglas D, Zhao C, Hawesa H, Ghose S, Tamminga CA (2009) In vivo detection of serine in the human brain by proton magnetic resonance spectroscopy (1H MRS) at 7 T. Magn Reson Med 62:1042–1046
Banerjee A, Ganji S, Hulsey K, Dimitrov I, Maher E, Ghose S, Tamminga C, Choi C (2012) Measurement of glycine in gray and white matter in the human brain in vivo by 1H MRS at 7 T. Magn Reson Med 68:325–331
Pan JW, Avdievich N, Hetherington HP (2010) J-refocused coherence transfer spectroscopic imaging at 7 T in human brain. Magn Reson Med 64:1237–1246
Deelchand DK, Van de Moortele PF, Adriany G, Iltis I, Andersen P, Strupp JP, Vaughan T, Ugurbil K, Henry PG (2010) In-vivo 1H NMR spectroscopy of the human brain at 9.4 T: initial results. J Magn Reson 206:74–80
Tkac I, Andersen P, Adriany G, Merkle H, Ugurbil K, Gruetter R (2001) In vivo 1H NMR spectroscopy of the human brain at 7T. Magn Reson Med 46:451–456
Boer VO, Siero JCW, Hoogduin H, van Gorp JS, Luijten PR, Klomp DWJ (2011) High-field MRS of the human brain at short TE and TR. NMR Biomed 24:1081–1088
Bogner W, Gruber S, Trattnig S, Chmielnik M (2012) High-resolution mapping of human brain metabolites by free induction decay 1H MRSI at 7T. NMR Biomed 25:873–882
Shajan G, Kozlov M, Hoffman J, Turner R, Scheffler K, Pohmann R (2013) A 16-channel dual-row transmit array in combination with a 31-element receive array for human brain imaging at 9.4 T. Magn Reson Med 71:870–879
Ogg RJ, Kingsley RB, Taylor JS (1994) WET, a T1 and B1 insensitive water suppression method for in vivo localized 1H NMR spectroscopy. J Magn Reson B 104(1):1–10
Yarnykh VL (2007) Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 57:192–200
Pohmann R, Scheffler K (2013) A theoretical and experimental comparison of different techniques for B1 mapping at very high fields. NMR Biomed 26:265–275
De Graaf RA (2007) Radiofrequency Pulses. In: De Graaf RA (ed) In vivo NMR spectroscopy, 2nd edn. Wiley, Chichester, pp 233–293
Provencher SW (1993) Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med 30:672–679
Govindaraju V, Young K, Maudsley AA (2000) Proton NMR chemical shifts and coupling constants for brain metabolites. NMR Biomed 13:129–153
Tkac I (2008) Refinement of simulated basis set for LCModel analysis. In: Proceedings of the 16th scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto, p 1624
Ernst T, Kreis R, Ross BD (1993) Absolute quantitation of water and metabolites in the human brain. I. Compartments and water. J Magn Reson Ser B 102:1–8
Hong ST, Balla DZ, Shajan G, Choi C, Ugurbil K, Pohmann R (2011) Enhanced neurochemical profile of the rat brain using in vivo 1H NMR spectroscopy at 16.4 T. Magn Reson Med 65:28–34
Terpstra M, Ugurbil K, Gruetter R (2002) Direct in vivo measurement of human cerebral GABA concentration using MEGA-editing at 7T. Magn Reson Med 47:1009–1012
Andreychenko A, Boer VO, de Castro CSA, Luijten PR, Klomp DWJ (2012) Efficient spectral editing at 7T: GABA detection with MEGA-sLASER. Magn Reson Med 68:1018–1025
Helms G (2000) A precise and user-independent quantification technique for regional comparison of single volume proton MR spectroscopy of the human brain. NMR Biomed 13:398–406
Wang Y, Li SJ (1998) Differentiation of metabolic concentrations between gray matter and white matter of human brain by in vivo 1H magnetic resonance spectroscopy. Magn Reson Med 39:28–33
Pan JW, Lo KM, Hetherington HP (2012) Role of high order and degree B0 shimming for spectroscopic imaging of the human brain at 7 Tesla. Magn Reson Med 68:1007–1017
Hetherington HP, Chu WJ, Gonen O, Pan JW (2006) Robust fully automated shimming of the human brain for high-field 1H spectroscopic imaging. Magn Reson Med 56:26–33
Shah S, Kellerman P, Greiser A, Weale PJ, Zuehlsdorff S, Jerecic R (2009) Rapid fieldmap estimation for cardiac shimming. In: Proceedings of the 17th scientific meeting, International Society for Magnetic Resonance in Medicine, Honolulu, Hawaii, USA, p 566
Gruetter R (1993) Automatic, localized in vivo adjustment of all first- and second-order shim coils. Magn Reson Med 29:804–811
Gruetter R, Tkac I (2000) Field mapping without reference scan using asymmetric echo-planar techniques. Magn Reson Med 43:319–323
Zhong X, Lyubich YM, De Vito T, Shah S, Knight-Scott J (2012) Improving in vivo 1H-MRS with robust automated shim techniques: a comparison study of FASTESTMAP and GRESHIM. In: Proceedings of the 20th scientific meeting, International Society for Magnetic Resonance in Medicine, Melbourne, Victoria, Australia, p 4397
Zhong X, Lyubich YM, De Vito T, Shah S, Knight-Scott J (2013) Quantitative comparison of shim algorithms for in vivo 1H-MRS. In: Proceedings of the 21th scientific meeting, International Society for Magnetic Resonance in Medicine, Salt Lake City, Utah, USA, p 3988
Avdievich NI, Pan JW, Beahring JM, Spencer DD, Hetherington HP (2009) Short echo spectroscopic imaging of the human brain at 7 T using transceiver arrays. Magn Reson Med 62:17–25
Andronesi OC, Gagoski BA, Sorensen AG (2012) Neurologic 3D MR spectroscopic imaging with low-power adiabatic pulses and fast spiral acquisition. Radiology 262(2):647–661
Hetherington HP, Avdievich NI, Kuznetsov AM, Pan JW (2010) RF shimming for spectroscopic localization in the human brain at 7 T. Magn Reson Med 63:9–19
Boer VO, van Lier ALHMW, Hoogduin JM, Wijnen JP, Luijten PR, Klomp DWJ (2011) 7-T 1H MRS with adiabatic refocusing at short TE using radiofrequency focusing with a dual-channel volume transmit coil. NMR Biomed 24:1038–1046
Emir UE, Auerbach EJ, Van de Moortele PF, Marjanska M, Ugurbil K, Terpstra M, Tkac I, Oez G (2012) Regional neurochemical profiles in the human brain measured by 1H MRS at 7 T using local B1 shimming. NMR Biomed 25:152–160
Boer VO, Klomp DWJ, Juchem C, Luijten PR, de Graaf RA (2012) Multislice 1H MRSI of the human brain at 7 T using dynamic B0 and B1 shimming. Magn Reson Med 68:662–670
Zhu H, Soher B, Ouwerkerk R, Schaer M, Barker PB (2013) Spin-echo magnetic resonance spectroscopic imaging at 7T with frequency-modulated refocusing pulses. Magn Reson Med 69:1217–1225
Tkac I, Starcuk Z, Choi IY, Gruetter R (1999) In vivo 1H NMR spectroscopy of rat brain at 1 ms echo time. Magn Reson Med 41:649–656
Acknowledgments
This work was founded (in part) by the Helmholtz Alliance ICEMED—Imaging and Curing Environmental Metabolic Diseases, through the Initiative and Networking Found of the Helmholtz Association.
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical standard
All volunteer studies have been approved by the local ethics committee and have therefore been performed in accordance with the ethical standards laid down in the 1964 Declaration of Helsinki and its later amendments.
Informed consent
All in vivo measurements described in this study were conducted exclusively on healthy volunteers. All volunteers gave their informed consent prior to their inclusion in the study.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Chadzynski, G.L., Pohmann, R., Shajan, G. et al. In vivo proton magnetic resonance spectroscopic imaging of the healthy human brain at 9.4 T: initial experience. Magn Reson Mater Phy 28, 239–249 (2015). https://doi.org/10.1007/s10334-014-0460-5
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10334-014-0460-5